Professional Documents
Culture Documents
25 Installation
25 Installation
gehealthcare.com
Technical
Publications
2244772
Revision 25
Operating Documentation
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 22 2244772
a
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 22 2244772
b
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 22 2244772
The equipment is sold with the understanding that the General Electric Company, Healthcare Group, its agents,
and representatives have no responsibility for injury or damage which may result from improper use of the equip-
ment.
Various protective material and devices are available. It is urged that such materials or devices be used.
All electrical installations that are preliminary to positioning of the equipment at the site prepared for the equipment
shall be performed by licensed electrical contractors. In addition, electrical feeds into the Power Distribution Unit
shall be performed by licensed electrical contractors. Other connections between pieces of electrical equipment,
calibrations, and testing shall be performed by qualified GE Medical personnel. The products involved (and the
accompanying electrical installations) are highly sophisticated, and special engineering competence is required.
In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s
electrical work on these products will comply with the requirements of the applicable electrical codes.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers, personnel of third–
party service companies with equivalent training, or licensed electricians) to perform electrical servicing on the
equipment.
c
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 22 2244772
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent, have notation “damage in
shipment” written on all copies of the freight or express bill before delivery is accepted or “signed for” by a Gen-
eral Electric representative or a hospital receiving agent. Whether noted or concealed, damage MUST be reported
to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the contents and con-
tainers held for inspection by the carrier. A transportation company will not pay a claim for damage if an inspection
is not requested within this 14 day period.
Call the Service Center, People’s Republic of China, (86) 800-810-8188 immediately after damage is found. At this
time be ready to supply name of carrier, delivery date, consignee name, freight or express bill number, item dam-
aged and extent of damage.
Complete instructions regarding claim procedure are found in Section “S” of the Policy & Procedure Bulletins.
CAUTION
Do not use the following devices near this equipment. Use of these devices near this equipment could cause
this equipment to malfunction.
Devices which intrinsically transmit radio waves such as; cellular phone, radio transceiver, mobile radio transmit-
ter, radio-controlled toy, etc.
Keep power to these devices turned off when near this equipment.
Medical staff in charge of this equipment is required to instruct technicians, patients and other people who may be
around this equipment to fully comply with the above regulation.
d
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
A
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
REVISION HISTORY
REV Date Primary Reason for Change
25 09/11/06 Added notice for oil hose connection between the Gantry and Table according to PQR13075752. (Sec. 4)
23 08/23/05 PSR 13050797: added content of attaching local language warning labels. (Sec. 8)
21 09/16/04 Updated Table 4-1 and Table 4-2: first row Cable Dia. (mm) to Conductors cross-sectional Area (mm2) (Sec. 4)
Updated 4-5 SYSTEM POWER AND LOAD DISTRIBUTION CHECK and 4-6 GRADE KEY / OPTION SOFTWARE IN-
STALLATION and added contents of product protocol MOD installation (Sec. 4)
20 03/31/04 Added illustration of remove Lock Plate and install Gantry Dolly in Step 1 for 3-2 GANTRY INSTALLATION (Sec. 3)
Added 4-5-1 Product Name Change (PTO) (Sec. 4)
Updated 4-14-2 Network Setting, Step 3 for DHCP IP Addresses according to PQR 13018606 (Sec. 4)
B
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
REVISION HISTORY
REV Date Primary Reason for Change
19 10/23/03 Updated: Info for CRT Gamma Value (Sec. 4)
Changed: “System Func. Check” to “Diagnostics” (Sec. 4)
Updated: Directory “/CDROM” in 4-12-2 JEDI Error Definition File Update & TnT Definition File Update(Sec. 4)
16 01/27/03 Added: a picture of rear connection for OC in Illustration 3-21 OC Cable Inner Connection (Sec. 3)
15 08/10/02 Added: Monitor adjustment procedure for CRT monitor 2320287 (Sec 4)
8 12/22/00 Changed: The hoist rings for PDU are removed. (Sec.3)
OC monitor Adjustment (Sec 4)
Gantry and Table anchoring torque (sec 7)
Table foot switch (sec 8)
Added: Cable rated and actual voltage (sec 4)
Updated the CDRH certification status of Generator. (Sec. 9)
Colored: Sec. 2, 8, Appendix A.
5 7/06/00 Torque conversion table, New Gantry Dolly, Global modem connection, PDU Ground cable check, Small type of a Mat
switch
4 5/09/00 HSDCD brush block, LFC procedure deleted, JEDI setting, System cable color code, OC label
C
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
D
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
INSTALLATION
TABLE OF CONTENTS
SECTION PAGE
i
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
TABLE OF CONTENTS
SECTION PAGE
ii
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
TABLE OF CONTENTS
SECTION PAGE
4-12 JEDI INFORMATION CORRECTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-56
4-12-1 JEDI Date & Time Correction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-56
4-12-2 JEDI Error Definition File Update & TnT Definition File Update . . . . . . . . . . 4-56
4-13 LASER IMAGER INSTALLATION (OPTION) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-57
4-14 ADVANTAGE WORKSTATION INSTALLATION (OPTION) . . . . . . . . . . . . . . . . . . . . 4-58
4-14-1 Hardware Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-58
4-14-2 Network Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-58
4-15 CONNECTING THE SYSTEM TO THE NETWORK . . . . . . . . . . . . . . . . . . . . . . . . . . 4-63
4-15-1 Hardware Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-63
4-15-2 Network Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-63
4-15-3 Remote Host Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-63
iii
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
TABLE OF CONTENTS
SECTION PAGE
7-11 CT NUMBER ADJUSTMENT. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-21
7-12 IMAGE PERFORMANCE CHECK . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-21
iv
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
1-1 GENERAL
The following is the typical installation and check/adjustment procedure.
Installation Completion
TOTAL: 21.5 H (BY TWO ENGINEERS) except for JANAN
TOTAL: 20 H (BY TWO ENGINEERS) for JAPAN
1-1
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
1-2
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
1-3
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
1-4
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
CAUTION
Installation must be performed in the proper order outlined in Section 1, INSTALLATION
STEERING GUIDE, of this manual. All steps outlined in the Preinstallation manual must also
be complete before proceeding with these installation procedures.
