SERVICE MANUAL
Cascada RO Lab Water Systems
100 to 240 VAC
50 to 60Hz RO-water
Model
INTRODUCTION
This service manual has been prepared for the Cascada™
RO-water system originally introduced in 2005.
INTRODUCTION 2
TABLE OF CONTENTS
SAFETY INSTRUCTIONS 4
Electrical 4
Pressure 4
Ultraviolet Light 5
Sanitization Chemicals 5
Control of Substances Hazardous to Health (COSHH) 5
PAL-CT1 Sanitization Tablet - Safety Information 5
ROUTINE SERVICE 6
TROUBLESHOOTING 7
ELECTRICAL CONNECTIONS 8
CALIBRATION PROCEDURE 9
CASCADA™ RO-WATER 11
SYSTEM PARTS LIST
Cascada RO-water System With Front Door 13
and All Covers Removed
Electrical
It is essential that the power switch on the Cascada Lab
Water system is turned off before any items are changed Main Power Supply
or maintenance work performed.
The ON/OFF switch is located at the left-hand side of Fuses
the unit. The main power cord is located just behind the On/Off Power switch
Power socket
ON/OFF switch.
WARNING! THIS APPLIANCE MUST BE
GROUNDED. PLUG INTO A PROPERLY Power cord
GROUNDED (GFCI) WALL OUTLET.
Pressure
The main water supply valve should be turned off and
residual pressure released prior to removal of any cartridges
or carrying out work on the unit.
CAUTION! Switching off the electrical power will
isolate the source of pressure, but
pressure trapped within the unit should
be released.
SAFETY INSTRUCTIONS 4
Ultraviolet Light
WARNING! UNDER NO CIRCUMSTANCES
SHOULD THE LAMP BE CONNECTED Cascada with UV Lamp
AND ACTIVATED WHEN OUTSIDE
THE HOUSING.
The Cascada™ RO-water system is fitted with an ultraviolet
lamp. The UV lamp is enclosed in a stainless steel chamber
ensuring that the operator will not be exposed to UV light.
UV lamp
Sanitization Chemicals
During the sanitization cycle, a PAL-CT1 sanitization tablet
is used and relevant safety information is included in this
handbook. A safety data sheet conforming to COSHH
and/or OSHA regulations is also provided with the tablets
and should be read before a tablet is used.
ROUTINE SERVICE 6
TROUBLESHOOTING
The table below gives guidance on what action to take should any problems arise. This section should be used
in conjunction with the Cascada™ RO-water System Operation Manual.
www.pall.com/lab 7 TROUBLESHOOTING
ELECTRICAL CONNECTIONS
For electrical connections, please see Wiring Diagram near
end of this manual.
Level control assembly circuit – fitted to all Pall reservoirs as
supplied.
R1
S1 220Ω
N/C 0.25W
R2
S2 220Ω
N/C 0.25W
1
R3 2
220Ω
Phone
0.25W
jack 3.5 mm
S3 connects to
N/C Cascada
R4
220Ω RO-water
0.25W system
1
2
3
ELECTRICAL CONNECTIONS 8
CALIBRATION PROCEDURE
To carry out a calibration of the Cascada™ RO-water 4. Set SUB-03 box to 5 and wait for a steady displayed
system, switch off main power. Remove electrical cover by value and PF1 icon to appear as follows: → ϖ ← .
removing single screw and sliding cover forward to 5. Press PF1 button and the icon fades for approximately
disengage securing clips. Locate electrical cover in the slot two seconds and then reappears indicating the
to the left of the electrical compartment. calibration setting is accepted. The display will change
80.0
to the exact value: HHHH .
TH10 TH9
6. Press PF2 button and the display changes to
"calibration 80 µS/cm".
TH12 TH10 TH9 TH8 7. Set SUB-03 box to 2 and wait for a steady displayed
::
:: value → ϖ ← .
JH1 Install
:: calibration 8. Press PF1 button and the icon fades approximately
Graphics
:: link
display for two seconds and then reappears indicating the
PAL-PCBS37987 calibration setting is accepted. The display will
80.0
change to the exact value: HHHH .
JH3
Touch
panel
TH TH TH TH TH TH TH
7 6 5 4 3 2 1
TH11
. 80 µS/cm
→ϖ← 80.2
Proceed as follows: HHHH
1. Install calibration link and connect SUB-07 setting
box to TH10 and SUB-03 setting box to TH9. PF1 PF2
Calculated quality
reading before 24 µS/cm
calibration
→ϖ← 24.2
HHHH
8µS/cm
→ϖ← 08.2
HHHH
PF1 PF2
Indicates
hexadecimal
PF1 PF2 A/D reading
Icon changes to Then
→ϖ←
Icon changes to Then
→ϖ←
12. Press PF2 button and the display changes to 18. Press PF2 button to return to PROCESS OFF.
"calibration 1ºC". 19. Switch off main power. Remove calibration link, SUB
13. Set SUB-07 box to position 1 and wait for a steady 03 and 07 setting boxes and reconnect transducers.
displayed value → ϖ ← . 20. Refit electrical cover taking care that the securing
14. Press PF1 button and the icon fades for approximately clips are properly engaged at the rear before replacing
two seconds and then reappears indicating the screw removed at the start of this procedure.
calibration setting is accepted and stored. The display 21. Switch on main power and revert to normal
changes to the exact value. operation.
