Professional Documents
Culture Documents
MT500
MT500
Published by
ZERA GmbH
Hauptstr. 392
53639 Königswinter Germany
Tel No +49 2223 704 0
Fax No +49 2223 704 70
www.zera.de
Copy rights © 2015 ZERA GmbH, All rights reserved.
Version: 8th December 2015
MT500_MAN_EXT_GB_V401
Moving Test MT500 2
MT500_MAN_EXT_GB_V401
Moving Test MT500 3
Table of Figure
Figure 1 Front panel of MT500................................................................................................................. 10
Figure 2 Connection terminal of MT500 ............................................................................................... 12
Figure 3 Side view of MT500 .................................................................................................................... 13
Figure 4 Start up screen (for example only) ....................................................................................... 15
Figure 5 Upper Menu Bar ........................................................................................................................... 15
Figure 6 Main Menu ...................................................................................................................................... 16
Figure 7 Setting Menu ................................................................................................................................ 17
Figure 8 Settings (2) Menu ........................................................................................................................ 18
Figure 9 Connection for Single Phase Meter Testing ..................................................................... 22
Figure 10 Connection for 3 Ph 4 wire ct/vt operated meters ...................................................... 23
Figure 11 Connection for 3 Ph 3wire ct/vt operated meters........................................................ 24
Figure 12 Connection for 3 Ph 4 wire ct/vt operated meters ...................................................... 25
Figure 13 Connection for 3 Ph 3wire ct/vt operated meters ...................................................... 26
Figure 14 Connection for 3 Ph 4 wire direct connected meters .................................................. 27
Figure 15 Connection for 3 Ph 3 wire direct connected meters .................................................. 27
Figure 16 Connection for 3 Ph 4 wire direct connected meters .................................................. 28
Figure 17 Actual Value display................................................................................................................. 32
Figure 18 Actual Value Display (Total Parameter) ........................................................................... 34
Figure 19 Total distortion ............................................................................................................................ 34
Figure 20 Vectorial display menu IEC 387 .......................................................................................... 35
Figure 21 Vectorial display menu as per DIN 410 ............................................................................ 35
Figure 22 Vectorial display in Delta mode ........................................................................................... 36
Figure 23 Vectorial display connection (open delta) ....................................................................... 36
Figure 24 Source Menu 4 wire mode .................................................................................................... 37
Figure 25 Source Menu ON ...................................................................................................................... 39
Figure 26 Source Menu OFF .................................................................................................................... 39
Figure 27 Source Menu, Save Settings ................................................................................................ 40
Figure 28 Dosage (time) ............................................................................................................................. 41
Figure 29 Curve Display ............................................................................................................................. 42
Figure 30 Harmonic measurement (Tabular Format) ..................................................................... 43
Figure 31 Harmonic measurement (Graphical Format) ................................................................. 44
MT500_MAN_EXT_GB_V401
Moving Test MT500 4
1.1 Warning
In order to avoid electrical shock or fire or damage to instrument and installations, pay attention
on below given information.
Don’t apply the voltage more than specified voltage rating for test voltage terminal (for
MT500 it is 300 V phase to neutral), for power supply adapter 265 V phase to neutral.
Remove all test leads and accessories that are not in use.
Do not insert metal objects into connectors.
Always use the isolated safety pins for voltage and/ or line having voltage more than 40 V.
Do not use exposed metal pins or banana plug connectors.
Always use the test tools only in the specified manner.
Do not use equipment in environmental condition for which they are not recommended or
specified, like rain, wet, excessive heat and do not leave unattended. Hands, shoes and
floor must be dry while making connections to power line.
Check all cables for breaks or cracks in the insulation before use. Replace immediately, if
defective.
User should wear proper personal proactive equipment, including safety glasses and
insulated gloves when making connection to power circuits.
Ensure that all connections are properly fixed, loose connection may cause sparking,
excessive heat, short circuit and may damage the equipment and/or installation.
MT500_MAN_EXT_GB_V401
Moving Test MT500 5
Red Box with text inside means that if ignored, damage to property or person can
occur.
