Download as pdf or txt
Download as pdf or txt
You are on page 1of 12

lnvited Review

Performance of Subcutaneously Implanted


Glucose Sensors: A Review
Martijn Gerritsen, MD, and
John A. Jansen, DDS, PhD ABSTRACT Despite a considerable amount of research attributed to
Department of Biomaterials, the development of an implantable glucose sensor, to date there is no
J Invest Surg Downloaded from informahealthcare.com by McGill University on 02/19/13

University of Nijmegen, clinically applicable concept for continuous glucose monitoring. Investi-
Nijmegen, The Netherlands
gations to validate the subcutaneous tissue for continuous glucose sens-
Alexander Kros and ing mostly comprise short-term implantations of glucose sensors. Most
Roeland 1. M. Nolte, PhD implanted glucose sensors showed a significant decay in sensitivity over
Department of Organic the implantation period. This bioinstability was not to be expected from
Chemistry, University of the in vitro performance of the sensors. In this article, the influence of
Nijmegen, Nijmegen, possible failure mechanisms on the poor in vivo performance of subcu-
The Netherlands
For personal use only.

taneously implanted glucose sensors is reviewed.


Jos A. Lutterman, MD, PhD
Department of Internal Medicine,
KEYWORDS continuous glucose sensing, implantable glucose sensor, subcuta-
neous tissue
University Hospital St. Radboud,
Nijmegen, The Netherlands

P
atients with diabetes mellitus have an increased morbidity and
mortality due to specific and nonspecific complications. Recently
the Diabetes Control and Complications Trial Research Group
has shown that there is a clear effect of intensive treatment of insulin-
dependent diabetes mellitus on the development and progression of
specific complications such as diabetic retinopathy, nephropathy, and
neuropathy [l].
Received 12 May 1998; Intensive treatment comprises several blood glucose measurements a
accepted 13 May 1998.
day with subsequent adjustment of the insulin dosage. Even with in-
This research was supported by the
Dutch Technology Foundation STW, tensive treatment, the number of measurements is still limited and will
the applied science division of the only provide information about blood glucose values at intermittent mo-
Dutch Science Foundation NWO, and
the technology program of the ments. Furthermore, intensive treatment for diabetes increases the risk
Ministry of Economic Affairs. of severe hypoglycemia. As a consequence, continuous glucose sensing
Address correspondence t o John A. could result in a more adequate insulin administration and could also
Jansen, DDS, PhD, Department of be of use in the early detection of hypoglycemia.
Biomaterials, Dental School, University
of Nijmegen, P.O. Box 9101,6500 HB Many investigators have already investigated the possibility of contin-
Nijmegen, The Netherlands. uous in vivo glucose monitoring by an implantable glucose sensor, using
E-mail: J.Jansen@dent.kun.nl.
a wide range of different approaches. Despite this considerable effort,
Journal o f Investigative SurgerK currently there is no clinical application available. The subcutaneous
11:163-174.1998
Copyright 0 1998 Taylor 81Francis tissue is regarded as the most appropriate site of implantation, because
0894-1939198 512.00 + .OO of good accessability for surgery and relatively easy replacement of the

163
Invited Re view

sensor in case of impaired function. However, de- of oxygen that is consumed by the oxidation of glu-
spite good in vitro sensor performance, it was ob- cose. The hydrogen peroxide-detecting sensor, most
served that subcutaneously implanted glucose sen- frequently, has a membrane containing glucose oxi-
sors showed a significant decay in sensitivity over the dase immobilized to an electrode. Glucose and oxy-
implantation period. Various explanations have been gen diffuse into the membrane, where the enzymatic
brought forward, but in general there was no struc- reaction produces hydrogen peroxide. This in turn
tural approach to assess the contributions of differ- diffuses to the electrode, where the oxidation of hy-
ent failure mechanisms to the functional instability drogen peroxide to oxygen produces an electrical
of implanted sensors. In this artic:le, the influence current proportional to the glucose concentration.
of possible failure mechanisms on the in vivo beha- Important advantages of this type of glucose sensor
vior of subcutaneously implanted glucose sensors is are easy fabrication and the possibility of construct-
reviewed. ing small sensors. The oxygen detecting sensor also
J Invest Surg Downloaded from informahealthcare.com by McGill University on 02/19/13

has a membrane containing immobilized glucose ox-


DEFINING GLUCOSE SENSORS idase coupled to an electrochemical oxygen sensor.
Glucose and oxygen diffuse through a membrane
Chemical sensors are miniaturized devices that and react, resulting in a reduction of the amount
detect a chemical compound selectively and reversi- of oxygen at the sensor site. The signal current is
bly. The detection results in a concentration-depen- related to a reference electrode without glucose ox-
dent electrical or optical signal. Chemical sensors idase. An advantage of the oxygen-based enzyme
that use biological molecules, such as enzymes, are electrode is the low electrochemical interference ow-
For personal use only.

called biosensors [2]. For a sensor to function as a ing to the fact that by using a nonporous hydropho-
device for continuously measuring a chemical com- bic membrane, only gaseous molecules can reach
pound at an intracorporal implantation site, certain the electrode. Also, hydrogen peroxidase could be
characteristics have to be met. These hnctional re- co-immobilized in the membrane, resulting in less
quirements can all be related to the biocompatibility glucose oxidase denaturation.
of the sensor, indicating the ability of the implant to Instead of oxygen, second-generation amperomet-
perform with an appropriate host response in a spe- ric glucose sensors make use of artificial electron car-
cific situation [3]. This means that the sensor has to riers to shuttle the electrons from the enzyme to the
be specific for the compound to be measured, needs electrode. Ferrocene and its derivatives are the most
to retain its stability for prolonged periods of time in commonly used mediators in in vivo experiments.
tissue, and has to be able to detect rapid fluctuations The oxidation of these mediators can be performed
in concentration of the compound at the implanta- at lower potentials than that of hydrogen peroxide,
tion site. thus lowering electrochemical interference.
Several techniques are available to design an im- Third-generation amperometric glucose sensors
plantable glucose sensor. The most frequently ap- are based on the principle of direct electron transfer
plied technique is based on the use of immobilized between the enzyme and the electrode. Conduct-
glucose oxidase, a redox enzyme, in an electrochem- ing organic salts or polymers have been used in the
ical sensor. In nature, this enzyme catalyzes the ox- construction of these electrodes. In this case, no co-
idation of glucose by molecular oxygen, producing substrates such as oxygen or mediators are required.
gluconolactone and hydrogen peroxide. Employing However, no subcutaneous implantation studies
this principle, three generations of electrochemical have been published using sensors based on this
glucose sensors have been developed. These are cat- principle.
egorized based on the mechanism of charge transfer
between enzyme and electrode [4]. SENSOR DESIGN
So-called first-generation glucose sensors estimate
the glucose concentration by measuring the amount The construction characteristics of the subcuta-
of hydrogen peroxide that is produced or the amount neously implanted glucose sensors as found in the