2-1
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
2-2
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
____ ____ Covers and/or plastic sheeting with tie ends, to protect equipment from dust
____ ____ Standard service person tool kit (as per pole policy)
2-3
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
2-4
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
2-5
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
2-6
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
2-7
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
1 N•m=0.737 lbf•ft
N•m 0 1 2 3 4 5 6 7 8 9
lbf•ft
10 7.370 8.107 8.844 9.581 10.318 11.055 11.792 12.529 13.266 14.003
20 14.740 15.477 16.214 16.951 17.688 18.425 19.162 19.899 20.636 21.373
30 22.110 22.847 23.584 24.321 25.058 25.795 26.532 27.269 28.006 28.743
40 29.480 30.217 30.954 31.691 32.428 33.165 33.902 34.639 35.376 36.113
50 36.850 37.587 38.324 39.061 39.798 40.535 41.272 42.009 42.746 43.483
60 44.220 44.957 45.694 46.431 47.168 47.905 48.642 49.379 50.116 50.853
70 51.590 52.327 53.064 53.801 54.538 55.275 56.012 56.749 57.486 58.223
80 58.960 59.697 60.434 61.171 61.908 62.645 63.382 64.119 64.856 65.593
90 66.330 67.067 67.804 68.541 69.278 70.015 70.752 71.489 72.226 72.963
100 73.700
2-8
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 20 2244772
Gantry
Lock Plate
Lock Plate
Adjuster
Screw (x4)
3-1
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 20 2244772
Gantry Dolly
Gantry
Gantry Dolly
Adjuster
Screw (x4)
2. Use the dollies to raise the gantry off the floor, then wipe off the floor surface where the Gantry will rest.
3. Position the Gantry over its mounting marks.
4. Set the four leveling adjusters to about 5 mm lower than the bottom surface of the Gantry base.
GANTRY BASE
approx. 5mm
ADJUSTER
3-2
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 20 2244772
SIDE PROTECTION
RAIL
DOLLY
or
3-3
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 20 2244772
STUDS
SERVICE
ANGLE
3-4
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 20 2244772
SDS BD
CN1
3-5
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 20 2244772
3-6
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 20 2244772
Safety Switches
19. Loosen the safety bracket mounting bolt.
20. Rotate the Gantry safety switch counterclockwise so that the Gantry can rotate.
21. Tighten the safety bracket mounting bolt.
Screws Bracket
NOTICE
Fragile part. The Photo Sensors are very sensitive. Be careful not to bend them.
23. Remove the two long hexa screws and the filter indicated in illustration 3-10.
Filter
FILTER REMOVAL
COLLIMATOR FRONT VIEW PATH (SIDE VIEW)
3-7
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 20 2244772
Upper Plate
TUBE
Mark
Mark
Lower Plate
25. Install the filter and collimator cover in the reverse order of removal.
3-8
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 20 2244772
Housing Filtration
Insert Filtration
3-9
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 20 2244772
NOTICE
Use ethyl alcohol only at a pure of 99.5% when cleaning the slip rings.
NOTICE
Keep grease or dust away from the brushes since grease or dust will considerably
deteriorate the life and performance of the brushes.
a. Moisten Kim Wipe or equivalent with ethyl alcohol (99.5% pure only).
b. Clean off surface of all the slip rings and slip ring groove.
If a kim wipe gets smudged during cleaning, replace it.
c. Clean the ring housing and the surface of the base frame located below the slip ring housing, using a dry
kim wipe.
28. Install the Brush Block Assy: (See Illustration 3-13.)
NOTICE
Never touch the brush surface.
a. Carefully unpack the brush block.
b. Remove the protect cover from the brush block.
c. Set the brush block to the mounting plate on the Gantry frame, aligning the dowel pins of the brush block
with the holes of the plate.
d. Tighten the brush block securely, using four brush block mounting bolts packed with the brush block.
e. Connect the power cables onto the terminals of the brush block. (The cable must be connected to the same
number of the terminal.)
Torque: 48 kg-cm (for M6 screw), 20 kg-cm (for M4 screw)
f. Connect the cable to the CN1 for signal brush.
g. Reinstall the protect cover to the brush block.
3-10
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 20 2244772
Power Cables
Torque: 48 kg-cm (M6 screw)
Brush Block Assy
CN1 Torque: 20 kg-cm (M4 screw)
Signal Brush
Board Protect Cover
3-11
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 20 2244772
CASTERS
3. Rotate the four level adjusters until they are approximately 5 mm (1/4 inch) below the bottom surface of the
Table base.
GANTRY
REAR VIEW
5 mm
5 mm
SIDE VIEW
3-12
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 20 2244772
NOTICE
When aligning the table with the Gantry, take sufficient care not to damage the Valve unit
installed on the Gantry side.
5. Remove the valve caps form the valve joints of the table and Gantry.
6. Position the Table so that the two mating slots on the Table and Gantry stud bolts are properly aligned, then
connect the valve joints securely.
Valve Joints
3-13
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 20 2244772
TABLE BASE
MOUNTING BOLT
CASTER LONGITUDINAL
FRAME
TABLE BASE
CAUTION
After removing the dolly and before anchoring, the table lacks balance. Therefore, do not
place anything onto the cradle until the table is fixed with the anchors.
3-14
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 20 2244772
CAUTION
Before installing anchors the table might be lack in balance when it moves up or down and
the cradle moves in or out. Therefore, the table mounting nuts must be tighten securely.
TABLE BASE
WASHER
TABLE
MOUNTING
NUT
10. Install the cradle base cover and table side cover in the reverse order of removal.
3-15
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 20 2244772
3-4 OC INSTALLATION
NOTICE
The OC contains the HDD, MOD drive, CD-ROM drive, and other fragile equipments.
Take care when moving the OC unit.
2. Remove the rear nest cover and rear cover to allow access to OC inner wiring.
3. Remove the OC right side cover, then remove the SCSI bay shipping lock from inside the OC.
When the system contains the ST1800 AC/DC converter, cut its tie-wraps to access the shipping lock screws.
Then after removing the shipping lock, hold the converter again using tie-wraps.
3-16
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 20 2244772
CAUTION
CRT WEIGHT: APPROX. 17.5 kg.
When moving the CRT, take care not to be injured.
1. Place the OC CRT, keyboard, and mouse on the OC table as shown in Illustration 3-21. This Illustration is an
example. You can place the OC at either of the left or right side of the OC table.
