Professional Documents
Culture Documents
User mamual-PC400
User mamual-PC400
User mamual-PC400
User Manual
Ref: 1300029947
User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Contents
Foreword..........................................................................................................................................5
1. Document Update.................................................................................................................. 6
2. Legal Information....................................................................................................................8
Introduction................................................................................................................................. 11
1. Warning and Precautions.................................................................................................. 12
2. Operational Conditions...................................................................................................... 16
3. Instrument Overview........................................................................................................... 21
5. Printer .......................................................................................................................................32
Specifications.............................................................................................................................33
1. Technical Specifications....................................................................................................34
2. Physical Specifications...................................................................................................... 38
3. Reagent Specifications...................................................................................................... 40
4. Analysis Specifications...................................................................................................... 42
5. Limitations............................................................................................................................... 43
Software......................................................................................................................................... 45
1. Software Overview............................................................................................................... 46
2. Menus Description............................................................................................................... 47
Quality Assurance................................................................................................................... 61
1. Quality Control Overview...................................................................................................62
User Manual i
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
2. Quality Control by Control................................................................................................ 63
4. Westgard Rules..................................................................................................................... 82
5. Logs............................................................................................................................................86
Workflow......................................................................................................................................101
1. Sample Materials................................................................................................................ 102
8. End of Day..............................................................................................................................212
Settings......................................................................................................................................... 217
1. Reagent Management...................................................................................................... 218
2. Maintenance......................................................................................................................... 339
3. Troubleshooting.................................................................................................................. 372
ii User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Alarms............................................................................................................................................399
1. System Warnings and Alarms....................................................................................... 400
1. Document Update.................................................................................................................. 6
1.1. Revisions........................................................................................................................................6
1.2. What's New?..................................................................................................................................6
2. Legal Information....................................................................................................................8
2.1. Declaration of Conformity.............................................................................................................. 8
2.2. Notice of Liability............................................................................................................................8
2.3. Trademarks.................................................................................................................................... 8
2.4. Graphics.........................................................................................................................................8
2.5. Document Symbols........................................................................................................................9
2.6. Typographical Conventions........................................................................................................... 9
2.7. Copyright © 2016 by HORIBA ABX SAS..................................................................................... 10
User Manual 5
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Foreword
Document Update
1. Document Update
1.1. Revisions
This document applies to the latest software version listed and higher versions.
When a subsequent software version changes the information in this document, a new electronic
edition (USB flash drive and/or online help) is released and supplied by HORIBA Medical.
To update a paper document, please contact your local HORIBA Medical representative.
6 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Foreword
Document Update
User Manual 7
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Foreword
Legal Information
2. Legal Information
This product complies with the Standards and Directives named in the Declaration of Conformity.
The latest version of the EC Declaration of Conformity for this product is available on www.horiba-
abx.com/documentation.
The information in this manual is distributed on an "As Is" basis, without warranty. While every
precaution has been taken in the preparation of this manual, HORIBA Medical will not assume any
liability to any persons or entities with respect to loss or damage, caused or alleged to be caused
directly or indirectly by not following the instructions contained in this manual, or by using the
computer software and hardware products described herein in a manner inconsistent with our
product labelling.
2.3. Trademarks
2.4. Graphics
All graphics including screens, printouts and photographs are for illustration purposes only and are
not contractual.
8 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Foreword
Legal Information
To alert the operator of potentially hazardous conditions, symbols described in this chapter are
provided wherever necessary throughout the manual.
Emphasizes information that must be followed to avoid hazard to either the operator or the
environment, or both.
Emphasizes information that must be followed to avoid possible damage to the instrument
or erroneous test results.
Emphasizes information that can be helpful to the operator before, during or after a
specific operational function.
Before you start using this documentation, you should become familiar with the following
typographical conventions.
Indicates, from the main screen, the
sequence of menus you have to go
through to begin the procedure.
User Manual 9
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Foreword
Legal Information
All rights reserved. No part of this publication may be reproduced or transmitted in any form or by any
means, electronic, mechanical, photocopying, recording, or otherwise, without the prior written
permission of HORIBA Medical.
10 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Introduction
2. Operational Conditions...................................................................................................... 16
2.1. Environment................................................................................................................................. 16
2.2. Location....................................................................................................................................... 16
2.3. Grounding.................................................................................................................................... 17
2.4. Humidity and Temperature Conditions........................................................................................ 17
2.5. Electromagnetic Environment Check...........................................................................................17
2.6. Main Power Supply...................................................................................................................... 18
2.7. Environmental Protection.............................................................................................................18
2.8. Storage Conditions and Transportation.......................................................................................18
2.9. Installation.................................................................................................................................... 19
2.10. Package..................................................................................................................................... 20
3. Instrument Overview........................................................................................................... 21
5. Printer .......................................................................................................................................32
User Manual 11
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Introduction
Warning and Precautions
Work safety reliability and general characteristics are guaranteed by HORIBA Medical under the
following conditions:
■ User manual must be entirely read, and personnel trained by a HORIBA Medical representative
before attempting to operate the instrument.
■ The user always operates with full knowledge and appreciation of instrument warnings and alarms.
■ Always refer to labelling and HORIBA Medical instructions in order to avoid compromising system
integrity.
This instrument must be operated as instructed in the user manual. Any other use might compromise
system integrity and might be hazardous for the operator.
This instrument complies with Standards and Directives named in the Declaration of Conformity. The
latest version of the Declaration of Conformity for this instrument is available online at www.horiba-
abx.com/documentation.
■ The reagents and accessories stipulated by HORIBA Medical have been validated in
accordance with the European Directive for in vitro medical devices (98/79/EC).
■ The use of any other reagents and accessories may place the performance of the
instrument at risk, thus engaging user responsibility. In this case, HORIBA Medical
takes no responsibility for the device nor for the results rendered.
■ Disposable gloves, eye protection and lab coat must be worn by the operator.
■ Local or national regulations must be applied in all the operations.
■ Mobile phones should not be used in proximity of the instrument.
■ All peripheral devices should comply with relevant standards.
The duration of warranty is stipulated in the Sales conditions associated with the purchase of this
instrument. To validate the warranty, ensure the following is adhered to:
■ The system is operated under the instructions of this manual.
■ Only software or hardware specified by HORIBA Medical is installed on the instrument. This
software must be the original copyright version.
■ Services must be done by recommendation from HORIBA Medical, provided by an authorized
technician using only approved spare parts and at least once per year or more, depending on the
number of samples.
■ The electrical supply of the laboratory adheres to national or international regulations.
■ The system is operated according to HORIBA Medical recommendations.
■ Specimens are collected and stored in normal conditions.
■ Reagents used are those specified or recommended by HORIBA Medical.
■ Proper tools are used when maintenance or troubleshooting operations are performed.
If this instrument has been supplied to you by anyone other than HORIBA Medical or an
authorized representative, HORIBA Medical cannot guarantee this product in terms of
specification, latest revision and latest documentation. Further information may be
obtained from your authorized representative.
12 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Introduction
Warning and Precautions
Operator injury may occur from an electric shock. Electronic components can shock and
injure the user. Do not dismantle the instrument nor remove any components (covers,
doors, panels, etc.) unless otherwise instructed within this document.
Danger of explosion if battery is not replaced correctly! When replacing the battery, always
use the same and/or equivalent type recommended by the manufacturer. Dispose of used
batteries according to the manufacturer specific instructions.
Moving parts: It is strictly forbidden to disable sensors as it may cause operator injuries.
Protection covers must not be opened during instrument operations. Opening the doors
and covers during instrument operations triggers the instrument emergency stop.
Consider all specimens, reagents, calibrators, controls, etc. that contain human
specimen extracts as potentially infectious! Use established, good laboratory working
practices when handling specimens. Wear protective gear, gloves, lab coats, safety
glasses and/or face shields, and follow other biosafety practices as specified in OSHA
Blood borne Pathogens Rule (29 CFR part 1910. 1030) or equivalent biosafety
procedures.
The manufacturer uses disinfectant products for instrument decontamination and highly recommends
it to decontaminate your instrument. Refer to the Maintenance and Troubleshooting > Maintenance >
Other Procedures > To Decontaminate your Instrument chapter to perform the instrument cleaning
and decontamination procedure.
Related information:
■ To Decontaminate your Instrument, p.368
User Manual 13
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Introduction
Warning and Precautions
Notice of environment-friendly
Up
use period
14 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Introduction
Warning and Precautions
Reagent Buffer
Calibrator Control
User Manual 15
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Introduction
Operational Conditions
2. Operational Conditions
2.1. Environment
The operation of the Pentra C400 should be restricted to indoor location use only.
The instrument is operational at an altitude of maximum 3000 m (9840 ft).
The instrument is designed for safety from voltage surges according to INSTALLATION CATEGORY II
and POLLUTION DEGREE 2 (IEC 61010-1).
Please contact your local representative for information regarding operation locations when it does
not comply with the recommended specifications.
2.2. Location
Keep in mind that the instrument weighs approximately 120 kg (265 lb).
To move the instrument, four persons are required.
The lifting handles provided in the installation kit must be used.
16 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Introduction
Operational Conditions
The Power switch and Power supply connection should always be accessible. When
positioning the system for operational use, leave the required amount of space for easy
access to these items.
2.3. Grounding
Proper grounding is required when installing the system. Check the wall outlet ground (earth) for
proper grounding to the facilities electrical ground. If you are unsure about the outlet grounding,
contact your facilities engineer to verify the proper outlet ground.
Instrument operating temperature: from +15°C (+59°F) to +32°C (+90°F). If the instrument is stored
at a temperature lower than +10°C (+50°F), it should stand for one hour at normal room temperature
before use.
Humidity Conditions: Relative humidity of 20% - 85% maximum, without condensation.
Temperature gradient: 2°C (3.6°F) per hour.
The instrument has been designed to produce less than the accepted level of electromagnetic
interference in order to operate in conformity with its destination, allowing the correct operation of
other instruments also in conformity with their destination.
In case of suspected electromagnetic noise, make sure that the instrument has not been placed in the
proximity of electromagnetic fields or short wave emissions, e.g. Radar, X-rays, Scanners, Cell
phones, etc.
Do not perform analysis while cover is open or not correctly fixed. Electromagnetic noise
can affect the data or disrupt a nearby instrument.
User Manual 17
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Introduction
Operational Conditions
Grounding is required. Make sure the earth wall-plug is correctly connected to the laboratory
grounding system. If there is no such system, a ground stake should be used.
Use only the main supply cable delivered with the instrument.
Main power supply voltage fluctuations must not exceed +/- 10% of the nominal voltage.
Instrument Disposal
This product should be disposed of and recycled at the end of the useful life in
accordance with European Directive 2002/96/EC on Waste Electrical and Electronic
Equipment (WEEE) and/or European Directive 2006/66/EC on batteries and
accumulators.
Instrument storage and transportation temperatures: from -20°C (-4°F) to +50°C (+122°F).
Analyzer exposure to rainfall and extended sunlight must be avoided. The outdoors storage of the
analyzer is prohibited.
18 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Introduction
Operational Conditions
Keep in mind that the instrument weighs approximately 120 kg (265 lb).
To move the instrument, four persons are required.
The lifting handles provided in the installation kit must be used.
Before instrument removal from use, transportation or disposal, perform a general cleaning and a
draining of your instrument.
Related information:
■ To Decontaminate your Instrument, p.368
2.9. Installation
For further information about the cooling unit, refer to the cooling unit manual.
Package content:
■ Pentra C400 with or without ISE module (option)
■ Cooling unit
■ Ethylene glycol 2 x 1 L
■ Keyboard (for computer connection)
■ Power supply cable
■ Power strip
■ Software version USB flash drive
■ Reagent Application USB flash drive
■ Documentation USB flash drive
■ Safety Information booklet
■ Installation kit
■ Accessory kit
■ Printer
■ Lifting handles
■ Waste tank
■ Water tank
■ External sample cover
User Manual 19
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Introduction
Operational Conditions
Only HORIBA Medical approved accessories should be used with the Pentra C400.
2.10. Package
Factory package of the analyzer Pentra C400 and its implements consists of firm corrugated
cardboard, polyethylene foil and inner foam plastic framework. Package protects analyzer and its
implements from adverse factors of outside environment.
20 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Introduction
Instrument Overview
3. Instrument Overview
User Manual 21
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Introduction
Instrument Overview
1 = ISE module
2 = Sample tray
3 = Reagent tray
4 = Sampling system
5 = Mixer
6 = Cuvette changer
7 = Computer connections
8 = Reaction tray
22 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Introduction
Labels and Connections
The Power switch and Power supply connection should always be accessible. When
positioning the system for operational use, leave the required amount of space for easy
access to these items.
User Manual 23
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Introduction
Labels and Connections
Related information:
■ Analyzer Power Problems, p.372
Condensation water evacuation is made by gravity. Make sure that the tube goes
downward and never upward, and that it does not bend.
24 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Introduction
Labels and Connections
Dispose of waste according to your local/national guidelines for biohazard waste disposal.
1 = Printer connection (parallel port) until the CCC014N (1201848014) computer reference
2 = Host connection
Since the 1300013734 computer reference, the printer should be connected into a USB
port.
User Manual 25
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Introduction
Labels and Connections
3 = VGA
4 = Mouse
5 = Keyboard
7 = USB
8 = Loudspeaker
3 = USB
5 = USB
6 = Loudspeaker
26 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Introduction
Labels and Connections
User Manual 27
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Introduction
Labels and Connections
On the needles
28 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Introduction
Labels and Connections
User Manual 29
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Introduction
Labels and Connections
1 = Barcode label
location
2 = 11.5 mm
3 = Reagent rack
Refer to the Maintenance and Troubleshooting > Maintenance > Consumables and Spare Parts
chapter for rack labels references.
30 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Introduction
Labels and Connections
6 = Sample rack
Refer to the Maintenance and Troubleshooting > Maintenance > Consumables and Spare Parts
chapter for rack labels references.
You should preferably use the sample rack labels from HAX0166 (1207230166) to
HAX0174 (1207230174). Nevertheless, if you use the 2 of 5 interleaved barcode type
without check digit, you must use the sample rack labels from HAX0273 (1207230273) to
HAX0281 (1207230281).
10 = Reagent stickers
11 = Control stickers
12 = Calibrator stickers
Related information:
■ Consumables and Spare Parts, p.369
■ To Register a Reagent, p.299
User Manual 31
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Introduction
Printer
5. Printer
Contact your local HORIBA Medical representative for more information about printer
compatibility and consumable part numbers.
32 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Specifications
1. Technical Specifications....................................................................................................34
1.1. Intended Use................................................................................................................................34
1.2. Analysis Methods.........................................................................................................................34
1.3. Throughput...................................................................................................................................34
1.4. Reagent........................................................................................................................................34
1.5. Sample......................................................................................................................................... 35
1.6. Calibrator and Control..................................................................................................................36
1.7. Computer Characteristics............................................................................................................ 36
1.8. Measurement............................................................................................................................... 36
2. Physical Specifications...................................................................................................... 38
2.1. Power Requirements....................................................................................................................38
2.2. Humidity and Temperature Conditions........................................................................................ 38
2.3. Dimension and Weight................................................................................................................. 39
2.4. Sound Level................................................................................................................................. 39
2.5. Water Requirement...................................................................................................................... 39
3. Reagent Specifications...................................................................................................... 40
3.1. Reagent Notices...........................................................................................................................40
3.2. Performance Data........................................................................................................................ 41
3.3. Waste Handling Precautions........................................................................................................41
4. Analysis Specifications...................................................................................................... 42
5. Limitations............................................................................................................................... 43
5.1. Maintenance.................................................................................................................................43
5.2. Interferences................................................................................................................................ 43
User Manual 33
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Specifications
Technical Specifications
1. Technical Specifications
The Pentra C400 system is a fully automated chemistry analyzer using colorimetry, turbidimetry and
potentiometry technologies. It is mostly meant to be used for in vitro diagnostic analyses based on
homogeneous samples such as serum, plasma, urine and whole blood.
1.3. Throughput
1.4. Reagent
Packaging accepted
■ Twin compartment cassette 30/10 from HORIBA Medical
■ Twin compartment cassette 50/50 from HORIBA Medical
■ Twin compartment cassette 70/30 from HORIBA Medical
■ Twin compartment cassette 80/10 from HORIBA Medical
■ Single compartment cassette 100 from HORIBA Medical
■ Reagent rack
34 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Specifications
Technical Specifications
On board conditions
■ Capacity: 52 cassettes (39 tests with reagent racks)
■ Temperature: 44 positions refrigerated at 4°C - 10°C (39°F - 50°F) and eight positions at room
temperature
Reagent management
■ Barcode reagent identification
■ Back-up for same reagent
■ Remaining volume calculation
■ Automated reagent on board stability check
Reagent sampling
■ Volume:
■ Minimum: 2 µL with sample needle and 15 µL with reagent needle
■ Maximum: 600 µL
■ Capacitive level detection
■ Insufficient volume detection
■ Reagent preheating at 37°C +/- 0.5°C (99°F +/- 0.9°F)
1.5. Sample
Sample types
■ Serum
■ Plasma
■ Urine
■ Cerebrospinal fluid (CSF)
■ Whole blood
■ Homogeneous liquid
On board conditions
■ Capacity: 60 samples
■ Continuous loading
Sample management
■ Barcode sample identification
■ Tube/cup detector
Sampling
■ Volume:
User Manual 35
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Specifications
Technical Specifications
■ Minimum: 2 µL
■ Maximum: 95 µL for one step and 380 µL for four steps
■ Automatic sample dilution: 1/2 to 1/22500
■ Automatic post-concentration: x2 to x10
■ Capacitive level detection
■ Shock detection
■ Clot detection and insufficient volume detection
Calibrator/Control positioning
Sample tray and reagent tray
1.8. Measurement
Reaction system
■ Reaction cuvettes: disposable acrylic cuvettes
■ Cuvette volume: 150 µL - 600 µL
■ Automatic loading and unloading of cuvettes
■ "New cuvette" holder capacity: 360 cuvettes
■ "Used cuvette" holder capacity: 360 cuvettes
■ Mixing: stirring paddle
■ Reaction temperature: 37°C +/- 0.2°C (99°F +/- 0.36°F), air bath controlled
■ Measurement cycle: 12 seconds
■ Sampling cycle: 12 seconds
■ Reaction time: 12 seconds - 20 minutes
36 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Specifications
Technical Specifications
Optical system
■ Measurement principle: absorbance measurement (Bichromatic or Monochromatic)
■ Light source: tungsten-halogen lamp
■ Diffraction: concave reflective grating spectrograph
■ Detector: photodiode array
■ Wavelengths: 340, 380, 405, 420, 455, 490, 505, 520, 550, 560, 580, 600, 620, 660, 700 nm
■ Optical linearity:
Algorithms supported
■ Factor
■ Slope average
■ Linear regression
■ Linear interpolation
■ LOGIT/LOG4
■ LOGIT/LOG5
■ Exponential
User Manual 37
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Specifications
Physical Specifications
2. Physical Specifications
■ Power supply:
■ Instrument: from 100 V to 240 V (+/- 10%), 50 Hz to 60 Hz
■ Cooling unit:
■ 100 V (+/- 10%), 50 Hz to 60 Hz
■ 115 V (+/- 10%), 60 Hz
■ 230 V (+/- 10%), 50 Hz
■ 230 V (+/- 10%), 60 Hz
■ Maximum power consumption:
■ Instrument: 250 VA
■ Cooling unit: 700 W
■ Maximum heat output (instrument + cooling unit): 1940 BTU/h (2010 kJ/h)
Fuses characteristics:
Slow-blow internal fuses having the following characteristics: 2 x T 6.3 A H 250 V (5x20 mm)
Printer
Refer to your printer manual.
Instrument operating temperature: from +15°C (+59°F) to +32°C (+90°F). If the instrument is stored
at a temperature lower than +10°C (+50°F), it should stand for one hour at normal room temperature
before use.
Humidity Conditions: Relative humidity of 20% - 85% maximum, without condensation.
Temperature gradient: 2°C (3.6°F) per hour.
38 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Specifications
Physical Specifications
■ De-ionized/distilled water
■ Water specifications:
■ Resistivity > 5 MOhm.cm
■ Conductivity < 0.2 µS/cm
User Manual 39
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Specifications
Reagent Specifications
3. Reagent Specifications
In order for the instrument to operate correctly, high-quality reagents must be used.
HORIBA Medical provides a full range of reagents.
These reagents are used for in vitro diagnostic.
Refer to the reagent notices for Pentra C400 available on the Pentra C400 reagent online help or
online at www.horiba-abx.com/documentation for all reagent specifications.
The reagents specified for this instrument have been approved in accordance with the
European Directive 98/79/EC (Annex III) for in vitro medical devices.
HORIBA Medical manufactures and markets reagents, calibrators and control bloods
specially designed for use with this analyzer. The use of products not recommended may
give erroneous results or cause instrument operation problems. For all information
regarding the recommended products, please contact your local representative.
Reagent, Control and Calibrator notices/MSDS can be displayed on the Pentra C400
reagent online help. Latest versions of these documents are available online at
www.horiba.com.
40 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Specifications
Reagent Specifications
For any information on reagent performance data such as Accuracy, Precision, Linearity,
etc., please refer to the reagent notices for Pentra C400 available on the Pentra C400
reagent online help or online at www.horiba.com.
When disposing of waste, protective clothing must be worn (lab coat, gloves, eye
protection, etc.). Follow your local and/or national guidelines for biohazard waste disposal.
■ At the beginning of each day, before startup, check if the waste container needs to be
emptied.
■ During instrument operation, do not remove the liquid waste tube under any
circumstance.
■ If required, waste can be neutralized before being discarded. Follow your laboratory protocol when
neutralizing and disposing of waste.
■ Dispose of the waste container according to your local and/or national regulatory requirements.
User Manual 41
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Specifications
Analysis Specifications
4. Analysis Specifications
Items Specifications
Number of applications Up to 999 applications
Reference range Three ranges: Man/Default, Woman, Child (< 12 years)
Rerun range Three ranges: Man/Default, Woman, Child (< 12 years)
Calibrators Unlimited number of calibrators
Controls Unlimited number of controls
Ratio Up to 200 ratios
Profile Unlimited number of profiles
Patient orders Up to 60 samples on board, continuous loading
Patient, calibration and control results Up to 100000 results in current database, unlimited results archiving
42 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Specifications
Limitations
5. Limitations
5.1. Maintenance
In the Maintenance and Troubleshooting section, specific maintenance procedures are listed. The
maintenance procedures identified are mandatory for proper use and operation of the Pentra C400.
Failure to execute any of these recommended procedures may result in poor reliability of
the system.
5.2. Interferences
The known interfering substances for each reagent are listed in the corresponding reagent notice
available on the Pentra C400 reagent online help or online at www.horiba.com.
User Manual 43
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Specifications
Limitations
44 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Software
1. Software Overview............................................................................................................... 46
2. Menus Description............................................................................................................... 47
User Manual 45
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Software
Software Overview
1. Software Overview
The Pentra C400 includes a control station with a software installed on the embedded computer. The
touch screen allows you to navigate easily in the application.
Either pressing the screen or using a computer mouse activates the keys.
46 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Software
Menus Description
2. Menus Description
Generic toolbar
User/Admin/Sales/Tech
Pause
Admin/Sales
Test Result
Admin/Sales/Tech
Sampling Exception
Test Review Sales/Tech
Error Report
Tech
Sequencing
Tray
Reagent Configuration
Configuration
Reagent Help
Exit/Back
All
Warning
System Warnings
Alarm
Alert
Help
Stop/Startup
User Manual 47
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Software
Menus Description
Main menu
User/Admin/Sales/Tech
Worklist Calibration
Admin/Sales
Control
Admin/Sales/Tech
Patient
Sales/Tech
Result Validation Calibration
Tech
Control
Worklist/SID Search
Archives
Search by PID/
Patient Name
Calibration
Calibration/Control Control
Logs
Application Configuration
Services Diagnostics
System Configuration
Customer Services
Session
Control Monthly
Annual
Quality Control
Session
Test Monthly
Annual
Cuvette Status
Timing
Reagent supply
Work Balance
Sampling Exception
48 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Software
Menus Description
Services menu
User/Admin/Sales/Tech
Reagent
Admin/Sales
Error
Admin/Sales/Tech
Logs System configuration
Sales/Tech
Applications
Tech
Sequencing
Maintenance
Applications
Ratio
Incompatibility
Reagents
Diagnostics
Analyser
Local Settings
Host connection
Printer
Audible alarm
Maintenance
Users
Channel configuration
Cycles
ISE (option)
Barcode
User Manual 49
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Software
Software Buttons Description
The generic toolbar is located at the top of the screen and is present in all screens.
Reagent
To open the Reagent Configuration menu.
Configuration
50 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Software
Software Buttons Description
Instrument
Indicates the instrument status.
status
Data Print/Send
To print data/to send data to the host.
to Host
Instrument status
Status Description
"Start-Up" The instrument initializes.
"Reagent processing" The instrument is performing samplings only on the reagent tray and is
performing analyses.
User Manual 51
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Software
Software Buttons Description
"Archive" A search for archived results in the archive files saved on the hard disk or
a USB key is in process.
Related information:
■ To Configure Sound Alarms, p.328
The menu access buttons are located on both sides of the main menu. These buttons allow you to
access the main functions of the system.
Button Name Action and Description
The Services menu also displays menu access buttons. These buttons give access to five submenus.
Button Name Action and Description
52 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Software
Software Buttons Description
Reagent and sample trays are symbolized in the middle of the main menu. They allow you to visualize
solutions and samples configured on the trays and to access their details.
The colored circle around the reagent tray indicates if the reagent tray is accessible (green) or in
progress (blue).
User Manual 53
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Software
Software Buttons Description
The reagent tray contains 52 positions, the status of each solution is indicated by a color code to
which a legend is available by pressing the blue background outside the reagent tray.
For detailed information concerning the solution status, refer to the Workflow > Reagent
Management > Reagent Status > Reagent Tray Description chapter.
You can access solution details by pressing a position on the reagent tray.
The sample tray contains 60 sample positions, the status of each sample is indicated by a color code
to which a legend is available by pressing the blue background inside the sample tray.
For detailed information concerning the sample status, refer to the Workflow > Patient
Samples > Sample Status > Sample Tray Description chapter.
You can access sample rack details by pressing a position on the sample tray.
Related information:
■ Reagent Tray Description, p.111
■ Sample Tray Description, p.178
The status buttons are located at the bottom of the main menu.
Button Name Action and Description
To open the Cuvettes window and to check cuvettes.
Green when there is no error on cuvettes.
Red when:
■ The "new cuvette" holder is empty.
■ The "used cuvette" holder is full.
Cuvette ■ There are less than four new cuvettes remaining on the reaction
Status tray.
The Cuvettes window displays:
■ if the "new cuvette" holder is OK or empty,
■ if the "used cuvette" holder is OK or full,
■ the number of cuvettes available on the reaction tray (taking
into account the cuvettes already reserved for ordered tests).
Timing ■ The finish time may change during analyses if samples are
automatically rerun.
■ The finish time is an estimation, and the exact finish time may
slightly vary from the one estimated by the instrument.
54 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Software
Software Buttons Description
ISE
To check the ISE module status and to open the ISE calibration
Module
results screen.
Status
OK To validate an action.
To duplicate data.
Duplicate
To change the lot number of a calibrator or a control.
Target values To display the target values configured for a calibrator or a control.
User Manual 55
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Software
Software Buttons Description
56 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Software
Using the Software
Buttons
Buttons are not always active, depending on the screen currently displayed, the instrument status or
the login profile.
Tooltips
A tooltip is a short piece of information describing a button. Place your mouse pointer over a button to
display a tooltip.
Dropdown lists
A dropdown list is a list of predefined items. Select one item from the list. Only one item can be
selected from the list.
Check boxes
Check boxes are options you can select. Click the check box to select the option. Several options can
be selected in a list of check boxes.
Radio buttons
Radio buttons are options you can select. Click the radio button to select the option. Only one option
can be selected in a list of radio buttons.
User Manual 57
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Software
Using the Software
Data fields
Data fields can have a predefined format, like a date field, or can be empty. Use the keyboard to enter
data.
Scroll bars
Scroll bars can be either vertical or horizontal. Use them to display hidden parts of the screen or a list.
Calendars
Calendars help you to select a date. To choose a month, use the left and right arrows. Then choose
the day. When done, click outside the calendar to close it.
In most of the Pentra C400 software lists (worklist, result list, etc.), you can modify the columns order
and size.
■ Drag and drop the column to move it.
■ Place your mouse pointer between two columns to display a double arrow, and drag the column
separator to change the column size.
The columns configuration is kept in memory.
In the Pentra C400 software, data fields that display a concentration value have the following format:
six numerical characters maximum + decimal position configured in the application with seven
significant digits maximum.
If the concentration value is too high to be displayed with this format, the scientific notation is used.
The value is then displayed: "1.234567E+38" for example.
58 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Software
Using the Software
Help is context-sensitive. The Help button allows you to open a help page corresponding to the
current screen.
If no help is available for the current screen, the home page is displayed.
Contents tab
Once you have entered the embedded help, you can navigate to find further information. In the left
frame, you can expand (+) or collapse (-) the entries of the table of contents. You can also use the
following buttons to navigate through the help application:
displays the table of contents.
synchronizes the table of contents with the page currently displayed in the right frame.
Index tab
The index provides an alphabetical list of terms. Each index entry is clickable and allows you to
display the page relating to the subject matter.
Search tab
The search functionality allows you to perform a search on all help pages. Matching results are
displayed in the left frame and provide direct links to display the pages required.
User Manual 59
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Software
Using the Software
60 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
4. Westgard Rules..................................................................................................................... 82
5. Logs............................................................................................................................................86
5.1. Calibration.................................................................................................................................... 86
5.2. Reagent........................................................................................................................................87
5.3. Error............................................................................................................................................. 87
5.4. System configuration................................................................................................................... 87
5.5. Applications................................................................................................................................. 88
5.6. Sequencing.................................................................................................................................. 88
5.7. Maintenance.................................................................................................................................90
User Manual 61
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Quality Control Overview
The Quality Control (QC) allows you to check the quality of the measurements performed by the
instrument.
Statistical analyses of control results are performed. These statistical analyses consist in calculating
the average, the standard deviation and the coefficient of variation of the measurements over a
defined period.
Three periods are selectable:
■ the current worklist,
■ a month,
■ a year.
Additionally, Westgard rules may be implemented by the user if required. Westgard rules are detailed
in the Quality Assurance > Westgard Rules chapter, and the procedure to configure them is described
in the Settings > System Configuration > Results Validation > To Configure Controls Automatic
Validation chapter.
A control can control several tests and similarly several controls can control a test. That is why the QC
is done by reviewing either the results and statistical analyses of a selected control for each test
controlled with this control (QC by control) or the results and statistical analyses of all controls used
for a selected test (QC by test).
In addition, the QCP Export tab allows you to export control results for the Quality Control Program
(QCP) of HORIBA Medical.
HORIBA Medical offers an Online Interlaboratory Comparison Program (QCP) which provides internet
access to:
■ Submit Internal Quality Control results online.
■ Monitor analytical performance and compare directly with hundreds of laboratories worldwide.
■ Obtain real time peer group statistical reports from QCP.
More information on http://qcp.horiba-abx.com.
Related information:
■ Quality Control by Control, p.63
■ Quality Control by Test, p.73
■ Westgard Rules, p.82
■ To Configure Controls Automatic Validation, p.326
■ To Export Control Results for QCP, p.71
62 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Quality Control by Control
Follow this procedure to check control results by control over the current worklist, a
month or a year.
1. Select a control (and lot number) from the Control dropdown list.
The controls that are expired or closed are not displayed by default in the Control dropdown list.
Only the current controls are displayed in the list. If you want to display all the controls, including
those that are expired or closed, select the All Controls option.
The control expiration date is displayed in the Expiration date field.
2. Press either Session, Monthly or Annual tab.
For the Monthly and Annual tabs, select a month and/or a year from the Month and Year
dropdown lists.
The statistical analyses over the current worklist, a month or a year for each test controlled with
the selected control are displayed in the table at the top of the screen.
3. Select an item from the table to display the detailed control results for a test.
The control results for the selected test are displayed as values (in the table at the bottom of the
screen) and as graphs.
4. Press either the Values, the Graph or the Both tab to display the results as values, graphs or
both.
User Manual 63
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Quality Control by Control
Related information:
■ Session Values Interpretation, p.64
■ Monthly Values Interpretation, p.65
■ Annual Values Interpretation, p.66
■ Graphs Interpretation, p.67
■ To Print Control Results by Control, p.69
Access: Main menu > Quality Control > Control > Session
For detailed information concerning the procedure to check control results by control,
refer to the Quality Control by Control > To Check Control Results by Control chapter.
Heading Description
Test Application local code for the controlled test
N Total number of controls performed over the current worklist
N Calculation Number of controls used for the statistical analyses
Mean Session mean value of control results
SD Session standard deviation of control results
CV Session coefficient of variation of control results
Target Value Control target value
Confidence Range Low and high limit values of confidence range
Heading Description
Run Date Date and time of control run
Validator Session Name User name or login used when validating the result
Final Result Final result
64 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Quality Control by Control
Heading Description
Quality flag Quality flag
Flag Analytical flag
If several analytical flags exist, the flag having priority is displayed.
Decision A (Accept) or D (Delete)
Unit Test unit configured in the application
The control results which have been excluded from the Result Validation menu are
grayed.
Related information:
■ To Check Control Results by Control, p.63
Access: Main menu > Quality Control > Control > Monthly
For detailed information concerning the procedure to check control results by control,
refer to the Quality Control by Control > To Check Control Results by Control chapter.
Heading Description
Test Application local code for the controlled test
N Total number of controls performed over the month
N Calculation Number of controls used for the statistical analyses
Mean Monthly mean value of control results
SD Monthly standard deviation of control results
CV Monthly coefficient of variation of control results
Target Value Control target value
Confidence Range Low and high limit values of confidence range
Heading Description
Run Date Date and time of control run
Validator Session Name User name or login used when validating the result
Final Result Final result
User Manual 65
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Quality Control by Control
Heading Description
Quality flag Quality flag
Flag Analytical flag
If several analytical flags exist, the flag having priority is displayed.
Decision A (Accept) or D (Delete)
Unit Test unit configured in the application
The control results which have been excluded from the Result Validation menu are
grayed.
Related information:
■ To Check Control Results by Control, p.63
Access: Main menu > Quality Control > Control > Annual
For detailed information concerning the procedure to check control results by control,
refer to the Quality Control by Control > To Check Control Results by Control chapter.
Heading Description
Test Application local code for the controlled test
N Total number of controls performed over the year
N Calculation Number of controls used for the statistical analyses
Mean Annual mean value of control results
SD Annual standard deviation of control results
CV Annual coefficient of variation of control results
Target Value Control target value
Confidence Range Low and high limit values of confidence range
Heading Description
Save Date Month and year of the displayed statistical analyses
N Total number of controls performed over the month
N Calculation Number of controls used for the statistical analyses
Mean Monthly mean value of control results
66 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Quality Control by Control
Heading Description
Unit Test unit configured in the application
SD Monthly standard deviation of control results
CV Monthly coefficient of variation of control results
Related information:
■ To Check Control Results by Control, p.63
For detailed information concerning the procedure to check control results by control,
refer to the Quality Control by Control > To Check Control Results by Control chapter.
Graphs display the control results for the selected test over the current worklist, a month or a year.
For a year, the monthly mean values are displayed. Two graphs are selectable:
■ Accuracy Cumulative Control
■ Precision Cumulative Control
User Manual 67
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Quality Control by Control
The control results which have been excluded from the Result Validation menu are
symbolized by a red triangle .
Drag your mouse pointer from left to right to zoom in or from right to left to zoom out.
The control results which have been excluded from the Result Validation menu are
symbolized by a red triangle .
Drag your mouse pointer from left to right to zoom in or from right to left to zoom out.
Related information:
■ To Check Control Results by Control, p.63
68 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Quality Control by Control
Follow this procedure to print control results by control over the current worklist, a month
or a year.
1. Select a control (and lot number) from the Control dropdown list.
The controls that are expired or closed are not displayed by default in the Control dropdown list.
Only the current controls are displayed in the list. If you want to display all the controls, including
those that are expired or closed, select the All Controls option.
The control expiration date is displayed in the Expiration date field.
2. Press either Session, Monthly or Annual tab.
For Monthly and Annual tabs, select a month and/or a year from Month and Year dropdown lists.
The results and statistical analyses over the current worklist, a month or a year for each test
controlled with the selected control are displayed.
3. Press Data Print/Send to Host from the generic toolbar.
The Print window is displayed.
For the Annual tab, Detailed by Accuracy and Detailed by Precision options are
unavailable.
c. To print the graphs, select Accuracy Graph and/or Precision Graph options.
5. Press OK to validate.
User Manual 69
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Quality Control by Control
70 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Quality Control by Control
Follow this procedure to export control results for the Quality Control Program (QCP) of
HORIBA Medical.
User Manual 71
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Quality Control by Control
Results are listed as a table which indicates the control name, the control lot number, the date of
the first result and of the last result, if the control results are selected to be exported.
3. By default, all control results are selected to be exported. If necessary, deselect the control results
that should not be exported.
Press the Select all button to select all the tests or the Deselect all button
4. Press Export:
A dialog box asks you to confirm that the USB key is in place.
5. Install your USB key on the instrument.
HORIBA Medical recommends you to make sure that all materials like CD-ROM or USB
key used on the Pentra C400 and non-referenced by HORIBA Medical are free of
viruses. These materials must imperatively be cleaned before being used on the
instrument.
6. Press OK to validate.
A dialog box informs you when the export is completed.
7. Press OK.
The selected control results are saved in an .xml file (one file per control) located in a folder named
"Export" on the USB key.
72 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Quality Control by Test
Follow this procedure to check control results by test over the current worklist, a month or
a year.
User Manual 73
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Quality Control by Test
Related information:
■ Session Values Interpretation, p.74
■ Monthly Values Interpretation, p.75
■ Annual Values Interpretation, p.76
■ Graphs Interpretation, p.77
■ To Print Control Results by Test, p.79
Access: Main menu > Quality Control > Test > Session
For detailed information concerning the procedure to check control results by test, refer to
the Quality Control by Test > To Check Control Results by Test chapter.
Heading Description
Name Control name
Lot Control lot number
N Total number of controls performed over the current worklist
N Calculation Number of controls used for the statistical analyses
Mean Session mean value of control results
SD Session standard deviation of control results
CV Session coefficient of variation of control results
Target Value Control target value
Confidence Range Low and high limit values of confidence range
Heading Description
Run Date Date and time of control run
Validator Session Name User name or login used when validating the result
74 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Quality Control by Test
Heading Description
Final Result Final result
Quality flag Quality flag
Flag Analytical flag
If several analytical flags exist, the flag having priority is displayed.
Decision A (Accept) or D (Delete)
Unit Test unit configured in the application
The control results which have been excluded from the Result Validation menu are
grayed.
Related information:
■ To Check Control Results by Test, p.73
Access: Main menu > Quality Control > Test > Monthly
For detailed information concerning the procedure to check control results by test, refer to
the Quality Control by Test > To Check Control Results by Test chapter.
Heading Description
Name Control name
Lot Control lot number
N Total number of controls performed over the month
N Calculation Number of controls used for the statistical analyses
Mean Monthly mean value of control results
SD Monthly standard deviation of control results
CV Monthly coefficient of variation of control results
Target Value Control target value
Confidence Range Low and high limit values of confidence range
User Manual 75
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Quality Control by Test
Heading Description
Run Date Date and time of control run
Validator Session Name User name or login used when validating the result
Final Result Final result
Quality flag Quality flag
Flag Analytical flag
If several analytical flags exist, the flag having priority is displayed.
Decision A (Accept) or D (Delete)
Unit Test unit configured in the application
The control results which have been excluded from the Result Validation menu are
grayed.
Related information:
■ To Check Control Results by Test, p.73
Access: Main menu > Quality Control > Test > Annual
For detailed information concerning the procedure to check control results by test, refer to
the Quality Control by Test > To Check Control Results by Test chapter.
Heading Description
Name Control name
Lot Control lot number
N Total number of controls performed over the year
N Calculation Number of controls used for the statistical analyses
Mean Annual mean value of control results
SD Annual standard deviation of control results
CV Annual coefficient of variation of control results
Target Value Control target value
Confidence Range Low and high limit values of confidence range
76 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Quality Control by Test
Heading Description
Save Date Month and year of the displayed statistical analyses
N Total number of controls performed over the month
N Calculation Number of controls used for the statistical analyses
Mean Monthly mean value of control results
Unit Test unit configured in the application
SD Monthly standard deviation of control results
CV Monthly coefficient of variation of control results
Related information:
■ To Check Control Results by Test, p.73
For detailed information concerning the procedure to check control results by test, refer to
the Quality Control by Test > To Check Control Results by Test chapter.
Graphs display the control results for the selected control over the current worklist, a month or a year.
For a year, the monthly mean values are displayed. Two graphs are selectable:
■ Accuracy Cumulative Control
■ Precision Cumulative Control
User Manual 77
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Quality Control by Test
The control results which have been excluded from the Result Validation menu are
symbolized by a red triangle .