164 M. GERRITSEN ET AL.


lnvited Review

TABLE 1 Construction of glucose sensors


Group Enzyme Inner Outer
[ref] Mediator Configuration Electrodes Isolation immobilization membrane membrane
Bessman [63] Oxygen Two electrodes Galvanic Nylon, GA None PP
Shichiri Peroxide Needle Pt, Ag, Ag Glass C-di-A None PUIPVA
[26,51,52, Ferrocene Needle Pt, Ag Resin Carboxaldehyde PU PVAI
68-71,741 MPCcoBMA
Fischer Peroxide Cannula Pt, Ag Resin Sepharosel PEICA CNPUIPE
[24,25,27, HSA, GA
36,44,53,
59,76,77]
Pickup Peroxide Cannula Pt, Ag Polyester CA PU
[28,72,73] Ferrocene Cannula Pt, Ag Epoxy Sepharose None C
Ferrocene Needle Gr, Ag Epoxy None PU
J Invest Surg Downloaded from informahealthcare.com by McGill University on 02/19/13

Moussy Peroxide Needle, skin Pt, Ag Vernace PPD Nafion


[14-161 button
Pfeiffer Peroxide Cannula Pt, Ag Epoxy CA, GA None PU
[22,38,75]
Updike Peroxide Plastic housing Pt, Ag, Pt Epoxy PU None Bioprotective
[10-12] membrane
Reach, Peroxide Wireheedle Pt, Ag Epoxy CNGA None PU/CA nafion
Wilson catheter BSNPBQ
[29-34,37,
42,43,
For personal use only.

62,781
Johnson [35] Peroxide Cannula Pt, Ag, Pt Polyimide BSA, GA None PU
Wilkins [I31 Peroxide Cannula, Pt, Ag, Pt Carbon None PCInafion
rechargeadle
Gough Oxygen Cylinder Pt, Ag, Pt SA, GA
[17,791 tissue chamber
Ward [9] Peroxide Disk shaped Pt, Ag Epoxy BSA, GA None PU (15 4 0 % )
Matthews Ferrocene Needle None
POI
Heller Redox Wire Pt, Ag, Allylamine, PVD None PC
[81,821 polymer gold aziridine
Koudelka Peroxide Planar Pt, Pt, Ag Epoxy BSA, GA None PU
[39-411 microcell
Clark [83] Peroxide Catheter Pt, Ag Cyanoacrylate GA A C
Vadaama Peroxide Needle Pt, SS Epoxy Polyether None PU
[6,56] copolymer
Note. PP, polypropylene; PE, polyethylene; PU, polyurethane; PC, polycarbonate; C, cellulose; A, acetate; GA, glutaraldehyde; (B)(H)SA, (bovine)
(human) serum albumin; PVA, polyvinylalcohol; PPD, poly(o-phenyldiamine); Gr, graphite; PVD, polyvinylpyridine; 55. stainless steel; PBQ, paraben-
zoquinone; MPC-co-BMA, 2-methcryloyloxethyl phosphorylcholine-co-n-butyl methcrylate.

literature are shown in Table 1. As can be seen, most resin was most frequently used for electrode insula-
glucose sensors used in subcutaneous implantation tion. Enzyme immobilization (glucose oxidase alone
studies were electrochemical enzyme electrodes ba- or mixed with bovine serum albumin) was often ac-
sed on the detection of hydrogen peroxide. Oxygen- complished by glutaraldehyde cross-linking. Inner
based enzyme electrodes and ferrocene-mediated membranes between the working electrode and en-
sensors were less frequently employed. A schematic zyme layer are known to reduce electrochemical in-
drawing illustrating the general layout, which is ap- terference at the electrode [2,5]. Remarkably, in most
plicable to many glucose sensors, is shown in implant designs no inner membrane was used.
Figure 1. In the majority of sensors two electrode Most investigators used an outer interfacial mem-
transducers were used, with platinum as working elec- brane to improve the performance and lifetime of
trode and Ag/AgCl as reference electrode. Epoxy their subcutaneously implanted glucose sensors.

SUBCUTANEOUS GLUCOSE SENSORS 165


Invited Review

used to enhance the performance of the implanted


sensors. For example, a rechargeable glucose sen-
sor has been developed that makes it possible to
replace immobilized glucose oxidase with fresh en-
zyme [13]. In another investigation, the implanted
sensors were connected to a skin button connector
to prevent infection and manipulation of the sen-
sor [14-161. The use of a tissue chamber to sup-
port the growth of vascularized subcutaneous tissue
around the implanted sensor was also described [ 171.
Figure 2 shows a percutaneous device as carrier of a
cannula-shaped implantable glucose sensor.
J Invest Surg Downloaded from informahealthcare.com by McGill University on 02/19/13

FIGURE 1 Aschematic drawing showing the general layout


of an implantable glucose sensor. a = anode; b = cathode;
c = electrode insulation; d = inner membrane; e = glucose
oxidase; f = outer membrane.
For personal use only.

Such an outer membrane has several functions [2,


5-71, First, it is regularly used for entrapment of glu-
cose oxidase at the electrode, thus preventing leak-
age of the enzyme. This is called microencapsula-
tion. Furthermore, its selective permeability slows
the transport of proteins and interferents to the elec-
trode. By restricting glucose diffusion to the elec-
trode, the linearity of the response is increased, and if
the membrane is more permeable to oxygen than to
glucose, the oxygen dependence will also be reduced.
Finally, this membrane is in direct contact with the
subcutaneous tissue and therefore it is an important
determinant for the biocompatibility of the sensor.
Membranes made of cellulose acetate, polycarbon-
ate, polyvinylalcohol, polyurethane, and the perflu-
orinated ionomer Nafion were most commonly used
for these purposes.
Considering the actual configuration of the im-
planted sensor, needle- or cannula-shaped sensors
were most popular. Microfabricated planar thin-film
cells [8] and disc-shaped sensors [9] were also used.
In only three studies, the glucose sensor was fully
implanted together with a transmitter system in an at-
tempt to develop a continuously implantable biosen-
sor [lo-121. FIGURE 2 (a) A cannula-shaped implantable glucose sen-
In addition to variations in Sensor CO~struCtion sor. (b) A percutaneous device in a New Zealand white rabbit
and configuration, various design approaches were as carrier of the sensor.

