CPA Triplex H EN 1

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 20

VACUUM

TECHNOLOGY

PROGRAMMING MANUAL
(Translation of the original instructions)

SIMPLEX, DUPLEX, TRIPLEX AND


TRIPLEX UNI EN ISO 7396-1
AUTOMATIC VACUUM SYSTEM

SIMPLEX – DUPLEX - TRIPLEX


CPA 1x25/100 - CPA 1x40/100
CPA 1x60/100 - CPA 1x106/100
CPA 1x25/300 - CPA 1x40/300
CPA 1x60/300 - CPA 1x106/300
CPA 1x40/500 - CPA 1x60/500
CPA 1x106/500 - CPA 1x205/500
CPA 1x305/500
CPA 2x25/300 - CPA 2x40/300
TRIPLEX UNI EN ISO 7396-1 CPA 2x60/300 - CPA 2x40/500
CPA 2x60/500 - CPA 2x106/500
CPA 2x205/1000V - CPA 2x305/1000V
CPA 3x25/300V - CPA 3Hx25/300V
CPA 3x25/500V - CPA 3Hx25/500V
CPA 3x40/500V - CPA 3Hx40/500V
CPA 3x60/500V - CPA 3Hx60/500V
CPA 3x106/500V - CPA 3Hx106/500V
CPA 3x106/1000V - CPA 3Hx106/1000V
CPA 3x151/1000V - CPA 3Hx151/1000V
CPA 3x205/1000V - CPA 3Hx205/1000V
CPA 3x305/1000V - CPA 3Hx305/1000V
Programming manual
CPA1 – CPA2 – CPA3 – CPA3H
EN

TECHNICAL SPECIFICATIONS
• Power supply 3~+PE 400V±10% 50-60Hz;
• Alarm warning light [ 1 ];
• Maintenance warning light [ 2 ];
• Emergency STOP button [ 3 ];
• Button to turn on centralized vacuum system (ON) [ 4 ]
• Button to turn off centralized vacuum system (OFF) [ 5 ]
• LCD with backlighting (128x64dots) [ 6 ];
• Interlocked isolating switch [ 7 ];
• Function keys [ 8 ];
• Manual/Automatic selector, emergency pump (for hospital cabinet only) [ 9 ];
• Absolute pressure sensor 1100÷0 mbar ±0.5% FS on tank;
• Current meter protection (TA) for industrial use;
• Protection fuses;
• Buffer battery (6 years life time with not supplied switchboard)
• Protection rating of cabinet IP44;
• Protection rating of control panel IP65;
• Operating temperature from +12°C to +40°C;
• Humidity: max. 50% at +40°C;
• Rated insulation voltage 450V~;
• Rated voltage of auxiliary circuits:
♦ Control circuits: 12-24V~
♦ Coils for energizing remote switches: 24V~
Numbers between brackets [ ] refer to the front view drawing of the switchboard cabinet illustrated below.
CABINET FOR INDUSTRIAL USE CABINET FOR HOSPITAL USE (UNI EN ISO 7396-1)

Centralized vacuum systems create and maintain vacuum in a system by means of information that comes from an
electronic control unit by means of an absolute pressure sensor installed on the tank.
Each system has a START parameter and a STOP parameter. The software installed in the PLC controls operation
of the vacuum pump(s) automatically depending on the type of system and guarantees even wear and tear of
pumps (for systems with more than one pump).
Programming and management of MENU parameters are protected by three password levels. Thanks to this
precaution, staff members can work on functions that fall within their area of qualification.
• Password level 0 (Operator)
• Password level 1 (Maintenance)
• Password level 2 (Factory)

8702032 – 01/09/2016 – R.9 1 www.dvp.it


Programming manual
CPA1 – CPA2 – CPA3 – CPA3H
EN

DISPAY VISUALIZATION
After the vacuum unit is supplied, the PLC display will show as follows:

DVP Vacuum Technology


!

OFF
999 mbar (Abs.)
I
MENU MIS.
0
F1 F2 F3 F4
After the first start of the vacuum unit we advise to press the emergency button before entering the configuration
parameters in order to familiarize with the use of PLC functions and setting and to avoid the unwilling start of the
unit.

Make sure that the emergency STOP button is not pressed and start the unit pressing the green button I (ON).
The unit will turn on and the display will show the following information:

P1 250 Stop
!
mbar
P2 310 (Abs.)
P3
I
350 Start !

START STOP MENU MIS.


0
F1 F2 F3 F4

P1 Pump off.

P1 Pump running.

P1 [pump icon flashing] number of startups/hour has been reached.

! Active alarm.

Maintenance signal.

Timer function on.

Connected remote panel [displayed in the same position of “Active alarm”].

Open START/STOP remote switch (see “PLC electric connections”).

