Download as pdf or txt
Download as pdf or txt
You are on page 1of 20

MedizinTechnik

... for a better life

English

ATMOS C 451
Surgical Suction Unit
Service Instructions

340.0300.B
IO.02
Table of contents

Page
1.0 General information
1.1 Valid operating instructions ...................................3
1.2 Information on manufacturer .................................3
1.3 Exclusion of liability ...............................................3

2.0 Unit function


2.1 Intended use .........................................................4
2.2 Function ................................................................5
2.3 Installation and operation ......................................5

3.0 Cleaning and care


3.1 General information on cleaning
and disinfection .....................................................6
3.2 Cleaning and sterilizing hoses and
collection jar ..........................................................6
3.3 Cleaning and sterilizing the unit surface ...............7
3.4 Recommended instrument disinfectants ...............7
3.5 Recommended surface disinfectants ....................7

4.0 Maintenance
4.1 General information ..............................................8
4.2 Safety information .................................................8
4.3 Changing fuse .......................................................8
4.4 Disassembly and assembly of the housing
resp. the pump module .........................................8
4.5 Demounting and reprocessing the pump ..............9
4.6 Exploded view .....................................................10
4.7 Changing the bellows..........................................10
4.8 Changing hose system of
the DDS-Switchover-Docking Station .................10

5.0 Inspection specifications


5.1 Visual inspections ...............................................11
5.2 Electrical inspection ............................................11
5.3 Function check ....................................................11
5.4 Technical Specifications ......................................12

6.0 Trouble-shooting list ........................................13

7.0 Spare parts


7.1 Spare parts .........................................................14
7.2 Spare parts for the pump ....................................14

8.0 Plans ..................................................................15

2
1.0 General information

1.1 Valid operating instructions 1.3 Exclusion of liability

 The operating instructions relating to the ATMOS C 451 ATMOS assumes responsibility for the unit, as far as safety,
(REF 340.0300.B) form part of these service instructions. dependability and functionality are concerned, only if assembly,
The persons authorized for service work must be familiar readjustment, modifications, extensions or repairs are carried
with the contents, the kind of installation, the start-up as out by ATMOS or by persons authorized by ATMOS.
well as the operation described therein.
Safety for the user and a trouble-free operation are only gua-
ranteed when using original ATMOS parts. ATMOS cannot
guarantee safe operation and function of the ATMOS C 451 if
third-party accessories or consumables are used.

1.2 Information on manufacturer

Further information, accessories and spare parts are available


from:
 Very important notes

ATMOS
MedizinTechnik GmbH & Co. KG
Ludwig-Kegel-Straße 16
79853 Lenzkirch
Deutschland/Germany

Tel. + 49 (0) 76 53 / 689-0


Fax: + 49 (0) 76 53 / 689-190
+ 49 (0) 76 53 / 689-292 (Service Center)

atmos@atmosmed.de
www.atmosmed.de

3
2.0 Unit function

2.1 Intended use


The surgical suction device ATMOS C 451 is a suction unit for
medical application. It is especially intended for suction and coll-
ection of secretions, body fluids, rinsing liquids and tissue during
surgical interventions. The main fields of application are:

– In the OPD, in the OR: for sucking off and collecting e.g.
drain pockets, abscesses, body and rinsing solutions
and during lipectomy.

– In endoscopy: e.g. to aspirate secretions or rinsing solutions


as well as for operative fixation.

– In gynaecology: for suction curettage and for vacuum


extraction.

– In ENT applications: to aspirate secretions, rinsing solu-


tions, cerumen or to extract foreign matters.

– In the ward, recovery ward and ICU: : for the spontaneous


aspiration of body fluids and foreign matters, e.g. from the
respiratory tract.

The ATMOS C 451 must NOT be used:

– in non-medical applications;
– to draw off combustible or explosive fluids or gases;
– for drainage in the low vacuum range (e.g. thorax drainage);
– for smoke evacuation in connection with HF-electrosurgery
or laser surgery.

4
2.0 Unit function

2.2 Function
 The ATMOS C 451 is a mains-operated surgical suction
unit, centering around a silent diaphragm-type pump which
generates a vacuum inside the hose and the container
system, allowing secretions to be withdrawn and collected.
Using a vacuum regulator and the vacuum-gauge, the
target vacuum and thus the air-flow rate can be precisely
adjusted.