166 M. GERRITSEN ET AL.


Invited Review

TABLE 2 Operating characteristics of glucose sensors


~~~ ~

In vitro In vivo
sensitivity sensitivity
Group Species Sterilization Duration Calibration (nA/mM) (nA/mM)
Bessman Dog Ethylene oxide 5 days In vitro 50%
Shichiri Dog/human Ethylene oxide 3-7,14 days In vitro 0.21-0.93, 1.4 0.30-0.69, 103%
Fischer Dog UV light, none 3-96 h In vitro/in vivo 0.20-1.70 0.40-3.48
Pickup Humadpig 2-4,8 h In vitroh vivo 0.4, 2780 0.05
Moussy Dog Dry heat 10-1 4 days In vitro 6-25 3-1 0
Pfeiffer Humadsheep Propanolol 7h In vitro 0.7 0.2-1.5
Updike Dog Thimerosal 20-144 days In vivo/in vitro 0.3 245%
Reach/ Rat/dog/human 2 h-10 days In vivo 0.9-3.6, 0.3-1.5, 4
Wilson 1000-3000
Johnson RabbiUhuman 30,72 h In vitro/in vivo
J Invest Surg Downloaded from informahealthcare.com by McGill University on 02/19/13

Wilkins Sheep lh In vitro 500


Gough Rat Glutaraldehyde 10 days In vitro
Ward Rat 1.5 h In vitro 0.043, 0.24 0.028, 0.080
Matthews Rat/human y-Irradiation 4.5, 6 h In vitro 40-50
Heller Rat/dog Ethylene oxide Hours In vitro/in vivo 0.2-0.3 0.15-0.2
Koudelka Rat y-Irradiation 3 h-8 days In vivo 1.5-3.1 0.6-2.0
Clark Cat/dog 4h In vitro
Vadgama Rat 4h In vitro
For personal use only.

IN VlVO EVALUATION 100% of the plasma glucose concentration. Over-


OF IMPLANTED SENSORS all, the subcutaneous glucose concentration in the
various studies closely followed alterations in blood
The in vivo operating characteristics of the glucose glucose values during steady-state conditions and
sensors as listed in Table 1 are shown in Table 2. oral glucose tolerance tests. However, rapid changes
This table demonstrates that glucose sensors have in blood glucose concentration, as induced by in-
been implanted in humans and a number of animal sulin and glucose infusions, were associated with de-
species like rat, rabbit, dog, pig, and sheep. Notice lays ranging from less than 1 min to approximately
that different procedures were employed for steriliza- 45 min.
tion of the sensors before implantation, like ethylene Frequently, in vitro calibration of the sensor prior
oxide, propanolol, thimerosal, dry heat, and y-irra- to implantation and after explantation was used to
diation. Occasionally, no sterilization was performed interpret the sensor current during implantation. Less
prior to implantation. Needle-shaped glucose sen- frequently, one- or two-point in vivo calibration pro-
sors based on hydrogen detection were most fre- cedures were used to relate variations in sensor cur-
quently used in subcutaneous implantation studies. rent to changes in glycemia. Almost all implanted
The sensors remained in situ for periods ranging glucose sensors generally showed a significant decay
from 1 h up to 144days. The majority of sensors, in sensitivity during the implantation period. By in
however, only functioned for several hours to sev- situ calibration or replacement, the lifetime of the
eral days. sensor was prolonged, but mostly the duration of
In vivo sensitivities to glucose ranged from as low sensor implantation did not even exceed the wound
as 0.2 to over 100% compared to in vitro values healing period. Further, taking into account that dif-
and, in general, were lower. Absolute sensor cur- ferent processes involved in sensor implantation, like
rents, if mentioned at all, ranged from 28 pA/mM the wound healing process and the host response to
to 50 nA/mM. Subcutaneous tissue glucose concen- the implant, are likely to cause a decay in sensor
trations ranged between approximately 20 and over sensitivity, the physiological relevance of the sensor

SUBCUTANEOUS GLUCOSE SENSORS 167


Invited Review

signal during this initial period has to be questioned. knowledge of the toxicity and carcinogenicity of the
In the next section the influences of possible failure new materials used in third-generation glucose sen-
mechanisms are discussed. sors also limited their in vivo application.

FAILURE MECHANISMS
Calibration Procedures
In general, subcutaneous implantation of a glu-
For the final clinical application, the output of a
cose sensor resulted in a gradual decrease in sensi-
subcutaneously implanted glucose sensor has to be
tivity, and eventually in a complete loss of sensor
related to the actual glucose concentration in the
function within hours or days. In this section, pos-
subcutaneous tissue. Frequently, an in vitro calibra-
sible contributions in sensor failure related to the
tion technique is used prior to implantation and after
properties of implanted sensors, calibration proce-
explantation to interpret the sensor current during
J Invest Surg Downloaded from informahealthcare.com by McGill University on 02/19/13

dures, healing phenomena, and interfacial reactions


implantation. In this procedure, it is assumed that
around subcutaneously implanted glucose sensors
the basal sensor current and the sensitivity of the
are addressed.
sensor are identical under both in vitro and in vivo
Sensor Properties conditions. This is not realistic, because we know
that after implantation sensor performance changes
The observed decay in sensor sensitivity after im- [ 11 16,22,23].
plantation can be related to the design of the sen- Since the final aim is to use a glucose sensor for
sor [4,18-211. The main problems associated with
For personal use only.

long implantation periods, frequent in vitro calibra-


hydrogen peroxide detection are limitation of the tion is no solution. Consequently, so called one- or
enzyme reaction due to an oxygen deficit, and elec- two-point in situ calibration methods have been de-
trochemical interference by small endogenous mole- veloped [10,11,16,24-421. In the one-point calibra-
cules. Oxygen limitation can for the greater part be tion procedure the sensor output at a certain time
avoided by implantation of a reference oxygen sen- afier implantation is related to the simultaneously
sor, reduction of the reaction area in relation to the measured plasma glucose level, resulting in an in
oxygen permeation area, the use of membranes with vivo sensitivity coefficient. In the two-point calibra-