Blue box with black text provides information, hints and references to further
documentation to use.
The gray box with black text is used in this manual to provide tips
Recycling
High voltage warning
information
MT500_MAN_EXT_GB_V401
Moving Test MT500 6
Abbreviation Description
MT500
MT500
Portable Three Phase Power Source
CT Current Transformer
PF Power Factor
F Frequency
Ø Phase angle
DC Direct current
MT500_MAN_EXT_GB_V401
Moving Test MT500 7
2 Getting Started
2.1 General Overview
The MT500 is a fully electronic three phase power source based on state of the art technology in
test voltage and test current generation for low voltage meter installations. Powerful digital
amplifiers combined with a user friendly operation concept are providing the highest possible
flexibility for a comprehensive load point simulation on metering installations on site.
The power source MT500 is located in a stable housing made of aluminum integrating a 6.4” LCD
display and built-in soft keys at the front panel. Combined with a user-friendly software this device
is especially applicable for operation on-site.
Version: This document is based on firmware version V6.03
2.2 Features
The MT500 is designed to have following easy and user friendly features
Three Phase Power generation up to 300 V/12 A
High accuracy through closed loop feedback control
Balanced/Unbalanced load conditions
Individual phase angle setting
Precise displaying of the generated values
Vector display of generated and measured values
2.3 Functions
The MT500 is designed for the following functions, useful in energy meter testing and analysis.
Power generation up to 12 A current and 300 V voltage for testing of electricity meter
installations (1 ph 2 wire, 3 ph 3 wire & 4 wire circuits)
Measurement and display all instantaneous parameters like delta voltage, star voltage,
current, PF, Ø, F, Power (active, reactive and apparent power), rotary field indication and
total harmonic distortion.
Vector diagram display
Distortion factor measurement
Frequency-, phase angle- and power factor measurement
Definition of load point and phase angles (symmetrical / asymmetrical)
MT500_MAN_EXT_GB_V401
Moving Test MT500 8
MT500_MAN_EXT_GB_V401
Moving Test MT500 9
4 Operation manual 1
MT500, English
For further details about accessories for the MT500 please see external price list.
1
By default calibration of this device will be at 50 Hz. Calibration at 60 Hz will be carried out only at demand or
specific countries where mains frequency is 60 Hz
MT500_MAN_EXT_GB_V401
Moving Test MT500 10
MT500_MAN_EXT_GB_V401
Moving Test MT500 11
Example 1: Select the Actual value indication and then press key to see total powers or
press key to see distortion in voltage and current circuits in each phase.
MT500_MAN_EXT_GB_V401
Moving Test MT500 12
1 2 3 4 5 6 7
11 10 9 8
Part No Details
9 Frequency input
10 RS 232 interface for external control via PC. Pin details for communication
Pin 2 = RXD , Pin 3 = TXD , Pin 5 = GND
11 Barcode/Serial number
MT500_MAN_EXT_GB_V401
Moving Test MT500 13
Press buttons at
both sides
simultaneously
to move handle.
MT500_MAN_EXT_GB_V401
Moving Test MT500 14
Safety
IP class according to DIN EN 60529 IP40
Declaration of conformity CE conform
Protection class according to DIN EN 61140 I
Overvoltage category voltage measurement CAT I 300 V
Overvoltage category current measurement CAT I 300 V
Source
Voltage min. max. 20 V … 300 V~
Voltage range(s) 75 V, 150 V, 300 V
Voltage max. output power 8) 15 VA
Voltage distortion < 0.5 %
Voltage resolution 5 Digits
Voltage accuracy < 0.2 %
Voltage stability 9) < 0.02 %
Voltage load dependence < 0.02 %
Current min. max. 4 mA … 12 A~
Current range(s) 12 A, 6 A, 3 A, 1.2 A, 0.6 A, 0.3 A, 0.12 A, 0.06 A, 0.03 A
Current max. voltage per range 1.3 V (12 A) - 2,7 V (6 A) -
5 V (3 A) - 6V (1,2 A .. 30 mA)
Current max. output power 8) 15 VA
Current distortion < 0.5 %
Current setting resolution 5 Digits
Current accuracy < 0.2 %
Current stability 9) < 0.02 %
Current long term stability < 0.02 %
Current load dependence < 0.02 %
Frequency range 45 … 65 Hz
Frequency accuracy 0.01 Hz
Frequency setting accuracy 0.01 Hz
Phase angle setting range 0.00 … 359.99°
Phase angle setting resolution 0.01°
Phase angle accuracy < 0.1°
Phase angle stability 9) < 0.01 °
8: Related of end of maximum range and end of range and ohmic load 28.05.2015
9: Stability over 1 hour (measurement with ti = 10 s)
Subjects to alteration.