 Several collection jars of different sizes are available for use


with the system (section 7.0 Spare parts). A hydrophobic
bacterial filter in the lid of the collection jar is implemented
to prevent that secretions enter the pump resp. bacteria
the interior of the unit.

 All parts of the collection jar system and the silicone ho-
ses (with the exception of the bacterial filter) that come
into contact with secretions may be autoclaved (up to
136°C).

 Trolleys with several accessories are available for mobile


use.

2.3 Installation and operation


 On this, please read the respective chapters in the valid
operating instructions.

5
3.0 Cleaning and care

3.1 General information on cleaning and disinfection

Prior to cleaning:

Medical devices like the ATMOS C 451 must always offer a maximum in safety and function.
We therefore recommend prior to each application:

if

necessary

3.2 Cleaning and sterilizing hoses and


collection jar
 The measures described herein regarding cleaning  Before using the device on a new patient be sure
and disinfection do not replace the regulations to clean and sterilize the following parts:
valid for operating the device! – suction hose including suction attachment or aspi-
ration set,
 The information regarding concentration of – DDS collection jar including DDS collection jar
cleaning liquids and disinfectant solutions stated by lid, DDS hose adapter, splash protector and DDS
the relevant manufacturers must strictly be
collection jar handle.
observed!

 Do NOT use  Unscrew all hose connectors, pull the DDS hose adap-
ter out of the DDS collection jar lid, open the lid, empty
• disinfectant solutions with organic or anorganic the collection jar and dispose of the suctioned material
acids and bases as these may cause corrosion. properly.
• disinfectant solutions with chloramides, phenol
derivates or anionic tensides as these may cause  Take the DDS bacterial filter out of the DDS collection jar
stress cracks with plastics. handle.

 All other parts, except the bacterial filter, must also


be thoroughly rinsed under running water. If you wish
you may add a detergent, which does not contain the
following ingredients: Organic solvents, fat, alcohol or
amines, acetate, acids, and alkaline watery solutions.
Cleaning in a dishwasher is also possible.

 Autoclave all the above mentioned parts


(up to 136 °C, 3 bar, 20 min max.).

 After sterilization, reassemble all parts (section 4.0


"Operation").

6
3.0 Cleaning and care
3.3 Cleaning and sterilizing the unit
surface

 Always disconnect the device from the power line,


before cleaning and disinfecting the surface.

 Wipe the surface clean with a cloth soaked in a cleaning


solution or disinfectant. Liquids must not enter the device.
All of the cleaning solutions and disinfectants listed below
can be used.

 Should liquids have penetrated into the device, it


must be inspected by an authorized service partner.
If secretion has penetrated the interior, all air-conducting
parts must be replaced and the interior must thoroughly
be cleaned and disinfected.
Please contact your ATMOS service partner.

3.4 Recommended disinfectants for instruments

Disinfectant Contents (in 100g) Manufacturer


Succindialdehyde
Dimethoxytetrahydrofurane
GIGASEPT FF 11,0 g
Corrosion inhibitors Schülke & Mayr, Norderstedt
(Application concentrate) 3,0 g
Non-ionic tensides and fra-
grances
Bis(3-aminopropyl)laurylamine 8,0 %
Alkyl dimethylbenzyl ammonium 19,0 %
Mucozit-T
chloride Merz & Co., Frankfurt/Main
(Application concentrate)
Cocospropylendiamin-1,5- 7,0 %
guanidiniumacetat

3.5 Recommended disinfectants for surfaces


For cleaning the surface the following disinfectant solutions were tested:

Disinfectant Contents (in 100g) Manufacturer


TERRALIN Benzalkonium chloride 20,0 g
Schülke & Mayr, Norderstedt
(Application concentrate) Phenoxypropanoles 35,0 g
Didecyl dimethyl
14,0 g
ammonium chloride
10,0 g
QUATOHEX Benzalkonium chloride
07,5 g Braun, Melsungen
(Application concentrate) Bi-guanidinium acetate
00,5 g
Polymer biguanide
Active cleaning substances
Glucoprotamine 26,00 g
Incidin Plus
Non-ionic tensides Ecolab, Düsseldorf
(Application concentrate)
Solvents, complexing agents
Pursept-A Ethanol 38,90 g
(Disinfectant spray or Glyoxal 00,10 g Merz & Co., Frankfurt
disinfectant cloths) QAV 00,05 g

■ Discolouration may result, if disinfectants containing aldehydes and amines are used on the same object.