8702032 – 01/09/2016 – R.9 2 www.dvp.it


Programming manual
CPA1 – CPA2 – CPA3 – CPA3H
EN

FUNCTION KEYS
Depending on the page you have opened, keys F1 – F2 – F3 – F4 have different meanings, which are shown on the
display.
• START: absolute pressure on the start of the unit.
• STOP: absolute pressure on the stop of the unit.
• MENU: access to PLC menu.
• MIS.: access to monitoring parameters of the unit.
• ENTER: data confirmation/access to parameter or submenu.
• ESC: exit from parameter or submenu.
• DEL: data cancellation.
• ALL: access to visualization of the signalled alarm.
• : move of the selected item or increment/decrement of the value of a parameter.
•  move of the selected item or increment/decrement of the value of a parameter.
When an alarm is active, the displayed function above button F4 starts flashing and “Measurement” (MIS) and
“Alarms” (ALL) appear alternately (see Section 9- Alarms).
When the “Remote Trminal” is connected and upon condition that there are no alarms, the display will be showing
the letters "RT" (Remote Terminal) in the alarm triangle [ Kit available upon demand ].
During the functioning of the unit it is possible to select the password level 0 parameters (Operator), directly from
the aboved described frame, in order to visualize or modify them.
• To modify START pressure, hold F1 down for two seconds (the setting will start flashing). Press F1 or F2 ( )
to modify the setting and press F3 (ENTER) to save your setting, which will stop flashing. The entering of this
value may also be set by START 1st level (see paragraph 4.3)
• To modify STOP pressure, hold F2 down for two seconds (the setting will start flashing). Press F1 or F2 ( ) to
modify the setting and press F3 (enter) to save your setting, which will stop flashing. The entering of this value
may also be set by STOP 1st level (see paragraph 4.2)
• Use button F4 (Measurements – MIS.) to display the following information for each pump:
 Oil temperature (*).
 Power draw.
 Hours of work.
 Oil level check (*).
 Oil and filter replacement.
 Air filter and exhaust filter replacement (*).
 Numbers of startups per hour.

(*) This parameter is displayed only if the sensor kit obtaining the exhaust filter, temperature and oil level efficiency
is installed [ it is an optional kit ].

ANTI-PANIC FUNCTION
Irrespective of the section of the menu, the pages or settings you are working on, if you wait 30 seconds without
pressing a key, you will return to the main page and none of your modifications will be saved.
During the display and/or setting stage, the PLC is active. This means that any alarms will be displayed in the main menu.
NOTE: the functioning of alarms prevents the unit to be activated till the alarms are not deactivated
(see 9 – Alarms).

ENERGY SAVING FUNCTION


After 10 minutes have passed since the last pressing of a key, the display's backlighting will go off to save energy. it
will come back on as soon as key is pressed again.

8702032 – 01/09/2016 – R.9 3 www.dvp.it


Programming manual
CPA1 – CPA2 – CPA3 – CPA3H
EN

MENU LIST
To gain access to the PLC configuration parameter setting page from the main menu, press the F3 key (MENU)
and scroll through the parameters you wish to display or modify using F1 and/or F2 keys ( or ).
Here below is a list showing the order of the menus and the relative password levels.
Note: with a level 0 password (Operator), you will not be allowed to open a sub-menu subject to a level 1
(Assistance) or 2 password (Factory). The message "Protected parameter" will appear on the display.
Similarly, it is not allowed to enter the sub-menu of level 2 password (Factory) from level 1 password (Assistance).

Menu item Description Password level


1 PASSWORD 0/1/2
1.1 Service 1/2
1.2 Factory 2
2 DISPLAY SETUP 0/1/2
3 UNIT SETUP 1/2
3.1 Pump capacity 1/2
3.2 Tank capacity 1/2
3.3 Pump working hours 1/2
3.4 Safety time 1/2
3.5 Power down time 1/2
3.6 Emergency pump time 1/2
3.7 Starts per hour 1/2
3.8 Starter priority 1/2
3.9 Pumps working time balance 1/2
3.10 Restart 1/2
3.11 Emergency pump 1/2
3.12 Second level 1/2
3.13 STOP delay 1/2
3.14 START delay 1/2
3.15 Relay status 1/2
3.16 TPA alarm setup 1/2
3.17 Manual operation 1/2
3.18 Enabling sensors 1/2
4 PRESSURES 1/2
4.1 High vacuum alarm 1/2
4.2 STOP 1° level 1/2
4.3 START 1° level 1/2
4.4 Low vacuum alarm 1/2
4.5 TPA probe offset 1/2
4.6 STOP 2° level 1/2
4.7 START 2° level 1/2
5 OIL TEMPERATURES 1/2
5.1 Max temperature 1/2
5.2 High temperature pre-alarm 1/2
5.3 High temperature alarm 1/2
5.4 Low temperature alarm 1/2
5.5 ST1 probe offset 1/2
5.6 ST2 probe offset 1/2
5.7 ST3 probe offset 1/2
6 PUMPS CURRENTS 1/2
6.1 Max current 1/2
6.2 Current pre-alarm 1/2
6.3 Current alarm 1/2
6.4 TA1 probe offset 1/2
6.5 TA2 probe offset 1/2
6.6 TA3 probe offset 1/2
7 TIMER 0/1/2
8 MAINTENANCE 1/2
8.1 Pomp off 1/2
8.2 Time setup 1/2
9 ALARMS 0/1/2
10 RESET - General Reset 1/2 - 2
11 SERIAL PORTS 1/2
11.1 Data download 1/2
11.2 Supervision 1/2
11.3 Modem 1/2
12 SERIAL NUMBERS 2

8702032 – 01/09/2016 – R.9 4 www.dvp.it


Programming manual
CPA1 – CPA2 – CPA3 – CPA3H
EN

1 - PASSWORD
1.1 - SERVICE
1.2 - FACTORY

1 - Password 1 - Password 1 - Password

1 - Service 1 – Service 2 – Factory


2 - Factory

0XX 0XXX
↑ ↓ ENTER ESC ↑ ↓ ENTER ESC ↑ ↓ ENTER ESC

From the main menu, press F1 and/or F2 ( or ) to select “Password" and confirm with F3 (ENTER).
Still using the same buttons, select the password level and press F3 to confirm. The password level is shown on the
first line. The second line features '0' (highlighted) and X X [Service] or 0 (highlighted) and X X X [Factory].
Default settings of password levels:
• Service password = 111 factory setting.
• Factory password = set by the manufacturer.
If the password is correct, the main menu will appear otherwise the message "Wrong password" will appear for 5
seconds and the page previously displayed will appear.
If you have forgotten your password, turn the system off and on again, and hold keys F3 down contemporaneously
and F4 for 5 seconds. During this lapse of time, the message "Password" will be flashing on the display; when the
message stops flashing, the default settings of the passwords have been resumed.
To modify password Assistance: once entered menu 1.1. – ASSISTANCE following the above mentioned
indications, the main menu is displayed again and the option 1 “Change Password” appears. Enter the sub-menu
pressing F3 (ENTER) and then write the new password.