high oxygen solubility, or selective reduction of glu- tion method the plasma glucose level and sensor out-
cose diffusion to the enzyme by a hydrophobic outer put reach a new plateau following insulin or glucose
membrane. Electrochemical interference can be re- infusion. It is possible to calculate an in vivo sen-
duced to a large extent by the use of selective mem- sitivity coefficient from the sensor currents during
branes or nonconducting electropolymerized films. the two steady-states. The obtained in vivo sensitiv-
The major disadvantage of oxygen-based enzyme ity coefficient is then used to determine an apparent
electrode is that the sensor output depends on the subcutaneous glucose concentration from the sensor
ambient oxygen concentration at the implantation output and to estimate its variations during changes
site. Implantation of a reference oxygen sensor is of blood glucose concentration.
needed to correct for these fluctuations. This, in turn, Generally, the two-point in vivo calibration pro-
leads to problems in the construction and miniatur- cedure is considered to be the most reliable [43].
ization of the sensors. However, since this method requires the induction of
A serious problem associated with second-genera- blood glucose alterations, its feasibility for daily clin-
tion glucose sensors is mediator leakage and toxi- ical practice can be questioned. As a consequence,
city. Improved immobilization of the mediator by other techniques like linear regression analysis of
adsorption or covalent binding to the electrode, or paired plasma/sensor current values have been pro-
entrapment in a polymer matrix or albumin matrix posed, which would not require induced adjust-
are possible solutions. This limited the application ments [44]. The clinical practability of all these cal-
of these sensors for glucose sensing in vivo. Lack of ibration methods is still unclear.

168 M.GERRITSEN ET AL.


Invited Review

Healing Phenomena Around and the proliferation of fibroblasts, leading to col-


Implanted Devices lagen synthesis. These events mark the start of the
repair phase. This form of repair continues until the
The subcutaneous implantation of a sensor will defect is closed. The tissue formed is called granu-
always result in a tissue response. This response is lation tissue. Later, remodeling of the wound takes
a combination of the healing process of the wound place, in which a new balance between collagen syn-
and the host response to the implant [3,45-471. The thesis and lysis is reached. Myofibroblasts are respon-
healing process comprises two phases, an inflamma- sible for reducing the wound surface through wound
tory phase followed by a repair phase. contraction. Finally, cell numbers will decrease, leav-
The inflammatory phase, which takes about 3 days, ing scar tissue behind.
starts after infliction of the wound. Implantation of The presence of an implant prevents a return to
the sensor damages cells, blood vessels, and con- the original situation. It can provide a continuous
nective tissue, initiating a number of healing mech-
J Invest Surg Downloaded from informahealthcare.com by McGill University on 02/19/13

inflammatory stimulus and result in a prolonged in-


anisms. Initially, the wound bed fills with blood, flammatory phase, associated with increased cellu-
followed by activation of the coagulation system. lar activity. However, it is still possible to achieve
Platelets bind to exposed collagen leading, to throm- a chronic phase in which no progressive biological
bus formation. The complement system is activated changes occur. This is called resolution [31. The time
by exposed collagen and extracellular ATP, induc- required to enter this steady state varies according
ing vasodilatation, changes in blood flow, and an to the host site and species, the trauma imposed
increase of the vascular permeability [48]. Th’is re- during implantation, the variables of normal wound
For personal use only.

sults in the formation of an extracellular exudate at healing, and the chemical composition, micro- and
the implantation site containing plasma proteins and macrostructure, and mechanical properties of the
other mediators. By increasing the vascular perme- implant. The possible ways by which resolution is
ability, the compliment system also facilitates the achieved are extrusion, resorption, integration, and
migration of phagocytic leukocytes to the damaged encapsulation. Encapsulation is the usual response
tissue [47,48].These cells, mainly polymorphonu- of the host to an implant. The fibrous capsule is
clear granulocytes and monocytes, are attracted and maintained by the presence of the implant and its
activated by complement mediators and chemotac- thickness is influenced by several factors, like chem-
tic substances released by platelets and damaged cells. ical properties of the implant material, shape of the
They start phagocytosis of injured tissue fragments implant, and mechanical factors at the tissue-implant
and microorganisms. Polymorphonuclear granulo- interface. Implanted electrodes-for example, glu-
cytes have a life span of only a few days in the tis- cose sensors-are known to additionally influence
sue. They are end-state, nondividing cells and obtain capsule formation through electrical currents,
their energy primarily from the anaerobic metabo- changes in local pH and oxygen tension, or the re-
lism of stored glycogen. This makes them able to lease of corrosion products [3]. The influence of
persist in areas with highly disturbed metabolic implant-host interactions and the host response on
states. After fulfilling their function, they die rapidly. the performance of implanted glucose sensors is dis-
The presence of large numbers of live neutrophils cussed next.
in tissue is evidence of continued inflammatory
challenge.
The monocytes differentiate into macrophages Interfacial Reactions Around
[47]. The activated macrophage is not an end-state Subcutaneously Implanted
cell; it has an aerobic glycolysis and is able to un- Glucose Sensors
dergo a number of transformations. Macrophages
phagocytize injured tissue, debris, and bacteria by
Initial Events
