The Accuracy and Stability of Quartz Watches
by Michael Lombardi 1-A 2-B
Quartz wristwatches are neither as
intricate nor as intriguing to many
collectors as their mechanical
counterparts, but with very few
exceptions, they do a considerably better
job of keeping time. At least one
manufacturer of low-priced quartz
watches specifies their accuracy as ±15
seconds per month, suggesting an
accumulated error of just a few minutes
per year. This type of accuracy is
sufficient for most people, who are
generally happy if their watch remains
within a minute or two of the correct time. 3-C 4-D
In fact, many quartz watch owners set
their watches only a few times per year –
typically when they change the battery,
change time zones, or switch to and from
daylight saving time. Unless their watch
is broken or the battery is dead, its
timekeeping accuracy is never in
question.
But for those among us who view even
the cheapest quartz watch as a precision
scientific instrument, rather than as a
piece of jewellery or as a disposable
consumer item, some questions remain. Like nearly all quartz watches, the frequency, or an ideal frequency with
For example, exactly how accurate is a four devices under test use 32.768 kHz zero uncertainty. For example, if a crystal
‘run-of-the-mill’ quartz wristwatch? Can (215 Hz) quartz crystals as their oscillator. with a nominal frequency of 32768 Hz is
they really keep time to within ±15 The quartz watch industry standardised measured at 32768.5 Hz, its frequency is
seconds per month? Does their accuracy on 32 kHz crystals in the early 1970s due said to be accurate to within 0.5 Hz.
vary over time? This article attempts to to their reliability, their compatibility with Both time accuracy and frequency
answer those questions. It characterises existing electronic circuits, their small accuracy are normally expressed as
the performance of four low-cost quartz dimensions, and their low power dimensionless values by using the
wristwatches by applying some consumption.1 Since their introduction, equations Δt/T and Δf/f, respectively.
measurement and data analysis watch manufacturers have continued to The two equations produce equivalent
techniques that are normally reserved for improve the timekeeping performance of answers when applied to the same
laboratory type frequency standards. quartz watches. Most of the advances device. Thus a time accuracy of 1.3 /
have been related to crystal and mount 86400 (seconds per day) and a
The Watches Under Test miniaturisation, better electronics, better frequency accuracy of 0.5 Hz / 32768 Hz
The four quartz watches chosen for manufacturing techniques, and most both result in a dimensionless accuracy
the test, 1-4, are members of the importantly, making the crystal frequency value of about 1.5 × 10-5.
author’s pedestrian collection. While less dependent on temperature.2 Stability indicates how well a device
none of them will make a watch can produce time or frequency with the
enthusiast’s heart beat faster, they do Accuracy versus Stability same accuracy over a given time
have the virtue of being common; and The performance of a timekeeping interval. It doesn’t indicate whether the
similar watches have found their way on device is usually stated in terms of its time or frequency produced by a device
to many wrists. Watch A is an ‘official’ accuracy and stability, and measuring is accurate or inaccurate, but only
Mickey Mouse watch, purchased at both characteristics was the goal of this whether it stays the same. In contrast,
Disneyland in California several years test. Accuracy is related to the difference accuracy indicates how well a clock has
ago for about $35 USD. Watch B is a between a measured value and an ideal been set on time or an oscillator has
Rolex ‘replica’, purchased from a street value. For example, a ‘perfect’ watch been set on frequency. To understand
vendor in South America for about $15 would agree exactly with Coordinated this difference, consider that an
USD, and somewhat surprisingly, still Universal Time (UTC), the international inaccurate device can be stable, and an
running some two years later. Watch C reference for time, time interval, and unstable device can be at least
is a 20-year old dress watch that frequency. If a watch was synchronised temporarily accurate. For example, a
originally sold (mid-1980s) for about to UTC and then found to be 1.3 seconds quartz watch that gains exactly 10.5
$100 USD, and was worn everyday for fast one day later, its time is said to be seconds every day is very inaccurate,
more than a decade. Watch D is a accurate to within 1.3 seconds per day. but very stable. It might be possible,
typical discount store watch, a new Frequency accuracy refers to the then, to adjust the frequency of the
(2007) Timex that sells for approximately difference between the measured crystal and make the watch both
$30 USD. frequency of an oscillator and its nominal accurate and stable. In contrast, a watch

Horological Journal February 2008 57


that fluctuates within a range of ±5 here, the mechanical quartz oscillations to support its measurement resolution.
