Tec 6 Plus Vaporizer

Tec 6 Plus Vaporizer

Features

• Electronic vaporizer which delivers the • For added convenience and safety, the Tec 6 Plus
anesthetic agent, Suprane® (desflurane) LED display indicates vaporizer status, including:

• Heats Suprane to maintain temperature and

Operational
No output
vapor pressure for consistent output
• Low agent
• Provides a concentration range of 1% to 18% • Warm-up
to help enhance titratability and depth regulation • Alarm battery low
benefits of desflurane
• Attaches to Datex-Ohmeda’s interlocking
Selectatec® manifold which allows you to
change vaporizer arrangement from case to
case or move vaporizers from suite to suite
without the need for tools
Tec 6 Plus Vaporizer

Physical specifications Electrical specifications


Dimensions Voltage requirement: Either 220 to 240 V or 100 to 120 V a.c.
50/60 Hz., depending on the voltage rating
Height: 25.0 cm/9.9 in
of the vaporizer.
Depth: 23.5 cm/9.3 in
Battery type: Duracel MN/1604
Width: 11.0 cm/4.3 in
Power consumption: 250 VA nominal
Weight: 9.5 kg/20 lbs, 15 oz empty
Environmental specifications
Pneumatic specifications
Operation
Operation
Temperature: 18° to 30°C/64° to 86°F
Concentration range: 1 to 18%
Humidity: 30 to 75% relative humidity (non-condensing)
Scale: 1% graduations from 1 to 10%
2% graduations from 10 to 18% Storage

Calibrated Temperature: –40° to 60°C/–40° to 140°F


flow range: 0.2 to 10 L/min (at 21°C) Humidity: 0 to 95% relative humidity at 40°C
Accuracy at (non-condensing)
5 L/min of O2: ±0.5% of delivered agent or ±15% of dial setting Effects of output at varied altitudes
(whichever is greater)
Decreased atmospheric pressure, with altitude, does not significantly
Maximum affect the concentration of agent delivered (V/V), but decreases the
permissible flow: 20 L/min partial pressure of the agent in the ratio of the atmospheric pressure to
Liquid capacity: 375 mL working volume the calibrated pressure of 760 mm Hg. To compensate for the reduction
250 mL refill indicator of vapor pressure output at altitude, the rotary valve must be advanced
50 mL reserve to maintain the required agent partial pressure.
Flow resistance < 50 cm H2O at 5 L/min O2 20

at 21°C: < 250 cm H2O at 15 L/min O2 18


Dial Setting for Required Output (V/V%)

Datex®, Ohmeda® and other trademarks are property of Instrumentarium Corp. or its subsidiaries. All other product and company names are property of their respective owners.
Calibration: Calibrated for the following v/v% levels at 15
s s
ter ter l
760 mm Hg and nominal 21°C using O2 at me me eve
000 000 ea L
2 1 S
5 L/min as the carrier gas: 10
1,2,3,4,5,6,7,8,9,10,12,14,16,18
Flow characteristics 5

The graph shows the nominal output of the Tec 6 Plus Vaporizer
at 760 mm Hg.
Effect of Flowrate at 21 deg. C 0 38 76 114 152

AN3307-A/1100 © 2000 Datex-Ohmeda Division Instrumentarium Corp. All rights reserved. Subject to change without notice.
19 Required Output (Vapor Pressure mmHg)
18
17 18
16 Warranty
15
14 Warranted to
Warranted to be
be free
free from
from functional
functionaldefects
defectsinin materials
materialsand
and
13
12
workmanshipfor
workmanship for aa period
periodof
of 12
24months
monthsfrom
fromthethedate
dateofoforiginal
original
% Desflurane Output

12
delivery.
delivery.
Dial Setting %

11
10
9
8
7 7
6
5 5
4
3 3
2
1 1

0 1 2 3 4 5 6 7 8 9 10

Flow (Liters/Minute O2)

Datex-Ohmeda Division • Instrumentarium Corp.


P.O. Box 900, FIN-00031 Datex-Ohmeda, Finland
Tel. +358 10 394 11 • Fax +358 9 146 3310

www.datex-ohmeda.com