LINETRONIC LTVB-474452-SA Low Temp Visco Bath
LINETRONIC LTVB-474452-SA Low Temp Visco Bath
LINETRONIC LTVB-474452-SA Low Temp Visco Bath
USER MANUAL
LINETRONIC TECHNOLOGIES SA
V i a On o ri o L o n g h i , 2
CH-6864 Arzo (CH)
Tel. +41/91 6300703 Fax +41/91 6300719
www.lin.tech.ch info@lin–tech.ch
LT/VB-47445-SA USER MANUAL N-0000-47445-000 Page 1
TABLE OF CONTENTS
MADE IN SWITZERLAND
LT/VB-47445-SA USER MANUAL N-0000-47445-000 Page 2
At the end of analysis and when the instrument is not performing analysis, the
cooling compressors are automatically put in stand-by mode, this will save power
and increase the compressor’s lifespan.
MADE IN SWITZERLAND
LT/VB-47445-SA USER MANUAL N-0000-47445-000 Page 3
Delivery minutes
This delivery minutes is issued from the CEE Directive No. 2006/42
annotated “Instrument Directives” and from the successive improvements
from the CEE Directive No. 2006/95 annotated “Low Voltage Directive”
and from the CEE Directive 2004/108 annotated “Electromagnetic
Compatibility” and following the EN 66.5 regarding the Safety Directive for
the Measuring and Control instrument for the use in laboratory.
MODEL LT/VB-47445/2-SA
IMPORTANT
Date : ______________
MADE IN SWITZERLAND
LT/VB-47445-SA USER MANUAL N-0000-47445-000 Page 4
Dear Customer,
Yours faithfully
Linetronic Technologies SA
MADE IN SWITZERLAND
LT/VB-47445-SA USER MANUAL N-0000-47445-000 Page 5
The technical particularity used in the Low Temperature range, is in the cooling mode
of the test room, the defrost system, the integrated light barriers, the panel pc with
managing software.
In the case of the Low Temperature, the gas directly cool the test chamber of approx.
30 litres. The important volume grant optimal bath stability on the chamber.
The unit at the power on will start only the first motor compressor, than after few
minutes the second motor will start. This procedure grant lower spikes generation and
Reduced power consumption.
MADE IN SWITZERLAND
LT/VB-47445-SA USER MANUAL N-0000-47445-000 Page 6
DIMENSIONS
With 650 mm
MADE IN SWITZERLAND
LT/VB-47445-SA USER MANUAL N-0000-47445-000 Page 7
CAUTION IMPORTANT
Read attentively and observe scrupulously the below general safety standards and take
note of the devices on the instrument in order to assure a safety use of this instrument.
REMAINS
During the use and the handling of the instrument, only a qualified staff must do the
intervention. Before to do the smallest type of handling, the 220/230V plug must be
removed. When the instrument is working, nobody non authorised can stay near the
instrument. The safety devices on the instruments must never be manipulated.
MADE IN SWITZERLAND
LT/VB-47445-SA USER MANUAL N-0000-47445-000 Page 8
THE INSTRUMENT DOESN'T CONSTITUTE A POTENTIAL DANGER FOR THE OPERATORS WHO
USE IT OR TO THOSE WHO WORK IN THE PROXIMITY ZONES; NOTWITHSTANDING THIS, WE
INVITE TO FOLLOW THE INDICATIONS BELOW.
MADE IN SWITZERLAND
LT/VB-47445-SA USER MANUAL N-0000-47445-000 Page 9
PREVENTIVE HANDLING
Each operation must be entrust to a trained staff, before to do the
intervention, it is necessary to remove the 230/220V plug from the
instrument or by switching OFF the main power switch.