KEYBOARD
OC
MOUSE
3-17
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 20 2244772
REAR
VIEW
SIGNAL
CABLE
POWER
CABLE
Mouse
Power
3-18
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 20 2244772
System cables
Keyboard cable
3-19
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 20 2244772
Note
The NPRM board contains 8 slots to install the optional NPRS boards. The optional NPRS board
must be installed at the slot next to existing NPRS board.
NPRM
NPRS 8 NPRS 1
NPRS 7 NPRS 2
NPRS 3 #7
NPRS 6 NEST
ASSY
NPRS 5 NPRS 4 #1
NPRM
3-20
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 20 2244772
Screws (Long)
2
NUTS
SCREWS
b. Remove the Protection plate from the transformer terminal, then set the PDU input RATING voltage by
changing voltage selection cable as shown in Illustrations 3-26 and 3-27.
3-21
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 20 2244772
Illustration 3-26 Transformer Coupling Terminal For 480V~380V system (Example: PDU1
Select
Proper
Input
Voltage.
PROTECTION COVER
TRANSFORMER
3-22
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 20 2244772
Illustration 3-27 Transformer Coupling Terminal For 200V system (Example: PDU1)
Select
200V.
TRANSFORMER
PROTECTION COVER
3-23
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 20 2244772
Ground Cables
3-24
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
4-1-1 General
This section describes the sequence of system cabling and power-on test. Complete as much of the Contractor
supplied wiring as possible before starting the system cabling. Some Contractor supplied wiring, for example
ground cables, must be pulled along with the system cables. In addition, some raceway work, such as cutting and
fitting covers at equipment locations, must be finished after completing the cabling and placing the equipment in
position.
All system cables use a system component designator for identification. The end of each system cable has a label
near the connector that identifies the CABLE NUMBER and CONNECTOR NUMBER.
4-1
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
4-2-1 Introduction
WARNING
ELECTROCUTION ! TURN OFF THE SWITCH OF THE POWER DISTRIBUTION BOX AND TAG
IT TO VERIFY THAT ALL POWER IS DISCONNECTED TO THE SYSTEM.
When performing system and ground wiring, ensure that all wiring conforms to the information contained in this
manual, in particular to the tables and illustrations contained in this section. The tables in this section list:
• All contractor supplied cables
• System furnished cables
• GEHW Supplied cable Part Numbers
• Destination connector or socket for the end of each cable
The illustrations provide detailed information on the actual connector and terminal strip locations. Illustrations
include :
• System Interconnect Diagram
• OC Cable Connection
• PDU Cable Connection
• Gantry Cable Connection
1. Verify that all ground wiring and other contractor wiring are complete to the point of equipment placement.
2. Check for reasonable cable routing. Don’t forget to leave enough cable take-up for equipment maintenance,
etc. Perform as neatly a job as possible.
Note
Keep signal and control cables away from power cables and power wiring. When laying cables in
a raceway, locate the signal cables in a separate section of the raceway.
3. Verify that both ends of each cable are clean cut and properly located before terminating and connecting them.
4. Ensure that all connections are tight. Use suitable tools. Check and tighten all visible connections, especially
ground connections.
NOTICE
Attach a tie-wrap to each system cable within 30 mm of each terminal end.
4-2
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
NOTICE
Check all the ground connections on the system for tightness and proper termination.
Check ground connection at the power supply transformer in a hospital. Check ground from
transformer to its earth connection.
Use the following standard identification in this Installation Manual.
• GROUP NUMBER-Is a number which is assigned to a group of cables or wires running between two
components. It may consist of only one cable, but usually contains more than one power and/or signal
cable. Group numbers are enclosed within a box. (ex; 3 means group number 3.)
• CABLE NUMBER-Is a unique number assigned to a single cable within a GROUP NUMBER. All
numbered cables are supplied with the system. Cable numbers are shown as numbers within a circle
(ex; 3 means cable number 3).
• ACTUAL LENGTH: total length of wire (see usable length).
• USABLE LENGTH: total length of a cable MINUS 1.5 m required for servicing and take up within
cabinets and the PDU.
Note
Actual usable length will depend upon the routing.
Yellow
Red Yellow
OC
PDU
Red Blue
Blue
GANTRY
Blue
TABLE
4-3
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
Cable Group FROM Insta TO Insta Cable Length Conduc- AWG# Part Rated Actual Notes
No. No. lled lled tors Number Voltage Voltage
cross-
Actual Usable section-
(m) (m) al Area
(mm2)
PDU
TB1-3 20 220240 AC600 UL14
21 1 PDB 18 9.5 -
or (Max.) 5 V 31
TB1-G13
PDU
SG
TB1-2 20 217694 AC600 UL14
22 2 Ground 18 12.1 #4 -
or (Max.) 2 V 31
Plate
TB1-G22
PDU
OC
TB1-1 20 217694 AC600 UL14
23 3 AC Box- 18 9.5 #6 -
or (Max.) 3 V 31
Ground
TB1-G11
Alre
SG ady PT
220307 AC600 UL14
25 4 Ground Inst Frame 3 3 9.5 #6 -
1 V 31
Plate alle Ground
d
4-4
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
4-5
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
Alrea
See SG AC115V
dy PT AC AC
4 Note NF2-3, 3 2 - 6 #14 2176621 (Power)
Install CN31 600V 115V
below. NF2-4 UL1015
ed
Alrea AC115V
SG
dy PT #14/ AC AC (Valve/
SUB - 7 2176622
Install CN32 18 600V 115V Pump)
BD CN4
ed UL1015
Alrea PT
SG DC
dy Table AC Control
TGP - 32 2176270 5V, DC
Install b’d 300V UL2464
CN11 24V
ed CN1
Alrea OC
SG Commun
dy REAR AC DC
5 300 TGP 19.5 18 10 31 2196035 ication /
Install CN1- 300V 12V
CN5 Intercom
ed J14
DAS data
SG Alrea
OC (Opt
RF dy
DASIF 12 34 2200197 - - cable,
RCV Install
N - J3 See
OPT ed
Notice.)
NoteNote
The cables in group No.4 run under the Gantry and do not require a raceway or duct.
NOTICE
Use extreme caution when installing the optical fiber cable. It is fragile and should not be
stepped on, kinked, or bent sharply.
4-6
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
TM2–3, 4 3 SW1–2,5
TM5–1, 2 4 BRK1–1,2
TB1–7,8,9 8 TM1–1,2,3
4
CN31 6 NF2–3,4
Signal Cable
Note
• All units must be isolated from building structures.
