5HSDLU,QVWUXFWLRQV *HQHUDOLQWURGXFWLRQ

 *HQHUDOLQWURGXFWLRQ 7DEOH Legend to Fig. 1


,WHP &RPSRQHQW
1 Connection board
2 IRIA cuvette
All rights reserved. Copyright reserved • R5132300T12_PGM.fm 05.04.02 • Repair Instructions • Version 1.0 • Released

3 IRIA, sensor head of anaesthetic gas analyzer


4 Bacterial filter
5 Fan
6 Water trap
7 Filter mat
8 Electrochemical O2 cell (fast O2 analysis)
9 MOPS PCB (electronics)
10 AMO IRIA PCB (anaesthetic gas analysis)
11 AMO O2 PUMP PCB (O2 analysis)
12 Pump (Pu)
13 Solenoid valve ILCA (V1, room air/sampling gas)
14 AMO FLOW ILCA PCB (actuation and monitoring
of valve, pump and flow values)
15 ILCA component carrier (for items 9 - 14)

)LJ Exploded view of PGM, for legend see Table


1

'UlJHU0HGLFDO$* &R.*D$‡3ULPXV 
5HSDLU,QVWUXFWLRQV *HQHUDOLQWURGXFWLRQ

 *HQHUDOLQWURGXFWLRQ

 7HVWLQJZLWKVHUYLFHPRGH

 7HVWLQJZLWKRXWVHUYLFHPRGH
All rights reserved. Copyright reserved • R5132300T12_PGM.fm 05.04.02 • Repair Instructions • Version 1.0 • Released

 &RQQHFWLRQGLDJUDPWXELQJGLDJUDP

)LJ PGM tubing diagram


)LJ PGM connection diagram

'UlJHU0HGLFDO$* &R.*D$‡3ULPXV 
5HSDLU,QVWUXFWLRQV *HQHUDOLQWURGXFWLRQ

 5HPRYLQJWKH3*0PRGXOH  5HPRYLQJ3*0KRXVLQJ

1. Remove the two PGM module fixing screws (see Fig. 1. Remove the three fixing screws (see Fig. 5).
4).
All rights reserved. Copyright reserved • R5132300T12_PGM.fm 05.04.02 • Repair Instructions • Version 1.0 • Released

)LJ Fixing screws

2. Remove the four side panel fixing screws (see Fig. 6).
)LJ PGM module fixing screws

2. Withdraw the PGM module from the Primus housing.

'UlJHU0HGLFDO$* &R.*D$‡3ULPXV 
5HSDLU,QVWUXFWLRQV *HQHUDOLQWURGXFWLRQ

2. Before mounting the top cover, check the following


points: Are all electrical connections plugged in? Are
the tubes all connected and without kinks? Are all
components firmly screwed down inside the PGM
module? Do all housing interlocks fit together (for
example see Fig. 7)?
All rights reserved. Copyright reserved • R5132300T12_PGM.fm 05.04.02 • Repair Instructions • Version 1.0 • Released

)LJ Side panel fixing screws

3. Tilt the side panel slightly (Fig. 7/1) and open up the top
cover.

 )LWWLQJWKH3*0KRXVLQJ

1. Refit the PGM housing in the reverse order of its


removal, as described in section 1.5.1 Removing PGM )LJ Example: housing interlocks
housing .
Note the following points during the various operations.

'UlJHU0HGLFDO$* &R.*D$‡3ULPXV 
5HSDLU,QVWUXFWLRQV *HQHUDOLQWURGXFWLRQ

 5HPRYLQJWKH,/&$

1. Remove the PGM module (see 1.5 Removing the PGM


module).
2. Unplug the three electrical connectors from the ILCA
All rights reserved. Copyright reserved • R5132300T12_PGM.fm 05.04.02 • Repair Instructions • Version 1.0 • Released

(see Fig. 8).

)LJ Disconnecting tube fittings

4. Disconnect one tube fitting (see Fig. 10).

)LJ ILCA electrical connections

3. Disconnect two tube fittings (see Fig. 9)

'UlJHU0HGLFDO$* &R.*D$‡3ULPXV 
5HSDLU,QVWUXFWLRQV *HQHUDOLQWURGXFWLRQ
All rights reserved. Copyright reserved • R5132300T12_PGM.fm 05.04.02 • Repair Instructions • Version 1.0 • Released

)LJ Pump tube fitting

6. Unplug the electrical connection from the IRIA (see Fig.


)LJ Disconnecting tube fitting 12).