seconds of the correct time is unstable, are acoustically recorded. The device The watches under test were each
but on occasion would have the correct can also capacitively record the stray measured for a period of at least 30
time and be considered accurate. electrical field from quartz oscillators days. During the test, the watch analyser
The Allan deviation (ADEV) is a with open movements or with cases produced readings every minute for both
statistic used internationally to estimate made of synthetic material. It is also the frequency of the quartz oscillator and
frequency stability.3 It differs from the possible to derive the quartz frequency the stepping motor. It also produced a
conventional standard deviation from the supply current if the analyser is temperature reading with a resolution of
because it does not use the average providing power to the watch.6 1 °C. The watch analyser was interfaced
accuracy of a device as a point of to a computer through its RS-232 port,
reference. Instead, it compares the The Watch analyser (with watch D and all of the readings were stored for
frequency accuracy of the device under resting on the sensor) later analysis.
test during a given measurement period To get a true picture of the timekeeping The readings returned by the watch
to its frequency accuracy during the capability of an analog quartz watch, analyser were expressed as seconds
previous measurement period. This simply measuring the quartz frequency per day. This was converted to
reveals how an oscillator’s frequency is is not adequate. It is also necessary to dimensionless frequency offset
changing over time due to effects such measure the stepping motor pulses, (accuracy) using the equation Δt/T.
as frequency drift and aging. ADEV is because many watches correct the Average frequency accuracy was
regularly used to estimate the stability of frequency of the stepping motor to computed by simply averaging all of the
devices ranging from high-performance compensate for the frequency offset of 1 minute samples collected during the
mechanical watches4,5 to the world’s the quartz oscillator. This correction entire test. Frequency stability was
best atomic oscillators, and will be system, sometimes called inhibition estimated by use of the Allan deviation
applied here to estimate the stability of compensation, can be implemented in as previously described. The
the quartz watches under test. ADEV, several different ways. One common dimensionless frequency offset values
expressed mathematically as σy(τ) is way is to design the oscillating circuit so served as the yi data series. Because a
computed as that the quartz crystal runs at a new value was obtained every minute,
frequency slightly higher than nominal. the base averaging time, τ0 was equal to
M −1 To compensate for this intentional 1 minute.
1
σ y (τ ) = ∑ ( yi+1 − yi ) 2
2( M − 1) i =1
frequency offset, a programmable
number of quartz oscillation pulses are Measurement Results
suppressed before they are sent to the Table 1 shows the measured accuracy
frequency divider that drives the of the watches under test, both as
where the yi series contains estimates stepping motor. This removes the dimensionless frequency accuracy, and
of the frequency accuracy of the device frequency offset, and makes time as time accuracy (seconds per day).
