GO2 RESULTS REPORT

Bradley S. Lambert, Ph.D – Report for PEEP Performance, LLC

COMPARISON OF GO2 TO CONTROL MOUTH PIECE & NO MOUTHPIECE


Protocol 1 (GXT): Graded exercise cycling test beginning at a workload of 150watts following unloaded 3minute
warm-up. During the test, stages progressed with increasing workloads of 30 Watts every two minutes until the
participant could no longer maintain within 10rpms of their self selected pedal cadence. Gas collection was
continuously collected and measurements of hemodynamics, heart rate, and perceived exertion were recorded
every two minutes during exercise and at 1, 3, and 5 minutes of unloaded recovery at the same self selected
pedal cadence.

Protocol 2 (SSXT): Steady state exercise testing set at a workload equivalent to measured anaerobic/ventilator
threshold assessed during GXT testing. Following an unloaded 3 minute warm-up, participants maintained a
constant pace until exhaustion. Gas collection was continuously collected and measurements of hemodynamics,
heart rate, and perceived exertion were recorded every two minutes during exercise and at 1, 3, and 5 minutes of
recovery.

Comparisons: GO2 vs CONTROL MOUTHPIECE (Battle), GO2 vs NO-MOUTHPIECE, GO2 vs COMBINED


CONTROLS

Analysis: Data were compared at each stage of exercise and each time-point during recovery using a two-tailed
paired samples t-test. To account for potential wide variance in the participant population with regards to aerobic
fitness and training history, a mixed model analysis of covariance (ANCOVA) was used where values were co-
varied on measures taken with the control mouthpiece. FIGURES and CONCLUSIONS are presented for both
types of analyses in this report. NOTE: Conclusions presented in this document represent statistical findings and
are not intended to the potential physiologic mechanisms responsible.
1
Figures: Figures presented here are for data whereby at least one significant measurement was detected.
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for
VO2max expressed as both absolute (LO2*min-1) and
and relative (mlO2*kg*min-1) units. *=Significantly

Summary (VO2max) Different at p<0.05.

GO2 vs CONTROL GO2 vs NO MP GO2 vs COMBINED CONTROL

3.70 3.70 3.50 *


* * 3.40
3.50
VO2max (L*min-1)

3.50 3.30

VO2max (L*min-1)
VO2max (L*min-1)
3.30 3.20
3.30
3.10
3.10
3.00
3.10
2.90 2.90
2.90 2.80
2.70
2.70
2.50 2.70 COMBINED GO2
CONTROL GO2 NO MP GO2 CONTROL

50 46 46
*
*
VO2max (ml*kg*min-1)

(ml*kg*min-1)

(ml*kg*min-1)
48
44 44
46
44 42
42
42
40
VO2max

40

VO2
40
38 38
36 38 COMBINED GO2 2
CONTROL GO2 NO MP GO2 CONTROL
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for VO2
(L/min) measured during graded exercise cycle testing. * =

Summary (VO2 L*min-1)


Significant difference between conditions (Control / GO2) at
the same measurement time-point (p<0.05)

3.8

*
CONTROL GO2
3.3
VO2 (L*min-1)

2.8
*

2.3

1.8

1.3

0.8
VO2 ST1 VO2 ST2 VO2 ST3 VO2 PEAK VO2 REC1 VO2 REC3 VO2 REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for VO2
(L/min) measured during graded exercise cycle testing. * =

Summary (VO2 L*min-1)


Significant difference between conditions (Control / GO2) at
the same measurement time-point (p<0.05)

4.0

*
3.5 NO MP GO2

3.0
VO2 (L*min-1)

2.5

2.0

1.5

1.0

0.5
VO2 ST1 VO2 ST2 VO2 ST3 VO2 PEAK VO2 REC1 VO2 REC3 VO2 REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for VO2
(L/min) measured during graded exercise cycle testing. * =

Summary (VO2 L*min-1)


Significant difference between conditions (Control / GO2) at
the same measurement time-point (p<0.05)

4.0

COMBINED CONTROL
3.5 *
GO2
3.0
VO2 (L*min-1)

2.5

2.0
*

1.5

1.0

0.5
VO2 ST1 VO2 ST2 VO2 ST3 VO2 PEAK VO2 REC1 VO2 REC3 VO2 REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for VO2
(ml/kg/min) measured during graded exercise cycle testing. *

Summary (VO2 ml*kg*min-1)


