Test 2 - GO2 vs. Nothing Test Results
Test 2 - GO2 vs. Nothing Test Results
Test 2 - GO2 vs. Nothing Test Results
Protocol 2 (SSXT): Steady state exercise testing set at a workload equivalent to measured anaerobic/ventilator
threshold assessed during GXT testing. Following an unloaded 3 minute warm-up, participants maintained a
constant pace until exhaustion. Gas collection was continuously collected and measurements of hemodynamics,
heart rate, and perceived exertion were recorded every two minutes during exercise and at 1, 3, and 5 minutes of
recovery.
Analysis: Data were compared at each stage of exercise and each time-point during recovery using a two-tailed
paired samples t-test. To account for potential wide variance in the participant population with regards to aerobic
fitness and training history, a mixed model analysis of covariance (ANCOVA) was used where values were co-
varied on measures taken with the control mouthpiece. FIGURES and CONCLUSIONS are presented for both
types of analyses in this report. NOTE: Conclusions presented in this document represent statistical findings and
are not intended to the potential physiologic mechanisms responsible.
1
Figures: Figures presented here are for data whereby at least one significant measurement was detected.
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for
VO2max expressed as both absolute (LO2*min-1) and
and relative (mlO2*kg*min-1) units. *=Significantly
3.50 3.30
VO2max (L*min-1)
VO2max (L*min-1)
3.30 3.20
3.30
3.10
3.10
3.00
3.10
2.90 2.90
2.90 2.80
2.70
2.70
2.50 2.70 COMBINED GO2
CONTROL GO2 NO MP GO2 CONTROL
50 46 46
*
*
VO2max (ml*kg*min-1)
(ml*kg*min-1)
(ml*kg*min-1)
48
44 44
46
44 42
42
42
40
VO2max
40
VO2
40
38 38
36 38 COMBINED GO2 2
CONTROL GO2 NO MP GO2 CONTROL
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for VO2
(L/min) measured during graded exercise cycle testing. * =
3.8
*
CONTROL GO2
3.3
VO2 (L*min-1)
2.8
*
2.3
1.8
1.3
0.8
VO2 ST1 VO2 ST2 VO2 ST3 VO2 PEAK VO2 REC1 VO2 REC3 VO2 REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for VO2
(L/min) measured during graded exercise cycle testing. * =
4.0
*
3.5 NO MP GO2
3.0
VO2 (L*min-1)
2.5
2.0
1.5
1.0
0.5
VO2 ST1 VO2 ST2 VO2 ST3 VO2 PEAK VO2 REC1 VO2 REC3 VO2 REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for VO2
(L/min) measured during graded exercise cycle testing. * =
4.0
COMBINED CONTROL
3.5 *
GO2
3.0
VO2 (L*min-1)
2.5
2.0
*
1.5
1.0
0.5
VO2 ST1 VO2 ST2 VO2 ST3 VO2 PEAK VO2 REC1 VO2 REC3 VO2 REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for VO2
(ml/kg/min) measured during graded exercise cycle testing. *
50
*
45 CONTROL GO2
VO2 (ml*kg*min-1)
40 *
35
30
25
20
15
10
VO2 ST1 VO2 ST2 VO2 ST3 VO2 PEAK VO2 REC1 VO2 REC3 VO2 REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for VO2
(ml/kg/min) measured during graded exercise cycle testing. *
50.0
*
45.0 NO MP GO2
VO2 (ml*kg*min-1)
40.0
*
35.0
30.0
25.0
20.0
15.0
10.0
VO2 ST1 VO2 ST2 VO2 ST3 VO2 PEAK VO2 REC1 VO2 REC3 VO2 REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for VO2
(ml/kg/min) measured during graded exercise cycle testing. *
50.0
* COMBINED CONTROL
45.0
GO2
VO2 (ml*kg*min-1)
40.0
35.0
30.0
25.0
*
20.0
15.0
10.0
VO2 ST1 VO2 ST2 VO2 ST3 VO2 PEAK VO2 REC1 VO2 REC3 VO2 REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for %
difference in VO2 (ml/kg/min) measured during graded
Summary (VO2max)
exercise cycle testing * = Significant difference between GO2
and listed conditions below.
