Professional Documents
Culture Documents
Tugas DR Nency
Tugas DR Nency
NIM : 406172091
FK UNTAR
1. Jelaskan mengenai prinsip circuit mesin anestesi: breathing ciricuit, aliran O2,
ambu bag, APL (applicable press loss)
1. The Adjustable Pressure Limiting (APL) valve allows a variable pressure within
the anaesthetic system using a one-way, spring-loaded valve. At a pressure
above the opening pressure of the valve, a controlled leak of gas is allowed
from the system, which enables control of the patient’s airway pressure. The
minimum pressure required to open the valve is 1cm H2O. A safety
mechanism exists to prevent pressure from exceeding 60cm H2O, however, be
aware that pressures below this can lead to barotrauma.
2. The reservoir bag allows collection of fresh gas flow during expiration, which in
turn minimises the amount of fresh gas required to prevent rebreathing. In
addition, it allows the anaesthetist to monitor the breathing pattern of a
spontaneously breathing patient. These are usually plastic or rubber, and can
come in sizes between 0.5 litres to 6 litres. However, the most common size in
the adult system is 2 litres. Laplace’s Law states that pressure is equal to twice
the radius divided by the radius of the bag. Therefore, as the bag increases, the
pressure within it reduces. This is an important safety measure as the
expansion of the bag to accommodate gas limits pressure within the system.
3. The inspiratory limb allows passage of fresh gas flow to the patient for inspiration.
The expiratory limb allows passage of expired gas from the patient. Although
tubing length varies depending on the system in use, the diameter is of
standard size: 22mm for adult and 18mm for paediatric systems.
CIRCLE SYSTEMS
A circle system improves the efficiency of anaesthetic gas delivery by recycling gas
that is expired from the patient and thus reducing the amount of fresh gas flow
required.
Fresh gas flow leaves the anaesthetic machine and passes to the patient via a one-way
valve during inspiration. As expiration occurs, the exhaled gases from the patient pass
via a one-way valve to the APL valve, then on to the reservoir bag (or ventilator).
Before this expired gas is mixed with the fresh gas flow and delivered to the patient, it
passes through soda lime, which absorbs carbon dioxide. Initially, a high fresh gas
flow is required in order to fill the breathing system with the desired mixture and to
equilibrate the system, after which a minimal flow of 0.5 litres per minute can be
used.
A circle system can be semi-closed or closed. In a semi-closed circle, the APL valve
is opened, and allows excess gas to be removed from the system and reduce the risk
of barotrauma. However, the relatively high fresh gas flow allows a vaporiser outside
circle (VOC) to be used, which can introduce a higher percentage and more precise
anaesthetic gas into the mixture.
In a closed circle, the APL valve is completely closed. Although this is the most
efficient anaesthetic breathing system, it leaves little margin for error as the fresh gas
flow must meet the exact patient requirements, and the soda lime must absorb all
expired carbon dioxide. The minimal flow in this system only allows the use of a
vaporiser inside circle to be used. A detailed description of vaporisers is beyond the
scope of this article.
One of the most important components of the circle system is soda lime. This is a
mixture of 80% calcium hydroxide, 4% sodium hydroxide, and 16% water. It also
contains a pH-sensitive dye, which indicates when the granules are exhausted. Soda
lime granule are described as 4-8 mesh, which means that each granule will fit
through a mesh that has 4 openings per inch, but not one that has 8. The following
exothermic reaction occurs:
The one-way valves can become stuck by water vapour within the system
leading to an increase in dead space.
The one-way valves increase the resistance to breathing in the system.
The lower the fresh gas flow rate, the longer it takes for changes made to the
anaesthetic gas mixture to occur.
Monitoring the gas composition within the circle is essential.
Use of sevoflurane at low flow rates below 1 litre/minute can generate
Compound A by reaction with soda lime. Although this is nephrotoxic in rat
models there is currently no evidence of harm in humans.
One must be familiar which the colour of the pH-sensitive dye as different
manufacturers use different colours to indicate that the soda lime is exhausted.
Uneven distribution of soda lime granules in the canister causes gas to flow
unevenly and reduces the efficiency of the soda lime.
GAS SUPPLY
Most machines have gas inlets for oxygen, nitrous oxide, and air. Compact models
often lack air inlets, whereas other machines may have a fourth inlet for helium,
heliox, carbon dioxide, or nitric oxide. Separate inlets are provided for the primary
pipeline gas supply that passes through the walls of health care facilities and the
secondary cylinder gas supply. Machines therefore have two gas inlet pressure gauges
for each gas: one for pipeline pressure and another for cylinder pressure.
Pipeline Inlets
Oxygen and nitrous oxide (and often air) are delivered from their central supply
source to the operating room through a piping network. The tubing is color coded and
connects to the anesthesia machine through a noninterchangeable diameter-index
safety system (DISS) fitting that prevents incorrect hose attachment.
Noninterchangeability is achieved by making the bore diameter of the body and that
of the connection nipple specific for each supplied gas. A filter helps trap debris from
the wall supply and a one-way check valve prevents retrograde flow of gases into the
pipeline supplies. It should be noted that most modern machines have an oxygen
(pneumatic) power outlet that may be used to drive the ventilator or provide an
auxiliary oxygen flowmeter. The DISS fittings for the oxygen inlet and the oxygen
power outlet are identical and should not be mistakenly interchanged. The
approximate pipeline pressure of gases delivered to the anesthesia machine is 50 psig.