7
4.0 Maintenance
4.1 General information
 Prior to each application, effect a visual check of unit incl. – The contact of liquid and other residues from the suction
hoses, collection jar and power cable. system with skin or clothes must be avoided.

 For hygienic reasons, the DDS bacterial filter / oversuction – During work it is most important not to hurt yourself
stop must be changed at least once a day! - DANGER OF INFECTION!

 Other regular maintenance work is not necessary. – Dismantled, contaminated parts which were replaced require
waterproof and airtight packing and must then be disposed
of professionally.
4.2 Safety information
– After you have completed your work you must thoroughly
(also see page 3) clean and disinfect the tools which you used. Please observe
the contact time prescribed by the manufacturer!
Opening the ATMOS C 451 as well as maintenance and re-
pair work is only to be carried out by persons authorized by – Furthermore please thoroughly clean and disinfect the
the manufacturer. In doing so, pay attention to technical and working area with surface disinfectant. Please observe the
hygienic precautionary measures. contact time prescribed by the manufacturer!

– At the end you must thoroughly clean and disinfect your


When mounting the shells of the unit pay attention not to hands.
crack or jam the cables. Mounting of the shells must be able
without effort. – For service technicians we recommend all required active
immunisations.
The manufacturer cannot be held liable for any damages re-
sulting from improper maintenance or repair work or from the 4.3 Change fuse
use of others than the original spare parts. If the unit is opened
during warranty period, repair or maintenance work will no  Remove mains cable.
longer be carried out under the terms of warranty.  Press the spring clips of the fuse holder together on
both sides with a small screwdriver and pull out the fuse
If the unit can no longer be used, disposal is to be effected acc. holder.
to local regulations. Components and replacement parts are  Replace the fuse and push the holder back in until both
to be given into the respective recycling containers. spring clips are locked into place.
 Then reconnect mains cable.
 When opening used units and effecting repair
work on hose and pump parts, it must always
be assumed that the parts are contaminated.
4.4 Disassembly and assembly
Avoid direct contact, protect yourself by using of the housing resp. the pump module
disposable gloves and pay attention to
a) Remove the power cable on the bottom of the unit, remove
hygienic precautionary measures! the 2 bottom screws and the 2 screws on the handle and
disperse the two shells.
 After having assembled the suction unit, check
its function acc. to the test instructions in b) Loosen the plug connectors and the hose connections on
section 5 ! the forward shell.

Prior to any service work and opening of the c) Remove the pump incl. support.
 device, please switch of the ATMOS C 451 and d) Remove corresponding hose connections and electrical
remove mains cable. connections.
 Hygiene measures during repair work
e) Perform maintenance work.
– Depending on the previous application the suction f) Assembly is done vice versa to a) - d); add again all cable
system and the waste water drainage system may and hose fixations (cable ties).
be highly contaminated.
 The adhesion of the connecting nipples on
– For all work performed on the suction system, the silencer serves only as fixation but not
gloves must be worn. The single-use gloves must be as absolute sealing!
changed prior to touching other things respectively parts
which are not contaminated.  Do not loose distance plates between diaphragm
and piston, if present!
– Please take care that no liquid or other residues
escape from the suction system respectively the Never remove the connecting rod from the motor
water drainage system. Any leaked fluids from the  spindle as this may result in vibrations or
suction system must be removed immediately with bad running !
a cloth and a disinfecting cleaning solution. After
wards thoroughly wipe the area with a suitable  Attention, supply voltage! Do not touch the pump!
surface disinfectant. Please observe the contact time
prescribed by the manufacturer!
8
4.0 Maintenance

4.5 Demounting and reprocessing


the pump
(see 4.6 Exploded view of pump and Fig 1. Pump)