• Cable specifications are found in tables 4-1 and 4-2.
• Ground cable No.21 are not furnished with the system.
• Numbers inside of a box represent group number and numbers within a circle represent cable number.
4-7
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
J3 OC 41 J15 J14
REAR CN1
Cable Number 2
TM2–1,2 2 BRK2–1,2
TM2–3, 4 3 SW1–2,5
TM5–1, 2 4 BRK1–1,2
TB1–7, 8, 9 8 TM1–1,2,3
4
CN31 6 NF2–3,4
4-8
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
Lasercam (ANALOG) OC
GROUND GROUND
AC BOX
Lasercam (DIGITAL) OC
LCAM
DATA PANEL
ASSY
CONTROL J7
GROUND GROUND
AC BOX
4-9
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
For Multi Tech Modem ONLY: Hold the shielding parts of the cable using
the metal clamp and screw.
AC/DC converter Modem
NOTICE
This system is shipped with the Multi Tech modem (Global Modem) . However, some
countries do NOT approve this modem. Before connecting this modem, check if it is
approved for your country on website at http://www.multitech.com/support/gems/ , Initial
Setup & Country Selection.
When your country does NOT approve the Global modem, a modem currently used in your
country must be selected instead of the Global modem.
4-10
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
Rear
CN2
panel
4-11
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
Between Units
Cable No. Remarks
From To OC Connector Panel:
31 Gantry (TGP CN5) REAR CN1-J14 Communication/Intercom
34 Gantry (RF RCV) DASIFN-J3 Optical cable
41 PDU (J3) REAR CN1-J15
NOTICE
Do NOT bend the optical fiber cables or they will break.
Code
R1
Between Units
Cable No. Remarks
From To AC Box:
5 PDU TM1 AC110V
23 PDU AC Box-Ground point Ground cable
4-12
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
CABLE HOLDER
23
5
to Ground
to TM1
4-13
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
J15
J14
31
34
41
No.31 to J14
No.41 to J15
to J3
(Opt.cable)
OPTICAL CABLE HANDLING:
After installing the rear cover, place the
fiber optical cable onto the optical cable
holder of the rear cover, so that it
makes as large circle as possible, in or-
der to avoid breaking it.
4-14
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
2 NUTS
3 1
4-15
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
CB2 for
400V PDU CB2 for
200V PDU
OF
OFF
4. Remove one of the cable hole covers from the PDU, depending on cable planning for your site.
(The PDU contains the three cable covers.) Then, you see holes and slots, through which cables will be routed.
Store the removed cover. (Do not discard it.)
4-16
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
21
22 40
POWER CABLES
23
3
41 5
front
1 8 4 2
POWER CABLES
front
for PDU2 (2298852/2298849)
2 21 1
40 3 22 8
41 4 23
5
4-17
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
NOTICE
When connecting a cable to a PDU cable terminal, cut away the shielding so that the
shielding is not clamped between the metal contact plates.
SHUT SCREW
CABLE SHIELDING
NG OK
4-18
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
BROWN RMTCNT
2
NO.4 CABLE TB1
1 LIGHT BLUE
TIEWRAP
TM2
TM5
4
for PDU2 (2298852 / 2298849) Rear View
1 2
TIEWRAP
TM2 TM5
TB1
4-19
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
a
b
TB1
TIEWRAP
RMTCNT
BLACK 9
NO.8 CABLE WHITE 8
RED 7
TIEWRAPS
BLACK 6
NO.21 CABLE 3
NO.22 CABLE 2 TM2
NO.23 CABLE 1
clamp clamp clamp
21 23 1
22 8
Fix the No.1 and No.8 Clamp for
cables onto the tie–wrap Signal cable
base.
9 8 7 6 5 4 9 8 7 6 5 4
White
Black
Red
White
Black
Red
White
White
Black
Black
Red
Red
No. 21
No. 22
No. 23
TM2 TM5
TB1
No. 1 No. 1
Cable Cable CLAMP CLAMP CLAMP
No. 23
No. 21
No. 22
No. 8 No. 8 TIEWRAPS
Cable Cable
8
23 22 21 1
4-20
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
RMTCNT
1 2 3 4 5 6 7 8 9 10 11 12 TB1
TIEWRAP
TM2
TM5
2 5
1 2 3 4 7 8 9 10 11 12
TIEWRAP
TM2 TM5
TB1
2 3 5
4-21
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
RMTCNT
J2
Using the tie–wrap base
T J3 installed at the side frame,
clamp these cables.
Clamp
Clamp
TM5 TM
C C C
40
41 Other type of the PDU exists
in the field.
J3
TM2
TM5
J2 TB1
41 40
7. Install the PDU left side cover, then install the top cover in the reverse order of removal. (See Illustration 4-8.)
Do not install the PDU front cover at this time.
4-22
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
OFF
2.Remove the rear base cover.
3.Remove the maintenance cover by removing the two screws to access
Rotate Slip Ring
the terminals of the TM1.
XG Power Table/Tilt
(Safety Loop) 115V
Maintenance cover
4-23
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
B
W
R
BREAKER
NO.8 CABLE
NO.22 CABLE
6. Tie the No.22 cable (Ground, yellow/green) with other cables, using one tie-wrap.
7. Tie the No.8 cable (AC400V, XG power) on the tie-wrap base, already installed, using two tie-wraps.
8. Reinstall the maintenance cover removed in step 3.
4-24
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
NOTICE
Comfirm Oil Hose joints between the Gantry and Table (shown as Illustration 4-18 above) are
connected tightly and correctly, otherwise return oil hose may burst.
2. Connect the No. 6, 7, 25, and 32 cables (Gantry-to-Table cable) to the table.
NOTICE
The No. 6 and No.7 Cables (AC115V) must be fixed using a tie-wrap so that they do not touch
the table board.
4-25
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
Ground Cable
4-26
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
4-3-1 System Power-ON LIght, Rotor-ON Light, X-ray-ON Light, and Simultaneous X-ray Exposure
Interlock for other equipment
A light can be connected to the Gantry to indicate when ’System Power ON’, ’Rotor ON’ or ’X-ray ON’ is initiated.
System controlled contacts (RL2, RL3, and RL4) are provided within the Gantry.
These terminal can also be used to disable simultaneous X-ray exposure for other equipment when two or more X-
ray equipments are installed in the same room.