= Significant difference between conditions (Control / GO2) at
the same measurement time-point (p<0.05)

50
*

45 CONTROL GO2
VO2 (ml*kg*min-1)

40 *

35

30

25

20

15

10
VO2 ST1 VO2 ST2 VO2 ST3 VO2 PEAK VO2 REC1 VO2 REC3 VO2 REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for VO2
(ml/kg/min) measured during graded exercise cycle testing. *

Summary (VO2 ml*kg*min-1)


= Significant difference between conditions (Control / GO2) at
the same measurement time-point (p<0.05)

50.0

*
45.0 NO MP GO2
VO2 (ml*kg*min-1)

40.0

*
35.0

30.0

25.0

20.0

15.0

10.0
VO2 ST1 VO2 ST2 VO2 ST3 VO2 PEAK VO2 REC1 VO2 REC3 VO2 REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for VO2
(ml/kg/min) measured during graded exercise cycle testing. *

Summary (VO2 ml*kg*min-1)


= Significant difference between conditions (Control / GO2) at
the same measurement time-point (p<0.05)

50.0

* COMBINED CONTROL
45.0
GO2
VO2 (ml*kg*min-1)

40.0

35.0

30.0

25.0
*
20.0

15.0

10.0
VO2 ST1 VO2 ST2 VO2 ST3 VO2 PEAK VO2 REC1 VO2 REC3 VO2 REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for %
difference in VO2 (ml/kg/min) measured during graded

Summary (VO2max)
exercise cycle testing * = Significant difference between GO2
and listed conditions below.

QUESTION: Does using the GO2 during GXT have an effect on VO2max?

ANSWER: YES.

10%
*
9%
8%
%Difference In VO2max

7% *
6% 6.69%
*
5% 5.28%
4%
4.02%
3%
2%
1%
0%
GO2 vs. GO2 vs. GO2 vs
CONTROL NO MP COMBINED
CONTROL
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for time
to fatigue (seconds) for the graded maximal cycling test and
during steady state testing at a workload associated with each

Summary (Time to Fatigue)


individual’s anaerobic threshold. *=Significantly Different at
p<0.05.

GO2 vs CONTROL GO2 vs NO MP GO2 vs COMBINED CONTROL

650 600 600


* * *
Time To Fatigue GXT

Time To Fatigue GXT

Time To Fatigue GXT


600 550
550
(Seconds)

(Seconds)

(Seconds)
550 500
500
500
450
450
450
400
400 400 COMBINED GO2
CONTROL GO2 NO MP GO2 CONTROL

1100 1100 1000

Time To Fatigue SSXT


1000
Time To Fatigue SSXT

1000
Time To Fatigue SSXT

900
900 900

(Seconds)
(Seconds)
(Seconds)

800 800 800


700 700
700
600 600
500 500 600
400 400 COMBINED GO2
CONTROL GO2 NO MP GO2 CONTROL
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for %
difference in time to exhaustion measured during graded

Summary (Time to Fatigue)


exercise cycle testing * = Significant difference between GO2
and listed conditions below.

QUESTION: Does using the GO2 during GXT and SSXT improve endurance?
ANSWER: YES FOR GXT; MAYBE FOR SSXT

9% 25%
8% * *
* *
20%

%Difference In TTE (SSXT)


%Difference In TTE (GXT)

7%
6%
5.81% 5.91% 5.86% 15%
5%
13.41% 13.70% 13.56%
4%
10%
3%
2% 5%
1%
0% 0%
GO2 vs. GO2 vs. GO2 vs GO2 vs. GO2 vs. GO2 vs
CONTROL NO MP COMBINED CONTROL NO MP COMBINED
CONTROL CONTROL
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for VO2
workload at which participants reached anaerobic threshold

Summary (Anaerobic Threshold)


* = Significant difference between GO2 and listed conditions
below.

37 33 34
* *
36 32
33
@ Anaerobic Threshold

35 31
VO2 (ml*kg*min-1)

32
34
30
33 31
29
32 30
28
31
29
30 27

26 28
29
28 25 27
CONTROL GO2 NO MP GO2 COMBINED GO2
CONTROL
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for %
difference in VO2 at which participants reached anaerobic

Summary (Anaerobic Threshold)


threshold * = Significant difference between GO2 and listed
conditions below.

