Professional Documents
Culture Documents
2 MC Prostate Presentation
2 MC Prostate Presentation
ALPHA-BETA RATIO
• The α/β ratio of most tumours is 10Gy; the α/β ratio of prostate cancer is 1.5Gy; rectum is 3Gy11
• Prostate cancer cells are more responsive to a higher single dose
HYPOFRACTIONATION
AIMS
• To summarise articles looking at intrafraction organ motion or volume change, and CTV-PTV margins
to determine if intrafraction motion necessitates larger PTV margins
METHODOLOGY
ARTICLE SEARCH
CHALLENGES
• Differences in data-reporting
• Differences in data collection
RESULTS
RESULTS
INTRAFRACTION MOTION
Study Patients Modality Measured point of interest (frequency) Statistic used Movement (mm)
AP SI LR
Mayyas et al., 2013 27 Electromagnetic Coordinates of EM transponders (real time, daily) Systematic 1.3 1.5 0.6
transponders (n=19) Random 2.6 2.4 1.4
Pre- and post-treatment KV 3 fiducial markers (daily) Mean -0.3 0.8 -0.2
(n=8) Systematic 2.4 2.7 2.1
Random 2.7 3.2 3.1
Tanyi et al., 2010 14 Electromagnetic Coordinates of EM transponders (real time, daily) Mean 0.22 0.26 0.12
transponders Direction Post Sup Right
Oehler, et al., 2014 20 Pre- and post-treatment 3 fiducial markers (#1-3, weekly) Systematic 1.39 1.36 0.92
CBCT Random 1.62 1.39 0.97
Olsen, et al., 2012 15 Electromagnetic Coordinates of EM transponders (real time – 0.1s Mean 0.39 0.69 0.64
transponders intervals, daily) (SD) (± 0.4) (± 0.6) (±0.5)
Mutanga, et al., 108 Pre- and post-treatment COM of 3-4 fiducial markers (daily) Mean of means 0.9 0.6 Not
2012 KV, and in-field MV Systematic 1.1 1.0 reported
Random 1.2 1.1
Thomas et al., 2013 54 Pre- and post-treatment Soft tissue contour (5#, weekly) Mean 0.5 0.5 0
MVCT Systematic 0.8 0.8 1.1
Random 2.2 2.0 1.3
Systematic - systematic error is calculated by taking the standard deviation of the mean set-up errors calculated for each patient
Random - random error is the root mean square of the standard deviation of set-up errors calculated for each patient
NR – Not reported; SD – Standard Deviation; COM – Centre of Mass; COG – Centre of Gravity
RESULTS
TRANSLATION
• 11 studies examined intrafraction motion, with data gathered from total of 626 patients16-18,19,7,20-25
• All studies delivered IMRT/VMAT – 9 with conventional fractionation; 2 with hypofractionation
• Range of imaging modalities
• Electromagentic transponders – 3
• EM transponders and pre- and post-treatment KV – 1
• Pre- and post-treatment CBCT, KV, or MV imaging with fiducials – 5
• CBCT or MVCT with soft tissue match – 2
• All of the studies found intrafraction gland motion to be largest in the anterior-posterior, or superior-inferior
direction
• Studies using fiducial markers with pre- and post-treatment KV imaging report the highest mean, systematic
error, and random error values
RESULTS
ROTATION
• 3 studies assessed prostate rotation, with a total of 81 patients
• Study of 15 patients, using electromagnetic transponders:
• Mean and standard deviation figures for magnitude of prostate rotation around the x-axis (pitch), y-axis
(roll), and z-axis (yaw) as 5.7° ± 5°, 2.0° ± 2°, and 1.6° ± 1°19
• Study of 27 patients, using MV image pairs for each treatment field, and fiducials:
• Systematic rotation - 0° ± 4.3°, 0° ± 1.6°, and 0° ± 1.3° (mean ± standard deviation) for rotation around the
x, y and z axes respectively
• Random variation - 2.7°, 1.1°, and 1.0° (mean standard deviation) around the x, y, and z axes
respectively.18
• Another study using in-field MV imaging showed LR rotations of 2.5° (mean of means) ± 2.3° (SD of means),
with a maximum rotation of 26.9° across 39 patients.26
RESULTS
TRANSLATION & ROTATION
Image showing planned dose cloud for a prostate +3mm PTV margin (grey). Translational shifts of 3mm
ANT and INF, and a pitch rotation of 20 degrees, were applied – with the new position of the prostate
(red) within the planned dose cloud shown.
RESULTS
TEMPORAL DEPENDENCE
• Three studies, of 180 patients in total, examined the temporal dependence of intrafraction prostate
motion18,21,22
• A variety of imaging methods: CBCT and EM transponders; MV image pairs and fiducials; and pre-and
post-treatment CBCTs with soft-tissue matching
• All studies found that the magnitude of motion increased with time
• One study of 10 patients performed pre- and post-treatment CBCTs, looking at COM prostate
displacement:
• Mean COM displacement ranged from 1.6—2.9 mm, increasing with elapsed time.
