Picosecond laser

USER MANUAL

1.Safety Precautions

Note: The person who operates or maintains this system must read this
manual thoroughly before operating this system. In this case, the laser beam
will cause serious injury to the operator, the patient and the surrounding
environment if improperly handled, which may result in burning, blindness,
fire, explosion, etc. In order to ensure safety
Please note the following safety issues:

Warning:
ND: YAG laser is invisible infrared light, the wavelength of 1064nm, all laser
operating area personnel (including patients) must wear protective glasses!

Warning:
The laser window is located at the end of the articulated arm outlet!
Avoid exposure to laser radiation from this window!

1. Do not look directly at the end of the articulation arm and any reflected
laser beam.
2 flammable, explosive narcotic drugs, liquids, gases (such as alcohol, ether,
oxygen) should be the principle of laser radiation area, and protective
measures.
3. Metal devices can cause laser reflection, surgery should avoid direct
exposure to these devices in the laser
4. On the instrument, and use non-reflective equipment.
5. Operation should be avoided with watches and necklaces, bracelets and
other jewelry can be reflected laser.
6. Must provide patients with strict and effective eye protection such as gauze
cover, wearing protective glasses, face block and so on.
7. The need to treat the affected area and adjust the light used in the target
surface, the use of laser irradiation is strictly prohibited elsewhere; during the
treatment of moving the arm and placed after the treatment of articulated arm,
the arm of the end of the exit must point to a safe place.
8. Laser treatment room entrance should be affixed with laser radiation
warning signs, laser access is strictly prohibited during the use of personnel.
9. Human tissue pigment absorption of different laser, the treatment dose of
choice should be small to large.

Note: Large spot corresponds to a smaller energy density, small spot


corresponds to a larger energy density

(2).Eye protection
The pulse width of laser is 10ns - 1ns. The laser beam emitted from the laser
window (the end of the arm) is converged at a distance of 70mm from the
outlet. The focal length of the focusing lens is 170mm. Without the focal
length lens, the laser output divergence angle of about 2mrad.
Any one of the treatment room, direct, such as intraocular laser enough to
make the eyes vitreous and retinal damage, resulting in long-blind!

Warning:
All personnel in the laser operating area must wear safety glasses!
Patients must be provided with strict and effective eye protection!

A. People in the operating area must wear laser goggles at a wavelength of


200-1100nm.
B. Laser goggles can not be safe protection Direct laser or glass, mirrors,
smooth metal surface reflection of the laser, it is strictly prohibited to wear
protective glasses directly above the direct light and reflected light.
C. Cover with gauze, wear protective goggles, face mask, etc. to protect the
patient's eyes are effective.

(3) Electrical safety The new skin beauty laser machine using 220V power
supply, the maximum output current of not more than 10A. With the new skin
beauty laser machine connected to the network power must meet the national
standard, well-grounded single-phase three-wire power cord blocks,
specifications for the 10A. New skin beauty laser machine running time in the
high-voltage storage capacitor ---- the charging voltage can be as high as
1.5KV.

Note: After turning off the power, high voltage devices may still have residual
voltage, free to open the chassis cover may have a high voltage shock
hazard!

Note: The power cord of the new skin cosmetic laser machine is equipped
with a standard 3-pin GB plug. Make sure that your power socket has a
protective grounding wire and ensure reliable grounding.

(4) Handling safe


New beauty laser machine for the optical-mechanical and electrical integration
of precision instruments, machine does not allow users to move long-distance
use. As the machine is modular design, the whole machine should be
disassembled during transportation. The power supply, laser and joint arm
must be packed and transported separately. During transportation, shock
absorption packaging must be used. Otherwise, the laser system may be out
of balance.

2.Treatment principle:
Rox Anderson and John Parrish, "the principle of selective photothermolysis",
the shorter the laser action time, the target tissue to absorb the accumulation
of laser energy is less likely to spread to the surrounding tissue, the energy
was limited extent In the target treatment needs to protect the surrounding
normal tissue, so the treatment of the stronger the selectivity. Picosecond
laser pulse width is only one hundredth of the traditional Q-switched
nanosecond laser, its excellent performance to ensure that picosecond laser
pulse crushing of pigment particles more thoroughly, while the surrounding
tissue damage

3.The main technical parameters, performance indicators:

(1) Working conditions:

A. Ambient temperature: 10~30℃;

B. Relative humidity: ≤ 75%

C. Atmospheric pressure: 860 ~1060hpa

D. Power supply: AC220V ± 10%, 50Hz ± 20%

E. Cooling water: Use distilled water or pure distilled water.

2. Basic parameters:
755PicoSure Cellular Picosecond Laser
Laser Type: Picosecond Laser
Laser wavelength: 755,1064,532
Power supply: 220V / 110V
The highest energy: 2000mj / 2000mj / 1000mj
Operating frequency: 1-10HZ
Pulse width: 10ms
Power: 1500W
Cooling system: air-cooled
Spot diameter: 2-10mm
Net amount: 40KG

Light energy performance


1. Single pulse energy: Jm ≥ 1500mj
2. Laser output instability: S ≤ ± 15%
3. Laser output energy reproducibility: Rj <10%
4. Output energy and energy display error: ≤ ± 20%
5. Surgery spot diameter: 1-4nm range continuously adjustable
6. Spot positioning focusing accuracy: the error ≤ ± 0.5mm
7. Multi-joint light guide arm adjustment range: ≥ 45 ° in the horizontal plane,
the vertical height from the ground between 60cm - 80cm.

4. Use the program, use the method


(A), installation and use of environmental requirements
1. The instrument should be installed in a separate studio.
2. The instrument room should be clean and dustproof
3. The temperature and humidity of the instrument should be relatively stable.
If it can not meet the requirements of working conditions, the user should be
equipped with air conditioner, dehumidifier and voltage regulator.
(B), the transport equipment
The machine is designed for modular power, laser, light guide arm needs a
separate transport, the user at the scene at the installation and
commissioning after delivery.
(C), equipment installation, commissioning
After the instrument is installed in the fixed, turn on the power according to
frequency, single pulse energy, laser output stability, energy reproducibility,
display error, spot diameter adjustment range, cursor positioning accuracy,
machine function can be adjusted and measured.
(D)Instrument operation
1, on the operator's request:
The operator of the instrument should be familiar with the knowledge of the
laser and be used by trained and qualified personnel
2, display control panel and security functions:

The LCD shows the contents:


Function choose:
1064nm interface:

755nm interface:
Note: The laser window should point to a safe place before turning on the
machine and after drilling the machine

3.Operation steps:
1) Power on
Turn on the power and twist the emergency switch button to the right

Then you can adjust the energy button ---- add and subtract; frequency
wavelength selection
Select the parameters, press the pre-burning button, until the lower left foot
light bulb, you can press the work key (work state), step on the foot switch,
you can light out.

The foot pedal is not allowed to be set, and all the parameters need to be
adjusted when the foot switch is disconnected. Any parameter adjustment
operation is not allowed.

2)Treatment

Set the parameters of the light guide arm output ports need to be treated at
the site, control light guide arm output port and the distance between the
treatment site that is controlled spot diameter size, depress the foot switch for
treatment.
3)Shutdown
First, open the light to stop the foot light, turn the emergency switch to the
right. Turn off the mains.

Note:not operate in accordance with the instructions in this manual, which


can cause dangerous laser radiation!

5.Indications

1)Indications
The instrument is mainly used for clinical tattoo removal, eyeliner, lip line
Epidermis and dermal pigment
Acquired Ota-like spot (both sides are stained)
freckle
Black spot disease
Inflammation after pigmentation
Age spots
Sunburn / simple spots
Coffee spots
2)Treatment interval
Three times as a course of treatment, one for about half a month, so that
crushed pigment particles have enough time to be absorbed.

3)Number of treatments
The number of treatment of specific patients by the concentration of the lesion
pigment concentration, depth, pigment composition, the amount of laser
treatment for each treatment, the patient's macrophage pigment particles on
the absorption of the body Circumstances and other factors. General
exogenous pigment to be more than 1 treatment, endogenous pigment to be
more than 3 times to obtain satisfactory results, a small number of patients
after repeated treatment is still not satisfactory.

6.Side effects of contraindications


1)For patients with deep skin, the use of this instrument treatment, because
the skin of patients with shallow pigment absorption of the laser, the skin will
be slightly damaged, it is recommended that the skin whitening and then use
the instrument treatment.