NOTICE
The Terminal Rating (TM4 and TM5):
240VAC, 5A and AWG22 to AWG12
Note
The following chart describes relationship between system status and contact open / close.
Terminals Contact
4-27
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
4-3-1 System Power-ON LIght, Rotor-ON Light, X-ray-ON Light, and Simultaneous X-ray Exposure
Interlock for other equipment (Continued)
1. Connect the System Power-ON light, Rotor-ON light, and X-ray-ON light to the terminals located at the Gantry
rear base:
Note
For connection of simultaneous X-ray exposure interlock, consider other equipment connected to
the CT scanner while following the terminal rating described on the previous page.
GANTRY
Breaker
TM5 TM4
X-RAY-ON
1 1
2 2
3 3
ROTOR-ON
4 4
5 5
6 6
SYSTEM POWER ON
TIEWRAP
4-28
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
4-3-2 Scan Room Door Safety Switch and Simultaneous X-ray Exposure Interlock for CT scanner
When the scan room door is opened, the X-ray exposure and Rotor-ON are NOT permitted.
This terminal can also be used to disable simultaneous X-ray exposure for CT scanner when two or more X-ray
equipments are installed in the same room.
NOTICE
The Terminal Rating (TM4): AWG22 to AWG12 ,
The Switch or Relay for other equipment: 24VDC, 200mA or more.
Note
The following chart describes relationship between system status and contact open / close.
TM4-3
TM4-3
4-29
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
4-3-2 Scan Room Door Safety Switch and Simultaneous X-ray Exposure Interlock for CT scanner
(Continued)
1. Connect the Scan Room Door Safety Switch or simultaneous X-ray exposure interlock relay to the terminal
located at the Gantry rear base:
Note
The example below shows when N.O. (Normal Open; with the switch OPEN at door closed) type
switch is connected. If N.C. (Normal Close; with the switch CLOSED at door closed) type is used,
change connection of the TM4-3 to the TM4-4.
Illustration 4-21 Connection for Scan Room Door Safety Switch or Simultaneous X-ray Exposure
Interlock Relay
GANTRY
Breaker
TIEWRAP
3 3
4 4
5 5
6 6
4-30
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
WARNING
VERIFY THAT WALL POWER (IN PDB) IS OFF, TAGGED, AND LOCKED.
Use an earth bonding test meter or equivalent device to measure the DC resistance between the following
components in the system, and verify that each is less than 0.1 W. Record resistance values in Table 4-3.
4-31
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
Note
For systems with PTO (Pick To Order) feature, the following steps should be performed to change
the product name.
1. After the Operating System (OS) is started, a red window appears:”Note: Activate the product name first
please”.
2. Insert the PTO CD-ROM (ProSpeed, CT/e Xtreme Series Product Naming). (Find the PTO CD-ROM in the
Unique Parts Box shipped with the product.)
3. “CT/e install utility” window appears. Click on Config button.
4. Reconfig window appears. Click on Accept button.
5. ”The system need to be rebooted for the changes to take effect. Do you wish to reboot now?” message ap-
pears. click on “Yes”.
6. The system reboots automatically.
7. Press Startup button to start application.
8. The system works normally. Press Service/Home button to verify the product name is correct.
Click the Shutdown Icon shutdown , located on the upper left of the screen. Then, click [OK] button in response to
“Attention: Shut down the system.”.
The system is automatically switched OFF.
4-32
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
4-33
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
Scan Parameter:
4-34
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
Tilt Arms
Adjustment Tools
4-35
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
4-36
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
Tilt Arm
4-37
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
ADJUSTER
Rear Base Cover
BUBBLE LEVEL POSITION
GANTRY
BASE
ADJUSTER
4-38
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
ADJUSTER
SHEET METAL BUBBLE LEVEL POSITION
BASE
Adjuster:
Turn the double-nuts
when adjusting the level.
Bubble Level:
Set the bubble level so that
it goes around the cables.
CAUTION
Crash Hazard. Tighten the two bolts between Gantry and Table securely. Otherwise, the
Gantry and Table will not be stable.
6. Tighten the two table mounting nuts securely. (See Illustration 4-22.)
4-39
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
Start
Color Temperature
Contrast
Adjustment
- Using SMPTE Pattern
Brightness (Minor Adjustment)
End
4-40
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
1 2
Reset
Brightness Menu Contrast LED Power SW
1. Press (menu) button on the front panel of the monitor to display setup menu.
4-41
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
% su <Enter>
2. Using or button of (Brightness), adjust brightness value so that view area of the monitor is the
same color as non-view area (edge section of monitor).
4-42
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
309mm~315 mm
Vertical: 41 lines
- Size / Center
Using this menu, you can adjust:
• Horizontal Position
• Vertical Position
• Horizontal Raster Size
• Vertical Raster Size
1. Press (auto sizing and centering) button on the front panel of the monitor to automatically sizing and
centering the picture.
Or, if the function above does not suit, perform the adjustment manually described bellow:
1. Press (menu) button on the front panel of the monitor to display setup menu.
4-43
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
- Geometry
Using this menu, you can adjust:
• Rotation
• V.Pincushion
• Trapezoidal
• Parallelogram
1. Press (menu) button on the front panel of the monitor to display setup menu.
4-44
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
1. Press (menu) button on the front panel of the monitor to display setup menu.
4-45
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
Contrast
1. Adjust the ambient room light to the level used by the operator during image review.
Brightness
5. Adjust Contrast and Brightness to the level an user desires so that a slight (not clear) distinction should be
made between 0% and 5% and between 95% and 100%.
5% 95%
0% 100%
6. Also check if the patterns are properly displayed.
4-46
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
Start
Color Temperature
Contrast
Adjustment
- Using SMPTE Pattern
Brightness (Minor Adjustment)
End
4-47
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
1. Power Switch
To turn the monitor on or off.
2. Power LED Indicator
As long as the monitor is powered, this indicator is lit. In normal operation, it lights green. While in power saving
mode, the indicator will turn to amber.
3. Adjust Keys
To adjust the parameter of the selected function.
To adjust the contrast level or brightness of the screen.
4. Menu Key
To display and select the "General", "Picture", "Geometry", "Color", "Advance", "Property" functions.
4-48
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
% su <Enter>
2. Press " " or " " (Manual Key) to open the On-Screen menu function. Six adjustment functions will appear
on the On-Screen menu. Press " " key to scroll selections from top-down sequence to select General as
below.