2 – DISPLAY SETUP

ENGLISH P mbar
P mbar °C Mon 01/01/2011
10:23
°C - +
Daylight time: Yes
- + Set clock
ENTER ESC ENTER ESC ENTER ESC

In this section it is possible to visualize (pressing F1 e F2 -  /  ) and/or modify the following parameters:
• Language (Italian / English / French / German / Spanish / Portuguese) – [ default setting: Italian ].
• Unit of measurement of pressure (mbar (Abs.) / PSI / mmHg) – [ Default setting: mbar (abs.) ].
• Temperature (°C / °F) – [ Default setting: °C ]
• Brightness of display.
• Setting time / day / month / year.

8702032 – 01/09/2016 – R.9 5 www.dvp.it


Programming manual
CPA1 – CPA2 – CPA3 – CPA3H
EN

3 – UNIT SETUP
3.1 - PUMP CAPACITY
Setting range: from 50 to 50,000 litres/minute (5 l/min. steps).
Default setting: 1000 l/min (to be modified depending on reference unit).

3.2 - TANK CAPACITY


Setting range: from 50 to 99,000 litres (step 10 litres).
Default setting: 1000 l (to be modified depending on reference unit).

3.3 - PUMP WORKING HOURS


Selecting this parameter the pumps of the unit and their hours of work are displayed:
Pompa n°1 = xxxx h
Pompa n°2 = xxxx h
Pompa n°3 = xxxx h
This parameter allows to enter the work hours of the pumps of the vacuum unit, when the PLC is installed or
replaced.

3.4 – SAFETY TIME


Setting range: from 2 to 300 sec. (5 sec. steps).
Default setting: 10 sec.
This parameter sets the time that must pass between turning the single pump off and turning it on again.

3.5 – POWER DOWN TIME


Setting range: from 0 to 3600 (5 sec. steps).
Default setting: 0 sec.
If this parameter is set choosing a value different from 0 second, when the unit is stopped by means of the 0 (OFF)
button, or timer or remote START/STOP, it allows the pumps to remain active till the set time is up and the display
shows OFF falshing to countdown the OFF time.
During this time countdown, any restarting commands cannot be accepted.
When time countdown ends, the message OFF will stop flashing and the centralised vacuum system goes off.
NOTE: while the turning off phase the emergency button remains active, so it is possible to stop the unit
pressing it.

3.6 – EMERGENCY PUMP TIME [this is only active on vacuum systems Duplex and Triplex]
Permissible setting: YES / NO.
Default setting: NO.
If setting is NO: pumps will work in accordance to the ON / OFF parameters that have been set.
If setting is YES: Setting range from 1 to 60 min. (default setting 5 min.). The vacuum system works as follow:
In case the air entering the system prevents the pump to reach the STOP pressure value, the reserve pump (the
one with less working hours) starts working after that the set time (counted from the turning on of the pump) has
elapsed. When the second pump starts working , the countdown of the set time begins again and if at the deadline
the pressure has not reached the STOP pressure value yet, the third pumps starts working.
In this way it is possible to equilibrate the working hours of the pumps.

3.7 – STARTS PER HOUR


Permissible setting: YES / NO.
Default setting: YES.
If this parameter is selected, , it allows the management and monitoring of the PLC software with reference to the
numbers of the startups per hour of each single pump installed on the vacuum unit.
If setting is NO: the starts per hour are not counted by the PLC, but only visualized in “MIS” menu (Measurement)
of the main display visualization.
If setting is YES: setting range 6 ÷ 20 starts/hour (default value 12). The 1 hour timer starts working from the very
first START (I – ON) of the vacuum unit, in this way the numbers of starts of each pump are counted and PLC
activates the pump with less working hours. In case the pumps have reached the set limit of starts/stops, when the
system activates them, the visualization of P1, P2, P3 flashes on the diplay, RL4 is de-energized (see chapter 9 –
Alarms) and the event is recorded in the alarm history.
When the hour is up, memorization resets to zero and a new one starts.

8702032 – 01/09/2016 – R.9 6 www.dvp.it


Programming manual
CPA1 – CPA2 – CPA3 – CPA3H
EN

3.8 – STARTER PRIORITY


If this parameter is selected, the following items appear in the PLC display:
PUMP NR.: possible settings 1 ÷ 3 (default setting: 1)
DELAY: possible settings 5 ÷ 240 sec. (default setting: 5 seconds)
The first item (pump nr.) establishes which pump starts working first each day.
The second item (Delay) establishes the time after which the second and third pump start working.
So this procedure avoids electrical input peaks during the starting of the unit.

3.9 – PUMPS WORKING TIME BALANCE [this is only active on vacuum systems Duplex and Triplex]
Permissible setting: YES / NO.
Default setting: NO.
This parameter allows the PLC software to manage the work of each pump in order to balance the working hours.

3.10 - RESTART
Permissible setting: MANUAL / AUTOMATIC.
Default setting: MANUAL.
If setting is MANUAL: if a power failure occurs and the switchboard goes off, when power is resumed, the PLC
puts itself in OFF condition and the "No Power" alarm message appears. The pumps of the vacuum system are off.
If setting is AUTOMATIC: if a power failure occurs and the switchboard goes off, when power is resumed, the
vacuum system starts automatically depending on the pressure of the system.

DANGER: considering the risk relating to use of this operation, read the "Instructions for use"
paragraph of the use and maintenance manual with particular care!