QUESTION: Does using the GO2 during GXT increase anaerobic Threshold?
ANSWER: YES (Compared to NO MP & ALL DATA COMBINED)

9% *
% Difference in Anaerobic Threshold

8%
*
7% 7.40%

6%
5.84%
5%
4%
4.09%
3%
2%
1%
0%
GO2 vs. GO2 vs. GO2 vs.
CONTROL NO MP COMBINED
CONTROL
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error heart rate
recorded during graded exercise testing (GXT). *=Significantly

Summary (HEART RATE)


Different at p<0.05.

200

190 CONTROL GO2

180
Heart Rate (BPM)

170

160

150

140

130

120

110
HR ST1 HR ST2 HR ST3 HR PEAK EX HR REC1 HR REC3 HR REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error heart rate
recorded during graded exercise testing (GXT). *=Significantly

Summary (HEART RATE)


Different at p<0.05.

190

180 NO MP GO2

170
Heart Rate (BPM)

160

150

140

130

120

110
HR ST1 HR ST2 HR ST3 HR PEAK EX HR REC1 HR REC3 HR REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error heart rate
recorded during graded exercise testing (GXT). *=Significantly

Summary (HEART RATE)


Different at p<0.05.

190
COMBINED CONTROL
180
GO2
170
Heart Rate (BPM)

160

150
*
140
*
130

120

110
HR ST1 HR ST2 HR ST3 HR PEAK EX HR REC1 HR REC3 HR REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error heart
rate recorded during steady state exercise testing (SSXT).

Summary (HEART RATE)


*=Significantly Different at p<0.05.

190
190 190 COMBINED
CONTROL * CONTROL
180 NO MP
180 180 GO2
170 GO2
GO2 170
170
Heart Rate (BPM)

160
160 160
150
150 150
140 140
140
*
130 130 * 130

120 120 120

110 110 110

100 100 100


AVG HR PEAK HR HR REC1 HR REC3 HR REC5 AVG HR PEAK HR HR REC1 HR REC3 HR REC5 AVG HR PEAK HR HR REC1 HR REC3 HR REC5
OBRL – GO2 Project Results
Summary (Heart Rate)
QUESTION: Does using the GO2 during GXT have an impact on stage-
matched heart rates? Does using the GO2 result in enhanced heart rate
recovery following peak exercise?

ANSWER: Maybe, when data from both trials were combined, HR was
found to be reduced during stage 1 of GXT (moderate intensity exercise).
However, HR was found to be elevated when using the GO2 compared
during recovery from GXT and SSXT. This may indicate that while the
GO2 may optimize performance, it may not be optimal for using during
recovery from exercise.
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for VCO2
(L/min) measured during graded exercise cycle testing. * =

Summary (VCO2 L*min-1)


Significant difference between conditions at the same
measurement time-point (p<0.05)

4.5

4.0 CONTROL GO2


VCO2 (L*min-1)

3.5

3.0

2.5

2.0

1.5

1.0

0.5
VCO2 ST1 VCO2 ST2 VCO2 ST3 VCO2 PEAK VCO2 REC1 VCO2 REC3 VCO2 REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for VCO2
(L/min) measured during graded exercise cycle testing. * =

Summary (VCO2 L*min-1)


Significant difference between conditions at the same
measurement time-point (p<0.05)

5.0

*
4.5
NO MP GO2

4.0
VCO2 (L*min-1)

3.5

3.0

2.5

2.0
* *
1.5

1.0

0.5
VCO2 ST1 VCO2 ST2 VCO2 ST3 VCO2 PEAK VCO2 REC1 VCO2 REC3 VCO2 REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for VCO2
(L/min) measured during graded exercise cycle testing. * =

Summary (VCO2 L*min-1)


Significant difference between conditions at the same
measurement time-point (p<0.05)

4.5
*
COMBINED CONTROL
4.0
GO2
VCO2 (L*min-1)

3.5

3.0

*
2.5

2.0
* *
1.5

1.0

0.5
VCO2 ST1 VCO2 ST2 VCO2 ST3 VCO2 PEAK VCO2 REC1 VCO2 REC3 VCO2 REC5
OBRL – GO2 Project Results
Summary (VCO2 & RER)
QUESTION: Does using the GO2 have a significant effect on the volume
of exhaled CO2 during GXT or SSXT?

ANSWER: Potentially. When analyzing VCO2, there was an effect of


GO2 at stage 1 of GXT, Peak GXT, and at 3 minutes of recovery
compared to NO MP or all control data combined. Similar to the HR data,
these data indicate that the GO2 may be beneficial during performance,
but not recovery.

