Professional Documents
Culture Documents
Analytical Target Profile
Analytical Target Profile
Analytical Target Profile
net/publication/282914684
CITATIONS READS
3 1,472
1 author:
Ron Snee
Snee Associates, LLC
403 PUBLICATIONS 6,151 CITATIONS
SEE PROFILE
Some of the authors of this publication are also working on these related projects:
Strategies of Formulations Development: A Step by Step Guide Using JMP View project
All content following this page was uploaded by Ron Snee on 19 November 2015.
The Use of the Analytical Target Profile in the Lifecycle of an Analytical Procedure
with an Example of for an HPLC Procedure
Jane Weitzel, Robert A. Forbes, and Ronald D. Snee
Introduction ufacturing processes and analytical quality by alytical procedure as described in the USP [9 actual results, are provided mainly for illustra-
The lifecycle management of an analytical design (aQbD) is that it will allow the applica- –11] and the ICH [12], will be discussed and il- tive purposes.
procedure applying Quality by Design (QbD) tion of the experience, knowledge, tools, and lustrated in the example. The terminology used In this example, we will focus on the preci-
concepts and using the analytical target profile statistics from QbD to be applied to aQbD. with both approaches will be compared and sion and uncertainty performance characteris-
(ATP) has been introduced by the USP stimu- Thus, all parties, laboratories, companies, Qual- explained. tics of the potency test method to demonstrate
li article “Lifecycle Management of Analytical ity Assurance, production, as well as regulatory The aQbD approach is illustrated using an how the gage R&R tool and experimentally de-
Procedures: Method Development, Procedure bodies can leverage their experience with QbD example of the potency test for a drug sub- signed intermediate precision studies are used
Performance Qualification, and Procedure Per- in developing an approach to aQbD. stance (DS) with specification limits of 98.0% for Procedure Performance Qualification and
formance Verification [1].” Current regulatory A key component of the aQbD paradigm to 102.0%. The potency test is one component Verification. The acceptance criterion for the
guidance around process validation is moving is the definition of the Analytical Target Pro- in the DS control strategy utilized to ensure the precision of the potency method will be set us-
away from a simple once-and-done validation, file (ATP) [1], analogous to the Quality Target safety and efficacy of a drug product (DP) man- ing the Lifecycle approach where the intended
and toward a continuous process verification Product Profile (QTPP) in the ICH [8]. The ufactured from the DS. The test ensures that the use of the method is defined using a decision
approach, also called a “lifecycle approach” [2]. ATP defines the requirements for the result of DS is of adequate potency, so that when used in rule. The decision rule defines the acceptable
The recent FDA method validation draft guid- a test method and is based on the suitability the DP, the patient will receive the appropriate level of probability of making an incorrect de-
ance also includes some discussion of a lifecycle for use of that result. This paper illustrates how dose of the active ingredient. In this way, poten- cision. For this discussion, the null hypothesis
approach for analytical methods [3]. the ATP is developed and applied to assess the cy is a critical quality attribute (CQA) of the DS is “The potencyof the sample is within specifi-
A number of articles have been written in- performance of the method at various stages of that is linked to a CQA for the DP [13]. For sim- cation,” so there are two different types of incor-
troducing the lifecycle management of analyti- its lifecycle. The use of traditional Gage R&R plicity, we will assume the limits are on the as-is rect decisions. First, a batch that has acceptable
cal procedures [4 – 7]. These articles introduced statistical analysis and control sample charting, basis, and no correction for water or solvents potency may test outside the specification limits.
the concept that analytical procedures are pro- which are commonly used in manufacturing is necessary. Obviously, there are other speci- Second, a batch with a potency truly outside of
cesses and, as such, the tools, approaches and process validation and improvement, will also fication tests (e.g. impurities, residual solvents, the specifications limits, could test within limits
statistical analyses for manufacturing process- be illustrated in the development of an analyt- etc.) that are required to ensure the DS meets its and be accepted. The first error can be consid-
es can be applied to analytical procedures. The ical procedure. The similarities, contrasts, and CQAs, but these will not be considered in this ered a false positive (positive test for unaccept-
benefit of recognizing and acknowledging the differences between the lifecycle approach, and illustration. It should also be noted that the DS able potency), or a type I error. The second er-
similarity between quality by design for man- the traditional approach to validation of an an- potency results used for this illustration, while rror would be a false negative (negative test for
www.ivtnetwork.com
Jane Weitzel, Robert A. Forbes, Ronald D. Snee
unacceptable potency) or a type II error. The the laboratory sample, continues with the treat-
consequences of each of these types of errors ment of that sample which may involve taking
are taken into consideration when developing a test-portion, comminution, extraction, disso-
the decision rule. The ATP uses the probability lution, etc. and concludes with the output, the
of each of these errors to define the maximum reportable result as shown in Figure 2.
