Metrohm - 702 Protocol of Measuring Amplifier Checkl
Metrohm - 702 Protocol of Measuring Amplifier Checkl
Metrohm - 702 Protocol of Measuring Amplifier Checkl
mvp702e051027.doc Page 1 of 3
Checking input Ind 2 (U/mV, pH)
Carry out on Titrino Remarks Carry out on Enter theoreti- Enter actual Difference2 Permitted OK
or on sensor 767 cal value from value from difference
767 Titrino !
1
cover display
1 Screw off cable from sensor, input Place Close cover
'Ind 2' (if no sensor or sensor without sensor in Sensor cable
plug-in head is connected: insert sleeve to socket (5)
6.2150.040 cable from accessories in
767 case to input 'Ind 2')
2 <param>: >measuring parameter
3x <enter>: measuring input
with <select>: select measuring input 2
<enter>
<quit>, <quit> mV value 5:
<meas/hold>: read display (0 mV) ................. mV ..............mV ±2 mV ....
3 Open cover mV value 5:
.................. mV ................. mV ..............mV ±3 mV ....
3
4 Sensor cable Diff. actual values
steps 3 and 4
to socket (4)
actual value of
step 3 ................. mV ..............mV ±2 mV ....
5 Observe Sensor cable mV value 6:
polarity to socket (6)
.................. mV ................. mV ..............mV ±2 mV ....
6 <meas/hold>: MEAS U (********)
7 Remove cable.
Carry out on Titrino Remarks Carry out on Enter theo- Enter actual Test OK
or on sensor 767 retical value value from sequence
from 767 Titrino !
cover display
1 Screw off cable from sensor (if no Place sensor in Close cover
sensor or sensor without plug head sleeve Sensor cable to
is connected: use 6.2150.040 Ca- socket (5)
ble from accessories in 767 case).
sensor cable into socket „Pol“
2 <enter> Polarizer test after a posi-
runs tive run there
(duration about appears:
16 s) polarizer test Test
ok4 passed ..
3 3x <clear>: MEAS U (********) Remove cable.
mvp702e051027.doc Page 2 of 3
Checking temperature (Pt 100/ Pt 1000)
Carry out on Titrino Remarks Carry out on Enter theo- Enter actual Difference2 Permitted OK
or on sensor 767 retical value value from differ- !
from 767 Titrino ence
cover display
1 Press <mode> (several times) until Cover (does not
MEAS appears (outer right-hand matter whether
side) open or closed)
<enter>
with <select>: select T
<enter>: MEAS T (********)
<meas/hold>
2 Use 2x banana cables On 767.0010: °C value:
(6.2150.000) to connect tempera-
ture measuring input of Titrino to Sockets (1) and (2) (Pt 100) → ................ °C ................ °C ...............°C ±0.7 °C .....
767.0010
Sockets (2) and (3) (Pt 1000) → ................ °C ................ °C ...............°C ±0.7 °C .....
Read the theoretical value for the sockets (1) + (3) Sockets (1) and (3) (Pt 1000) → ................ °C ................ °C ...............°C ±0.7 °C
from the white field 1 - 3 on the cover
.....
Remove cables and reconnect sensors.
Please put back to the case the original cables belonging to 767.
1
Wait until display remains stable.
2
The given permitted difference applies for normal room temperatures (20 - 30 °C) and warmed-up instruments. Alternatively these values can be
determined from the technical specifications of the 767 and the Titrino.
3
If the variation is too large please try exchanging the electrode cable
4
In case that 'error 100' appears then please check if 767 is correctly connected to the socket 'Pol'.