QUESTION: Does using the GO2 have a significant effect on RER


(VO2/VCO2) during GXT or SSXT during recovery?

ANSWER: Maybe, RER was found to be elevated compared to NO MP


during 5min of recovery. (NO MP: 1.059 ± 0.031; GO2: 1.174 ± 0.036)
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for
respiratory rate (RR) measured during graded exercise cycle

Summary (Respiratory Rate)


testing. * = Significant difference between conditions at the
same measurement time-point (p<0.05)

50
RESPIRATORY RATE (Breaths*min-1)

45 CONTROL GO2

40

35

30

25 *

20

15
RR ST1 RR ST2 RR ST3 RR PEAK RR REC1 RR REC3 RR REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for
respiratory rate (RR) measured during graded exercise cycle

Summary (Respiratory Rate)


testing. * = Significant difference between conditions at the
same measurement time-point (p<0.05)

55
RESPIRATORY RATE (Breaths*min-1)

50 NO MP GO2

45

40

35

30

25

20

15
RR ST1 RR ST2 RR ST3 RR PEAK RR REC1 RR REC3 RR REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for
respiratory rate (RR) measured during graded exercise cycle

Summary (Respiratory Rate)


testing. * = Significant difference between conditions at the
same measurement time-point (p<0.05)

50
RESPIRATORY RATE (Breaths*min-1)

COMBINED GO2
45 CONTROL

40

35

30

*
25
*
20

15
RR ST1 RR ST2 RR ST3 RR PEAK RR REC1 RR REC3 RR REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for
respiratory rate (RR) measured during steady state exercise

Summary (Respiratory Rate)


testing (SSXT). * = Significant difference between conditions at
the same measurement time-point (p<0.05)

50 55 50 COMBINED
CONTROL NO MP CONTROL
GO2
50
RESPIRATORY RATE (Breaths*min-1)

45 45
GO2 GO2
45
40 40

* 40 *
35 35
35 *

30 30
30

25 25
25

20 20 20
RR RR RR RR RR RR RR RR RR RR RR AVG RR RR RR RR
AVG PEAK REC1 REC3 REC5 AVG PEAK REC1 REC3 REC5 PEAK REC1 REC3 REC5
OBRL – GO2 Project Results
Summary (Respiratory Rate, RR)
QUESTION: Does using the GO2 have a significant effect on respiratory
rate (breaths / min) during GXT or SSXT.

ANSWER: YES, a decrease was observed at stage 1 of GXT ( 14 ±


5%, -3.3±1.3 breaths*min). This was also observed when observing the
GO2 against both control groups (NO MP & CONTROL) combined for
stage 1 and 2 of GXT. When looking at the combined data, average RR
during SSXT was reduced when using the GO2.

QUESTION: Does using the GO2 have a significant effect on respiratory


rate (breaths*min) during recovery from exercise?

ANSWER: Potentially. When adjusting for control RR, a decrease was


observed at 1 minute following SSXT(9 ± 4%, -3.1±1.5 breaths*min).
This was not observed following maximal graded exercise. This trend
was also observed when analyzing the combined data sets.
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for
ventilation (vt) measured during graded exercise cycle testing.

Summary (Ventilation (L*min-1)


* = Significant difference between conditions at the same
measurement time-point (p<0.05)

155
VENTILATION (L*min-1)

135 CONTROL GO2

115

95

75

55

35

15
Vt ST1 Vt ST2 Vt ST3 Vt PEAK Vt REC1 Vt REC3 Vt REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for
ventilation (vt) measured during graded exercise cycle testing.

Summary (Ventilation (L*min-1)


* = Significant difference between conditions at the same
measurement time-point (p<0.05)

155

135 NO MP GO2
VENTILATION (L*min-1)

115

95

75
*
55
*
35

15
Vt ST1 Vt ST2 Vt ST3 Vt PEAK Vt REC1 Vt REC3 Vt REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for
ventilation (vt) measured during graded exercise cycle testing.

Summary (Ventilation (L*min-1)


* = Significant difference between conditions at the same
measurement time-point (p<0.05)

155

* COMBINED GO2
135 CONTROL
VENTILATION (L*min-1)

115

95
*

75

*
55
* *
35

15
Vt ST1 Vt ST2 Vt ST3 Vt PEAK Vt REC1 Vt REC3 Vt REC5
OBRL – GO2 Project Results
Summary (Ventilation)
QUESTION: Does using the GO2 have a significant effect on ventilation (L*min-
1) during GXT?