Drag your mouse pointer from left to right to zoom in or from right to left to zoom out.
78 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Quality Control by Test
The control results which have been excluded from the Result Validation menu are
symbolized by a red triangle .
Drag your mouse pointer from left to right to zoom in or from right to left to zoom out.
Related information:
■ To Check Control Results by Test, p.73
Follow this procedure to print control results by test over the current worklist, a month or a
year.
User Manual 79
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Quality Control by Test
For the Annual tab, Detailed by Accuracy and Detailed by Precision options are
unavailable.
c. To print the graphs, select the Accuracy Graph and/or the Precision Graph options.
5. Press OK to validate.
80 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Quality Control by Test
User Manual 81
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Westgard Rules
4. Westgard Rules
Westgard rules are additional checks that apply to default controls for validation.
For detailed information concerning the procedure to configure Westgard rules, refer to
the Settings > System Configuration > Results Validation > To Configure Controls
Automatic Validation chapter.
Definitions
■ When configuring a control for a test, the target value and the low and high limit values of the
confidence range are entered. These three values allow you to define the following thresholds:
■ target value +/- 1SD (Standard Deviation),
■ target value +/- 2SD,
■ target value +/- 3SD,
■ target value +/- 4SD,
where 1SD = (high limit of confidence range - low limit of confidence range) / 4.
■ Control level: Up to three default controls can be configured for a test. Each default control is a
control level for the test.
■ Intra-control rule: The rule applies to the last results of each control level independently of other
control levels.
■ Inter-control rule: The rule applies to the last results of all control levels.
Westgard rules
Westgard rules allow you to characterize whether a control result outside the confidence range is a
random error non specific to the system or if it is due to a problem on the system.
■ If the control result is outside the confidence range and no Westgard rule is true, then the result is
flagged with CONF_RANGE_HIGH_W / LOW_W but the automatic validation is not blocked.
■ If the control result is outside the confidence range and the Westgard rule 1, 2, 3, 4, 5 or 6 is true,
then the result is flagged with CONF_RANGE_HIGH_W1-6 / LOW_W1-6 and the automatic
validation is blocked.
When Westgard rules are enabled, the following rules are checked one after the other.
Starting rule, 12S rule
The last control result is higher than + 2SD or lower than - 2SD.
QC values
26
+ 3SD
24
high limit = + 2SD
22
+ 1SD
Values
20 target value
- 1SD
18
low limit = - 2SD
16
- 3SD
14
1 2 3 4 5 6 7 8 9 10
Runs
82 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Westgard Rules
QC values
26
+ 3SD
24
high limit = + 2SD
22
+ 1SD
Values
20 target value
- 1SD
18
low limit = - 2SD
16
- 3SD
14
1 2 3 4 5 6 7 8 9 10
Runs
■ If true, then the result is flagged with CONF_RANGE_HIGH_W1 / LOW_W1.
■ If false, then the rule 2 is checked.
QC values
26
+ 3SD
24
high limit = + 2SD
22
+ 1SD
Values
20 target value
- 1SD
18
low limit = - 2SD
16
- 3SD
14
1 2 3 4 5 6 7 8 9 10
Runs
■ If true, then the result is flagged with CONF_RANGE_HIGH_W2 / LOW_W2.
■ If false, then the rule 3 is checked.
User Manual 83
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Westgard Rules
Inter-control rule: The difference between the last results of two different control levels is higher than
4SD.
QC values
26
+ 3SD
24
high limit = + 2SD
22
+ 1SD
Values
20 target value
- 1SD
18
low limit = - 2SD
16
- 3SD
14
1 2 3 4 5 6 7 8 9 10
Runs
■ If true, then the result is flagged with CONF_RANGE_HIGH_W3 / LOW_W3.
■ If false, then the rule 4 is checked.
QC values
26
+ 3SD
24
high limit = + 2SD
22
+ 1SD
Values
20 target value
- 1SD
18
low limit = - 2SD
16
- 3SD
14
1 2 3 4 5 6 7 8 9 10
Runs
■ If true, then the result is flagged with CONF_RANGE_HIGH_W4 / LOW_W4.
■ If false, then the rule 5 is checked.
84 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Westgard Rules
Inter-control rule: The last ten control results, regardless of the control level, are all higher than the
target value or all lower than the target value.
QC values
26
+ 3SD
24
high limit = + 2SD
22
+ 1SD
Values
20 target value
- 1SD
18
low limit = - 2SD
16
- 3SD
14
1 2 3 4 5 6 7 8 9 10
Runs
■ If true, then the result is flagged with CONF_RANGE_HIGH_W5 / LOW_W5.
■ If false, then the rule 6 is checked.
Rule 6, 7T rule
The last seven control results are strictly increasing or decreasing.
QC values
26
+ 3SD
24
high limit = + 2SD
22
+ 1SD
Values
20 target value
- 1SD
18
low limit = - 2SD
16
- 3SD
14
1 2 3 4 5 6 7 8 9 10
Runs
■ If true, then the result is flagged with CONF_RANGE_HIGH_W6 / LOW_W6.
■ If false, then the result is flagged with CONF_RANGE_HIGH_W / LOW_W that means the control
result is outside the confidence range and no Westgard rule is true.
Related information:
■ To Configure Controls Automatic Validation, p.326
User Manual 85
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Logs
5. Logs
Logs list events notified while the instrument is operating. Seven logs are available:
■ Calibration
■ Reagent
■ Error
■ System configuration
■ Applications
■ Sequencing
■ Maintenance
Events are listed in a table format which indicates the event date and time, the user name or login
used when the event occurred, the event name and description except for Error and Sequencing
logs which display specific information.
Events of the last six months are stored from the most recent to the oldest except for the Sequencing
log which displays events of the last seven days from the oldest to the most recent.
5.1. Calibration
86 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Logs
The Calibration log displays the events by test. Select a test from the Name dropdown
list.
5.2. Reagent
5.3. Error
The Error log lists the system warnings and alarms triggered while the instrument is operating.
The system warnings and alarms are listed in a table format which indicates the event level (warning
or alarm), the event date and time, the user name or login used when the event occurred, the event
code and message.
For detailed information on system warnings and alarms, refer to the Alarms > System
Warnings and Alarms chapter.
Related information:
■ System Warnings and Alarms, p.400
Access: Main menu > Services > Logs > System configuration
User Manual 87
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Logs
■ every system parameter modification from the Analyser, the Local Settings and the Results
validation tabs,
■ every host connection modification from the Host connection tab,
■ every printer settings modification from the Printer tab,
■ every modification of the sound alarms configuration from the Audible alarm tab,
■ every maintenance configuration modification from the Maintenance tab,
■ every user account creation, deletion or modification from the Users tab,
■ every modification of the user channels number (configured by your local HORIBA Medical
representative).
5.5. Applications
For an application provided by HORIBA Medical or your local representative, the following
events are specified:
■ every modification of the application version number,
■ every modification of a user parameter.
5.6. Sequencing
Heading Description
Date Sampling date and time
Nature Sampling nature:
■ for a test: Application local code,
■ for a dilution: Dilution,
■ for a cleaning: Cleaner.
Reagent needle Solution pipetted by the reagent needle
Volume Volume pipetted by the reagent needle (µL)
88 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Logs
Heading Description
Sampling position Position where the solution is pipetted:
■ on the reagent tray: R (for reagent tray) + position number on the tray +
position number in the cassette or on the reagent rack.
Type Container type where the solution is pipetted:
■ for a cassette: Cassette,
■ for a reagent rack: Rack ID + position number on the reagent rack.
Dispensing position Position where the solution is dispensed:
■ on the reaction tray: C (for cuvette) + position number on the tray +
position number in the cuvette segment.
Sample Arm
Heading Description
Date Sampling date and time
Nature Sampling nature:
■ for a test: Application local code,
■ for an ISE test: ISE,
■ for a dilution: Dilution,
■ for a cleaning: Cleaner.
Sample needle Solution pipetted by the sample needle
Volume Volume pipetted by the sample needle (µL)
H2O Distilled water volume pipetted by the sample needle (µL) to push the sample
volume
Sampling position Position where the solution is pipetted:
■ on the sample tray: S (for sample tray) + rack number + position number
on the rack,
■ on the reagent tray: R (for reagent tray) + position number on the tray +
position number in the cassette or on the reagent rack,
■ on the reaction tray: C (for cuvette) + position number on the tray +
position number in the cuvette segment.
Type Container type where the solution is pipetted:
■ for a sample tube: Tube,
■ for a sample cup: Sample cup,
■ for a cassette: Cassette,
■ for a reagent rack or a calibrator/control reagent rack: Rack ID + position
number on the reagent rack,
■ for a cuvette: Cuvette.
Dispensing position Position where the solution is dispensed:
■ on the reaction tray: C (for cuvette) + position number on the tray +
position number in the cuvette segment,
■ in the ISE module sample cup: ISE bowl.
Other
Heading Description
Date Date and time of mixing or cuvette segment loading/unloading
Mixing speed Mixing speed used to homogenize solutions into the cuvette
Cuvette loader Cuvette segment loading/unloading: Loading or Unloading + position
number on the tray
Events of the last seven days are stored from the oldest to the most recent.
User Manual 89
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Logs
5.7. Maintenance
The Maintenance log lists the maintenance procedures that are performed on the instrument. These
procedures must be configured in Main menu > Services > System Configuration > Maintenance
to be listed in the log.
Most of the maintenance procedures use a software function from the Customer Services menu.
These procedures are automatically notified in the log when they are performed.
Some maintenance procedures do not use a software function. These procedures must be manually
notified in the log when they are performed.
The Maintenance log displays all maintenance procedures by default. You can select a
frequency from the Occurrence dropdown list to display the maintenance procedures by
frequency.
The maintenance procedures are listed in a table format which indicates the event date and time, the
user level, and the user name or login used when the event occurred, the event name and description
and the event comment.
An item in red means that the maintenance procedure has not been performed on time.
Related information:
■ Configuring Maintenance Alerts, p.329
■ To Add a Maintenance Procedure, p.90
■ To Add a Comment, p.91
When a maintenance procedure (which does not use a software function) is performed on the
instrument, you must manually notify this procedure in the log.
90 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Logs
Only the maintenance procedures which do not use a software function are selectable.
The maintenance procedures which use a software function are automatically notified
in the log when they are performed.
User Manual 91
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Test Counter (On Request)
Contact your local HORIBA Medical representative to activate the Test Counter.
Follow this procedure to monitor your activity balance for all the applications.
Access: Main menu > Services > Customer Services > Test Counter
92 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Test Counter (On Request)
Press Select all to select all the applications or Deselect all to deselect all the
applications.
For detailed information concerning test statistics interpretation, refer to the Test Counter
(On Request) > Test Statistics Interpretation chapter.
Related information:
■ Test Statistics Interpretation, p.94
User Manual 93
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Test Counter (On Request)
For detailed information concerning the procedure to check test statistics, refer to the Test
Counter (On Request) > To Check Test Statistics chapter.
Heading Description
Selection To select or to deselect an application.
If selected, the tests for this application are taken into account in the
cumulative test statistics table.
Channel # Application channel number
Current code Application local code
Control Number of control tests performed
Calibrator Number of calibration tests performed
Patient Number of patient tests performed
Reportable Number of patient test results having the decision Accept or Modify
Rerun Number of patient tests with a rerun (manual or automatic)
% rerun Percentage of reruns compared with the number of patient test results
having the decision Accept or Modify (Reportable column)
Reportable deleted Number of patient test results having the decision Delete
Failure Number of failed patient tests, calculated as follows:
Failure column = Patient column - (Reportable + Rerun + Reportable
deleted) columns.
Total Total number of patient, calibration and control tests
Heading Description
Control Number of control tests performed
Calibrator Number of calibration tests performed
Patient Number of patient tests performed
Reportable Number of patient test results having the decision Accept or Modify
Rerun Number of patient tests with a rerun (manual or automatic)
% rerun Percentage of reruns compared with the number of patient test results
having the decision Accept or Modify (Reportable column)
Reportable deleted Number of patient test results having the decision Delete
94 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Test Counter (On Request)
Heading Description
Failure Number of failed patient tests, calculated as follows:
Failure column = Patient column - (Reportable + Rerun + Reportable
deleted) columns.
Total Total number of patient, calibration and control tests
Related information:
■ To Check Test Statistics, p.92
Follow this procedure to monitor your activity balance for one application.
Access: Main menu > Services > Customer Services > Test Counter
User Manual 95
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Test Counter (On Request)
3. Select the application in the test statistics table and press Detail.
Results are displayed in the detailed test statistics table.
4. To sort the results, select either the Year, the Month or the Day option.
The selectable options depend on the defined period.
5. To display cumulative test statistics for several periods, select the periods in the detailed test
statistics table.
Press Select all to select all the periods or Deselect all to deselect all the periods.
6. To display detailed test statistics for several periods as a graph, select the periods in the detailed
test statistics table (100 periods maximum are selectable) and press Graph.
Results are displayed in a graph format.
7. To print detailed test statistics for several periods, proceed as follows:
a. Select the periods in the detailed test statistics table and press Data Print/Send to Host from
the generic toolbar.
A dialog box asks for user confirmation.
b. Press OK to validate or Cancel to cancel.
96 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Test Counter (On Request)
For detailed information concerning detailed test statistics interpretation, refer to the Test
Counter (On Request) > Detailed Test Statistics Interpretation chapter.
Related information:
■ Test Statistics Interpretation, p.94
■ Detailed Test Statistics Interpretation, p.97
For detailed information concerning the procedure to check test statistics, refer to the Test
Counter (On Request) > To Check Detailed Test Statistics for One Application chapter.
Heading Description
Selection To select or to deselect a period.
If selected, the tests for this period are taken into account in the cumulative
test statistics table.
Time Period Period
Control Number of control tests performed
Calibrator Number of calibration tests performed
Patient Number of patient tests performed
Reportable Number of patient test results having the decision Accept or Modify
Rerun Number of patient tests with a rerun (manual or automatic)
User Manual 97
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Test Counter (On Request)
Heading Description
% rerun Percentage of reruns compared with the number of patient test results
having the decision Accept or Modify (Reportable column)
Reportable deleted Number of patient test results having the decision Delete
Failure Number of failed patient tests, calculated as follows:
Failure column = Patient column - (Reportable + Rerun + Reportable
deleted) columns.
Total Total number of patient, calibration and control tests
Heading Description
Control Number of control tests performed
Calibrator Number of calibration tests performed
Patient Number of patient tests performed
Reportable Number of patient test results having the decision Accept or Modify
Rerun Number of patient tests with a rerun (manual or automatic)
% rerun Percentage of reruns compared with the number of patient test results
having the decision Accept or Modify (Reportable column)
Reportable deleted Number of patient test results having the decision Delete
Failure Number of failed patient tests, calculated as follows:
Failure column = Patient column - (Reportable + Rerun + Reportable
deleted) columns.
Total Total number of patient, calibration and control tests
Graph
The graph displays detailed test statistics for the periods selected in the detailed test statistics table.
98 User Manual
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Test Counter (On Request)
Related information:
■ To Check Detailed Test Statistics for One Application, p.95
User Manual 99
Ref: 1300029947 - Int. Ref. Doc.: RAB281EEN
Quality Assurance
Test Counter (On Request)
8. End of Day..............................................................................................................................212
8.1. To Check the Patient Worklist....................................................................................................212
8.2. To Check the Instrument............................................................................................................213
8.3. To Shut Down the Instrument.................................................................................................... 214
8.4. To Switch the Printer Off............................................................................................................216
1. Sample Materials
Samples such as serum, plasma, urine, cerebrospinal fluid (CSF), whole blood and homogeneous
liquid can be used on the Pentra C400.
Ensure that a sufficient sample volume without air bubbles or foam is placed in the
instrument, whatever the sample type (calibrator, control, patient sample).
Refer to each reagent notice to determine the sample volume required (refer to the Performance on
Pentra C400 > Sample Volume paragraph) and take into account the following dead volumes:
■ primary and secondary tubes: 1 mL,
■ sample cups: 100 µL.
For any further information or special care regarding samples such as pretreatment or sample
stability, please refer to the corresponding reagent notice.
The size of a barcode varies within certain limits depending on print conditions. The width of the
symbol must be selected so that each bar corresponds to the entire number of the printer dots except
when the width of the barcode bars varies within +/- 1 printer dot and especially for low printing
resolution with larger size dots. The margins on the left and right sides of the barcode must always be
proportional to the width of the symbol. Therefore, ensuring high quality of each barcode as well as
accuracy and correct data structure.
2. Workflow Overview
Start of day
End of day
3. Start of Day
At the beginning of each day (before startup) or at the end of each day (before shutdown),
check the following items:
■ distilled water level,
■ waste level,
■ "new cuvette" holder,
■ "used cuvette" holder,
■ levels of ISE reagents.
1. Check the distilled water level in the tank. Fill the tank if necessary.
2. Check the waste level in the tank. Empty the tank if necessary.
When disposing of waste, protective clothing must be worn (lab coat, gloves, eye
protection, etc.). Follow your local and/or national guidelines for biohazard waste
disposal.
■ At the beginning of each day, before startup, check if the waste container needs to
be emptied.
■ During instrument operation, do not remove the liquid waste tube under any
circumstance.
■ Do not wash and re-use cuvettes. They are designed for single use only.
■ Cuvette segments with at least one used cuvette must never be loaded in the
reaction tray.
Dispose of used cuvettes according to your local and/or national guidelines for
biohazard waste disposal.
7. For Pentra C400 with ISE module (option), open the ISE module cover to check levels of ISE
reagents.
8. If an ISE reagent needs to be replaced, refer to the Maintenance and Troubleshooting >
Maintenance > Other Procedures > To Replace ISE Reagents (Option) chapter.
Related information:
■ To Replace ISE Reagents (Option), p.363
1. Check if the printer has enough paper for daily operations. If not, add some paper following the
instructions of the printer user guide.
If the printer does not work properly, refer to its user guide.
The Pentra C400 hardware has to remain on 24 hours a day. This is necessary to keep the cooling
unit functional and therefore maintain the temperature in the refrigerated area of the reagent tray. Only
the Pentra C400 software must be started, at the beginning of the day, either manually or
automatically if automatic startup is programmed.
Follow this procedure to start the instrument at the beginning of the day, either manually
or automatically if automatic startup is programmed.
1. If the automatic startup is programmed, the instrument is automatically started every day at the
time specified.
2. If not, press the button located on the right side of the instrument for a few seconds.
Related information:
■ To Check the Instrument, p.105
■ To Configure the Automatic Startup, p.305
■ To Log in to the Application, p.108
Follow this procedure to log in to the application with your user name and password.
1. Select your user name or login from the Name dropdown list.
2. Type your password in the Password field.
3. The New Worklist option is selected by default.
This option allows you to begin a new day of work. This means that a new worklist is created and
the previous worklist is archived.
Samples of the previous worklist are either deleted if they are ordered or pending, or archived if
they are incomplete, for validation or validated.
Sample identifiers (sample ID or rack position, depending on sample identification mode) are
cleared for the new worklist.
Related information:
■ To Start the Instrument, p.107
■ To Create a User Account, p.332
Follow this procedure to perform an ISE module activation at the beginning of the day
before running samples or during the day of work after an ISE module cleaning.
Access: Main menu > Services > Customer Services > ISE
The ISE module activation consists in running a serum or plasma sample in order to condition the
electrodes before running samples.
1. Select an available rack position on the sample tray from the Rack and Pos. dropdown lists.
2. Press ISE activation.
A dialog box asks you to place the activator at the selected rack position.
3. Place a serum or plasma sample on the sample tray at the selected rack position.
4. Press OK.
The activator is pipetted and dispensed into the ISE module sample cup. At the end, a dialog box
asks you to remove the activator from the selected rack position.
5. Remove the activator from the sample tray.
6. Press OK.
4. Reagent Status
The reagent tray is symbolized in the middle of the main menu. It allows you to visualize solutions
configured on the tray and to access their details.
The instrument checks the solutions on board (barcodes reading) during the initialization
and when closing the reagent tray cover.
The colored circle around the reagent tray indicates if the reagent tray is accessible or in progress.
■ Green when the reagent tray is accessible (solutions can be loaded/unloaded).
■ Blue when the reagent tray cover is locked.
The reagent tray contains 52 positions including 44 refrigerated at 4°C - 10°C (39°F - 50°F) and eight
positions at room temperature. The reagent tray can hold reagent cassettes, reagent racks and
calibrator/control reagent racks.
The status of each solution is indicated by a color code to which a legend is available by pressing the
blue background outside the reagent tray.
Low Volume There is less than 10% of the initial volume remaining
(only for cassettes).
To be checked The solution needs to be checked because of one of the
following reasons:
■ The cassette is empty.
■ The solution is unknown (bad barcode for example).
■ Not enough solution for the pending analyses.
■ The solution is inactive.
■ The solution has reached its use limit date or has
expired.
Available Empty position
You can access the solution details by pressing a position on the reagent tray. The information below
is displayed depending on whether the solution is in a cassette, in an open cassette, on a reagent
rack or on a calibrator/control reagent rack.
Cassette
Heading Description
Solution Name Solution short name
Position Position number on the reagent tray
Tests Available Number of tests that can be performed with the solution volume remaining
in the cassette.
Tests to be done Number of tests to be performed on the samples present in the instrument.
Expiry Date Expiration date before opening.
Open cassette
Heading Description
Solution Name Solution short name
Position Position number on the reagent tray
Reagent rack
Heading Description
Reagent Rack Rack ID
Tray Position First position number on the reagent tray
Solution Name Solution short name
Position Rack sector: A, B or C
Tests to be done Number of tests to be performed on the samples present in the instrument.
Heading Description
Reagent Rack Rack ID
Position Position number on the reagent tray
Name Calibrator/control name
Lot Number Calibrator/control lot number
Expiry Date Calibrator/control expiration date
Follow this procedure to check the status of each solution configured on the reagent tray.
1 = Reagent tray
2 = Solution requiring an additional check
1. Check the status of each solution configured on the reagent tray from the main menu.
The status of each solution is indicated by a color code on the reagent tray. The solutions
appearing in red require an additional check.
2. Physically check the solutions in red and replace them if necessary.
The colored circle around the reagent tray on the main menu must be green. This
means that the reagent tray is accessible.
When you use reagent racks or calibrator/control reagent racks, you must ensure that a
sufficient solution volume without air bubbles or foam is placed in the instrument.
Related information:
■ Reagent Tray Description, p.111
Every day before running samples, you have to perform a new calibration for the tests that require one
and you have to check all of your tests with the appropriate control materials.
The ISE module remains on 24 hours a day and is calibrated regularly. At the beginning of each day
before running samples, you have to check the ISE module status.
2. In case of error on the ISE module, perform a global initialization of the ISE module as follows:
a. Enter Main menu > Services > Customer Services > ISE.
b. Press Global Initialization.
A global initialization of the ISE module is performed. It takes a few minutes.
If the error persists, refer to the Maintenance and Troubleshooting > Troubleshooting > ISE Module
(Option) chapter.
3. In case of error on the ISE calibration, run a new ISE calibration.
Refer to the ISE Calibration (Option) > To Run an ISE Calibration chapter.
Related information:
■ To Enable or Disable ISE Module, p.308
■ To Run an ISE Calibration, p.115
■ ISE Module (Option), p.386
Follow this procedure to run a 2-point calibration in case of error on the ISE calibration.
For detailed information concerning ISE calibration results interpretation, refer to the
ISE Calibration (Option) > ISE Calibration Results chapter.
3. Press Calibrate.
The 2-point calibration is performed. It takes a few minutes.
4. Wait for the end of the 2-point calibration.
The new ISE calibration results are displayed.
5. Make sure that no ISE analytical flag has been triggered in 1 Point Cal Error and 2 Point Cal
Error fields.
6. If an error persists on the ISE calibration, refer to the Maintenance and Troubleshooting >
Troubleshooting > ISE Module (Option) > Error on the ISE Calibration chapter.
Related information:
■ Error on the ISE Calibration, p.388
■ ISE Calibration Results, p.116
The ISE calibration results screen displays the date and time of the last ISE calibration (at the top of
the screen) as well as the last ISE calibration results for each electrode.
Heading Description
Slope Slope (A) of the regression line y = A * x + B
Intercept Intercept (B) of the regression line y = A * x + B
1 Point Cal Error ISE analytical flag on the low standard value.
mV(1) Low standard value in mV.
2 Point Cal Error ISE analytical flag on the high standard value.
mV(2) High standard value in mV.
For detailed information concerning ISE analytical flags, refer to the Alarms > Analytical
and Quality Flags > ISE Analytical Flags chapter.
The expected values of the slope and the sensitivity (difference between the low and high standard
values) for each electrode are given in the following table:
Related information:
■ ISE Analytical Flags (Option), p.445
The calibration worklist lists all calibration requests that are ordered, pending or incomplete. As soon
as samplings of an incomplete calibration are finished, the request is deleted from the worklist. It is
then transferred to the result list.
Calibration requests are sorted by date and are listed in a table format which gives the following
information.
Heading Description
Status Calibration status: Ordered, Pending or Incomplete
Test Application local code for the test to be calibrated
Position Rack position of calibrator
Name Calibrator name
Lot Calibrator lot number
Order Date Request date
The Calibration tab allows you to add and to delete calibration requests.
Related information:
■ To Add Calibration Requests, p.118
■ To Delete Calibration Requests, p.119
The test is not requested and the The test cannot be requested (except from
corresponding reagent is inactive. the host).
The time lapse must be previously programmed in Main menu > Services >
System Configuration > Analyser.
3. If the All calibrations expired only or All calibrations expired in window option is chosen, the
tests that require a new calibration are selected by default. You can deselect the tests that you do
not wish to calibrate.
4. If the All parameters option is chosen, select the tests that you want to calibrate.
For detailed information concerning tests selection, refer to the Adding Calibration
Requests > Tests Selection in the Request Capture Screen chapter.
5. Press OK to validate.
■ For tests using the Factor calibration mode, a reagent blank is carried out.
For tests using the other calibration modes, a reagent blank is performed with the
calibration if configured in the application.
■ If, for a test, the Control required option is selected in the application and the
required default controls are configured, then the default controls are automatically
performed with the calibration.
Related information:
■ To Program the Time Lapse for Displaying Calibrations that Will Expire, p.310
■ Tests Selection in the Request Capture Screen, p.117
The control worklist lists all control requests that are ordered, pending or incomplete. As soon as
samplings of an incomplete control are finished, the request is deleted from the worklist. It is then
transferred to the result list.
Control requests are sorted by date and are listed in a table format which gives the following
information.
Heading Description
Status Control status: Ordered, Pending or Incomplete
Test Application local code for the test to be controlled
Position Rack position of control
Name Control name
Lot Control lot number
Order Date Request date
The Control tab allows you to add and to delete control requests.
Related information:
■ Adding Control Requests, p.120
■ To Delete Control Requests, p.123
The test is not requested and the The test cannot be requested (except from
corresponding reagent is inactive. the host).
If, for a test, the Control required option is selected in the application and the required
default controls are configured, then the default controls are automatically performed with
the calibration.
■
Press Select all to select all the tests or Erase to deselect all the
tests.
■ For detailed information concerning tests selection, refer to the Adding Control
Requests > Tests Selection in the Request Capture Screen chapter.
4. Press OK to validate.
If several default controls are configured for a test, then a control request is created for
each default control.
Related information:
■ Tests Selection in the Request Capture Screen, p.117
The control name, the rack position of control, the control lot number and expiration date are
displayed.
■
Press Select all to select all the tests or Erase to deselect all the
tests.
■ For detailed information concerning tests selection, refer to the Adding Control
Requests > Tests Selection in the Request Capture Screen chapter.
4. For each selected test, configure the number of replicates (from 1 to 20 replicates) as follows:
a. Select the test.
b. Press the up or down arrows (right side of the screen) to increase or to decrease the
number of replicates.
The number of replicates is set to 1 by default.
5. Press OK to validate.
Related information:
■ Tests Selection in the Request Capture Screen, p.117
Follow this procedure to place calibrators and controls on the sample tray and/or on the
reagent tray.
Ensure that a sufficient sample volume without air bubbles or foam is placed in the
instrument, whatever the sample type (calibrator, control, patient sample).
1 = Configured positions
2. Place the required calibrators on racks according to the configured positions.
3. Press the Control tab to view the control worklist.
4. Place the required controls on racks according to the configured positions.
5. Load the racks on the sample tray and/or on the reagent tray.
■ Before loading racks on the sample tray, ensure that the sample tray LED is green.
If not, press Pause to stop samplings on the sample tray and wait for the LED to
turn green.
■ If automatic reruns must be performed, the instrument turns to "Sampling" status
and the LED turns to orange for a few seconds before turning red. Never load
racks on the sample tray when the LED is orange or red.
Before loading racks on the reagent tray, make sure that the colored circle around the
reagent tray on the main menu is green. This means that the reagent tray is accessible.
Related information:
■ Sample Type and Volume, p.102
The Work Balance menu allows you to assess the reagents on board relative to the analyses to be
performed.
This menu is divided into two tabs:
■ Reagent supply
■ Sampling Exception
The Reagent supply tab lists the solutions necessary for the pending tests.
You can display either the pending tests for the samples present on the instrument or those for the
samples programmed in the worklist.
■ Tubes on tray: To display the pending tests for the samples present on the instrument.
■ Worklist: To display the pending tests for the samples programmed in the worklist.
Heading Description
Solution requested Solution short name
Capacity requested ■ Cassette: Number of tests to be performed.
■ Open cassette or reagent rack: Solution volume necessary for the tests
to be performed (without taking into account the dead volume).
Present on reagent tray ■ Reagent: Number of tests that can be performed with the solution
volume remaining in the cassette.
■ Diluent or cleaner: Solution volume remaining in the cassette.
Only for cassette.
Lot # Solution lot number
Position ■ Cassette or open cassette: Position number on the reagent tray.
■ Reagent rack: First position number on the reagent tray + rack sector.
Heading Description
Status Solution status: Active or Inactive
On tray If selected, the solution is present on the reagent tray.
The Host request button allows you to manually ask the host if there are
additional tests programmed for a validated patient sample. This function is available only
if the Sending validated samples option is enabled in Main menu > Services > System
Configuration > Host Connection.
Related information:
■ To Configure the Host Connection, p.319
The Sampling Exception tab displays the pending tests for which a sampling alarm will be triggered.
You can display either the pending tests for the samples present on the instrument or those for the
samples programmed in the worklist.
■ Tubes on tray: To display the pending tests for the samples present on the instrument.
■ Worklist: To display the pending tests for the samples programmed in the worklist.
The tests are listed in a table format which gives the following information:
Heading Description
Sample ID ■ Blank: "BLANK" on yellow background.
■ Calibration: Calibrator name on yellow background.
■ Control: Control name on green background.
■ Patient sample: Sample ID.
Position Rack position of the sample tube
Test Application local code for the test
Sampling Alarms Sampling alarm
To display additional information concerning the sampling alarm.
Sampling Alarm Date Sampling alarm date and time
The Host request button allows you to manually ask the host if there are
additional tests programmed for a validated patient sample. This function is available only
if the Sending validated samples option is enabled in Main menu > Services > System
Configuration > Host Connection.
Related information:
■ To Configure the Host Connection, p.319
Follow this procedure to assess the reagents on board relative to the analyses to be
performed.
The instrument checks the solutions on board (barcodes reading) on reagent and sample trays. It
takes a few minutes.
The colored circle around the reagent tray on the main menu must be green. This
means that the reagent tray is accessible.
Related information:
■ Reagent supply, p.125
■ Sampling Exception, p.126
■ Sampling Alarms, p.420
1. Press Start.
Before performing the analyses, the reagent tray temperature is checked and if it is out of range, a
dialog box is displayed: The current reagent tray temperature [temperature] is out of the range
(acceptable range). The results can be affected by this malfunction. We recommend you to perform
controls. Do you still want to run analysis?
This message means that the reagents refrigeration does not work properly and this
can generate erroneous test results.
■ Press Cancel to cancel the analyses without any warning. This will allow you to perform the
appropriate controls before performing the analyses.
■ Press OK to perform the analyses in spite of this message. The analyses are performed and
the warning #642 Worklist performed with reagent tray temperature [temperature] out of the
range (acceptable range) is triggered.
The instrument turns to the "Sampling" status, then to the "Analysing" status.
1 = Instrument status
Related information:
■ To Check the Instrument, p.105
The Test Review menu allows you to review the tests in progress, to check sampling alarms and
analytical flags that may have been triggered and to manually export results to a USB key at each end
of run.
This menu is divided into four tabs:
■ Test Result
■ Sampling Exception
■ Error Report
■ Sequencing
The Test Result tab displays the tests in progress as soon as they are pipetted.
The tests are sorted by run date.
■ Up to 200 items are displayed from the oldest to the most recent. Beyond this number,
the oldest are deleted.
■ The tab is cleared at instrument startup, when starting analyses and when a new
worklist is created.
The tests are listed in a table format which gives the following information:
Heading Description
Status Ordered, Pending, Incomplete, For validation or Validated
For Ordered and Pending status, at least one test of the sample is finished.
Sample ID ■ Blank: "BLANK" on yellow background.
■ Calibration: Calibrator name on yellow background.
■ Control: Control name on green background.
■ Patient sample: Sample ID.
Position Rack position of the sample tube
Test Application local code for the test
Result(s) ■ Blank and Calibration: Absorbance value.
■ Control: Final result. On red background if out of confidence range.
■ Patient sample: Final result.
"--.-----" means that the result is not given because of a sampling alarm or
an analytical flag.
QF Quality flag (only the flag having the highest priority is displayed).
D/C The sample was automatically rerun with post-dilution (Dil) or post-
concentration (Conc).
Only for patient sample.
Analysis flag Sampling alarm or analytical flag (only the flag having the highest priority is
displayed).
If both are triggered, only the sampling alarm is displayed.
Dec. Decision:
■ Blank: Accept or Delete.
■ Calibration: Not displayed.
■ Control: Accept or Delete.
■ Patient sample: Accept, Rerun, Modify or Delete.
Unit Test unit configured in the application
Run date Sample run date and time
You can access the sample details by selecting a test and pressing Details.
You can also search for a sample either by sample ID or by position using the search tools at the
bottom of the screen.
The Export button allows you to manually export results to a USB key. Refer to
the Test Review > To Manually Export Results to a USB Key chapter.
Related information:
■ To Manually Export Results to a USB Key, p.137
The Sampling Exception tab displays the tests for which a sampling alarm is triggered.
The tests are listed in a table format which gives the following information:
Heading Description
Sample ID ■ Blank: "BLANK" on yellow background.
■ Calibration: Calibrator name on yellow background.
■ Control: Control name on green background.
■ Patient sample: Sample ID.
Position Rack position of the sample tube
Test Application local code for the test
Sampling Alarms Sampling alarm
To display additional information concerning the sampling alarm.
Sampling Alarm Date Sampling alarm date and time
Run date Sample run date and time
You can access the sample details by selecting a test and pressing Details.
You can also search for a sample either by sample ID or by position using the search tools at the
bottom of the screen.
The Error Report tab displays the tests for which an analytical flag or a quality flag blocking the
automatic validation is triggered.
The tests are sorted by run date.
The tests are listed in a table format which gives the following information:
Heading Description
Sample ID ■ Blank: "BLANK" on yellow background.
■ Calibration: Calibrator name on yellow background.
■ Control: Control name on green background.
■ Patient sample: Sample ID.
Position Rack position of the sample tube
Test Application local code for the test
Result(s) ■ Blank and Calibration: Absorbance value.
■ Control: Final result. On red background if out of confidence range.
■ Patient sample: Final result.
"--.-----" means that the result is not given because of a sampling alarm or
an analytical flag.
QF Quality flag (only the flag having the highest priority is displayed).
D/C The sample was automatically rerun with post-dilution (Dil) or post-
concentration (Conc).
Only for patient sample.
Analysis flag Analytical flag (only the flag having the highest priority is displayed).
Unit Test unit configured in the application
Run date Sample run date and time
You can access the sample details by selecting a test and pressing Details.
You can also search for a sample either by sample ID or by position using the search tools at the
bottom of the screen.
5.6.4. Sequencing
■ Up to 200 items are displayed from the oldest to the most recent. Beyond this number,
the oldest are deleted.
■ The tab is cleared at instrument startup, when starting analyses and when a new
worklist is created.
Heading Description
Date Sampling date and time
Nature Sampling nature:
■ for a test: Application local code,
■ for a dilution: Dilution,
■ for a cleaning: Cleaner.
Reagent needle Solution pipetted by the reagent needle
Volume Volume pipetted by the reagent needle (µL)
Heading Description
Sampling position Position where the solution is pipetted:
■ on the reagent tray: R (for reagent tray) + position number on the tray +
position number in the cassette or on the reagent rack.
Type Container type where the solution is pipetted:
■ for a cassette: Cassette,
■ for a reagent rack: Rack ID + position number on the reagent rack.
Dispensing position Position where the solution is dispensed:
■ on the reaction tray: C (for cuvette) + position number on the tray +
position number in the cuvette segment.
Sample Arm
Heading Description
Date Sampling date and time
Nature Sampling nature:
■ for a test: Application local code,
■ for an ISE test: ISE,
■ for a dilution: Dilution,
■ for a cleaning: Cleaner.
Sample needle Solution pipetted by the sample needle
Volume Volume pipetted by the sample needle (µL)
H2O Distilled water volume pipetted by the sample needle (µL) to push the sample
volume
Sampling position Position where the solution is pipetted:
■ on the sample tray: S (for sample tray) + rack number + position number
on the rack,
■ on the reagent tray: R (for reagent tray) + position number on the tray +
position number in the cassette or on the reagent rack,
■ on the reaction tray: C (for cuvette) + position number on the tray +
position number in the cuvette segment.
Type Container type where the solution is pipetted:
■ for a sample tube: Tube,
■ for a sample cup: Sample cup,
■ for a cassette: Cassette,
■ for a reagent rack or a calibrator/control reagent rack: Rack ID + position
number on the reagent rack,
■ for a cuvette: Cuvette.
Dispensing position Position where the solution is dispensed:
■ on the reaction tray: C (for cuvette) + position number on the tray +
position number in the cuvette segment,
■ in the ISE module sample cup: ISE bowl.
Other
Heading Description
Date Date and time of mixing or cuvette segment loading/unloading
Mixing speed Mixing speed used to homogenize solutions into the cuvette
Cuvette loader Cuvette segment loading/unloading: Loading or Unloading + position
number on the tray
Follow this procedure to check sampling alarms and analytical flags that may have been
triggered.
The Test Review button becomes red when a sampling alarm or an analytical flag is triggered.
The Sampling Exception tab displays the tests for which a sampling alarm is triggered.
Related information:
■ Sampling Exception, p.132
■ Error Report, p.133
■ Analytical and Quality Flags, p.423
■ Sampling Alarms, p.420
Follow this procedure to manually export results to a USB key at the end of run.
Results are automatically exported to a USB key at each end of run if the Activate export
of the results on USB key option is selected in Main menu > Services > System
Configuration > Analyser. Refer to the Settings > System Configuration > Analyser > To
Enable or To Disable Automatic Export of Results chapter.
HORIBA Medical recommends you to make sure that all materials like CD-ROM or USB
key used on the Pentra C400 and non-referenced by HORIBA Medical are free of
viruses. These materials must imperatively be cleaned before being used on the
instrument.
All the completed test results are saved in a .csv file located in a folder named "Export" on the USB
key. ISE test results and ratio results are not exported.
Related information:
■ To Enable or To Disable Automatic Export of Results, p.310
The calibration result list displays all calibration results that are incomplete, for validation or validated.
You can select the calibration results to be displayed from the following options.
■ For validation: To display the calibration results that are incomplete or for validation.
■ Current: To display the calibration results that are validated.
■ All: To display all calibration results.
Calibration results are listed in a table format which gives the following information.
Heading Description
Status Calibration status: Incomplete, For validation or Validated
Test Application local code for the calibrated test
Position ■ Blank: Reagent position
■ Calibration: Rack position of calibrator
Name ■ Blank: "BLANK"
■ Calibration: Calibrator name
Lot ■ Blank: Reagent lot number
■ Calibration: Calibrator lot number
Flag Calibration analytical flag.
QF Quality flag (only the flag having the highest priority is displayed).
Run Date Date and time of calibration run
■ For tests using the Factor calibration mode, a reagent blank is carried out.
■ An item in red means that the calibration is flagged with a calibration analytical flag or
a quality flag blocking the automatic validation.
The Calibration tab allows you to validate, to rerun or to delete a calibration or a reagent blank.
For detailed information concerning the procedure to check calibration results, refer to the
Calibration Results > To Validate a Calibration chapter.
Calibration results
The result validation screen displays the new calibration results as well as the current calibration
results (at the bottom of the screen).
New calibration results
Heading Description
Test Application local code for the calibrated test
Unit Test unit configured in the application
Position Rack position of calibrator
Name Calibrator name
Lot Calibrator lot number
Calibration Flag Calibration analytical flag.
Displayed in red.
New Calibration Date Date and time of calibration run
New blank (ΔO.D.) Absorbance value of reagent blank (if configured in the application).