4-49
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
4. Press ”+” or ”-” button on the front panel of the monitor, adjust brightness value so that view area of the monitor
is the same color as non-view area (edge section of monitor).
4-50
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
309mm~315 mm
Vertical: 41 lines
- Size / Center
Using this menu, you can adjust:
• Horizontal Position
• Vertical Position
• Horizontal Raster Size
• Vertical Raster Size
1. Press " " or " " (Manual Key) to open the On-Screen menu function
2. Press " " key to scroll selections from top-down sequence to select Picture as below.
4-51
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
- Geometry
Using this menu, you can adjust:
• Side-pin
• Balance
• Keystone
• Parallel
• Rotate
1. Press " " or " " (Manual Key) to open the On-Screen menu function.
2. Press " " key to scroll selections from top-down sequence to select Geometry as below.
4-52
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
• Trapezoidal:
Make the screen image narrower or wider at the top so that all lines are perpendicular.
• Parallel:
Tilt the screen image leftward or rightward so that all lines are perpendicular.
• Rotate:
Tilt the image left or right so that the image can be horizontal to the monitor edge.
1. Press " " or " " (Manual Key) to open the On-Screen menu function..
2. Press " " key to scroll selections from top-down sequence to select Color as below.
4-53
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
Contrast
1. Adjust the ambient room light to the level used by the operator during image review.
2. Press " " or " " (Manual Key) to open the On-Screen menu function. Press " " key to scroll selections
to select General.
4. Press "+" or "-" button on the front panel of the monitor, adjust contrast value to 100.
Brightness
6. Press "+" or "-" button on the front panel of the monitor, adjust brightness value to 60.
4-54
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
5% 95%
0% 100%
9. Also check if the patterns are properly displayed.
4-55
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
4-12-2 JEDI Error Definition File Update & TnT Definition File Update
1. Insert the JEDI software CD-ROM into the CD-ROM drive of the OC.
2. Verify that the Generator test screen appears. If it does not, click on Service icon to display the Service Menu
desktop, then select Diagnostics -> Generator Test .
3. Select Error Definition file update, then Apply.
4. In Filter entry, select /CDROM/*.csv (for Irix OC) or /mnt/cdrom/*.csv (for Linux OC), then press the <Enter>
key.
5. In Directories entry, select CDROM/cdrompxxx (for Irix OC) or /mnt/cdrom/cdrompxxx (for Linux OC), then
press the <Enter> key.xxx : JEDI software version
6. In Files entry, select efilexxx.csv , then click on OK.xxx : JEDI software version
7. Verify that the message “xxxxx successfully” appears. (It takes a few seconds to complete an update.)
8. Click on Close.
9. Select TnT Definition file update, then Apply.
10. In Filter entry, select /CDROM/*.csv (for Irix OC) or /mnt/cdrom/*.csv (for Linux OC), then press the <Enter>
key.
11. In Directories entry, select CDROM/cdrompxxx (for Irix OC) or /mnt/cdrom/cdrompxxx (for Linux OC), then
press the <Enter> key..xxx : JEDI software version
12. In Files entry, select tntpxxx.csv , then click on OK..xxx : JEDI software version
13. Verify that the message “xxxxx successfully” appears. (It takes a few seconds to complete an update.)
4-56
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
4-12-2 JEDI Error Definition File Update & TnT Definition File Update (Continued)
14. Click on Close.
15. Click on Exit to exit from the Generator test. This shut down the system automatically.
16. Verify that the message “Okay to power off the system now. Press any key to restart.” appears.
17. Press the eject button of the CD-ROM drive to remove the CD-ROM from the OC.
18. Click on Restart.
19. Verify that the system starts up with no errors.
4-57
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
4-58
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
GATEWAY
IP ADDRESS
IP ADDRESS
HOST NAME
CT
4-59
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
00:1B:AC:36:E9
L E
M P
A
EX
GATEWAY
IP ADDRESS
IP ADDRESS
CT IP ADDRESS/MAC Address
4-60
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
255.255.255.0 gw 192.168.0.254
4-61
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
4-62
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 25 2244772
4. Click Save.
5. Verify that it is available to transmit the image data from the CT scanner.
6. Verify that it is available to execute Query/Retrieve from the DICOM workstation side.
4-63
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
4-64
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
5-1
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
blank
5-2
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
6-1
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
blank
6-2
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
NOTICE
Be sure to have the appropriate hospital personnel approve the location of the table/Gantry
before starting to drill any anchor holes.
7-2-1 Precautions
• Complete the Table to Gantry Alignment before Anchoring the Gantry and Table.
(Refer to Functional Check/Adjustment, Gantry, Table and Gantry, TABLE POSITIONING.)
• Prior to anchoring, after drilling the anchoring holes, check that the Gantry and Table are tightly fixed
together by two bolts.
(Refer to Section 4-8, LEVELING OF GANTRY AND TABLE).
DANGER
CRASH HAZARD!! THE TABLE AND GANTRY MUST BE ANCHORED.
7-2-2 General
CAUTION
Unstable condition. Keep the Table at its highest position. Don’t tilt the Gantry except if
necessary.
7-1
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
NOTICE
The minimum Tensile load strength of the anchor must be 13000 N.
1. Attach a piece of tape to the 12 mm concrete drill bit (2203081: HILTI, TE-CX 12/17 or equivalent) in order to
assure a minimum of 95 mm depth of anchor hole.
95mm or more
TAPE
7-2
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
When making rear holes, the Gantry Rotor section might disturb this work. If it does, rotate the Gantry to make
space.
NOTICE
During drilling, always keep the drill bit perpendicular to the floor. With one person drilling
the hole, a second person can visually insure perpendicularity throughout the drilling
operation. Clear the hole of debris several times while drilling to prevent the drill bit from
binding.
Note
If any obstruction is found under the proper anchoring position, stop drilling and use the alternate
position (see Illustration 7-2). Use the alternate positions only in the case where its corresponding
proper position is not possible.
Table Side
Rear
7-3
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
ADJUSTER
Rear Base Cover
BUBBLE LEVEL POSITION
GANTRY
BASE
ADJUSTER
7-4
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
Anchor Nut
Insulating Plate
Washer
9. Hammer the anchor into the hole made in step 2 until the insulating plate touches the frame.
10. Tighten the anchor nut to the specified torque to fix the anchor bolts into the floor.
TORQUE: 500 kg-cm
NOTICE
If specified torque cannot be achieved, another anchor point must be used. See Illustration
7-2 for alternative anchor position.