ANSWER: YES (compared to NO MP & all control data combined).

QUESTION: Does using the GO2 have a significant effect on ventilation (L*min-
1) during SSXT?

ANSWER: No

QUESTION: Does using the GO2 have a significant effect on ventilation (L*min-
1) during recovery?

ANSWER: Potentially. When adjusting for ventilation using the control


mouthpiece, ventilation was observed to be higher following GXT when compared
to control and NO MP.
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for
systolic blood pressure (SBP) measured during graded exercise

Summary (Systolic Blood Pressure)


cycle testing. * = Significant difference between conditions at
the same measurement time-point (p<0.05)

210

200 *
Systolic Blood Pressure (mmHg)

CONTROL GO2
190

180

170

160

150

140

130

120
SBP ST1 SBP ST2 SBP ST3 SBP PEAK SBP REC1 SBP REC3 SBP REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for
systolic blood pressure (SBP) measured during graded exercise

Summary (Systolic Blood Pressure)


cycle testing. * = Significant difference between conditions at
the same measurement time-point (p<0.05)

190
Systolic Blood Pressure (mmHg)

180 NO MP GO2

*
170

160 *

150

140

130

120
SBP ST1 SBP ST2 SBP ST3 SBP PEAK SBP REC1 SBP REC3 SBP REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for
systolic blood pressure (SBP) measured during graded exercise

Summary (Systolic Blood Pressure)


cycle testing. * = Significant difference between conditions at
the same measurement time-point (p<0.05)

200
COMBINED CONTROL
*
Systolic Blood Pressure (mmHg)

190
* GO2
180

170
*

160

150 *

140

130

120
SBP ST1 SBP ST2 SBP ST3 SBP PEAK SBP REC1 SBP REC3 SBP REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for
systolic blood pressure (SBP) measured during steady state

Summary (Systolic Blood Pressure)


exercise testing. * = Significant difference between conditions
at the same measurement time-point (p<0.05)

200 190 190 COMBINED


CONTROL
NO MP CONTROL
190 180 180
GO2
Systolic Blood Pressure (mmHg)

* GO2 GO2
180 170 170

170 160 160

160 150 150

150 140 140

140 130 130

130 120 120

120 110 110


SBP AVG SBP PEAK SBP REC1 SBP REC3 SBP REC5 SBP AVG SBP PEAK SBP REC1 SBP REC3 SBP REC5
SBP AVG SBP PEAK SBP REC1 SBP REC3 SBP REC5
OBRL – GO2 Project Results
Summary (Systolic Blood Pressure)
QUESTION: Does using the GO2 have a significant effect on systolic
blood pressure during GXT compared to control or No MP?

ANSWER: YES. During GXT, the exercising with GO2 was observed to
elicit a lower SBP at various points during exercise.

QUESTION: Does using the GO2 have a significant effect on SBP


during SSXT compared to control?

ANSWER: Maybe, an elevated mean SBP was observed during SSXT


when comparing the GO2 to the control mouthpiece. However, this
difference was not observed compared to NO MP or when all control data
were combined.
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for
diastolic blood pressure (DBP) measured during graded

Summary (Diastolic Blood Pressure)


exercise cycle testing. * = Significant difference between
conditions at the same measurement time-point (p<0.05)

78

76
Diastolic Blood Pressure (mmHg)

* CONTROL GO2

74

72

70

68

66

64

62

60
DBP ST1 DBP ST2 DBP ST3 DBP PEAK DBP REC1 DBP REC3 DBP REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for
diastolic blood pressure (DBP) measured during graded

Summary (Diastolic Blood Pressure)


exercise cycle testing. * = Significant difference between
conditions at the same measurement time-point (p<0.05)

75
Diastolic Blood Pressure (mmHg)

NO MP GO2
70

65

60

55

50
DBP ST1 DBP ST2 DBP ST3 DBP PEAK DBP REC1 DBP REC3 DBP REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for
diastolic blood pressure (DBP) measured during graded

Summary (Diastolic Blood Pressure)


exercise cycle testing. * = Significant difference between
conditions at the same measurement time-point (p<0.05)

75
COMBINED CONTROL
73
Diastolic Blood Pressure (mmHg)

71 GO2

69

67
*
65

63

61

59

57

55
DBP ST1 DBP ST2 DBP ST3 DBP PEAK DBP REC1 DBP REC3 DBP REC5
OBRL – GO2 Project Results
Summary (Diastolic Blood Pressure)
QUESTION: Does using the GO2 have a significant effect
on diastolic blood pressure during GXT compared to control
or NO MP?

