CALIBRATION PROCEDURE Date: 01-03-2019
ASF-QC-CAL-001 Page: 1 of 7
CALIBRATION PROCEDURE
ASF-QC-CAL-001
CALIBRATION PROCEDURE
ASF-QC-CAL-001
QUALITY CONTROL DEPARTMENT REV. 1
CALIBRATION PROCEDURE Date: 01-03-2019
ASF-QC-CAL-001 Page: 2 of 7
LIST OR REVISIONs
Rev. Date Comment Prepared By Reviewed By Approved By
0 01-01-2019 For Implementation Salem Abdel Naby Ahmed Safwat
1 01-03-2019 Change in the procedure number Salem Abdel Naby Ahmed Safwat
CALIBRATION PROCEDURE
ASF-QC-CAL-001
QUALITY CONTROL DEPARTMENT REV. 1
CALIBRATION PROCEDURE Date: 01-03-2019
ASF-QC-CAL-001 Page: 3 of 7
INDEX
No. DESCRIPTION PAGE
1 OBJECTIVES 4
2 SCOPE 4
3 REFERENCE 4
4 DEFINITIONS 4
5 RESPONSIBILITIES 4
6 PROCEDURE 4
7 RECORDS 7
8 ATTACHMENTS 7
CALIBRATION PROCEDURE
ASF-QC-CAL-001
QUALITY CONTROL DEPARTMENT REV. 1
CALIBRATION PROCEDURE Date: 01-03-2019
ASF-QC-CAL-001 Page: 4 of 7
1. OBJECTIVES
The purpose of this document is to describe the necessary information required to effectively manage the working tools
calibrated and maintained properly and fits for use
2. SCOPE
This procedure establishes system for planning and performing calibrations of measuring tools and equipment either
internally or externally calibrated according to National and International standards and/or ASME requirements as
appropriate.
3. REFERENCE
3.1. ISO 9001:2015
3.2. AWS D1.1 – Steel Structural Welding Code
3.3. EN 1090 - Technical Requirements of the execution of the steel structure
3.4. ASME – ASME Boilers and Pressure Vessel Codes
3.5. API – American Petroleum Institute
3.6. ASF-QC-CAL-001 – this procedure
4. DEFINITION
QCM: Quality Control Manager
RCS: Recognized Calibration Society/Institute
QC: Quality control Engineer which is the responsible person to carry out the calibration activity
NS: National/International Standard
RT: Radiographic testing
MT: Magnetic Testing
5. RESPONSIBILITIES
QC engineer is the responsible person to carry out the calibration activity including witnessing the testing, carrying out
the testing or preparing the reports to record all the data assuring that the equipment running in a proper way.
6. PROCEDURE
6.1. REFERENCE TOOLS AND EQUIPMENT
6.1.1. Reference equipment/tools that will be used for internal calibration and other equipment/tools that shall be
calibrated externally in an RCS against relevant national or international standards are listed in table 1
6.1.2. In case of conflict between the tolerances specified in table 1, and the standards used by the RCS, the most
restricted one shall prevail.
6.1.3. Reference equipment/tools calibration intervals shall follow the frequencies specified in table 1, and should be
performed in case of damage or deterioration raised against the required accuracy level.
6.1.4. Calibration certificate of reference equipment/tools shall be issued by the RCS and shall be maintained and
approved by the QCM this certificate shall at least specifies the following details:
- The calibration dates
- The serial no. of the equipment/tool under calibration
- A complete description of the calibration equipment/tool used for calibration (including related
description, serial no., manufacturer, and calibration due date).
- Reference calibration standard
- Calibration readings
- The recorded accuracy of equipment/tool under calibration.
6.1.5. Calibrated Reference equipment/tools shall only be used for internal calibration purposes (not as daily working
tools) and shall be kept by the QCM in suitable containers to protect them from being damaged.
CALIBRATION PROCEDURE
ASF-QC-CAL-001
QUALITY CONTROL DEPARTMENT REV. 1
CALIBRATION PROCEDURE Date: 01-03-2019
ASF-QC-CAL-001 Page: 5 of 7
6.1.6. Calibrated Reference equipment/tools shall be listed in the Equipment Calibration Log form no. ASF-QC-CAL-
002
6.1.7. All devices shall be calibrated over the full range of use
6.1.8. At least all the devices mentioned in the Spanish regulation issued by ministry of industry, energy and tourism
regarding calibration of devices shall be calibrated or verified as per table 2
CALIBRATION PROCEDURE
ASF-QC-CAL-001
QUALITY CONTROL DEPARTMENT REV. 1
CALIBRATION PROCEDURE Date: 01-03-2019
ASF-QC-CAL-001 Page: 6 of 7
6.4.4. The Densitometer is acceptable if the density readings do not vary by more than +0.05 density units from the
actual density stated on the national standard step tablet or step wedge calibration film.
6.4.5. Shift calibration verification checks shall be performed at the beginning of each shift, after 8hr of continuous
use, or after change of apertures, whichever comes first.
6.4.6. Densitometer calibration readings shall be recorded in the densitometer calibration and shall be approved by
QCM
6.4.7. Shift verification readings required above do not have to be recorded.
CALIBRATION PROCEDURE
ASF-QC-CAL-001
QUALITY CONTROL DEPARTMENT REV. 1
CALIBRATION PROCEDURE Date: 01-03-2019
ASF-QC-CAL-001 Page: 7 of 7
7. Records
All the data collected during the calibration activity shall be recorded in the issued certificates/reports, then labeled on
the item/instrument/machine.
8. ATTACHMENT
8.1. Equipment Calibration list – form no. ASF-QC-CAL-002
8.2. Internal Equipment calibration certificate – form no. ASF-QC-CAL-003
8.3. Calibration Label – form no. ASF-QC-CAL-004