• Time between CBCTs ranged from 4 to 16 minutes (mean 7.3 min).
• Significant dose differences were observed between CBCTs, increasing with elapsed time. A
difference of up to 10% for an 8 minute treatment was observed at the V95% point for the CTV.22
RESULTS
INTRAFRACTION MOTION
Study Patients Modality Measured point of interest Notes Suggested margin (mm)
(frequency) AP SI LR
Mayyas et al., 2013 27 Electromagnetic Coordinates of EM Intrafraction motion only 5.0 5.4 2.5
transponders (n=19); pre- transponders (real time,
and post-treatment KV daily); 3 fiducial markers Intrafraction motion + residual setup error 6.6 6.8 3.9
(n=8) (daily)
Tanyi et al., 2010 14 Electromagnetic Coordinates of EM Skin mark localisation with intrafraction motion 16.3 11.4 7.52
transponders transponders (real time, CBCT marker alignment with intrafraction
daily) motion 3.23 3.68 2.81
Calypso-based alignment
2.28 2.64 1.36
Oehler, et al., 2014 20 Pre- and post-treatment 3 fiducial markers (#1-3, Low-risk patients, fiducial match on CBCT Ant – 7.18 7.09 5.47
CBCT weekly) Post – 5.47
High-risk patients, fiducial match on CBCT Ant – 10.52 9.67 7.39
Post – 6.39
Thompson et al., 151 Pre- and post-treatment KV Daily BMI – normal 4.61 4.13 2.54
2011 BMI – overweight 4.24 4.15 2.94
BMI – obese 3.7 3.35 2.4
BMI – severely obese 4.08 3.34 2.99
BMI – morbidly obese 4.06 3.41 2.29
Cramer, et al., 2013 143 Electromagnetic Daily (0.1s intervals) Transponder only localisation (IMAT/IMRT) 2.0-2.1 2.0 1.0-1.1
transponders
Transponder + CBCT localisation (IMAT/IMRT) 2.5 2.8-2.9 1.5-1.6
Thomas et al., 2013 54 Pre- and post-treatment Soft tissue contour (5#,
MVCT weekly) 2.1 2.1 2.2
RESULTS
CTV-PTV MARGINS
• In order to cover the CTV with a minimum dose of 95%, for 90% of patients, a CTV to PTV margin can be
calculated using systematic and random error, and Van Herk’s formula: 2.5Σ +0.7σ.29
• Six studies of 409 patients report suggested PTV margins calculated using the data obtained from
assessing prostate motion16,17,20,21,23,24
• All studies suggested anisotropic margins, ranging from 2.0 mm—16.3 mm vertically, 2.0 mm—11.4 mm
longitudinally, and 1.0 mm—7.52 mm laterally
• Studies assessing prostate motion with soft-tissue matching, found that CTV delineation uncertainties of
between 1—3.5 mm contributed most to the PTV margin,17 and inter-observer matching uncertainties
could potentially be as much as σ = 2 mm and Σ = 1 mm.24
• PTV margins of 0 mm were found to be insufficient at ensuring the PTV received no less than 98% of the
planned minimum dose19
DISCUSSION
DISCUSSION
• Intrafraction motion of the prostate is considered negligible compared to interfraction
uncertainties,30 however it is not insignificant
• All studies show that intrafraction motion is greatest in the SI and AP directions, meaning that a
tight PTV margin between the prostate and the rectum should be applied with caution
• Rotational displacement greater than 5° had a significant effect on the dose delivered to the PTV19
• Rotational displacement is not routinely assessed. Due to shape and position of seminal vesicles,
rotational displacement could potentially result in inadequate PTV coverage
• The studies show that prostate motion has a temporal dependence. VMAT, was shown to have
significantly reduced instances and magnitude of prostate motion compared to IMRT. Larger PTV
margins may be required for longer treatment times
DISCUSSION
• Studies that used real-time tracking, or in-field imaging, noted that there was a significant element of
residual set-up error - in some cases, another 1.6 mm to the planning margin was needed23
• Recommended that verification image acquisition, correction, and repositioning be as streamlined and
quick as possible, so as not to negate the pre-treatment corrections
• Despite blinding or averaging contours, studies suggested that a residual margin of 3 mm is needed to
compensate for delineation errors,26 which can approximate 2.03 mm in the longitudinal direction17
• No significant correlation between prostate deformation and hormonal treatment was found28
• Fiducial marker migration was rare – over 5 years, 5 out of 342 patients had lost a marker; 10 patients
had to be replanned because of migration >4mm
LIMITATIONS