The LT/VB-47445/2-SA doesn’t need particular handling. In order to keep it in
good conditions, it is periodically necessary to do the following controls :
CONTROL FREQUENCY
Cleaning of the aeration grille Monthly
Cleaning of the capillary support At Each use
Cleaning of the defrost windows from humidity At Each use
Control of the liquid level At Each use
Replacement of the liquid in case of water
Monthly
presence
The instruments that we manufacture, do not include dangerous components for the
nature, during the START-UP please observe the following prescriptions :
TO BE SURE THAT ANY OPERATION CONCERNING THIS INSTRUMENT WILL BE
ENTRUST TO SPECIALISED AND AUTHORISED COMPANIES.
CLAIM COPIES OF THE PRESCRIBED AUTHORISATION AND CONSERV THEM IN THE
RECORDS.
In compliance with art. 13 of the Legislative Decree dated 25 July 2005, no. 151
“Implementation of the Directives 2002/95/EC, 2002/96/EC and 2003/108/EC on the
restriction of the use of certain hazardous substances in electrical and electronic
equipment, as well as waste disposal”. The symbol of the crossed-out wheeled bin on
the equipment or its packaging indicates that the product must be disposed of
separately from other waste at the end of its operational life. The differentiated
collection of this equipment arrived at the end of life is organized and managed by the
manufacturer. The user that intends to trash this equipment will have to contact the
manufacturer and follow the procedure that it has taken to allow separate collection
of the apparatus arrived at the end of life. Appropriate differentiated collection for the
dismantled appliance being subsequently sent out for recycling, treatment and for
environmentally friendly disposal, contributes to the prevention of
possible negative effects on the environment and on human
health, and encourages re-employment and/or recycling of the
materials the appliance is
made of. Specific administrative sanctions provided for by current
regulations will be applied for illicit disposal of the product by the
user.
MADE IN SWITZERLAND
LT/VB-47445-SA USER MANUAL N-0000-47445-000 Page 10
The Linetronic Technologies instruments are controlled according our internal quality
control procedures. The components quality as the component assembly are
controlled and tested before the assembling of the instrument. The components
checks are done in function of their own specificity.
Once the instrument is assembled, it goes through a first test of general functionality
control, allowing to verify if every device works properly in accordance with its
expected functions.
Then the measurement devices are electronically calibrated according to the physical
parameters that they must quantify and to the measurement scales defined for each
application.
Finally the instrument is tested with a certified reference sample on the parameter
that we want to measure. These tests are done according our internal procedure. The
test results are consigned in the manual delivered with the instrument.
The instrument is referring to the normalized methods where, the product nature
which has to be tested as well as the procedure to follow, are clearly defined. The
certified reference samples used for the final tests respond to these characteristics.
Any discrepancy on the results found on other products, must be done only on the
basis of certified reference sample. We kindly ask the customer to send us this
sample with its certificate. Linetronic Technologies SA will immediately undertake an
act procedure for the control and, if necessary, the correction.
The instruments are tested in a laboratory where the ambient temperature can vary
from +15°C to +25°C. The power supply is 220 Volts ±5%, Hot, Neutral, Ground,
uninterrupted, without electrical noise and perturbation, equipped with ground fault
intensity devices (30 mA).
MADE IN SWITZERLAND
LT/VB-47445-SA USER MANUAL N-0000-47445-000 Page 11
INSTALLATION
right right
screw driver with insulated handle
RIGHT FALSE
MADE IN SWITZERLAND
LT/VB-47445-SA USER MANUAL N-0000-47445-000 Page 12
GUARANTEE
The warranty period is for 12 months for the mechanical parts and 6 months for
the electrical and electronic parts, at the condition that the use and the handling
for each instrument are respected. Guarantee period longer than the standard
must be accepted and confirmed by the manufacturer.