Displayed in red if flagged.
QF Quality flag (only the flag having the highest priority is displayed).
Target Target value for each calibrator level
Run1 Absorbance value of the first run for each calibrator level.
Displayed in red if flagged.
Run2 Absorbance value of the second run (if configured in the application) for
each calibrator level.
Displayed in red if flagged.
Run3 Absorbance value of the third run (if configured in the application) for each
calibrator level.
Displayed in red if flagged.
Dev_Rep Deviation concerning replicates in %.
Displayed in red if it exceeds the authorized value.
"--.--" means that the deviation cannot be calculated because only one run
is configured.
Dev_C Deviation to the curve in %.
Displayed in red if it exceeds the authorized value.
"--.--" means that the deviation cannot be calculated because only one
calibrator level is configured.
■ If several runs are configured, you can deactivate a defective result by double-clicking
the result. The result is then displayed in gray and the calibration results are updated.
■ For detailed information concerning analytical flags or quality flags, refer to the Alarms
> Analytical and Quality Flags chapter.
Heading Description
In Progress Calibration Date Date and time of calibration run
Previous blank (ΔO.D.) Absorbance value of reagent blank (if configured in the application).
Displayed in red if flagged.
Run1 Mean absorbance value for each calibrator level.
Calibration curves
You can display the calibration curves by selecting a result in the new calibration results table.
The graph (right side of the screen) displays the following data:
■ X-axis: concentration,
■ Y-axis: rate,
■ Green continuous line: new calibration curve,
■ Blue dotted line: current calibration curve.
■ The calibration curve points are displayed in green for the new calibration curve or in blue for the
current calibration curve. Click a point to display the point concentration and rate.
■ For linear calibration modes (Slope average, Linear regression), the calculated calibration factor is
displayed in green for the new calibration curve or in blue for the current calibration curve.
■ For the Linear regression calibration mode, the intercept (B) of the regression line y = A * x + B is
displayed in green for the new calibration curve or in blue for the current calibration curve.
Drag your mouse pointer from left to right to zoom in or from right to left to zoom out.
The graph (right side of the screen) displays the following data:
■ X-axis: time,
■ Y-axis: absorbance value,
■ Purple line: kinetic curve of the first run,
■ Yellow line: kinetic curve of the second run,
■ Green line: kinetic curve of the third run.
Drag your mouse pointer from left to right to zoom in or from right to left to zoom out.
The graph (right side of the screen) displays the following data:
■ X-axis: time,
■ Y-axis: absorbance value,
■ Green or blue lines: kinetic curves for each calibrator level as well as for the reagent blank (if
configured in the application).
■ If the absorbance deviation check on linear regression is configured in the application, the
correlation coefficient r2 is calculated and displayed for each calibrator level. Click a kinetic curve
to display the corresponding correlation coefficient r2.
Drag your mouse pointer from left to right to zoom in or from right to left to zoom out.
Data of kinetic curves for each calibrator level as well as for the reagent blank (if configured in the
application) are also displayed in a table format (left side of the screen).
If analytical flags are triggered, they are displayed in a second table.
Related information:
■ Analytical and Quality Flags, p.423
■ Calibration Parameters, p.260
■ Calculation Parameters, p.267
■ To Validate a Calibration, p.142
If the calibration automatic validation is configured in Main menu > Services > System
Configuration > Results validation, calibration results are automatically validated on condition that:
■ No point used for calculation is flagged (except with the analytical flag CALC_RANGE_HIGH /
LOW).
■ For each calibrator level, deviations (Dev_Rep and Dev_C) do not exceed the authorized values.
■ The calibration factor is not out of the programmed values.
■ The reagent blank (if configured in the application) is not flagged.
1. Select the For validation option to display the calibration results that are incomplete or for
validation.
2. Select the calibration results that you want to review and press Details.
This function is unavailable if the selected calibration is incomplete.
The result validation screen is displayed.
4 = Edit
5 = Accept
6 = OK
3. Check the calibration results and make sure that no analytical flag or quality flag has been
triggered.
4. Press Edit.
5. Select Accept and press OK to validate.
■ When the calibration cannot be calculated, it cannot be validated and must be deleted. If
several runs are configured, you can deactivate a defective result in order to calculate the
calibration.
■ OK is unavailable if a default control for this test is in progress.
Related information:
■ To Configure Calibrations Automatic Validation, p.326
■ To Rerun a Calibration or a Reagent Blank, p.147
■ To Delete a Calibration or a Reagent Blank, p.148
■ Calibration Results Interpretation, p.138
■ Analytical and Quality Flags, p.423
For detailed information concerning the procedure to check reagent blank results, refer to
the Calibration Results > To Validate a Reagent Blank chapter.
Heading Description
Test Application local code for the calibrated test
Unit Test unit configured in the application
Run Date Date and time of calibration run
Expiration date Reagent expiration date.
Only for cassette.
Position Reagent position
Lot Reagent lot number
Flag Calibration analytical flag.
QF Quality flag (only the flag having the highest priority is displayed).
Report Absorbance value of the new reagent blank
Previous Report Absorbance value of the previous reagent blank
For detailed information concerning analytical flags or quality flags, refer to the Alarms >
Analytical and Quality Flags chapter.
Kinetic curve
You can display the reagent blank kinetic curve by pressing the OD tab.
The graph (right side of the screen) displays the following data:
■ X-axis: time,
■ Y-axis: absorbance value,
■ Green line: reagent blank kinetic curve.
Data of the reagent blank kinetic curve are also displayed in a table format (left side of the screen).
If analytical flags are triggered, they are displayed in a second table.
Related information:
■ Analytical and Quality Flags, p.423
■ To Validate a Reagent Blank, p.145
If calibrations automatic validation is configured in Main menu > Services > System Configuration >
Results validation, reagent blank results are automatically validated on condition that they are not
flagged.
1. Select the For validation option to display the calibration results that are incomplete or for
validation.
2. Select the reagent blank results that you want to review and press Details.
This function is unavailable if the selected reagent blank is incomplete.
The result validation screen is displayed.
4 = Edit
5 = Accept
6 = OK
3. Check the reagent blank results and make sure that no analytical flag or quality flag has been
triggered.
■ For detailed information concerning reagent blank results interpretation, refer to the
Calibration Results > Reagent Blank Results Interpretation chapter.
■ In case an analytical flag or a quality flag has been triggered, do not validate the
reagent blank. Refer either to the Calibration Results > To Rerun a Calibration or a
Reagent Blank chapter to rerun the reagent blank or to the Calibration Results > To
Delete a Calibration or a Reagent Blank chapter to delete the reagent blank.
4. Press Edit.
5. Select Accept and press OK to validate.
■ If the reagent blank is flagged with HIGH_ABS, it cannot be validated and must be deleted.
■ OK is unavailable if a default control for this test is in progress.
Related information:
■ To Configure Calibrations Automatic Validation, p.326
■ To Rerun a Calibration or a Reagent Blank, p.147
■ To Delete a Calibration or a Reagent Blank, p.148
■ Reagent Blank Results Interpretation, p.144
■ Analytical and Quality Flags, p.423
1. Select the For validation option to display the calibration results that are incomplete or for
validation.
2. Select the calibration that you want to rerun and press Rerun:
This function is unavailable if the calibration is incomplete (unless it is flagged with a sampling
alarm) or validated.
A dialog box informs you that the calibration will be deleted and a new calibration will be ordered.
3. Press OK to validate or Cancel to cancel.
■ In case the calibration is deleted, the previous calibration will be the current calibration
and all following results are flagged with C (CAL_ERROR).
■ If default controls are required for this calibration, results are calculated using the
current calibration. If no current calibration exists, the default controls are deleted.
1. Select the For validation option to display the calibration results that are incomplete or for
validation.
2. Select the calibration that you want to delete and press Delete.
This function is unavailable if the calibration is incomplete or validated or if a default control for this
test is in progress.
■ In case the calibration is deleted, the previous calibration will be the current calibration
and all following results are flagged with C (CAL_ERROR).
■ If default controls are required for this calibration, results are calculated using the
current calibration. If no current calibration exists, the default controls are deleted.
Calibration results are automatically printed when validated if configured in Main menu > Services >
System Configuration > Printer.
1. Select the Current option to display the calibration results that are validated.
2. Select the calibration results that you want to print and press Data Print/Send to Host from the
generic toolbar.
The Print window is displayed.
3. Select the printout format from the following options:
■ standard,
■ detailed.
If reagent blank results are selected, the printout format is not selectable. The reagent
blank kinetic curve will be printed by default.
4. Press OK to validate.
The selected calibration results are printed.
On the printouts, flagged results are indicated by an asterisk (*) in the column marked
F.
Related information:
■ To Configure Automatic Printouts, p.323
Follow this procedure to modify some application parameters in order to convert the
current calibration results as well as the associated patient and control results.
1. Select the Current option to display the calibration results that are validated.
2. Select the calibration results that you want to convert and press Details.
The result validation screen is displayed.
c. Press OK to validate.
5. Once you modified the application parameters, press the Calibration tab to review the converted
calibration results.
6. Press the Patients and Controls tab to review the converted patient and control results.
■ Only the patient and control results that are for validation are displayed.
■ Ratio results and test results used in ratios are not displayed.
7. Modify some patient and control results by the converted results as follows:
a. Select a result.
b. Press OK.
c. Repeat steps a and b for each result.
The converted results are then flagged with R (CAL_CONV).
Related information:
■ Absorbance Applications Parameters, p.253
The control result list displays all control results that are incomplete, for validation or validated.
Control results are sorted per control block. A control block contains either all the required default
controls for a test (up to three default controls can be configured for a test) or a single control if the
control is not configured as a default control for the test. Alternation of white and blue materializes the
separation between control blocks.
You can select the control results to be displayed from the following options.
■ Block for validation: To display the control results that are incomplete or for validation.
■ Current block: To display the control results that are validated.
Control results are listed in a table format which gives the following information.
Heading Description
Status Control status: Incomplete, For validation or Validated
Test Application local code for the controlled test
Name Control name
Run date Date and time of control run
Lot Control lot number
Position Rack position of control
Flag Sampling alarm or analytical flag (only the flag having the highest priority is
displayed).
If both are triggered, only the sampling alarm is displayed.
QF Quality flag (only the flag having the highest priority is displayed).
An item in red means that the control is flagged with an analytical flag or a quality flag
blocking the automatic validation.
For detailed information concerning the procedure to check control results, refer to the
Control Results > To Validate a Control chapter.
Control results
Heading Description
Name Control name
Lot Control lot number
Test Application local code for the controlled test
Unit Test unit configured in the application
Position Rack position of control
Run Date Date and time of control run
Report Final result. On red background if out of confidence range.
"--.-----" means that the result is not given because of a sampling alarm or
an analytical flag.
Target Value Control target value
Confidence Range Low and high limit values of confidence range
Flag Sampling alarm or analytical flag (only the flag having the highest priority is
displayed).
If both are triggered, only the sampling alarm is displayed.
Quality flag Quality flag (only the flag having the highest priority is displayed).
Decision Accept or Delete
For detailed information concerning analytical flags or quality flags, refer to the Alarms >
Analytical and Quality Flags chapter.
Kinetic curve
You can display the control kinetic curve by pressing the OD tab.
The graph (right side of the screen) displays the following data:
■ X-axis: time,
■ Y-axis: absorbance value,
■ Green line: control kinetic curve.
■ If the absorbance deviation check on linear regression is configured in the application, the
correlation coefficient r2 is calculated and displayed.
Data of the control kinetic curve are also displayed in a table format (left side of the screen).
If analytical flags are triggered, they are displayed in a second table.
Related information:
■ Analytical and Quality Flags, p.423
■ Calculation Parameters, p.267
■ To Validate a Control, p.152
If controls automatic validation is configured in Main menu > Services > System Configuration >
Results validation, control results are automatically validated on condition that they are not flagged
(except with the analytical flag CONF_RANGE_HIGH_W / LOW_W).
1. Select the Block for validation option to display the control results that are incomplete or for
validation.
2. Select the control results that you want to review and press Details.
This function is unavailable if the selected control is incomplete or, for a control block, if at least
one default control of the selected block is incomplete.
The result validation screen is displayed.
4 = Edit
5 = Accept
6 = OK
3. Check the control results and make sure that no analytical flag or quality flag has been triggered.
4. Press Edit.
5. Select Accept and press OK to validate.
When the control result cannot be calculated, it cannot be validated and must be deleted.
Related information:
■ To Configure Controls Automatic Validation, p.326
■ To Rerun a Control, p.154
■ To Delete a Control, p.155
■ Control Results Interpretation, p.151
■ Analytical and Quality Flags, p.423
1. Select the Block for validation option to display the control results that are incomplete or for
validation.
2. Select the control that you want to rerun and press Rerun:
1. Select the Block for validation option to display the control results that are incomplete or for
validation.
2. Select the control that you want to delete and press Delete.
This function is unavailable if the control is incomplete or validated.
However, an incomplete control can be deleted if it is flagged with a sampling alarm or if the
instrument is in "Ready" or "Emergency Stop" status.
When you delete a control result, this one is excluded from the statistical analyses but it is
still present in the Quality Control.
Control results are automatically printed when validated if configured in Main menu > Services >
System Configuration > Printer.
1. Select the Block for validation option to display the control results that are incomplete or for
validation or the Current block option to display the control results that are validated.
2. Select the control results that you want to print and press Data Print/Send to Host from the
generic toolbar.
This function is unavailable if the selected control is incomplete or, for a control block, if at least
one default control of the selected block is incomplete.
The Print window is displayed.
3. Select the printout format from the following options:
■ condensed,
■ standard,
■ detailed.
4. Press OK to validate.
The selected control results are printed.
Related information:
■ To Configure Automatic Printouts, p.323
Control results are automatically sent by test when validated if configured in Main menu > Services >
System Configuration > Host Connection.
1. Select the Current block option to display the control results that are validated.
2. Select the control results that you want to send and press Data Print/Send to Host from the
generic toolbar.
This function is unavailable if the selected control is incomplete or, for a control block, if at least
one default control of the selected block is incomplete.
The Print window is displayed.
3. Press Send.
The Send window is displayed.
4. Press OK to validate.
The selected control results are sent.
Related information:
■ To Configure the Host Connection, p.319
6. Patient Samples
The patient worklist lists all patient sample requests that are ordered, pending or incomplete. As soon
as samplings of an incomplete patient sample are finished, the request is deleted from the worklist. It
is then transferred to the result list.
Patient sample requests are sorted by patient ID and by date. Alternation of white and blue
materializes the separation between patient ID.
Patient sample requests are listed in a table format which gives the following information.
Heading Description
Status Sample status: Ordered, Pending or Incomplete
Sample ID Sample ID
Position Rack position of the sample tube
Order Date Request date
Patient ID Patient ID
Last Name Patient last name
First Name Patient first name
The Patient tab allows you to add, to modify and to delete patient sample requests.
Related information:
■ Creating a Patient Worklist, p.161
■ To Modify a Patient Sample Request, p.168
■ To Delete Patient Sample Requests, p.169
Patient demographic data is registered in patient files. Each patient file is identified by a unique
patient number named patient ID. This patient ID is either captured by the operator or automatically
assigned by the instrument: "AUTO_PIDXXXXXXXX".
■ If the operator enters a patient ID that already exists, the patient demographic data is
displayed.
■ At least one field of the patient demographic data must be captured so that the
instrument automatically assigns a patient ID.
The patient demographic data can be modified from the instrument only if it was manually entered
from the instrument. If it was programmed from the host, it must be modified from the host. When a
patient file is modified (either from the instrument or from the host), all validated and archived sample
ID for this patient are flagged with an asterisk.
Patient sample requests are identified in the worklist by a unique sample number named sample ID.
This sample ID is either captured by the operator or automatically assigned by the instrument:
"AUTO_SIDXXXXXXXX".
If the operator enters a sample ID that already exists in the worklist, all information about
this sample ID is displayed.
As soon as the analysis process begins on a sample tube, the associated request is searched in the
worklist upon two criteria:
■ the sample ID (identification mode),
■ the rack position of the sample tube (position mode).
When the corresponding request is found, the instrument automatically performs the analyses
requested in the worklist. As soon as samplings are finished, the request is deleted from the worklist.
It is then transferred to the result list.
The association between a sample tube and a request can be done only on the two following
conditions.
■ The sample tube must be present on the instrument.
If a request exists in the worklist but the associated sample tube is not found on the instrument,
the sample tube and the request are unmatched and the request stays pending.
■ The request must exist in the worklist.
If a sample tube is read on the tray and no associated request is found in the worklist, the sample
Identification mode
The identification mode is recommended on the Pentra C400 as it ensures security and flexibility.
As soon as the instrument reads a sample tube barcode, the associated sample ID is searched in the
worklist.
For example, the read barcode is "123" and the sample ID "123" is found in the worklist. The sample
tube and the request are matched. The instrument fills in the rack position of sample tube in the
worklist.
The identification mode is made to use the sample ID, however it is possible to force a rack position
from the request capture screen (refer to the Patient Worklist > Creating a Patient Worklist > Patient
Sample Request Fields chapter). In that case, the following cases can occur.
1. For example, the rack position "01.01" is forced in the worklist and a tube without barcode is
found on this position. The sample tube and the request are matched.
2. The rack position "01.01" is forced in the worklist and a tube with barcode is found on this
position.
a. The sample ID was not entered by the operator, it is then filled in by the instrument. The
sample ID is identical to the read barcode. The sample tube and the request are matched.
b. The sample ID was not entered by the operator, it is then filled in by the instrument. The
sample ID is different from the read barcode. The sample tube and the request are
mismatched.
c. The sample ID was entered by the operator and it is identical to the read barcode. The sample
tube and the request are matched.
d. The sample ID was entered by the operator and it is different from the read barcode. The
sample tube and the request are mismatched.
When a rack position is forced, you must ensure that the sample tube is physically
positioned as programmed in the worklist.
Position mode
Samples are identified by rack position. In this mode, the rack position of the sample tube must be
filled in by the operator. The sample ID can be entered by the operator but it is not mandatory.
Working with the position mode, you must ensure that the sample tube is physically
positioned as programmed in the worklist.
As soon as the instrument reads a sample tube, the associated rack position is searched in the
worklist. The following cases can occur.
1. For example, a tube without barcode is read on the rack position "01.01" and this position is
found in the worklist. The sample tube and the request are matched.
2. A tube with barcode is read on the rack position "01.01" and this position is found in the worklist.
a. The sample ID was not entered by the operator, it is then filled in by the instrument. The
sample ID is identical to the read barcode. The sample tube and the request are matched.
b. The sample ID was not entered by the operator, it is then filled in by the instrument. The
sample ID is different from the read barcode. The sample tube and the request are
mismatched.
c. The sample ID was entered by the operator and it is identical to the read barcode. The sample
tube and the request are matched.
d. The sample ID was entered by the operator and it is different from the read barcode. The
sample tube and the request are mismatched.
Normally, barcodes are not used in this mode. However, when a tube with barcode is
read, the barcode has priority.
Related information:
■ Patient Sample Request Fields, p.161
■ To Configure Loading Mode, p.302
Sample characteristics
In order to ensure security of patient identification and sample identification, the space
character is allowed in the patient ID and the sample ID only when this one is placed
inside the capture. The space character is not authorized and is systematically deleted
when this one is the first character or the last character of the capture.
This rule also applies to the host connection and the sample barcode types Code 39 and
Code 128.
Also refer to the Settings > System Configuration > Host Connection > To Configure the
Host Connection chapter.
Related information:
■ To Configure the Host Connection, p.319
The test is not requested and the The test cannot be requested (except from
corresponding reagent is inactive. the host).
The sample identification mode must be previously configured in position mode. Refer to the Settings
> System Configuration > Analyser > To Configure Loading Mode chapter.
Working with the position mode, you must ensure that the sample tube is physically
positioned as programmed in the worklist.
For detailed information concerning patient sample request fields, refer to the Creating
a Patient Worklist > Patient Sample Request Fields chapter.
■ For detailed information concerning tests selection, refer to the Creating a Patient
Worklist > Tests Selection in the Request Capture Screen chapter.
■ Profiles must be previously configured in Main menu > Services > Application
Configuration > Profiles.
■ When a profile is selected, all tests included in the profile are selected (unless
the corresponding reagent is inactive).
5. For each selected test or ratio, configure the number of replicates (from 1 to 20 replicates) as
follows:
Related information:
■ To Configure Loading Mode, p.302
■ To Add a Profile, p.292
■ Patient Sample Request Fields, p.161
■ Tests Selection in the Request Capture Screen, p.117
Do not fill in the patient demographic data. If at least one field of the patient
demographic data is captured, you will not be able to order samples by batch.
However, once the patient samples are ordered, it is possible to add the patient
demographic data to each sample. Refer to the Patient Worklist > To Modify a Patient
Sample Request chapter.
2. Depending on sample identification mode, proceed as follows to select the rack position of the
first sample tube:
a. In position mode, select the rack position from the Rack and Pos. dropdown lists.
b. In identification mode, press the Rack Pos. button to force the position mode. Then, select the
rack position from the Rack and Pos. dropdown lists.
Working with the position mode, you must ensure that the sample tube is physically
positioned as programmed in the worklist.
3. Configure the number of samples of the batch from the Batch spin box.
The maximum value of this field is restricted by the number of consecutive rack positions available
that follow the rack position of the first sample tube.
4. Fill in the sample characteristics.
In position mode, the sample ID can be entered by the operator but it is not mandatory.
In addition, to use the sample ID in batch mode, you have to enter the sample ID of the first
sample tube and this one must contain only digit characters. This sample ID will automatically be
incremented by one for each sample of the batch.
For detailed information concerning patient sample request fields, refer to the Creating
a Patient Worklist > Patient Sample Request Fields chapter.
■ For detailed information concerning tests selection, refer to the Creating a Patient
Worklist > Tests Selection in the Request Capture Screen chapter.
■ Profiles must be previously configured in Main menu > Services > Application
Configuration > Profiles.
■ When a profile is selected, all tests included in the profile are selected (unless
the corresponding reagent is inactive).
7. For each selected test or ratio, configure the number of replicates (from 1 to 20 replicates) as
follows:
a. Select the test or ratio.
b. Press the up or down arrows (right side of the screen) to increase or to decrease the
number of replicates.
The number of replicates is set to 1 by default.
8. Press OK to validate.
9. Repeat steps 2 to 8 for each batch.
Related information:
■ To Configure Loading Mode, p.302
■ To Add a Profile, p.292
■ Patient Sample Request Fields, p.161
■ Tests Selection in the Request Capture Screen, p.117
■ To Modify a Patient Sample Request, p.168
■ The sample identification mode must be previously configured in identification mode. Refer to the
Settings > System Configuration > Analyser > To Configure Loading Mode chapter.
■ Sample tubes must be identified by barcode labels.
When a rack position is forced, you must ensure that the sample tube is physically
positioned as programmed in the worklist.
For detailed information concerning patient sample request fields, refer to the Creating
a Patient Worklist > Patient Sample Request Fields chapter.
■ For detailed information concerning tests selection, refer to the Creating a Patient
Worklist > Tests Selection in the Request Capture Screen chapter.
■ Profiles must be previously configured in Main menu > Services > Application
Configuration > Profiles.
■ When a profile is selected, all tests included in the profile are selected (unless
the corresponding reagent is inactive).
5. For each selected test or ratio, configure the number of replicates (from 1 to 20 replicates) as
follows:
a. Select the test or ratio.
b. Press the up or down arrows (right side of the screen) to increase or to decrease the
number of replicates.
The number of replicates is set to 1 by default.
6. Press OK to validate.
7. Repeat steps 2 to 6 for each patient sample.
Related information:
■ To Configure Loading Mode, p.302
■ To Add a Profile, p.292
■ Patient Sample Request Fields, p.161
■ Tests Selection in the Request Capture Screen, p.117
To order patient samples from the host, the following conditions must be met:
■ The sample identification mode must be previously configured in identification mode. Refer to the
Settings > System Configuration > Analyser > To Configure Loading Mode chapter.
■ The host connection must be configured. Refer to the Settings > System Configuration > Host
Connection > To Configure the Host Connection chapter.
If the two conditions above are met, the patient sample requests programmed from the host are sent
to the instrument and added in the worklist.
Related information:
■ To Configure Loading Mode, p.302
■ To Configure the Host Connection, p.319
■ For detailed information concerning tests selection, refer to the Creating a Patient
Worklist > Tests Selection in the Request Capture Screen chapter.
6. For each selected test or ratio, configure the number of replicates (from 1 to 20 replicates) as
follows:
a. Select the test or ratio.
b. Press the up or down arrows (right side of the screen) to increase or to decrease the
number of replicates.
The number of replicates is set to 1 by default.
7. Press OK to validate.
Related information:
■ Patient Sample Request Fields, p.161
■ Tests Selection in the Request Capture Screen, p.117
Follow this procedure to delete patient sample requests from the worklist.
Ensure that a sufficient sample volume without air bubbles or foam is placed in the
instrument, whatever the sample type (calibrator, control, patient sample).
1 = Configured positions
Working with the position mode, you must ensure that the sample tube is physically
positioned as programmed in the worklist.
In identification mode, the sample tubes with barcode labels can be placed randomly
on the sample tray.
■ Before loading racks on the sample tray, ensure that the sample tray LED is green.
If not, press Pause to stop samplings on the sample tray and wait for the LED to
turn green.
■ If automatic reruns must be performed, the instrument turns to "Sampling" status
and the LED turns to orange for a few seconds before turning red. Never load
racks on the sample tray when the LED is orange or red.
Related information:
■ Sample Type and Volume, p.102
■ To Configure Loading Mode, p.302
The Work Balance menu allows you to assess the reagents on board relative to the analyses to be
performed.
This menu is divided into two tabs:
■ Reagent supply
■ Sampling Exception
The Reagent supply tab lists the solutions necessary for the pending tests.
You can display either the pending tests for the samples present on the instrument or those for the
samples programmed in the worklist.
■ Tubes on tray: To display the pending tests for the samples present on the instrument.
■ Worklist: To display the pending tests for the samples programmed in the worklist.
Heading Description
Solution requested Solution short name
Capacity requested ■ Cassette: Number of tests to be performed.
■ Open cassette or reagent rack: Solution volume necessary for the tests
to be performed (without taking into account the dead volume).
Heading Description
Present on reagent tray ■ Reagent: Number of tests that can be performed with the solution
volume remaining in the cassette.
■ Diluent or cleaner: Solution volume remaining in the cassette.
Only for cassette.
Lot # Solution lot number
Position ■ Cassette or open cassette: Position number on the reagent tray.
■ Reagent rack: First position number on the reagent tray + rack sector.
Status Solution status: Active or Inactive
On tray If selected, the solution is present on the reagent tray.
The Host request button allows you to manually ask the host if there are
additional tests programmed for a validated patient sample. This function is available only
if the Sending validated samples option is enabled in Main menu > Services > System
Configuration > Host Connection.
Related information:
■ To Configure the Host Connection, p.319
The Sampling Exception tab displays the pending tests for which a sampling alarm will be triggered.
You can display either the pending tests for the samples present on the instrument or those for the
samples programmed in the worklist.
■ Tubes on tray: To display the pending tests for the samples present on the instrument.
■ Worklist: To display the pending tests for the samples programmed in the worklist.
The tests are listed in a table format which gives the following information:
Heading Description
Sample ID ■ Blank: "BLANK" on yellow background.
■ Calibration: Calibrator name on yellow background.
■ Control: Control name on green background.
■ Patient sample: Sample ID.
Position Rack position of the sample tube
Test Application local code for the test
Sampling Alarms Sampling alarm
To display additional information concerning the sampling alarm.
Sampling Alarm Date Sampling alarm date and time
The Host request button allows you to manually ask the host if there are
additional tests programmed for a validated patient sample. This function is available only
if the Sending validated samples option is enabled in Main menu > Services > System
Configuration > Host Connection.
Related information:
■ To Configure the Host Connection, p.319
Follow this procedure to assess the reagents on board relative to the analyses to be
performed.
The instrument checks the solutions on board (barcodes reading) on reagent and sample trays. It
takes a few minutes.
The colored circle around the reagent tray on the main menu must be green. This
means that the reagent tray is accessible.
Related information:
■ Reagent supply, p.125
■ Sampling Exception, p.126
■ Sampling Alarms, p.420
1. Press Start.
Before performing the analyses, the reagent tray temperature is checked and if it is out of range, a
dialog box is displayed: The current reagent tray temperature [temperature] is out of the range
(acceptable range). The results can be affected by this malfunction. We recommend you to perform
controls. Do you still want to run analysis?
This message means that the reagents refrigeration does not work properly and this
can generate erroneous test results.
■ Press Cancel to cancel the analyses without any warning. This will allow you to perform the
appropriate controls before performing the analyses.
■ Press OK to perform the analyses in spite of this message. The analyses are performed and
the warning #642 Worklist performed with reagent tray temperature [temperature] out of the
range (acceptable range) is triggered.
The instrument turns to the "Sampling" status, then to the "Analysing" status.
1 = Instrument status
Related information:
■ To Check the Instrument, p.105
Follow this procedure to run additional patient samples or a stat sample when the
instrument is in "Sampling" status.
Ensure that a sufficient sample volume without air bubbles or foam is placed in the
instrument, whatever the sample type (calibrator, control, patient sample).
1. Press Pause to stop samplings on the sample tray and wait for the LED to turn green.
2. Place the additional patient samples or the stat sample on the sample tray.
■ Before loading racks on the sample tray, ensure that the sample tray LED is green.
If not, press Pause to stop samplings on the sample tray and wait for the LED to
turn green.
■ If automatic reruns must be performed, the instrument turns to "Sampling" status
and the LED turns to orange for a few seconds before turning red. Never load
racks on the sample tray when the LED is orange or red.
Working with the position mode, you must ensure that the sample tube is physically
positioned as programmed in the worklist.
In identification mode, the sample tubes with barcode labels can be placed randomly
on the sample tray.
■ For detailed information concerning tests selection, refer to the Creating a Patient
Worklist > Tests Selection in the Request Capture Screen chapter.
■ Profiles must be previously configured in Main menu > Services > Application
Configuration > Profiles.
■ When a profile is selected, all tests included in the profile are selected (unless
the corresponding reagent is inactive).
8. For each selected test or ratio, configure the number of replicates (from 1 to 20 replicates) as
follows:
a. Select the test or ratio.
b. Press the up or down arrows (right side of the screen) to increase or to decrease the
number of replicates.
The number of replicates is set to 1 by default.
9. Press OK to validate.
The additional patient samples or the stat sample are run. For a stat sample, this one is run in priority
before any sample remaining in the worklist.
If the instrument is in "Reagent processing" or "Analysing" status, skip steps 1 and 11.
Related information:
■ Sample Type and Volume, p.102
■ Patient Sample Request Fields, p.161
■ Tests Selection in the Request Capture Screen, p.117
Status Description
Ordered The sample is requested in the worklist.
Pending The sample is read on the tray.
Incomplete At least one test of the sample is pipetted.
For validation All tests of the sample are finished and for validation.
Validated All tests of the sample are validated.
Archived The sample is archived.
The sample tray is symbolized in the middle of the main menu. It allows you to visualize samples
configured on the tray and to access their details.
The sample tray contains 60 sample positions and can hold sample racks, sample cup racks and
calibrator/control sample racks.
The status of each sample is indicated by a color code to which a legend is available by pressing the
blue background inside the sample tray.
You can access the sample rack details by pressing a position on the sample tray.
Heading Description
Position Rack position of the sample tube
Status Pending, Incomplete, For validation or Validated
Run date Sample run date and time
Sample ID Sample ID
Patient ID Patient ID
Last Name Patient last name
Sample Type Serum/Plasma, Urine or Other.
Follow this procedure to check the status of each sample configured on the sample tray.
1 = Sample tray
2 = Sample requiring an additional check
1. Check the status of each sample configured on the sample tray from the main menu.
The status of each sample is indicated by a color code on the sample tray. The samples appearing
in red require an additional check because of a sampling alarm.
2. Check the samples in red by referring to the Patient Samples > Test Review > To Check Sampling
Alarms and Analytical Flags chapter.
Related information:
■ To Check Sampling Alarms and Analytical Flags, p.136
■ Sample Tray Description, p.178
The Test Review menu allows you to review the tests in progress, to check sampling alarms and
analytical flags that may have been triggered and to manually export results to a USB key at each end
of run.
This menu is divided into four tabs:
■ Test Result
■ Sampling Exception
■ Error Report
■ Sequencing
The Test Result tab displays the tests in progress as soon as they are pipetted.
The tests are sorted by run date.
■ Up to 200 items are displayed from the oldest to the most recent. Beyond this number,
the oldest are deleted.
■ The tab is cleared at instrument startup, when starting analyses and when a new
worklist is created.
The tests are listed in a table format which gives the following information:
Heading Description
Status Ordered, Pending, Incomplete, For validation or Validated
For Ordered and Pending status, at least one test of the sample is finished.
Sample ID ■ Blank: "BLANK" on yellow background.
■ Calibration: Calibrator name on yellow background.
■ Control: Control name on green background.
■ Patient sample: Sample ID.
Position Rack position of the sample tube
Test Application local code for the test
Result(s) ■ Blank and Calibration: Absorbance value.
■ Control: Final result. On red background if out of confidence range.
■ Patient sample: Final result.
"--.-----" means that the result is not given because of a sampling alarm or
an analytical flag.
QF Quality flag (only the flag having the highest priority is displayed).
D/C The sample was automatically rerun with post-dilution (Dil) or post-
concentration (Conc).
Only for patient sample.
Analysis flag Sampling alarm or analytical flag (only the flag having the highest priority is
displayed).
If both are triggered, only the sampling alarm is displayed.
Dec. Decision:
■ Blank: Accept or Delete.
■ Calibration: Not displayed.
■ Control: Accept or Delete.
■ Patient sample: Accept, Rerun, Modify or Delete.
Unit Test unit configured in the application
Run date Sample run date and time
You can access the sample details by selecting a test and pressing Details.
You can also search for a sample either by sample ID or by position using the search tools at the
bottom of the screen.
The Export button allows you to manually export results to a USB key. Refer to
the Test Review > To Manually Export Results to a USB Key chapter.
Related information:
■ To Manually Export Results to a USB Key, p.137
The Sampling Exception tab displays the tests for which a sampling alarm is triggered.
The tests are listed in a table format which gives the following information:
Heading Description
Sample ID ■ Blank: "BLANK" on yellow background.
■ Calibration: Calibrator name on yellow background.
■ Control: Control name on green background.
■ Patient sample: Sample ID.
Position Rack position of the sample tube
Test Application local code for the test
Sampling Alarms Sampling alarm
To display additional information concerning the sampling alarm.
Sampling Alarm Date Sampling alarm date and time
Run date Sample run date and time
You can access the sample details by selecting a test and pressing Details.
You can also search for a sample either by sample ID or by position using the search tools at the
bottom of the screen.
The Error Report tab displays the tests for which an analytical flag or a quality flag blocking the
automatic validation is triggered.
The tests are sorted by run date.
The tests are listed in a table format which gives the following information:
Heading Description
Sample ID ■ Blank: "BLANK" on yellow background.
■ Calibration: Calibrator name on yellow background.
■ Control: Control name on green background.
■ Patient sample: Sample ID.
Position Rack position of the sample tube
Test Application local code for the test
Result(s) ■ Blank and Calibration: Absorbance value.
■ Control: Final result. On red background if out of confidence range.
■ Patient sample: Final result.
"--.-----" means that the result is not given because of a sampling alarm or
an analytical flag.
QF Quality flag (only the flag having the highest priority is displayed).
D/C The sample was automatically rerun with post-dilution (Dil) or post-
concentration (Conc).
Only for patient sample.
Analysis flag Analytical flag (only the flag having the highest priority is displayed).
Unit Test unit configured in the application
Run date Sample run date and time
You can access the sample details by selecting a test and pressing Details.
You can also search for a sample either by sample ID or by position using the search tools at the
bottom of the screen.
6.7.4. Sequencing
■ Up to 200 items are displayed from the oldest to the most recent. Beyond this number,
the oldest are deleted.
■ The tab is cleared at instrument startup, when starting analyses and when a new
worklist is created.
Heading Description
Date Sampling date and time
Nature Sampling nature:
■ for a test: Application local code,
■ for a dilution: Dilution,
■ for a cleaning: Cleaner.
Reagent needle Solution pipetted by the reagent needle
Volume Volume pipetted by the reagent needle (µL)
Heading Description
Sampling position Position where the solution is pipetted:
■ on the reagent tray: R (for reagent tray) + position number on the tray +
position number in the cassette or on the reagent rack.
Type Container type where the solution is pipetted:
■ for a cassette: Cassette,
■ for a reagent rack: Rack ID + position number on the reagent rack.
Dispensing position Position where the solution is dispensed:
■ on the reaction tray: C (for cuvette) + position number on the tray +
position number in the cuvette segment.
Sample Arm
Heading Description
Date Sampling date and time
Nature Sampling nature:
■ for a test: Application local code,
■ for an ISE test: ISE,
■ for a dilution: Dilution,
■ for a cleaning: Cleaner.
Sample needle Solution pipetted by the sample needle
Volume Volume pipetted by the sample needle (µL)
H2O Distilled water volume pipetted by the sample needle (µL) to push the sample
volume
Sampling position Position where the solution is pipetted:
■ on the sample tray: S (for sample tray) + rack number + position number
on the rack,
■ on the reagent tray: R (for reagent tray) + position number on the tray +
position number in the cassette or on the reagent rack,
■ on the reaction tray: C (for cuvette) + position number on the tray +
position number in the cuvette segment.
Type Container type where the solution is pipetted:
■ for a sample tube: Tube,
■ for a sample cup: Sample cup,
■ for a cassette: Cassette,
■ for a reagent rack or a calibrator/control reagent rack: Rack ID + position
number on the reagent rack,
■ for a cuvette: Cuvette.
Dispensing position Position where the solution is dispensed:
■ on the reaction tray: C (for cuvette) + position number on the tray +
position number in the cuvette segment,
■ in the ISE module sample cup: ISE bowl.
Other
Heading Description
Date Date and time of mixing or cuvette segment loading/unloading
Mixing speed Mixing speed used to homogenize solutions into the cuvette
Cuvette loader Cuvette segment loading/unloading: Loading or Unloading + position
number on the tray
Follow this procedure to check sampling alarms and analytical flags that may have been
triggered.
The Test Review button becomes red when a sampling alarm or an analytical flag is triggered.
The Sampling Exception tab displays the tests for which a sampling alarm is triggered.
Related information:
■ Sampling Exception, p.132
■ Error Report, p.133
■ Analytical and Quality Flags, p.423
■ Sampling Alarms, p.420
Follow this procedure to manually export results to a USB key at the end of run.
Results are automatically exported to a USB key at each end of run if the Activate export
of the results on USB key option is selected in Main menu > Services > System
Configuration > Analyser. Refer to the Settings > System Configuration > Analyser > To
Enable or To Disable Automatic Export of Results chapter.
HORIBA Medical recommends you to make sure that all materials like CD-ROM or USB
key used on the Pentra C400 and non-referenced by HORIBA Medical are free of
viruses. These materials must imperatively be cleaned before being used on the
instrument.
All the completed test results are saved in a .csv file located in a folder named "Export" on the USB
key. ISE test results and ratio results are not exported.
Related information:
■ To Enable or To Disable Automatic Export of Results, p.310
The patient result list displays all patient sample results that are incomplete, for validation or validated,
as well as the patient sample results that are ordered or pending and having at least one test of the
sample finished.
Patient sample results are sorted by patient ID and by date. Alternation of white and blue materializes
the separation between patient ID.
You can select the patient sample results to be displayed from the following options:
■ Incomplete: To display patient sample results that are incomplete as well as patient sample
results that are ordered or pending and having at least one test of the sample finished.
■ For validation: To display patient sample results that are for validation.
■ Validated: To display patient sample results that are validated.
■ All: To display all patient sample results.
Patient sample results are listed in a table format which gives the following information.
Heading Description
Status Sample status: Ordered, Pending, Incomplete, For validation or Validated
For Ordered and Pending status, at least one test of the sample is finished.
Patient ID Patient ID
Sample ID Sample ID
Position Rack position of the sample tube
Run Date Sample run date and time
Collection Date Sample collection date and time
The Host request button allows you to manually ask the host if there are
additional tests programmed for a validated patient sample. This function is available only
if the Sending validated samples option is enabled in Main menu > Services > System
Configuration > Host Connection.
For detailed information concerning the procedure to check patient results, refer to the
Patient Results Validation > To Make a Decision on Patient Results chapter.
For detailed information concerning patient sample request fields, refer to the Creating
a Patient Worklist > Patient Sample Request Fields chapter.
Patient results
Heading Description
Pos. Rack position of the sample tube
To display several results for a test in case of rerun or result modification.
Test Application local code for the test
Result(s) Final result.
"--.-----" means that the result is not given because of a sampling alarm or
an analytical flag.
QF Quality flag (only the flag having the highest priority is displayed).
D/C The sample was automatically rerun with post-dilution (Dil) or post-
concentration (Conc).