11. Loosen all of the anchor nuts, then re-tighten the anchor nuts securely (Approx. 50kg-cm).
NOTICE
Do not tighten the anchor nuts to the specified torque (500 kg-cm), because the anchor
bolts, tighten in previous step, has been fixed securely into the floor. So, to over-tighten
nuts will damage the frame of the Gantry and also bring an out-of-level Gantry.
7-5
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
Anchor nut
31mm or
less
Gantry base
Floor Floor
95mm or
more
12mm
7-6
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
NOTICE
The minimum Tensile load strength of the anchor must be 13000 N.
1. Verify that the table flapper is removed to make space for anchoring. If it is not, remove the table flapper by
removing the four flapper Assy mounting bolts.
Flapper Bolts
Table Flapper
7-7
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
NOTICE
During drilling, always keep the drill bit perpendicular to the floor. With one person drilling
the hole, a second person can visually insure perpendicularity throughout the drilling
operation. Clear the hole of debris several times while drilling to prevent the drill bit from
binding.
Note
If any obstruction is found under the proper anchoring position, stop drilling and use the alternate
position (see Illustration 7-8). Use the alternate positions only when the case its corresponding
proper position is not possible.
7-8
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
ADJUSTER
SHEET METAL BUBBLE LEVEL POSITION
BASE
Adjuster:
Turn the double-nuts
when adjusting the level.
Bubble Level:
Set the bubble level so that
it goes around the cables.
Bubble Level:
Set the bubble level so that
it goes around the cables.
7-9
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
NOTICE
Be sure to have the appropriate hospital personnel approve the location of the OC/PDU
before starting to drill any anchor holes.
7-3-1 General
The following hardwares are provided with optional anchor kit.
• Optional Anchor Kit For OC:
Anchor bolts (2 pcs)
Washers (2 pcs)
Insulating Plates (2 pcs)
Anchor nuts (including cone wedge) (2 pcs)
Anchor Stick (1 pcs) (Anchor nut hammering tool) (Outside Japan)
• Optional Anchor Kit For PDU:
Anchor Plate (2 pcs)
Anchor Plate mounting bolt (4 pcs)
Anchor bolts (2 pcs)
Washers (2 pcs)
Insulating Plates (2 pcs)
Anchor nuts (including cone wedge) (2 pcs)
Anchor Stick (1 pcs) (Anchor nut hammering tool) (Outside Japan)
NOTICE
The minimum Tensile load strength of the anchor must be 13000 N.
7-10
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
7-3-2 OC Anchoring
1. Turn the two adjusters counterclockwise so that they do not touch floor surface.
Adjuster
2. Select positions of the two anchor plates from the following two ways. (See Illustration 7-11.)
Note
Two anchor plates must be positioned diagonally.
3. Mark the two anchor points on the floor using a pen, then move the OC to make the holes at the marked
positions.
POSITION I POSITION II
Anchor
Plate
Front
Anchor
Plate Front
7-11
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
5cm
TAPE
5. Make holes at the anchor position marked in step 3, using electric drill with the bit prepared in the previous
step, so that the lower edge of the tape just touch the floor surface.
NOTICE
During drilling, always keep the drill bit perpendicular to the floor. With one person drilling
the hole, a second person can visually insure perpendicularity throughout the drilling
operation. Clear the hole of debris several times while drilling to prevent the drill bit from
binding.
6. Attach a piece of tape around the 18 mm concrete drill bit (Illustration 7-13) in order to assure 6 cm depth of
anchor nut hole.
6 cm
TAPE
7-12
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
HAMMER
(1.3 kg head Recommended)
ANCHOR
STICK Nut Floor
Edge
HOLE
7-13
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
NOTICE
Do not over-tighten the anchor bolt. Otherwise, the OC frame will be broken.
INSULATING PLATE
OC FRAME
ANCHOR NUT
7-14
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
Adjuster
2. Select positions of the two anchor plates from the following four ways. (See Illustration 7-17.)
Note
Two anchor plates must be positioned diagonally.
3. Mark the two anchor points on the floor using a pen, then move the PDU to make the holes at the marked
positions.
7-15
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
POSITION I POSITION II
Anchor
Plate
Anchor
Plate
7-16
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
7-17
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
Adjuster
7-18
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
Lock Nut
Adjuster
7-19
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
7-6 INTERCOMMUNICATION
Refer to :
Functional Check/Adjustment, OC, OC Operation, AUDIO FUNCTION.
NOTICE
The characters of alphanumerics, underscore, and period ONLY can be used. If other
characters would be used, the system will not start up. (For the system of V/R 2.5 or more,
other characters are rejected to input even if they are input.)
Refer to :
Functional Check/Adjustment, Introduction, Basic Operation, PERFORMING THE SYSTEM CONFIGURATION.
7-20
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 18 2244772
7-10 CALIBRATION
Refer to :
Functional Check/Adjustment, System, System Calibration, HILIGHT CALIBRATION and AIR AND PHANTOM
CALIBRATIONS.
7-21
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
blank
7-22
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 23 2244772
Slice Number
Rotation Number
SERVO AMP
X-ray slice coutner
sub
board
TGP BOARD
Select Reset Total Gantry Revolution Count, then enter the same number as Rotation Number recorded in
the previous step.
8-1
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 23 2244772
8-2-1 OC
Install the following covers in the reverse order of removal. For detailed information, refer to Section 3–4 OC
INSTALLATION.
• Rear cover
8-2-2 Gantry
Install the following covers in the reverse order of removal. For detailed information, refer to Section 3–2 GANTRY
INSTALLATION.
Note
The left and right side cover must be installed before front or rear cover installation.
• Rear cover
• Front cover
• Left and Right side covers
NOTICE
When installing the Gantry front cover, first cover the front edge of the cradle with a cloth
to prevent the front cover from scratches or damage.
Cloth
CAUTION
If the service angles (cover stand) fall over, they can injure persons. Therefore, after using
the service angles, keep them in a safe place and/or secure them from falling over even when
the customer or other persons accidentally hit the service angles.
8-2
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 23 2244772
8-2-3 Table
Flapper
1. Insert two(2) plastic rings into the left side of flapper and the right side of flapper.
2. Attach two flapper brackets to the flapper.
For Table part number 2297024 (LCT Table), follow the next steps:
a. Remove the two lock plates from the left/right side of the flapper bracket by four screws.
b. Insert 2 plastic wings into the left side and the right side of the flapper.
c. Put the flapper in the hatches in the flapper bracket.
d. Attach the lock plates to the flapper bracket securely.