1 ºC
→ϖ← 01.2
HHHH
PF1 PF2
35 ºC
→ϖ← 35.2
HHHH
PF1 PF2
CALIBRATION PROCEDURE 10
CASCADA™ RO-WATER SYSTEM PARTS LIST
Part Number Description Quantity Part Number Description Quantity
ADAP32798 Adaptor straight 8 mm 1 PAL-PLAT37941 Plate – Door 1
OD tube x 1/4" BSP
PINS36834 Pin dowel 4 mm dia x 1
ADAP35438 Adaptor straight 6 mm 1 14 mm (bottom door
OD tube x 1/4" BSP hinge)
CABL36864 Ribbon connector 1 POWE34241 Power inlet module 1
microprocessor to display 240V 6A
CONN32791 Connector bulkhead 3 POWE35964 65W switch mode 1
8 mm OD power supply (SMPS)
COVR36773 Cover rear 1 PUMP36858 Diaphragm pump 1
(boost pump)
COVR38020 Cover dust – EMI/RFI 1
shielded REDU35441 Reducer 8 mm OD 5
stem x 6 mm OD
DISP36777 Display LCD 1
RING36843 ‘O’ ring 1
ELBW30903 Elbow stem 5/16" OD 1
RIVT35962 Rivet snap 5 mm dia x 2
ELBW34263 Elbow swivel 6 mm 2 12 mm long
OD x 1/8" BSP
RIVT36352 Rivet snap 4 mm dia x 4
ELBW35604 Elbow stem 6 mm 2 11 mm long
OD x 6 mm OD tube
SP713 Thermistor assembly 1
ELBW35957 Elbow stem 3/8" 2
OD x 5/16" OD tube SP714 Line cell 380 mm
with plug 1
ELBW36122 Elbow swivel 8 mm 3
OD x 1/4" BSP SPRI36849 Spring 3.7 mm dia. x 1
9.5 mm (bottom door pin)
ELBW36616 Elbow swivel 6 mm 2
OD x 1/4" BSP (para) STRA20564 Strap cable small black 6
FANS36899 Fan ball bearing 24 volt 1 STRA30673 Strap cable releasable 2
black
FOOT36782 Foot rubber, self adhesive 4
SUPP36850 Support vessel 2
FUSE36107 Fuse semi-delay
20 mm 3.15A 2 TEES31702 Tee equal 6 mm OD tube 2
HOUS35913 Line cell housing 1 TEES36857 Tee swivel 6 mm OD 1
tube x 1/4" BSP off-set
HOUS36787 Housing disinfection port 1
TUBE31957 Tube hipure 8 mm OD 0.3 mtr
PAL-COVR37937 Cover front 1 x 5 mm ID
PAL-COVR37942 Cover top 1 TUBE35949 Tube 6 mm OD LDPE 2.0 mtr
PAL-DOOR37936 Door front 1 natural
PARTS LIST 12
Cascada™ RO-water System With Front Door and All Covers Removed
PAL-LC143
ELBW34263 x 3
SUPP36850
REDU35441
PUMP36858
ELBW35957
REDU35441
REDU35441
VALV36853 CONN32791
ELBW36616
ELBW36122
TUBE35949
VALV36905
CONN32791 ADAP32798
ELBW36616
FILT36123
BUSH36141 PAL-LC140
COUP31006 ELBW36616
ELBW36122 x 2 ADAP35438
WASH10534
SP714
CAPS36763 HOUS35913
HOLD36786
RING36843
TEES36857
HOUS36787
ELBW34263 x 2
SP713
Sanitization port
Test links:
Top pair = calibration
DISP6777 TH10 TH12 TH8 TH9 2nd pair =
1
JH1
SYSTEM WIRING DIAGRAM
Graphics Processor
1 PAL-PCBS37987
JH3
Touch panel TB1 alarm output connections:
6 1 = common
TO 2 = common
PAL-PANL37984-03 3 = normally open - closed when
3
unit alarms.
14
2 TB1
1
red
F2 black
TH7 TH6 TH5 TH4 TH3 TH2 TH1
1 8
TH11 F1 65 watt SMPS
POWE35964
blue
Fan N L
black brown Aurorinse sol. valve
red
Boost pump (optional) black brown 1 Power inlet
brown blue 2 module
Inlet sol. valve blue fused at 3.15A
Earth wire to be
soldered to SMPS
KEY TO CONTROL PANEL
Icons Alarm Conditions
Icon Description Icon Alarm Conditions
(Flashing)
(Flashing)
(Flashing)
(Flashing)
(Flashing)
(Flashing)
(Flashing)
(Flashing)
(Flashing)
Printer
Quality Alarms
Screen Description
Intermittent Recirculation
Outlet Purity Alarm
Replacement Date
Temperature Alarm
© 2005, Pall Corporation. Pall, , and Cascada are trademarks of Pall Corporation.