releasing degrading enzymes. Furthermore, they re- The initial step in the complex cascade of interac-
lease substances that stimulate neovascular growth tions between the implant and the host is contact of

SUBCUTANEOUS GLUCOSE SENSORS 169


Invited Review

the sensor with blood and extracellular tissue fluid. in implantation studies, this problem needs further
Biological molecules from the exudate that forms consideration.
around the implant, mainly proteins, are attracted to
the implant-tissue interface. Because of their chem-
ical heterogeneity, proteins have the tendency to
lntermediate Events
adsorb at interfaces [23,46,49,50]. This results in cov- In the inflammatory phase of the wound healing
erage of the implant surface within seconds to min- response, the output of implanted glucose sensors
utes, followed by a competitive exchange of different could be influenced by the actions of inflammatory
proteins, called remodeling. Mostly, the adsorbed cells. After attachment to the surface of the implant,
layer does not exceed a monolayer. In protein ad- these cells release various reactive oxygen radicals in
sorption from blood this is called the Vroman effect. the so-called respiratory burst. This is followed by
In general, the composition of the adsorption layer is the secretion of a multitude of proteolytic enzymes,
J Invest Surg Downloaded from informahealthcare.com by McGill University on 02/19/13

strongly dependent on the characteristics of the sur- resulting in a decrease of the tissue pH [48,57]. The
face of a particular sensor [23,49]. To what extent this adsorption of cells and proteolytic enzymes on the
deposition acts as a diffusive barrier for glucose is un- sensor surface could prevent glucose from enter-
known. The influence of the remodeling process on ing the sensor. Another possibility is that these
sensor performance is also unclear. Postimplantation products may damage components of the sensor,
in vitro evaluation frequently showed sensor sensitiv- like glucose oxidase, and impair sensor function-
ities similar to preimplantation values [9,11,14-16, ing [23]. O n the other hand, glucose oxidase has
22,26-28,32,36,37,40,41,43,5 1-54]. This indicates been shown to be very resistent to proteolysis and
For personal use only.

that the formation ofa protein layer does not prevent pH changes [58].
glucose diffusion to the sensor. However, it cannot The observation that most explanted glucose sen-
be not excluded that the removal of the sensor from sors are still capable of glucose detection could also
the implantation site disrupted the integrity of the indicate that low concentrations of analyte were
adsorption layer and, in this way, influenced sensor present at the implantation site. Inflammatory cells,
performance. Occasionally, the in vitro sensitivity mainly monocytes and macrophages, are known to
of explanted probes was lower than before implan- alter the glucose concentration of the exudate fluid
tation, but returned gradually in buffered glucose at the implantation site [59]. The damage inflicted
solutions [11,16,221. This could have been caused on the subcutaneous tissue upon implantation can
by a gradual desorption of adsorbed molecules from also result in a limitation of the blood supply to
the sensor surface [23]. the sensor surroundings, and can cause a decrease in
Open-flow microperfusion has been used to im- sensor output through poor availability of glucose
prove sensor stability. This technique uses slow flow and oxygen. Sensor output was shown to drop to
of isotonic phosphate buffer over the sensor tip to approximately background levels in the first days of
reduce electrode fouling by proteins. It was demon- implantation, after which the output gradually rose
strated that upon implantation only a minor reduc- to stabilize at a new level. This increase was attributed
tion in output occurred, which was not progressive. to the development of sufficient blood supply to the
After implantation, the in vitro sensitivitywas found tissue surrounding the sensor [ 10,113.
to be equal to the sensitivitybefore implantation, in- The rate of glucose oxidation is also dependent
dicating little or no surface fouling [55,56]. upon the availability of oxygen. For every oxygen
Thrombus formation due to damage of small cap- molecule in the interstitial fluid there will be an ex-
illaries could further inhibit glucose diffusion by cov- cess ofglucose molecules present [60]. This can cause
erage of the sensor surface [16,17]. Manipulations of oxygen limitation of the enzyme reaction. Insuffi-
the sensor during implantation can cause repetitive cient blood supply, resulting in a lack of oxygen at
small hemorrhages at the implantation site, with a the implantation site, is likely to further increase the
subsequent drift in sensor signal [8,22,38]. Since the oxygen dependence ofglucose oxidase based sensors.
needle-type sensor configuration is frequently used In in vitro experiments an oxygen independence of
170 M. GERRITSEN ET AL.
Invited Review

the implantable sensors was shown down to oxygen [ 10,16,17,29,67]. Also, sufficient oxygen tensions
tensions that are thought to be present in the subcu- have been reported to exist within a fibrous capsule
taneous tissue. The influence of tissue oxygenation [10,16]. In one study, a layer of granulation tissue
on in vivo sensor performance has not been studied with numerous small vessels was found surrounding
extensively. Occasionally, implantable oxygen sen- the sensor several weeks afier implantation, indicat-
sors were used to investigate the oxygen dependence ing a persistent inflammatory reaction [ l l ] . From
of implanted sensors. In general, it was found that these results it can be assumed that sufficient vas-
the output of the sensors was not influenced by cularization of the fibrous capsule is necessary for a
alterations in oxygen tension at the subcutaneous rapid response of encapsulated glucose sensors. Im-
implantation site [ 17,60-621. Nevertheless, Bessman proved sensor performance is needed to assess the
et al. [63] noticed a distinct influence of tissue oxy- physiological relevance of this response.