During this period, the assistance is provided free of charge upon our head office in
Arzo (Switzerland) or upon our local distributor / agent, (if any) present on
customer’s territory.During the warranty period the manufacturer will replace
and/or will repair, free of charges, all the components that will present a
manufacturing defect or malfunction. Spent the terms above, the warranty declines
and the assistance will be realised debiting the cost of the components, of the
technical intervention, of the transport of the materials and the bed and board and
travelling expenses if any. In case of assistance on customer’s site, the client will
admit the call cost and the travelling expenses. The intervention cost, during the
warranty period, are also subordinated to our availability and to the eventual spare
parts.
The manufacturer is not responsible for the possible damages caused to the device
due to wrong connections to the power supply.
In case of extraordinary maintenance, personally realised by the customer, during
the warranty period, the guarantee will immediately be suspended.
The manufacturer reserves himself the right to bring in any moment and without
warning: changes and innovations, respecting the ASTM norms, without nobody
can advance complaint.
The warranty is not valid in the case of damages caused by:
wrong use of the device
negligence of the operator
atmospheric and geophysical agents, vandalism, wartime events
transport
use of unoriginal spare parts.
IMPORTANT
MADE IN SWITZERLAND
LT/VB-47445-SA USER MANUAL N-0000-47445-000 Page 13
INTERVENTIONS LIST
In order to control the instrument handling, we advise to consign with details all
the interventions done on the LT/VB-47445/2-SA with S/N 0419-47445-659
P.S. The signature of the technician who did the intervention must be readable.
MADE IN SWITZERLAND
LT/VB-47445-SA USER MANUAL N-0000-47445-000 Page 14
PACKING
The complete instrument is properly packed by qualified and trained staff of Linetronic
Technologies SA with suitable packaging materials to ensure the highest safety
during handling and transport.
The packing is made following the more recent internal norms of the company. The
packing procedure are the following :
The instrument is standard packed with plastic film protection.
The glass cell are mounted on the analytical head with paper film protection
The packing is made on a carton box or in a suitable wooden box filled with anti-
shocks foaming.
Dedicated labels for identification, Up side, Fragile Handle With Care, Customer
Name and Address, Gross and Net Weight, etc,
Wooden Box are fumigated as for international health rules.
Additional plastic strapping if required
The shipping documentation may can be change following the contract’s point.
SHIPMENT
CAUTION
The forwarder must take all the necessary actions to assure the stability of his
shipment whatever is his speed, his accelerations, his direction changes and the
journey characteristics.
He must moreover respect every notes written on the packing.
RECEPTION
MADE IN SWITZERLAND
LT/VB-47445-SA USER MANUAL N-0000-47445-000 Page 15
SOFTWARE FEATURES
Features:
* Windows and Excel are registered trademark of Microsoft Corporation in the United States and other countries.
MADE IN SWITZERLAND
LT/VB-47445-SA USER MANUAL N-0000-47445-000 Page 16
1. TECHNICAL CHARACTERISTICS
General specifications
Specifications Low Temperature Bath
Testing span -75°C to +5°C
Measuring span -100°C to +50°C
Scope Viscosity analysis at low temperature
Measuring parameters
- Low temperature Viscosity as for D445, D2532 and or D2983 method
- Temperature in °C/°F
Cooling unit
- Internal refrigeration, double stage up to - 75°C or lower temperature
specification given at ambient temperature +22°C
- Equipped with over temperature cut-out at +5°C
Push buttons
- One main switch
MADE IN SWITZERLAND
LT/VB-47445-SA USER MANUAL N-0000-47445-000 Page 17
ROTATIONAL
VISCOMETER D2983
Head can be different
HEAD ASSEMBLY
support only forD2983
INSPECTION
WINDOWS
WITH
DEFROSTING
SYSTEM
BATH
MAIN
SWITCH DRAINING TAP
BATH PLUG
PANEL PC
USB PORTS
VENTILATION
GRILL
DRIVING WHEELS
MADE IN SWITZERLAND
LT/VB-47445-SA USER MANUAL N-0000-47445-000 Page 18
4. EQUIPMENT SUPPLIED
LT/VB-47445/2-SA , INCLUDES :
6. SOFTWARE INTERFACE
MADE IN SWITZERLAND
LT/VB-47445-SA USER MANUAL N-0000-47445-000 Page 19
Please referee to the test method ASTM D445 D2532 and or D2983 test procedure
TOUCH THIS
FIELD TO SET
THE
ANALISYS
PARAMETERS
TOUCH THIS
BUTTON TO
TOUCH THIS
ACCESS THE
BUTTON TO
DIAGNOSTIC
TURN ON THE
PANEL
BACKLIGHT
TOUCH THIS
BUTTON TO
START THE
ANALYSIS
MADE IN SWITZERLAND
LT/VB-47445-SA USER MANUAL N-0000-47445-000 Page 20
The Heat button allows to test the heaters inside the bath.