3.11 – EMERGENCY PUMP [this is only active on vacuum systems Duplex and Triplex]
Permissible setting: YES / NO.
Default setting: NO.
If setting is YES: setting range from 1 to 3 (default setting: 1). This parameter allows the selection of the pump,
which is NOT used during the normal functioning of the unit, but only in case a working pump fails or a pump is
momentarily stopped for maintenance.
NOTE: during the first start of the day of the unit the reserve pump is activated for a sigle cycle in order to
verify its correct functioning.

3.12 – SECOND LEVEL


Permissible setting: YES / NO.
Default setting: NO.
If setting is YES: this parameter permits to manage the set values for 2° level START and STOP of the vacuum
unit (see 4.6 and 4.7).
NOTE: the parameter “Second Level” is only controlled by the activated “Timer” function (parametr 7).

3.13 – STOP DELAY


Setting range: from 0 to 60 min.
Default setting: 0 sec.
Thanks to this set parameter the PLC makes the “Turn Off” command active when the set STOP value is reached
and activates the delay, which is performed , if the control algorithm allows the turning off.

3.14 – START DELAY


Setting range: from 0 to 60 min.
Default setting: 0 min.
Thanks to this set parameter the PLC makes the “Turn On” command active when the set START value is reached
and activates the delay, which is performed, if the control algorithm allows the turning on.

3.15 – RELAY STATUS


This parameter controls relay alarms (see 9 – Alarms)
Pressing F3 (ENTER) the parameter lights up and relay RL4, RL5 and RL6 and their status are displayed.
Pressing F1 or F2 ( / ) the relay to check is selected and confirming with F3 the relay changes status for 5
seconds, after this time the initial menu will appear.

8702032 – 01/09/2016 – R.9 7 www.dvp.it


Programming manual
CPA1 – CPA2 – CPA3 – CPA3H
EN

3.16 – TPA ALARM SETUP (Absolute Pressure Transducer)


Permissible setting: YES / NO.
Default setting: NO.
If setting is NO: in case of failure of the absolute pressure transducer the unit stops.
If setting is YES: in case of failure of the absolute pressure transducer the unit works in Manual mode (see 3.17 –
Manual functioning) visualizing the alarm on the display and recording it in the alarm memory.

3.17 – MANUAL OPERATION


Permissible setting: YES / NO.
Default setting: NO.
If setting is NO: The PLC works in “Automatic” mode controlling the functioning of the vacuum unit by means of
set parameters START (4.2) and STOP (4.2).
If setting is YES: The PLC works in “Manual” mode. Starting the unit by pressing button I (ON) and/or by the
opening of ID2 (remote START/STOP) or by the clock signal ( see parameter 7 – Timer in ON mode).
NOTE: in this mode pressure is not checked by the transducer, as a consequence the pump (or pumps)
works continuously and STOP and START settings do not appear in the display to signal the “Manual”
mode of the pump/s.

3.18 – ENABLING SENSORS (only if sensor kit is installed, signalling the efficiency of the exhaust filter, oil
temperature and level – available upon demand)
Permissible setting: YES / NO.
Default setting: NO.
If setting is NO: the sensor kit is not fitted to the pump/s and the PLC will not be displaying the following
parameters:
• Temperature parameter and all temperature controls (parameter 5 - Temperature);
• In the main visualization of Measurement menu (MIS) is not active the visialization of the oil temperature of the
pump/s;
• Temperature alarm parameter (STx) (Parameter 9 - Alarms);
• Oil level alarm parameter (LOx) (Parameter 9 - Alarms);
• Pressure switch alarm parameter (PRx) (Parameter 9 - Alarms);
If setting is YES: the sensor kit is fitted to the pump/s and the PLC will be displaying all following parameters are
displayed:
• Temperature parameter (parameter 5 - Temperature);
• In the main visualization of Measurement menu (MIS) is active the visialization of the oil temperature of the
pump/s;
• Temperature alarm parameter (STx) (Parameter 9 - Alarms);
• Oil level alarm parameter (LOx) (Parameter 9 - Alarms);
• Pressure switch alarm parameter (PRx) (Parameter 9 - Alarms);

ATTENTION: if there is no sensor kit fitted to the pump, this parameter must be set to NO otherwise the
PLC will not give the authorisation to start (I - ON) at the vacuum system.

8702032 – 01/09/2016 – R.9 8 www.dvp.it


Programming manual
CPA1 – CPA2 – CPA3 – CPA3H
EN

4 – PRESSURES
4.1 - High vacuum alarm (P1)
900
Setting range: from 0 to (P2-25) mbar (Abs.). ALL - P4

Default setting: 0 mbar (Abs.).


875
START - P3 START - P7
4.2 - STOP 1st level (P2)
MAX MAX
Setting range: (P1+25) ÷ (P3-25) mbar (Abs.). STOP - P2
850
STOP - P6
Default setting: 250 mbar (Abs.).
825
4.3 - START 1st level (P3)
Setting range: (P2+25) ÷ (P4-25) mbar (Abs.).
Default setting: 350 mbar (Abs.).

4.4 - Low vacuum alarm (P4)


Setting range: (P7+25) ÷ 900 mbar (Abs.).
Default setting: 900 mbar (Abs.). ALL - P4
100

4.5 - TPA probe Offset (P5) 75

Setting range: from -20 to +20 mbar (Abs.).


Default setting: 0 mbar (Abs.). 50
START - P7

4.6 - STOP 2nd level (P6) MIN


25
Setting range: (P2+25) ÷ (P7-25) mbar (Abs.). START - P3 STOP - P6

Default setting: 380 mbar (Abs.). MIN


0
ALL - P1 / STOP - P2

4.7 - START 2nd level (P7) mbar (Abs.)


Setting range: (P6+25) ÷ (P4-25) mbar (Ass.). 1° Level 2° Level
Default setting: 450 mbar (Abs.).

5 – OIL TEMPERATURES
The following parameters refer to the temperature of the oil in the tank of the pump/s and allow the visualization of
the temperarture on PLC display only if the sensor kit is installed, showing the efficiency of the exhaust filter, oil
temperature and level.
NOTE: the values of the parameters must be T1>T2>T3>T4.