QUESTION: Does using the GO2 during GXT have an effect on VO2max?
ANSWER: YES.
10%
*
9%
8%
%Difference In VO2max
7% *
6% 6.69%
*
5% 5.28%
4%
4.02%
3%
2%
1%
0%
GO2 vs. GO2 vs. GO2 vs
CONTROL NO MP COMBINED
CONTROL
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for time
to fatigue (seconds) for the graded maximal cycling test and
during steady state testing at a workload associated with each
(Seconds)
(Seconds)
550 500
500
500
450
450
450
400
400 400 COMBINED GO2
CONTROL GO2 NO MP GO2 CONTROL
1000
Time To Fatigue SSXT
900
900 900
(Seconds)
(Seconds)
(Seconds)
QUESTION: Does using the GO2 during GXT and SSXT improve endurance?
ANSWER: YES FOR GXT; MAYBE FOR SSXT
9% 25%
8% * *
* *
20%
7%
6%
5.81% 5.91% 5.86% 15%
5%
13.41% 13.70% 13.56%
4%
10%
3%
2% 5%
1%
0% 0%
GO2 vs. GO2 vs. GO2 vs GO2 vs. GO2 vs. GO2 vs
CONTROL NO MP COMBINED CONTROL NO MP COMBINED
CONTROL CONTROL
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for VO2
workload at which participants reached anaerobic threshold
37 33 34
* *
36 32
33
@ Anaerobic Threshold
35 31
VO2 (ml*kg*min-1)
32
34
30
33 31
29
32 30
28
31
29
30 27
26 28
29
28 25 27
CONTROL GO2 NO MP GO2 COMBINED GO2
CONTROL
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for %
difference in VO2 at which participants reached anaerobic
QUESTION: Does using the GO2 during GXT increase anaerobic Threshold?
ANSWER: YES (Compared to NO MP & ALL DATA COMBINED)
9% *
% Difference in Anaerobic Threshold
8%
*
7% 7.40%
6%
5.84%
5%
4%
4.09%
3%
2%
1%
0%
GO2 vs. GO2 vs. GO2 vs.
CONTROL NO MP COMBINED
CONTROL
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error heart rate
recorded during graded exercise testing (GXT). *=Significantly
200
180
Heart Rate (BPM)
170
160
150
140
130
120
110
HR ST1 HR ST2 HR ST3 HR PEAK EX HR REC1 HR REC3 HR REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error heart rate
recorded during graded exercise testing (GXT). *=Significantly
190
180 NO MP GO2
170
Heart Rate (BPM)
160
150
140
130
120
110
HR ST1 HR ST2 HR ST3 HR PEAK EX HR REC1 HR REC3 HR REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error heart rate
recorded during graded exercise testing (GXT). *=Significantly
190
COMBINED CONTROL
180
GO2
170
Heart Rate (BPM)
160
150
*
140
*
130
120
110
HR ST1 HR ST2 HR ST3 HR PEAK EX HR REC1 HR REC3 HR REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error heart
rate recorded during steady state exercise testing (SSXT).
190
190 190 COMBINED
CONTROL * CONTROL
180 NO MP
180 180 GO2
170 GO2
GO2 170
170
Heart Rate (BPM)
160
160 160
150
150 150
140 140
140
*
130 130 * 130
ANSWER: Maybe, when data from both trials were combined, HR was
found to be reduced during stage 1 of GXT (moderate intensity exercise).
However, HR was found to be elevated when using the GO2 compared
during recovery from GXT and SSXT. This may indicate that while the
GO2 may optimize performance, it may not be optimal for using during
recovery from exercise.