Analysis flag Sampling alarm or analytical flag (only the flag having the highest priority is
displayed).
If both are triggered, only the sampling alarm is displayed.
DC Delta check flag (D+ or D-)
Dec. Decision: Accept, Rerun, Modify or Delete
Unit Test unit configured in the application
Run date Sample run date and time
Previous Previous result
Previous date Sample run date and time
Reference Range Low and high limit values of the Man/Default reference range
■ For a ratio result, the ratio is displayed on the first line and then its included tests and
sub-ratio (if any) are displayed in gray on the following lines.
■ For detailed information concerning analytical flags or quality flags, refer to the Alarms
> Analytical and Quality Flags chapter.
The O.D. Values tab is unavailable for ISE test results, ratio results and for results coming
from rerun or result modification.
The graph (right side of the screen) displays the following data:
■ X-axis: time,
■ Y-axis: absorbance value,
■ Green line: kinetic curve.
■ If the absorbance deviation check on linear regression is configured in the application, the
correlation coefficient r2 is calculated and displayed.
Data of the kinetic curve is also displayed in a table format (left side of the screen).
If analytical flags are triggered, they are displayed in a second table.
The Control tab is unavailable for ratio results and for results coming from rerun or result
modification.
Heading Description
Default If selected, the control is a default control for this test.
Name Control name
Run date Date and time of control run
On orange background if the control and the sample are not performed the
same day or if the control and the sample do not use the same calibration.
Result(s) Final result. On red background if out of confidence range.
"--.-----" means that the result is not given because of a sampling alarm or
an analytical flag.
QF Quality flag (only the flag having the highest priority is displayed).
Analysis flag Sampling alarm or analytical flag (only the flag having the highest priority is
displayed).
If both are triggered, only the sampling alarm is displayed.
Dec. Decision: Accept or Delete
Unit Test unit configured in the application
Target Value Control target value
Confidence Range Low and high limit values of confidence range
For detailed information concerning analytical flags or quality flags, refer to the Alarms >
Analytical and Quality Flags chapter.
Related information:
■ Patient Sample Request Fields, p.161
■ Analytical and Quality Flags, p.423
■ To Make a Decision on Patient Results, p.191
Follow this procedure to manually validate or to manually modify a patient result, as well
as to rerun or to delete a patient test.
Patient results are automatically validated depending on the patients automatic validation configured
in Main menu > Services > System Configuration > Results validation.
1. Select the For validation option to display the patient sample results that are for validation.
2. Select the patient results that you want to review and press Details.
The result validation screen is displayed.
4 = Edit
5 = Test selection
6 = Result selection
7 = Decision selection
8 = OK
3. Check the patient results and make sure that no analytical flag or quality flag has been triggered.
For detailed information concerning patient results interpretation, refer to the Patient
Results Validation > Patient Results Interpretation chapter.
4. Press Edit.
5. Make a decision for each test as follows:
a. Select a test.
b. In case of rerun or result modification, press to display all results for the test and select a
result by double-clicking the corresponding colored button.
The colored button is:
■ Green when the result is selected,
■ Orange when the result is selectable,
■ Red when the result is not selectable.
The selected result becomes the final result.
c. Select a decision from the following options:
■ Accept: To validate the test result.
■ Rerun: To rerun the test. Up to three reruns can be performed (manual or automatic).
■ Modify: To manually modify the test result.
You can make a decision on a ratio result but not on its included tests and sub-ratio (if
any).
6. Press OK to validate.
Related information:
■ To Configure Patients Automatic Validation, p.327
■ Patient Results Interpretation, p.188
Follow this procedure to add tests on a patient sample from the result validation screen.
■ For detailed information concerning tests selection, refer to the Creating a Patient
Worklist > Tests Selection in the Request Capture Screen chapter.
■ Profiles must be previously configured in Main menu > Services > Application
Configuration > Profiles.
■ When a profile is selected, all tests included in the profile are selected (unless
the corresponding reagent is inactive).
6. For each selected test or ratio, configure the number of replicates (from 1 to 20 replicates) as
follows:
a. Select the test or ratio.
b. Press the up or down arrows (right side of the screen) to increase or to decrease the
number of replicates.
The number of replicates is set to 1 by default.
7. Press OK to validate.
Related information:
■ To Add a Profile, p.292
■ Tests Selection in the Request Capture Screen, p.117
Patient results are automatically printed when validated if configured in Main menu > Services >
System Configuration > Printer.
1. Select the For validation option to display the patient sample results that are for validation or the
Validated option to display the patient sample results that are validated.
2. Select the patient results that you want to print and press Data Print/Send to Host from the
generic toolbar.
The Print window is displayed.
3. Select the Order option to print all results of the selected patient (results of all samples) or the
Selected Sample option to print only results of the selected sample.
4. Select the printout format from the following options:
■ condensed,
■ standard,
■ detailed,
■ portrait.
Detailed and portrait printout formats are unavailable if the Order option is selected.
5. Press OK to validate.
The validated results of the selected patient (results of all samples) or of the selected sample are
printed.
Related information:
■ To Configure Automatic Printouts, p.323
Follow this procedure to print all patient results of a working day as a condensed list.
Press the Select all button to select all the tests or the Deselect all button
Patient results are automatically sent when validated if configured in Main menu > Services >
System Configuration > Host Connection.
1. Select the Validated option to display the patient sample results that are validated.
2. Select the patient results that you want to send and press Data Print/Send to Host from the
generic toolbar.
The Print window is displayed.
3. Press Send.
The Send window is displayed.
4. Select the Order option to send all results of the selected patient (results of all samples) or the
Selected Sample option to send only results of the selected sample.
5. Press OK to validate.
The validated results of the selected patient (results of all samples) or of the selected sample are
sent.
Related information:
■ To Configure the Host Connection, p.319
Related information:
■ To Search for Archived Results by Worklist/SID, p.200
■ To Search for Archived Results by Patient, p.201
■ To Search for Results in the Archive Files, p.202
■ To Export Archived Results to a USB Key, p.204
Follow this procedure to search for archived results in the current database by worklist
and sample ID.
For detailed information concerning patient results interpretation, refer to the Patient
Results Validation > Patient Results Interpretation chapter.
Related information:
■ Patient Results Interpretation, p.188
Follow this procedure to search for archived results in the current database by patient.
2. Select the run date that you want to review and press Details.
The result validation screen is displayed.
For detailed information concerning patient results interpretation, refer to the Patient
Results Validation > Patient Results Interpretation chapter.
Related information:
■ Patient Results Interpretation, p.188
Follow this procedure to search for results in the archive files saved on the hard disk or a
USB key.
HORIBA Medical recommends you to make sure that all materials like CD-ROM or
USB key used on the Pentra C400 and non-referenced by HORIBA Medical are free
of viruses. These materials must imperatively be cleaned before being used on the
instrument.
2. Select the patient that you want to review from the Patient List area (left side of the screen).
The archive files available for the selected patient are displayed in the Archives List area (right
side of the screen).
If no patient is selected from the Patient List area, all available archive files are
displayed in the Archives List area.
3. Select the archive file that you want to open and press Details.
Once you have selected an archive file, you can consult logs and control results of the selected
archive file by pressing respectively Logs or Quality Control.
Results of the selected patient are displayed at the bottom of the screen. Results are listed by run
date in a table format which indicates the tests performed, the test unit, the test result and the
number of replicates.
Refer to the Patient Results Archives > To Search for Archived Results by Patient chapter.
If no patient is selected from the Patient List area, worklists of the selected archive file
are displayed.
Refer to the Patient Results Archives > To Search for Archived Results by Worklist/SID
chapter.
Related information:
■ To Search for Archived Results by Worklist/SID, p.200
■ To Search for Archived Results by Patient, p.201
■ To Save the Archive Files on a USB Key, p.203
Follow this procedure to save the archive files from the hard disk on a USB key.
HORIBA Medical recommends you to make sure that all materials like CD-ROM or USB
key used on the Pentra C400 and non-referenced by HORIBA Medical are free of
viruses. These materials must imperatively be cleaned before being used on the
instrument.
4. Press Export:
Follow this procedure to export archived results from the current database in a .csv file to
a USB key.
Press the Select all button to select all the tests or the Deselect all button
6. Select the test results that you want to export from the Results found table.
By default, all test results are selected to be exported. If necessary, deselect the test results that
should not be exported.
7. Select the Include OD option to export the absorbance values of kinetic curves.
8. Press Export:
A dialog box asks you to confirm that the USB key is in place.
9. Install your USB key on the instrument.
HORIBA Medical recommends you to make sure that all materials like CD-ROM or USB
key used on the Pentra C400 and non-referenced by HORIBA Medical are free of
viruses. These materials must imperatively be cleaned before being used on the
instrument.
The selected test results are saved in a .csv file located in a folder named "Export" on the USB key.
The instrument can be switched to standby mode either manually or automatically if automatic
standby mode is programmed.
The instrument standby mode consists in switching the spectrophotometer lamp to standby mode.
1 = Stand by option
2 = ISE Cleaning option
3 = ISE Cleaning dropdown list
4 = System Cleaning option
5 = System Cleaning dropdown list
6 = OK
■ The ISE module should be cleaned with ABX Pentra Etching CP A11A01769 every
day only if more than 20 samples are run daily.
■ If less than 20 samples are run daily, deselect the ISE Cleaning option and perform
a complete ISE module cleaning every week. Refer to the Maintenance and
Troubleshooting > Maintenance > Weekly Procedures > To Clean the ISE Module
(Option) chapter.
4. If you programmed a needle cleaning during the instrument shutdown by default, you can clean
the needles before entering in standby mode as follows:
a. Select the System Cleaning option.
b. Choose DEPR (ABX Pentra Deproteinizer CP A11A01754) from the System Cleaning
dropdown list.
5. Press OK to validate.
Related information:
■ To Configure Automatic Standby Mode, p.207
■ To Configure Automatic Cleanings, p.306
Access: Main menu > Services > System Configuration > Analyser
If an automatic standby mode is programmed, the instrument automatically enters standby mode after
a specified inactivity time.
1. Press Edit.
2. Select the Standby mode check option to enable automatic standby mode.
3. Specify the inactivity time.
4. Press OK to validate.
1. Select your user name or login from the Name dropdown list.
2. Type your password in the Password field.
3. Press OK to validate and to enter the application.
Follow this procedure to change the user name or login during the day of work without
instrument initialization.
1 = Change User
2 = OK
2. Select the Change User option.
3. Press OK to validate.
The Startup window is displayed.
Follow this procedure to create a new worklist during the day of work without instrument
initialization.
1 = Change User
2 = OK
2. Select the Change User option.
3. Press OK to validate.
The Startup window is displayed.
8. End of Day
Before performing end of day procedure, make sure that no analysis is pending in the
patient worklist.
Related information:
■ To Delete Patient Sample Requests, p.169
■ To Make a Decision on Patient Results, p.191
At the beginning of each day (before startup) or at the end of each day (before shutdown),
check the following items:
■ distilled water level,
■ waste level,
■ "new cuvette" holder,
■ "used cuvette" holder,
■ levels of ISE reagents.
1. Check the distilled water level in the tank. Fill the tank if necessary.
2. Check the waste level in the tank. Empty the tank if necessary.
When disposing of waste, protective clothing must be worn (lab coat, gloves, eye
protection, etc.). Follow your local and/or national guidelines for biohazard waste
disposal.
■ At the beginning of each day, before startup, check if the waste container needs to
be emptied.
■ During instrument operation, do not remove the liquid waste tube under any
circumstance.
■ Do not wash and re-use cuvettes. They are designed for single use only.
■ Cuvette segments with at least one used cuvette must never be loaded in the
reaction tray.
Dispose of used cuvettes according to your local and/or national guidelines for
biohazard waste disposal.
7. For Pentra C400 with ISE module (option), open the ISE module cover to check levels of ISE
reagents.
8. If an ISE reagent needs to be replaced, refer to the Maintenance and Troubleshooting >
Maintenance > Other Procedures > To Replace ISE Reagents (Option) chapter.
Related information:
■ To Replace ISE Reagents (Option), p.363
Follow this procedure to shut down the instrument at the end of the day.
■ Automatic cleanings must be programmed in Main menu > Services > System Configuration >
Analyser.
■ The main menu must be displayed.
■ The instrument must be in the "Ready" status.
1 = Shutdown option
2 = Startup date
3 = ISE Cleaning option
4 = ISE Cleaning dropdown list
5 = System Cleaning option
6 = System Cleaning dropdown list
7 = OK
2. Select the Shutdown option.
3. If an automatic startup is programmed, the instrument automatically starts the next day at a
specified time. In case of day off, change the startup date.
4. If you programmed by default an ISE module cleaning at the instrument shutdown, the ISE
Cleaning option is selected by default. Choose ETCH (ABX Pentra Etching CP A11A01769) from
the ISE Cleaning dropdown list.
■ The ISE module should be cleaned with ABX Pentra Etching CP A11A01769 every
day only if more than 20 samples are run daily.
■ If less than 20 samples are run daily, deselect the ISE Cleaning option and perform
a complete ISE module cleaning every week. Refer to the Maintenance and
Troubleshooting > Maintenance > Weekly Procedures > To Clean the ISE Module
(Option) chapter.
5. If you programmed by default a needles cleaning at the instrument shutdown, the System
Cleaning option is selected by default. Choose DEPR (ABX Pentra Deproteinizer CP
A11A01754) from the System Cleaning dropdown list.
6. Press OK to validate.
ISE module and/or needles are automatically cleaned before the instrument shutdown:
■ reaction tray check (used cuvette segments are unloaded from the reaction tray),
■ temperature regulation switching off,
■ computer and spectrophotometer lamp switching off.
7. Wait for the end of instrument shutdown. If the ABX Pentra Etching CP A11A01769 was
previously placed on the reagent tray, remove it from the tray, close it and store it at room
temperature.
8. Ensure that position 36 of the reagent tray is in front of the yellow triangle.
The Pentra C400 hardware has to remain on 24 hours a day. This is necessary to keep the
cooling unit functionnal and therefore maintain the temperature in the refrigerated area of
the reagent tray. Do not switch off neither the Pentra C400 nor the cooling unit.
If you want to switch the Pentra C400 and the cooling unit off for several days, remove the
reagents from the tray, close them and place them in a refrigerated area. For Pentra C400
with ISE module (option), refer to the Maintenance and Troubleshooting > Maintenance >
Other Procedures > ISE Module Shutdown (Option) chapter.
Related information:
■ To Configure Automatic Cleanings, p.306
■ To Configure the Automatic Startup, p.305
■ To Clean the ISE Module (Option), p.341
1. Reagent Management
Tray
The 52 positions of the reagent tray are symbolized in the Tray tab. It allows you to visualize solutions
configured on the tray and to access their details.
The instrument checks the solutions on board (barcodes reading) during the initialization
and when closing the reagent tray cover.
The status of each solution is indicated by a color code to which a legend is available by pressing the
blue background.
Low Volume There is less than 10% of the initial volume remaining
(only for cassettes).
To be checked The solution needs to be checked because of one of the
following reasons:
■ The cassette is empty.
■ The solution is unknown (bad barcode for example).
■ Not enough solution for the pending analyses.
■ The solution is inactive.
■ The solution has reached its use limit date or has
expired.
Available Empty position
You can access solution details by pressing a position. The displayed information varies depending
on whether the solution is in a cassette, in an open cassette, on a reagent rack or on a calibrator/
control reagent rack.
■ Cassette: Refer to the Reagent Management > Managing Cassettes > Cassette Screen chapter.
■ Open cassette: Refer to the Reagent Management > Managing Reagents in an Open Cassette >
Open Cassette Screen chapter.
■ Reagent rack: Refer to the Reagent Management > Managing Reagents on a Reagent Rack >
Reagent Rack Screen chapter.
■ Calibrator/control reagent rack: Displays the rack ID and the position number on the reagent
tray as well as the characteristics (name, lot number and expiration date) of each calibrator/control
configured on the rack.
Configuration
The Configuration tab lists the solutions configured on the instrument. Solutions are either
automatically configured after installation on board (for solutions in a cassette or in an open cassette)
or manually configured on a reagent rack.
You can select the solutions to display from the following options.
■ Current: To display all solutions currently configured on the instrument.
■ Active current: To display only the active solutions currently configured on the instrument.
■ Inactive current: To display only the inactive solutions currently configured on the instrument.
■ ABX cassettes run out: To display the deleted cassettes.
Cassettes are automatically deleted when they are empty or definitively deactivated.
The solutions are listed in a table format which gives the following information.
Heading Description
Position Position number on the reagent tray
Solution Solution type: Reagent, Diluent or Cleaner
Rack ID Cassette barcode or rack ID
Rack Type Container type: Cassette or Reagent Rack
Status Solution status: Active or Inactive
Reagent Short Name Solution short name
Lot # Solution lot number
Configuration Number of reagents: Reagent 1, Reagent 2 or Reagent 3
Filling Volume R1 initial volume.
Only for cassette.
Expiration Date Expiration date before opening.
Only for cassette.
Installation Date Installation on board date and time.
Only for cassette.
Heading Description
Use Limit Date Expiration date after installation on board.
Only for cassette.
Available Tests Number of tests that can be performed with the solution volume remaining
in the cassette.
Only for cassette.
You can access solution details by selecting a solution and pressing Details. The displayed
information varies depending on whether the solution is in a cassette, in an open cassette or on a
reagent rack.
■ Cassette: Refer to the Reagent Management > Managing Cassettes > Cassette Screen chapter.
■ Open cassette: Refer to the Reagent Management > Managing Reagents in an Open Cassette >
Open Cassette Screen chapter.
■ Reagent rack: Refer to the Reagent Management > Managing Reagents on a Reagent Rack >
Reagent Rack Screen chapter.
Related information:
■ Cassette Screen, p.221
■ Open Cassette Screen, p.225
■ Reagent Rack Screen, p.228
The solutions provided by HORIBA Medical in a cassette are automatically configured in Reagent
Configuration menu after installation on board.
This screen is accessible either from the Tray tab by pressing a position or from the
Configuration tab by selecting a solution and pressing Details.
The Cassette screen displays the selected cassette (left side of the screen) and the selected solution
characteristics (right side of the screen).
Heading Description
ID Cassette barcode
Tray Sector Position number on the reagent tray
Solution Type Solution type: Reagent, Diluent or Cleaner
Reagent Solution Name Solution short name
Lot Solution lot number
Analysis Method Number of reagents: Reagent 1 or Reagent 2
Reagent Status Solution status: Active or Inactive
Installation Date Installation on board date and time.
Expiration Date Expiration date before opening.
Use Limit Date Expiration date after installation on board.
Volume R1 initial volume.
When a new cassette is installed on the reagent tray, it is automatically activated unless a cassette of
the same solution is already active.
Several cassettes of the same solution cannot be active simultaneously on the reagent tray. If several
cassettes of the same solution are installed on the reagent tray, then the first installed cassette is
active and the next cassettes are inactive.
If the automatic back-up cassette option is enabled, the next cassettes are called back-up cassettes.
Back-up cassettes are automatically activated when the cassette in use is empty.
Refer to the Settings > System Configuration > Analyser > To Configure the Automatic
Back-up Cassette Option chapter to configure the automatic back-up cassette option.
■ If Installation date of cassette (n+1) < Installation date of cassette (n), then the current
calibration is disabled and a new calibration must be performed with the default
controls if configured in the application.
■ If Installation date of cassette (n+1) ≥ Installation date of cassette (n), then the current
calibration stays valid and the default controls, if configured in the application, must be
performed.
Related information:
■ Calibration Parameters, p.260
■ To Configure the Automatic Back-up Cassette Option, p.309
Follow this procedure to activate a cassette once it has been installed on the reagent tray.
When a new cassette is installed on the reagent tray, it is automatically activated unless a cassette of
the same solution is already active. In this case, you have to deactivate the active cassette before
activating the new cassette.
1. Select the Active current option to display only the active solutions currently configured on the
instrument.
2. Select the solution that you want to deactivate and press Details.
The Cassette screen is displayed.
3. Press Edit.
This function is unavailable if a calibration for this test is ordered, pending, incomplete or for
validation.
4. Select Inactive from the Reagent Status dropdown list.
5. Press OK to validate.
6. Press Back to go back to the Configuration tab.
7. Select the Inactive current option to display only the inactive solutions currently configured on
the instrument.
8. Select the solution that you want to activate and press Details.
The Cassette screen is displayed.
9. Press Edit.
10. Select Active from the Reagent Status dropdown list.
Related information:
■ Cassette Screen, p.221
The solutions in an open cassette are automatically configured in the Reagent Configuration menu
after installation on board.
However, they must be previously registered in Main menu > Services > Application Configuration
> Reagents and identified by barcode labels (PC400/P400 Open Cas. sticker sheet1 HAX0334
(1207230334) and PC400/P400 Open Cas. sticker sheet2 HAX0335 (1207230335)) provided by
HORIBA Medical.
The barcode label must match the reagent number previously registered.
1 = Barcode label
location
2 = 11.5 mm
Related information:
■ To Register a Reagent, p.299
This screen is accessible either from the Tray tab by pressing a position or from the
Configuration tab by selecting a solution and pressing Details.
The Open Cassette screen displays the selected cassette (left side of the screen) and the selected
solution characteristics (right side of the screen).
Heading Description
ID Cassette barcode
Tray Sector Position number on the reagent tray
Solution Type Solution type: Reagent, Diluent or Cleaner
Reagent Solution Name Solution short name
Lot Solution lot number
Analysis Method Number of reagents: Reagent 1 or Reagent 2
Reagent Status Solution status: Active or Inactive
When a new open cassette has been installed on the reagent tray, it stays inactive. Therefore, it is
necessary to activate it before performing analyses on this new cassette.
Several open cassettes of the same solution cannot be installed on the reagent tray.
Empty open cassettes are automatically deactivated.
Follow this procedure to activate an open cassette once it has been installed on the
reagent tray.
When a new open cassette has been installed on the reagent tray, it stays inactive. Therefore, it is
necessary to activate it before performing analyses on this new cassette.
1. Select the Inactive current option to display only the inactive solutions currently configured on
the instrument.
2. Select the solution that you want to activate and press Details.
The Open Cassette screen is displayed.
3. Press Edit.
4. Type the lot number (16 characters maximum).
5. Select Active from the Reagent Status dropdown list.
6. Press OK to validate.
Related information:
■ Open Cassette Screen, p.225
Follow this procedure to change the lot number of a solution in an open cassette.
The colored circle around the reagent tray on the main menu must be green. This
means that the reagent tray is accessible.
After you changed the solution lot number, a new calibration must be performed with the default
controls if configured in the application.
Related information:
■ Open Cassette Screen, p.225
The solutions on a reagent rack must be manually configured in the Reagent Configuration menu.
They must also be previously registered in the Main menu > Services > Application Configuration >
Reagents if they are not already pre-configured on the instrument.
Related information:
■ To Add a Reagent on a Reagent Rack, p.229
■ To Register a Reagent, p.299
This screen is accessible either from the Tray tab by pressing a position or from the
Configuration tab by selecting a solution and pressing Details.
The Reagent Rack screen displays the selected rack (left side of the screen) and the selected
solution characteristics (right side of the screen).
Heading Description
ID Rack ID
Rack Sector Rack sector: A, B or C
Tray Sector First position number on the reagent tray
Solution Type Solution type: Reagent, Diluent or Cleaner
Reagent Solution Name Solution short name
Lot Solution lot number
Analysis Method Number of reagents: Reagent 1, Reagent 2 or Reagent 3
Reagent Status Solution status: Active or Inactive
When a new solution is configured on the reagent tray, it is automatically activated unless an identical
solution is already active.
Several identical solutions cannot be active simultaneously on the reagent tray. If several identical
solutions are configured on the reagent tray, then the first configured solution is active and the next
solutions are inactive.
Empty solutions are automatically deactivated.
The solution is configured on the reagent tray, it is automatically activated unless an identical solution
is already active.
Related information:
■ Reagent Rack Screen, p.228
Follow this procedure to activate a solution on a reagent rack once it is configured on the
reagent tray.
When a new solution is configured on the reagent tray, it is automatically activated unless an identical
solution is already active. In this case, you have to deactivate the active solution before activating the
new solution.
1. Select the Active current option to display only the active solutions currently configured on the
instrument.
2. Select the solution that you want to deactivate and press Details.
The Reagent Rack screen is displayed.
3. Press Edit.
4. Select Inactive from the Reagent Status dropdown list.
5. Press OK to validate.
6. Press Back to go back to the Configuration tab.
7. Select the Inactive current option to display only the inactive solutions currently configured on
the instrument.
8. Select the solution that you want to activate and press Details.
The Reagent Rack screen is displayed.
9. Press Edit.
10. Select Active from the Reagent Status dropdown list.
11. Press OK to validate.
Related information:
■ Reagent Rack Screen, p.228
Follow this procedure to change the lot number of a solution on a reagent rack.
After you changed the solution lot number, a new calibration must be performed with the default
controls if configured in the application.
Related information:
■ Reagent Rack Screen, p.228
The Calibration tab lists the calibrators configured on the instrument. Calibrators are listed in a table
format which indicates the calibrator name, the rack position of calibrator, the calibrator lot number
and expiration date.
To configure a calibrator, three steps are needed:
■ Enter the calibrator characteristics,
■ Add the target values,
■ Configure the calibrator position.
Related information:
■ To Enter the Calibrator Characteristics, p.232
■ To Add Target Values, p.233
■ To Configure the Calibrator Position, p.234
Follow this procedure to enter the calibrator name, lot number and expiration date.
After you entered the calibrator characteristics, add the target values and configure the calibrator
position.
Related information:
■ To Add Target Values, p.233
■ To Configure the Calibrator Position, p.234
To add the calibrator target values, the calibrator characteristics must have already been entered.
Always ensure that the entered target values correspond to the calibrator name and lot
number used. Refer to the calibrator annex.
If the application features a calibration pre-dilution with Main direct or Main indirect
dilution type and a sample pre-dilution is programmed in the application, the calibrator
target values entered in the Target Values Entry screen have to be multiplied by the
pre-dilution factor. Example: If a sample pre-dilution with a pre-dilution factor of 21 is
programmed in the application and the calibrator target value mentions 2.22 g/L, the
value that must be entered is 2.22 * 21 = 46.62 g/L.
5. If you want to perform a recalibration with a single calibrator level, select the Standard Correction
option and choose the calibrator level from the spin box.
This function is unavailable if:
■ The calibration mode configured in the application is Factor or Slope average,
■ A calibration pre-dilution with Main direct or Main indirect dilution type is configured in the
application,
■ No calibration was previously performed for the test.
6. Press OK to validate.
After you added the target values, configure the calibrator position.
Related information:
■ To Enter the Calibrator Characteristics, p.232
■ To Configure the Calibrator Position, p.234
To configure the calibrator position, the calibrator characteristics and the target values must have
already been entered.
The racks used on the instrument must be previously configured in the Main menu >
Services > System Configuration > Analyser.
The selected rack is displayed (left side of the screen). The status of each position is indicated by
a color code.
5. Press OK to validate.
Related information:
■ To Enter the Calibrator Characteristics, p.232
■ To Add Target Values, p.233
■ To Configure Racks, p.307
Follow this procedure to import the calibrator configuration (calibrator characteristics and
target values) provided by HORIBA Medical.
A dialog box asks you if you want to open or save the file.
5. Click Save and save the CiblesCalibrant.xml file in the root directory of a USB key.
Given that the .xml file has always the same name, you must import the calibrators one
by one.
6. On the Pentra C400, enter Main menu > Calibration/Control > Calibration.
7. Press Import:
HORIBA Medical recommends you to make sure that all materials like CD-ROM or USB
key used on the Pentra C400 and non-referenced by HORIBA Medical are free of
viruses. These materials must imperatively be cleaned before being used on the
instrument.
■ After you imported the calibrator configuration (calibrator characteristics and target values),
configure the calibrator position.
■ After you modified the target values, a new calibration must be performed with the default controls
if configured in the application.
Related information:
■ To Configure the Calibrator Position, p.234
Follow this procedure to modify the calibrator name, lot number and expiration date.
Follow this procedure to change the calibrator lot number and expiration date.
After you changed the calibrator lot number, modify the target values.
Related information:
■ To Add Target Values, p.233
Always ensure that the entered target values correspond to the calibrator name and lot
number used. Refer to the calibrator annex.
3. Press Edit.
This function is unavailable if associated patient or control tests are ordered, pending, incomplete
or for validation.
4. Modify either the main standard value or the target value for each calibrator level.
a. If the application features a calibration pre-dilution with Main direct or Main indirect dilution
type, type the main standard value.
The target value for each calibrator level is automatically calculated.
b. If the application does not feature a calibration pre-dilution or features a calibration pre-dilution
with Factor diluent dilution type, type the target value for each calibrator level in ascending or
descending order.
If the application features a calibration pre-dilution with Main direct or Main indirect
dilution type and a sample pre-dilution is programmed in the application, the calibrator
target values entered in the Target Values Entry screen have to be multiplied by the
pre-dilution factor. Example: If a sample pre-dilution with a pre-dilution factor of 21 is
programmed in the application and the calibrator target value mentions 2.22 g/L, the
value that must be entered is 2.22 * 21 = 46.62 g/L.
5. If you want to perform a recalibration with a single calibrator level, select the Standard Correction
option and choose the calibrator level from the spin box.
This function is unavailable if:
■ The calibration mode configured in the application is Factor or Slope average,
■ A calibration pre-dilution with Main direct or Main indirect dilution type is configured in the
application,
■ No calibration was previously performed for the test.
6. Press OK to validate.
If a calibration for this test is ordered, pending, incomplete or for validation or in case a current
calibration exists for this test, a dialog box informs you that the calibration will be lost.
7. Press OK to validate or Cancel to cancel.
After you modified the target values, a new calibration must be performed with the default controls if
configured in the application.
The Control tab lists the controls configured on the instrument. Controls are listed in a table format
which indicates the control name, the rack position of control, the control lot number and expiration
date.
Related information:
■ To Enter the Control Characteristics, p.242
■ To Add Target Values, p.242
■ To Configure the Control Position, p.244
Follow this procedure to enter the control name, lot number and expiration date.
After you entered the control characteristics, add the target values and configure the control position.
Related information:
■ To Add Target Values, p.242
■ To Configure the Control Position, p.244
To add the control target values, the control characteristics must have already been entered.
Always ensure that the entered target values correspond to the control name and lot
number used. Refer to the control annex.
After you added the target values, configure the control position.
Related information:
■ To Enter the Control Characteristics, p.242
■ To Configure the Control Position, p.244
To configure the control position, the control characteristics and the target values must already be
entered.
The racks used on the instrument must be previously configured in the Main menu >
Services > System Configuration > Analyser.
The selected rack is displayed (left side of the screen). The status of each position is indicated by
a color code.
Related information:
■ To Configure Racks, p.307
■ To Enter the Control Characteristics, p.242
■ To Add Target Values, p.242
Follow this procedure to import the control configuration (control characteristics and
target values) provided by HORIBA Medical.
A dialog box asks you if you want to open or save the file.
5. Click Save and save the CiblesControle.xml file in the root directory of a USB key.
Given that the .xml file has always the same name, you must import the controls one
by one.
6. On the Pentra C400, enter Main menu > Calibration/Control > Control.
7. Press Import:
HORIBA Medical recommends you to make sure that all materials like CD-ROM or USB
key used on the Pentra C400 and non-referenced by HORIBA Medical are free of
viruses. These materials must imperatively be cleaned before being used on the
instrument.
After you imported the control configuration (control characteristics and target values), configure the
control position.
Related information:
■ To Configure the Control Position, p.244
Follow this procedure to modify the control name, lot number and expiration date.
Follow this procedure to change the control lot number and expiration date.
After you changed the control lot number, modify the target values.
Related information:
■ To Add Target Values, p.242
Always ensure that the entered target values correspond to the control name and lot
number used. Refer to the control annex.
3. Press Edit.
This function is unavailable if associated tests are ordered, pending, incomplete or for validation.
4. Modify the control target value and confidence range.
5. To configure the control as a default control for this test, select the Default Control option.
Up to three default controls can be configured for a test on condition that the default controls are
required in the application.
The default controls already configured for the test are listed at the bottom of the screen.
6. Press OK to validate.
If the control was configured as a default control for this test, a new default control must be configured
for this test.
Related information:
■ To Add Target Values, p.242
If the control was configured as a default control for one or several tests, a new default control must
be configured for these tests.
Related information:
■ To Add Target Values, p.242
To operate correctly, the ISE module remains on, 24 hours a day, and is calibrated regularly whether
the Pentra C400 is operational or shut down.
■ The 2-point calibration determines a slope and an intercept using the low and high standards (of
known concentrations). The determined slope and intercept are then used to calculate the
concentrations of patient samples.
The 2-point calibration should be performed every 2 hours whether the Pentra C400 is operational
or shut down.
■ The 1-point calibration checks the electrode stability and recalculates the intercept using the low
standard.
This calibration should be performed every 15 minutes only when the Pentra C400 is operational.
■ The Ready cycle ensures the electrode stability when the Pentra C400 is operational.
This cycle circulates low standard through the electrodes and should be performed every 5
minutes.
■ The Keep ready cycle ensures the electrode stability when the Pentra C400 is shut down.
This cycle circulates low standard through the electrodes and should be performed every 15
minutes.
The ISE module calibrations as well as the Ready and Keep ready cycles are configured
by default as mentioned above. To modify this configuration, refer to the ISE Calibration
Configuration > To Modify ISE Calibration Configuration chapter.
Related information:
■ To Modify ISE Calibration Configuration, p.251
The ISE module calibrations as well as the Ready and Keep ready cycles are configured
by default. Follow this procedure to modify the default configuration.
1. Press Edit.
2. To enable or to disable a calibration or a cycle, select or deselect the corresponding check box.
3. To change the frequency of a calibration or a cycle, press the up or down arrows from the
corresponding spin box.
Frequency
Parameter Configurable range Default value (recommended)
1-point calibration 5-480 min. 15 min.
2-point calibration 15-480 min. 120 min.
Keep ready 10-480 min. 15 min.
Ready 5-15 min. 5 min.
4. Press OK to validate.
3. Application Configuration
Access: Main menu > Services > Application Configuration > Applications
Many applications are provided by HORIBA Medical. These applications are maintained and updated
by HORIBA Medical with the Reagent Application media (USB flash drive).
Your local representative may also provide you with some applications.
The applications provided by HORIBA Medical or your local representative are configured on reserved
channels. Only a few parameters are modifiable in these applications.
User channels allow you to add your own applications. The number of user channels is limited and
configured by your local HORIBA Medical representative. All parameters are configurable and
modifiable in these applications.
The applications are listed in a table format which gives the following information.
Heading Description
Code Application short name.
Local Code Application short name defined by the user.
Enable If selected, the application is enabled and then selectable in the request
capture screen.
Unit Unit used to display the results.
Reference Range Low Low limit value of the Man/Default reference range
Reference Range High High limit value of the Man/Default reference range
Calibrator Used Name of the calibrator used for this test.
Contr. Required If selected, default controls are required to validate the calibration.
Up to three default controls can be configured.
Control Used Name of the default controls used for this test
Reagent Short Name Short name of the reagent used for this test.
Cassette If selected, the reagent is in a cassette.
Release Application version number.
Modified on Application modification date and time.
■ The version number of the last Reagent Application media installed on the instrument
is displayed in the Current revision field.
■ An item in gray means that the application was not updated during the last Reagent
Application media installation.
Related information:
■ Managing Absorbance Applications, p.272
Access: Main menu > Services > Application Configuration > Applications > General Parameters
■ Channel: The channels from 1 to 800 are reserved for the applications provided by
HORIBA Medical. Your local representative may also provide you with some applications on the
channels from 801 to "800 + X" where X is the number of channels reserved for your local
representative. Depending on the number of user channels configured on your instrument, the
user channels are from "1000 - Y" to 999 where Y is the number of user channels configured on
your instrument. Example: If the number of channels reserved for your local representative is set to
10 and the number of user channels is set to 5, the channels from 801 to 810 ("800 + 10") are
reserved for your local representative and the user channels are from 995 ("1000 - 5") to 999.
Characteristics
Pre-dilution
For detailed information concerning the dilution procedure, refer to the Description and
Technology > Sampling System > Dilutions Management chapter.
■ Incubation time (in cycles): For applications not linked to a ratio, this incubation time represents
the minimum time of delay between the dilution and its use in the analysis sequence.
For applications linked to a ratio, this time is fixed. The analysis sequence cycles and cleaning
cycles (if a needle cleaning is required) must be taken into account for the incubation time
configuration.
That is illustrated by the example below.
Cycle # 1 2 3 4 5 6 7 8
R1 R2 + Sample
Incubation time
Sample
pre-dilution
Result
■ Unit: Units must be previously configured in Main menu > Services > Application Configuration
> Applications > Unit Parameters.
■ Manual Patient Validation: If unselected, only patient samples that are not automatically
validated must be manually validated. It then depends on the patients automatic validation
configured in Main menu > Services > System Configuration > Results Validation.
Correlation
The Correlation area allows the user to convert results in order to correlate them with an alternative
method or another temperature. This result correlation area is dedicated to the user.
The intercept is always expressed in the reference unit. If an alternative unit is chosen to
display the results, it is necessary to make the following calculation before programming
the intercept: B = b / FU where:
■ B is the intercept to be programmed in the application,
■ b is the intercept to be applied to the final result,
■ FU is the unit conversion factor. The unit conversion factor is displayed in Main menu
> Services > Application Configuration > Applications > Unit Parameters.
For example, in an application the reference unit is g/L and the alternative unit mmol/L is
used. 1 g/L converts to 10 mmol/L, so in this example FU = 10. The user wishes to add an
intercept of 20 mmol/L to all results. In this case, B = b / FU = 20 / 10 = 2. 2 must be
programmed in the Intercept field.
Linearity
■ Linearity range check: The sample result is checked to ensure that it is within the programmed
linearity range. If not, the analytical flag LINEARITY_HIGH or LINEARITY_LOW is triggered and an
automatic rerun with post-dilution or post-concentration is performed if configured in the
application.
Automatic Rerun
■ Post-dilution:
■ Post-dilution without pre-dilution configured:
If, at each analysis line where a "SAMPLE" pipetting is programmed, a H2O volume to push the
sample volume is programmed and SV ≥ FPost-dil * 2, then the dilution is performed in the
needle and no cuvette is used for dilution.
If one of the two conditions above is not met, then the dilution procedure applies with
FD = FPost-dil
■ Post-dilution with pre-dilution configured:
The dilution procedure applies with FD = FPost-dil * FPre-dil
FPost-dil * FPre-dil must be ≤ 22500.0.
For detailed information concerning the dilution procedure, refer to the Description and
Technology > Sampling System > Dilutions Management chapter.
FD = FPre-dil / FPost-conc
If FPost-conc ≤ FPre-dil / 1.2, then the dilution procedure applies with FD = FPre-dil / FPost-conc
If FPost-conc = FPre-dil then FD = 1 and no dilution is performed.
■ Post-concentration without pre-dilution and with H2O volume configured:
One of the following conditions must be met:
■ FPost-conc ≤ (SV + H2OV - 2) / SV or
■ FPost-conc = (SV + H2OV) / SV
FD = 1 / FPost-conc and the dilution (or concentration) is performed in the needle and no cuvette
is used for dilution.
■ Post-concentration with pre-dilution and with H2O volume configured:
One of the following conditions must be met:
■ FPost-conc ≤ FPre-dil * (SV + H2OV - 2) / SV and FPost-conc ≥ 2 * FPre-dil / SV or
■ FPost-conc = FPre-dil * (SV + H2OV) / SV or
■ FPost-conc ≤ FPre-dil / 1.2
FD = FPre-dil / FPost-conc
If FD < 1.2, then the dilution (or concentration) is performed in the needle and no cuvette is
used for dilution.
If FPost-conc ≤ FPre-dil / 1.2, then the dilution procedure applies with FD = FPre-dil / FPost-conc
For detailed information concerning the dilution procedure, refer to the Description and
Technology > Sampling System > Dilutions Management chapter.
Delta Check
■ Delta check: The delta check is used to check the variation between the current result and the
previous result to ensure that it is not higher than the authorized absolute and/or relative
variations. If it is, the delta check flag (D+ or D-) is triggered.
The delta check is performed only if:
Run date of the current result - Run date of the previous result ≤ Delta Check Validity
where the previous result is a result for this test, previously run and validated, for the same patient
but for a different sample.
Reference Range
■ Low reference value check: The sample result is checked to ensure that it is not lower than the
low limit value of the reference range. If it is, the analytical flag REF_RANGE_LOW is triggered.
If the Low Check option is selected, at least the low limit value of the Man/Default reference range
must be programmed.
■ High reference value check: The sample result is checked to ensure that it is not higher than the
high limit value of the reference range. If it is, the analytical flag REF_RANGE_HIGH is triggered.
If the High Check option is selected, at least the high limit value of the Man/Default reference
range must be programmed.
Rerun Range
■ Low critical value check: The sample result is checked to ensure that it is not lower than the low
limit value of the critical range. If it is, the analytical flag CRITICAL_LOW is triggered and the
sample is automatically rerun without post-concentration (a new test is requested).