ANSWER: Potentially, DBP was observed to be reduced


during stage 1 of GXT when compared to the control MP but
not compared to NOMP. When all control data were
combined, DBP was observed to be reduced at 5 minutes of
recovery.

QUESTION: Does using the GO2 have a significant effect


on DBP during SSXT compared to control?

ANSWER: No.
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for mean
arterial pressure (MAP) measured during graded exercise

Summary (Mean Arterial Pressure)


cycle testing. * = Significant difference between conditions at
the same measurement time-point (p<0.05)

120
Mean Arterial Pressure (mmHg)

115 CONTROL GO2

110

105

100

95

90

85

80
MAP ST1 MAP ST2 MAP ST3 MAP PEAK MAP REC1 MAP REC3 MAP REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for mean
arterial pressure (MAP) measured during graded exercise

Summary (Mean Arterial Pressure)


cycle testing. * = Significant difference between conditions at
the same measurement time-point (p<0.05)

115
Mean Arterial Pressure (mmHg)

NO MP GO2
110

105
*
100

*
95

90

85

80

75
MAP ST1 MAP ST2 MAP ST3 MAP PEAK MAP REC1 MAP REC3 MAP REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for mean
arterial pressure (MAP) measured during graded exercise

Summary (Mean Arterial Pressure)


cycle testing. * = Significant difference between conditions at
the same measurement time-point (p<0.05)

115
COMBINED CONTROL
Mean Arterial Pressure (mmHg)

110
GO2

105

100

95
*
90
*
85

80

75
MAP ST1 MAP ST2 MAP ST3 MAP PEAK MAP REC1 MAP REC3 MAP REC5
OBRL – GO2 Project Results
Summary (Mean Arterial Pressure)
QUESTION: Does using the GO2 have a significant effect
on mean arterial pressure during GXT compared to control or
NO MP?

ANSWER: Potentially. MAP was observed to be reduced


during stage 2 and at 1 minute of recovery compared to NO
MP. When compared to all control data combined, MAP was
reduced at 3 and 5 minutes of recovery.

QUESTION: Does using the GO2 have a significant effect


on DBP during SSXT compared to control?

ANSWER: NO
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for mean
pulse pressure (PP) measured during graded exercise cycle

Summary (Pulse Pressure)


testing. * = Significant difference between conditions at the
same measurement time-point (p<0.05)

130

120 CONTROL GO2


Pulse Pressure (mmHg)

110

100

90

80

70

60
PP ST1 PP ST2 PP ST3 PP PEAK PP REC1 PP REC3 PP REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for mean
pulse pressure (PP) measured during graded exercise cycle

Summary (Pulse Pressure)


testing. * = Significant difference between conditions at the
same measurement time-point (p<0.05)

120

*
NO MP GO2
110
Pulse Pressure (mmHg)

100

90

80

70

60
PP ST1 PP ST2 PP ST3 PP PEAK PP REC1 PP REC3 PP REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for mean
pulse pressure (PP) measured during graded exercise cycle

Summary (Pulse Pressure)


testing. * = Significant difference between conditions at the
same measurement time-point (p<0.05)

130
COMBINED CONTROL
*
120
Pulse Pressure (mmHg)

* GO2

110
*
*
100

90 *

80

70

60
PP ST1 PP ST2 PP ST3 PP PEAK PP REC1 PP REC3 PP REC5
OBRL – GO2 Project Results
Summary (Pulse Pressure)
QUESTION: Does using the GO2 have a significant effect
on pulse pressure during GXT compared to control or NO
MP?

ANSWER: Overall – YES; When comparing GO2 to NO MP


or to all control data combined PP appears to be reduced at
certain time points during GXT.

QUESTION: Does using the GO2 have a significant effect


on DBP during SSXT compared to control or NO MP?

