MACSQuant® Instrument quick guide

Start-up, shutdown, maintenance


Instrument start-up
1. Ensure the MACSQuant® Instrument main power switch
is in the on-position ‘I’.
2. Tap the touchscreen monitor to activate the
Figure 2: Instrument status bar indicating the MACSQuant Instrument
MACSQuantify™ Software. is in data analysis mode (top) or acquisition mode (bottom)
3. After logging on to MACSQuantify Software, the
instrument is in data analysis mode. Instrument shutdown
4. Prime the instrument (power the lasers, activate fluidics)
1. 
The MACSQuant Instrument can be configured
by clicking the icon and choose ‘Acquisition mode’
to automatically power down the lasers by
(figure 1).
pre-selecting a time under
5. Optical bench requires at least 30 minute warm up Edit > Options (default) > Timers > Standby Timer.
before any fluorescence measurement can be made.
2. 
To manually shutdown the instrument from acquisition
 Note: While the instrument is warming, it is best
mode, click the icon.
practice to perform a flush cycle followed by a clean.
Refer to section ‘Rinse, flush, and clean programs’ Choosing ‘Instrument off’ will initiate a fluidic washing
3. 
below. cycle, fill fluidics with storage solution, power down
the lasers, and shutdown the MACSQuantify Software.
6. Instrument can be monitored by the status bar
(figure 2) or hardware monitor (figure 3). Choosing ‘Data Analysis Mode’ will initiate a fluidic
4. 
For detailed information refer to the user manual. washing cycle, fill fluidics with storage solution, and
power down the lasers. The software will stay active
for data analysis or other tasks.
5. 
Upon full instrument shutdown, the monitor will be
in stand-by mode, until the touchscreen is tapped.
This will initiate the MACSQuantify Software to restart.

Figure 1: Priming instrument

Figure 3: Monitoring of optical bench under View > Hardware


Basic maintenance of the Rinse, flush, and clean programs
The MACSQuant Instrument has several rinsing programs
MACSQuant® Instrument to clean fluidics between samples, to remove partial clogs,
MACSQuant® Buffers and solutions or to decontaminate fluidics after running biohazardous
Three fluidic containers are housed on the MACSQuant material.
Instrument to supply MACSQuant Buffers to the instrument. Rinse program
A fourth container collects the waste. For washing of fluidics after a poor quality sample
(aggregates, dead cells) a rinse program can be chosen by
clicking the icon on the status bar. This program uses
the MACSQuant Running Buffer.
Running Buffer (# 130-092-747)
Sterile, ready-to-use buffer with preservative. Flush cycle
Used as sheath fluid, buffer A or B, and during For clearing fluidics of partial or full clogs a flush cycle can
‘Rinse’ program. be used by right-clicking the icon on the status bar and
choosing ‘Flush’. This 16’ program uses the MACSQuant
Washing Solution to incubate the flow cell and back flush
the solution through the tubing system.
Clean program
For decontamination of the instrument, a clean cycle can
Washing Solution (# 130-092-749)
be performed by right-clicking the icon on the status
Sterile, ready-to-use basic solution bar and choosing ‘Clean’. The program can also be used
containing detergent. Used during ‘Flush’ to remove clogs in combination with a Flush cycle. This
and ‘Shutdown’ programs.
10’ process uses on-screen instructions to walk the user
through the procedure. A tube with 0.5 mL of MACS Bleach
Solution is required.

Periodic maintenance
Column exchange (optional)
Exchange the MACSQuant Column (# 130-094-458)
Storage Solution (# 130-092-748) every three months. For detailed information refer
Sterile, ready-to-use solution. Used for
overnight and long term storage of the to the user manual.
instrument. Automatically fills fluidics Hydrophobic air filter exchange
upon instrument shut down.
Exchange filters when they come in contact with liquid,
and at least once a year.
Cleaning of uptake needle
The instrument should be turned OFF before needle
cleaning. To prevent contamination or blockages, wipe
Waste collection bottle needle with tissue soaked with 70% ethanol.
If measuring biohazardous material, it is
recommended to add 100 mL of MACS Bleach
Solution (# 130-093-663) to waste container.
Are you in need of additional assistance?
Visit www.miltenyibiotec.com/locations to find
your nearest Miltenyi Biotec contact. 130-106-087

Miltenyi Biotec provides products and services worldwide. Visit www.miltenyibiotec.com/local to find your nearest Miltenyi Biotec contact.

MACS, the MACS logo, MACSQuant, and MACSQuantify are registered trademarks or trademarks of Miltenyi Biotec GmbH. Unless otherwise specifically indicated,
Miltenyi Biotec products and services are for research use only and not for therapeutic or diagnostic use. Copyright © 2014 Miltenyi Biotec GmbH. All rights reserved.