If the Low Check option is selected, at least the low limit value of the Man/Default critical range
must be programmed.
■ High critical value check: The sample result is checked to ensure that it is not higher than the
high limit value of the critical range. If it is, the analytical flag CRITICAL_HIGH is triggered and the
sample is automatically rerun without post-dilution (a new test is requested).
If the High Check option is selected, at least the high limit value of the Man/Default critical range
must be programmed.
Related information:
■ Unit Parameters, p.271
■ Dilutions Management, p.459
■ To Register a Reagent, p.299
■ To Configure Patients Automatic Validation, p.327
Access: Main menu > Services > Application Configuration > Applications > Calibration Parameters
Calibration
■ Calibration mode and Level: The calibration modes and the number of calibrator levels that they
require are described in the following table.
■ Dev_Rep (%) and Dev_C (%): Two types of deviation are checked.
■ Deviation concerning replicates (Dev_Rep): For each calibrator level, the absorbance values
of replicates are compared with the mean value. The maximum deviation is calculated
(Dev_Rep) and then checked to ensure that it is not higher than the authorized Dev_Rep. If it is,
the analytical flag DEVIATION is triggered.
■ Deviation to the curve (Dev_C): For each calibrator level, the concentration of the mean value
is compared with the theoretical value. The deviation is calculated (Dev_C) and then checked
to ensure that it is not higher than the authorized Dev_C. If it is, the analytical flag DEVIATION
is triggered.
Validity
The Validity area allows you to configure the calibration validity. Two options are available:
■ On Request: The current calibration is valid until a new calibration is requested.
■ Time Validity: The current calibration is valid for the defined period. When the defined period has
expired, the current calibration is disabled. As soon as a test for this application is requested, a
new calibration is automatically ordered.
Factor calibration
The Factor calibration area allows you to configure the calibration factor checks. These checks apply
to the calculated calibration factor and are only available for Slope average and Linear regression
calibration modes.
■ Low limit check: The calculated calibration factor is checked to ensure that it is not lower than
the authorized low limit value. If it is, the analytical flag FACTOR_FLAG is triggered.
■ High limit check: The calculated calibration factor is checked to ensure that it is not higher than
the authorized high limit value. If it is, the analytical flag FACTOR_FLAG is triggered.
■ Relative variation check: The relative variation check consists in checking the variation between
the calculated calibration factor and the previous validated calibration factor to ensure that it is not
higher than the authorized relative variation. If it is, the analytical flag FACTOR_FLAG is triggered.
Control required
■ If the Control required option is selected, the required default controls must be configured for this
test in Main menu > Calibration/Control > Control.
The default controls are automatically performed with the calibration. Ordered sample tests are not
carried out until the calibration and the default controls are validated.
To deselect the required default controls, the default controls configured for this test must be
previously disabled in Main menu > Calibration/Control > Control.
Validity Backup
The Validity Backup area allows you to configure a period during which the current calibration stays
valid when a new cassette, with the same lot number as the previous one, is activated (manually or
automatically). This period is called back-up time frame.
If the Backup time frame without calibration required option is selected and when a new cassette
(n+1), with the same lot number as the previous one (n), is activated (manually or automatically), the
following checks are performed:
■ If Activation date of cassette (n+1) - Installation date of cassette (n) > Back-up time frame, then the
current calibration is disabled and a new calibration must be performed with the default controls if
configured in the application.
■ If Activation date of cassette (n+1) - Installation date of cassette (n) ≤ Back-up time frame, then the
installation date of cassette (n+1) is checked.
■ If Installation date of cassette (n+1) < Installation date of cassette (n), then the current
calibration is disabled and a new calibration must be performed with the default controls if
configured in the application.
■ If Installation date of cassette (n+1) ≥ Installation date of cassette (n), then the current
calibration stays valid and the default controls if configured in the application must be
performed.
Pre-dilution
■ Main direct: The main standard is diluted in order to obtain a series of standards. Each dilution
is performed starting from the main standard. Dilution factors must be captured by the user.
■ Main indirect: The main standard is diluted in order to obtain a series of standards. Each
dilution is performed starting from the previous one. Dilution factors must be captured by the
user.
■ Factor diluent: This dilution type is only available if a sample pre-dilution has been configured
in the application. The series of standards is pre-diluted as patient samples and controls (with
the same dilution factor and incubation time). Factor fields are unavailable.
■ Factor 1 to Factor 8: For Main direct and Main indirect dilution types, dilution factors must be
captured by the user. The number of dilution factors to be captured equals the number of
calibrator levels configured in the application.
Factors are configured as follows:
■ 0: This means that no dilution has to be performed for this level, pure diluent is dispensed.
■ 1: This means that no dilution has to be performed for this level, pure main standard is
dispensed.
■ From 2.0 to 150.0: The corresponding dilution is performed with the configured dilution factor,
the associated concentration is calculated accordingly.
Using the Main direct dilution type, dilution factors must be captured in ascending or descending
order.
■ In ascending order, only the first factor can be equal to 1 and only the last factor can be equal
to 0.
■ In descending order, only the first factor can be equal to 0 and only the last factor can be equal
to 1.
Using the Main indirect dilution type, the only restrictions on dilution factors are that only the first
factor can be equal to 1 and only the last factor can be equal to 0.
Checks
The Checks area is only available if the reagent blank has been previously programmed in Main
menu > Services > Application Configuration > Applications > Analysis Parameters.
■ Reagent range check: All reagent blank absorbance values used for calculations are checked to
ensure that they are not out of the programmed reagent range. If they are, the analytical flag
REAG_RANGE_HIGH or REAG_RANGE_LOW is triggered.
■ Blank range check: The absorbance variation of the reagent blank is checked to ensure that it is
not out of the programmed blank range. If it is, the analytical flag BLK_RANGE is triggered.
Related information:
■ Analysis Parameters, p.265
■ To Add Target Values, p.233
■ To Add Target Values, p.242
Access: Main menu > Services > Application Configuration > Applications > Analysis Parameters
Cleaner
The Cleaner area allows you to require a needle cleaning and to configure this cleaning.
■ If a needle cleaning is required, the cleaning cycles must be taken into account for the analysis
sequence cycles configuration. In addition, if a sample pre-dilution is programmed, the cleaning
cycles must also be taken into account for the incubation time configuration.
Wavelength
Blank
■ Reagent Blank: If unselected, the Checks area in Main menu > Services > Application
Configuration > Applications > Calibration Parameters is unavailable.
Analysis Sequence
Up to four analysis lines are programmed in the analysis sequence. Each analysis line is configured
with the following parameters.
■ Cycle: For the first analysis line, the cycle number is set to 1 and cannot be modified. For the
other analysis lines, the cycle numbers must be captured in ascending order.
If a needle cleaning is required, the cleaning cycles must be taken into account for the analysis
sequence cycles configuration. In addition, if a sample pre-dilution is programmed, the incubation
time of pre-diluted sample must also be taken into account for the analysis sequence cycles
configuration.
If a needle cleaning is required and the cleaning cycle is carried out after R1 has been
dispensed, then the cycle number following R1 dispensing must be:
■ ≥ R1 dispensing cycle number + 2, if one cleaning cycle is performed,
■ ≥ R1 dispensing cycle number + 3, if two cleaning cycles are performed.
■ Reagent Needle and Sample Needle: For the first analysis line, a solution must be selected from
the Reagent Needle dropdown list.
R1, R2 and/or R3 are pipetted once in the analysis sequence.
A solution cannot be pipetted twice in the same analysis line.
Moreover, if no sample is pipetted in the analysis sequence, sample pre-dilution, automatic rerun
and calibration pre-dilution must not be configured in the application.
■ Volume (µL) and H2O Vol (µL): For the first analysis line, the volume pipetted by the reagent
needle must be ≥ 95.0 µL.
For each analysis line, the total volume pipetted by the sample needle must be ≤ 95.0 µL.
A "SAMPLE" pipetting can be programmed at each analysis line. In this case, the total maximum
sample volume is 4 * 95.0 µL = 380 µL.
The addition of all volumes pipetted in the analysis sequence must be included between 150.0 and
600.0 µL.
Mixing Speed
HORIBA Medical recommends not to use mixing speeds lower than 40%.
Related information:
■ General Parameters, p.254
■ Calibration Parameters, p.260
Access: Main menu > Services > Application Configuration > Applications > Calculation Parameters
Correlation Factor
The Correlation Factor area allows the conversion of results in order to correlate them with an
alternative method or another temperature. This result correlation area is dedicated to
HORIBA Medical.
Reaction Direction
■ Reaction direction check: The reaction direction is checked to ensure that it corresponds to the
expected reaction direction. If not, the analytical flag SIGN is triggered.
■ Sample limit check: The sample limit value is used to determine the sample-specific absorbance.
The reagent blank absorbance is subtracted from the sample absorbance, both measured at the
defined cycle. The calculated value is checked to ensure that it is not higher than the programmed
sample limit value. If it is, the analytical flag SAMPLE_LIMIT is triggered.
The antigen excess check is configurable only if the application answers to the three
following criteria:
■ One of the following calibration modes is used: Linear regression, Linear interpolation,
LOGIT/LOG4, LOGIT/LOG5 or EXPONENT5.
■ Only one calculation step is configured.
■ The Endpoint calculation type is used.
■ Antigen excess check: The antigen excess check is used to detect the presence of excess
antigen. The concentration measured at the defined cycle (CCP) is divided by the concentration
measured at the last reading cycle (C) and multiplied by 100. The calculated ratio is checked to
ensure that it is higher than the programmed antigen excess limit value. If not, the analytical flag
AG_EXCESS is triggered and an automatic rerun with post-dilution is performed if configured in
the application.
Please note that the concentrations used for this calculation do not take into account
the result correlations (user or HORIBA Medical) nor the unit conversion.
The areas Definition, Steps and Formula allow you to configure the reaction rate calculation.
■ The reaction can be divided in up to four calculation steps. The number of calculation steps is
defined in the Steps area.
■ For each calculation step, the step rate calculation is configured in the Definition area.
■ Then, the global reaction rate calculation is defined in the Formula area.
Steps
The Steps area allows you to define the number of calculation steps for the reaction rate calculation.
Definition
The Definition area is divided into four tabs: Step A, Step B, Step C and Step D. This allows you to
configure the step rate calculation for each calculation step according to the number of calculation
steps defined in the Steps area.
■ Calculation Type: For detailed information concerning the calculation types, refer to the
Description and Technology > Data Analysis > Absorbance Measurements > Calculation Types
chapter.
■ Reaction limit check: The reaction limit value is used to determine substrate depletion. For the
Kinetic calculation type, the absorbance measured at the defined cycle is subtracted from the
absorbance value of the first reading point. For the Kinsearch calculation type, the absorbance
measured at the defined cycle is subtracted from the absorbance value of the first point within the
linear range. The calculated value is checked to ensure that it is not higher than the programmed
reaction limit value. If it is, the analytical flag REAC_LIMIT is triggered and an automatic rerun with
post-dilution is performed if configured in the application.
■ First Reading Cycle: In order to read correctly, the volume into the cuvette must be greater than
150 µL.
■ If the total volume pipetted in the first analysis line is lower than 150 µL, then the first reading
cycle must be ≥ 2.
■ If the total volume pipetted in the first analysis line is greater than 150 µL but the volume
pipetted by the reagent needle is lower than 150 µL, then the first reading cycle must be ≥ 1.
■ If the volume pipetted by the reagent needle in the first analysis line is greater than 150 µL,
then the first reading cycle could be 0. This means that the first reading takes place after the
reagent needle dispensing and before the sample needle dispensing.
■ Last Reading Cycle: The last reading cycle should answer to the three following criteria:
■ last reading cycle > first reading cycle,
■ last reading cycle ≥ last cycle number of the analysis sequence,
■ for Kinetic and Kinsearch calculation types, last reading cycle - first reading cycle ≥ 4.
OD deviation check
The OD deviation check area allows you to configure the absorbance deviation checks. These
checks are performed during the reaction rate calculation in order to flag optical density jumps and
are configurable for each calculation step according to the number of calculation steps defined in the
Steps area.
■ Absorbance deviation check on linear regression: The correlation coefficient r2 and the
standard deviation SD are calculated. For the Kinetic calculation type, the measurement points
used to calculate r2 and SD are those included between the first reading cycle and the last reading
cycle. For the Kinsearch calculation type, the measurement points used to calculate r2 and SD are
those included in the linear range found with the linear search program. If r2 is lower than the
configured r2 threshold and SD is higher than the configured SD threshold, then the analytical flag
DIV_ABS is triggered.
■ Absorbance deviation check on first point: Two different algorithms can be used for this check:
one if the first reading cycle < 4, another if the first reading cycle ≥ 4.
■ If the first reading cycle < 4, a calculated absorbance difference is checked to ensure that it is
included in the configured blank range. If not, the analytical flag DIV_ABSi is triggered.
The blank range must be programmed in Main menu > Services > Application
Configuration > Applications > Calibration Parameters.
■ If the first reading cycle ≥ 4, three calculated absorbance differences are checked to ensure
that they are lower than the configured threshold. If at least two absorbance differences are
greater than or equal to the configured threshold, the analytical flag DIV_ABSi is triggered.
The first point threshold is used and must then be configured only when the first
reading cycle ≥ 4.
■ Absorbance deviation check on last point: The difference between absorbance values at time n
and at time n-1 (where n = last reading cycle) is calculated and checked to ensure that it is
included in a calculated range. If not, the analytical flag DIV_ABSf is triggered.
This algorithm applies only when the last reading cycle is ≥ 11.
For detailed information concerning the absorbance deviation checks, refer to the
Description and Technology > Data Analysis > Absorbance Measurements > Calculation
Types > Absorbance Deviation Checks chapter.
Formula
The Formula area allows you to define the global reaction rate calculation when several calculation
steps are configured.
Related information:
■ General Parameters, p.254
■ Calibration Parameters, p.260
■ Absorbance Deviation Checks, p.473
The Unit Parameters tab allows you to configure several units that can be used to display the results.
These units are then selectable from the Unit dropdown list (in Main menu > Services > Application
Configuration > Applications > General Parameters).
The first Unit and Conversion Factor fields correspond to the reference unit for this application. The
reference unit is mandatory and selectable from the Unit dropdown list. The conversion factor is set
to 1.000000 and cannot be modified.
The five other Unit and Conversion Factor fields correspond to the other units configured for this
application. Each unit is selectable from the Unit dropdown list and the conversion factor between the
reference unit and the chosen unit must be captured in the Conversion Factor field (from 10-6 to 107).
This multiplication factor is then used to display the results in the chosen unit.
For the applications provided by HORIBA Medical or your local representative, the reference unit and
up to five other units are configured and cannot be modified.
For the user applications, the reference unit and one other unit are programmable.
The table below lists the units that can be configured on the instrument.
Related information:
■ General Parameters, p.254
Access: Main menu > Services > Application Configuration > Applications
4. Press OK to validate.
Related information:
■ Absorbance Applications Parameters, p.253
Access: Main menu > Services > Application Configuration > Applications
1. Select the application of which you want to copy parameters and press Duplicate.
This function is unavailable if all user channels are already configured or no reagent was registered
by your local representative or yourself.
The absorbance applications configuration screen is displayed with the selected application
parameters.
2. Type the test name (15 characters maximum), the application code (7 characters maximum) and
the application local code (7 characters maximum).
The test name, the application code and the application local code must be unique.
3. Change the selected application parameters where necessary in the five tabs of the screen:
■ General Parameters
■ Calibration Parameters
■ Analysis Parameters
■ Calculation Parameters
■ Unit Parameters
4. Press OK to validate.
Related information:
■ Absorbance Applications Parameters, p.253
Access: Main menu > Services > Application Configuration > Applications
Only a few parameters are modifiable in the applications provided by HORIBA Medical or your
local representative. These parameters are called user parameters. For detailed information
concerning user parameters, refer to the Application Configuration > Updating the Applications >
Major, Minor and User Parameters chapter.
All parameters are modifiable in the user applications. However, the following restrictions apply:
■ The number of calibrator levels and the calibration pre-dilution (in Main menu > Services >
Application Configuration > Applications > Calibration Parameters) cannot be modified
without deleting the calibrator target values for this test first.
■ If the application is linked to a profile, the sample type (in Main menu > Services >
Application Configuration > Applications > General Parameters) cannot be modified
without deleting the profile first.
■ In case a current calibration exists for this test, the calibration will be lost if a major parameter
is modified. For detailed information concerning major parameters, refer to the Application
Configuration > Updating the Applications > Major, Minor and User Parameters chapter.
4. Press OK to validate.
In case a current calibration exists for this test and if a major parameter was modified, a dialog box
informs you that the calibration will be lost.
5. Press OK to validate or Cancel to cancel.
Related information:
■ Absorbance Applications Parameters, p.253
■ Major, Minor and User Parameters, p.282
Access: Main menu > Services > Application Configuration > Applications
Access: Main menu > Services > Application Configuration > Applications
For detailed information concerning the procedure to modify the ISE applications, refer to
the ISE Applications > To Modify an ISE Application chapter.
Characteristics
Result
■ Manual Patient Validation: If unselected, only patient samples that are not automatically
validated must be manually validated. It then depends on the patients automatic validation
configured in Main menu > Services > System Configuration > Results Validation.
Correlation
The Correlation area allows the user to convert results in order to correlate them with an alternative
method or another temperature. This result correlation area is dedicated to the user.
The intercept is always expressed in the reference unit. If an alternative unit is chosen to
display the results, it is necessary to make the following calculation before programming
the intercept: B = b / FU where:
■ B is the intercept to be programmed in the application,
■ b is the intercept to be applied to the final result,
■ FU is the unit conversion factor.
For example, in an application the reference unit is g/L and the alternative unit mmol/L is
used. 1 g/L converts to 10 mmol/L, so in this example FU = 10. The user wishes to add an
intercept of 20 mmol/L to all results. In this case, B = b / FU = 20 / 10 = 2. 2 must be
programmed in the Intercept field.
Reference Range
■ Low reference value check: The sample result is checked to ensure that it is not lower than the
low limit value of the reference range. If it is, the analytical flag REF_RANGE_LOW is triggered.
If the Low Check option is selected, at least the low limit value of the Man/Default reference range
must be programmed.
■ High reference value check: The sample result is checked to ensure that it is not higher than the
high limit value of the reference range. If it is, the analytical flag REF_RANGE_HIGH is triggered.
If the High Check option is selected, at least the high limit value of the Man/Default reference
range must be programmed.
Rerun Range
■ Low critical value check: The sample result is checked to ensure that it is not lower than the low
limit value of the critical range. If it is, the analytical flag CRITICAL_LOW is triggered and the
sample is automatically rerun without post-concentration (a new test is requested).
If the Low Check option is selected, at least the low limit value of the Man/Default critical range
must be programmed.
■ High critical value check: The sample result is checked to ensure that it is not higher than the
high limit value of the critical range. If it is, the analytical flag CRITICAL_HIGH is triggered and the
sample is automatically rerun without post-dilution (a new test is requested).
If the High Check option is selected, at least the high limit value of the Man/Default critical range
must be programmed.
Delta Check
■ Delta check: The delta check is used to check the variation between the current result and the
previous result to ensure that it is not higher than the authorized absolute and/or relative
variations. If it is, the delta check flag (D+ or D-) is triggered.
The delta check is performed only if:
Run date of the current result - Run date of the previous result ≤ Delta Check Validity
where the previous result is a result for this test, previously run and validated, for the same patient
but for a different sample.
Control required
■ If the Control required option is selected, the required default controls must be configured for this
test in Main menu > Calibration/Control > Control.
Ordered sample tests are not carried out until the default controls are validated.
To deselect the required default controls, the default controls configured for this test must be
previously disabled in Main menu > Calibration/Control > Control.
Linearity
■ Linearity range check: The sample result is checked to ensure that it is within the programmed
linearity range. If not, the analytical flag LINEARITY_HIGH or LINEARITY_LOW is triggered.
Related information:
■ To Modify an ISE Application, p.279
■ To Configure Patients Automatic Validation, p.327
■ To Add Target Values, p.242
Access: Main menu > Services > Application Configuration > Applications
For detailed information concerning ISE applications parameters, refer to the ISE
Applications > ISE Applications Parameters chapter.
Only a few parameters are modifiable in the ISE applications (provided by HORIBA Medical).
These parameters are called user parameters. For detailed information concerning user
parameters, refer to the Application Configuration > Updating the Applications > Major, Minor and
User Parameters chapter.
4. Press OK to validate.
Related information:
■ Major, Minor and User Parameters, p.282
■ ISE Applications Parameters, p.275
Follow this procedure to configure the applications order, which is then used to display
the tests in the request capture screen, in the result validation screen and to print patient
results.
Access: Main menu > Services > Application Configuration > Applications
1. Press Order.
The applications order configuration screen is displayed.
2. Select the applications to be displayed from the following options.
■ Serum/Plasma: To display the applications which sample type is Serum/Plasma.
■ Urine: To display the applications which sample type is Urine.
■ Other: To display the applications which sample type is Other.
3. Press Edit.
4. To modify the applications order:
a. Select an application.
b. Press Up to move the application up or press Down to move the application down.
c. Repeat steps a and b for each application.
When a new application is configured on the instrument, this one takes the last
position in the applications order.
5. Press OK to validate.
Follow this procedure to update the applications provided by HORIBA Medical on your
instrument.
Access: Main menu > Services > Application Configuration > Applications
1. Press Import:
This function is unavailable if tests are ordered, pending, incomplete or for validation.
A dialog box allows you to select the support type:
■ CD-ROM,
■ USB key.
2. Install your support on the instrument.
The applications for which an update is available on your support are listed in two tables: one for
major parameter updates, another for minor parameter updates.
■ When a major parameter is updated in an application, the current calibration for this test is lost.
A new calibration must be performed. If the number of calibrator levels or the calibration pre-
dilution is modified, the calibrator target values for this test are also deleted and must be
reconfigured.
■ When a minor parameter is updated in an application, the current calibration for this test is
kept in memory. A new calibration is not required.
For detailed information concerning major and minor parameters, refer to the Updating the
Applications > Major, Minor and User Parameters chapter.
Each table indicates the application code, the current application version number and modification
date and time, the update application version number, if the application is selected to be updated.
4. Press Edit.
5. By default, all applications are selected to be updated. If necessary, deselect the applications that
should not be updated.
If an application is already present on the instrument with the same test name, the
same application code or the same application local code but a different channel
number, then the application is unselected by default and cannot be updated. The
following message is displayed: At least one application has the same name or code
with different channel in the media. In this case, the test name, the application code or
the application local code must be modified before updating this application.
6. Press OK to validate.
The restrictions below apply when updating the applications.
■ If an application is linked to a ratio and the parameters modified in this application are
incompatible with the ratio, then the application cannot be updated. In this case, the ratio must
be deleted before updating the applications.
■ If an application is linked to a profile and the sample type is modified in this application, then
the application cannot be updated. In this case, the profile must be deleted before updating
the applications.
■ If a reagent is already present on the instrument with the same reagent short name but a
different reagent number, then the reagent cannot be updated. In this case, the reagent short
name must be modified before updating the applications.
■ If the container type is changed for a solution having a reagent number between 600 and 799,
a dialog box informs you that the solution will be deleted from the Reagent Configuration
menu.
■ Press OK to validate. In this case, after the applications update, you will have to
reconfigure the container type in Main menu > Services > Application Configuration >
Reagents and the solution in Reagent Configuration menu.
■ Press Cancel to cancel the applications update.
■ When an update of the reagent online help is available on your support and at least one
application is unselected, a dialog box informs you that the whole reagent online help is going
to be updated whereas at least one application will not be updated.
■ Press OK to validate.
■ Press Cancel to cancel the applications update.
The selected applications are updated as well as:
■ the reagents used in these applications,
■ all diluents and cleaners,
■ the incompatibilities related to these applications and configured by HORIBA Medical,
■ the reagent online help (when an update is available on your support).
It takes a few minutes.
User parameters are never affected when the applications are updated. For detailed information
concerning user parameters, refer to the Updating the Applications > Major, Minor and User
Parameters chapter.
A dialog box informs you when the update is completed.
7. Press OK.
Related information:
■ To Modify an Absorbance Application, p.274
■ To Delete a Ratio, p.289
■ To Delete a Profile, p.293
■ To Modify a Reagent, p.299
■ Major, Minor and User Parameters, p.282
3.6. Ratios
Access: Main menu > Services > Application Configuration > Ratio
The Ratio tab lists the ratios configured on the instrument. Ratios are listed in a table format which
gives the following information.
Heading Description
Ratio name Ratio name
Enable If selected, the ratio is enabled and then selectable in the request capture
screen.
Unit Unit used to display the results.
Reference Range Low Low limit value of the Man/Default reference range
Reference Range High High limit value of the Man/Default reference range
Test A Name of the test A used in the ratio.
Test B If enabled, name of the test B used in the ratio.
Test C If enabled, name of the test C used in the ratio.
Test D If enabled, name of the test D used in the ratio.
Access: Main menu > Services > Application Configuration > Ratio
For detailed information concerning the procedures to manage the ratios, refer to the
Ratios > Managing Ratios chapter.
Definition
■ Enable: If one of the applications used in the ratio is disabled, then the ratio is automatically
disabled.
■ Sample Type: The applications used in a same ratio must have the same sample type configured.
Formula
The Formula area allows you to define the ratio calculation with the tests enabled.
Result
Reference Range
■ Low reference value check: The sample result is checked to ensure that it is not lower than the
low limit value of the reference range. If it is, the analytical flag REF_RANGE_LOW is triggered.
If the Low Check option is selected, at least the low limit value of the Man/Default reference range
must be programmed.
■ High reference value check: The sample result is checked to ensure that it is not higher than the
high limit value of the reference range. If it is, the analytical flag REF_RANGE_HIGH is triggered.
If the High Check option is selected, at least the high limit value of the Man/Default reference
range must be programmed.
Rerun Range
■ Low critical value check: The sample result is checked to ensure that it is not lower than the low
limit value of the critical range. If it is, the analytical flag CRITICAL_LOW is triggered and the
sample is automatically rerun without post-concentration (a new test is requested).
If the Low Check option is selected, at least the low limit value of the Man/Default critical range
must be programmed.
■ High critical value check: The sample result is checked to ensure that it is not higher than the
high limit value of the critical range. If it is, the analytical flag CRITICAL_HIGH is triggered and the
sample is automatically rerun without post-dilution (a new test is requested).
If the High Check option is selected, at least the high limit value of the Man/Default critical range
must be programmed.
Delta Check
■ Delta check: The delta check is used to check the variation between the current result and the
previous result to ensure that it is not higher than the authorized absolute and/or relative
variations. If it is, the delta check flag (D+ or D-) is triggered.
The delta check is performed only if:
Run date of the current result - Run date of the previous result ≤ Delta Check Validity
where the previous result is a result for this test, previously run and validated, for the same patient
but for a different sample.
Related information:
■ Managing Ratios, p.287
Access: Main menu > Services > Application Configuration > Ratio
For detailed information concerning ratios parameters, refer to the Ratios > Ratio
Parameters chapter.
5. Press OK to validate.
When validating the ratio, a check is performed on the cycles involved in pre-dilutions
of all the applications used in the ratio. A dialog box informs you if the pre-dilution
cycles are incompatible. For detailed information on pre-dilution cycles, refer to the
Application Configuration > Absorbance Applications > Absorbance Applications
Parameters > General Parameters chapter.
Related information:
■ Ratio Parameters, p.285
■ General Parameters, p.254
Access: Main menu > Services > Application Configuration > Ratio
1. Select the ratio of which you want to copy parameters and press Duplicate.
The ratios configuration screen is displayed with the selected ratio parameters.
2. Type the ratio name (7 characters maximum).
The ratio name must be unique and different from test names, application codes and application
local codes configured on the instrument.
3. Enter the ratio channel number.
The ratio channel number must be unique.
4. Change the selected ratio parameters where necessary.
For detailed information concerning ratios parameters, refer to the Ratios > Ratio
Parameters chapter.
5. Press OK to validate.
When validating the ratio, a check is performed on the cycles involved in pre-dilutions
of all the applications used in the ratio. A dialog box informs you if the pre-dilution
cycles are incompatible. For detailed information on pre-dilution cycles, refer to the
Application Configuration > Absorbance Applications > Absorbance Applications
Parameters > General Parameters chapter.
Related information:
■ Ratio Parameters, p.285
■ General Parameters, p.254
Access: Main menu > Services > Application Configuration > Ratio
For detailed information concerning ratios parameters, refer to the Ratios > Ratio
Parameters chapter.
4. Press OK to validate.
When validating the ratio, a check is performed on the cycles involved in pre-dilutions
of all the applications used in the ratio. A dialog box informs you if the pre-dilution
cycles are incompatible. For detailed information on pre-dilution cycles, refer to the
Application Configuration > Absorbance Applications > Absorbance Applications
Parameters > General Parameters chapter.
Related information:
■ Ratio Parameters, p.285
■ General Parameters, p.254
Access: Main menu > Services > Application Configuration > Ratio
Follow this procedure to add, in an existing ratio, a secondary ratio calculation using the
same tests than the main ratio. This avoids creating another main ratio with the same tests
and therefore this avoids running the same tests twice.
Access: Main menu > Services > Application Configuration > Ratio
For detailed information concerning ratios parameters, refer to the Ratios > Ratio
Parameters chapter.
7. Press OK to validate.
The sub-ratio is created but it is neither displayed in the Ratio tab nor selectable in the request
capture screen. Only the main ratio is displayed and selectable.
Related information:
■ Ratio Parameters, p.285
Access: Main menu > Services > Application Configuration > Ratio
For detailed information concerning ratios parameters, refer to the Ratios > Ratio
Parameters chapter.
4. Press OK to validate.
Related information:
■ Ratio Parameters, p.285
Access: Main menu > Services > Application Configuration > Ratio
3.7. Profiles
Access: Main menu > Services > Application Configuration > Profiles
The Profiles tab lists the profiles configured on the instrument. Profiles are listed in a table format
which indicates the profile name and the profile sample type.
Access: Main menu > Services > Application Configuration > Profiles
The applications included in a same profile must have the same sample type
configured.
Access: Main menu > Services > Application Configuration > Profiles
Access: Main menu > Services > Application Configuration > Profiles
3.8. Incompatibilities
Access: Main menu > Services > Application Configuration > Incompatibility
Test samplings are said to be incompatible when they can pose contamination problems if they are
performed one after the other.
A test sampling is defined by the application local code and the solution pipetted. To configure an
incompatibility, a first and a second test sampling must be specified in the order in which they must
not be performed. The following restrictions apply:
■ An incompatibility must be unique.
■ Both applications must be different.
■ Both applications must not be used in a same ratio.
■ ISE applications having the same sample type cannot configure an incompatibility.
The configured incompatibilities are taken into account when the instrument performs samplings.
Samplings are organized in order to avoid performing incompatible test samplings one after the other.
In case of incompatibility, the instrument inserts by order of priority a compatible test sampling or a
cleaning between both incompatible test samplings.
In case the cleaning is not performed, the second test sampling is not carried out and the test is
flagged with the sampling alarm relative to the cleaning failure. If both tests are already in progress,
the second test is flagged with I (INCOMPATIBILITY).
Incompatibilities are either configured by HORIBA Medical or added by the user. Those configured by
HORIBA Medical are maintained and updated at the same time as the applications by
HORIBA Medical with the Reagent Application media (refer to the Application Configuration >
Updating the Applications > To Update the Applications chapter).
The Incompatibility tab lists the incompatibilities configured on the instrument. Incompatibilities are
listed in a table format which gives the following information.
Heading Description
Needle Type Needle that performs both incompatible test samplings.
Reagent Needle or Sample Needle.
Code 1 Application local code of the first test sampling
Sampling Type 1 Solution pipetted of the first test sampling
Code 2 Application local code of the second test sampling
Heading Description
Sampling Type 2 Solution pipetted of the second test sampling
Cleaner Cleaner used when the instrument inserts a cleaning between both
incompatible test samplings.
Related information:
■ To Update the Applications, p.280
Access: Main menu > Services > Application Configuration > Incompatibility
For detailed information concerning the procedures to manage the incompatibilities, refer
to the Incompatibilities > Managing Incompatibilities chapter.
A test sampling is defined by the application local code and the solution pipetted. To configure an
incompatibility, a first and a second test sampling must be specified in the order in which they must
not be performed. The following restrictions apply:
■ An incompatibility must be unique.
■ Both applications must be different.
■ Both applications must not be used in a same ratio.
■ ISE applications having the same sample type cannot configure an incompatibility.
First Sequence Incompatibility
Cleaner
The Cleaner dropdown list allows you to configure which cleaner is used when the instrument inserts
a cleaning procedure between both incompatible test samplings.
Related information:
■ Managing Incompatibilities, p.295
Access: Main menu > Services > Application Configuration > Incompatibility
3. Press OK to validate.
Related information:
■ Incompatibility Parameters, p.294
Access: Main menu > Services > Application Configuration > Incompatibility
4. Press OK to validate.
Related information:
■ Incompatibility Parameters, p.294
Access: Main menu > Services > Application Configuration > Incompatibility
3.9. Reagents
Access: Main menu > Services > Application Configuration > Reagents
Any reagent to be used on the Pentra C400 must be previously registered on the instrument.
HORIBA Medical provides a full range of reagents. These reagents are already pre-configured on the
instrument.
Your local representative may also provide you with some reagents and register them on your
instrument.
The Reagents tab allows you to add your own reagents.
The Reagents tab lists the solutions registered on the instrument. Solutions are listed in a table
format which gives the following information.
Heading Description
Reagent Short Name Solution short name
Reagent # Reagent number.
Solution type Solution type: Reagent, Diluent or Cleaner
Solution information Solution information: Deproteinization or Etching solution (only for cleaner).
On Board Stability Reagent on board stability in days (only for cassette).
Number of Reagents Number of reagents: Reagent 1, Reagent 2 or Reagent 3
Cassette If selected, the reagent is in a cassette.
Dead volume Dead volume used to calculate the solution initial volume into the cassette
(only for cassette).
Sensibility Sensitivity level used for the reagent level detection.
Access: Main menu > Services > Application Configuration > Reagents
For detailed information concerning the procedures to manage the reagents, refer to the
Reagents > Managing Reagents chapter.
■ Reagent #: The reagent numbers from 1 to 799 are reserved for the reagents provided by
HORIBA Medical. These reagents are already pre-configured on the instrument. The reagent
numbers from 800 to 999 can be used to register new reagents on the instrument.
Related information:
■ Managing Reagents, p.299
Access: Main menu > Services > Application Configuration > Reagents
After you registered a new reagent, configure this reagent in the Reagent Configuration menu.
■ If the reagent is on a reagent rack, refer to the Reagent Management > Managing Reagents on a
Reagent Rack > To Add a Reagent on a Reagent Rack chapter.
■ If the reagent is in an open cassette, refer to the Reagent Management > Managing Reagents in an
Open Cassette chapter.
Related information:
■ Managing Reagents in an Open Cassette, p.224
■ To Add a Reagent on a Reagent Rack, p.229
Access: Main menu > Services > Application Configuration > Reagents
■ If the solution was registered by HORIBA Medical, the reagent short name and the
solution type cannot be modified.
■ If the solution is programmed in an application or an incompatibility, the solution
type cannot be modified.
■ If the solution type is Reagent 3, the container type cannot be modified.
■ The reagent number and the sensitivity are never modifiable.
For detailed information concerning reagents parameters, refer to the Reagents >
Reagent Parameters chapter.
4. Press OK to validate.
If you changed the container type, a dialog box informs you that all associated calibrations
(ordered, pending, incomplete, for validation or validated) will be lost. In addition, the solution will
be deleted from the Reagent Configuration menu.
5. Press OK to validate or Cancel to cancel.
If you changed the container type, the solution must be reconfigured in the Reagent Configuration
menu and a new calibration must be performed with the default controls if configured in the
application.
Related information:
■ Reagent Parameters, p.298
Access: Main menu > Services > Application Configuration > Reagents
4. System Configuration
4.1. Analyser
Access: Main menu > Services > System Configuration > Analyser
Follow this procedure to configure the sample identification mode according to your
laboratory working method.
Access: Main menu > Services > System Configuration > Analyser
This parameter cannot be modified if tests are ordered, pending, incomplete or for validation.
1. Press Edit.
2. Select the sample identification mode from the Loading Mode dropdown list.
■ Identification: Select this mode if all tubes are identified by barcode (default option).
■ Position: Select this mode if tubes without barcode are used on the instrument.
3. Press OK to validate.
Follow this procedure to enter the instrument serial number in the Pentra C400 software.
Access: Main menu > Services > System Configuration > Analyser
1. Press Edit.
2. Type your instrument serial number in the Serial # field.
You will find your instrument serial number on the serial label located at the back of the
instrument.
3. Press OK to validate.
Follow this procedure to enable or to disable the sample and reagent level detections.
Access: Main menu > Services > System Configuration > Analyser
1. Press Edit.
2. To enable or to disable the sample level detection, select or deselect the Sample option in the
Level detection area.
■ If selected, the sample level detection is enabled.
■ If unselected, the sample level detection is disabled.
If the sample level detection is disabled, only sample cups must be used.
3. To enable or to disable the reagent level detection, select or deselect the Reagent option in the
Level detection area.
■ If selected, the reagent level detection is enabled.
■ If unselected, the reagent level detection is disabled.
Working with the reagent level detection disabled can pose important
contamination problems, and erroneous test results may occur.
4. Press OK to validate.
Access: Main menu > Services > System Configuration > Analyser
1. Press Edit.
2. To enable or to disable the cuvette changer, select or deselect the Cuvettes changer option.
■ If selected, the cuvette changer is enabled.
■ If unselected, the cuvette changer is disabled.
If the cuvette changer is disabled, cuvettes must be manually loaded and unloaded
(refer to the Maintenance and Troubleshooting > Troubleshooting > Analytical
Module > Cuvette Changer > To Manually Load Cuvettes chapter).
3. Press OK to validate.
Related information:
■ To Manually Load Cuvettes, p.383
Follow this procedure to enable or to disable the pressure detection of sample and
reagent needles.
Access: Main menu > Services > System Configuration > Analyser
1. Press Edit.
2. To enable or to disable the sample needle pressure detection, select or deselect the Sample
detection option in the Pressure Detections area.
■ If selected, the sample needle pressure detection is enabled.
■ If unselected, the sample needle pressure detection is disabled.
3. To enable or to disable the reagent needle pressure detection, select or deselect the Reagent
detection option in the Pressure Detections area.
■ If selected, the reagent needle pressure detection is enabled.
■ If unselected, the reagent needle pressure detection is disabled.
4. Press OK to validate.
Follow this procedure to enable or disable the sample and reagent barcode readers.
Access: Main menu > Services > System Configuration > Analyser
1. Press Edit.
2. To enable or disable the sample barcode reader, select or deselect the Sample option in the
Barcode Reader area.
■ If selected, the sample barcode reader is enabled.
■ If unselected, the sample barcode reader is disabled.
This parameter cannot be modified if tests are ordered, pending, incomplete or for
validation.
3. To enable or to disable the reagent barcode reader, select or deselect the Reagent option in the
Barcode Reader area.
■ If selected, the reagent barcode reader is enabled.
■ If unselected, the reagent barcode reader is disabled.
Working with the reagent barcode reader disabled, you must ensure that solutions
are physically positioned as programmed in the Reagent Configuration menu.
4. Press OK to validate.
Access: Main menu > Services > System Configuration > Analyser
If an automatic startup is programmed, the instrument automatically starts every day at a specified
time.
1. Press Edit.
2. Select the Automatic startup option to enable automatic startup.
3. Specify the startup time.
4. Press OK to validate.
Access: Main menu > Services > System Configuration > Analyser
If an automatic standby mode is programmed, the instrument automatically enters standby mode after
a specified inactivity time.
1. Press Edit.
2. Select the Standby mode check option to enable automatic standby mode.
3. Specify the inactivity time.
4. Press OK to validate.
Access: Main menu > Services > System Configuration > Analyser
If automatic cleanings are programmed, ISE module and/or needles are automatically cleaned at the
instrument shutdown.
1. Press Edit.
2. Select the ISE cleaning option to program by default an ISE module cleaning at the instrument
shutdown.
3. Select the System cleaning option to program by default a needles cleaning at the instrument
shutdown.
4. Press OK to validate.
■ If you selected the ISE cleaning option, you will have to choose the cleaner from the ISE Cleaning
dropdown list at the instrument shutdown.
■ If you selected the System cleaning option, you will have to choose the cleaner from the System
Cleaning dropdown list at the instrument shutdown.
Refer to the Workflow > End of Day > To Shut Down the Instrument chapter.
Related information:
■ To Shut Down the Instrument, p.214
Access: Main menu > Services > System Configuration > Analyser
1. Press Edit.
2. Select a rack type.
Rack numbers displayed in the list of available racks and the list of racks used on the instrument
are updated.
A rack number cannot be removed from the list of racks used on the instrument if a
solution is still configured on this rack number.
Access: Main menu > Services > System Configuration > Analyser
This parameter cannot be modified if associated tests are ordered, pending, incomplete or for
validation.