The pump is maintenance-free, when the device is used in Never remove the connecting rod from the motor! This may
line with the operating instructions. There is no preventive result in unbalance which would deteriorate start-up of the
maintenance required. Nevertheless, the pump should be re- pump and the noise level.
processed (recommended every 3 years) at regular intervals
depending on its use. This guarantees operational reliability In case secretion penetrated the crank case due to a perfo-
and hygienic safety. rated diaphragm, a repair/reprocessing of the pump makes
no sense.
However, if the pump was oversucked, it must be reprocessed
immediately by the ATMOS service. Further operation of the In case the pump was not oversucked, the pump heads
pump may be dangerous! (15, 16), the pump chambers (12) and the crank case (8) may
be cleaned in an ultrasonic bath, the openings blown out with
Demounting the pump compressed air. Then the parts must be disinfected with a
recommended disinfectant.
– Remove the electrical connections.
– Loosen the 4 fixation screws from the pump. In case a pump was oversucked, the pump heads (15, 16)
must be exchanged.
Opening the pump heads
Mounting the crank case
– Remove the hose connection (27) with check valve (28)
from the pump heads.
– Loosen the 4 fixation screws (6) from the pump heads.
– Remove the pump heads.
– Twist off the pump head (15) from the motor
condenser (2).
– Separate the pump chamber (12) from the pump head (15)
by loosening the 2 screws (18).
– Remove all gaskets (13, 14) and rubber parts. A guide guarantees that the crank case can only be connected
to the motor in one position.
Removing the diaphragms
– Mounting the crank case. Fixing the new diaphragms. The
– The diaphragms (23) are connected to the connecting rod threads must be secured with thread locker. The diaphragms
(19) by a threaded fitting which is cast. For twisting off the should preferably be fixed with the special wrench.
diaphragms, a special wrench is required.
– Lift the rubber part of the diaphragm and attach the wrench Mounting the pump heads
to the threaded fitting.
– It is also possible to loosen the diaphragms with a gripper, – Insert the new rubber seals (14) into the pump chamber
but then they will be destroyed. (12).
– It is most important not to attach the wrench to the connec – Connect the pump chambers (12) with the pump heads
ting rod! (15, 16). Both parts can only be assembled in one
– Loosen the diaphragms. position.

Final assembly of the pump

– Unscrew the motor condenser (2) on the pump head without


hose connections. The threaded fitting can be removed from
the condenser to prevent the condenser from being
destroyed.
– Insert new rubber seals (13).
– Mounting the pump heads. Fix the screws cross-wise. The
correct position of the pump heads is guaranteed by special
guides.
Demounting the crank case – Connect the pump heads to new hoses (27) and new check
valves (28). Make sure that the arrows on the check valves
It must be observed that the motor windings are not point away from the pump head with the hose con-
damaged! nections.
– Loosen the 4 screws (9) from the crank case. – Attach the insulating foil and the rubber feet.
– Remove the crank case (8).
– Remove the rubber feet and the insulation foil.

9
4.0 Maintenance
4.6 Exploded view of pump
1 Motor
2 Condenser 069.0200.0
3 Tooth lock washer
4 Accessories for wiring
5 Flange
6 Screws M 4x60
7 Washers
8 Console, parallel
9 Screws M 4x60
10 Ball bearing
11 Silicone ring
12 Chamber
13 Seal *
14 Seal *
15 Head with connections
16 Head with connections
A-G1/4 B-G 1/4
17 Spring washers
18 Screws M 4x10
19 Connecting rod
20 Excentric 3.0
21 Headless pin
22 Distance plate
23 Diaphragms *
24 Fan
25 Lock washer
26 Grommet
27 Hoses
28 Check valves *
29 Washers
Fig 1. Pump
30 Hose connection
31 Female plug/pin 4.7 Changing the bellows
32 Label
33 Protective cap The bellows can be added to the vacuum attachment by turning
it with the thumb.
* included in spare parts kit REF 069.0201.0

4.8 Changing the hose system of the


DDS-Changeover-Docking Station
a) Remove the hose end sticking out of the DDS-Changeover-
Docking Station.

b) Remove the two bellows.

c) Remove the operating lever and press out the excenter


with the two rolls downwards.

d) Remove complete hose harness.

e) Assembly is done vice versa from d) to a).