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for VCO2
(L/min) measured during graded exercise cycle testing. * =
4.5
3.5
3.0
2.5
2.0
1.5
1.0
0.5
VCO2 ST1 VCO2 ST2 VCO2 ST3 VCO2 PEAK VCO2 REC1 VCO2 REC3 VCO2 REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for VCO2
(L/min) measured during graded exercise cycle testing. * =
5.0
*
4.5
NO MP GO2
4.0
VCO2 (L*min-1)
3.5
3.0
2.5
2.0
* *
1.5
1.0
0.5
VCO2 ST1 VCO2 ST2 VCO2 ST3 VCO2 PEAK VCO2 REC1 VCO2 REC3 VCO2 REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for VCO2
(L/min) measured during graded exercise cycle testing. * =
4.5
*
COMBINED CONTROL
4.0
GO2
VCO2 (L*min-1)
3.5
3.0
*
2.5
2.0
* *
1.5
1.0
0.5
VCO2 ST1 VCO2 ST2 VCO2 ST3 VCO2 PEAK VCO2 REC1 VCO2 REC3 VCO2 REC5
OBRL – GO2 Project Results
Summary (VCO2 & RER)
QUESTION: Does using the GO2 have a significant effect on the volume
of exhaled CO2 during GXT or SSXT?
50
RESPIRATORY RATE (Breaths*min-1)
45 CONTROL GO2
40
35
30
25 *
20
15
RR ST1 RR ST2 RR ST3 RR PEAK RR REC1 RR REC3 RR REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for
respiratory rate (RR) measured during graded exercise cycle
55
RESPIRATORY RATE (Breaths*min-1)
50 NO MP GO2
45
40
35
30
25
20
15
RR ST1 RR ST2 RR ST3 RR PEAK RR REC1 RR REC3 RR REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for
respiratory rate (RR) measured during graded exercise cycle
50
RESPIRATORY RATE (Breaths*min-1)
COMBINED GO2
45 CONTROL
40
35
30
*
25
*
20
15
RR ST1 RR ST2 RR ST3 RR PEAK RR REC1 RR REC3 RR REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for
respiratory rate (RR) measured during steady state exercise
50 55 50 COMBINED
CONTROL NO MP CONTROL
GO2
50
RESPIRATORY RATE (Breaths*min-1)
45 45
GO2 GO2
45
40 40
* 40 *
35 35
35 *
30 30
30
25 25
25
20 20 20
RR RR RR RR RR RR RR RR RR RR RR AVG RR RR RR RR
AVG PEAK REC1 REC3 REC5 AVG PEAK REC1 REC3 REC5 PEAK REC1 REC3 REC5
OBRL – GO2 Project Results
Summary (Respiratory Rate, RR)
QUESTION: Does using the GO2 have a significant effect on respiratory
rate (breaths / min) during GXT or SSXT.
155
VENTILATION (L*min-1)
115
95
75
55
35
15
Vt ST1 Vt ST2 Vt ST3 Vt PEAK Vt REC1 Vt REC3 Vt REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for
ventilation (vt) measured during graded exercise cycle testing.
155
135 NO MP GO2
VENTILATION (L*min-1)
115
95
75
*
55
*
35
15
Vt ST1 Vt ST2 Vt ST3 Vt PEAK Vt REC1 Vt REC3 Vt REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for
ventilation (vt) measured during graded exercise cycle testing.
155
* COMBINED GO2
135 CONTROL
VENTILATION (L*min-1)
115
95
*
75
*
55
* *
35
15
Vt ST1 Vt ST2 Vt ST3 Vt PEAK Vt REC1 Vt REC3 Vt REC5
OBRL – GO2 Project Results
Summary (Ventilation)
QUESTION: Does using the GO2 have a significant effect on ventilation (L*min-
1) during GXT?