5.1 - Max. temperature (T1)


Setting range: 130°C ÷ (T2+2°C)
Default setting: 110°C.
5.2 - High temperature pre-alarm (T3)
Setting range: (T2-2°C) ÷ (T4+2°C).
Default setting: 95°C.
5.3 - High temperature alarm (T2)
Setting range: (T1-2°C) ÷ (T3+2°C).
Default setting: 100°C.
5.4 - Low temperature alarm (T4)
Setting range: 0°C ÷ +10°C.
Default setting: 0°C.

5.5 - ST1 probe offset (T5)


Setting range: 10°C ÷ +10°C.
Default setting: 0°C.
5.6 - ST2 probe offset (T6)
Setting range: 10°C ÷ +10°C.
Default setting: 0°C.

5.7 - ST3 probe offset (T7)


Setting range: 10°C ÷ +10°C.
Default setting: 0°C.

8702032 – 01/09/2016 – R.9 9 www.dvp.it


Programming manual
CPA1 – CPA2 – CPA3 – CPA3H
EN

6 – PUMPS CORRENTS
NOTE: the setting must be entered in PLC following this order 1, A2. A3 and respecting this proportion
A1>A2>A3.

6.1 - Max. current (A1)


Setting range: 30A÷1A.
Defaul setting: Plate value x 1.7 (factory setting).

6.2 - Current pre-alarm (A3)


Setting range: (A2-0.5A) ÷ 1A.
Defaul setting: Plate value x 1.3 (factory setting).

6.3 - Current alarm (A2)


Setting range: (A1-0.5A) ÷ (A3+0.5A).
Defaul setting: Plate value x 1.4 (factory setting).

6.4 - TA1 probe offset(A4)


Setting range: -2.0A ÷ +2.0A.
Default setting: 0A.

6.5 - TA2 probe offset (A5)


Setting range: -2.0A ÷ +2.0A.
Default setting: 0A.

6.6 - TA3 probe offset (A6)


Setting range: -2.0A ÷ +2.0A.
Default setting: 0A.

7 - TIMER
Permissible setting: YES / NO.
Default setting: NO.
Thanks to this parameter it is possible to set 3 different daily time slots during which the starting (I – ON) and stop
(0 – Off) of the unit may be planned together with the programming of the 2 different pressure levels (1° and 2°
level; see 4 – Pressures) assigned to each time slot.
If setting is NO: the parameter is deactivated.
If setting is YES: the symbol of the clock (see Display visualizations) appears in the PLC display, either the unit is
working or not.
In case the pressure parameter 2° level is activated (see 3.12 – 4.6 – 4.7) in TIMER menu together with ON and
OFF parameters, the LIV settings will be displayed allowing the control of the 3 time slots of START/STOP
assigned both to 1° and 2° level of pressure.

DVP Vacuum Technology 7 - Timer


! ! Mon ON OFF Lev.
OFF
999 mbar (Abs.)
I
08:00 12:00
14:00 18:00
1
2 I
18:01 24:00 1
MENU MIS. ENTER ESC
0 0
F1 F2 F3 F4 F1 F2 F3 F4

Activating this parameter, that is, the activation of START (I –ON)/STOP (0 –OFF) by means of the timer, in any
case it is possible to turn the unit on bypassing this set parameter (TIMER), holding I (ON) button down for 5
seconds. When the unit is turned on, it automatically re-starts working respecting the previously saved time slots.
Similarly, press 0 button (OFF) to turn the unit off (in case of unscheduled stop) in TIMER mode.
NOTE: the signal comes from the remote emergency button START/STOP (see 4.8.2 in Use and
Maintenance Manual) and it is active only when TIMER parameter is in NO mode. In other words, when
TIMER parameter is in YES mode, the remote START/STOP function is not active.

8702032 – 01/09/2016 – R.9 10 www.dvp.it


Programming manual
CPA1 – CPA2 – CPA3 – CPA3H
EN

8 – MAINTENANCE

8.1 – Pump OFF


Permissible setting: YES / NO.
Default setting: NO.
Selecting this parameter all the unit pumps are visualized in the PLC display showing all the saved settings (ON and
OFF).
If setting is ON: The pump in ON mode is active and it is involved in the work cycle of the unit.
If setting in OFF: The pump in OFF mode is not active and, as a consequence, it is not involved in the work cycle
of the unit.

8.2 – Time setup


This parameter allows the management of the counters signalling the kinds and intervals of maintenance of pump/s
of the unit.
Selecting this parameter all the pumps of the unit are visualized and the following items are displayed for each
pump:
• Oil check (M1) – [Oil level check]
Setting range: 100 ÷ 3.000 h.
Default setting: 100 h.
• Oil and oil filter (M2) – [Change of oil and oil filter]
Setting range: 100 ÷ 9.900 h.
Default setting: 500 h.
• Inlet and exhaust filter (M3) – [Change of inlet and exhaust filter]
Setting range: 100 ÷ 9.900 h.
Default setting: 1000 h.

If the new setting is lower than the countdown underway, the latter is reset.
The countdown refers to the hours of work of each pump (hours of work are counted down); when the countdown
reaches 0, the relative maintenance alarm is given and the countdown is continued in negative.
Hours are memorised every 15 minutes. If a power failure occurs during the countdown, the 15-minute lapse of time
will not be considered.

With the maintenace message on the display the yellow signal of the optional alarm visualization kit turns on and
RL5 relay contact is closed. (see section 9 – Alarms and vacuum unit maintenance manual).
NOTE: cancel the displayed alarms, once maintenance is carried out. After cancellation of the alarms the
calculation starts again to signal the next maintenance alarm.