5. Disconnect the tube fitting from the pump (see Fig. 11).

'UlJHU0HGLFDO$* &R.*D$‡3ULPXV 
5HSDLU,QVWUXFWLRQV *HQHUDOLQWURGXFWLRQ
All rights reserved. Copyright reserved • R5132300T12_PGM.fm 05.04.02 • Repair Instructions • Version 1.0 • Released

)LJ O2-Sensor connection


)LJ Connection to IRIA

8. Unscrew the four ILCA fixing screws (see Fig. 14).


7. Unplug the electrical connection to the O2 sensor from
the ILCA (see Fig. 13).

'UlJHU0HGLFDO$* &R.*D$‡3ULPXV 
5HSDLU,QVWUXFWLRQV *HQHUDOLQWURGXFWLRQ

3. Refit the electrical connections and tubing in the


reverse order of their removal (see section 1.4
Connection diagram/tubing diagram).
4. Fit the PGM housing (see section 1.5.2 Fitting the PGM
housing).
All rights reserved. Copyright reserved • R5132300T12_PGM.fm 05.04.02 • Repair Instructions • Version 1.0 • Released

 5HPRYLQJWKH,5,$

1. Remove the PGM module (see section 1.5 Removing


the PGM module).
2. Unplug the electrical connector from the IRIA (see Fig.
12).
3. Remove the four IRIA fixing screws (see Fig. 15 ).
)LJ ILCA fixing screws

 )LWWLQJWKH,/&$

1. Place the ILCA on a flat surface and transfer the


remaining tubing to the new ILCA (see section 1.4
Connection diagram/tubing diagram).

In the following fitting operation, first insert the


insulating foil between the ILCA and the housing.

2. Secure the new ILCA with the four fixing screws

'UlJHU0HGLFDO$* &R.*D$‡3ULPXV 
5HSDLU,QVWUXFWLRQV *HQHUDOLQWURGXFWLRQ

 )LWWLQJWKH,5,$

1. Disconnect the tubing from the IRIA and transfer it to


the new IRIA (see Fig. 16).
All rights reserved. Copyright reserved • R5132300T12_PGM.fm 05.04.02 • Repair Instructions • Version 1.0 • Released

)LJ IRIA fixing screws

4. Place the IRIA on a flat surface.

)LJ IRIA tubing

2. Disconnect the two tube fittings from the IRIA and


transfer them to the new IRIA (see Fig. 17)

'UlJHU0HGLFDO$* &R.*D$‡3ULPXV 
5HSDLU,QVWUXFWLRQV *HQHUDOLQWURGXFWLRQ
All rights reserved. Copyright reserved • R5132300T12_PGM.fm 05.04.02 • Repair Instructions • Version 1.0 • Released

)LJ IRIA tubing

3. Lay out two tubes and the O2 sensor cable underneath


the IRIA (see Fig. 18). )LJ Tubing underneath IRIA

4. Screw in the four IRIA fixing screws (see Fig. 14).


5. Plug in the electrical connector to the IRIA (see Fig.
12).
6. Fit the PGM housing (see section 1.5.2 Fitting the PGM
housing).

'UlJHU0HGLFDO$* &R.*D$‡3ULPXV 


5HSDLU,QVWUXFWLRQV *HQHUDOLQWURGXFWLRQ

 5HPRYLQJWKHIDQ

1. Remove the PGM module (see section 1.5 Removing


the PGM module).
2. Take out the dust filter on the rear of the module and
All rights reserved. Copyright reserved • R5132300T12_PGM.fm 05.04.02 • Repair Instructions • Version 1.0 • Released

unscrew the two fixing screws (see Fig. 19).

)LJ Fan direction of flow

)LJ Fan fixing screws 4. Plug the electrical connector of the new fan into the
ILCA (see section 1.4 Connection diagram/tubing
diagram).
3. Fit the new fan in keeping with the direction of flow (see
Fig. 20). 5. Fit the PGM housing (see section 1.5.2 Fitting the PGM
housing).

'UlJHU0HGLFDO$* &R.*D$‡3ULPXV 


5HSDLU,QVWUXFWLRQV *HQHUDOLQWURGXFWLRQ

 5HPRYLQJWKH2VHQVRUKRXVLQJ  )LWWLQJWKH2VHQVRUKRXVLQJ

1. Remove the PGM module (see section 1.5 Removing 1. Fit the tube connections one after the other (see
the PGM module). section 1.4 Connection diagram/tubing diagram).
2. Remove the IRIA (see section 1.7 Removing the IRIA). 2. Plug in the electrical connection between the ILCA and
All rights reserved. Copyright reserved • R5132300T12_PGM.fm 05.04.02 • Repair Instructions • Version 1.0 • Released

3. Take the cover plate out of the housing (see Fig. 21). the O2 sensor housing (see Fig. 13 or section 1.4
Connection diagram/tubing diagram).
3. Lay the tube to the pump, to the T-piece of S2 and the
O2 sensor cable underneath the IRIA (see Fig. 18).
4. Screw in the four IRIA fixing screws (see Fig. 14 ).
5. Plug in the electrical connector to the IRIA (see Fig.
12).
6. Fit the PGM housing (see section 1.5.2 Fitting the PGM
housing).

)LJ Cover plate and fixing screws

4. Slacken the two fixing screws enough for the O2


sensor housing to be removed (see Fig. 21).

'UlJHU0HGLFDO$* &R.*D$‡3ULPXV 