gen tension on sensor readings and stated that all
J Invest Surg Downloaded from informahealthcare.com by McGill University on 02/19/13

glucose electrodes based on glucose oxidase will re-


quire correction for oxygen tension by a reference PROSPECTS
oxygen electrode. The fact that second-generation
Despite a considerable effort, there is no clinical
glucose sensors, where output is not influenced by
concept available for continuous subcutaneous glu-
variations in oxygen tension, also showed a decay in
cose monitoring. In general, good in vitro results
sensitivity upon implantation indicates that oxygen
were achieved with differently designed implantable
limitation is not the only problem that has to be
glucose sensors. Upon implantation, however, these
overcome.
devices showed a progressive loss of sensor function.
For personal use only.

This can be regarded as poor bioperformance of the


Late Events
implanted sensors. The fact that the in vitro function
In the repair phase of the healing process, the for- of explanted sensors was ofien unaffected indicates
mation of a fibrous capsule around the sensor cre- that the environment surrounding the sensor plays
ates a diffusional barrier to glucose. This is thought an pivotal role in the behavior of implanted sensors.
to influence sensor response time, dependent on the Questions concerning the influences of wound heal-
thickness of the capsule [64]. Further, it can be ques- ing, host response to the implant, and structure and
tioned whether the transsudate within the fibrous blood supply of the surrounding tissue on sensor
capsule surrounding implanted sensors is at all in dy- performance need to be addressed systematically. In
namic equilibrium with glycemia and provides phys- addition, more insight is needed in the physiological
iologically relevant data [7,65]. By using a difhsion processes at the sensor-tissue interface.
model, Sharkawy et al. [66] estimated that the for- The achievement of this knowledge implies an in-
mation of a fibrous capsule will increase the lag-time terdisciplinary approach at a basic scientific level.
of a sensor to detect 95% of the blood analyte con- In vitro experiments can reproduce many elements
centration from approximately 20 min (no fibrous of the in vivo situation and should be used more
capsule) to 60 min (with a 200-pm-thick fibrous cap- often to bridge the gap with implantation studies. In
sule). Clearly, this estimated increase is dependent vivo experiments can then be reserved for the more
on the fibrosity and vascularity of the capsule sur- promising approaches. Animal models seem more
rounding the sensor. suitable in this respect than human subjects, since in-
In only a few studies the performance of implanted terventions and histological processing are more
glucose sensors was related to the structure of the easily performed. Only with a better understanding
subcutaneous tissue surrounding implanted sensors. of the processes involved in sensor inactivation is it
Relatively thin, well-vascularized fibrous capsules possible to develop adequate strategies to improve in
were found, and a rapid adjustment of the glucose vivo sensor performance. It also seems worthwhile to
concentration in the interstitial fluid within the cap- simultaneously investigate possibilities of influenc-
sule to blood glucose changes was demonstrated ing the tissue reaction to implanted devices.

SUBCUTANEOUS GLUCOSE SENSORS 171


Invited Review

REFERENCES cose and oxygen sensors. J Biomed Eng. 1989; 11:362-


368.
1. Diabetes Control and Complications'TrialResearch Group. 18. Cammann K. Implantable electrochemical glucose sen-
The effect of intensive treatment of diabetes on the de- sors: state of the art. Horm Metab Res Suppl. 1988;20:4-
velopment and progression of long-term complications 8.
in insulin-dependent diabetes mellitus. N Engl J Med. 19. Gough DA, Armour 1C. Development of the implantable
1993;329:977-986. glucose sensor: what are the prospectsandwhy is it taking
2. Eggins B. Biosensors: An Introduction. Chichester: Wiley so long? Diabetes. 1995;44: 1005-1 009.
and Teubner; 1996. 20. Wilkins ES. Towards implantableglucose sensors: a review.
3. Black J. Biological Performance of Materials, Fundamen- J BiomedEng. 1989;11:354-361.
tals o f Biocompatibility 2nd ed. New York: Marcel Dekker; 21. Fischer U. Fundamentals of glucose sensors. Diabetic
1992. Med. 1991;8:309-321.
4. Wilkins E, Atanasov P. Glucose monitoring: state of the art 22. Kerner W, Kiwit M, Linke B, Keck F, Zier H, Pfeiffer
and future possibilities. Med Eng Phys. 1996; 18(4):273- EF. The function of a hydrogen peroxide-detecting elec-
288. troenzymatic glucose sensor is markedly impaired in hu-
5. Pickup JC, Thevenot DR. European achievements in sensor man subcutaneous tissue and plasma. Biosens Bioelectc
J Invest Surg Downloaded from informahealthcare.com by McGill University on 02/19/13

research dedicated to in vivo monitoring. In Advances in 1993;8:473-482.


Biosensors, supplement 1. London: JAI Press, 1993:201- 23. Fraser DM. Biosensors in the Body Continuous In Vivo
225. Monitoring. Chichester: John Wiley & Sons; 1997.
6. Pickup JC. In vivo glucose monitoring: sense and sensor- 24. Abel P, Muller A, Fischer U. Experiencewith an implantable
bility. Diabetes Care. 1993; 16(2):535-539. glucosc sensor as a prerequisite of an artificial beta cell.
7. Kyrolainen M, Rigsby P, Eddy S, Vadgama P. Bio-/ Biomed Biochim Acta. 1984;43:577-584.
haemocompatibility: implications and outcomes for sen- 25. Abel P, Fischer U, Brunstein E, Ertle R. The GOD-HLOZ-
sors? Acta Anaesthesiol Scand. 1995;39(suppl. 104):55- electrode as an approach to implantable glucose sensors.
60. Horm Metab Res Suppl. 1988;20:26-29.
For personal use only.

8. Koudelka M, Rohner-Jeanrenaud F, Terrettaz J, Bobbioni- 26. Shichiri M, Kawamori R, Yamasaki Y. The develop-
Harsch E, de Rooij NF, Jeanrenaud B. In vivo behaviour ment of artificial endocrine pancreas. Med Prog Techno/.
of hypodermicallyimplanted microfabricatedglucose sen- 1987; 12:39-50.
sors. Biosens Bioelectc 1991;6:3 1-36. 27. Rebrin K, Fischer U, von Woedtke T, Abel P, Brunstein