1. Press Edit.
2. Select or deselect the Activate ISE Module check box to enable or disable the ISE module.
■ If selected, the ISE module is enabled.
■ If unselected, the ISE module is disabled.
If the ISE module is disabled, the ISE Module Status button of the main menu is
unavailable, the ISE applications are not displayed anymore in the request capture
screen nor in Main menu > Services > Application Configuration > Applications,
all ISE functionalities of Customer Services menu are not displayed anymore.
3. Press OK to validate.
Access: Main menu > Services > Customer Services > ISE
1. Press Edit.
2. To activate or deactivate an electrode, select or deselect the corresponding check box in the
Activated electrode area.
■ If selected, the electrode is activated.
■ If unselected, the electrode is deactivated.
3. Press OK to validate.
A dialog box asks for user confirmation.
4. Press OK to validate or Cancel to cancel.
Follow this procedure to enable or to disable the automatic back-up cassette option.
Access: Main menu > Services > System Configuration > Analyser
1. Press Edit.
2. To enable or to disable the automatic back-up cassette option, select or deselect the Activate
dynamic cassette change option.
■ If selected, the automatic back-up cassette option is enabled.
If several cassettes of the same solution are installed on the reagent tray, the first cassette
installed is active and the next cassettes are inactive. If the automatic back-up cassette option
is enabled, the next cassettes are called back-up cassettes. Back-up cassettes are
automatically activated when the cassette in use is empty.
■ If unselected, the automatic back-up cassette option is disabled.
3. Press OK to validate.
Access: Main menu > Services > System Configuration > Analyser
1. Press Edit.
2. To enable or disable the cooling unit, select or deselect the Activate Cold Area option.
■ If selected, the cooling unit is enabled.
■ If unselected, the cooling unit is disabled.
3. Press OK to validate.
4.1.14. To Program the Time Lapse for Displaying Calibrations that Will
Expire
Follow this procedure to configure the time lapse for displaying calibrations that will
expire.
Access: Main menu > Services > System Configuration > Analyser
When adding a calibration request, you can display calibrations that will expire in a programmed time
lapse (refer to the Workflow > Calibration and Control > Calibration and Control Worklists > To Add
Calibration Requests chapter). The Calibrations expired time windows area allows you to configure
this time lapse (from 0 to 96 hours).
1. Press Edit.
2. Configure the time lapse from the Hour(s) spin box in the Calibrations expired time windows
area.
3. Press OK to validate.
Related information:
■ To Add Calibration Requests, p.118
Follow this procedure to enable or to disable the automatic export of results to a USB key.
Access: Main menu > Services > System Configuration > Analyser
1. Press Edit.
2. To enable or to disable the automatic export of results to a USB key, select or deselect the
Activate export of the results on USB key option.
■ If selected, the automatic export of results is enabled.
If the automatic export of results is enabled, all the completed test results are saved in a .csv
file on a USB key at each end of run. ISE test results and ratio results are not exported.
HORIBA Medical recommends you to make sure that all materials like CD-ROM or
USB key used on the Pentra C400 and non-referenced by HORIBA Medical are free
of viruses. These materials must imperatively be cleaned before being used on the
instrument.
3. Press OK to validate.
Access: Main menu > Services > System Configuration > Local Settings
Follow this procedure to select one of the three available date formats.
Access: Main menu > Services > System Configuration > Local Settings
1. Press Edit.
Follow this procedure to select one of the two available time formats.
Access: Main menu > Services > System Configuration > Local Settings
1. Press Edit.
2. Choose a time format from the Time Format dropdown list.
■ hh:mm:ss ampm
■ HH:mm:ss
hh stands for hours, mm stands for minutes and ss stands for seconds.
3. Press OK to validate.
The instrument is restarted after user confirmation.
Follow this procedure to change the date and time of the instrument.
Access: Main menu > Services > System Configuration > Local Settings
The date and time cannot be modified if tests are ordered, pending, incomplete or for validation.
1. Press Edit.
2. Modify the date as follows:
a. Press the arrow from the Date field to open the calendar.
b. To choose a month, use the left and right arrows. Then choose the day.
c. When done, click randomly outside the calendar to close it.
3. Set the hours, minutes and seconds in the Time field.
4. If the hh:mm:ss ampm time format is selected, either choose am or pm.
5. Press OK to validate.
Follow this procedure to change the language of user interface, printouts and data sent to
host.
Access: Main menu > Services > System Configuration > Local Settings
1. Press Edit.
2. Choose the software language from the Language List dropdown list.
■ English
■ French
■ German
■ Italian
■ Spanish
■ Portuguese
■ Greek
■ Polish
3. Press OK to validate.
The instrument is restarted after user confirmation.
Access: Main menu > Services > System Configuration > Local Settings
1. Press Edit.
2. Press Change Configuration.
A dialog box is displayed.
3. Choose the keyboard language in the dialog box.
4. Press OK in the dialog box.
5. Press OK to validate.
The instrument is restarted after user confirmation.
Follow this procedure to configure the barcode types used in your laboratory.
Access: Main menu > Services > System Configuration > Local Settings
1. Press Edit.
2. Select the barcode types used in your laboratory from the following options.
■ ITF 2/5 (2 of 5 interleaved)
■ Code 39
■ Code 128
■ Codabar
3. If the ITF 2/5 barcode type is selected, configure the ITF 2/5 Check Digit option.
■ If unselected, the ITF 2/5 barcode type is used without check digit.
You should preferably use the sample rack labels from HAX0166 (1207230166) to
HAX0174 (1207230174). Nevertheless, if you use the 2 of 5 interleaved barcode
type without check digit, you must use the sample rack labels from HAX0273
(1207230273) to HAX0281 (1207230281).
■ If selected, the ITF 2/5 barcode type is used with check digit.
4. If the Code 39 barcode type is selected, configure the Code 39 Check Digit option.
■ If unselected, the Code 39 barcode type is used without check digit.
■ If selected, the Code 39 barcode type is used with check digit.
5. Press OK to validate.
Follow this procedure to update the Pentra C400 online help (user manual).
Access: Main menu > Services > System Configuration > Local Settings
The Pentra C400 online help is maintained and updated by HORIBA Medical with the Documentation
media (USB flash drive).
The contents of this media can also be viewed on any computer having the Adobe™
Acrobat™ Reader software (version 6.0 or later).
2. Press Edit.
3. Press Update help.
The Documentation media application is launched.
4. Choose your language.
5. Press Update Help.
Access: Main menu > Services > System Configuration > Local Settings
The Serial # field in Main menu > Services > System Configuration > Analyser must be completed.
Refer to the System Configuration > Analyser > To Enter the Instrument Serial Number chapter.
The client parameters backup includes the current database, the columns configuration, the
parameters of System Configuration menu and Customer Services menu. The parameters
configured by your local representative and the default printer are not saved.
1. Press Edit.
2. Press Save in the Save client parameters area.
A dialog box allows you to select the support type:
■ hard disk (internal),
■ USB key.
3. If you use a USB key, install it on your instrument.
HORIBA Medical recommends you to make sure that all materials like CD-ROM or USB
key used on the Pentra C400 and non-referenced by HORIBA Medical are free of
viruses. These materials must imperatively be cleaned before being used on the
instrument.
The client parameters are saved in a file named with the backup date and time and the instrument
serial number.
Related information:
■ To Enter the Instrument Serial Number, p.302
Follow this procedure to restore your instrument configuration previously saved on the
hard disk or a USB key.
Access: Main menu > Services > System Configuration > Local Settings
1. Press Edit.
2. Press Restore in the Restore client parameters area.
A dialog box allows you to select the support type:
■ hard disk (internal),
■ USB key.
3. If you use a USB key, install it on your instrument.
HORIBA Medical recommends you to make sure that all materials like CD-ROM or USB
key used on the Pentra C400 and non-referenced by HORIBA Medical are free of
viruses. These materials must imperatively be cleaned before being used on the
instrument.
This parameter cannot be restored if tests are ordered, pending, incomplete or for
validation.
■ Database: If selected, the database is updated (if necessary) then restored, and the warning
#628 Restore database saved on [date and time] by [user name] is triggered.
7. Press OK to validate.
■ If necessary, the database is updated after user confirmation.
■ The instrument is restarted after user confirmation.
The selected client parameters are restored and the corresponding warnings are triggered.
Access: Main menu > Services > System Configuration > Local Settings
Each time a new worklist is created, the current database is saved on the instrument. Up to eight
databases are saved and numbered from the most recent to the oldest.
1. Press Edit.
2. Press Restore database.
The list of databases saved on the instrument is displayed.
3. Select the database that you want to restore.
4. Press OK to validate.
The instrument is restarted after user confirmation.
The selected database is restored and the warning #628 Restore database saved on [date and time]
by [user name] is triggered.
Access: Main menu > Services > System Configuration > Host Connection
Access: Main menu > Services > System Configuration > Host Connection
Host connection must be configured by an authorized technician using the Output format
for host connection document. This document is available online at www.horiba.com.
1. Press Edit.
2. Configure the connection parameters as follows:
a. Select the speed in Baud from the following options: 1200, 2400, 4800, 9600, 19200, 38400,
57600, 115200.
Set to 9600 by default.
b. Select the parity from the following options: None, Odd, Even.
Set to None by default.
c. Select the stop bit from the following options: 1 Stop, 2 Stop.
Set to 1 Stop by default.
d. Select the protocol from the following options: Xon/Xoff, None.
Set to Xon/Xoff by default.
3. Configure the Query Mode option.
■ If selected, the Query Mode option is enabled. If a sample tube is read on the tray and no
associated request is found in the worklist, the instrument automatically asks the host if there
is a programmed request for this tube.
■ If unselected, the Query Mode option is disabled.
This option is selectable only if the sample identification mode is set to Identification
in the Main menu > Services > System Configuration > Analyser.
4. If the Query Mode option is enabled, configure the Sending validated samples option.
■ If selected, the Sending validated samples option is enabled. The Host request button is
available in the Work Balance menu or in the result list. This button allows you to manually ask
the host if there are additional tests programmed for a validated patient sample.
■ If unselected, the Sending validated samples option is disabled.
5. Configure the Handling spaces on the left side of SID option.
■ If selected, the space character is allowed on the left side of the sample ID. In this case, the
space character is indicated by a rectangle on printouts.
■ If unselected, the space character is not allowed on the left side of the sample ID. In this case,
the space character is systematically deleted.
6. Configure the automatic transmission of patient results to the host from the following options.
■ Nothing: If selected, patient results are not automatically sent.
■ Order: If selected, patient results are automatically sent by request after validation of all results
of the request.
■ Test: If selected, patient results are automatically sent by test after result validation.
7. Configure the automatic transmission of control results to the host from the following options.
■ Nothing: If selected, control results are not automatically sent.
■ Test: If selected, control results are automatically sent by test after result validation.
8. Enter the instrument identification number for the host in the Analyser Id field.
9. Press OK to validate.
Follow this procedure to check the host connection and to reset it if necessary.
Access: Main menu > Services > System Configuration > Host Connection
1. Press Edit.
2. Press the Reset connection button:
Follow this procedure to delete all pending data (receipt and transmission).
Access: Main menu > Services > System Configuration > Host Connection
1. Press Edit.
2. Press the Purge host orders button:
Access: Main menu > Services > System Configuration > Host Connection
1. Press Edit.
2. Press the Test communication button:
The instrument sends a message to the host and waits for the host to confirm the message
receipt. The Received Character field displays the test result:
Message Description
ACK The host confirms the message receipt.
XXX The host does not confirm the message receipt.
Time out The host did not confirm within the expected time.
Process NOK The host connection does not work.
Follow this procedure to view the communication frames and to search for a message
within these frames.
Access: Main menu > Services > System Configuration > Host Connection
1. Press Edit.
2. Press the Connection Spy button:
The communication frames (data receipt and transmission) are displayed from the oldest to the
most recent. The screen is continuously refreshed.
5. Select the Case Sensitive option to distinguish between uppercase and lowercase letters.
6. Select the Whole Words option to search for the complete message.
7. Choose the search direction from the following options.
■ Search Forward: To search from the oldest to the most recent.
■ Search Backward: To search from the most recent to the oldest.
8. Press Next to launch the search.
■ The message is highlighted if found.
■ A dialog box informs you if the message was not found.
4.4. Printer
Access: Main menu > Services > System Configuration > Printer
Access: Main menu > Services > System Configuration > Printer
1. Press Edit.
2. Fill in the Header fields to customize the header of the condensed, standard and detailed printout
formats.
Six fields of 20 characters maximum each are available.
3. Fill in the Portrait header fields to customize the header of the portrait printout format.
Eight fields of 40 characters maximum each are available.
4. Add a logo to customize the header of the portrait printout format as follows:
a. Save your logo on a CD-ROM or a USB key.
Logo characteristics:
■ Size: 55 x 45 mm (Width x Height)
■ Format: .jpeg file
HORIBA Medical recommends you to make sure that all materials like CD-ROM or
USB key used on the Pentra C400 and non-referenced by HORIBA Medical are free
of viruses. These materials must imperatively be cleaned before being used on the
instrument.
5. Press OK to validate.
Access: Main menu > Services > System Configuration > Printer
1. Press Edit.
2. Configure automatic printouts of patient results as follows:
a. Select the Complete option to automatically print patient results when validated.
b. Select the order in which patient results are printed from the following options.
■ Sort by Order: If selected, results are sorted by patient.
■ Sort by sample result: If selected, results are sorted by run date.
Detailed and portrait printout formats are unavailable if the Sort by Order option is
selected.
Access: Main menu > Services > System Configuration > Printer
1. Press Edit.
2. Select the printer of which you want to configure the properties from the Printer List dropdown
list.
3. Press Printer Properties.
The selected printer properties window is displayed.
4. Change the selected printer properties where necessary and press OK.
5. Press OK to validate.
Access: Main menu > Services > System Configuration > Printer
1. Press Edit.
2. Select the printer that you want to configure by default from the Printer List dropdown list.
3. Press Set Default Printer.
The selected printer becomes the default printer.
4. Press OK to validate.
Access: Main menu > Services > System Configuration > Printer
1. Press Edit.
2. Select the printer that you want to remove from the Printer List dropdown list.
3. Press Delete Printer.
The selected printer is removed from the list.
4. Press OK to validate.
Access: Main menu > Services > System Configuration > Results validation
Access: Main menu > Services > System Configuration > Results validation
1. Press Edit.
2. Select the Automatic validation option in the Calibrator area to enable calibrations automatic
validation.
If selected, calibration results are automatically validated on condition that:
■ No point used for calculation is flagged (except with the analytical flag CALC_RANGE_HIGH /
LOW).
■ For each calibrator level, deviations (Dev_Rep and Dev_C) do not exceed the authorized
values.
■ The calibration factor is not out of the programmed values.
3. Select the Quality flag option in the Calibrator area to block the automatic validation when the
calibration is flagged with a quality flag.
4. Press OK to validate.
Follow this procedure to configure automatic validation of control results and to configure
Westgard rules.
Access: Main menu > Services > System Configuration > Results validation
1. Press Edit.
2. Select the Automatic validation option in the Control area to enable controls automatic
validation.
If selected, control results are automatically validated on condition that they are not flagged
(except with the analytical flag CONF_RANGE_HIGH_W / LOW_W).
3. Select the Quality flag option in the Control area to block the automatic validation when the
control is flagged with a quality flag.
4. Select the Westgard rule validation option to enable Westgard rules.
This parameter cannot be modified if tests are ordered, pending, incomplete or for validation.
For detailed information concerning Westgard rules, refer to the Quality Assurance >
Westgard Rules chapter.
If Westgard rules are enabled, the Apply to values outside confidence range option is selected
by default and cannot be unselected. In this case, Westgard rules apply when the control results
are outside the confidence range.
5. If Westgard rules are enabled, you can select the Apply rules 4, 5 and 6 (drift) to values inside
confidence range option to apply the Westgard rules 4, 5 and 6 to control results inside the
confidence range (suspicion of drift).
This parameter cannot be modified if tests are ordered, pending, incomplete or for validation.
6. Press OK to validate.
Related information:
■ Westgard Rules, p.82
Access: Main menu > Services > System Configuration > Results validation
1. Press Edit.
2. Choose the validation mode of patient results from the following options.
■ Automatic validation: If selected, all patient results are automatically validated.
■ Review of all results: If selected, all patient results must be manually validated.
■ Review of exception: If selected, patient results are automatically validated on defined
criteria.
3. If the Review of exception option is selected, choose the criteria on which patient results are
automatically validated.
■ Analytical flag: Selected by default and not modifiable, patient results are automatically
validated on condition that they are not flagged (except with the analytical flag
REF_RANGE_HIGH / LOW).
■ Reference range flag: If selected, patient results flagged with REF_RANGE_HIGH / LOW must
be manually validated.
■ Delta check flag: If selected, patient results flagged with the delta check flag must be
manually validated.
■ Quality flag: If selected, patient results flagged with a quality flag must be manually validated.
4. Press OK to validate.
Access: Main menu > Services > System Configuration > Audible alarm
1. Press Edit.
2. Choose the sound alarms language from the dropdown list at the top of the screen.
■ English
■ French
■ German
■ Italian
■ Spanish
■ Portuguese
■ Greek
■ Polish
3. Select the events for which you want to hear a sound alarm from the following options.
■ System alarm: If selected, an alarm sounds when a system alarm is triggered.
■ Warning: If selected, an alarm sounds when a system warning is triggered.
■ Pre-analytical alarm: If selected, an alarm sounds when a sampling alarm is triggered.
■ Manual validation: If selected, an alarm sounds when a manual validation is required.
■ Maintenance warning: If selected, an alarm sounds when a maintenance alert is triggered.
■ Reagent processing: If selected, an alarm sounds when the instrument turns to "Reagent
processing" status.
■ Ready: If selected, an alarm sounds when the instrument turns to "Ready" status.
■ Sample tray reading: If selected, an alarm sounds when the sample tray is not accessible
anymore.
For each event, press Test to hear the corresponding alarm sound.
4. For each selected event, press the up or down arrows from the Repeat spin box to configure the
number of times the alarm should sound (from one to five times).
5. Set the sound volume using the slider.
6. Press OK to validate.
When several sound alarms are simultaneously triggered, only the alarm having the
highest priority sounds.
Sound alarms priority (from highest to lowest): System alarm > Warning > Pre-analytical
alarm > Manual validation > Maintenance warning > Reagent processing > Ready >
Sample tray reading.
Follow this procedure to configure the maintenance procedures that must be performed
on the instrument and when they must be performed.
Access: Main menu > Services > System Configuration > Maintenance
Once the maintenance procedures are configured, the instrument checks that they are performed
when necessary. If a maintenance procedure is not performed on time, a maintenance alert is
triggered.
Moreover, the configured maintenance procedures are listed in the Maintenance log when they are
performed.
Most of the maintenance procedures use a software function from the Customer Services menu.
These procedures are automatically notified in the log when they are performed.
Some maintenance procedures do not use a software function. These procedures must be manually
notified in the log when they are performed.
1. Press Edit.
2. Select the maintenance procedures that must be performed on your instrument from the Active
area.
3. For each selected procedure, configure when it must be performed from the Occurrence area.
The following frequencies are selectable:
■ On request,
■ Daily,
■ Weekly,
■ Monthly,
■ Every 2 months,
■ Every 6 months,
■ Annually.
4. Press OK to validate.
Refer to the Configuring Maintenance Alerts > Maintenance Alerts Default Configuration
chapter to know the maintenance alerts default configuration.
Related information:
■ Maintenance, p.90
■ Maintenance Alerts Default Configuration, p.330
The table below gives the maintenance alerts default configuration and indicates, for each procedure,
if it is automatically notified in the Maintenance log when performed or if it must be manually notified.
For detailed information concerning the procedure to configure maintenance alerts, refer
to the Configuring Maintenance Alerts > To Configure Maintenance Alerts chapter.
Related information:
■ Maintenance, p.90
■ To Configure Maintenance Alerts, p.329
4.8. Users
Access: Main menu > Services > System Configuration > Users
Each user account has a personal and unique user name or login and an associated password.
Four user levels are available in the Pentra C400 software.
Access: Main menu > Services > System Configuration > Users
Only administrators (Admin) can create, modify and delete user accounts.
Access: Main menu > Services > System Configuration > Users
Only administrators (Admin) can create, modify and delete user accounts.
Access: Main menu > Services > System Configuration > Users
Only administrators (Admin) can create, modify and delete user accounts.
2. Maintenance......................................................................................................................... 339
2.1. Daily Procedures........................................................................................................................ 339
2.2. Weekly Procedures.................................................................................................................... 341
2.3. Monthly Procedures................................................................................................................... 344
2.4. Bimonthly Procedures (Every Two Months)............................................................................... 352
2.5. Other Procedures.......................................................................................................................355
2.6. Consumables and Spare Parts.................................................................................................. 369
2.7. Maintenance Schedule...............................................................................................................371
3. Troubleshooting.................................................................................................................. 372
3.1. Analyzer Power Problems.......................................................................................................... 372
3.2. Printer Operation Problems....................................................................................................... 373
3.3. Reagent and Sample Trays........................................................................................................373
3.4. Sampling System....................................................................................................................... 378
3.5. Analytical Module.......................................................................................................................380
3.6. ISE Module (Option)................................................................................................................... 386
3.7. Water and Waste Tanks............................................................................................................. 389
3.8. Cooling Unit............................................................................................................................... 390
3.9. Result Problems.........................................................................................................................392
The table below gives a summary of what can be achieved from the Customer Services menu.
Cycles tab
ISE tab
Analyser tab
Barcode tab
The Barcode tab allows you to make sure that the sample barcode reader correctly reads a barcode
label. Refer to the Troubleshooting > Reagent and Sample Trays > Sample Tray > To Check Correct
Reading of a Barcode Label chapter.
Related information:
■ Maintenance, p.339
■ Troubleshooting, p.372
■ Test Counter (On Request), p.92
■ To Activate or Deactivate an Electrode, p.308
■ To Perform ISE Module Activation (Option), p.110
2. Maintenance
The maintenance procedures identified in this chapter are mandatory for proper use and operation of
the Pentra C400.
Failure to execute any of these recommended procedures may result in poor reliability of
the system.
HORIBA Medical recommends you to perform maintenance procedures at the end of the
day.
Clean the ISE module with the ABX Pentra Etching CP A11A01769 on a daily basis if
more than 20 samples are run daily.
Access: Main menu > Services > Customer Services > ISE
■ The ISE module should be cleaned with ABX Pentra Etching CP A11A01769 every
day only if more than 20 samples are run daily.
■ If less than 20 samples are run daily, perform a complete ISE module cleaning every
week. Refer to the Maintenance and Troubleshooting > Maintenance > Weekly
Procedures > To Clean ISE Module (Option) chapter.
1. Choose ETCH (ABX Pentra Etching CP A11A01769) from the ISE Cleaning Solution dropdown
list.
2. Press ISE Start to clean the ISE module.
The ISE module can be automatically cleaned during instrument shutdown if programmed
in Main menu > Services > System Configuration > Analyser.
Related information:
■ To Clean the ISE Module (Option), p.341
■ To Configure Automatic Cleanings, p.306
■ To Perform ISE Module Activation (Option), p.110
Clean the needles with ABX Pentra Deproteinizer CP A11A01754 on a daily basis.
Access: Main menu > Services > Customer Services > Cycles
1. Choose the ABX Pentra Deproteinizer CP A11A01754 from the Needle Cleaning Solution
dropdown list.
2. If the Mixer cleaning option is selected, clear it.
3. Press Needle Cleaning to clean the needles.
The needles can be automatically cleaned during the instrument shutdown if programmed
in Main menu > Services > System Configuration > Analyser.
Related information:
■ To Configure Automatic Cleanings, p.306
Access: Main menu > Services > Customer Services > ISE
■ Consumables and spare parts: ABX Pentra Deproteinizer CP A11A01754, ABX Pentra Etching
CP A11A01769, 200 µL of serum or plasma into a sample cup, ABX Pentra Dummy Electrode
A11A01851, Cleaning kit for Chloride electrode XEA935AS (1209159935)
■ The ABX Pentra Deproteinizer CP A11A01754 must be previously configured in the Reagent
Configuration menu and placed on the reagent tray.
■ The ABX Pentra Etching CP A11A01769 must be previously configured in the Reagent
Configuration menu and placed on the reagent tray at room temperature.
■ The ISE module should not be busy.
■ The ISE module status must be ≥ 4 (module sleeping).
The steps to be carried out during the maintenance are listed. The status of each step is indicated
by a color code:
■ Gray: The step must be performed.
■ Orange: The step is in progress.
■ Green: The step has successfully been completed.
■ Red: An error appeared during the step.
2. Select an available rack position on the sample tray from the Rack and Pos. dropdown lists.
3. Put 200 µL of serum or plasma into a sample cup and place it in the selected rack position.
4. Press Next to launch step 1.
The instrument checks the following items:
■ the required solutions are present on the reagent and sample trays,
■ the required cleaners are active and there remains a sufficient volume.
Then, the ISE module drains the reagents from the electrodes.
If an error appears during the step, check and correct the error before launching again either the
step or the whole maintenance.
5. Wait for the end of step 1 and replace the chloride electrode with the ABX Pentra Dummy
Electrode A11A01851 as follows:
a. Open the ISE module cover.
b. Open the Faraday cage by loosening the fixing screw.
c. Replace the chloride electrode with the ABX Pentra Dummy Electrode A11A01851.
d. Close the Faraday cage by tightening the fixing screw.
e. Close the ISE module cover.
6. Press Next to launch step 2.
The ISE module is cleaned with the ABX Pentra Deproteinizer CP A11A01754. It takes a few
minutes.
If an error appears during the step, check and correct the error before launching again either the
step or the whole maintenance.
7. Manually clean the chloride electrode as follows:
a. Use one thread provided in the Cleaning kit for Chloride electrode XEA935AS (1209159935).
b. Pull one thread only once slowly and carefully through the sample channel of the electrode in
the direction indicated below.
Do not wash and re-use the threads. They are designed for single use only.
c. Make sure that the O-ring is well-positioned on one side of the electrode.
8. Wait for the end of step 2 and replace the ABX Pentra Dummy Electrode A11A01851 with the
chloride electrode.
9. Press Next to launch step 3.
The following substeps are carried out:
■ the ISE module is cleaned with the ABX Pentra Etching CP A11A01769,
■ the ISE module activation is performed,
■ the ISE module rinses the fluidic system,
■ the 2-point calibration is performed.
It takes a few minutes.
If an error appears during the step, check and correct the error before launching again either the
step or the whole maintenance.
10. Wait for the end of step 3 and press the Quit button:
For detailed information concerning ISE calibration results interpretation, refer to the
ISE Calibration (Option) > ISE Calibration Results chapter.
Related information:
■ ISE Calibration Results, p.116
Clean needles and mixer with the ABX Pentra Deproteinizer CP A11A01754 on a weekly
basis.
Access: Main menu > Services > Customer Services > Cycles
1. Choose the ABX Pentra Deproteinizer CP A11A01754 from the Needle Cleaning Solution
dropdown list.
2. Select the Mixer cleaning option.
3. Press Needle Cleaning to clean needles and mixer.
Access: Main menu > Services > Customer Services > Cycles
Clean wash towers with ABX Pentra Deproteinizer CP A11A01754 on a monthly basis.
Access: Main menu > Services > Customer Services > Cycles
Consumables and spare parts: A cotton swab, ABX Pentra Deproteinizer CP A11A01754
To ensure good pipetting precision, replace the syringe plunger tips on a monthly basis.
Access: Main menu > Services > Customer Services > Cycles
Consumables and spare parts: 1000 µL + 500 µL O'rings B8086583 (1228086583), 1000 µL Teflon
Seal B8086605 (1228086605), 100 µL Teflon Seal B8086648 (1228086648), Seal Replacement Tool
B8088454 (1228088454)
1 = Reagent syringe
2 = Sample syringe
4. Replace the reagent syringe plunger tip as follows:
a. Remove the plunger from the glass barrel.
b. Pull off the tip as well as the O-ring.
3 = O-ring
4 = Tip
5 = Tip
6 = Replacement tool
d. Carefully put the plunger back into the glass barrel.
6. Manually fill and empty each syringe with distilled water several times.
Ensure that each syringe is filled with water and that no air bubbles are visible in the syringe.
7. Reassemble the syringes in reverse order.
8. Close the main cover and press OK.
9. Press Priming Cycle and make sure that the reagent and sample needles correctly dispense
distilled water for a few seconds in their respective wash towers.
10. Calibrate the pressure detection of reagent and sample needles as follows:
Related information:
■ To Configure the Pressure Detection, p.304
Check the pipetting accuracy and precision once a month. Depending on the control
solution you have, follow either the To Perform the T1 Test procedure (ABX Pentra
Qualitest Solution A11A01758) or the To Perform the P1 Test procedure (ABX Pentra
Precitest Solution 1300017438).
Follow this procedure to check pipetting accuracy and precision on a monthly basis.
The control name, the rack position of control, the control lot number and expiration date are
displayed.
d. Select the T1 test and configure 15 replicates by pressing the up or down arrows (right
side of the screen) to increase or to decrease the number of replicates.
e. Press OK to validate.
6. Press Start.
The instrument turns to the "Sampling" status, then to the "Analysing" status.
7. Wait for the end of analyses.
The instrument turns to the "Ready" status.
8. If the control results are not automatically validated, refer to the Workflow > Calibration and
Control > Calibration and Control Results Validation > Control Results > To Validate a Control
chapter.
9. Check the control results as follows:
a. Enter Main menu > Quality Control > Test.
b. Select the test T1 from the Select Test dropdown list.
The Session Mean table displays the statistical analyses over the current worklist for the T1
test.
c. Make sure that the session coefficient of variation (CV) is ≤ 1.0%.
If not, refer to the Troubleshooting > Sampling System > Reproducibility Problems chapter.
Related information:
■ To Validate a Calibration, p.142
■ To Validate a Control, p.152
■ Reproducibility Problems, p.379
Follow this procedure to check pipetting accuracy and precision on a monthly basis.
The control name, the rack position of control, the control lot number and expiration date are
displayed.
d. Select the P1 test and configure 15 replicates by pressing the up or down arrows (right side of
the screen) to increase or to decrease the number of replicates.
e. Press OK to validate.
6. Press Start.
The instrument turns to the "Sampling" status, then to the "Analysing" status.
7. Wait for the end of analyses.
The instrument turns to the "Ready" status.
8. If the control results are not automatically validated, refer to the Workflow > Calibration and
Control > Calibration and Control Results Validation > Control Results > To Validate a Control
chapter.
9. Check the control results as follows:
a. Enter Main menu > Quality Control > Test.
b. Select the P1 test from the Select Test dropdown list.
The Session Mean table displays the statistical analyses over the current worklist for the P1
test.
c. Make sure that the session coefficient of variation (CV) is ≤ 1.0%.
If not, refer to the Troubleshooting > Sampling System > Reproducibility Problems chapter.
Related information:
■ To Validate a Calibration, p.142
■ To Validate a Control, p.152
■ Reproducibility Problems, p.379
To maintain a good cooling capacity, clean the cooling unit condenser on a monthly basis.
Access: Main menu > Services > System Configuration > Analyser
When you close the ventilation grid, take care to arrange the drainage tube up as
indicated below.
6. Connect the power supply cable and switch the cooling unit on.
Wait ten minutes minimum, between switching off and switching on the cooling unit.
Check the cooling unit glycol level on a bimonthly basis (every two months).
Access: Main menu > Services > System Configuration > Analyser
1. Make sure that the glycol level indicator is between the high and low marks.
2. If the glycol level indicator is below the low mark, follow steps 3 to 9 to add distilled water.
3. Disable the cooling unit as follows:
a. Press Edit.
b. Deselect the Activate Cold Area check box.
c. Press OK to validate.
4. Switch the cooling unit off and disconnect the power supply cable.
5. Slightly push to unlock and to open the lid as shown below.
6. Add distilled water until the glycol level indicator is between the high and low marks.
Wait ten minutes minimum, between switching off and switching on the cooling unit.
Replace the filter at the back of the instrument on a bimonthly basis (every two months).
Access: Main menu > Services > Customer Services > Cycles
■ Consumables and spare parts: Liq. Elem + Gasket Ring Filter B8089078 (1228089078)
■ The instrument must be in the "Ready" status.
2. Tighten both parts of the filter assembly and unscrew to open it.
Follow this procedure to replace either the reagent syringe or the sample syringe or both.
Access: Main menu > Services > Customer Services > Cycles
Consumables and spare parts: Reagent Syringe (1000 µL) B8078955 (1228078955), Sample Syringe
(100 µL) B8076812 (1228076812)
1 = Reagent syringe
2 = Sample syringe
4. Manually fill and empty each new syringe with distilled water several times.
Ensure that each syringe is filled with water and that no air bubbles are visible in the syringe.
5. Reassemble the new syringes in reverse order.
6. Close the main cover and press OK.
7. Press Priming Cycle and make sure that the reagent and sample needles correctly dispense
distilled water for a few seconds in their respective wash towers.
8. Calibrate the pressure detection of the reagent and sample needles as follows:
a. Press Reagent pressure calibration.
This function is unavailable if the pressure detection is disabled. Refer to the Settings > System
Configuration > Analyser > To Configure Pressure Detection chapter.
The reagent needle pressure detection is calibrated. The slope (A) and the intercept (B) of the
regression line y = A * x + B are displayed.
If the calibration fails, the slope (A) and the intercept (B) are displayed in red and the warning
#627 Pressure Reagent calibration failed is triggered.
b. Press Sample pressure calibration.
This function is unavailable if the pressure detection is disabled. Refer to the Settings > System
Configuration > Analyser > To Configure Pressure Detection chapter.
The sample needle pressure detection is calibrated. The slope (A) and the intercept (B) of the
regression line y = A * x + B are displayed.
If the calibration fails, the slope (A) and the intercept (B) are displayed in red and the warning
#626 Pressure Sample calibration failed is triggered.
Related information:
■ To Configure the Pressure Detection, p.304
Access: Main menu > Services > Customer Services > Cycles
5. Unscrew the two plastic collars maintaining the reagent needle teflon tubing.
6. Disconnect the level detection cable and loosen the two screws to free the needle.
13. Press Priming Cycle and make sure that the reagent and sample needles correctly dispense
distilled water for a few seconds in their respective wash towers.
14. Press Reagent pressure calibration.
This function is unavailable if the pressure detection is disabled. Refer to the Settings > System
Configuration > Analyser > To Configure Pressure Detection chapter.
The reagent needle pressure detection is calibrated. The slope (A) and the intercept (B) of the
regression line y = A * x + B are displayed.
If the calibration fails, the slope (A) and the intercept (B) are displayed in red and the warning #627
Pressure Reagent calibration failed is triggered.
Related information:
■ To Configure the Pressure Detection, p.304
Access: Main menu > Services > Customer Services > Cycles
5. Unscrew the two plastic collars maintaining the sample needle teflon tubing.
6. Disconnect the level detection cable and loosen the screw to free the needle.
■ Before the sample arm cover reassembly: position, as much as possible, the
sample needle locking screw towards the sample arm inside.
■ Take care that the tubing is correctly positioned in the sample arm cover and be
careful not to clamp the tubing nor the wires of the sample arm when you tighten
the screw on the sample arm cover.
If the calibration fails, the slope (A) and the intercept (B) are displayed in red and the warning #626
Pressure Sample calibration failed is triggered.
Related information:
■ To Configure the Pressure Detection, p.304
Access: Main menu > Services > Customer Services > Cycles
Make sure that the mixer paddle is pushed up to the maximum in its support.
Access: Main menu > Services > Customer Services > Cycles
Consumables and spare parts: Lamp bulb GBM1326 (1203601326), Light Bulb Dismantling Allen
Key MAB086A (1207821086)
4. Disconnect the lamp supply cable by pressing the clip and pulling it. Loosen the light lock screws
with the Light Bulb Dismantling Allen Key MAB086A (1207821086) and turn the light lock to
remove the lamp.
1 = Press
2 = Pull
3 = Light lock screws
5. Replace the lamp and reassemble in reverse order.
■ Do not touch the lamp with fingers. If necessary use a soft cloth.
■ When reassembling, ensure that the lamp is correctly positioned as indicated
above.
Access: Main menu > Services > Customer Services > ISE
■ Consumables and spare parts: ABX Pentra Standard 1 A11A01717, ABX Pentra Standard 2
A11A01718, ABX Pentra Reference 280 mL A11A01901
■ The ISE module should not be busy.
Access: Main menu > Services > Customer Services > ISE
■ Consumables and spare parts: ABX Pentra Sodium-E A11A01738, ABX Pentra Chloride-E
A11A01739, ABX Pentra Potassium-E A11A01740 or ABX Pentra Reference-E A11A01741
■ The ISE module should not be busy.
■ The ISE module status must be ≥ 3 (mechanical part initialised).
■ Please note that the electrodes can be installed up to the date mentioned on the
packaging label.
■ After opening, the electrodes are stable for:
■ 12 months for the reference electrode,
■ 12 months for the sodium electrode,
■ 4 months for the chloride electrode,
■ 6 months for the potassium electrode.
6. For the reference electrode, remove the protection from one side of the electrode and disconnect
the tubing between the two connectors.
7. Make sure that the O-ring is well-positioned on one side of the electrode and make sure that the
electrode is clean.
8. Install the new electrode.
■ When facing the ISE module, the electrodes have to be positioned as follows (from
left to right): reference electrode > sodium electrode > chloride electrode >
potassium electrode > air sensor.
■ For the reference electrode, reconnect the two tubings as follows:
■ The tubing labeled U must be connected to the upper position of the reference
electrode.
■ The tubing labeled D must be connected to the lower position of the reference
electrode.
Follow this procedure to shut down the ISE module within three days.
Access: Main menu > Services > Customer Services > ISE
Related information:
■ End of Day, p.212
Follow this procedure to shut down the ISE module for over three days.
Access: Main menu > Services > Customer Services > ISE
After opening, ISE reagents are stable for one month if stored at 5 - 30°C and
protected from light.
5. Press Drain.
The ISE module drains the reagents from the fluidic system.
6. Plunge the straws into a distilled water bottle and press Fluidic Initialization.
The ISE module primes the distilled water in the fluidic system.
7. Remove the straws from the distilled water bottle and wrap them into absorbent paper.
8. Press Drain.
The ISE module drains the distilled water from the fluidic system.
9. You can now switch the Pentra C400 off.
Refer to the Workflow > End of Day chapter.
10. Open the Faraday cage by loosening the fixing screw.
11. Remove the four electrodes as well as the air sensor. For the reference electrode, disconnect the
two tubings.
■ For the reference electrode, connect a tube filled with reference solution between the two
connectors and seal the inner tube with a cellophane tape.
■ For the sodium, chloride and potassium electrodes as well as for the air sensor, ensure there is
no liquid in the inner tube.
Store the electrodes in a dry and clean area in their original packaging.
Related information:
■ End of Day, p.212
■ To Put the Electrodes Back after an ISE Module Shutdown, p.366
Follow this procedure to start up the ISE module after a shutdown exceeding three days.
■ When facing the ISE module, the electrodes have to be positioned as follows (from
left to right): reference electrode > sodium electrode > chloride electrode >
potassium electrode > air sensor.
■ For the reference electrode, reconnect the two tubings as follows:
■ The tubing labeled U must be connected to the upper position of the reference
electrode.
■ The tubing labeled D must be connected to the lower position of the reference
electrode.
Related information:
■ To Shut Down the ISE Module for Over 3 Days, p.365
■ To Perform ISE Module Activation (Option), p.110
See also the WHO (World Health Organization) guidelines: "Laboratory Biosafety
Manual, 3rd edition" for more information.
Related information:
■ To Clean the Reagent Needle (External Part), p.344
■ To Clean Needles and Mixer with Deproteinizer, p.343
■ To Clean the ISE Module (Option), p.341
The table below provides you with a complete list of consumables and spare parts that may be
needed on the instrument.
