Metrohm - 702 Protocol of Measuring Amplifier Checkl

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 3

702, 716, 718, 719, 720, 736, 794:

Protocol of measuring amplifier check


Instrument: 1.7..................Serial / Fabr.-No.: ........................... ID-No (if available): ....................................
Test agent: 2.767.0010 Serial / Fabr.-No.: ........................... ID-No (if available): ....................................
Next Cal.: .......................................... Cal. Master: ................................................
Type: ................................................. Certificate-No.: ...........................................
SCS Cal. Serv. Reg. Nr.: ...................
Check carried out on:.................................. Name: ...................................... Signature:................................
Checked instrument ! yes
meets requirements: ! no Decision: .................................................................................................

Please read the notes on the last page first.

Checking input Ind 1 (U/mV, pH)


Carry out on Titrino Remarks Carry out on Enter theoreti- Enter actual Difference2 Permitted OK
or on sensor 767 cal value from value from difference
767 cover Titrino !
1
display
1 Screw off cable from sensor at input Place Sensor cable
'Ind 1' (if no sensor or sensor without sensor in to socket (5)
plug-in head is connected: insert sleeve
6.2150.040 cable from accessories in
767 case in 'Ind 1')
2 Switch on instrument. Line
<mode> „mode“
appears.
3 Press <mode> (several times) until Close cover
MEAS appears in display.
<enter>: MEAS (?)
with <select>: select U
<enter>: MEAS U (********)
4 <param>: >measuring parameter
3x <enter>: measuring input
with <select>: select measuring input 1
<enter>
<quit>,<quit> mV value 5:
<meas/hold>: read display (0 mV) .............. mV ..............mV ±2 mV ...
5 Open cover mV value 5:
.................. mV .............. mV ..............mV ±3 mV ...
3
6 Sensor cable Diff. actual values
steps 5 and 6
to socket (4)
actual value of
step 5 .............. mV ..............mV ±2 mV ...
7 Observe Sensor cable mV value 6:
polarity to socket (6)
.................. mV .............. mV ..............mV ±2 mV ...
8 <meas/hold>: MEAS U (********)
9 Remove cable.

mvp702e051027.doc Page 1 of 3
Checking input Ind 2 (U/mV, pH)
Carry out on Titrino Remarks Carry out on Enter theoreti- Enter actual Difference2 Permitted OK
or on sensor 767 cal value from value from difference
767 Titrino !
1
cover display
1 Screw off cable from sensor, input Place Close cover
'Ind 2' (if no sensor or sensor without sensor in Sensor cable
plug-in head is connected: insert sleeve to socket (5)
6.2150.040 cable from accessories in
767 case to input 'Ind 2')
2 <param>: >measuring parameter
3x <enter>: measuring input
with <select>: select measuring input 2
<enter>
<quit>, <quit> mV value 5:
<meas/hold>: read display (0 mV) ................. mV ..............mV ±2 mV ....
3 Open cover mV value 5:
.................. mV ................. mV ..............mV ±3 mV ....
3
4 Sensor cable Diff. actual values
steps 3 and 4
to socket (4)
actual value of
step 3 ................. mV ..............mV ±2 mV ....
5 Observe Sensor cable mV value 6:
polarity to socket (6)
.................. mV ................. mV ..............mV ±2 mV ....
6 <meas/hold>: MEAS U (********)
7 Remove cable.

Checking Pol. current


(We check now the built-in polarizer using the diagnosis routines).

Prepare Titrino for the diagnosis procedure:


Switch off mains,
wait for 10 s and then switch on again
depress at once key <9>
(and keep it depressed until the test pattern disappears) diagnose press key 0...9

<7> analog input test 1...5

<5> polarizer test

<enter> dummy resistor 14.3k ?

Carry out on Titrino Remarks Carry out on Enter theo- Enter actual Test OK
or on sensor 767 retical value value from sequence
from 767 Titrino !
cover display
1 Screw off cable from sensor (if no Place sensor in Close cover
sensor or sensor without plug head sleeve Sensor cable to
is connected: use 6.2150.040 Ca- socket (5)
ble from accessories in 767 case).
sensor cable into socket „Pol“
2 <enter> Polarizer test after a posi-
runs tive run there
(duration about appears:
16 s) polarizer test Test
ok4 passed ..
3 3x <clear>: MEAS U (********) Remove cable.

mvp702e051027.doc Page 2 of 3
Checking temperature (Pt 100/ Pt 1000)
Carry out on Titrino Remarks Carry out on Enter theo- Enter actual Difference2 Permitted OK
or on sensor 767 retical value value from differ- !
from 767 Titrino ence
cover display
1 Press <mode> (several times) until Cover (does not
MEAS appears (outer right-hand matter whether
side) open or closed)
<enter>
with <select>: select T
<enter>: MEAS T (********)
<meas/hold>
2 Use 2x banana cables On 767.0010: °C value:
(6.2150.000) to connect tempera-
ture measuring input of Titrino to Sockets (1) and (2) (Pt 100) → ................ °C ................ °C ...............°C ±0.7 °C .....
767.0010
Sockets (2) and (3) (Pt 1000) → ................ °C ................ °C ...............°C ±0.7 °C .....
Read the theoretical value for the sockets (1) + (3) Sockets (1) and (3) (Pt 1000) → ................ °C ................ °C ...............°C ±0.7 °C
from the white field 1 - 3 on the cover
.....
Remove cables and reconnect sensors.
Please put back to the case the original cables belonging to 767.

Hints for the preparation of the measuring amplifier check


• Place Calibrated Reference on bench near sensor. Ensure that light reaches solar cell without interference (no shadows from
cables or accessories). Switch on room lighting if necessary.
• The electrode cap must not necessarily be firmly screwed onto sockets (4), (5), and (6) of the 767; plugging it in is sufficient.
• Do not unplug any sensor at the measuring input 'Ind'. Unplug any other sensors or interconnections to other instruments.
• The exchange unit does not necessarily need to be mounted, but has no influence to the result.
• Please mind also the introduction to the measuring amplifier check that is to be found in the Metrohm home page, under <sup-
port> <validation and QM> <useful forms> (German and English) <introduction>.

1
Wait until display remains stable.
2
The given permitted difference applies for normal room temperatures (20 - 30 °C) and warmed-up instruments. Alternatively these values can be
determined from the technical specifications of the 767 and the Titrino.
3
If the variation is too large please try exchanging the electrode cable
4
In case that 'error 100' appears then please check if 767 is correctly connected to the socket 'Pol'.

mvp702e051027.doc Page 3 of 3