3. Tighten it to the Table rear bottom by four (4) support bracket screws.
There is a gap between the flapper and the table arm as shown in the Illustration below. When the flapper is
installed on the table base, the left gap and the right gap must be equal using four (4) support bracket screws.
TABLE ARM
GAP
FLAPPER
SUPPORT
BRACKET
SCREW
PLASTIC RING
FLAPPER BRACKET
FLAPPER .
8-3
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 23 2244772
Left Mat / Pedal Switch (Example, Mat switch Right Mat Switch
type) Connector (Inside the case) Connector
Note:
The length of Pedal Switch
cables outside Table must be
approximately 20 to 30 cm.
Tie-wrap
8-4
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 23 2244772
Bottom Covers
7. Install the front bottom cover using four screws. After installation, install the screw caps onto the screws.
8. Install the rear bottom cover so that its tabs are aligned with the frame pins.
9. Pull the cover rearwards, then mount it using two screws.
After installation, install the screw caps onto the screws.
FRONT
BOTTOM
COVER
8-5
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 23 2244772
Block
8-6
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 23 2244772
WARNING LABEL
KEYBOARD
8-7
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 23 2244772
3. Check and attach appropriate language labels.
Before attaching label, thoroughly clean the machine surface where to attach the label.
Refer to the table below for proper location and attaching method.
NOTICE
In order to avoid any violation of each National Regulation, the appropriate language warn-
ing labels MUST be attached.
Quan-
Description PN Pictorial Location and Method
tity
PDU Front
Cover label - 5149894 1
Estop
PDU Front
Cover label - 5149878 1
Power lamp
8-8
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
Quan-
Description PN Pictorial Location and Method
tity
Head Holder
Warning Label 2182119
(34 Kg)
8-9
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 23 2244772
8-4-2 OC Emblem
4. The two OC emblem ( “ProSpeed AI” and “ProSpeed FI” ) comes with the system.
Select the proper one and attach it to the location shown in Illustration 8-5.
a. Power ON the OC.
b. Check the scan time of your system on the scan screen.
c. When 1.5 sec exist as a scan time, select “ProSpeed AI” or “CT/e”.
When 1.0 sec exist as a scan time, select “ProSpeed FI”.
5. To correctly attach the emblem,
a. clean the surface on the upper left section of the left CRT,
b. remove the backing paper,
c. carefully position the label (see Illustration 8-5),
d. rub adequately on the label surface so that the letters properly adhere to the CRT surface,
e. remove carefully the front paper.
6. The caution label in English has been attached on the OC. Since the system contains the same label with other
languages, replace it with an appropriate label.
EMBLEM
8-10
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 23 2244772
70mm
ProSpeed AI
35mm
70mm
CT/e
27.5mm
8-11
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 23 2244772
8-12
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 23 2244772
6. Initiate Reconfiguration:
a. Click on Accept after completing the configuration.
“Installing Product sw” window appears.
The System configuration procedures are started, then automatically login as root. It takes approx. 3
minutes.
8-13
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 23 2244772
8-7-1 Preparation
• One MOD (3.5 inches 230MB)
8-7-2 Procedures
1. Verify that the system is powered ON. If it is not, switch ON the main switch of the OC. The system is started
up, then the main screen appears.
2. Insert the MOD into the MOD drive of the OC.
3. Click on Service menu icon.
4. Select:
Utilities -> Save System State.
5. Click on Continue several times according to the messages to be displayed.
The MOD auto-format phase will be started, then the system state data are saved on the MOD.
6. “System State Saved Successfully” appears. Click on Continue.
8-14
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 24 2244772
OR
9-1
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 24 2244772
HV Tank 2186730 NO
9-2
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 24 2244772
9-3
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 24 2244772
9-4
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 8 2244772
Direct Current
Emergency Stop
A-1
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 8 2244772
Type B
417-5009 Standby
Start
Table Set
Abort
Intercom
Warning sign :
Radiation of Laser Apparatus
A-2
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 8 2244772
Symbol Description
Microphone (Mic)
Contrast
Brightness
-10°C
Humidity 5-95%
Excluding Condensation
A-3
CT/e, CT ProSpeed AI/FI Series
GE HEALTHCARE INSTALLATION
REV 8 2244772
CLASS 1 EQUIPMENT
Any permanently installed equipment containing operator or patient accessible surfaces must provide backup
protection against electric shock,in case the BASIC INSULATION fails. In addition to BASIC INSULATION,Class1
equipment contains a direct connection to a PROTECTIVE(EARTH) CONDUCTOR which prevents shocks when a
person touches a broken piece of equipment or touches two different equipment surfaces simultaneously.
TYPE B EQUIPMENT
CLASS I, II, or III EQUIPMENT or EQUIPMENT with INTERNAL ELECTRICAL POWER SOURCES provide an
adequate degree of protection against electric shock arising from (allowable) LEAKAGE CURRENTS or a
breakdown in the reliability of the protective earth connection.
ORDINARY EQUIPMENT
Enclosed EQUIPMENT without protection against the ingress of water.
OPERATION 0f EQUIPMENT
CONTINUOUS OPERATION WITH INTERMITTENT LOADING.
Operation in which EQUIPMENT is connected continuously to the SUPPLY MAINS.The stated permissible loading
time is so short that the long term on-load operating temperature is not attained.The ensuing interval in loading is,
however, not sufficiently long for cooling down to the long term no-load operating temperature.
EQUIPMENT not suitable for use in the presence of a FLAMMABLE ANESTHETIC MIXTURE WITH AIR or WITH
OXYGEN or NITROUS OXIDE
CLEANING
The system is NOT WATERPROOF. It is NOT designed to protect internal components against the ingress of
liquids.Clean external system surfaces(Gantry,table,consoles and accessories)with a soft cloth dipped in hot water
and wrung DAMP/DRY. (NOT dripping!) IF NECESSARY, use only mild (dish washing liquid) soap to remove dirt.
NOTICE
Avoid damage to equipment! Some îspray and wipeîcleaners etch and permanently cloud
clear plastic surfaces!! Use only warm water and mild liquid soap to clean surfaces.
A-4
www.gehealthcare.com
Imagination at work