APNA-370 Maintenance Manual e PDF

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 35

Converter efficiency check

After maintenance of APNA you have to check and adjust the converter efficiency:

• Change to Supervisor Modus 8/9 FACTORY CALIBRATION

• Introduce zero gas and calibrate zero point

• Introduce NO span gas and calibrate span point

• Introduce GPT this means NO is converted to NO2 by using ozone

• If the measured value is stable, adjust the number of the EFFICIENCY until the
NOx value is the same like before without GPT.

You might also like