under test, M is the number of values in derived from the stepping motor more Due to inhibition compensation, all of the
the yi series, and the data are equally accurate than time derived from the free watches are accurate to much better
spaced in segments τ seconds long. running quartz. The duration of the than 1 second per day. In response to
inhibition period, usually 10 or 60 our initial question, only one of the
The Measurement Method seconds, is automatically detected by watches under test failed to meet the
To estimate their accuracy and the watch analyser. Quartz pulses might ±15 seconds per month specification
stability, the watches were measured also be added or suppressed to that was discussed earlier. That was
with a commercial watch analyser, 5. compensate for the aging rate of the Watch C, the oldest watch in the test,
This versatile device can simultaneously quartz crystal, or for temperature and it missed by only a few seconds per
measure the frequency of both the changes. month. The quartz oscillators in watches
quartz oscillator and the stepping motor The watch analyser displays A, B, and C are not particularly accurate,
pulses. The watch analyser sensor can measurements of both the quartz with frequency offsets (perhaps
automatically detect the quartz frequency and the stepping motor with a intentionally introduced) ranging from
frequency through several available resolution of 0.01 seconds/day. The 5.9 to 10 parts in 105. In contrast, the
methods. If the watches have metal measurements are referenced to the quartz oscillator in Watch D was a stellar
cases, as did all of the watches tested time base oscillator inside the watch performer, with an average frequency
analyser, and to support this resolution, offset of just 8 × 10-7, or less than 1 part
5 the time base oscillator must have a per million. The accuracy of Watch D’s
frequency accuracy of better than about stepping motor was nearly identical to
1.2 × 10-7. The watch analyser was the accuracy of its quartz oscillator, so it
calibrated before and after it was used to is not clear if inhibition compensation is
measure the watches under test. The used in the design. However, the watch
calibration was done by locking a analyser detected an inhibition period of
synthesised signal generator to the 10 seconds, as reported in Table 1.
United States national frequency The stability estimates for the four
standard, and then deriving a reference watches are summarised in Table 2 and
32768 Hz signal from the signal illustrated in 6. The eight lines on the
generator that was accurate to parts in graph show the stability of both the
1013 or better. When this reference signal quartz oscillator and the stepping motor
was applied to the watch analyser for each of the four watches. The graph
sensor, it was correctly found to be within is an ‘all-tau’ graph, meaning that it
0.01 seconds/day. This indicated that shows stability estimates for all possible
the watch analyser was accurate enough values of τ, ranging from 1 minute to 1

58 February 2008 Horological Journal


Watch Inhibition Dimensionless Time Accuracy Temperature
Period Frequency Accuracy (seconds per day) during test (° C)
(seconds) Stepping Quartz Stepping Quartz Range Average
Motor Oscillator Motor Oscillator
week (in 1-minute increments). It is
A 10 5.3 × 10-6 7.9 × 10-5 0.46 6.79 22 to 25 23.3
interesting to note that the watches were B 60 2.1 × 10-6 5.9 × 10-5 0.18 5.09 21 to 25 23.8
most stable at τ = 1 hour, when all of the C 60 6.7 × 10-6 1.0 × 10-4 0.58 8.76 22 to 26 23.9
devices were stable to within less than D 10 7.8 × 10-7 8.0 × 10-7 0.07 0.07 22 to 26 23.3
2.5 × 10-8. At τ = 1 day, all of the devices Table 1: The accuracy of the watches under test.
were stable to at or near 3×10-8,
Watch Stability (Allan deviation)
suggesting that their accuracy will vary 1 minute 1 hour 1 day 1 week
by only a few milliseconds per day. Motor Quartz Motor Quartz Motor Quartz Motor Quartz
A 2.9 × 10-8 4.6 × 10-8 1.5 × 10-8 1.3 × 10-8 2.2 × 10-8 1.8 × 10-8 2.6 × 10-8 2.2 × 10-8
The Allan deviation graph for the B 4.1 × 10-8 8.4 × 10-8 1.3 × 10-8 1.3 × 10-8 2.9 × 10-8 3.2 × 10-8 4.0 × 10-8 4.0 × 10-8
C 3.4 × 10-8 6.7 × 10-8 2.3 × 10-8 2.0 × 10-8 2.9 × 10-8 3.1 × 10-8 4.8 × 10-8 5.4 × 10-8
watches under test D 2.5 × 10-8 6.6 × 10-8 1.2 × 10-8 1.1 × 10-8 2.9 × 10-8 1.7 × 10-8 2.2 × 10-8 1.6 × 10-8
While inhibition compensation Table 2: The stability of the watches under test.