allowed measurement variation (uncertainty)
for the reported result, which translates into an The FDA definition for process validation is
Figure 3 QbD for analytical procedures is analogous to that terms for the major steps in the QbD approach.
for production processes. This figure shows the analogous
side the acceptance zone comprise the rejection The consequences for a type I error will de-
zone. This simple decision rule is illustrated in termine what probability is acceptable. Since
Figure 4. there is little flexibility to increase the accep-
This simple decision rule can be stated as: tance zone, the measurement uncertainty is the
only parameter that can be adjusted in order to
The product will be considered compliant if the meet the acceptable error rate.
measurement result is within the acceptance
zone. Measurement Uncertainty
Measurement uncertainty is defined in the in-
With this decision rule, the probability for ternational vocabulary of metrology (VIM) as a
either type of error is a function of both the pro- “non-negative parameter characterizing the dis-
Figure 5 shows the normal distribution in relation to the specifications of 98.0% to 102.0%. The width of the normal distribu-
cess variability and the measurement uncertain- persion of the quantity values being attributed
tion is chosen so that the probability the true result is below98.0% is 2.5% and above 102.0% is 2.5%.
www.ivtnetwork.com
Jane Weitzel, Robert A. Forbes, Ronald D. Snee
a normal distribution over the decision rule as Target Profile. Target measurement uncertainty method precision to avoid failing to meet the able batch will fall outside this range must be
shown in Figure 5. In this example, the true drug (TMU) is defined in the VIM as the “measure- TMU. Larger studies with more runs are more NMT 5%.
substance potency is assumed to be 100.0%, ment uncertainty specified as an upper limit costly in terms of time and resources, so one
which establishes the center of the normal dis- and decided on the basis of the intended use of must weigh the benefit of greater confidence in The analytical target profile can now be de-
tribution. The width of the normal distribution the measurement results” [19]. Any analytical the precision estimate, versus the additional cost termined. The laboratory qualifies the analytical
is determined by the standard deviation which procedure that yields a reportable result with a of larger studies. procedure for a concentration range wider than
was set to 1% to coincide with a 2.5% probabili- standard uncertainty of 1.0% or less, given there the specification to allow for variability in sam-
ty of obtaining values each above and below the is no bias, will be fit for its intended use. Relationship Between Accuracy and Target Mea- ple concentration. The matrix and measurand
limits. As shown in Figure 5, 5% of the potency The probability associated with the decision surement Uncertainty are defined clearly. There is only one known
results for the sample will fall outside the limits rule is based on the theoretical normal distri- The probability of not meeting the specifica- impurity, Impurity A, and it occurs below 0.1%.
due to measurement variation (type I error rate) bution and the specified standard deviation. tion depends on where the true distribution is The ATP can now be written:
for the assumed drug substance potency and It does not take into account the variability in centered. If the true distribution is not centered
precision described above. This distribution the estimates of the uncertainty that will be at 100.0%, as in the above example, then a dif- The analytical procedure must be able to quan-
only captures the measurement uncertainty and obtained in the qualification and verification ferent proportion of the measured results are tify the drug substance in the presence of impu-
does not include process variation (variation in experiments. That is dealt with in the experi- expected to be outside the specifications. When rity A, and potential degradation products, over
the true batch to batch potency), which was as- mental designs of the development and qualifi- this occurs, the target measurement uncertain- a range of 80% to 120% of the nominal concen-
sumed to be negligible. cation experiments, and the desired confidence ty must be improved in order to maintain the tration with an accuracy and uncertainty so that
If a 5% type I error rate is acceptable, the in their resulting precision estimates. Generally, same type I error rate. However, the analytical the reportable result falls within ±2% of the true
standard deviation of 1.0% becomes the target higher confidence limits will require more ex- procedure could have a bias that cannot be value with at least a 95% probability.
measurement uncertainty for the Analytical perimental runs to provide better estimates of removed in method development. If it is not
corrected for, then the TMU will have to be ad- Performance Characteristics for the
justed. Or if the drug substance does not have a Analytical Procedure
potency of 100.0%, this will impact the degree The ATP can now be used to determine the
of precision that the method will need to attain. performance characteristics for the analytical
For example, if the DS has an actual potency procedure. The following performance char-
value of 99.0%, then the TMU must be reduced acteristics need to be included in the method
to 0.6% (Figure 6). qualification: accuracy, precision, specificity,
linearity, and range.