MT500_MAN_EXT_GB_V401
Moving Test MT500 15
1 2 3 4
Figure 5 Upper Menu Bar
MT500_MAN_EXT_GB_V401
Moving Test MT500 16
*optional
Press the Func key then list of menus will appear on display.
Use the arrow key () to go up to Settings menu and then press the Enter key.
Setting menu will appear on display.
MT500_MAN_EXT_GB_V401
Moving Test MT500 17
Press key, select the desired language in pull down menu by using keys and
then press Enter key. MT500 supports the following languages:
German, English, Italian, French, Czech, Spanish, Turkish, Hungarian and Serbian.
Press key and select the angle definition IEC 387 or DIN 410 from pull down menu
by using keys and then press Enter key.
The angle definition will decide how vectorial display will appear. In DIN 410 voltage
vector is fixed (on Y axis) and current vector is moving while in IEC 387 current vector is
fixed (on X axis) and voltage vector is moving. Please refer the IEC 62053-23 Figure C.1
& C.2.
Press key and select the mode of control from pull down menu by using
keys and then press Enter key. You can choose between Dual, Device and
External. If you select Dual you can control the MT500 as well via soft keys as
via external device connected to the interface. Select Device to control the MT500
exclusively via soft keys. If you select External the MT500 can exclusively be
controlled via an external device connected to the interface. Then the soft keys of
the MT500 are locked.
MT500_MAN_EXT_GB_V401
Moving Test MT500 18
Press key then cursor will be in date field. Enter the correct date and time by using
numeric key and press Enter key. Use keys to move cursor left and right.
Press key then cursor will be at S field to set the apparent power calculation for
individual phase apparent power. Choose the calculation between Arithmetic (S= Urms x
Irms) and Geometric (S = P 2 Q 2 ) by using arrow key and press enter key.
After pressing the Enter key cursor will be automatically at ∑S field to set the apparent
power calculation for total apparent power. Choose the calculation between Arithmetic
and Geometric by using arrow key and Enter key.
Please refer calculation of single phase and total apparent power in 5.3.1.3
Press key and select the printer from pull down menu. Press Enter key to
activate the selected printer. The MT500 supports Seiko and Mobile Pro® thermal
printer.
MT500_MAN_EXT_GB_V401
Moving Test MT500 19
The MT500 has a voltage range 300 V but a range selection is not required because the
MT500 selects the best suitable range automatically.
Voltage ranges
300 V
150 V
75 V
Auto
The MT500 has internally the below indicated current ranges but a range selection is not
required because the MT´500 selects the best suitable range automatically.
Current ranges
12 A
6A
3A
1.2 A
600 mA
300 mA
120 mA
60 mA
30 mA
Auto
MT500_MAN_EXT_GB_V401
Moving Test MT500 20
Wrong selection of meter circuit type will affect s wrong load point setting.
MT500_MAN_EXT_GB_V401
Moving Test MT500 21
3 Meter Connection
This chapter explains various connections which may be occur in meter installations. Here the
given connection diagrams are just a few examples and does not necessarily cover all types of
meters around the world.
MT500_MAN_EXT_GB_V401
Moving Test MT500 22
The presented connection diagram here is an example and showing the connection
of the MT500 power source with a portable calibration standard of the MT3xx series.