10
5.0 Inspection specifications

5.1 Visual inspection

After assembly all mechanical parts must be visually inspected 5.2 Electrical inspection
as to their proper fastening. Moving parts must be inspected
for their positive allowance. Bearing parts must be visually acc. to DIN VDE 0751-1:
inspected for any damages.
– Measurement of spare leakage current of appliance
The following items should be observed, if applicable: < 0,75 mA.

– The outside and interior of the unit must be inspected on


recognizable damages (e.g. damages on cables, connec-
tions, hoses, filter, pumps, silencer, protective caps, regu- 5.3 Function check
lating valve, manometer,...). As to units of protection class
I, especially check the grounding conductors and their OPERATING SWITCH OF THE UNIT:
connections (pay attention to wiring diagrams in service
instructions). Correct function of the operating switch / signal lamp must be
checked by operating it repeatedly.
– Inspect completeness and legibility of all labels (above all
safety-relevant labels). Optional foot switch: Test foot switch while power switch is set
to "0". Inspect On/Off switch (multiple switch) for correct func-
– Inspect fuses for correct rates acc. to specifications on the tion (positions and corresponding indications, signal lamp).
type plate.
Operating switch set to ON
– Inspect unit for corrosion damages and contaminations. – Unit permanently in operation (lamp is lit),
foot switch produces no effect.
– Inspect hose connections for leaks and kinks.
Operating switch set to OFF
– Inspect gaskets for proper seat. – Unit can be started with foot switch.

– Inspect filter for permeability and proper seat resp. correct


direction of flow (also see operating instructions). CHECK SUCTION:

Performance for vacuum and litre/min must be measured at


the unit attachment (see values in Technical Specifications,
chapter 5.4).

 Pay also attention to the respectively – Readout of measuring device must be compared with
applicable standards for service work ATMOS C 401 / C 361 readout (class 1.6).
in your country.
(e.g. the VDE 0751 in Germany). – Movement and function of vacuum regulating valve must
be checked.
 The inspections must be documented.

11
5.0 Inspection specifications
5.4 Technical specificationsons

Air flow rate of pump 45 l/min +3/-5 l/min


Max. vacuum -91 kPa ( -910 mbar or 682,5 mmHg)*@ NN
Vacuum readout -1...0 bar ± 16 mbar (class 1.6) ø 63 mm
Additional air regulation mechanical regulating valve
Collection jar 1.5 l, 3 l or 5 l jars made of polysulphone or Receptal®- system 1.5 l, 2 l or
3 l, with the trolley also 2 jars can be used simultaneously
Suction hose connections ø 6 mm or ø 10 mm
Voltage 230 V~ (+/-10%) 50/60 Hz
Special voltage:
115 V ~ (+/-10%) 60 Hz
Current input (max.) max. 0,75 A at 230 V~
max. 1,3 A at 115 V~
Operating time > 12 h continuous operation without interruption, within 24 h
Fuse T 800 mA/H for 230 V~
T 1,6 A/H for 100 - 127 V~
Protective earth conductor resistance —
Earth leakage current —
Enclosure leakage current N.C. < 0,1 mA
Patient leakage current —
Heat emission max. 173 J/s
Noise level ≤ 48 dB (A) @ 1 m (ISO 7779) at max. vacuum
Ambient conditions
Transport/storage -30...+50°C
5...90 % air humidity,
non-condensing air pressure 700...1060 hPa

Operation +5...+35°C
20...80 % air humidity,
non-condensing air pressure 700...1060 hPa
Dimensions HxWxD H 330 x W 240 x D 360 mm (with secretion container)
H 1010 x W 360 x D 440 mm (with system trolley)
Weight Ca. 6.7 kg (with collection jar)
Regular safety relevant No
inspections
Protection class (EN 60601-1) II
Degree of protection
Type BF

Protection category IPX 1


Classification acc. to Appendix IX IIa (acc. to EC Directive 93/42 EEC)
EC Directive 93/42/EEC
CE marking CE 0124
Rules applied EN 60601-1: 2007
EN ISO 10079-1: 2000
UMDNS-Code 10 - 217
REF 340.0300.0 230 V
340.0300.2 115 V