2)Laser irradiation, the skin will appear redness, subcutaneous oozing and
inflammatory response, postoperative treatment of frozen parts, and the use
of place after burns and scars ointment.
3)Some of the metal ions (such as iron ions) of the pigment, the laser
irradiation will appear not easily absorbed by the black chemicals (such as
ferrous oxide) to deepen the color of the treatment site, at this time due to
different pigment composition caused, and Non-quality of the instrument itself.
4)After treatment, the treatment site should be kept clean and dry, to avoid
infection.
5)After the treatment of knot furuncle, should be the other off their own, to
prevent people from forcibly cut scar.
6)The treatment of the same parts of the prohibition of continuous repeated
exposure, otherwise the accumulation of heat will cause damage to the
organization.

7.Precautions and Explanations


1)The actual use of the parameters of the product has been adjusted good
lock, with special requirements of the components in the factory to be sealed.
Users are not free to unseal, disassemble, adjust and change the parameters
of the lock, otherwise, the consequences caused by the user is responsible
for their own.
2)Key switch, beam stop device, etc., is set for laser radiation safety
protection, the user shall not be any change demolition of these devices, so
as not to affect the protection of security.
3)New laser skin treatment instrument set when the light, machine, electricity
in one of the precision instruments. Therefore, the operator must be trained,
non-trained personnel shall not operate the machine.
4)The course of treatment must pay attention to the laser treatment of non-
treatment site.
8.Maintenance

1)This product consists of precision laser generator and microelectronic


instruments. Its maintenance and repair should be consistent with precision
optical instruments of microelectronic devices General provisions. Storage
and use to prevent bumps, collision, extrusion, vibration. After each use to be
reset, with a special day after the use of packaging materials covered, dust
storage.
2)Clean the output lens: the end of the arm
If there is, please put the arm straight up, remove the handle, with cotton
swab dipped in absolute ethanol or saline gently wipe the lens .

Warning: Do not check and clean the output lens during power-on. 9.
List of product configurations

9.Parts List
Host X1
Joint arm X1
Foot switch X1
Laser protection glasses X1
Power cable X1
User manual X1

What is picosecond laser?


Time units of the conversion relationship:
1 second =103 Millisecond
=106 Microsecond
=109 Nanosecond
=1012 Picosecond

So picosecond laser, that is, each laser pulse duration (pulse width) to
picosecond-level laser.
Rox Anderson and John Parrish, "the principle of selective photothermolysis",
the shorter the laser action time, the target tissue to absorb the accumulation
of laser energy is less likely to spread to the surrounding tissue, the energy
was limited extent In the target treatment needs to protect the surrounding
normal tissue, so the treatment of the stronger the selectivity. Picosecond
laser pulse width is only one hundredth of the traditional Q-switched
nanosecond laser, its excellent performance to ensure that picosecond laser
pulse crushing of pigment particles more thoroughly, while the surrounding
tissue damage less.

755 honeycomb picosecond refers to the picosecond honeycomb laser, using


755 wavelength laser, with honeycomb instantaneous lens will focus 70%
laser energy pool, each hair laser light to provide 10% of the skin tissue
efficient shock wave effect, shattered skin layer Of the melanin can also pass
the shock wave to the deep dermis, stimulate collagen production, accelerate
metabolism, to achieve the purpose of beauty.

Basic information

Name: 755PicoSure Cellular Picosecond Laser


755PicoSure cellular picosecond laser, a new laser treatment, the use of 755
emerald treatment head, 3 times the efficacy of traditional laser, ice star
strongly recommend, with a landmark skin beauty significance.
Full name: 755PicoSure cellular Picosecond laser
Place of birth: United States
Date of birth: December 4, 2012, 755PicoSure beehive picosecond laser to
get the US FDA listing license, to become the world's first to get a "birth
certificate" picosecond laser equipment.
MEETINGS: On September 2, 2015, Picosure was officially approved by the
CFDA, which means that the first picosecond laser equipment has entered the
domestic market.

Scope of application

1)mild, stubborn chloasma treatment


2) deep pigmentary lesions, such as Ota, blue mole
3)superficial pigmented nevus, such as freckles, age spots and so on
4)eliminate all kinds of color tattoos and tattoo, pattern eyeliner and so on
5)shrink large pores, remove small telangiectasia, effectively improve the
auxiliary rejuvenation function such as acne
Advantage

Safe and efficient


Treatment rate than the Q-laser 20% increase, the process safe and
convenient, short course of treatment, saving your skin treatment time, with
the rule with the go;

Professional assurance
755PicoSure cellular picosecond laser is a professional medical beauty
equipment, treatment doctors are trained through the US headquarters
training certification, skin care treatment to ensure the safety, effectiveness,
scientific.