9. Ward WK, Wilgus W, Troupe JE. Rapid detection of hyper- E. Automated feedback control of subcutaneous glu-
glycaemia by a subcutaneously implanted glucose sensor cose concentration in diabetic dogs. Diabetologia. 1989;
in the rat. Biosens Bioelectr. 1994;9:423-428. 32:573-576.
10. Updike SJ, Shults MC, Rhodes RK, Gilligan BJ, Luebow JO, 28. Pickup JC, Shaw GW, Claremont DJ. In vivo molec-
von Heimburg D. Enzymatic glucose sensors: improved ular sensing in diabetes mellitus: an implantable glu-
long-term performance in vitro and in vivo. ASAIO J. cose sensor with direct electron transfer. Diabetologia.
1994;40:157-163. 1989;32:213-217.
11. Gilligan BJ, Shults MC, Rhodes RK, Updike SJ. Evaluation 29. Moatti-Sirat D, Capron F, Poitout V, Reach G, Bindra
of a subcutaneous glucose sensor out to 3 months in a DS, Zhang Y, Wilson GS, Thevenot-DR. Towards con-
dog model. Diabetes Care. 1994; 17:882-887. tinuous glucose monitoring: in vivo evaluation of a
12. Shults MC, Rhodes RK, Updike SJ, Gilligan BJ. Reining miniaturized glucose sensor implanted for several days
WN. A telemetry-instrumentation system for monitoring in rat subcutaneous tissue. Diabetologia. 1992;35:224-
multiple subcutaneously implanted glucose sensors. I€€€ 230.
Trans Biomed Eng. 1994;41:937-942. 30. Moatti-Sirat D, Vehlo G, Reach G. Evaluating in vitro and
13. Wilkins E, Atanasov t? Integrated implantable device in vivo the interference of ascorbate and acetaminophen
for long-term glucose monitoring. Biosens Bioelectr. on glucose detection by a needle-type glucose sensor.
1995; 10:485-494. Biosens BioelectE 1992;7:345-352.
14. Moussy F, Harrison DJ, Rajotte RV. A miniaturized Nafion- 31. Moatti-Sirat D, Poitout V, Thome V, Gangnerau MN,
based glucose sensor: in vitro and in vivo evaluation in Zhang Y, Hu Y, Wilson GS, Lemonnier F, Klein JC, Reach G.
dogs. Int J Artif Organs. 1994; 17:88-94. Reduction of acetaminophen interference in glucose sen-
15. Moussy F, Harrison DJ, O'Brein DW, Rajotte RV. Perfor- sors by a composite Nafion membrane: demonstration in
mance of subcutaneously implanted needle-type glucose rats and man. Diabetologia. 1994;37:610-616.
sensors employing a novel trilayer coating. Anal Chem. 32. Poitout V, Moatti-Sirat D, Reach G, Wilson GS, Lemonnier
1993;65:2072-2077. F, Klein JC. A glucose monitoring system for on line esti-
16. Moussy F, Jakeway S, Harrison JD, Rajotte RV. In vitro mation in man of blood glucose concentration using a
and in vivo performance and lifetime of perfluorinated miniaturized glucose sensor implanted in the subcuta-
ionomer-coated glucose sensors after high-temperature neous tissue and a wearable control unit. Diabetologia.
curing. Anal Chem. 1994;66:3882-3888. 1993;36:658-663.
17. Ertefai S, Gough DA. Physiological preparation for study- 33. Poitout V, Moatti D, Vehlo G, Reach G, Sternberg R,
ing the response of subcutaneously implanted glu- Thevenot DR, Bindra DS, Zhang Y, Wilson GS. In vifro

172 M. GERRITSEN ET AL.


Invited Review

and in vivo evaluation in dogs of a miniaturized glucose and Applications. Washington, DC: American Chemical
sensor. ASAIO Trans. 1991;37:M298-M300. Society; 1995.
34. Poitout V, Moatti-Sirat D, Reach G. Calibration in dogs 51. Shichiri M, Kawamori R, Hakui N, Asakawa N, Yamasaki
of a subcutaneous miniaturized glucose sensor using a Y, Abe H. The development of wearable-type artificial
glucose meter for blood glucose determination. Biosens endocrine pancreas and its usefulness in glycaemic con-
Bioelectr: 1992;7:587-592. trol of human diabetes mellitus. Biomed Biochim Acta.
35. Johnson KW, Mastrototaro JJ, Howey DC, Brunelle RL, 1984;5:561-568.
Burden-Brady PL, Bryan NA, Andrew CC, Rowe HM, Allen 52. Shichiri M, Yamasaki Y, Nao K, Sekiya M, Ueda N. In vivo
DJ, Noffke BW, McMahan WC, Morff RJ, Lipson D, Nevin characteristics of needle-type glucose sensor: measure-
RS. In vivo evaluation of an electroenzymaticglucose sen- ments of subcutaneous glucose concentrations in human
sor implanted in subcutaneous tissue. Biosens Bioelectr: volunteers. Horm Metab Res Suppl. 1988;20: 17-20.
1992;7:709-7 14. 53. Muller A, Abel P, Fischer U. Continuous monitoring of
36. Von Woedtke T, Fischer U, Brunstein E, Rebrin K, Abel P. subcutaneousglucose concentration using implanted en-
Implantable glucose sensors: comparison between in vitro zyme electrodes. Biomed Biochim Acta. 1986;45:769-
and in vivo kinetics. IntJArtifOrgans. 1991;14:473-481. 777.
37. Velho G, Froguel P, Thevenot DR, Reach G. In vivo calibra- 54. Brunstein E, Abel P, Gens A, Eich K, von Woedtke T. Prepa-
J Invest Surg Downloaded from informahealthcare.com by McGill University on 02/19/13

tion of a subcutaneous glucose sensor for determination ration and validation of implantable electrodes for the
of subcutaneous glucose kinetics. Diabetes Nutr Metab. measurement of oxygen and glucose. Biomed Biochim
1988;3:227-233. Ada. 1989;48:91 1-917.
38. Pfeiffer EF, Meyerhoff C, Bischof F, Keck FS, Kerner W. On 55. Rigby GP, Crump P, Vadgama P. Open flow microperfusion:
line continuous monitoring of subcutaneous tissue glu- approach to in vivo glucose monitoring. Med Biol Eng
cose is feasible by combining portable glucose sensor with Comput. 1995;33:231-234.
microdialysis. Horm Metab Res. 1993;25: 12 1-1 24. 56. Rigby GP, Crump P, Vadgama P. Stabilizedneedleelectrode
39. Koudelka M, Rohner-Jeanrenaud F, Terrettaz J, Bobbioni- system for in vivo glucose monitoring based on open flow
Harsch E, de Rooij NF, Jeanrenaud B. In vivo response microperfusion. Analyst. 1996; 12 1:871-875.
For personal use only.

of microfabricated glucose sensors to glycaemia changes 57. Labow RS, Erfle DJ, Santerre JP. Neutrophil-mediated
in normal rats. Biomed Biochim Acta. 1989;48:953- degradation of segmented polyurethanes. Biomaterials.
956. 1995;16:51-59.
40. Koudelka M, Rohner-Jeanrenaud F, Terrettaz J, Bobbioni- 58. Wilson R, Turner APF. Glucose oxidase: an ideal enzyme.
Harsch E, de Rooij NF, Jeanrenaud B. In vivo behaviour Biosens BioeIectc 1992;7:165-185.
of hypodermicallyimplanted microfabricated glucose sen- 59. Rebrin K, Fischer U, Hahn von Dorsche H, von Woetke
sors. Biosens Bioelectr: 1991;6:31-36. T, Abel P, Brunstein E. Subcutaneous glucose monitoring