MT500_MAN_EXT_GB_V401
Moving Test MT500 23
The presented connection diagram here is an example and showing the connection
of the MT500 current source with a portable calibration standard of the MT3xx series.
MT500_MAN_EXT_GB_V401
Moving Test MT500 24
The presented connection diagram here is an example and showing the connection
of the MT500 current source with a portable calibration standard of the MT3xx series.
MT500_MAN_EXT_GB_V401
Moving Test MT500 25
The presented connection diagram here is an example and showing the connection
of the MT500 current source with a portable calibration standard of the MT3xx series.
MT500_MAN_EXT_GB_V401
Moving Test MT500 26
The presented connection diagram here is an example and showing the connection
of the MT500 current source with a portable calibration standard of the MT3xx series.
MT500_MAN_EXT_GB_V401
Moving Test MT500 27
The presented connection diagram here is an example and showing the connection
of the MT500 current source with a portable calibration standard of the MT3xx series.
MT500_MAN_EXT_GB_V401
Moving Test MT500 28
For further information to make settings and to perform the dosage via MT3x0 see
external operation manual MT300 - MT310 - MT320.
MT500_MAN_EXT_GB_V401
Moving Test MT500 29
MT500_MAN_EXT_GB_V401
Moving Test MT500 30
It is necessary to make the following settings in the WinSAM Software to connect the
power source MT500 with MT3000 reference meter via PC properly.
2.
MT500_MAN_EXT_GB_V401
Moving Test MT500 31
4 Software Features
This chapter explains the software functions/menus available in the MT500 to meet the various
requirements on site. The list of menu and their tasks are as follows.
Name of Menu Chapter Tasks
Actual Values 4.1.1 To indicate the instantaneous values of 3 phase network
Vector 4.1.2 To display the vectorial diagram of connection
Curve* 4.2.1 To see the waveform of supply
Harmonics* 4.2.2 To measure harmonics up to 40th harmonics
Source 4.1.3 To set the required load point parameters
Dosage (time) 4.1.4 To perform a dosage test
Settings 2.7.5 To set language, angle definition
Settings (2) 2.7.6 To set date and time, apparent power calculation method
* optional features
MT500_MAN_EXT_GB_V401
Moving Test MT500 32
See hint in
chapter 4.1.3,
page 34
- > Src
MT500_MAN_EXT_GB_V401
Moving Test MT500 33
Angle measurement U, IU UI depends on the angle definition set in Setting
menu.
UI is measured phase angle w.r.t. voltage when DIN 410 is selected, IU
measured phase angle w.r.t. current when IEC 387 is selected.
In case of 3 ph 3 wire mode L2 current and IU, UI , S, of all phases will be not be
displayed
In case of 1 ph 2 wire mode L2, L3 currents and IU, UI , S, of all phases will be
not be displayed
Symbol Description
To see the total parameter like Total Active, Reactive and Apparent power,
average PF, F, Phase sequence.
To see individual phase parameter.
Off . By pressing the OFF key, measurement and current generation will be stopped.
MT500_MAN_EXT_GB_V401
Moving Test MT500 34
In the 3 phase 3 wire mode total apparent power (∑S) will be always geometrically
based.
MT500_MAN_EXT_GB_V401
Moving Test MT500 35
This feature allows the operator to find out a wiring fault or verify the meter connection/network
easily. Operator can select the vectorial display as per IEC 387 and DIN 410.
Make connection of MT500 refer chapter 3.
Make load point setting refer chapter 4.1.3
Press Func. key and then select Vector from list of menu and press Enter key.
Press key to change the angle definition. Select between IEC 387 and DIN 410 from
pull down menu.
MT500_MAN_EXT_GB_V401
Moving Test MT500 36
MT500_MAN_EXT_GB_V401
Moving Test MT500 37
4.1.3 Source
This feature allows the operator to define and to set the required load point for supply to the meter
under test. The operator can select the phase to be set, current value and angles as per IEC 387
and DIN 410.
