Performance characteristics
Measuring range
“The test has been developed to determine LDL-Cholesterol
Concentration until 400 mg/6L, When values exceed this
ange samples should be diuted 144 with NaCl-Solution
(8 g/L) and the result multiplied by 5,
Specificity /Interferences
No interference was observed by bilirubin up to 30 mg/dl.
‘and hemoglobin up to 800 mg/dL.
Limit of detection
‘The lower limit of detection Is 2 ma/Ah.
Precision
intra-assay | Mean 35 wr
n= 20, Ima/dt)_| _tmgydty [1
[Sample 20, 0.81 a
| Sample 2 57 247 43
Sample 3 aa 139 10
Trter-assay | Mean 30 VK
n= 10 [ma/st) | tma/suy 6)
‘sample T 62 1.90 3.0)
Sample 2 Bt 280 21
Sample 3 283 2.09 o7
Method comparison
'A comparison of LDL-Cholesterol values obtained with the
DiaSys LDL precipitant (y) with the Friedewald calculation
(x) using 49 samples gave following results:
y= 1121 x~ 9.62 mg/dl; r= 0.947
Reference range [4]
LDL cholestero!
Desirable £130 mg/dL (3.4 mmol/L)
Borderline high risk 130 -160 mg/di. (3.4 ~ 4.1 mmol/L)
High risk > 160 mg/dl. (> 4.1 mmol/L)
Each laboratory should check if the reference ranges are
transferable to its own patient population and determine
own reference ranges if necessary.
Clinical Interpretation (2)
Epidemiological studies have observed that low HOL-
Cholesterol concentrations < 39 mg/dL (0.9 mmol/L) in
men and < 43 mg/dl. (1.0 mmol/L) in women) especially if
associated with fasting triglycerides > 180 mg/dl.
(2 mmol/L), predict @ high risk of coronary heart disease
The European Task Force on Coronary Prevention
recommends to lower TC concentration to less than
$90 mg/dl (5.0 mmol/L) and LDL-cholesterol to less than
115 mg/at (3.0 mmol/L}.
LDL Preciptant- Page 2
Literature
1
Rifai N, Bachorik PS, Albers 3- Lipids, lipoproteins and
Spolipoproteins. In: Burtis CA, Ashwood ER, editors
Tletz Textbook "of Clinical’ Chemistry. 3 ed.
Philadelphia: W.8 Saunders Company; 1999. p. 809-
61,
Recommendation of the Second Joint Task Force of
European and other Societies on Coronary Prevention
Prevention of coronary heart disease in clinical
practice, Gur Heart } 1998;19: 1434-503.
Copes-Virela MF, Stone Py Elis S, Colwell 3A.
Cholesterol determination in high-density lipoproteins
‘separated by. three. different methods. Clin Chem
1977;23,982-4.
Schaefer £), McNamara 3. Overview of the diagnosis
tnd treatment of lipid disorders. In: Rifai N, Warnick
GR, Dominezak MH, eds. Handbook of lipoprotein
testing. Washington: RACC Press;1997.p.25-48.
Guder WG, Zawta B et al, The Quality of Diagnostic
Samples. 1" ed. Darmstadt: GIT Verlag; 200.
e223.
Manufacturer
Diasys Diagnostic Systems GmbH
Alte Strasse 9 65558 Holzhelm Germany
1844 4330 10 02 00 sanwary 2009/2