Cylinder Inlets
Cylinders attach to the machine via hanger-yoke assemblies that utilize a pin index
safety system to prevent accidental connection of a wrong gas cylinder. The yoke
assembly includes index pins, a washer, a gas filter, and a check valve that prevents
retrograde gas flow. The gas cylinders are also color-coded for specific gases to allow
for easy identification. In North America, the following color-coding scheme is used:
oxygen = green, nitrous oxide = blue, carbon dioxide = gray, air = yellow, helium =
brown, nitrogen = black. In the United Kingdom, white is used for oxygen and black
and white for air. The E-cylinders attached to the anesthesia machine are a high-
pressure source of medical gases and are generally used only as a backup supply in
case of pipeline failure. Pressure of gas supplied from the cylinder to the anesthesia
machine is 45 psig. Some machines have two oxygen cylinders so that one cylinder
can be used while the other is changed. At 20°C, a full E-cylinder contains 600 L of
oxygen
stated capacity.130 Although most bags adhere to these standards, some latex-free
bags have exceeded this maximal value.129 Classically, the reservoir bag was
excluded from the breathing circuit when the ventila- tor was in use. However, on
some contemporary work- stations, such as the Dräger Fabius and Dräger Apollo, the
reservoir bag is integral to circuit function during mechanical ventilation, where it
excludes or closes the valve.127 Several other common names exist for these devices,
including “pop-off” valve and pressure relief valve.122 The two basic types of
pressure-limiting valves are the variable-resistor (or variable-ori ce) type and the
pressure-regulating type. The variable-type functions as a needle valve, much like a
ow control valve (Fig. 29-31). The operator adjusts the outlet ori ce size, so the
resultant breathing system pressure at any given adjust- ment is directly related to the
fresh gas ow rate. Modern machines now mostly use pressure-regulating–type APL
valves (Fig. 29-32). This type of APL valve has an adjust- able internal tension spring
and an external scale indicat- ing approximate or relative opening pressure. When the
pressure in the system exceeds spring tension, a disk opens and gas is vented (see Fig.
29-32, B). In this manner, the operator can adjust the circuit pressure, which remains
stable even as the fresh gas ow is increased. Continu- ous positive airway pressure
(CPAP) can be more reliably controlled using this type of APL valve; however,
circuit pressure should be carefully monitored. This type of valve usually has a fully
open position for spontaneous breath- ing whereby the valve is open to the
atmosphere (see Fig. 29-32, C). Gas is prevented from returning from the scav-
enging system by a downstream check valve. When the valve is placed in the manual
mode, spring pressure is applied to the valve in an amount proportional to the desired
https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3273374/pdf/ijgm-5-117.pdf
3. Jelaskan mengenai recovery position
Recovery position mengacu ke salah satu dari serangkaian variasi pada berbaring
lateral atau posisi tiga perempat pronasi tubuh, di mana korban tidak sadar tetapi
bernapas, dan ditempatkan sebagai bagian dari pertolongan pertama.
Tujuan Recovery position untuk mencegah kekurangan oksigen karena obstruksi jalan
napas, yang bisa terjadi pada pasien terlentang dan tidak sadar.
4. Setting ventilator
Pengaturan Ventilasi Mekanik ( Setting)
Parameter yang harus ditetapkan sangat bervariasi tergantung pada mode ventilasi
yang digunakan. Beberapa parameter tersebut antara lain:
Pada sebagian besar kasus, FiO2 harus 100% pada saat pasien diintubasi dan
dihubungkan dengan ventilator untuk pertama kali. Ketika penempatan pipa
endotrakea sudah ditetapkan dan pasien telah distabilisasi, FiO2 harus
diturunkan sampai konsentrasi terendah yang masih dapat mempertahankan
saturasi oksigen hemoglobin , karena konsentrasi oksigen yang tinggi dapat
menyebabkan toksisitas pulmonal. Tujuan utama ventilasi adalah
mempertahankan nilai saturasi 90 % atau lebih. Kadang-kadang nilai tersebut
bisa berubah, misalnya pada keadaan-keadaan yang membutuhkan suatu
proteksi terhadap paru-paru dari volume tidal, tekanan dan konsentrasi
oksigen yang terlalu besar. Pada keadaan ini, target saturasi oksigen dapat
diturunkan sampai 85% saat faktor-faktor yang berperan pada penyaluran
oksigen sedang dioptimalkan.
Laju pernapasan = 10
Waktu siklus respirasi = 6 detik
I:E ratio = 1:2
Waktu inspirasi = 2 detik
Waktu ekspirasi = 4 detik
Volume tidal = 500 ml
Laju aliran = volume/ waktu inspirasi = 500 ml tiap 2 detik
h. Perbandingan waktu inspirasi terhadap waktu ekspirasi
Sejalan dengan laju aliran inspirasi, ahli terapi respirasi mengatur I:E ratio
tanpa permintaan dari dokter. Tetapi para klinisi dituntut untuk mengerti
tentang perubahan ini yang dapat mempengaruhi mekanika sistem respirasi
dan kenyamanan pasien. I:E ratio yang umum digunakan adalah 1:2. Pada
gagal napas hipoksemia akut, perbandingan ini dapat meningkat dengan
adanya pemanjangan waktu inspirasi, tekanan jalan napas rata-rata atau alveoli
yang terisi cairan yang dapat memperbaiki oksigenasi. Pada hipoksemia berat,
I:E ratio kadang-kadang terbalik menjadi 2:1, sehingga kewaspadaan harus
dipertahankan untuk mengatasi akibat yang merugikan terhadap hemodinamik
7
dan integritas paru-paru.
5. Pemeriksaan fisik berdasarkan American Society of Anesthesiologist
6. Difficult airway management