41. Bobbioni-HarschE, Rohner-JeanrenaudF, Koudelka M, de by means of electrochemical sensors: fiction or reality?
Rooij N, Jeanrenaud B. Lifespan of subcutaneous glucose J Biomed Eng. 1992; 14:33-40.
sensors and their performancesduring glycaemia changes 60. Updike SJ, Shults M, Ekman B. Implanting the glucose en-
in rats. J Biomed fng. 1993; 15:457-463. zyme electrode: problems, progress, and alternative solu-
42. Sternberg R, Barrau MB, Gangiotti L, Thevenot DR, Bindra tions. Diabetes Care. 1982;5(3):207-2 12.
DS, Wilson GS, Vehlo G, Froguel P, Reach G. Study and de- 61. Fischer U, Hidde A, Herrmann 5 , von Woedtke T,
velopment of multilayer needle-type enzyme-based glu- Rebrin K, Abel P. Oxygen tension at the subcutaneous im-
cose microsensors. Biosenson. 1988;4:27-40. plantation site of glucose sensors. Biomed Biochim Acta.
43. Velho G, Froguel P, Thevenot DR, Reach G. Strategies 1989;48:965-971,
for calibrating a subcutaneous glucose sensor. Biomed 62. Zhang Y, Wilson GS. In vitro and in vivo evaluation of
Biochim Acta. 1989;48:957-964. oxygen effects on a glucose oxidase based implantable
44. Von Woedtke T, Rebrin K, Fischer U, Abel P, Wilke W, Vogt glucose sensor. Anal Chim Acta. 1993;281:513-520.
L, Albrecht G. In situ calibration of implanted electrochern- 63. Bessman SP, Thomas LJ, Koijma H, Sayler DF, Layne EC. The
ical glucose sensors. BiomedBiochimActa. 1989;48:943- implantation of a closed-loop artificial beta cell in dogs.
952. Trans Am Soc Artif Intern Organs. 1981;27:7-17.
45. Park JB, Lakes RS. Biomaterials, An Introduction. 2nd ed. 64. Reach G, Feijen J, Alcock A. BIOMED concerted action
New York: Plenum Press; 1992. chemical sensors for in vivo monitoring. The biocompati-
46. Williams DF. Tissue-biomaterial interactions. J Mater Sci. bility issue. Biosens Bioelectc 1994;9(6):21-28.
1987;22:3421-3445. 65. Woodward SC. How fibroblasts and giant cells encapsu-
47. Spector M, Cease C, Tong-Li X. The local tissue response late implants: considerations in design of glucose sensors.
to biomaterials. Crit Rev Biocompatibility 1989;5(3):269- Diabetes Care. 1982;5:278-281.
295. 66. Sharkawy AA, Klitzman B, Truskey GA, Reichert WM. En-
48. Remes A, Williams DF. Immune response in biocompati- gineering the tissue which encapsulates subcutaneous
bility. Biomaterials. 1992;13(11):731-743. implants. 1. Diffusion properties. J Biomed Mater Res.
49. Brash JL, Wojciechowski PW. Interfacial Phenomena and 1997;37:401-412.
Bioproducts. New York: Marcel Dekker; 1996. 67. Chang KW, Aisenberg 5 , Soeldner JS, Hiebert JM. Val-
50. Horbett TA, Brash JL. Proteinsat Interfaces: fundamentals idation and bioengineering aspects of an implantable

SUBCUTANEOUS GLUCOSE SENSORS 173


Invited Review

glucose sensor. Trans Am Soc Artif Intern Organs. measurement of subcutaneous glucose concentrations
1973; 19:352-360. with an enzymatic glucose sensor and a wick method.
68. Shichiri M, Yamasaki Y, Kawamori R, Hakui N. Wear- Klin Wochenschr. 1989;67:491-495.
able artificial endocrine pancreas with needle-type glu- 76. Fischer U, Ertle R, Abel P, Brunstein E, Hahn von Dorsche H,
cose sensor. lancet. 1982;20:1129-1130. Freyse EJ. Assessment of subcutaneous glucose concen-
69. Shichiri M, Asakawa N, Yamasaki Y, Kawamori R, Abe tration: validation of the wick technique as a referencefor
H. Telemetry glucose monitoring device with needle-type implanted electrochemicalsensors in normal and diabetic
glucose sensor: a useful tool for blood glucose monitoring dogs. Diabetologia. 1987;30:940-945.
in diabetic individuals. Diabetes Care. 1986;9:298-301. 77. Fischer U, Ertle R, Rebrin K, Freyse EJ. Wick technique:
70. Shichiri M, Kawamori R, Goriya Y, Yamasaki Y, Nomura reference method for implanted glucose sensors. Artif
M, Hakui N, Abe H. Glycaemic control in pancreatec- Organs. 1989; 13:453-458.
tomized dogs with a wearable artificial endocrine pan- 78. Vehlo G, Froguel P, Sternberg R, Thevenot DR, Reach G.
creas. Diabetologia. 1983;24: 179-1 84. In vitro and in vivo stability of electrode potentials in
71. Nishida K, Sakakida M, lchinose K, Uemura T, Uehara M, needle-type glucose sensors. Influence of needle mate-
Kajiwara K, Miyata T, Shichiri M, lshihara K, Nakabayashi rial. Diabetes. 1989;38:164-171.
J Invest Surg Downloaded from informahealthcare.com by McGill University on 02/19/13

N. Developmentof a ferrocene-mediated needle-typeglu- 79. McKean BD, Gough DA. A telemetry-instrumentation sys-
cose sensor covered with newly designed biocompatible tem for chronically implanted glucose and oxygen sensors.
membrane, 2-methacryloyloxyethyl phosphorylcholine- IEEE Trans Biomed Eng. 1988;35:526-532.
co-n-butyl methacrylate. MedProg Techno/. 1995;21:91- 80. Matthews DR, Bown E, Beck TW, Plotkin E, Lock L, Gosden
103. E, Wickham M. An amperometric needle-type glucose
72. Claremont DJ, Sambrook IE, Penton C, Pickup JC. Sub- sensor tested in rats and man. DiabeticMed. 1988;5:248-
cutaneous implantation of a ferrocene-mediated glucose 252.
sensor in pigs. Diabetologia. 1986;29:8 17-821. 81. Quinn CP, Pishko MV, Schmidtke DW, lshikawa M,
73. Pickup JC, Claremont DJ, Shaw GW. Responses and Wagner JG, Raskin P, Hubbell JA, Heller A. Kinetics of
calibration of amperometric glucose sensors implanted glucose delivery to subcutaneous tissue in rats measured
For personal use only.

in the subcutaneous tissue of man. Acta Diabetologia. with 0.3-mm amperometric microsensors. A m J Physiol.
1993;30:143-148. 1995;269:E155-E161.
74. Shichiri M, Kawamori R, Hakui N, Yamasaki Y, Abe H. 82. Linke B, Kerner W, Kiwit M, Pishko M, Heller A. Amper-
Closed-loop glycaemic control with a wearable artificial ometric biosensor for in vivo glucose sensing based on
endocrine pancreas: variations in daily insulin require- glucose oxidase immobilized in a redox hydrogel. Biosens
ments to glycaemic response. Diabetes. 1984;33: 1200- BioelectK 1994;9: 15 1-1 58.
1202. 83. Clark LC Jr, Duggan CA. Implanted electroenzymatic glu-
75. Bruckel J, Kerner W, Zier H, Steinbach G, Pfeiffer EF. In vivo cose sensors. Diabetes Care. 1982;5(3): 174-1 80.

174 M. GERRITSEN ET AL.

You might also like