TRANSCAL

You might also like

Download as docx, pdf, or txt
Download as docx, pdf, or txt
You are on page 1of 3

Error

T/C or
Sl. No. Description Make Model ID No. DUC Cert. No. Std. Cert. No. STD. Used Model Make ID. No. Set. Point Error Max.
CH No.
Allowed
INSTRUMENTS/THERMOCOUPLE USED TO CALIBRATE FURNACE
TSC/18-
TSC/19- Digital Multimeter 34461A Keysight MY57206241 100 -0.1 --
01 Handy Calibrator Yokogawa CA71 TSC/TCAL/81 19/INH/ECAL/165-1 ±0.40
20/INH/TCAL-81-1
CR/PCAL/50225 Multi-Product Calibrator 5520A Fluke 5520A 500 -0.2 ±2.00 --
200.012 0.288 ±0.80
54A
500.015 0.485 ±2.00
200.012 0.388 ±0.80
54B
500.015 0.385 ±2.00
200.012 0.188 ±0.80
CC284019000001881F SPRT with Readout 5609/1524 Fluke 01161/2024133 54C
500.015 0.485 ±2.00
200.012 0.288 ±0.80
54D
500.015 0.285 ±2.00
200.012 0.188 ±0.80
TSC/19- 54E
500.015 0.485 ±2.00
02 Thermocouple -- N-type TSC/TCAL/54A to 54J 20/INH/TCAL/54A-
200.012 0.288 ±0.80
54J-1 54F
500.015 0.485 ±2.00
200.012 0.588 ±0.80
54G
500.015 0.585 ±2.00
200.012 0.488 ±0.80
CC284019000007242F S Type T/C with Readout 5650-20/1524 Fluke 9813/2024133 54H
500.015 0.485 ±2.00
200.012 0.188 ±0.80
54I
500.015 0.385 ±2.00
200.012 0.288 ±0.80
54J
500.015 0.485 ±2.00
100 0.100 ±2.00
CH-1
500 0.300 ±0.40
100 0.200 ±2.00
CH-2
500 0.200 ±0.40
100 0.100 ±2.00
CH-3
500 0.200 ±0.40
100 0.100 ±2.00
CH-4
500 0.200 ±0.40
100 0.100 ±2.00
CH-5
500 0.300 ±0.40
100 0.100 ±2.00
CH-6
500 0.200 ±0.40
100 0.100 ±2.00
CH-7
500 0.200 ±0.40
100 0.100 ±2.00
CH-8
500 0.100 ±0.40
100 0.200 ±2.00
CH-9
TSC/19- TSC/19-20/INH/ECAL- 500 0.100 0.40
03 Paperless Recorder Brainchild VR18 TSC/TCAL/58 Multi-Product Calibrator 5520A Fluke 9175011
20/INH/TCAL-58-1 35-1 100 0.100 ±2.00
CH-10
500 0.300 ±0.40
100 0.200 ±2.00
CH-11
500 0.200 ±0.40
100 0.200 ±2.00
CH-12
500 0.200 ±0.40
100 0.100 ±2.00
CH-13
500 0.200 ±0.40
100 0.100 ±2.00
CH-14
500 0.300 ±0.40
100 0.100 ±2.00
CH-15
500 0.100 ±0.40
100 0.100 ±2.00
CH-16
500 0.300 ±0.40
100 -0.100 ±2.00
CH-17
500 -0.200 ±0.40
100 -0.100 ±2.00
CH-18
500 -0.100 ±0.40
100 0.200 ±2.00
CH-1
500 0.300 ±0.40
100 0.200 ±2.00
TSC/18- CH-2
04 Temperature Recorder Yokogawa DX 106-1-2 TSC/TCAL/49 CR/PCAL/50225 Multi-Product Calibrator 5520A Fluke 9175011 500 0.100 ±0.40
19/INH/TCAL-49-1
100 -0.100 ±2.00
CH-3
500 -0.200 ±0.40
100 0.100 ±2.00 CH-4
500 0.100 ±0.40
100 -0.100 ±2.00
CH-5
500 -0.100 ±0.40
100 0.100 ±2.00
CH-6
500 0.100 ±0.40
100 0.100 ±2.00
CH-7
500 0.200 ±0.40
100 0.100 ±2.00
CH-8
500 0.100 ±0.40
100 0.200 ±2.00
CH-9
500 0.200 ±0.40
100 0.200 ±2.00
CH-10
500 0.100 ±0.40
100 0.100 ±2.00
CH-11
500 0.200 ±0.40
100 0.200 ±2.00
CH-12
500 0.200 ±0.40

Note: Accuracy of the instruments / thermocouples used to calibrate furnace is within acceptable limits as per Table-1 of AMS2750E
INSTRUMENTS/THERMOCOUPLS USED TO CALIBRATE CALIBRATION INSTRUMENTS/THERMOCOUPLE
Error
T/C or
Sl. No. Description Make Model ID No. DUC Cert. No. Std. Cert. No. STD. Used Model Make ID. No. Set. Point Error Max.
CH No.
Allowed
TSC/18-
01 Digital Multimeter Keysight 34461A MY57206241 19/NH/ECAL-165-1 TSC/18-19/INH
(ETDC)
Multi-Product CR/PCAL/50225
02 Fluke 5520A 5520A 100 100.02
Calibrator (ETDC)
500 500.03
03 SPRT with Readout Fluke 5609/1524 01161/2024133
S Type T/C with
04 Fluke 5650-20/1524 9813/2024133
Readout
Multi-Product
05 Fluke 5520A 9175011
Calibrator
Instruments / Thermocouples Instruments / Thermocouples Certificate No. of Instruments /
Sl. No. Remark
used to calibrate Furnace (DUC) used to calibrate DUC Thermocouples used to calibrate DUC
Digital Multimeter TSC/18-19/INH/ECAL/165-1
01 Handy Calibrator
Multi-Product Calibrator CR/PCAL/50225
SPRT with Readout CC284019000001881F
02 Thermocouple - 54A-54J-1
CC284019000007242F Please provide these certificates
03 Paperless Recorder Multi-Product Calibrator TSC/19-20/INH/ECAL-35-1
04 Temperature Recorder Multi-Product Calibrator CR/PCAL/50225

You might also like