8702032 – 01/09/2016 – R.9 11 www.dvp.it


Programming manual
CPA1 – CPA2 – CPA3 – CPA3H
EN

9 - ALARMS
The access to this menu section (password level 0 – Operator) allows the visualization of the alarms with relative
codes, meaning, progressive number, cause, date and time of the event.
The memory may contain till 100 alarms and message nr. 101 cancels the first alarm and so on.
If the memory is empty, the message “EMPTY MEMORY” appears.
When an alarm is obtained, in the low right part of the display the symbol of “Alarm Presence” (triangle) appears
and the red Led (Alarm Presence) turns on, RL6 relay on NC is closed and button F4 alternatively changes meaning
from Measurement to Alarm (ALL). Selecting F4 (when on ALL) it is possible to visualize the generated alarm.

DVP Vacuum Technology 250 Stop


P1
! !

OFF
999 mbar (Abs.)
I
P2
P3
mbar
310 (Abs.)
I
! 350 Start !

MENU ALL START STOP MENU ALL


0 0
F1 F2 F3 F4 F1 F2 F3 F4

To reset an alarm, select it (button F1/F2 - /) and press F3 (DEL).


Then press F4 (ESC) to exit and go back to the previous menu.
NOTE: the alarm cancels only once that its cause is solved. Otherwise clear tension and apply it again.

RELAY STATUS
• RL4 (machine status): relay RL4 is energized when the system is ON or OFF, when maintenance signal (M) is
active and when no alarm is activated.; it is de-energized when block (B) and/or visual alarms (W) are active.
• RL5 (maintenance): relay RL5 is energized when the maintenance signal (M) is active; it is de-energized when
the system is ON or OFF and when block (B) and/or visual alarms (W) are active.
• RL6 (block alarm): relay RL6 is energized when the system is ON or OFF, when maintenance signal (M) is active
and when no alarm is activated; it is de-energized when block (B) and/or visual alarms (W) are active.

RELAY STATUS RL 4 RL 5 RL 6
OFF ON (n.o.) OFF ON
START ON (n.o.) OFF ON
WARNING (W) OFF (n.c.) OFF ON
BLOCK (B) OFF (n.c.) OFF OFF (n.c.)
MAINTENANCE (M) ON (n.o.) ON (n.o.) OFF (n.c.)

MEANING OF THE ALARM CODES

AL.B.xx Alarm that blocks the system – (B)


AL.B.xx-P.x Alarm that blocks a single pump – (B)
AL.W.xx Alarm that does not block the system – (W)
AL.W.xx-P.x Alarm that does not block a single pump – (W)
AL.M.xx-P.x Maintenance alarm of a single pump – (M)

8702032 – 01/09/2016 – R.9 12 www.dvp.it


Programming manual
CPA1 – CPA2 – CPA3 – CPA3H
EN
ALARM THAT BLOCK THE SYSTEM: AL.B.xx

Cod. Input Meaning of alarm


AL.B.00 Loss of setting data
AL.B.01 ID1 aperto Emergency STOP button
AL.B.03 TPA Pressure transducer is faulty
AL.B.04 Power No power
AL.B.05 ST Temperature probes are faulty
AL.B.06 TA Selected TA ammeter is faulty

NOTE: The AL.B.04 “No power” alarm is displayed only if the system is set up for manual restarting. On the
contrary, the message “Power OFF” appears.

ALARM THAT BLOCK A SINGLE PUMP: AL.B.xx-P.x


The events causing this kind of alarms stop of the relative pump.
In case the unit installs one pump, this latter stops.
In case the unit installs more than one pump, the PLC finds the one no longer available and it automatically controls
the functioning pumps.

Cod. Input Meaning of alarm


AL.B.01-P.1 ST1 > T2 High oil temperature of pump n°1
AL.B.01-P.2 ST2 > T2 High oil temperature of pump n°2
AL.B.01-P.3 ST3 > T2 High oil temperature of pump n°3
AL.B.02-P.1 ST1 < T4 Low oil temperature of pump n°1
AL.B.02-P.2 ST2 < T4 Low oil temperature of pump n°2
AL.B.02-P.3 ST3 < T4 Low oil temperature of pump n°3
AL.B.03-P.1 ST1 Oil temperature probe of pump n°1 out of order
AL.B.03-P.2 ST2 Oil temperature probe of pump n°2 out of order
AL.B.03-P.3 ST3 Oil temperature probe of pump n°3 out of order
AL.B.04-P.1 TA1 T.A or contactor of pump n°1 out of order
AL.B.04-P.2 TA2 T.A or contactor of pump n°2 out of order
AL.B.04-P.3 TA3 T.A or contactor of pump n°3 out of order
AL.B.05-P.1 T.A.1 > A2 Current alarm of pump n°1
AL.B.05-P.2 T.A.2 > A2 Current alarm of pump n°2
AL.B.05-P.3 T.A.3 > A2 Current alarm of pump n°3

NOTE: Alarm AL.B.04-P.1÷3 can be caused by:


1) While the pump works, the current absorption is not registered.
2) When the pump is off, the current absorption is registered.

ALARM THAT DOES NOT BLOCK THE SYSTEM: AL.W.xx

Cod. Input Meaning of alarm


AL.W.01 P > P4 Low vacuum level
AL.W.02 P < P1 High vacuum level

NOTE: Alarm AL.W.01 is displayed when the vacuum system is turned on and when the first pressure check is
performed.