* 1 bar = 750,06 mm Hg =1000 hPa / depending on ambient air pressure Technical data issued on: 26.05.2008

12
6.0 Trouble-shooting

Problem Possible causes Remedy

Unit does not run No line voltage Check main fuse

Power plug is loose Check connection to wall socket

Fuse is defective Replace the fuse

Cables are defective Check all cables and replace,


if necessary

Switch is defective Replace the switch

Motor is defective Replace the pump

Additional air valve does not Additional air valve is damaged Replace additional air valve
close

Vacuum readout cannot be Hoses are jammed Open the housing, properly lay
changed the hoses and reassemble the
housing

Vacuum gauge is defective Replace the vacuum gauge

Suction performance is too weak Hoses are jammed Open the housing, properly lay
the hoses and reassemble the
housing

Connecting nipple on pump leaks Connecting nipple must be glued


again

Valves/diaphragms are defective Install new valves/diaphragms

Secretion was sucked into the Open the pump and clean with
pump spirit

Leakages within the hose system Check if hoses and collection jar lid
or the collection jar lid are correctly fixed

Filter is blocked (vacuum gauge Replace the filter; check hose


shows a vacuum value) connections

Secretion was sucked into the Suction procedures have been Replace hoses
pump effected without bacterial filter /
overflow safety Clean pump with spirit and check
bacterial filter, overflow safety,
sealing rings, valve plates and
silencer and replace, if necessary

13
7.0 Spare parts
7.1 Accessories

Description REF

 DDS-jar handle, PSU ................................. 340.0055.0
 DDS-bacterial filter / oversuction stop,
hydrophobic, disposable part ..................... 340.0054.0
 DDS-hose adaptor set, Ø 6 + 10 mm ........ 340.0057.0
 DDS-jar lid with gaskets ............................. 340.0053.0
  DDS-splash protection ............................... 340.0056.0
  DDS-collection jar, polysulphone, 1.5 l ...... 340.0050.0
 DDS-collection jar, polysulphone, 3.0 l ...... 340.0051.0
 DDS-collection jar, polysulphone, 5.0 l ...... 340.0052.0
 Fig 2. DDS-jar system

 Connecting nipple ...................................... 312.1165.0


 3DS-internal bag, 1,0 l ............................... 312.1180.0
 3DS-internal bag, 2,0 l ............................... 312.1181.0
  3DS-internal bag, 3,0 l ............................... 312.1182.0
 3DS-external container, 1,0 l ..................... 312.1170.0
 3DS-external container, 2,0 l ..................... 312.1171.0
 3DS-external container, 3,0 l ..................... 312.1172.0
 Adapter ring ............................................... 312.1166.0
Fig 3. 3DS-jar system

Suction hose, silicone, Ø 10 mm, 2 m ............. 000.0243.0


Suction hose, silicone, Ø 6 mm, 2 m ............... 000.0361.0
,, Suction hose, silicone, Ø 6 mm, 1,30 m .......... 000.0013.0
Single-use suction hose,
Ø 6 mm, length 1,30 m .................................... 006.0057.0
Single-use suction hose,
Ø 6 mm, length 2,10 m .................................... 006.0059.0
Bellows, silicone rubber ................................... 000.0739.0
Fig 2. DDS-jar system Fuse 230 V T 0,63 A/H .................................... 008.0634.0
Fuse 115 V T 1,25 A/H..................................... 008.0720.0
Mains cable angle-angle, 5 m.......................... 008.0818.0