Consumables
Designation Reference
4 mL Nalgene Vials A11A01572
ABX Pentra Deproteinizer CP A11A01754
ABX Pentra Clean-Chem CP A11A01755
ABX Pentra Qualitest Solution A11A01758
ABX Pentra Precitest Solution 1300017438
Sample Cup - Blue A11A01765
Sample Cup - Green A11A01766
Sample Cup - White A11A01767
Sample Cup - Yellow A11A01768
ABX Pentra Clean-Chem 99 CP A11A01789
30/10 cassettes kit (x 6) A11A01922
10 mL Reag Cup B1034626
4 mL Reag Cup B1034634
15 mL Reag Cup B1037307
Cuvette Segments A11A01891
Sample Syringe (100 µL) B8076812 (1228076812)
Reagent Syringe (1000 µL) B8078955 (1228078955)
1000 µL + 500 µL O'rings B8086583 (1228086583)
1000 µL Teflon Seal B8086605 (1228086605)
100 µL Teflon Seal B8086648 (1228086648)
Seal Replacement Tool B8088454 (1228088454)
Liq. Elem + Gasket Ring Filter B8089078 (1228089078)
Mixer Paddle GBM0134 (1203600134)
Lamp bulb GBM1326 (1203601326)
Sample Rack Stickers - # 01 to 10 HAX0166 (1207230166)
Sample Rack Stickers - # 11 to 20 HAX0167 (1207230167)
Sample Rack Stickers - # 21 to 30 HAX0168 (1207230168)
Sample Rack Stickers - # 31 to 40 HAX0169 (1207230169)
Sample Cup Rack Stickers - # 41 to 50 HAX0170 (1207230170)
Sample Cup Rack Stickers - # 51 to 60 HAX0171 (1207230171)
Sample Cup Rack Stickers - # 61 to 70 HAX0172 (1207230172)
Calibrator Sample Rack Stickers - # 71 to 84 HAX0173 (1207230173)
Control Sample Rack Stickers - # 85 to 99 HAX0174 (1207230174)
Reagent Rack Stickers - # 60 to 69 HAX0175 (1207230175)
Reagent Rack Stickers - # 70 to 79 HAX0176 (1207230176)
Calibrator Reagent Rack Stickers - # 01 to 15 HAX0188 (1207230188)
Control Reagent Rack Stickers - # 30 to 44 HAX0190 (1207230190)
Control Reagent Rack Stickers - # 45 to 59 HAX0191 (1207230191)
Sticker, Reagent Rack P400 HAX0230 (1207230230)
Designation Reference
Sample Rack Stickers - # 01 to 10 (2 of 5 Int.) HAX0273 (1207230273)
Sample Rack Stickers - # 11 to 20 (2 of 5 Int.) HAX0274 (1207230274)
Sample Cup Rack Stickers - # 41 to 50 (2 of 5 Int.) HAX0277 (1207230277)
Sample Cup Rack Stickers - # 51 to 60 (2 of 5 Int.) HAX0278 (1207230278)
Sample Cup Rack Stickers - # 61 to 70 (2 of 5 Int.) HAX0279 (1207230279)
Calibrator Sample Rack Stickers - # 71 to 84 (2 of 5 Int.) HAX0280 (1207230280)
Control Sample Rack Stickers - # 85 to 99 (2 of 5 Int.) HAX0281 (1207230281)
PC400/P400 Open Cas. sticker sheet1 HAX0334 (1207230334)
PC400/P400 Open Cas. sticker sheet2 HAX0335 (1207230335)
Water Container LBH004A (1207791004)
Sample Cup Rack XDA805B (1209132805)
Reagent Rack + Adaptors XDA849A (1209131849)
Cal/Ctrl Reagent Rack + Adaptors XDA850A (1209131850)
Tube Adaptors XDA968A (1209131968)
Sample Rack + Adaptors XEA778CS (1209159778)
4 mL Vial Adaptors XEA780AS (1209159780)
10 mL Vial Adaptors XEA781AS (1209159781)
Sample Cup Adaptors XEA782AS (1209159782)
Waste Container XEA783AS (1209159783)
You should preferably use the sample rack labels from HAX0166 (1207230166) to
HAX0174 (1207230174). Nevertheless, if you use the 2 of 5 interleaved barcode type
without check digit, you must use the sample rack labels from HAX0273 (1207230273) to
HAX0281 (1207230281).
Spare parts
Designation Reference
Light Bulb Dismantling Allen Key MAB086A (1207821086)
Reagent Needle XBA564DT (1209119564)
Sample Needle XBA847AT (1209119847)
Designation Reference
ABX Pentra Standard 1 A11A01717
ABX Pentra Standard 2 A11A01718
ABX Pentra Sodium-E A11A01738
ABX Pentra Chloride-E A11A01739
ABX Pentra Potassium-E A11A01740
ABX Pentra Reference-E A11A01741
ABX Pentra Etching CP A11A01769
ABX Pentra Dummy Electrode A11A01851
ABX Pentra Reference 280 mL A11A01901
Cleaning kit for Chloride electrode XEA935AS (1209159935)
User Manual
MONTH:
DAILY
DAY: 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
ISE module activation (beginning of the day)
ISE module cleaning with Etching (shutdown if > 20 samples)
Needles cleaning with Deproteinizer (shutdown)
WEEKLY
WEEK:
DATE DATE
Reagent needle cleaning (external part) Glycol level checking
Wash towers cleaning Filter replacement
Syringe plunger tips replacement
DATE CV (%)
Test T1/P1
Cooling unit condenser cleaning
ON REQUEST
371
Maintenance
Maintenance and Troubleshooting
Maintenance and Troubleshooting
Troubleshooting
3. Troubleshooting
This chapter provides instructions for solving problems that may occur when using the Pentra C400
analyzer.
1. Make sure the main switch, at the back of the analyzer is on (I).
2. Make sure the power cord is properly connected.
3. Make sure the instrument is receiving power from the wall outlet.
4. Check the main fuses:
a. Make sure the main switch, at the back of the instrument is off (0).
b. Unplug the power cable from the analyzer.
c. Pull out the fuse holder.
d. Check the two fuses.
e. If necessary replace the fuses by new ones provided in the accessory kit.
If the problem persists, please contact your local HORIBA Medical representative.
If the problem persists, please contact your local HORIBA Medical representative.
Follow this procedure if you have a cassette or a rack displayed in red on the main menu.
d. Make sure that each calibrator/control configured on the rack has not expired.
Related information:
■ To Configure Racks, p.307
■ To Activate a Cassette, p.223
■ To Activate an Open Cassette, p.226
■ To Activate a Reagent on a Reagent Rack, p.229
■ To Register a Reagent, p.299
Follow this procedure if you have a cassette or a rack not displayed on the main menu
whereas it is present on the reagent tray.
1. Make sure that the barcode label is well-positioned and not dirty.
2. The reagent barcode reader may be out of service.
Please contact your local HORIBA Medical representative.
If the problem persists, please contact your local HORIBA Medical representative.
Follow this procedure if the warning #305 Reagent tray temperature exceeds limits or the
alarm #055 Refrigerated Compartment Temperature regulation failed is triggered. This
means that the temperature in the refrigerated area of the reagent tray is out of range or
the temperature regulation of the reagent tray is out of service.
1. Check the temperature in the refrigerated area of the reagent tray in Main menu > Services >
Customer Services > Analyser.
Expected value: 4°C - 10°C (39°F - 50°F)
2. Check the glycol level. Refer to the Maintenance > Bimonthly Procedures (Every Two Months) > To
Check Glycol Level chapter.
If the problem persists, please contact your local HORIBA Medical representative.
Related information:
■ To Check Glycol Level, p.352
Follow this procedure if you have a rack appearing totally in red on the main menu.
1. Check if a rack having the same rack number is placed on the sample tray.
2. Remove it if necessary.
Follow this procedure if you have a rack not displayed on the main menu whereas it is
present on the sample tray.
1. Make sure that the rack was previously configured in Main menu > Services > System
Configuration > Analyser.
2. Make sure that the barcode label is well-positioned and not dirty.
3. Make sure that the barcode type of the label is configured in Main menu > Services > System
Configuration > Local Settings.
4. The sample barcode reader may be out of service.
Please contact your local HORIBA Medical representative.
Related information:
■ To Configure Racks, p.307
■ To Configure Sample Barcode Reader, p.315
■ To Check the Sample Barcode Reader, p.376
Follow this procedure if you have a rack displayed full on the main menu whereas it is
empty on the sample tray.
Related information:
■ To Check the Sample Barcode Reader, p.376
Follow this procedure if you have a patient sample, a calibrator or a control displayed in
red on the main menu.
Related information:
■ To Clean Needles with Deproteinizer, p.340
■ To Replace the Sample Needle, p.358
Follow this procedure to make sure that the sample barcode reader works properly.
Access: Main menu > Services > Customer Services > Analyser
Press a sample tray position (left side of the area) to display the corresponding rack
positions (right side of the area).
If the sample barcode reader does not work properly, please contact your local HORIBA Medical
representative.
Follow this procedure to make sure that the sample barcode reader correctly reads a
barcode label.
Access: Main menu > Services > Customer Services > Barcode
1. Select an available rack position on the sample tray from the Sector and Position spin boxes.
2. Place a sample tube with barcode at the selected rack position.
3. Press Polling.
■ The sample tube is positioned in front of the sample barcode reader.
■ The sample barcode reader starts continuous reading of the barcode label.
■ The barcode, the barcode type as well as the correct reading percentage are displayed. "No
code" means that no barcode label is read.
4. Press Stop to stop the continuous reading of the barcode label.
If the sample barcode reader does not work properly, make sure that the barcode label is well-
positioned and that it is not dirty.
If the problem persists, please contact your local HORIBA Medical representative.
Follow this procedure if you have problems with the reagent syringe, the sample syringe or
both.
Related information:
■ To Replace Syringes, p.355
Follow this procedure if you have problems with the reagent needle, the sample needle or
both.
1. If you have problems with the reagent needle, this one may be damaged. In this case, replace the
reagent needle by referring to the Maintenance > Other Procedures > To Replace the Reagent
Needle chapter.
2. If you have problems with the sample needle, this one may be damaged. In this case, replace the
sample needle by referring to the Maintenance > Other Procedures > To Replace the Sample
Needle chapter.
If the problem persists, please contact your local HORIBA Medical representative.
Related information:
■ To Replace Reagent Needle, p.356
■ To Replace the Sample Needle, p.358
Follow this procedure if the warning #124 Heating needle temperature exceeds limits or
the alarm #054 Heating Needle Temperature regulation failed is triggered. This means that
the reagent needle temperature is out of range or the temperature regulation of reagent
needle is out of service.
1. Check the reagent needle temperature in Main menu > Services > Customer Services >
Analyser.
Expected value: 37°C +/- 0.5°C (99°F +/- 0.9°F)
2. The temperature regulation of the reagent needle may be out of service.
Please contact your local HORIBA Medical representative.
Follow this procedure if you have a coefficient of variation (CV) > 1.0% when performing
the precision test (T1 or P1).
If the problem persists, please contact your local HORIBA Medical representative.
Related information:
■ To Replace Syringe Plunger Tips, p.346
■ To Replace Syringes, p.355
■ Performing the Precision Test, p.348
Follow this procedure if a warning or an alarm Used Cuvette Holder Missing is triggered.
If the problem persists, please contact your local HORIBA Medical representative.
Follow this procedure if a warning or an alarm Used Cuvette Holder Full is triggered.
Dispose of used cuvettes according to your local and/or national guidelines for
biohazard waste disposal.
3. Make sure that the used cuvette sensor is not dirty and clean it if necessary.
If the problem persists, please contact your local HORIBA Medical representative.
Disable the cuvette changer and manually load cuvettes by referring to the Cuvette Changer > To
Manually Load Cuvettes chapter.
Related information:
■ To Manually Load Cuvettes, p.383
Follow this procedure if the warning #301 Cuvette Holder Empty is triggered.
■ Do not wash and re-use cuvettes. They are designed for single use only.
■ Cuvette segments with at least one used cuvette must never be loaded in the
reaction tray.
3. If the problem persits, the new cuvette sensor may be out of service.
Please contact your local HORIBA Medical representative.
Disable the cuvette changer and manually load cuvettes by referring to the Cuvette Changer > To
Manually Load Cuvettes chapter.
Related information:
■ To Manually Load Cuvettes, p.383
Follow this procedure if you hear noise in the reaction tray and if a warning or an alarm is
triggered.
If not, press Stop and wait for the instrument to turn to "Emergency Stop" status.
3. Manually turn the reaction tray and make sure that each cuvette segment is correctly loaded in the
reaction tray by pressing it down.
4. Close the reaction tray door.
5. Put the cuvettes hatch back.
Follow this procedure to make sure that the cuvette changer works properly.
Access: Main menu > Services > Customer Services > Cycles
1. Select the first position of the reaction tray from the Sector spin box.
2. Press Unload to unload a used cuvette segment from the first position of the reaction tray.
This function is unavailable if the "used cuvette" holder is full.
3. Press Load to load a new cuvette segment into the first position of the reaction tray.
This function is unavailable if the "new cuvette" holder is empty or if the used cuvette holder is full.
4. Make sure that there is no abnormal noise when loading/unloading cuvette segments in the
reaction tray.
5. Repeat steps 1 to 4 for each reaction tray position.
If the cuvette changer does not work properly, please contact your local HORIBA Medical
representative.
Disable the cuvette changer and manually load cuvettes by referring to the Cuvette Changer > To
Manually Load Cuvettes chapter.
Related information:
■ To Manually Load Cuvettes, p.383
Follow this procedure if you have problems with the cuvette changer and you have to
disable it.
Access: Main menu > Services > System Configuration > Analyser
c. If necessary, unload the used cuvette segment from the reaction tray.
Dispose of used cuvettes according to your local and/or national guidelines for
biohazard waste disposal.
■ Ensure that the cuvette segment is correctly loaded in the reaction tray by
pressing it down.
■ Do not wash and re-use cuvettes. They are designed for single use only.
■ Cuvette segments with at least one used cuvette must never be loaded in the
reaction tray.
Follow this procedure if a warning or an alarm Manual Cuvette Loading Door Open is
triggered.
If the problem persists, please contact your local HORIBA Medical representative.
Follow this procedure if the warning #123 Reaction tray temperature exceeds limits or the
alarm #053 Reaction Tray Temperature regulation failed is triggered. This means that the
reaction tray temperature is out of range or the temperature regulation of reaction tray is
out of service.
1. Check the reaction tray temperature in Main menu > Services > Customer Services > Analyser.
Expected value: 37°C +/- 0.2°C (99°F +/- 0.36°F)
2. Check the room temperature.
3. The temperature regulation of the reaction tray may be out of service.
Please contact your local HORIBA Medical representative.
Follow this procedure if the alarm #505 Light stability Not OK or the alarm #517 Light
intensity out of range is triggered.
If the problem persists, please contact your local HORIBA Medical representative.
Related information:
■ To Replace the Lamp, p.361
Follow this procedure to check the spectrophotometer gains for each wavelength.
Access: Main menu > Services > Customer Services > Analyser
Spectrophotometer gains are adjusted and checked for the fifteen wavelengths at each instrument
initialization.
1. Check the spectrophotometer gains for each wavelength from the Spectro Gain area.
Spectrophotometer gains are expressed in % and should be included between 1% and 95%.
■ If one of the gains is included between 95% and 98%, the warning #629 The lamp is dying, it is
still operational but you will need to change it soon. [wavelength nm / gain in % of Gain]. is
triggered. This means that spectrophotometer gains are reaching limits. You will have to
replace the lamp soon.
■ If one of the gains is < 1% or > 98%, the alarm #520 The lamp has failed the gain check.
[wavelength nm / gain in % of Gain] Please change the lamp. is triggered. This means that
spectrophotometer gains are out of range. You have to replace the lamp.
2. To replace the lamp, refer to the Maintenance > Other Procedures > To Replace the Lamp chapter.
Related information:
■ To Replace the Lamp, p.361
3.6.1. To Check the ISE Module Status and ISE Error Code from
Customer Services Menu
Access: Main menu > Services > Customer Services > ISE
Code Message
000 Module not initialised
001 Module initialised
002 Module configured
003 Mechanical part initialised
004 Module sleeping
Code Message
005 Fluidic part initialised
006 Module calibrated
2. If the ISE module status is lower than expected, press Global Initialization.
A global initialization of the ISE module is performed. It takes a few minutes.
3. If you have the ISE Module Status button displayed in red on the main menu, check the ISE error
code which has been triggered in the ISE Status area.
For detailed information concerning ISE error codes, refer to the Alarms > ISE Error
Codes chapter.
4. In case of mechanical error on the ISE module, refer to the ISE Module (Option) > Mechanical Error
on the ISE Module chapter.
5. In case of reagent/fluidic error on the ISE module, refer to the ISE Module (Option) > Reagent/
Fluidic Error on the ISE Module chapter.
Related information:
■ Mechanical Error on the ISE Module, p.387
■ Reagent/Fluidic Error on the ISE Module, p.387
Follow this procedure if you have a mechanical error on the ISE module.
Access: Main menu > Services > Customer Services > ISE
Follow this procedure if you have a reagent/fluidic error on the ISE module.
Access: Main menu > Services > Customer Services > ISE
a. If an ISE reagent needs to be replaced, refer to the Maintenance > Other Procedures > To
Replace ISE Reagents (Option) chapter.
b. Press Global Initialization.
A global initialization of the ISE module is performed. It takes a few minutes.
2. If you have just replaced an electrode:
a. Make sure that you removed the protection on each side of the electrode.
b. If not, remove it and press Global Initialization.
A global initialization of the ISE module is performed. It takes a few minutes.
c. Make sure that the O-ring is well-positioned on one side of the electrode, make sure that the
electrode is well-positioned by referring to the Maintenance > Other Procedures > To Replace
Electrodes (Option) chapter.
d. If not, reinstall the electrode and press Global Initialization.
A global initialization of the ISE module is performed. It takes a few minutes.
3. If the ISE module has been shut down for a long period, salt-crystals may have formed.
a. Remove the electrodes by referring to the Maintenance > Other Procedures > To Replace
Electrodes (Option) chapter.
b. Clean the electrodes with distilled water.
c. Reinstall the electrodes and press Global Initialization.
A global initialization of the ISE module is performed. It takes a few minutes.
4. If the problem persists after several global initializations of the ISE module, please contact your
local HORIBA Medical representative.
Related information:
■ To Replace ISE Reagents (Option), p.363
■ To Replace Electrodes (Option), p.363
Follow this procedure if you have the ISE Module Status button displayed in blue with a
red circle on the main menu. This indicates an error in the ISE calibration.
1. Make sure that the ISE module maintenance procedures have been performed on time.
If not, refer to the Maintenance > Daily Procedures > To Clean the ISE Module with Etching
(Option) chapter for daily maintenance or to Maintenance > Weekly Procedures > To Clean the ISE
Module chapter for weekly maintenance.
2. Check the volume and expiration date of Standard 1, Standard 2 and Reference bottles.
a. If an ISE reagent needs to be replaced, refer to the Maintenance > Other Procedures > To
Replace ISE Reagents (Option) chapter.
b. Press Global Initialization.
A global initialization of the ISE module is performed. It takes a few minutes.
3. Make sure that the Reference bottle and the Standard 1 bottle are not inverted by referring to the
Maintenance > Other Procedures > To Replace ISE Reagents (Option) chapter.
a. If necessary, put the bottles back in place.
b. Press Global Initialization.
A global initialization of the ISE module is performed. It takes a few minutes.
4. An electrode may have expired.
a. Press the ISE Module Status button from the main menu.
The ISE calibration results screen is displayed.
For detailed information concerning ISE calibration results interpretation, refer to the
ISE Calibration (Option) > ISE Calibration Results chapter.
c. If the ISE calibration results are bad for one electrode, the concerned electrode may have
expired. Replace the concerned electrode by referring to the Maintenance > Other Procedures
> To Replace Electrodes (Option) chapter.
d. If the ISE calibration results are bad for all electrodes, the reference electrode may have
expired. Replace the reference electrode by referring to the Maintenance > Other Procedures >
To Replace Electrodes (Option) chapter.
5. If you have just replaced the electrodes, run successively five 2-point calibrations to stabilize the
electrodes.
Refer to the ISE Calibration (Option) > To Run an ISE Calibration chapter.
Related information:
■ To Clean the ISE Module with Etching (Option), p.339
■ To Clean the ISE Module (Option), p.341
■ To Replace ISE Reagents (Option), p.363
■ To Replace Electrodes (Option), p.363
■ To Run an ISE Calibration, p.115
Follow this procedure if the warning #610 Water Tank Empty is triggered.
Follow this procedure if the water tank is empty or if the waste tank is full and no warning
is triggered.
Follow this procedure if the warning #307 Water level of the cold group is insufficient is
triggered.
1. Check the glycol level. Refer to the Maintenance > Bimonthly Procedures (Every Two Months) > To
Check Glycol Level chapter.
2. The glycol level sensor may be out of service.
Please contact your local HORIBA Medical representative.
Related information:
■ To Check Glycol Level, p.352
3.9.1.
Yes Yes Remove bubbles
Alarm on blank Reagent bubbles (reagent or diluent)
No No
maintenance
Result Problems
No
Use new reagent
Calibration Results
Yes
Yes Yes No
Factor alarm Correct lot/target Fresh calibrator Use new calibrator
No No
Change lot/target
Yes
Deviation alarm Well-mixed calibrator
Yes
No No
Mix calibrator
Remove bubbles
Analytical alarm*
Yes
No
Run control
Follow this troubleshooting diagram if calibration results are flagged.
* Refer to Alarms > Analytical and Quality Flags > Analytical Flags chapter.
No
Analytical alarm* Run patient sample
Yes
Correct Yes Yes Yes Yes Correct configuration Yes Run T1/P1 and check
confidence range Same result on rerun New calibration New reagent of application maintenance
No No No No No
Rerun calibration
Well-mixed and Yes Enough volume Yes Run T1/P1 and check
reconstituted control in sample cup maintenance
No No
Use new sample cup
Mix control
Remove bubbles
Follow this troubleshooting diagram if control results are flagged.
Check lot/target
* Refer to Alarms > Analytical and Quality Flags > Analytical Flags chapter.
393
Troubleshooting
Maintenance and Troubleshooting
394
PROBLEM ACTION
Troubleshooting
3.9.3.
No No No
Rerun sample
Maintenance and Troubleshooting
Yes Correct configuration Yes Correct sample type Yes Yes Check patient history
Same result on rerun of application (serum, plasma) Control ok
to validate
Patient Results
No No No No
Rerun control,
calibration
Refer to reagent
notice
Check application
No No No
Transfer sample
in sample cup
Follow this troubleshooting diagram if patient results are flagged.
Centrifuge correctly
Remove bubbles,
fibrin
No
ISE control problem Run patient sample
Yes
Well-mixed and Yes Enough volume Yes Run T1/P1 and check
reconstituted control in sample cup maintenance
ISE Control Results (Option)
No No
Use new sample cup
Before any investigation, check the following items:
Mix control
Remove bubbles
Use new control
Check lot/target
Follow this troubleshooting diagram if ISE control results are flagged.
395
Troubleshooting
Maintenance and Troubleshooting
Maintenance and Troubleshooting
Troubleshooting
ISE patient sample Yes Yes Correct configuration Yes Correct sample type Yes Yes Check patient history
problem Same result on rerun of application (serum, plasma) Control ok
to validate
No No No No No
Refer to ISE control
problem
No No
Transfer sample
in sample cup
Centrifuge correctly
Remove bubbles,
fibrin
397
Troubleshooting
Maintenance and Troubleshooting
Maintenance and Troubleshooting
Troubleshooting
The System Warnings menu lists the system warnings and alarms as well as the maintenance alerts
that are triggered during instrument operation.
The All tab displays the system warnings and alarms that are triggered.
The system warnings and alarms are sorted by date and are listed in a table format which indicates
the event level (warning or alarm), the event date and time, the event code and message.
You can display either the system warnings or the system alarms, or both by pressing Warning,
Alarm or All tab.
For detailed information on system warnings and alarms, refer to the System Warnings
and Alarms > Warnings and Alarms chapter.
Alert tab
The Alert tab displays the maintenance procedures that must be performed on your instrument.
These procedures must be configured in Main menu > Services > System Configuration >
Maintenance.
The maintenance procedures are sorted by frequency and are listed in a table format which indicates
the procedure name and frequency, when the procedure must be performed. If a maintenance
procedure is not performed on time, a maintenance alert is triggered and indicated by a red
exclamation mark.
Related information:
■ Configuring Maintenance Alerts, p.329
■ Warnings and Alarms, p.401
The table below lists the system warnings and alarms that can appear during instrument operation.
When a system warning or a system alarm is triggered, the System Warnings button becomes red
until you press it to open the System Warnings menu.
■ System warnings do not block the instrument and are displayed to warn the user. When a warning
is triggered, the instrument is not stopped and turns to "Ready" status.
■ System alarms block the instrument. When an alarm is triggered, the instrument is stopped and
turns to "Emergency Stop" status.
When a system warning or a system alarm is triggered, follow the instructions given in the table
below.
If the problem persists, please contact your local HORIBA Medical representative.
Related information:
■ Temperature Warnings and Alarms, p.417
Related information:
■ Temperature Warnings and Alarms, p.417
■ Compressor Overload Alarm, p.417
■ Reagent Level Detection Board Errors, p.418
Related information:
■ Sampling Alarms, p.420
■ ISE Error Codes, p.448
■ Compressor Overload Alarm, p.417
The table below describes how the temperature warnings and alarms are triggered.
* Every hour from the start of the cooling unit, the temperature in the refrigerated area of
the reagent tray (if it is still out of the acceptable range) is checked to ensure that it is
decreasing by at least two degrees per hour. If not, the alarm #055 Refrigerated
Compartment Temperature regulation failed is triggered.
ON
OFF
0 5 10 15 20 25 Time
ON
OFF
0 5 10 15 20 25 Time
The table below lists the errors that can appear on the reagent level detection board and can trigger
the alarm #157 Error card reagent detection [error code: message].
During the instrument initialization, the following dialog box may be displayed: Transmission not
successful, do you want to continue to process queued messages?. This message means that data
sent to the host is pending.
■ Press OK to send the pending data to host as soon as the connection is running.
■ Press Cancel to delete the pending data.
2. Sampling Alarms
The table below lists the sampling alarms that can appear during instrument operation.
When a sampling alarm is triggered, the Test Review button becomes red until you press it to open
Test Review menu.
When a sampling alarm is triggered, follow the instructions given in the table below.
If the problem persists, please contact your local HORIBA Medical representative.
This chapter describes the analytical flags that apply to kinetic curves for all cuvettes (Blank,
Calibrator, Control, Patient sample).
HIGH_ABS
Message HIGH_ABS
Definition At least one absorbance point used in the calculation has exceeded 3.0 A if the
wavelength is 340 nm, or 3.5 A for other wavelengths.
Applies on Blank, Calibration, Control, Patient sample
Implications The result is not displayed.
■ Blank and calibration: The calibration is then flagged with CALCHK_ERROR.
■ Control: The control cannot be validated.
■ Patient sample: The sample is automatically rerun with post-dilution if programmed in
the application.
Actions ■ Blank and calibration: Reassay with fresh reagent.
■ Control: Reassay with fresh control. Check pipetting process (needles, run T1 or P1).
■ Patient sample: If no post-dilution is programmed in the application, manually dilute
the sample and reassay.
normal run
OD run with flag
OD > 3.0
at 340 nm
SAMPLE_LIMIT
Message SAMPLE_LIMIT
Definition The sample limit value is used to determine the sample-specific absorbance.
The reagent blank absorbance is subtracted from the sample absorbance, both measured
at the defined cycle. The calculated value is checked to ensure that it is not higher than the
programmed sample limit value.
Applies on Patient sample
Implications The result is displayed and flagged.
Actions Check the sample visually for any interferences; it may be icteric, hemolytic or lipemic.
Dilute it manually and reassay.
normal run
run with flag
OD blank
Sample
limit
REAG_RANGE_HIGH / LOW
Message REAG_RANGE_HIGH
REAG_RANGE_LOW
Definition All reagent blank absorbance values used for calculations are checked to ensure that they
are not out of the programmed reagent range.
Applies on Blank
Implications The result is displayed and flagged.
The calibration is then flagged with CALCHK_FLAG.
Actions Reassay with fresh reagent. Check the water quality and verify the maintenance. Check
the quality of the cuvettes.
Reagent
range OD
BLK_RANGE
Message BLK_RANGE
Definition The absorbance variation of the reagent blank is checked to ensure that it is not out of the
programmed blank range.
Applies on Blank
Implications The result is displayed and flagged.
The calibration is then flagged with CALCHK_FLAG.
Actions Reassay with fresh reagent. Check the water quality and verify the maintenance. Check
the quality of the cuvettes.
ΔOD
Blank range
NON_LIN
Message NON_LIN_A, NON_LIN_B, NON_LIN_C, NON_LIN_D
Definition The Kinsearch calculation type failed to find a linear portion of the reaction curve or less
than 5 data points are within the linear portion. This flag applies to each calculation step
(Steps A, B, C and D).
Applies on Blank, Calibration, Control, Patient sample
Implications The result is not displayed.
■ Blank and calibration: The calibration is then flagged with CALCHK_ERROR.
■ Control: The control cannot be validated.
■ Patient sample: The sample is automatically rerun with post-dilution if programmed in
the application.
Actions ■ Blank and calibration: Reassay with fresh reagent.
■ Control: Reassay with fresh control and/or reagent.
■ Patient sample: If no post-dilution is programmed in the application, manually dilute
the sample and reassay.
Reassay with fresh reagent if more than one sample gets the alarm.
normal run
run with flag
OD
Reading zone
NOISE
Message NOISE_A, NOISE_B, NOISE_C, NOISE_D
Definition Only for the Kinsearch calculation type. One or more data points within the linear portion of
the reaction curve exceed a calculated difference from the regression line. This flag applies
to each calculation step (Steps A, B, C and D).
Applies on Blank, Calibration, Control, Patient sample
Implications The result is displayed.
■ Blank and calibration: Does not flag nor block the calibration calculation.
■ Control: The control cannot be automatically validated. If a default control is manually
validated, all following results are flagged with Q (CTRL_ERROR).
■ Patient sample: The result is flagged.
Actions ■ Blank and calibration: Rerun if necessary.
■ Control: Reassay with fresh control and/or reagent.
■ Patient sample: Reassay with fresh reagent if more than one sample gets the alarm.
normal run
OD run with flag
Reading zone
HIGH_ACTIV
Message HIGH_ACTIV
Definition This flag appears if the reaction limit check and the reaction direction check fail
simultaneously.
Applies on Control, Patient sample
Implications The result is not displayed.
■ Control: The control cannot be validated.
■ Patient sample: The sample is automatically rerun with post-dilution if programmed in
the application.
Actions ■ Control: Reassay with fresh control. Check pipetting process (needles, run T1 or P1).
■ Patient sample: If no post-dilution is programmed in the application, manually dilute
the sample and reassay.
REAC_LIMIT
Message REAC_LIMIT_A, REAC_LIMIT_B, REAC_LIMIT_C, REAC_LIMIT_D
Definition Only for Kinetic and Kinsearch calculation types.
The reaction limit value is used to determine substrate depletion.
■ For the Kinetic calculation type: the absorbance measured at the defined cycle is
subtracted from the absorbance value of the first reading point.
■ For the Kinsearch calculation type: the absorbance measured at the defined cycle is
subtracted from the absorbance value of the first point within the linear range.
The calculated value is checked to ensure that it is not higher than the programmed
reaction limit value. This flag applies to each calculation step (Steps A, B, C and D).
Applies on Control, Patient sample
Implications The result is not displayed.
■ Control: The control cannot be validated.
■ Patient sample: The sample is automatically rerun with post-dilution if programmed in
the application.
Actions ■ Control: Reassay with fresh control. Check pipetting process (needles, run T1 or P1).
■ Patient sample: If no post-dilution is programmed in the application, manually dilute
the sample and reassay.
OD normal run
run with flag
Reaction
limit Abs
SIGN
Message SIGN
Definition The absorbance change for the flagged sample minus the absorbance change of the
reagent blank does not agree with the reaction direction programmed in the application.
Applies on Calibration, Control, Patient sample
Implications The result is not displayed.
■ Calibration: The calibration is then flagged with CALCHK_ERROR.
■ Control: The control cannot be validated.
■ Patient sample: The sample is flagged.
Actions ■ Calibration: Check no air bubbles. Check pipetting process (needles, run T1 or P1).
■ Control: Check no air bubbles. Check pipetting process (needles, run T1 or P1).
■ Patient sample: Check no air bubbles, no fibrin clot. Check sample type (serum,
plasma, EDTA).
Reassay with fresh reagent if more than one sample gets the alarm.
normal run
run with flag
OD
+
DIV_ABS
Message DIV_ABS_A, DIV_ABS_B, DIV_ABS_C, DIV_ABS_D
Definition Only for Kinetic and Kinsearch calculation types.
The correlation coefficient r2 and the standard deviation SD are calculated.
■ For the Kinetic calculation type, the measurement points used to calculate r2 and SD
are those included between the first reading cycle and the last reading cycle.
■ For the Kinsearch calculation type, the measurement points used to calculate r2 and
SD are those included in the linear range found by the linear search program.
If r2 is lower than the configured r2 threshold and SD is higher than the configured SD
threshold, then the analytical flag DIV_ABS is triggered. This flag applies to each
calculation step (Steps A, B, C and D).
Applies on Calibration, Control, Patient sample
DIV_ABS
Implications The result is displayed and flagged.
■ Calibration: The calibration is then flagged with CALCHK_FLAG.
■ Control: The control cannot be automatically validated. If a default control is manually
validated, all following results are flagged with Q (CTRL_ERROR).
■ Patient sample: The result is flagged.
Actions Check the absorbances. Make sure that there are no air bubbles. Reassay.
normal run
run with flag
OD
r2 = 0.99
r2 = 0.70
For detailed information concerning the absorbance deviation checks, refer to the
Description and Technology > Data Analysis > Absorbance Measurements > Calculation
Types > Absorbance Deviation Checks chapter.
DIV_ABSi
Message DIV_ABSi_A, DIV_ABSi_B, DIV_ABSi_C, DIV_ABSi_D
Definition Only for the Endpoint calculation type.
Two different algorithms can be used for this check: one if the first reading cycle < 4,
another if the first reading cycle ≥ 4.
■ If the first reading cycle < 4, a calculated absorbance difference is checked to ensure
that it is included in the configured blank range.
■ If the first reading cycle ≥ 4, three calculated absorbance differences are checked to
ensure that they are lower than the configured threshold. If at least two absorbance
differences are greater than or equal to the configured threshold, the analytical flag
DIV_ABSi is triggered.
This flag applies to each calculation step (Steps A, B, C and D).
Applies on Calibration, Control, Patient sample
Implications The result is displayed and flagged.
■ Calibration: The calibration is then flagged with CALCHK_FLAG.
■ Control: The control cannot be automatically validated. If a default control is manually
validated, all following results are flagged with Q (CTRL_ERROR).
■ Patient sample: The result is flagged.
Actions Check the absorbances. Make sure that there are no air bubbles. Reassay.
Reagent blank
OD Normal run
Run with flag
A0
An
Blank range
0 12 24 36 48 60 72
Time (s)
Normal run
Run with flag
OD
A0
D1
For detailed information concerning the absorbance deviation checks, refer to the
Description and Technology > Data Analysis > Absorbance Measurements > Calculation
Types > Absorbance Deviation Checks chapter.
DIV_ABSf
Message DIV_ABSf_A, DIV_ABSf_B, DIV_ABSf_C, DIV_ABSf_D
Definition Only for the Endpoint calculation type.
The difference between absorbance values at time n and at time n-1
(where n = last reading cycle) is calculated and checked to ensure that it is included in a
calculated range. This flag applies to each calculation step (Steps A, B, C and D).
Applies on Calibration, Control, Patient sample
DIV_ABSf
Implications The result is displayed and flagged.
■ Calibration: The calibration is then flagged with CALCHK_FLAG.
■ Control: The control cannot be automatically validated. If a default control is manually
validated, all following results are flagged with Q (CTRL_ERROR).
■ Patient sample: The result is flagged.
Actions Check the absorbances. Make sure that there are no air bubbles. Reassay.
normal run
run with flag
OD
n
D1
For detailed information concerning the absorbance deviation checks, refer to the
Description and Technology > Data Analysis > Absorbance Measurements > Calculation
Types > Absorbance Deviation Checks chapter.
Related information:
■ Calibration Analytical Flags, p.433
■ Quality Flags, p.444
■ Absorbance Deviation Checks, p.473
This chapter describes the analytical flags that apply when calculating calibration curves.
CALCHK_FLAG
Message CALCHK_FLAG
Definition At least one calibrator level is flagged with REAG_RANGE_HIGH / LOW, BLK_RANGE,
DEVIATION, DIV_ABS, DIV_ABSi or DIV_ABSf.
Applies on Calibration
CALCHK_FLAG
Implications The calibration is calculated and flagged. The calibration cannot be automatically
validated. If manually validated, all following results are flagged with C (CAL_ERROR).
Actions Check the absorbances. Make sure that there are no air bubbles and that the calibrator is
properly mixed. Check the pipetting precision (run T1 or P1). Reassay with fresh calibrator
and/or reagent. Check the water quality and verify the maintenance. Check the quality of
cuvettes.
CALCHK_ERROR
Message CALCHK_ERROR
Definition At least one calibrator level is flagged with HIGH_ABS, NON_LIN or SIGN.
Applies on Calibration
Implications The calibration cannot be calculated and must be deleted. Then, the previous calibration
will become the current calibration and all following results are flagged with C
(CAL_ERROR).
Actions Make sure that there are no air bubbles. Check pipetting process (needles, run T1 or P1).
Reassay with fresh reagent.
CALC_ERROR
Message CALC_ERROR
Definition Three cases trigger this analytical flag:
■ For linear calibration modes (Slope average, Linear regression), the calculated
calibration factor is negative.
■ For non-linear calibration modes, the calculation algorithm cannot generate an
acceptable standard non-linear curve fit.
■ For mathematical algorithms (LOGIT/LOG4, LOGIT/LOG5, EXPONENT5), the curve fit
is not monotonic.
Applies on Calibration
Implications The calibration cannot be calculated and must be deleted. Then, the previous calibration
will become the current calibration and all following results are flagged with C
(CAL_ERROR).
Actions Check the calibrator position. Reassay with fresh calibrator and/or reagent.
CAL_NOT_OPTIMISED
Message CAL_NOT_OPTIMISED
Definition For non-linear calibration modes, a maximum of 150 iterations is performed to optimize the
curve fit. If the number of 150 iterations is reached, the calibration is flagged. This means
that the calibration has been calculated but not with the best optimization.
Applies on Calibration
Implications The calibration is calculated and flagged. The calibration can be validated manually or
automatically. If validated, all following results are flagged with C (CAL_ERROR).
Actions Reassay with fresh calibrator and/or reagent.
FACTOR_FLAG
Message FACTOR_FLAG
Definition For linear calibration modes (Slope average, Linear regression), the calculated calibration
factor failed one of the three calibration factor checks.
Applies on Calibration
FACTOR_FLAG
Implications The calibration is calculated and flagged. The calibration cannot be automatically
validated. If manually validated, all following results are flagged with C (CAL_ERROR).
Actions Ensure that the entered target values correspond to the calibrator name and lot number
used. Reassay with fresh calibrator and/or reagent.
ΔOD
Calibration curve
Factor low
Factor high
Calibrator Concentrations
concentration
Related information:
■ Kinetic Analytical Flags, p.423
■ Result Analytical Flags, p.435
■ Quality Flags, p.444
DEVIATION
Message DEVIATION
Definition Two types of deviation are checked.
■ Deviation concerning replicates (Dev_Rep): For each calibrator level, the
absorbance values of replicates are compared to the mean value. The maximum
deviation is calculated (Dev_Rep) and then checked to ensure that it is not higher than
the authorized Dev_Rep.
■ Deviation to the curve (Dev_C): For each calibrator level, the concentration of the
mean value is compared to the theoretical value. The deviation is calculated (Dev_C)
and then checked to ensure that it is not higher than the authorized Dev_C.
Applies on Calibration
Implications The result is displayed and flagged.
The calibration is then flagged with CALCHK_FLAG.
Actions Make sure that there are no air bubbles and that the calibrator is properly mixed. Check
pipetting precision (run T1 or P1). Reassay with fresh calibrator and/or reagent.
ΔOD
- Replicate 1'
- Replicate 2'
Calibrator Concentrations
concentration
Deviation concerning replicates
ΔOD
calibration
curve
Calculated
concentration
RESULT_CALC_ERR
Message RESULT_CALC_ERR
Definition For the Linear regression calibration mode and the non-linear calibration modes, at least
one result could not be calculated using the selected algorithm.
Applies on Control, Patient sample
RESULT_CALC_ERR
Implications The result is not displayed.
■ Control: The control cannot be validated.
■ Patient sample: The sample is flagged.
Actions ■ Control: Reassay with fresh control. Check pipetting process (needles, run T1 or P1).
■ Patient sample: Reassay with fresh reagent if more than one sample gets the alarm.
LINEARITY_HIGH / LOW
Message LINEARITY_HIGH
LINEARITY_LOW
Definition The sample result is checked to ensure that it is within the programmed linearity range.
Applies on Control, Patient sample
Implications The result is displayed and flagged.
■ Control: The control cannot be automatically validated. If a default control is manually
validated, all following results are flagged with Q (CTRL_ERROR).
■ Patient sample: The sample is automatically rerun with post-dilution or post-
concentration if programmed in the application.
Actions ■ Control: Reassay with fresh control. Check pipetting process (needles, run T1 or P1).
■ Patient sample: In case of LINEARITY_HIGH flag and if no post-dilution is programmed
in the application, dilute the sample manually and reassay.
ΔOD
ΔOD
Linearity
High
calibration
curve
Linearity
Low
CALC_RANGE_HIGH / LOW
Message CALC_RANGE_HIGH
CALC_RANGE_LOW
Definition Only for non-linear calibration modes.
■ Calibration: If, for the highest calibrator level, one absorbance value is higher than the
theoretical value, the analytical flag CALC_RANGE_HIGH is triggered. If, for the lowest
calibrator level, one absorbance value is lower than the theoretical value, the analytical
flag CALC_RANGE_LOW is triggered.
