Periodic maintenance
Column exchange (optional)
Exchange the MACSQuant Column (# 130-094-458)
Storage Solution (# 130-092-748) every three months. For detailed information refer
Sterile, ready-to-use solution. Used for
overnight and long term storage of the to the user manual.
instrument. Automatically fills fluidics Hydrophobic air filter exchange
upon instrument shut down.
Exchange filters when they come in contact with liquid,
and at least once a year.
Cleaning of uptake needle
The instrument should be turned OFF before needle
cleaning. To prevent contamination or blockages, wipe
Waste collection bottle needle with tissue soaked with 70% ethanol.
If measuring biohazardous material, it is
recommended to add 100 mL of MACS Bleach
Solution (# 130-093-663) to waste container.