8702032 – 01/09/2016 – R.9 13 www.dvp.it


Programming manual
CPA1 – CPA2 – CPA3 – CPA3H
EN
ALARM THAT DOES NOT BLOCK A SINGLE PUMP: AL.W.xx-P.x

Cod. Input Meaning of alarm


AL.W.01-P.1 ID3 open Oil level of pump n°1
AL.W.01-P.2 ID4 open Oil level of pump n°2
AL.W.01-P.3 ID5 open Oil level of pump n°3
AL.W.02-P.1 ID6 open Pressure switch of pump n°1
AL.W.02-P.2 ID7 open Pressure switch of pump n°2
AL.W.02-P.3 ID8 open Pressure switch of pump n°3
AL.W.03-P.1 ST1 > T3 Oil temperature pre-alarm of pump n°1
AL.W.03-P.2 ST2 > T3 Oil temperature pre-alarm of pump n°2
AL.W.03-P.3 ST3 > T3 Oil temperature pre-alarm of pump n°3
AL.W.04-P.1 TA1 > A3 Current pre-alarm of pump n°1
AL.W.04-P.2 TA2 > A3 Current pre-alarm of pump n°2
AL.W.04-P.3 TA3 > A3 Current pre-alarm of pump n°3
The established number of startups/hour for pump n°1 has been
AL.W.05-P.1 St./h
reached.
The established number of startups/hour for pump n°2 has been
AL.W.05-P.2 St./h
reached.
The established number of startups/hour for pump n°3 has been
AL.W.05-P.3 St./h
reached.
NOTE:
1) Alarms AL.W.03-P1÷3 are automatically reset when temperature drops below the T3 – 2°C.
2) Alarms AL.W.04-P1÷3 are automatically reset when current drops below A3 – 0.5 A.
3) Alarms AL.W.02-P.1÷3 are activated only when the temperature of the relative pump is over 30°C.

MAINTENANCE ALARMS: AL.M.xx-P.x


This kind of alarms turns the red led (Maintenance) on.

Cod. Input Meaning of alarm


AL.M.01-P.1 C-L-O.1 Check oil level of pump n°1
AL.M.01-P.2 C-L-O.2 Check oil level of pump n°2
AL.M.01-P.2 C-L-O.3 Check oil level of pump n°3
AL.M.02-P.1 C-O/F.1 Replace oil and oil filter of pump n°1
AL.M.02-P.2 C-O/F.2 Replace oil and oil filter of pump n°2
AL.M.02-P.3 C-O/F.3 Replace oil and oil filter of pump n°3
AL.M.03-P.1 C-F-D/A.1 Chain exh.filter and air filter of pump n°1
AL.M.03-P.2 C-F-D/A.2 Chain exh.filter and air filter of pump n°2
AL.M.03-P.3 C-F-D/A.3 Chain exh.filter and air filter of pump n°3

NOTE: When there is a maintenance alarm, press F3 to reset alarm display. However, if the alarm is not reset in
menu 10 - maintenance reset - the alarm message will reappear after 8 hours.

8702032 – 01/09/2016 – R.9 14 www.dvp.it


Programming manual
CPA1 – CPA2 – CPA3 – CPA3H
EN

10 - RESET
Entering this section it is possible to reset the below listed parameters as they were programmed by the factory.

• Maintenance
Selecting this parameter all the pumps of the unit are visualized. Choosing a pump the items of the “Maintenance “
menu (check of oil, oil filter and inlet/exhaust filter) appear in the display. Now it is possible to reset the intended
parameter.

• Hours of work
Selecting this parameter all the pumps of the unit are visualized. Choosing a pump the display shows YES/NO for
the relative setting.
NOTE: the resetting of the hours of work has to be used only in case of replacement of a pump or PLC. In case the
counter of the replaced pump is reset to zero, the PLC will make it work till that pump will reach the same working
hours of the other pumps (if settting is YES in “Hour balancing” parameter, see 3.9).

• Alarms
Permissible setting: YES / NO.
Default setting: NO.
Selecting NO: You will return to the previous page.
Selecting YES: The resetting of the alarms is carried out.

• General (2 level password - Factory)


Permissible setting: YES / NO.
Default setting: NO.
Selecting NO: You will return to the previous page.
Selecting YES: You will confirm the general resetting procedure. This function is only accepted when the system is
OFF and the controller acquires all default settings. Alarms can thus be reset.

NOTE: When resetting has been confirmed, the message “RESETTING” starts flashing. When resetting has been
completed, the message will stop flashing and previous page will appear.

11 – SERIAL PORTS

11.1 – Data download


NOT active parameter.

11.2 – Supervision
NOT active parameter.

11.3 - Modem
This parameter allows the entering of telephone numbers of GSM modem.
This parameter is not active.

12 – SERIAL NUMBERS
In this menu section (protected 2 level password – Factory) the values of the following parameters are entered:
• Serial nr. P1
• Serial nr. P2
• Serial nr. P3
• Vacuum system serial number
• CPU software release
• PLC software release (factory settings)

8702032 – 01/09/2016 – R.9 15 www.dvp.it


Programming manual
CPA1 – CPA2 – CPA3 – CPA3H
EN

ELECTRICAL PLC CONNECTIONS


SERIAL RS232
CONNECTOR
CONNECTION TO + T R G
CONTROL PANEL 15 X X N
V D
CONNECTOR TO
I/O AXPANSION

1
M11 M10 M2 ID5 LO3

OIL LEVEL
M5 ID4 LO2
PROBES
TA1

ID3 LO1
AMPERE
TRANSDUCERS
TA2 M9 1

REMOTE
ID2 STR
TA3 START/STOP

KD1
1
ID8 PR3

NO
M6
PRESSURE
ID7 PR2
PROBES

RL3
ST1
ID6 PR1
TEMPERATURE 1
RL2
ST2 Relay 3
PROBES
M8 Relay 2
M3
ST3 Relay 1
RL1