The Moto compressor button allows to test the motor compressors used to cooling down the liquid bath.
The Cold button allows to power ON the solenoid used to start the cooling procedure.
The Stirrer button allows to power ON the motors used to stir the liquid inside the bath.
The Windows’s Defrost button allows to power ON the defrost of the window.
The Clock button allows to change the actual time according your local time *.
The Calendar button allows to change the actual date according your local data *.
The Shutdown button let you power off the instrument in the correct way avoiding losing data**.
MADE IN SWITZERLAND
LT/VB-47445-SA USER MANUAL N-0000-47445-000 Page 21
6d. CALIBRATION
OFFSET
SETTING
The graph will show the sensor reading in real time in both Value and Volt, the calibration shape
represented by the black ascending line is the actual graphic representation of the calibration itself,
where the white circles are spotting the provided calibration points.
The calibration can be performed on each single point of the curve, meaning that each single degree of
the scale can be calibrated and fine tuned up to 100 individual calibration points.
By default, the five calibration points shown are the reference calibration points, they can be
tuned/recalibrated as needed , but unlike the user-defined calibration points we are about to describe,
they cannot be removed; the calibration of an existing reference point can be performed by both a
resistor box or by the mean of an external comparison thermometer, provided a common temperature
media between the comparison thermometer and the probe to calibrate in case of the bath temperature,
in case of the force, you can use an external dynamometer to calibrate the force cell.
There is two method to calibrate each sensor :
1. Advanced Calibration;
2. Calibration Wizard;
MADE IN SWITZERLAND
LT/VB-47445-SA USER MANUAL N-0000-47445-000 Page 22
As mentioned, the calibration can be standard ( five points) or dynamic (up to 100 points, dynamically
added); performed by mean of a resistor box or by a reference temperature, by using a resistor box it is
necessary to unplug the sample’s PT100 from the board and then connect the resistor box on the
revealed socket; to perform the calibration by mean of an reference thermometer, will be necessary to
provide a suitable temperature media as a beaker filled by a liquid, or poured liquid into the heating
bath (1) so that both probe to calibrate and reference thermometer are measuring the same
temperature media.
(1) A note of caution: by using as reference an external thermometer to calibrate the bath, ensure to
maintain the temperature at a stable point for an extended amount of time (3~5 minutes) to allow
the stabilization and balancing of the mass of the stove’s temperature.
MADE IN SWITZERLAND
LT/VB-47445-SA USER MANUAL N-0000-47445-000 Page 23
Once the desired calibration temperature has been reached then it is necessary to introduce it prior to
confirm the calibration point; in the following example we are introducing a new calibration spot, please
note that the operation is identical while updating an existing calibration point.
The new calibration point has been added by mean of the Add Calibration Point button, a new white spot
is appeared on the curve
and the buttons on the
bottom row are showing
the new -65 value, the 50
value is not currently
visible but will be shown
by touching the rightmost
button in the same row.
MADE IN SWITZERLAND
LT/VB-47445-SA USER MANUAL N-0000-47445-000 Page 24
As shown in the picture above, five calibration points were added; as mentioned, it is possible to
introduce up to 95 individual calibration spots in addition to the 5 reference points for a total of 100
calibration points.