Analytical Target Profile The acceptance criteria for specificity (i.e.,
Based on a risk analysis that includes consider- resolution of the main peak from impurity A
ations such as the cost of OOS investigations, and potential degradation products) and linear-
the known characteristics of the product (the ity are determined based on their impact on the
drug substance is known to have few impuri- accuracy and precision. As long as their impact
ties and the potency is close to 100%), and the on the accuracy and precision is such that the
capability of the analytical technique (HPLC) a target measurement uncertainty can be met,
company can decide the probability it is willing they are acceptable. This is a key advantage of
to accept associated with making an incorrect this approach; it defines the performance that is
decision (type I error) concerning a product “good enough.” One can confidently accept the
specification. In the below example, the proba- performance because it is directly linked to the
bility selected is 5%. intended use.
The decision rule can be more detailed now: The performance characteristics and the ac-
ceptance criteria are summarized as follows:
The batch of drug substance will be considered
Figure 6 (a) shows that if the normal distribution is centered over 99.0% the probability of making a wrong decision is in-
creased to 16.3%. (b) shows the target measurement uncertainty must be reduced to 0.61% in order that the probability of compliant if the test result is within the range of The target uncertainty is NMT 1.0%. The com-
making a wrong decision remains at <5%. 98.0 to 102.0%. The probability that an accept- bination of bias and uncertainty need to be such
www.ivtnetwork.com
Jane Weitzel, Robert A. Forbes, Ronald D. Snee
and to support process improvement efforts. Samples from five API batches were tested on
The variance component estimates are of- four setups by two analysts using each of two
ten used to determine how many within-run instruments (each instrument used a unique
replicates and/or setups are required to give the column for total to two columns). New mobile
desired method intermediate precision. If the phase was prepared for each setup and each
within-run variability (repeatability) is high, analyst prepared a fresh set of standards for
then more within-run replicates maybe re- each run. Samples were prepared and analyzed
quired to generate a reportable result with the in duplicate on each run, producing a total of
desired precision level. If setup-to-setup vari- 5x2x2x2 = 40 Potency(%) test results. The data
ability (reproducibility) is high, then multiple are summarized in Table 2.
setups including one or more analysts, instru- The results of the variance components
ments, and columns maybe required to gener- analysis for the potency measurements are
ate a reportable result with the desired preci- summarized in Table 3 and shown in Figure
sion level. The reportable result consists of the 7. In this study, multiple analysts, instruments
average of multiple within run and/or setups as and columns were used to obtain more realistic
prescribed in the method and is the value that is variability estimates, and could not be separate- Figure 7. Variability Chart showing the results of the Gage R&R study to verify performance of the method in a new labo-
compared to the specification and applied to the ly evaluated for statistically significant effects. ratory.
decision rule. This kind of a replication strategy The variance components show that 84% of
can be employed to reduce the variability test variability is due to the setup and the remaining
method that currently has a higher than desired 16% due to repeatability (within-setup).
level of variability to continue to release product The standard deviation of the test method
until a better test method can be developed. In (intermediate precision for two preparations on
this way production doesn’t have to be stopped a single setup) is 0.71% with an upper 80% con- Table 3. Gage R&R Study Repeatability and Reproducibility Variance Components
(or not initiated) because of a highly variable fidence limit of1.16%. The intermediate preci- proach incorporating a confidence limit is used. ber of setups and within setup replicates.