■ Control and Patient sample: If the absorbance value is between the calculated mean
value and the theoretical value of the highest calibrator level, the analytical flag
CALC_RANGE_HIGH is triggered. If the absorbance value is between the calculated
mean value and the theoretical value of the lowest calibrator level, the analytical flag
CALC_RANGE_LOW is triggered.
Applies on Calibration, Control, Patient sample
CALC_RANGE_HIGH / LOW
Implications In case of CALC_RANGE_HIGH flag, the result equals the highest calibrator level
concentration. In case of CALC_RANGE_LOW flag, the result equals the lowest calibrator
level concentration.
■ Calibration: Does not flag nor block the calibration calculation.
■ Control: The control cannot be automatically validated. If a default control is manually
validated, all following results are flagged with Q (CTRL_ERROR).
■ Patient sample: The sample is automatically rerun with post-dilution or post-
concentration if programmed in the application.
Actions ■ Control: Reassay with fresh control. Check pipetting process (needles, run T1 or P1).
■ Patient sample: In case of CALC_RANGE_HIGH flag and if no post-dilution is
programmed in the application, dilute the sample manually and reassay.
ΔOD
Cal range
high
calibration
curve
TEST_RANGE_HIGH / LOW
Message TEST_RANGE_HIGH
TEST_RANGE_LOW
Definition Only for non-linear calibration modes.
If the absorbance value is higher than the calculated mean value of the highest calibrator
level, the analytical flag TEST_RANGE_HIGH is triggered. If the absorbance value is lower
than the calculated mean value of the lowest calibrator level, the analytical flag
TEST_RANGE_LOW is triggered.
Applies on Control, Patient sample
Implications The result is not displayed.
■ Control: The control cannot be validated.
■ Patient sample: The sample is automatically rerun with post-dilution or post-
concentration if programmed in the application.
Actions ■ Control: Reassay with fresh control and/or reagent.
■ Patient sample: In case of TEST_RANGE_HIGH flag and if no post-dilution is
programmed in the application, dilute the sample manually and reassay.
ΔOD
Test range
high
calibration
curve
AG_EXCESS
Message AG_EXCESS
Definition The antigen excess check is used to detect the presence of excess antigen. The
concentration measured at the defined cycle (CCP) is divided by the concentration
measured at the last reading cycle (C) and multiplied by 100. The calculated ratio is
checked to ensure that it is higher than the programmed antigen excess limit value.
Applies on Patient sample
Implications The result is displayed and flagged.
The sample is automatically rerun with post-dilution if programmed in the application.
Actions If no post-dilution is programmed in the application, dilute the sample manually and
reassay.
Ccp C
OD
Last reading
First reading
AGE
CONF_RANGE_HIGH / LOW
Message CONF_RANGE_HIGH
CONF_RANGE_LOW
Definition The control result is outside the programmed confidence range. For a default control,
Westgard rules are disabled.
Applies on Control
CONF_RANGE_HIGH / LOW
Implications The result is displayed and flagged.
The control cannot be automatically validated. If a default control is manually validated, all
following results are flagged with Q (CTRL_ERROR).
Actions Ensure that the entered target values correspond to the control name and lot number
used. Reassay with fresh control. Rerun calibration.
Concentration
CONF range High
High value
TARGET
Low value
Date
CONF_RANGE_HIGH_W / LOW_W
Message CONF_RANGE_HIGH_W
CONF_RANGE_LOW_W
Definition The default control result is outside the programmed confidence range and no Westgard
rule is true.
Applies on Control
Implications The result is displayed and flagged.
Actions Ensure that the entered target values correspond to the control name and lot number
used. Reassay with fresh control. Rerun calibration.
CONF_RANGE_HIGH_W1-6 / LOW_W1-6
Message CONF_RANGE_HIGH_W1-6
CONF_RANGE_LOW_W1-6
Definition The default control result is outside the programmed confidence range and the Westgard
rule 1, 2, 3, 4, 5 or 6 is true.
Applies on Control
Implications The result is displayed and flagged.
The control cannot be automatically validated. If a default control is manually validated, all
following results are flagged with Q (CTRL_ERROR).
Actions Ensure that the entered target values correspond to the control name and lot number
used. Reassay with fresh control. Rerun calibration.
WESTGARD_RULE_4-6
Message WESTGARD_RULE_4
WESTGARD_RULE_5
WESTGARD_RULE_6
Definition The default control result is inside the programmed confidence range and the Westgard
rule 4, 5 or 6 is true.
Applies on Control
WESTGARD_RULE_4-6
Implications The result is displayed and flagged.
The control cannot be automatically validated. If a default control is manually validated, all
following results are flagged with Q (CTRL_ERROR).
Actions Reassay with fresh control. Check pipetting precision (run T1 or P1).
For detailed information concerning Westgard rules, refer to the Quality Assurance >
Westgard Rules chapter.
REF_RANGE_HIGH / LOW
Message REF_RANGE_HIGH
REF_RANGE_LOW
Definition The result is outside the programmed reference range.
Applies on Patient sample
Implications The result is displayed and flagged.
Actions Check the patient history.
ΔOD
REF_RANGE_LOW REF_RANGE_HIGH
Normal range
CRITICAL_RANGE_HIGH / LOW
Message CRITICAL_RANGE_HIGH
CRITICAL_RANGE_LOW
Definition The result is outside the programmed critical range.
Applies on Patient sample
Implications The result is displayed and flagged.
The sample is automatically rerun without post-dilution or post-concentration (a new test is
requested).
Actions Check the critical range programmed in the application. Check the patient history.
Related information:
■ Calibration Analytical Flags, p.433
■ Quality Flags, p.444
■ Westgard Rules, p.82
When several analytical flags are triggered for the same result, only the flag having the highest priority
is displayed.
Analytical flags priority (from highest to lowest):
■ HIGH_ABS
■ SAMPLE_LIMIT
■ REAG_RANGE_HIGH / LOW
■ NON_LIN
■ HIGH_ACTIV
■ REAC_LIMIT
■ NOISE
■ AG_EXCESS
■ BLK_RANGE
■ SIGN
■ DIV_ABS, DIV_ABSi, DIV_ABSf
■ TEST_RANGE_HIGH / LOW
■ RESULT_CALC_ERR
■ DEVIATION
■ LINEARITY_HIGH / LOW
■ CONF_RANGE_HIGH / LOW, CONF_RANGE_HIGH_W / LOW_W, CONF_RANGE_HIGH_W1-6 /
LOW_W1-6, WESTGARD_RULE_4-6
■ CRITICAL_RANGE_HIGH / LOW
■ REF_RANGE_HIGH / LOW
■ CALC_RANGE_HIGH / LOW
Expired current calibrations, calibrators, controls and solutions can be used on your
instrument only if configured by HORIBA Medical or your local representative.
When several quality flags are triggered for the same result, only the flag having the highest priority is
displayed.
Quality flags priority (from highest to lowest):
■ R (CAL_CONV)
■ C (CAL_ERROR)
■ Q (CTRL_ERROR)
■ KE (CAL_EXPIRED)
■ CE (CALIBRATOR_EXPIRED)
■ QE (CTRL_EXPIRED)
■ SE (SOL_EXPIRED)
■ I (INCOMPATIBILITY)
Related information:
■ Kinetic Analytical Flags, p.423
■ Calibration Analytical Flags, p.433
■ Result Analytical Flags, p.435
Sensibility error
Message Sensibility error
Definition The difference between the low and high standard values is out of the sensitivity normal
range.
This flag is triggered by the ISE error code 050.
Applies on Calibration
Sensibility error
Implications The ISE Module Status button is blue with a red circle. This means that the ISE module is
not calibrated.
Actions Run a 2-point calibration. Perform the ISE module maintenance. Change the concerned
electrode.
Offset error
Message Offset error
Definition The low or high standard value is out of the acceptable range.
This flag is triggered by the ISE error code 051.
Applies on Calibration
Implications The ISE Module Status button is blue with a red circle. This means that the ISE module is
not calibrated.
Actions Run a 2-point calibration. Perform the ISE module maintenance. Change the concerned
electrode.
Repeatability error
Message Repeatability error
Definition Repeatability error on the low standard value.
This flag is triggered by the ISE error code 052.
Applies on Calibration
Implications The ISE Module Status button is blue with a red circle. This means that the ISE module is
not calibrated.
Actions Make sure that there are no air bubbles in the Standard 1 bottle. Change the Standard 1
bottle. Perform the ISE module maintenance. Change the concerned electrode. Check the
spring contact of the concerned electrode.
Response error
Message Response error
Definition The potential difference during the three last seconds of a measurement is ≥ 0.40 mV.
This flag is triggered by the ISE error code 055.
Applies on Calibration
Implications The ISE Module Status button is blue with a red circle. This means that the ISE module is
not calibrated.
Actions Make sure that there are no air bubbles in the Standard 1 or Standard 2 bottle. Change the
Standard 1 or Standard 2 bottle. Perform the ISE module maintenance. Change the
concerned electrode. Check the spring contact of the concerned electrode.
ISE_ERROR
Message ISE_ERROR
Definition Two cases trigger this flag:
■ Repeatability error on the low standard value. The flag is triggered by the ISE error
code 052.
■ The potential difference during the three last seconds of a measurement is ≥ 0.40 mV.
The flag is triggered by the ISE error code 055.
Applies on Control, Patient sample
Implications The result is not displayed.
Actions Reassay with fresh control or patient sample.
Check no air bubbles in the Standard 1 or Standard 2 bottle. Change the Standard 1 or
Standard 2 bottle. Perform the ISE module maintenance. Change the concerned electrode.
Check the spring contact of the concerned electrode.
For detailed information concerning ISE error codes, refer to the Alarms > ISE Error Codes
chapter.
Related information:
■ ISE Error Codes, p.448
The table below lists the errors that can appear on the ISE module.
When an ISE error code is triggered, follow the instructions given in the table below.
If the problem persists, please contact your local HORIBA Medical representative.
1. Instrument Description.....................................................................................................452
1.1. Reagent and Sample Trays........................................................................................................452
1.2. Sampling System....................................................................................................................... 458
1.3. Analytical Module.......................................................................................................................461
1.4. ISE Module (Option)................................................................................................................... 463
1.5. Analysis Cycle Diagram............................................................................................................. 465
1. Instrument Description
The reagent tray contains 52 positions including 44 refrigerated at 4°C - 10°C (39°F - 50°F) and eight
positions at room temperature. The temperature in the refrigerated area of reagent tray is maintained
at 4°C - 10°C (39°F - 50°F) 24 hours a day by a cooling unit with glycol circulation.
The reagent tray can hold reagent cassettes, reagent racks and calibrator/control reagent racks.
■ The cassettes and the open cassettes occupy one position per cassette on the tray. The cassettes
are either single compartment or twin compartment containers. One cassette is dedicated to a
single solution: either a reagent, a diluent or a cleaner.
■ The reagent racks occupy four positions per rack on the tray. The reagent racks are cut into three
sectors (each of three positions) and can accomodate either three reagents, diluents or cleaners or
a combination of the above (for example: one reagent and two diluents) in reagent vials of 4 mL,
10 mL and 15 mL.
■ The calibrator reagent racks occupy one position per rack on the tray. These racks can
accomodate up to three different calibrators in sample cups (with adaptors).
■ The control reagent racks occupy one position per rack on the tray. These racks can accomodate
up to three different controls in sample cups (with adaptors).
Reagents are loaded/unloaded on the reagent tray by the reagent tray cover.
The colored circle around the reagent tray on the main menu indicates if the reagent tray
is accessible (green) or in progress (blue).
The instrument checks the solutions on board (barcodes reading) during the initialization
and when closing the reagent tray cover.
■ The cassettes are identified by a single barcode which allows the solutions automatic
configuration in Reagent Configuration menu. However, solutions in an open cassette must be
previously registered in Main menu > Services > Application Configuration > Reagents and
identified by barcode labels (PC400/P400 Open Cas. sticker sheet1 HAX0334 (1207230334) and
PC400/P400 Open Cas. sticker sheet2 HAX0335 (1207230335)) provided by HORIBA Medical.
■ The reagent racks are identified by a specific barcode which indicates the rack ID. The solutions
on the reagent racks are manually configured in Reagent Configuration menu.
■ The calibrator/control reagent racks are identified by a specific barcode which indicates the rack
ID and if the rack is dedicated to calibrators or controls.
Related information:
■ Consumables and Spare Parts, p.369
■ To Register a Reagent, p.299
The sample tray contains six positions, each position can hold a ten-sample rack for a total capacity
of 60 samples.
The sample tray can hold sample racks, sample cup racks and calibrator/control sample racks.
■ The sample racks have ten positions including five external positions and five internal positions.
The external positions can accomodate sample tubes of 7 and 10 mL (diameter: 16 mm) and the
internal positions sample tubes of 4 and 5 mL (diameter: 13 mm). Samples tubes of 4 and 5 mL
(diameter: 13 mm) can also be positioned on the external positions with adaptors.
■ The sample cup racks have ten positions and can accomodate only sample cups.
■ The calibrator/control sample racks have ten positions and can accomodate only sample cups.
■ Do not wash or re-use sample tubes or sample cups. They are designed for single use
only.
■ When filling sample tubes or sample cups, dispense the sample slowly in order to
avoid the formation of foam.
■ Sample tubes or sample cups must always be uncapped before being positioned on
the sample tray or on the reagent tray.
Dispose of used sample tubes or sample cups according to your local/national guidelines
for biohazard waste disposal.
The sample tray access allows continuous loading of patient samples, calibrators and controls.
■ Before loading racks on the sample tray, ensure that the sample tray LED is green. If
not, press Pause to stop samplings on the sample tray and wait for the LED to turn
green.
■ If automatic reruns must be performed, the instrument turns to "Sampling" status and
the LED turns to orange for a few seconds before turning red. Never load racks on the
sample tray when the LED is orange or red.
The sample tray is equipped with a tube/cup detector and with a barcode reader which allows sample
identification.
The sample barcode reader reads barcodes from two to 16 digits of the following types:
■ ITF 2/5 (2 of 5 interleaved) with or without check digit
■ Code 39 with or without check digit
■ Code 128
■ Codabar
The sample barcode reader frequency (550 Hz) allows the reading of high resolution
barcodes (higher than or equal to 0.16 mm). However, this frequency does not allow the
reading of higher resolution barcodes (lower than 0.16 mm).
In order to ensure security of patient identification and sample identification, the space
character is allowed in the patient ID and the sample ID only when this one is placed
inside the capture. The space character is not authorized and is systematically deleted
when this one is the first character or the last character of the capture.
This rule also applies to the host connection and the sample barcode types Code 39 and
Code 128.
Also refer to the Settings > System Configuration > Host Connection > To Configure the
Host Connection chapter.
■ The sample racks are identified by a specific barcode which indicates the rack number. For
samples identified by a barcode, the barcode allows to automatically associate a sample tube on
the tray with a request in the worklist.
■ The sample cup racks are identified by a specific barcode which indicates the rack number.
■ The calibrator/control sample racks are identified by a specific barcode which indicates the rack
number and whether the rack is dedicated for calibrators or controls.
1 = Barcode label
location
2 = 11.5 mm
3 = Reagent rack
Refer to the Maintenance and Troubleshooting > Maintenance > Consumables and Spare Parts
chapter for rack labels references.
6 = Sample rack
Refer to the Maintenance and Troubleshooting > Maintenance > Consumables and Spare Parts
chapter for rack labels references.
You should preferably use the sample rack labels from HAX0166 (1207230166) to
HAX0174 (1207230174). Nevertheless, if you use the 2 of 5 interleaved barcode type
without check digit, you must use the sample rack labels from HAX0273 (1207230273) to
HAX0281 (1207230281).
10 = Reagent stickers
11 = Control stickers
12 = Calibrator stickers
Related information:
■ Consumables and Spare Parts, p.369
■ To Register a Reagent, p.299
The sampling system includes a reagent syringe and a reagent needle for pipetting, preheating at
37°C +/- 0.5°C (99°F +/- 0.9°F) and dispensing reagents; a sample syringe and a sample needle for
pipetting and dispensing samples and reagents. Both needles are washed in their respective wash
towers between each sampling.
1 = Reagent syringe
2 = Sample syringe
3 = Sample needle
4 = Reagent needle
1.2.1. Syringes
Each syringe assembly consists of a glass barrel, a metal plunger fitted with a teflon tip, a valve and a
screw fitting to hold the syringe head.
Reagent syringe
This 1000 µL syringe controls the volume of reagents. It must be filled with water at all times and no
air bubbles should be visible in the syringe or tubing. This syringe performs samplings from 15 µL to
600 µL by steps of 0.1 µL.
Sample syringe
This 100 µL syringe controls the volume of sample, distilled water and additional reagents. It must be
filled with water at all times and no air bubbles should be visible in the syringe or tubing. This syringe
performs samplings from 2 µL to 95 µL by steps of 0.1 µL.
1.2.2. Needles
Needles are sharp and may contain biohazardous materials. When opening the main
cover, wait for the arms to stop before proceeding.
Reagent needle
The reagent needle is a stainless steel needle with an external teflon coating and an internal sleeving
made of a teflon capillary tube from one end to the other.
The reagent needle preheats reagents at 37°C +/- 0.5°C (99°F +/- 0.9°F).
The reagent arm is equipped with capacitive level detection and insufficient volume detection.
The reagent needle is washed between each sampling in its wash tower.
Sample needle
The sample needle is a stainless steel needle with an external and internal teflon coating.
The sample arm is equipped with shock detection, the sample needle does not pierce neither sample
tubes nor sample cups.
Sample tubes or sample cups must always be uncapped before being positioned on the
sample tray or on the reagent tray.
The sample arm is also equipped with capacitive level detection, clot detection and insufficient
volume detection.
The sample needle is washed between each sampling in its wash tower.
Definitions
All volumes are in micro-litres (µL).
FD is the dilution factor.
FPre-dil is the pre-dilution factor.
FPost-dil is the post-dilution factor.
FPost-conc is the post-concentration factor.
SV is the sample volume.
H2OV is the H2O volume pipetted by the sample needle to push the sample volume (programmed in
the application).
DV is the diluent volume.
The sample needle can aspirate from 2.0 to 95 µL.
When possible, 300 µL of diluted material is prepared when using a cuvette. For example: if FD = 10.0,
30 µL of sample is pipetted and mixed with 270 µL of diluent. The total volume is 300 µL.
In some cases, this is not possible due to the limitations of pipetting volumes. For example: if
FD = 2.0, 95 µL of sample is pipetted and mixed with 95 µL of diluent. The total volume is not 300 µL
but 190 µL. This exception is true when 2.0 ≤ FD < 3.1.
1 = SV
2 = DV
■ If FD > 150.0, then two cuvettes are used in series for dilution.
If two cuvettes are required, the first one is performed as above. For the second one, SV2 is
pipetted in the first cuvette and then mixed with DV2 in the second one as shown in the figure
below.
1 = SV1
2 = DV1
3 = SV2
4 = DV2
Dilution procedure
■ If 2.0 ≤ FD < 3.1, then one cuvette is used for dilution and:
SV = 95
DV = 95 * (FD - 1)
The total volume in the cuvette is then: 95 * FD
■ If 3.1 ≤ FD ≤ 150.0, then one cuvette is used for dilution and:
SV = 300 / FD
DV = SV * (FD - 1)
The total volume in the cuvette is then: 300 µL
■ If 150.0 < FD ≤ 22500.0, then two cuvettes are used in series for dilution and:
FDCuv = √ (FD)
SV = 300 / FDCuv
DV = SV * (FDCuv - 1)
The total volume in each cuvette is then: 300 µL
Cuvette segments
Cassettes of 15 cuvette segments are used to fill the "new cuvette" holder (Cuvette Segments
A11A01891). Each disposable cuvette segment consists of 12 identically moulded cuvettes. The total
volume in a cuvette must be included between 150 µL to 600 µL.
■ Care should be taken in the storage and handling of the cuvette segments in order to
prevent breakage, damage or contamination.
■ Cuvette segments should be left in the plastic wrapping until use to prevent dust
contamination.
■ Cuvette segments should be handled only by the upper part.
■ Do not wash and re-use cuvettes. They are designed for single use only.
■ Cuvette segments with at least one used cuvette must never be loaded in the reaction
tray.
Dispose of used cuvettes according to your local and/or national guidelines for biohazard
waste disposal.
The reaction tray contains six positions, each position can hold a 12-cuvette segment for a total
capacity of 72 cuvettes.
The reaction tray is located in a chamber of which the temperature is regulated at 37°C +/- 0.2°C
(99°F +/- 0.36°F) by air circulation. A temperature sensor controls the temperature within the reaction
tray chamber. If the reaction tray temperature is out of range, the warning #123 Reaction tray
temperature exceeds limits or the alarm #053 Reaction Tray Temperature regulation failed is triggered.
1.3.3. Mixer
1 = Mixer paddle
2 = Mixer wash tower
1.3.4. Spectrophotometer
The Pentra C400 spectrophotometer disposes of 15 wavelengths: 340, 380, 405, 420, 455, 490, 505,
520, 550, 560, 580, 600, 620, 660, 700 nm.
Principle
White light from the tungsten-halogen lamp is collected by the condensing lens.
Once reflected by the folding mirror and shaped by the second lens, the white light beam passes
through the cuvette where it travels through the reactive solution.
White light emerging from the cuvette is then coupled with the spectrograph entrance slit by the third
lens.
The concave reflective grating spreads the incoming white light into monochromatic radiations and
reflects them onto the photodiode array.
1 = Tungsten-halogen lamp
2 = Condensing lens
3 = Folding mirror
4 = Shaping lens
5 = Cuvette
6 = Coupling lens
7 = Spectrograph entrance slit
8 = Concave reflective grating
9 = Photodiode array
The ISE module (option) is designed to determine sodium, potassium and chloride concentrations in
serum, plasma or urine samples. It uses three Ion Selective Electrodes (ISE) for the measurement of
sodium, potassium and chloride concentrations and one reference electrode.
■ The sodium electrode consists of a glass membrane selective to Na+ ions.
■ The potassium electrode consists of a PVC crown ether membrane which is selective to K+ ions.
■ The chloride electrode is a plastic membrane electrode selective to Cl- ions.
■ The reference electrode consists of an open liquid junction.
The four electrodes are compact and located in a Faraday cage to minimize the influence of
electromagnetic fields. The electrodes are supplied ready to be used and no assembly is required.
To minimize environmental interferences, the ISE module cover must always be closed
during ISE module operation.
Principle
Sodium, potassium and chloride are measured on:
■ undiluted serum and plasma samples,
■ diluted urine samples.
For serum and plasma samples, 60 µL of sample is pipetted by the sample needle and 50 µL is
dispensed into the ISE module sample cup. For urine samples, 20 µL of sample is pipetted by the
sample needle and 10 µL is dispensed and diluted with 60 µL of low standard directly into the ISE
module sample cup.
The sample flows through the air sensor to reach the ISE block. The air detection by air sensor
indicates the end of sample transfer from the sample cup to the ISE block. The reference solution
flows through the reference electrode and merges with the sample downstream of the ISE block.
The electrical potential is measured between each ion selective electrode and the reference electrode.
This measured potential is function of the ion concentration.
Low standard
Reference
Waste
Syringe 2 Syringe 1
Valve 6
High standard
Low standard
Reference
Air sensor
Sample cup
Valve 5
Isolator
Air
Waste
The figure below illustrates the Pentra C400 analysis cycle giving a mono-reagent/sample cycle
example.
Time (sec.) 0 1 2 3 4 5 6 7 8 9 10 11 12
Reagent Washing
Reagent arm dispensing
Reagent
pipetting
Reagent syringe
Mixing
Mixer
Automatic loading or
unloading of a cuvette
segment, if necessary
Reaction tray
Absorbance
reading Reading of
T0 Reading cuvettes 1 to 72
Reagent tray
Tray motion
Sample tray
Tray motion and
sample identification
if necessary
2. Data Analysis
Absorbance
A = log [(I0 - IB) / (I - IB)]
Where:
A is the absorbance.
I is the intensity measured.
I0 is the reference intensity. The reference intensity is measured for each wavelength during the
instrument initialization, it corresponds to the lamp intensity measured through a water-filled cuvette.
IB is the black intensity. The black intensity is measured for each wavelength during the instrument
initialization, it corresponds to the residual intensity measured in the black (the lamp is masked).
Bichromatism
The instrument measures the absorbance at two wavelengths.
■ The primary wavelength (λ1) is used to measure the component absorbance.
■ The secondary wavelength (λ2) is used to measure an absorbance due to a cuvette defect, a
variation of lamp intensity or some sample interferences.
AMeas = Aλ1 - Aλ2
Where:
AMeas is the absorbance measured.
Aλ1 is the absorbance measured at the primary wavelength.
Aλ2 is the absorbance measured at the secondary wavelength.
Volume correction
Absorbance values are corrected for volume differences if more than one pipetting step is performed
in a cuvette.
AEnd = AMeas * (VMeas / VEnd)
Where:
AMeas is the absorbance measured.
AEnd is the absorbance based on final total volume in the cuvette.
VMeas is the volume at time of measurement.
Checks
The three checks below are performed after the volume correction:
■ HIGH_ABS
■ SAMPLE_LIMIT
■ REAG_RANGE_HIGH / LOW
2.1.2.1. Endpoint
Two absorbance readings are used to determine the reaction rate (ΔA).
ALast
ΔA
AFirst
Time
T0 Tn
A = ALast - AFirst
Where:
A is the absorbance.
AFirst is the first absorbance reading.
ALast is the last absorbance reading.
The checks below are performed for Endpoint calculation type:
■ DIV_ABSi
■ DIV_ABSf
2.1.2.2. Kinetic
Linear regression analysis is performed on all the measurement points included between the first and
the last absorbance readings. The slope of the regression line is the reaction rate (ΔA/min).
ALast
ΔA/min
AFirst
Time
TFirst TLast
2.1.2.3. Kinsearch
For the determination of enzymatic activities, a linear search is performed on each cuvette to find the
linear range. Linear regression is then done on all the measurement points within this linear range. The
slope of the regression line is the reaction rate (ΔA/min).
If the change in absorbance between the first and the last absorbance readings is less
than 0.002 absorbance units, a linear regression (Kinetic calculation type) will be
performed instead of a linear search (Kinsearch calculation type) to determine the ΔA/min.
Special program
To find the end of the linear range, the consecutive absorbance changes are compared until the
following absorbance change becomes greater than or equal to the previous one.
Δa > Δb
Δb > Δc
Δc ≤ Δd
The measurement point used in both intervals is established as point I.
An auxiliary point end (APE) is set 12 seconds prior to I in time and at the same absorbance as I.
To find the end of the linear range, straight lines are drawn from the point APE to each absorbance
point, in sequence starting with An. The tangents of the angles formed by each of these lines are
sequentially compared.
Tangent of angle = ΔAbsorbance / ΔTime
Example: tn (αn) = (An - A3) / (Tn - T3 + 12 sec)
tn (αn) < tn (αn-1)
tn (αn-1) < tn (αn-2)
tn (αn-2) < tn (αn-3)
tn (αn-3) < tn (αn-4)
tn (αn-4) ≥ tn (αn-5)
When the tangent of the compared angle becomes less than or equal to the previous tangent, the
point used to draw the associated line is defined to be the end (E) of the linear range.
A0
Δa
A1
Δb
A2
Δc APE I
A3
Δd
A4
αn-4 αn-2 αn
αn-5 αn-3 αn-1
E
12 sec
An
Time
T0 T1 T2 T3 T4 ... Tn-4 Tn-3 Tn-2 Tn-1 Tn
To find the beginning of the linear range, an auxiliary point beginning (APB) is set 12 seconds prior to
E in time and at the same absorbance as E.
Straight lines are drawn from the point APB to each absorbance point, in sequence starting with A0.
The tangents of the angles formed by each of these lines are sequentially compared.
Tangent of the angle = ΔAbsorbance / ΔTime
Example: tn (α0) = (An-5 - A0) / (Tn-5 - T0 - 12 sec)
tn (α0) > tn (α1)
tn (α1) ≤ tn (α2)
When the tangent of the compared angle becomes greater than or equal to the previous tangent, the
point used to draw the associated line is defined to be the beginning (B) of the linear range.
A0
A1 B
A2
α0
α1
α2
E
APB
12 sec
An
Time
T0 T1 T2 T3 T4 ... Tn-4 Tn-3 Tn-2 Tn-1 Tn
Normal program
To find the end of the linear range, an auxiliary point end (APE) is set 12 seconds prior to A0 in time
and at the same absorbance as A0.
Straight lines are drawn from the point APE to each absorbance point, in sequence starting with An.
The tangents of the angles formed by each of these lines are sequentially compared.
Tangent of the angle = ΔAbsorbance / ΔTime
Example: tn (αn) = (An - A0) / (Tn - T0 + 12 sec)
tn (αn) < tn (αn-1)
tn (αn-1) < tn (αn-2)
tn (αn-2) ≥ tn (αn-3)
When the tangent of the compared angle becomes less than or equal to the previous tangent, the
point used to draw the associated line is defined to be the end (E) of the linear range.
A
An
αn
αn-1
αn-2
αn-3
APE
A0
12 sec
Time
T0 T1 T2 T3 T4 ... Tn-4 Tn-3 Tn-2 Tn-1 Tn
To find the beginning of the linear range, an auxiliary point beginning (APB) is set 12 seconds after An
in time and at the same absorbance as An.
Straight lines are drawn from the point APB to each absorbance point, in sequence starting with A0.
The tangents of the angles formed by each of these lines are sequentially compared.
Tangent of the angle = ΔAbsorbance / ΔTime
Example: tn (α0) = (An - A0) / (Tn - T0 + 12 sec)
tn (α0) < tn (α1)
tn (α1) < tn (α2)
tn (α2) < tn (α3)
tn (α3) ≥ tn (α4)
When the tangent of the compared angle becomes less than or equal to the previous tangent, the
point used to draw the associated line is defined to be the beginning (B) of the linear range.
APB
An
E
α1 α2 α3 α4 α5
12 sec
A0
Time
T0 T1 T2 T3 T4 ... Tn-4 Tn-3 Tn-2 Tn-1 Tn
The absorbance deviation checks are performed during the reaction rate calculation in order to flag
optical density jumps.
DIV_ABSi
This check is only available for the Endpoint calculation type.
The goal of this check is to verify that the absorbance difference between the first point and a
reference point is included in a defined range.
Two different algorithms can be used for this check: one if the first reading cycle < 4, another if the
first reading cycle ≥ 4.
■ First reading cycle < 4
Reagent blank
A Normal run
An of reagent blank
blank range high
blank range low
A0 of run
T0 Tn Time
The difference |An of reagent blank - A0 of run| is calculated and checked to ensure that it is
included in the configured blank range. If not, the analytical flag DIV_ABSi is triggered.
Where:
A0 is the absorbance value at the first reading cycle.
An is the absorbance value at the last reading cycle.
■ First reading cycle ≥ 4
An
A0-1
A0
D1
T0 Tn Time
DIV_ABSf
This check is only available for the Endpoint calculation type.
The goal of this check is to verify that the absorbance difference between the last point and the
previous one is included in a calculated range.
A
D1
An-6 An-4 An-2 An
This algorithm applies only when the last reading cycle is ≥ 11.
DIV_ABS
This check is only available for Kinetic and Kinsearch calculation types.
The goal of this check is to calculate the correlation coefficient r2 of the reaction and to flag any
reaction with a r2 lower than 0.80.
An
r2
T0 Tn Time
The correlation coefficient r2 and the standard deviation SD are calculated as follows:
■ Correlation coefficient r2:
r = (b * ∑XY - ∑X * ∑Y) / √ {[b * ∑X2 - (∑X)2] * [b * ∑Y2 - (∑Y)2]}
Where:
r2 is the correlation coefficient.
b is the number of measurement points used for the linear regression.
X is the absorbance values of the used measurement points.
Y is the times in seconds of the used measurement points.
∑X is the sum of absorbance values of the used measurement points.
∑Y is the sum of times in seconds of the used measurement points.
∑X2 is the sum of squares of absorbance values of the used measurement points.
∑Y2 is the sum of squares of times in seconds of the used measurement points.
∑XY is the sum of absorbance values multiplied by times in seconds of the used measurement
points.
■ Standard deviation SD:
SD = √ {[b * ∑X2 - (∑X)2] / [b * (b - 1)]}
Where:
SD is the standard deviation.
b is the number of measurement points used for the linear regression.
X is the absorbance values of the used measurement points.
∑X is the sum of absorbance values of the used measurement points.
∑X2 is the sum of squares of absorbance values of the used measurement points.
■ For the Kinetic calculation type, the measurement points used to calculate r2 and SD
are those included between the first reading cycle and the last reading cycle.
■ For the Kinsearch calculation type, the measurement points used to calculate r2 and
SD are those included in the linear range found within the linear search program.
If r2 is lower than the configured r2 threshold (set to 0.80 by default) and SD is higher than the
configured SD threshold, then the analytical flag DIV_ABS is triggered.
2.1.3. Rates
The reagent blank absorbance change is stored in memory until another calibration is
performed.
Sign correction
If the reaction direction is programmed as decreasing in the application, the calculated rate is
multiplied by (- 1) to make the rate a positive number.
This correction applies to rates of calibrator, control and patient sample and occurs before the
reaction direction check.
R' = R * (- 1)
Where:
R is the rate of calibrator, control or patient sample.
R' is the corrected rate of calibrator, control or patient sample.
Checks
The three checks below are performed after rate corrections:
■ BLK_RANGE
■ SIGN
■ HIGH_ACTIV
Calibration modes
The calibration modes used on the Pentra C400 are described in the chapters below:
■ Factor
■ Slope average
■ Linear regression
■ Linear interpolation
■ Multiparameter mathematical algorithms (LOGIT/LOG4, LOGIT/LOG5 or EXPONENT5)
Checks
The checks below are performed on calibrations:
■ CALCHK_FLAG
■ CALCHK_ERROR
■ CALC_ERROR
■ CAL_NOT_OPTIMISED
■ FACTOR_FLAG
Related information:
■ Factor, p.478
■ Slope Average, p.479
■ Linear Regression, p.480
■ Linear Interpolation, p.480
■ Multiparameter Mathematical Algorithms (LOGIT/LOG4, LOGIT/LOG5 or EXPONENT5), p.481
2.1.4.1. Factor
A theoretical factor is entered in the application to multiply the calculated rate to obtain test results.
■ For enzyme activities:
U/L = ΔA/min * F
Where:
F is the theoritical factor.
ΔA/min is the rate.
U/L is the result.
■ For concentrations:
C = ΔA * F
Where:
F is the theoretical factor.
ΔA is the rate.
C is the result.
This calibration mode is used for assays when the observed change in absorbance and the
concentration of the analyte have a linear relationship.
One to three standards can be programmed to derive the factor required in order to calculate test
results.
The example below illustrates the use of three standards:
Rate
RStd-3
RS
RStd-2
RStd-1
Conc.
CStd-1 CStd-2 CS CStd-3
For linear calibration curves, two to eight standards can be programmed to derive the curve
parameters required in order to calculate test results.
Linear regression is performed through all data points to find the most corrected line.
Rate
Individual
data point
R0
Conc.
R = a * C + R0
C = (R - R0) / a
Where:
R is the rate of control or patient sample.
a is the slope of the regression line.
R0 is the intercept of the regression line.
C is the concentration of control or patient sample.
For non-linear calibration curves, three to eight standards can be programmed to derive the curve
parameters required in order to calculate test results.
Linear interpolation amounts to draw a straight line from standard to standard.
Rate
Std-3
Std-2
RS
Std-1
Conc.
CS
These calibration modes are used for assays when the observed change in absorbance and the
concentration of the analyte does not follow a linear relationship.
Where:
R0 is the predicted rate for a standard with zero concentration.
R is the rate of control or patient sample.
KC is the scale parameter for LOGIT/LOG4 and LOGIT/LOG5 (= difference between the predicted rate
for a standard with infinite concentration and R0).
K is the scale parameter for EXPONENT5.
a, b, c are various parameters which define the non-linear elements of each mathematical function.
C is the concentration of control or patient sample.
2.1.5. Results
Concentration calculation
The conversion of rates to concentrations is based on a function which depends on the calibration
mode programmed. The function parameters are derived from the standard rates and concentrations
except for the factor calibration mode.
For detailed information concerning the calibration modes, refer to the Absorbance
Measurements > Calibration Modes chapter.
Result correlation
A correlation factor and an offset can be entered in the application to convert results in order to
correlate them with an alternative method or another temperature. The correlation factor and the
offset correspond respectively to the slope (A) and the intercept (B) of the regression line y = A * x + B
determined by a correlation study between the two methods.
y=A*x+B
Where:
A is the correlation factor or slope (A) of the regression line.
B is the offset or intercept (B) of the regression line.
x is the concentration of calibrator, control or patient sample.
y is the correlated concentration of calibrator, control or patient sample.
Two result correlations are programmable in the application: one dedicated to the user, another
dedicated to HORIBA Medical.
Checks
The checks below are performed on results:
■ DEVIATION
■ RESULT_CALC_ERR
■ LINEARITY_HIGH / LOW
■ CALC_RANGE_HIGH / LOW
■ TEST_RANGE_HIGH / LOW
■ AG_EXCESS
■ CONF_RANGE_HIGH / LOW
CONF_RANGE_HIGH_W / LOW_W
CONF_RANGE_HIGH_W1-6 / LOW_W1-6
WESTGARD_RULE_4-6
■ REF_RANGE_HIGH / LOW
■ CRITICAL_RANGE_HIGH / LOW
Related information:
■ Calibration Modes, p.478
2.2.1. Principle
The measurement principle of the ISE module is based on the interaction between moveable free ions
in a sample solution and an active sensing unit (ion selective sensing electrode).
An ion selective membrane separates the sample solution, where the electrolyte concentration is
unknown, from the electrode electrolyte, where the concentration is known.
1 = Reference electrode
2 = Selective electrode (for the ion measured)
3 = Electrolyte (known concentration)
4 = Voltage measurement
5 = Selective membrane
In order to measure this potential change, a reference electrode with a fixed potential is immersed in
the same solution. The potential of the reference electrode is kept fixed and constant, by filling it with
an electrolyte that has a constant concentration.
The electrical potential is measured between each ion selective electrode and the reference electrode.
This measured potential is function of the ion concentration.
The potential of an ion selective electrode depends on the logarithm of the activity of this ion. This
relationship is described by the Nernst equation:
E = E0 + S * log ai
Where:
E is the measured potential between ion selective electrode and reference electrode.
E0 is the standard potential of electrode assembly.
S is the electrode slope (Nernst factor).
ai is the activity of measured ion.
S = (2.303 * R * T) / (n * F)
Where:
S is the electrode slope (Nernst factor).
2.303 is the conversion factor for ln to log.
R is the gas constant.
T is the absolute temperature (°K).
n is the Valency of the ion.
F is the Faraday constant.
If the ionic strength of the calibration solution is equal to that of the sample, the electrode system can
be calibrated with accuracy using concentration instead of activity.
E = E0 + S * log C
Where:
E is the measured potential between ion selective electrode and reference electrode.
E0 is the standard potential of electrode assembly.
S is the electrode slope (Nernst factor).
C is the concentration of measured ion.
Relative measurements are performed to eliminate temperature influence.
2.2.2. Calibration
The slope of the measuring system is defined as being the variation in voltage obtained when the ion
concentration is multiplied by 10. It is determined by measuring two different standard solutions.
Taking into account the logarithm relationship between concentration and voltage in the Nernst
equation, the slope is calculated as follows.
S = (EStd-1 - EStd-2) / log (CStd-1 / CStd-2)
Where:
EStd-1, EStd-2 is the measured potential of Standard 1 or Standard 2.
CStd-1, CStd-2 is the known concentration of Standard 1 or Standard 2.
S is the slope.
The slope is checked to ensure that it is included in the slope normal range.
2.2.3. Results
Concentration calculation
The unknown concentration is calculated by comparing the measured potential with the potential of
the standard solution which has a known concentration.
C = CStd-1 * 10^[(E - EStd-1) / S]
Where:
EStd-1 is the measured potential of Standard 1.
E is the measured potential of control or patient sample.
CStd-1 is the known concentration of Standard 1.
C is the concentration of control or patient sample.
S is the slope.
Result correlation
A correlation factor and an offset can be entered in the application to convert results in order to
correlate them with an alternative method or another temperature. The correlation factor and the
offset correspond respectively to the slope (A) and the intercept (B) of the regression line y = A * x + B
determined by a correlation study between the two methods.
y=A*x+B
Where:
A is the correlation factor or slope (A) of the regression line.
B is the offset or intercept (B) of the regression line.
x is the concentration of calibrator, control or patient sample.
y is the correlated concentration of calibrator, control or patient sample.
Checks
The checks below are performed after the unit conversion:
■ LINEARITY_HIGH / LOW
■ CONF_RANGE_HIGH / LOW
CONF_RANGE_HIGH_W / LOW_W
CONF_RANGE_HIGH_W1-6 / LOW_W1-6
WESTGARD_RULE_4-6
■ REF_RANGE_HIGH / LOW
■ CRITICAL_RANGE_HIGH / LOW