ANSWER: Potentially. Average and Peak PP was observed


to be reduced when comparing the GO2 to NO MP during
SSXT.
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for mean
rate pressure product (RPP) measured during graded exercise

Summary (Rate Pressure Product)


cycle testing. * = Significant difference between conditions at
the same measurement time-point (p<0.05)

40
Rate Pressure Product (HR*mmHg*10-3)

CONTROL GO2
35 *

30

25

20

15

10
RPP ST1 RPP ST2 RPP ST3 RPP PEAK RPP REC1 RPP REC3 RPP REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for mean
rate pressure product (RPP) measured during graded exercise

Summary (Rate Pressure Product)


cycle testing. * = Significant difference between conditions at
the same measurement time-point (p<0.05)

35
Rate Pressure Product (HR*mmHg*10-3)

NO MP GO2
30

25
*

20

15

10
RPP ST1 RPP ST2 RPP ST3 RPP PEAK RPP REC1 RPP REC3 RPP REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for mean
rate pressure product (RPP) measured during graded exercise

Summary (Rate Pressure Product)


cycle testing. * = Significant difference between conditions at
the same measurement time-point (p<0.05)

40
Rate Pressure Product (HR*mmHg*10-3)

COMBINED CONTROL

35
GO2
*
30
*
*
25

20

15

10
RPP ST1 RPP ST2 RPP ST3 RPP PEAK RPP REC1 RPP REC3 RPP REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for mean
rate pressure product (RPP) measured during steady state

Summary (Rate Pressure Product)


exercise testing. * = Significant difference between conditions
at the same measurement time-point (p<0.05)

40 CONTROL 35 35 COMBINED
NO MP CONTROL
GO2
Rate Pressure Product (HR*mmHg*10-3)

35 GO2
30 GO2 30
*
30
25 25
25
20 20
20

15 * 15
15

10 10 10
RPP RPP RPP RPP RPP RPP RPP RPP RPP RPP RPP RPP RPP RPP RPP
AVG PEAK REC1 REC3 REC5 AVG PEAK REC1 REC3 REC5 AVG PEAK REC1 REC3 REC5
OBRL – GO2 Project Results
Summary (Rate Pressure Product)
QUESTION: Does using the GO2 have a significant effect on rate pressure
product during GXT compared to control or NO MP?

ANSWER: YES. RPP tended to be lower at several stages when comparing


GO2 to control, NO MP, or combined controls.

QUESTION: Does using the GO2 have a significant effect on DBP during SSXT
compared to control?

ANSWER: Potentially. While mean RPP was higher when comparing GO2 to
the control mouth piece, the difference was not observed elsewhere. Compared
to NO MP, the RPP was higher at 5min of recovery.
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for mean
O2 pulse measured during steady state exercise testing. * =

Summary (O2 Pulse)


Significant difference between conditions at the same
measurement time-point (p<0.05)

0.30

* CONTROL GO2
0.25
O2 Pulse (VO2/HR)

0.20

0.15

0.10

0.05
O2P ST1 O2P ST2 O2P ST3 O2P PEAK O2P REC1 O2P REC3 O2P REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for mean
O2 pulse measured during steady state exercise testing. * =

Summary (O2 Pulse)


Significant difference between conditions at the same
measurement time-point (p<0.05)

0.27

*
0.25
NO MP GO2

0.23
O2 Pulse (VO2/HR)

0.21

0.19

0.17

0.15

0.13

0.11

0.09
O2P ST1 O2P ST2 O2P ST3 O2P PEAK O2P REC1 O2P REC3 O2P REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for mean
O2 pulse measured during steady state exercise testing. * =

Summary (O2 Pulse)


Significant difference between conditions at the same
measurement time-point (p<0.05)

0.27

* COMBINED CONTROL
0.25

GO2
0.23
*
0.21
O2 Pulse (VO2/HR)

0.19

0.17

0.15

0.13

0.11

0.09
O2P ST1 O2P ST2 O2P ST3 O2P PEAK O2P REC1 O2P REC3 O2P REC5
OBRL – GO2 Project Results
Summary (O2 Pulse)
QUESTION: Does using the GO2 during GXT have
an effect on O2 pulse?

ANSWER: Yes, during GXT at high intensity


exercise and maximal exercise.

QUESTION: Does using the GO2 during SSXT


have an effect on O2 pulse?

ANSWER: No
OBRL – GO2 Project Results
Summary (Perceived Exertion)
QUESTION: Does using the GO2 during GXT OR
SSXT have an effect on rates of perceived exertion.

ANSWER: No