QUESTION: Does using the GO2 have a significant effect on ventilation (L*min-
1) during SSXT?
ANSWER: No
QUESTION: Does using the GO2 have a significant effect on ventilation (L*min-
1) during recovery?
210
200 *
Systolic Blood Pressure (mmHg)
CONTROL GO2
190
180
170
160
150
140
130
120
SBP ST1 SBP ST2 SBP ST3 SBP PEAK SBP REC1 SBP REC3 SBP REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for
systolic blood pressure (SBP) measured during graded exercise
190
Systolic Blood Pressure (mmHg)
180 NO MP GO2
*
170
160 *
150
140
130
120
SBP ST1 SBP ST2 SBP ST3 SBP PEAK SBP REC1 SBP REC3 SBP REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for
systolic blood pressure (SBP) measured during graded exercise
200
COMBINED CONTROL
*
Systolic Blood Pressure (mmHg)
190
* GO2
180
170
*
160
150 *
140
130
120
SBP ST1 SBP ST2 SBP ST3 SBP PEAK SBP REC1 SBP REC3 SBP REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for
systolic blood pressure (SBP) measured during steady state
* GO2 GO2
180 170 170
ANSWER: YES. During GXT, the exercising with GO2 was observed to
elicit a lower SBP at various points during exercise.
78
76
Diastolic Blood Pressure (mmHg)
* CONTROL GO2
74
72
70
68
66
64
62
60
DBP ST1 DBP ST2 DBP ST3 DBP PEAK DBP REC1 DBP REC3 DBP REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for
diastolic blood pressure (DBP) measured during graded
75
Diastolic Blood Pressure (mmHg)
NO MP GO2
70
65
60
55
50
DBP ST1 DBP ST2 DBP ST3 DBP PEAK DBP REC1 DBP REC3 DBP REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for
diastolic blood pressure (DBP) measured during graded
75
COMBINED CONTROL
73
Diastolic Blood Pressure (mmHg)
71 GO2
69
67
*
65
63
61
59
57
55
DBP ST1 DBP ST2 DBP ST3 DBP PEAK DBP REC1 DBP REC3 DBP REC5
OBRL – GO2 Project Results
Summary (Diastolic Blood Pressure)
QUESTION: Does using the GO2 have a significant effect
on diastolic blood pressure during GXT compared to control
or NO MP?
ANSWER: No.
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for mean
arterial pressure (MAP) measured during graded exercise
120
Mean Arterial Pressure (mmHg)
110
105
100
95
90
85
80
MAP ST1 MAP ST2 MAP ST3 MAP PEAK MAP REC1 MAP REC3 MAP REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for mean
arterial pressure (MAP) measured during graded exercise
115
Mean Arterial Pressure (mmHg)
NO MP GO2
110
105
*
100
*
95
90
85
80
75
MAP ST1 MAP ST2 MAP ST3 MAP PEAK MAP REC1 MAP REC3 MAP REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for mean
arterial pressure (MAP) measured during graded exercise
115
COMBINED CONTROL
Mean Arterial Pressure (mmHg)
110
GO2
105
100
95
*
90
*
85
80
75
MAP ST1 MAP ST2 MAP ST3 MAP PEAK MAP REC1 MAP REC3 MAP REC5
OBRL – GO2 Project Results
Summary (Mean Arterial Pressure)
QUESTION: Does using the GO2 have a significant effect
on mean arterial pressure during GXT compared to control or
NO MP?