Stable without recurrence


755PicoSure cellular picosecond laser beam soft, will not cause strong
stimulation of the skin, better treatment stability, eliminate the effect of
repeated phenomenon, eliminating the need for repeated treatment time and
costs

Personalized exclusive
Covering the world's powerful skin beauty case data, multi-wavelength, multi-
pulse width, multi-head adjustment, different skin types of treatment more
targeted;
Accurate freckle
No skin cooling, 755PicoSure cellular picosecond laser "selective
photothermal effect" technology can intelligently detect the depth of pigment
blocks directly against the root of a hit that, while the normal skin will not
cause any impac

Lasting effect
One can quickly effect dilute spots, simple melasma cure rate of 99% years of
clinical treatment results show that, 755PicoSure cellular picosecond laser on
more than 85% of the stain treatment effect is outstanding, no recurrence.
No side effects
Integration of the most authoritative experts and the use of the United States
"safety cap" technology, the dual advantage of freckle energy density
uniformity, not easy to cause skin bleeding, will not produce pigmentation and
hypopigmentation and so on.
Treatment faster
5 minutes to complete a treatment, with the rule with the walk, easy and
convenient! Each time interval shortened to 7-15 days, greatly improving the
efficiency of treatment.
99.9% satisfaction rate
Safe freckle at the same time the perfect rejuvenation, senior skin beauty
experts Zuozhen, thousands of US 755PicoSure hives to accept the United
States Picosecond laser treatment for the US survey feedback, satisfaction
reached 99.9%.
Precautions
Preoperative
1)Preoperative attention continued sunscreen, Eat less greasy spicy food, to
maintain a good mood.
2)After the first 3 to 4 days of surgery, pay attention to "repair" work, try to use
non-irritating facial cleanser to keep the affected area clean, fresh, and rub
some doctors issued repairing ointment, eat oral anti-inflammatory drugs,
antibiotics .
3)A week later, the wound will start scab off, this time the sun is absolutely
essential steps to go out to wipe SPF20 to 30 sunscreen, sun at least 3 to 6
months.
4)Have any abnormal conditions have to find your doctor asked, do not
arbitrarily call the shots, so as to avoid irreversible situation.
5.755PicoSure cellular Picosecond laser beauty is definitely not an ordinary
life beauty, technology and equipment have very high requirements, so be
sure to go to the formal professional plastic surgery hospital.

Postoperative
1)Use 755PicoSure honeycomb picosecond laser freckle treatment, the skin
by the energy of the burst, so there may be swelling of the phenomenon, but
this is normal, and sometimes some slight heat, but soon subsided.
2)In the 755PicoSure cellular picosecond laser freckle every day to wipe anti-
inflammatory cream, the wound will gradually form scab, about a week after
peeling, when the skin was pink. In this process should pay attention to keep
clean, and avoid being sun, otherwise it will affect the 755PicoSure cellular
Picosecond laser after freckle effect.
3)755PicoSure cellular Picosecond laser if the pigment has not yet subsided
after the freckle, you do not have to worry about, about 2 months or so can be
the next treatment.

Whether it will rebound


755PicoSure cellular picosecond laser freckle is a combination of a variety of
light sources in a high-tech revolutionary freckle technology. In fact,
755PicoSure cellular picosecond laser technology to achieve more than twice
the effect of traditional photon therapy, in-depth damage to tissue treatment,
755PicoSure cellular Picosecond laser freckle effect so that the skin condition
will be more rapid improvement.
755PicoSure cellular Picosecond laser freckle treatment is to start from the
damaged cells, the treatment is generally not rebound after the completion.
755PicoSure cellular picosecond laser technology, can penetrate the skin
deep, selective role in the subcutaneous pigment or blood vessels,
decomposition of color, close the abnormal bloodshot, lifting your skin on a
variety of blemishes, while 755PicoSure cellular Picosecond laser freckle
effect Stimulate subcutaneous collagen proliferation, resistance to pigment
formation.
755PicoSure cellular picosecond laser freckle skin beauty instrument, light
energy on the skin epidermis photothermolysis and biological stimulation, to
achieve the elimination of all kinds of stain, bloodshot, hair removal.

You might also like