Pay Attention :
The calibration of the sensor must be perform in Celsius degree scale.
MADE IN SWITZERLAND
LT/VB-47445-SA USER MANUAL N-0000-47445-000 Page 25
DATE SETTING
BUTTON
TIME SETTING
BUTTON
MADE IN SWITZERLAND
LT/VB-47445-SA USER MANUAL N-0000-47445-000 Page 26
In the diagnostic panel there is button to close the application in a safety mode, avoiding losses of data.
OFF
BUTTON
By clicking the red button, the software will ask if the whole instrument must be turned off.
By clicking on "Yes", the panel PC will start the procedure and, only when the screen will
all black it will be possible to turn off the analyser by acting on the main switch button placed on
the left bottom side on the front panel.
MADE IN SWITZERLAND
LT/VB-47445-SA USER MANUAL N-0000-47445-000 Page 27
7. ANALYSIS PARAMETERS
The same thing happens by clicking black field closed to FILE NAME .
The FILE NAME field is also used to create a report file of the analysis.
MADE IN SWITZERLAND
LT/VB-47445-SA USER MANUAL N-0000-47445-000 Page 28
ASTM D2983
The software allow you to choose the temperature kept inside the liquid bath during whole analysis.
By touching the Turn ON backlight after soak time button, the software will automatically turn on the
backlight after the soak time is expired.
A little resume of analysis appear at the bottom of the settings panel.
MADE IN SWITZERLAND
LT/VB-47445-SA USER MANUAL N-0000-47445-000 Page 29
Once the analysis has been setup as described in previous chapter, the START ANALYSIS button
present in the main panel can be pressed.
Automatically the software starts to regulate the bath at the highest chosen temperature between all the
analysis in progress.
The status each test place is showed under its name.
Now the button that previously was named START ANALYSIS becomes now STOP ANALYSIS.
Now the operator can load the sample and wait until the soak time will expire. Once the temperature is
reached, the soak time starts automatically.
NOTE:
The above pictures are referred to the ASTM D2983, 6 test positions as example. The same are present
on the ASTM D2532 5 test places position software application.
MADE IN SWITZERLAND
LT/VB-47445-SA USER MANUAL N-0000-47445-000 Page 30
When the soak time expired, a result panel will appear, furthermore, the backlight light will be enlighten
if its button has been selected during the analysis setup.
As long as the operator does not press the start button, the instrument will perform an acoustic alert to
inform the finish of soak time. In this panel will be displayed the most important information as : place,
temperature of the test, soak time, file name and operator.
By pressing start, the software will show a keyboard allowing the operator to introduce the result. If a
typing mistake occurs, please press again Start and modify the result. After pressing finish, the analysis
will be automatically stored in the result browser.
MADE IN SWITZERLAND
LT/VB-47445-SA USER MANUAL N-0000-47445-000 Page 31
9. RESULTS BROWSER
The result browser may be accessed even while the analysis is in progress.
On the main screen will be displayed all the analysis performed by the instrument with the more recent
at first.
The analysis can be searched by name or by data touching the relevant button.
MADE IN SWITZERLAND
LT/VB-47445-SA USER MANUAL N-0000-47445-000 Page 32
OIL-LAB 81 Calibration kit (set of connectors and cables for the calibration)
MADE IN SWITZERLAND
LT/VB-47445-SA USER MANUAL N-0000-47445-000 Page 33
START-UP
√ The analyser is powered up properly √
√ The analyser is calibrated √ Direct comparison with thermormeter (0.1°C)
√ The cryostat is working properly √ -75°C min. temp, reach. at ambient temp: 22°C
√ Standard Resolution: 0.01°C √
Advanced Resolution: 0.001°C
According the Start-up procedure and the tests above, the analyser is considered as 100 %
working according the LIN–TECH SPECIFICATIONS.
MADE IN SWITZERLAND