Make connection of the MT500 refer chapter 3 and follow the instructions below.
Press Func. key and then select Source from list of menu and press Enter key.
If this mode is
indicated, the
dosage mode is
controlled
externally. As a
result the test
current can not Indication of U-
be switched on and I-range
in this menu. differs if only
See also current or only
chapter 7.2 voltage or both
values are
Figure 24 Source Menu 4 wire mode switched on.
Press MM key and then select according the meter to be supplied the correct type of
meter circuit from the list of the menu and press Enter key.
Press key and choose between Symmetry and Frequency by pressing keys
and press the Enter key. If you select Symmetry you can choose between
Symmetrically and Not symetrically from another list of the menu and press Enter key.
In case of Symetrically it is only required to set the values for current and angle in phase
L1. The values for L2 and L3 will change accordingly.
In case of Not symetrically it is possible to set the current, voltage and angle values for
the phases L1, L2 and L3 individually.
After you have chosen between both possibilities, the cursor automatically goes to the
next field. Now select Angle (phase angle) or PF (power factor).
If you have select Frequency after pressing the key, the cursor goes to the field
F:. Select between Synchronously and Variable. If you select Variable you have to
enter the frequency manually. E. g. enter 50 Hz and press the Enter key.
MT500_MAN_EXT_GB_V401
Moving Test MT500 38
If you have selected Not symetrically before, press key to enter the current,
voltage and the phase angle to be generated by using the alphanumeric keys and press
Enter key. After pressing the key one field is indicated with a frame, e. g. U L2 (see
picture above). Now you can select a field/value to be modified by using the cursor keys.
Press the Enter key and insert your values via alphanumeric keys and press the Enter
key again. Choose the next field/value which you want to modify again and so on. If you
have finished your input, press the key again to get back to the source menu.
- > Src
If you have select PF before, instead of phase angle indication you can enter additional
values. In field P/Q choose between P (active power) and Q (reactive power). In field PF
enter the power factor between 0 and 1. In field Q select a quadrant (I, II, III or IV).
Corresponding with the setting you have done before not all 4 quadrant are available.
Press key, if you want to store the entered source values for future use (see also
next chapter.)
Press key and then Enable or Disable the corresponding Phase by using the
Arrow on the soft key pad and press the Enter key.
Press On key and then the MT500 will switch on the load point accordingly. While
load point setting the following message will appear temporarily.
MT500_MAN_EXT_GB_V401
Moving Test MT500 39
After setting of the load point the MT500 will change automatically to the Actual Values
display explained in chapter 4.1.1 in order to display the generated values.
Press Off key and then the MT500 will switch off the load point accordingly. While
load point switching off the following message will appear temporarily.
MT500_MAN_EXT_GB_V401
Moving Test MT500 40
After you have pressed the key, you will get to the sub-menu as follows:
Selected loadpoint
(here: Pos. 1) will
be displayed also in
the main source
menu.
Press the Save key to save the settings on the internal memory card.
Press the Load key to select settings which you have saved before. Select a setting
with the function key (forwards) or (backwards). After that press the [Load]
key and the software changes to the source menu again. The selected setting is indicated
at the red marked field, here: Pos. 1.
MT500_MAN_EXT_GB_V401
Moving Test MT500 41
In case that the range changes over, the measurement has to be stopped and
restarted.
The max. value for dosage energy you can enter is 74.5640 kWh. If you enter a
higher value the MT500 set the value automatically to the max. value when starting
the dosage mode.
Press MODE key and select the dosage mode from the pulldown menu and
press Enter key.
MT500_MAN_EXT_GB_V401
Moving Test MT500 42
desired channel press key then cursor will be automatically at CH-1 field,
select the desired channel from pull down list and press Enter key. Cursor will
automatically move to CH-2 field, select the desired channel, same as CH-1.
MT500_MAN_EXT_GB_V401
Moving Test MT500 43
To see other phases press key and then select the desired phase from pull
down menu. Press Enter key to activate the selection.