Push-in foot for housing................................... 505.0337.0


Fixing screw for
 system trolley (star handle) ............................. 000.0726.0
Clamping ring for fixing screw ......................... 000.0727.0
Casing front part C451 cpl.. ............................. 340.0310.0
Casing back part C451 cpl. ............................ 340.0315.0
Front foil C 451 ................................................ 060.0528.0
Foot RD 16 X 10,5 .......................................... 505.0337.0
Vacuum gauge C451 cpl ................................. 340.0311.0
,, PCB on / off ..................................................... 999.1026.0
Plug SE KALTGER 2P TYP KEA.................... 008.0695.0
Electrical connector, extra safe........................ 008.0695.1
Fuse, slow-blow 800mA H ............................... 008.0081.0
Quik coupling W1/4“ ....................................... 000.0738.0
Suction nozzle ................................................. 340.0014.0
Bellows ............................................................ 000.0739.0
Diaphragm pump 40L/90% .............................. 000.0917.0
Cabel harness C451 ........................................ 340.0314.0
,, Hose harness C451 ......................................... 340.0312.0
Silencer............................................................ 999.1032.0
Fuse plate T 800mA ...................................... 060.0151.0
Cable harness foot switch ............................... 340.0021.0
Front foil with glass for vacuum gauge ............ 340.0010.2
Spare part kit for vacuum gauge ..................... 340.0010.5
Spare part kit for pump ................................... 069.0201.0
Capacitor 4uF ................................................. 069.0200.0
 Double castor with locking (for trolley)............. 340.0070.1
Double castor without locking (for trolley)........ 340.0070.2
Fuse plate T 1,6 A (115V) ................................ 060.0149.0
Fuse, slow-blow 115 V T 1,6 A......................... 008.0749.0
Fig 3. 3DS-jar system Operating instructions...................................... 340.0300.B

14
8.0 Plans

Wiring Diagram

L N
T800mAH

1 2 3 4
ST2 1
L N 2 Microschalter
ST1 (Option Fußschalter)
Platine Microswitch
Ein-/ 3
4 (Optional footswitch)
Ausschalter
Circuit Board
ON/OFF switch

230V ; 50/60Hz

-0.2 +0.2

Unbemaßte Kanten: Schutzvermerk DIN 34-1-D


Undimensioned
FA:
edges FAF: FAM: 1 ME: QT: 1 Zul. Abweichung Maßstab ./.
für Maße ohne
Toleranzangabe:
DIN 7168 m
über 30mm: +/-0,3
2007
Bearb.
Datum
18.10. Birkenb.
Name
Verdrahtungsplan
Gepr. Wiring Diagram
Norm
C451
ATMOS Medizintechnik Issue: 19.11.2007
Blatt
Ludwig-Kegel-Str. 16 340.0300.0
79853 Lenzkirch
15
Bl.
Zust. Änderung Datum Name 3655 AlterName
8.0 Plans
Function plan (pneumatics plan)

Schalldämpfer
Silencer

Platine mit
Ein/Aus-Schalter
PCB with
ON/OFF switch

Pumpe M

Nebenluftventil
Additional air vent

Vakuumanzeige
Vacuum display

CPC-Kupplung
CPC-clutch

Verbindungsstelle
Connection
Bakterienfilter
bacterial filter
Patient

Sekretbehälter
secretion container
elektrische Verbindung
electric connection
pneumatische Verbindung
pneumatic connection

Issue:19.11.2007

16
011.0029.0 340.0314.0

340.0318.0
8.0 Plans
Hose system

340.0313.0

006.0009.0; 160mm lg.

340.0312.0

Issue: 26.11.2007

17
8.0 Plans
Exploded view

340.0319.0
(Satz)

340.0010.1

340.0317.0
340.0010.0

060.0528.0 340.0311.0
340.0321.0
340.0310.0
053.0039.0; 2x

051.0212.0; 2x

060.0414.0

053.0089.0
000.0739.0 340.0316.0
340.0014.0 053.0088.0

340.0318.0
051.0288.0; 2x
999.1026.0
000.0702.0;
2x 340.0315.0

000.0674.0; 999.1032.0
2x
051.0340.0; 4x

340.0327.0

008.0482.0
000.0441.0 060.0415.0
(2tlg.)

000.0917.0 060.0151.0

000.0441.0 000.0738.0
(2tlg.)
051.0275.0; 053.0013.0; 2x
2x 053.0005.0; 2x
051.0059.0; 2x

053.0039.0; 4x 505.0337.0; 4x

053.0001.0; 4x
051.0008.0; 4x
fix with mit Schrauben-
loctite sicherung Loctite
fixieren
060.0374.0

008.0695.0 060.0231.0

008.0081.0
008.0695.1

Issue:05.12.2007

18
8.0 Plans
Exploded view DDS-Switchover-Docking Station

Issue:23.07.2001

19

You might also like