CUSTOMER : M/S FONTUS WATER PVT. LTD.

D-4, FREEDOM FIGHTER ENCLAVE, NEB SARAI, IGNOU ROAD NEW DELHI
Calibration Certificate of : DIGITAL WEIGH BALANCE Certificate No. – HTCS/03169
Make : INTERFACE Service Request No. : HTCS/03169
Range : 0-30 Kg Service Request Date : 25.04.2012
L.C. : 02 gm Date of Calibration : 26.04.2012
I.D. No. : FWPL/WB-01 Suggested due Date : 26.04.2013
Sl. No. : 9110075 Page : 1 of 1
: :
Visual Inspection OK Calibration Method WI-28
:
Environmental Condition of Lab.- : Uncertainty of Measurement : ±5.8 g
Temperature : (20+/-1) Deg. C (At 95% Confidence Level,
Humidity (50+/-10)% RH k=1.96)

DETAILS OF STANDARD EQUIPMENTS USED FOR CALIBRATION


Name of the Instrument : STANDARD WEIGHTS : :
Make : WEIGH INDIA : :
Certificate No. : WMCL/M/2008/25 : :
Sl. No./ID No. : HTCS/SW/01 : :
: 18/11/2012 : :
Calibration Due Date
: WEIGHTRONICS : :
Traceable By

Calibration Results
Sl. Standard Observed Value Sl. Standard Observed Value
No. Value (UUC) No. Value (UUC)
01 0.10 Kg 0.100 Kg 08 15.00 Kg 14.998 Kg
02 0.20 Kg 0.200 Kg 09 20.00 Kg 19.997 kg
03 0.50 Kg 0.500 Kg 10 30.00 Kg 29.997 kg
04 1.00 Kg 1.000 Kg
05 2.00 Kg 2.000 Kg
06 5.00 Kg 4.999 Kg
07 10.00 Kg 9.998 Kg

Remarks:-

Notes :-
1. This calibration results reported in this certificate are valid at the time of & under the stated conditions of measurements.
2. This report should not be reproduced except in full without our prior permission in writing.
3. This certificate relates only to the item calibrated.
4. Format No.-F01(5.10/01).

Calibrated By Approved By

_____________________________________________________________________________________
B-19, Saraswati Enclave, Gali No.4, Khandsa Road, (Near Kadipur Chowk) Gurgoan
Mob. : 9810842522, 9999518005, E-mail : hitech.calibration@yahoo.com