test method. sion estimate of 0.71% meets the acceptance The method may not be good enough for its Table 4 demonstrates that increasing the
To illustrate the utility of this approach for criterion of NMT 1.0% that was defined based intended use. To address this, additional set- number of within setup measurements (re-
verification of the method transfer to the new upon the ATP. However, the upper confidence ups could be performed to obtain a better SD peatability) from one to two has little effect on
laboratory, a Gage R&R study was performed. limit exceeds 1.0%, indicating that we cannot estimate with a tighter CI. For example, if the the method intermediate precision. This is be-
say with 80% confi- results for all the batches are pooled, with the cause the within setup (repeatability) variance
dence that the standard assumption that there are no significant differ- is much smaller than the setup-to-setup vari-
deviation is not NMT ences between them, a standard deviation of ance (reproducibility). Generally, the standard
1.0%. If the verification 0.63% with an 80% UCL of 1.03% (rounds to practice for a DS potency method is to prepare
protocol utilized an 1.0%) is obtained, which meets the acceptance samples in duplicate. While the decrease in
acceptance criterion of criterion with an 80% confidence level. variation for this method is marginal, the major
NMT 1.0% for the 80% If the method still did not meet the accep- benefit of the two preparations is to control for
UCL of the SD, these tance criterion with additional runs, the test preparation errors by setting an acceptance cri-
results would not pass method variation can be reduced by employing terion on their difference. Since reproducibility
the criterion. This study a replication strategy according to Equation 1. was the larger component of variance, analyzing
provides an interest- Table 4 shows the effect of increasing the num- the samples on multiple setups would be more
ing example where the effective at reducing the method
method would pass a standard deviation, than preparing
traditional assessment, more samples for the same setup.
but does not pass when By including multiple setups (n=2)
a more rigorous ap- the method intermediate precision
Table 2. Gage R&R Study Results for DS Potency Method Transfer. could be reduced to approximately
www.ivtnetwork.com
Jane Weitzel, Robert A. Forbes, Ronald D. Snee
Recommendations for Interpreting the observed, it may be appropriate to consider
Control Charts performing a robustness evaluation of the test
When all the results are within the control limits method as discussed above. A test method
and no OOT patterns are observed the meth- that is not robust can be expected to exhibit in-
od is performing as expected. However, over creased repeatability, intermediate precision, or
the course of time OOC and/or OOT may be both over time.
observed. The course of action will depend on
the specific situation but some general guidance Product Quality Review – Summarize
can be given. The following recommendations Test Method Performance
are similar to those regarding process monitor- The performance of the test method should be
ing when using control charts. reviewed periodically in accordance with the
First when out-of-control (OOC) or out-of- reviews for the process and product. As part of
trend (OOT) signals are obtained the data that review, the results of the test method con-
Figure 8. I-MR Control Charts for Sample Averages (n=2): All Samples collection, analysis, and control charting proce- tinued process verification system would be
dures are examined to ensure that no mistakes provided. The control charts of the test results
have occurred. If mistakes are found the data for the control samples and summary statistics
should be corrected and the analysis repeated. which document the test method performance
When mistakes are ruled out then an inves- including the repeatability and intermediate
tigation should be considered to find the cause precision standard deviations and relative stan-
of the OOC or OOT signals. The investigation dard deviations, results of recent calibration
will likely examine if that signal can be traced to studies and other information depending on
a specific change in materials, reagents, instru- the test method and situation should be includ-
ments, columns, reference standards, or ana- ed in the summary. The performance summary
lysts. The action taken will then depend on the documentation should be stored and dissemi-
cause and the specific situation. nated in accordance with the knowledge man-
However, sometimes the cause of the OOC agement expectations of the ICH [25].
or OOT cannot be determined. If this is the case,
the practical significance of the signal should be Conclusion
considered. For example, the associated results In this paper, we have illustrated the application
may be close to the control chart limits and iso- of the lifecycle management approach for an
Figure 9. I-MR Control Charts for Sample Averages (n=2): Samples 1-23 lated with no OOC or OOT points before or af- analytical procedure (a drug substance potency
ter those results. The process may be perform- test) and showed how a decision rule was used
ing at a very high level relative to specifications, to develop the analytical target profile (ATP).
suggesting that measurement variation may not The ATP was then used to define the fitness of
be a problem. the method for its intended use, which in turn
In the case when the variation has changed enables an objective assessment of the method
– increased or decreased – revised limits should performance during initial validation, transfer
be considered along with the appropriate ap- to the quality control laboratory (QCL), and its
provals and change control actions taken. Such ongoing performance in the QCL.
changes should be accompanied by appropriate The lifecycle approach is essentially applying
training of the analyst and supervision involved. Quality by Design (QbD) principles to the exe-
A control plan should also be created and put cution of the method to produce a result, in an
in place to ensure that the new procedures are analogous way that QbD is applied to a manu-
followed. facturing process producing a product. We have
Figure 10. I-MR Control Charts for Sample Averages (n=2): Samples 24-48 If unacceptable variation continues to be also demonstrated the use of the Gage R&R
www.ivtnetwork.com
Jane Weitzel, Robert A. Forbes, Ronald D. Snee