1
- NO
PRESSURE
TPA Relay 6
+ NC
RL6

TRANSDUCERS
1
NO
N
M4 NC Relay 5
RL5

POWER SUPPLY L12


M1 NO
L24
NC Relay 4
RL4

1 1

RELAY STATUS
WITHOUT
POWER SUPPLY

CONNECTION KEY
Terminal Description Connections
n°1 - Neutral (0 Vac)
n°2 - L12V (12Vac)
M1 Power supply n°3 - L24V (24Vac)
n°4 - L24V EM
n°1 = GND
n°2 = RX
M2 Serial output n°3 = TX
n°4 = +15 Vcc
n°1 = P. n°1
Relay outputs for pump
M3 n°2 = P. n°2
control n°3 = P. n°3
RL4 = n°1 N.C. e n°2 N.O.
Relay output alarms
M4 RL5 = n°3 N.C. e n°4 N.O.
(relay contacts: 2A - 24Vac) RL6 = n°5 N.C. e n°6 N.O.
n°1e2 = ID3 (LO1)
M5 Safety devices n°3e4 = ID4 (LO2)
n°5e6 = ID5 (LO3)
n°1e2 = ID6 (PR1)
n°3e4 = ID7 (PR2)
M6 Safety device n°5e6 = ID8 (PR3)
n°7e8 = ID2 (STR) [remote START/STOP – Only for cabinet for industrial use]
n°1e2 = ST1
n°3e4 = ST2
Temperature probes
M8 n°5e6 = ST3
Pressure transducer n°7 = TPA (negative - white)
n°8 = TPA (positive - brown)
n°1e2 = TA1
M9 Transducers, Ammeters n°3e4 = TA2
n°5e6 = TA3
M10 Front card
M11 Expansion

8702032 – 01/09/2016 – R.9 16 www.dvp.it


Programming manual
CPA1 – CPA2 – CPA3 – CPA3H
EN

DIAGRAM FOR CONNECTION TO REMOTE PANEL


NOTE: When remote panel is connected, the message “RT” appears on the display.

+15V

GND
RX
TX

1
M11 M10 M2

M5

M9 1
KD1

1
NO

M6
RL3

1
RL2

M8
M3
RL1

1
RL6

M4
RL5

M1
RL4

1 1

CPU BOARD REMOTE OPERATOR PANEL

8702032 – 01/09/2016 – R.9 17 www.dvp.it


GENERAL CONDITIONS OF SALE
D.V.P. Vacuum Technology s.p.a. supplies products exclusively for professional clientele, hence, excluding
consumers.

PRODUCT WARRANTY TERMS AND CONDITIONS


D.V.P. Vacuum Technology s.p.a. guarantees that the product is free from material or manufacturing
defects for a period of 24 months of normal use from the shipping date. This period is of 6 months of normal
use for products subject to repair not under warranty.
Normal use means an operating cycle of 8 hours per day for a maximum of 5000 operating hours in the 24
months covered by the warranty.

Warranty means the free replacement or repair at its own assistance network of any components of the
product that are found to be faulty from the start due to manufacturing defects.

In the event of repair, D.V.P. Vacuum Technology s.p.a. guarantees, exclusively to its own customer, the
identical spare parts for 24 months from the shipping date; once this period has passed, the pieces may no
longer be available on the market, therefore the repairs, even under warranty, may require the payment of a
difference between the product purchased and that installed during the repair. This price will be indicated
to the customer before the repair is carried out, for acknowledgement and acceptance.
D.V.P. Vacuum Technology s.p.a. will do everything reasonable within its power to respect the assistance
times and standard response (20 working days), which may vary according to the distance and accessibility
of the place where the product is located and the availability of the components.
D.V.P. Vacuum Technology s.p.a. will not be held responsible for any direct or indirect losses caused by its
failure to respect the assistance times and will not have any responsibility or contractual or civil obligation
for product faults or for failure to repair the faults in a reasonable period of time.

In the event of irreparable faults, the product will be replaced. The replacement will cause the original
warranty to be extended to the new product, until its expiry date.

The warranty does not cover any parts that appear to be faulty due to negligence and/or carelessness
during use (failure to observe the equipment operating instructions, lack of maintenance), incorrect
installation and/or maintenance, maintenance carried out by unauthorized staff, damage due to transport, or
circumstances which, in any case, cannot be attributed to manufacturing faults on the equipment.
The warranty also excludes all components of the product that have been modified or repaired without prior
written authorisation from D.V.P. Vacuum Technology s.p.a.

The warranty also excludes any faults deriving from improper use, normal wear, galvanic and electrostatic
currents, chemical corrosion, tampering, replacement or elimination of the registration plate.
The warranty does not cover, in any case, faults generated by external causes, such as accidents and
fortuitous events.

D.V.P. Vacuum Technology s.p.a. declines all responsibility to anyone for any damage and, consequence, of
any kind and/or reason, that may derive from the use of the product, as well as for any faults that it may
present.
By way of non-limiting example, it declines all responsibility:
• for any damage that could, directly or indirectly, be caused to people, objects and animals, due to failure
to observe all the instructions indicated in the relevant use and maintenance manual, especially the
indications on the installation, use and maintenance of the equipment;
• for any damage and/or loss caused by faults of deficiencies of products repaired by D.V.P. Vacuum
Technology s.p.a.;
• for any indirect or consequential damage such as, by way of non-limiting example, loss of business,
profits, salaries, payments etc.;
• losses that could have been avoided by the customer by following the advice and instructions from
D.V.P. Vacuum Technology s.p.a..
In any case, the customer waives the right to claim any right and/or demand as well as raising any objection
or promoting any action, inherent to the use of the product.

The warranty is not extended to consumable parts, or faults deriving from: filtering cartridges, blades,
membranes or sealing rings, as well as third party products that are part of the final product.

The transport, removal and subsequent re-installation costs of the repaired or replaced product are,
however, to be entirely borne by the customer.
Cod. 8702032 – 01/09/2016 – R.9 – ( EN )

D.V.P. Vacuum Technology s.p.a.


Via Rubizzano, 627
40018 San Pietro in Casale (BO) – Italy
Ph +3905118897101
Fx +3905118897170
www.dvp.it

You might also like