Professional Documents
Culture Documents
Calibration Manual GD FH Tue IR and Buveco 12 1 E
Calibration Manual GD FH Tue IR and Buveco 12 1 E
This instruction shows how to use the Calibration kit 5201013-00A for the range of
ST200/400/600/700 range of detectors as well as Salwico GD FH Tue IR infrared
detectors. The kit contains a regulator with fixed flow suitable for a range of
detectors, a rubber hose and a Flow-cap with a coupling, for connection to the
ST200/600/700 range of detectors.
In addition, a suitable Gas bottle must be supplied with suitable gas-mixture for the
detectors onboard. There are various types available, see separate list.
The Salwico GD FH Tue IR detectors are normally calibrated simply by connecting the hose
from the regulator to the calibration connection at the end of unit. Remove both covering
caps if they are fitted. To avoid unwanted trips and alarms, go to the Central panel and
disconnect outgoing alarm functions. At the detector, open the bottle regulator valve and
start the gas- flow. At the control-panel, see the reading in % LEL. The Salwico GD FH Tue
IR signal output can only be adjusted with the hand-held calibration tool. See separate
instructions!
Pic 2.
The ST200 and ST600-EX detectors for 0-100% LEL Hydrocarbon can be calibrated and
out-signal adjusted. Reading at 0% is 4,0 mA, and full range 100% LEL is 20mA. Tools
required for this operation are:
- A small terminal screw driver, with thin straight tip.
- A Multi-meter with mA (Mili-ampere) capacity.
Connect the Flow-cap to the sensor head as shown in Pic 5, and the measuring tips of the
multi-meter to the test-points in the detector. Plus is red, minus is black. Picture shows a
ST200, the ST600EX has identical connections and calibration, only fitted in an EX proof
housing
Pic 4.
Before gas Calibration is made, a Zero-adjustment should be made. Connect the mA meter
to the two test-points as shown in Picture 5.
Pic 5.
The display readout should be 4,0 mA with normal air / zero gas exposed to the sensor.
Adjust to this reading with the “Zero” pot, marked in red where the screwdriver is held.
4/10
Remove the cover from the detector. The calibration push buttons is situated at the
PCB below the display. In this type of detector no Multi-meter required, the mA
signal is adjusted by pressing the keys.
Zero adjustment:
• Apply Zero gas (or simply clean air, make sure there is no LEL content in the
area before work starts) to the sensor.
• Press SW3 (Enter) to activate the zero menu.
• Press SW3 (Enter) to perform the ZERO calibration. Reading should now
stabilize to zero. One can press SW3 several times and wait a little to make
sure the reading is stabile, half a minute or so.
• Press SW4 (Menu) – The display will return to its standard mode operation
after displaying some numbers, the calibration factor.
6/10
Span calibration:
• Enter the USER password in the detector. See above for Zero adjustment.
• Select the menu E2 for span calibration by pressing SW1 (Down).
• Press SW3 (Enter) to activate the menu.
• Use the SW2 (UP) and SW1 (DOWN) buttons to adjust the display reading
according to the calibration gas concentration. 50% LEL is the standard gas in our
gas bottles, if any other bottle is used the corresponding LEL concentration must be
chosen in the display.
• Press SW3 (Enter) to confirm the calibration gas concentration setting.
• Now apply the calibration gas to the sensor by opening the regulator valve.
• Wait until the display value is stable. (The value does not correspond to the
concentration most likely.) On newly fitted sensor the reading may very well be zero
or very much higher than 50% LEL before the next step is done.
• Press SW3 (Enter) to confirm the SPAN calibration. Now the reading should go to
50% LEL. This can be done several times and it will take up to a minute or more for
the reading to stabilize.
• Remove the calibration gas and wait until the gas has disappeared and reading is
back. Press SW1 (Menu) – The display will, after displaying some figures
corresponding to sensor factor curves, return to its standard mode operation again
and the correct mA signal will be sent out. One can also press SW1 while gas is still
flowing and reading is high, but the detector will then immediately after restart send
out the high mA signal and the central go into alarm condition. This may or may not
be a problem, depending on preparations before the calibration commenced.
The H2S sensor is Electrochemical and measures H2S content by means of chemical
reaction. H2S – Hydrogen Sulphide – is a sticky gas and smells very distinctly from
rotten eggs.Its toxic and dangerous in also rather low concentrations. Normal measuring
range is 0-30 ppm, our standard calibration bottle 5201005-00A contains 20 ppm of gas
in air. For SW2020 a mix-bottle is often used, with H2S 20 ppm as well as N2 99% and
C3H8 50% LEL, it’s the article 5201010-01A. Consilium Replacement article number
for ST400 H2S detectors is 004462-S.
Zero adjustment:
Connect the Multi-meter set to mA as shown in Pic 9, and adjust with the “ZERO”pott
to 4,0 mA signal output.
Pic 9
Span calibration:
To verify which type is in use, make a span calibration as below and verify the
display reading in the central cabinet. If the readout at 14,6 mA is less than 20 ppm a
calibration with 16,8 mA set-point must be done.
Connect the gas-bottle with flow-cap to the sensor head, and open the regulator
valve. When the reading is stabilized, adjust the readout to 14,6 mA (or 16,8 mA)
with the “SPAN” pot. If this reading cannot be reached, the sensor-head should
be replaced. Note that due the sticky nature of the H2S as short hoses as possible
is to be used, and response can be estimated to several minutes in some cases.
8/10
Pic 10
Connect a mA meter to the test-terminals and adjust the readout on the “SPAN” pot to
17,4 mA when normal breathing air is present. In a SW2020 flow-system, make sure the
flow is stabile at about 40 ln/h during this operation.
9/10
Pic 11.
Pic 12.
Typical signs for when lifespan of the O2 sensor is up is drifting values, instable
readings and un-ability to set the 17,4 mA signal even with the SPAN pot in max
position. A new sensor is supplied without the metal housing and must be fitted in the
housing like below. This applies to both H2S and O2 sensors of ST400 type.
10/10
Pic 13.
This calibration instruction is intended for the range of calibration bottles in use
from 2011 from our supplier STG. If older equipment is used the procedure still
applies, and set-points of mA signals etc are identical in the most cases.
For calibration support etc, contact your local Consilium office, or forward your
requests to:
service@consilium.se
For spare part orders and calibration bottle orders, contact your local Consilium
office, or:
spares@consilium.se