dramatically improved the timekeeping the wrist, and about 8 hours off the wrist can maintain time accurate to within less
accuracy of three of the four watches each day. If the watch is left off the wrist than 1 second per day with the aid of
(Table 1), it seemed to only significantly for extended periods, its accuracy can be inhibition compensation. And due to the
improve the stability at short averaging expected to degrade. The angle of cut of surprisingly good stability of 32 kHz
times. At longer averaging times, the the crystal resonator used in quartz crystal oscillators, the accuracy of
stability of the stepping motor was about wristwatches is such that the zero quartz wristwatches can be expected to
the same or worse as the stability of the temperature coefficient is usually in the change by only a small amount over
quartz crystal. The crossover point range of 25 °C to 28 °C (27 °C is typical), time.
where the stability of the quartz oscillator which is warmer than the laboratory
began to meet or exceed the stability of temperature during the test. The author is an employee of a US
the stepping motor occurred at less than Aging is the systematic change in Institute making this article a
25 minutes for watches A and C, and frequency with time due to internal contribution of the United States
near 1 hour for watches B and D. changes in an oscillator. All quartz government, and not subject to
As might be expected, the variation in oscillators age, but the aging rate often copyright. The illustrations and
frequency for watches A, B, and C was depends upon its surface area to volume descriptions of commercial products
larger at one week than it was at one ratio of the crystal; and in theory, small, are provided only as examples of the
day, due to the effects of frequency drift low frequency crystals will age slowly.7 technology discussed, and this
and aging. Frequency drift is generally The results seem to support this, as the neither constitutes nor implies
attributed to factors external to the crystal in the watches under test all were endorsement by the NATIONAL
oscillator, including environmental stable to within about 5 × 10-8 or better at INSTITUTE OF STANDARDS AND
factors such as temperature, vibration, τ = 1 week, and watch D was nearly as TECHNOLOGY (NIST).
and humidity. These factors were stable at one week as it was at one day.
reasonably well controlled in the The frequency stability of watch D
laboratory environment, and the watches suggests that its timekeeping accuracy
were certainly subjected to fewer would change by less than 2
environmental changes than they would milliseconds per day over the course of a
have been during normal use. However, week. Thus, in response to one of our References
it should be noted that the laboratory questions, quartz watches do change 1 J. Engdahl and H. Matthey, 32 kHz Quartz
temperature during the tests (Table 1) their accuracy slightly over time, but the Crystal Unit for High Precision Wrist Watch,
was lower than optimal. Quartz watches change is small and will probably not be Proceedings of the 1975 Frequency Control
are optimised to work best at a noticeable to the owner of the watch. Symposium, May 1975, pp. 187-194.
temperature that reflects the expected Summary 2 E. Momosaki, A Brief Review of Progress
temperature of the watch in normal Based on these tests, it seems likely in Quartz Tuning Fork Resonators,
operation. If the watch is worn as that even the humblest quartz wristwatch Proceedings of the 1997 IEEE Frequency
intended, this means about 16 hours on Control Symposium, May 1997, pp. 552-
565.
3 IEEE, IEEE Standard Definitions of
6
Physical Quantities for Fundamental
Frequency and Time Metrology - Random
Instabilities, IEEE Standard 1139-1999,
March 1999.
4 J. A. Frieman, Performance of a Railroad
Watch, HJ, 141(7), July 1999, pp. 243-246.
5 P. Woodward, Performance of the Daniels
Coaxial Escapement, HJ, 146(8), August
2004, pp. 283-285.
6 Witschi Electronics Ltd., analyser Q1:
Instruction Manual, Witschi Document
Number 26.6410D35e, Rel. 1.0, 2005.
7 J. R. Vig and T. R. Meeker, The Aging of
Bulk Acoustic Wave Resonators, Filters,
and Oscillators, Proceedings of the 1991
IEEE Frequency Control Symposium, May
1991, pp. 77-101.

Horological Journal February 2008 59