HPLC Pump L-7100

High Precision
Solvent Delivery
The LaChrom L-7100 HPLC pump provides the highest possible
degree of precision and reliability for high quality analytical results.
Corresponding to the requirements of modern quality assurance, it is
equipped with intelligent
function control and in-
corporates a documen-
tation facility. Versatile
in its application, it has
become a universal pump
module in the HPLC
laboratory.
Apart from its control
by the D-7000 HPLC
System Manager, the
L-7100 can also function
as a stand-alone system
controller.
LaChrom2000

Precision and Reliability the manufacturer's test report docu- ■ High-pressure gradient pump in
ments that all functions and specifi- conjunction with one or two
The L-7100 has been designed along cations have been fulfilled. further L-7100 pumps and as
the lines of the proven serial dual- ■ Column packing device at con-
piston principle, characterised by an stant pressure.
extremely constant flow rate, low Versatile in Application The L-7100 pump can store
pulsation and reliable delivery of all ■ 9 time programmes, in which
commonly used solvents. Moreover, User-friendly programming via key- mobile phase composition, flow
the L-7100 is able to form solvent board and LCD dialogue enable the rate and four events can be pro-
gradients with high precision and L-7100 to be used as grammed and varied over a period
accuracy. Each individual component ■ Isocratic HPLC pump of time.

and the finished instrument are ■ Quaternary low-pressure gradient


subject to stringent quality control pump in conjunction with a gra-
measures. A validation certificate and dient accessory
HPLC Pump L-7100
■ Via a Link function, programmes ■ All parameters, programmes and Maintenance is facilitated by the
for multi-method applications gradient curves of the L-7100 ■ Maintenance logbook which auto-
can be linked. pump can be printed out by the matically registers the total volume
LaChrom data systems. delivered. This provides an indica-
A shut-down routine can be pro- ■ The pressure signal of the tion as to when the next replace-
grammed to be carried out at the L-7100 pump is constantly moni- ment of the pump seal or the next
end of an analysis series by tored as part of the reliability cleaning of the pump valves
■ Delayed pump stop and a check. Any deviations from the should take place.
■ System flush programme. normal pressure profile, caused
for example by air bubbles or
Due to minimised internal volume, a malfunction of pump valves, are Controller Function
highly constant flow-rate and excel- registered during the course of
lent gradient reproducibility at low the analysis. In stand-alone operation, the system
flow-rates, the L-7100 pump is ■ The Pump pressure profile can be control function of the
■ Optimally suited for narrow-bore printed out in the analysis report L-7100 enables external instruments
applications. and hence serve in the assessment to be controlled by providing
of the quality of the analytical ■ Four event and other control
For applications involving sensitive results. signals.
materials requiring operation at low
temperatures
■ The cold room compatibility of
the L-7100 is advantageous. Technical Data
Premature wear caused by salt- Delivery system Series configured double piston pump with electronic compensation of
containing eluents is prevented by solvent compressibility and elimination of residual pulsation
the fitting of a
■ Piston cleaning facility, whereby
Maximum pressure 400 bar
the piston wash procedure can be Flow rate 0.001 – 9.999 ml/min
carried out automatically in Flow stability ≤ 0,2% RSD
combination with the LaChrom
autosamplers. Gradient modes Quaternary low-pressure gradient, binary or ternary high-pressure gradient
Special functions Programme link, column pack programme, delayed pump stop,
system flush programme
GLP Functions GLP functions Pump logbook, display of pressure profile deviations, key-board lock
In view of the requirements of qua- Documentation Print-out of pump parameters, gradient programmes, gradient curves,
lity assurance systems, all modules
pump logbook via LaChrom data systems
of the LaChrom system have been Outputs 4 time-progammable event contacts, analogue pressure signal,
fitted with comprehensive facilities busy/ready, D-line
for status display, reliability monito- Inputs Start, stop, pump on, D-line
ring and the provision of correspon-
Materials Stainless steel, ruby, sapphire, zirconium oxide-ceramic, PTFE, carbon
ding documentation and traceability.

Ordering Information
Cat.No. Designation
1.18880.0001 HPLC Pump Model L-7100
LaChrom2000

1.18881.0001 Quaternary low-pressure gradient accessory


1.18882.0001 Binary high-pressure gradient accessory
1.18886.0001 Ternary high-pressure gradient accessory
1.18885.0001 Dynamic mixer (for fitting to the L-7100 pump)
1.18982.0001 Membrane degasser (4 channels) Model L-7612
1.18115.0001 Helium degasser (3 channels)
1.18980.0001 Solvent organiser
1.18887.0001 Fixing parts for helium degasser
For more information please contact
1.18883.0001 Plunger washing mechanism
Merck KGaA
64271 Darmstadt / Germany 1.18888.0001 Column holder
Telefax: +49 6151 - 72 74 95 Technical data may be altered at any time without prior notice.
LaChrom® is a registered trademark of Merck KGaA, Darmstadt, Germany.
or your local Merck company.
W 219161
http://www.merck.de/chromatography LaChrom 2000: A development of Merck KGaA and Hitachi Gruppe X ‘99