Operating Procedures For PMI
Operating Procedures For PMI
Introduction
The WAS AG Metal Analyser is used to quantitatively determine the concentrations of metal
compositions in a metal sample.
The Metal Analyser consists of a cabinet with excitation section, spectrometer, electronics section and a
computer, equipped with a printer.
The spark is a superficial discharge that remelts the surface ad vaporises a part of the sample. Because
of the high energy in the spark, plasma is built and this plasma is emitting light. This light is dismantled
by an optical grating in the optic and measured by CCD’s (Charge Coupled Devices, photosensitive
IC’s). Out of the light-data collected from the CCD’s, the internal computer calculates the analyse.
Inspection of PMI Shall be performed by the qualified person.
Scope
Apparatus
PMI-MASTER PLUS Metal Analyzer
Serial No. : 02I0013
Country of Origin : GERMANY
Maker : WAS Worldwide Analytical Systems AG
Type : PMI-Master Plus
Readout System : Jukeluke
Source : Arc / Spark
Main Voltage : 220V / 60Hz
Calibration : Fe-Base ; Cu-Base ; Al-Base and Ni-Base
Application : Fe-Base ; Cu-Base ; Al-Base and Ni-Base
Preparation
The surface of the materials for test should be clean, dry and free from any condition which may
affect results.
In order to avoid contamination during sample preparation, separate grinding disks are used for the
individual matrices.
Sample Marking
Material need to be PMI test shall be inspect for the identification. Samples are marked according to the
sample marking in the Project Specification FW Project Specification 3080-8440-00-SP0001 Rev. A2.
If the sample identification is not stated by the customer, the testing officer will mark the sample
according to MTS own reference number, to ensure there is no confusion in handling the sample
during the analysis.
.
Procedure
Before using the test machine for PMI, determine its calibration by using the standardized test blocks.
If measured values are out of the reference range, perform recalibration as per process.
Safety First
1. Always switch the source off when you are working at the probe head or the electrode to avoid
electric shock.
2 Always switch off the source when you are not working with the instrument to avoid accidents.
Measurement Results
The PMI test results shall be recorded and summated to customer.
Acceptance standard for this chemical analysis shall be based on material specification or Project
Specification FW Project Specification 3080-8440-00-SP0001 Rev. A2.
Customer shall be notified on rejected items.