Harmonic table will indicate the nth harmonic along with amplitude and phase
angle. By default it will display up to 10th harmonic and to see further press
until desired value appears on table.
To see harmonic in graphical format (bar type), press key. By default it will
display up to 10th Harmonic and to see further press key until desired nth
appears on graph.
MT500_MAN_EXT_GB_V401
Moving Test MT500 44
MT500_MAN_EXT_GB_V401
Moving Test MT500 45
U (1,2,3)
1 N
u i (1,2,3)
2 ui(1,2,3) =Sampled value of the voltage in phase L1,L2, L3
N i 0
N = Total number of sample
5.3 Power
5.3.1 Power in measuring mode 4 wire
MT500_MAN_EXT_GB_V401
Moving Test MT500 46
In Arithmatic mode
S (1,2,3) = U(1,2,3) x I(1,2,3) S(1,2,3) = Apparent power of the phases L1, L2, L3
P(1,2,3) = Active power but also mean power of the phase
In Geometic mode
L1, L2, L3 (see 5.3.1.1)
2 2
S(1,2,3) = P (1,2,3 ) Q (1,2,3 ) ) Q(1,2,3): Reactive power but also mean power of the phase
L1, L2,L3 (see 5.3.1.2)
cos (1,2,3) = (1,2,3) cos (1,2,3) : Active power factor of the phases L1, L2, L3
(1,2,3) : Power factor (see 6.3.1.4.)
The Formula stated are only applicable for sinusoidal variable (kU/I ≈ 1)
MT500_MAN_EXT_GB_V401
Moving Test MT500 47
5.4 Angles
5.4.1 Voltage Angles Upn1Upn
MT500_MAN_EXT_GB_V401
Moving Test MT500 48
7 Trouble Shooting
This chapter explains the common problems which can occur during the operation in the product
life of equipment.
MT500_MAN_EXT_GB_V401
Moving Test MT500 49
Current output can Dosage mode inside the MT500 is still active.
not be activated by Reset by power OFF/ON.
MT500 (source
menu)
MT500_MAN_EXT_GB_V401
Moving Test MT500 50
Cables should be regularly cleaned because dirt will impair or reduce the insulation
properties and can cause the electrical shock or short circuit in wet or humid environment.
Cleaning frequency depends on the quality of environment and dirt. Cleaning can be done
by fine damp cloth. Do not use abrasives, solvents or alcohol. These may damage the
device or remove the text on the tools.
All products from ZERA GmbH, Germany are warranted to the original purchaser against
defective material or workmanship during the warranty period. ZERA GmbH will repair or replace,
at its options, all defective equipment that is returned, freight prepaid, during the warranty period.
There will be no charge for repair provided there is no evidence that the equipment has been
misused, mishandled or abused. The warranty shall not apply to any defect resulting from
improper or inadequate maintenance or handled serviced by unskilled person.
The warranty shall not apply to periodical maintenance, calibration as required by local rules and
regulation. The warranty shall not apply to consumable items like rechargeable batteries and
other components which will naturally drift due to physical properties of the material and uses.
ZERA shall not be liable for any special, direct, indirect, incidental or consequential damages or
losses, including loss of data whether arising from breach of warranty or based on contract, tort,
reliance or any other theory.
This warranty shall remain valid for 12 months from the date of installation or 15 months from
shipment date, whichever is earlier shall be valid.
9.3 Reliability
The information given in the manual has been checked and is believed to be reliable. However
ZERA will not assumed any liability for any misinterpretation or inaccuracies and resulted to any
kind of direct/indirect losses. All data, steps, procedures and explanations given in the manual is
only for information purposes and is subject to change without prior notices. This manual should
not be view as evidence of compliance statement.
MT500_MAN_EXT_GB_V401
Moving Test MT500 51
9.6 Software
The user software, developed by ZERA is at customer disposals to operate test equipment’s. The
software remains in ZERA's proprietary. It is prohibited to copy or transfer to other users.
MT500_MAN_EXT_GB_V401
Moving Test MT500 52
MT500_MAN_EXT_GB_V401