ANSWER: NO
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for mean
pulse pressure (PP) measured during graded exercise cycle
130
110
100
90
80
70
60
PP ST1 PP ST2 PP ST3 PP PEAK PP REC1 PP REC3 PP REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for mean
pulse pressure (PP) measured during graded exercise cycle
120
*
NO MP GO2
110
Pulse Pressure (mmHg)
100
90
80
70
60
PP ST1 PP ST2 PP ST3 PP PEAK PP REC1 PP REC3 PP REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for mean
pulse pressure (PP) measured during graded exercise cycle
130
COMBINED CONTROL
*
120
Pulse Pressure (mmHg)
* GO2
110
*
*
100
90 *
80
70
60
PP ST1 PP ST2 PP ST3 PP PEAK PP REC1 PP REC3 PP REC5
OBRL – GO2 Project Results
Summary (Pulse Pressure)
QUESTION: Does using the GO2 have a significant effect
on pulse pressure during GXT compared to control or NO
MP?
40
Rate Pressure Product (HR*mmHg*10-3)
CONTROL GO2
35 *
30
25
20
15
10
RPP ST1 RPP ST2 RPP ST3 RPP PEAK RPP REC1 RPP REC3 RPP REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for mean
rate pressure product (RPP) measured during graded exercise
35
Rate Pressure Product (HR*mmHg*10-3)
NO MP GO2
30
25
*
20
15
10
RPP ST1 RPP ST2 RPP ST3 RPP PEAK RPP REC1 RPP REC3 RPP REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for mean
rate pressure product (RPP) measured during graded exercise
40
Rate Pressure Product (HR*mmHg*10-3)
COMBINED CONTROL
35
GO2
*
30
*
*
25
20
15
10
RPP ST1 RPP ST2 RPP ST3 RPP PEAK RPP REC1 RPP REC3 RPP REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for mean
rate pressure product (RPP) measured during steady state
40 CONTROL 35 35 COMBINED
NO MP CONTROL
GO2
Rate Pressure Product (HR*mmHg*10-3)
35 GO2
30 GO2 30
*
30
25 25
25
20 20
20
15 * 15
15
10 10 10
RPP RPP RPP RPP RPP RPP RPP RPP RPP RPP RPP RPP RPP RPP RPP
AVG PEAK REC1 REC3 REC5 AVG PEAK REC1 REC3 REC5 AVG PEAK REC1 REC3 REC5
OBRL – GO2 Project Results
Summary (Rate Pressure Product)
QUESTION: Does using the GO2 have a significant effect on rate pressure
product during GXT compared to control or NO MP?
QUESTION: Does using the GO2 have a significant effect on DBP during SSXT
compared to control?
ANSWER: Potentially. While mean RPP was higher when comparing GO2 to
the control mouth piece, the difference was not observed elsewhere. Compared
to NO MP, the RPP was higher at 5min of recovery.
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for mean
O2 pulse measured during steady state exercise testing. * =
0.30
* CONTROL GO2
0.25
O2 Pulse (VO2/HR)
0.20
0.15
0.10
0.05
O2P ST1 O2P ST2 O2P ST3 O2P PEAK O2P REC1 O2P REC3 O2P REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for mean
O2 pulse measured during steady state exercise testing. * =
0.27
*
0.25
NO MP GO2
0.23
O2 Pulse (VO2/HR)
0.21
0.19
0.17
0.15
0.13
0.11
0.09
O2P ST1 O2P ST2 O2P ST3 O2P PEAK O2P REC1 O2P REC3 O2P REC5
OBRL – GO2 Project Results KEY: Data are presented as means ± standard error for mean
O2 pulse measured during steady state exercise testing. * =
0.27
* COMBINED CONTROL
0.25
GO2
0.23
*
0.21
O2 Pulse (VO2/HR)
0.19
0.17
0.15
0.13
0.11
0.09
O2P ST1 O2P ST2 O2P ST3 O2P PEAK O2P REC1 O2P REC3 O2P REC5
OBRL – GO2 Project Results
Summary (O2 Pulse)
QUESTION: Does using the GO2 during GXT have
an effect on O2 pulse?
ANSWER: No
OBRL – GO2 Project Results
Summary (Perceived Exertion)
QUESTION: Does using the GO2 during GXT OR
SSXT have an effect on rates of perceived exertion.
ANSWER: No