pH measurement in
industrial processes
Selection and engineering guide
for your industry and application
pH measurement in industrial processes
Step by step
pH determination is an essential ultra pure water in power stations or the
measurement in all industries. Monitoring life science industry. The lifetime of a pH
of product quality or of chemical reactions electrode depends on these conditions,
is often done by means of pH measurement. but as well on cleaning, calibration,
The pH value is related to the hydrogen ion regeneration intervals and on the right
(H+) concentration in an aqueous solution choice of sensor type. A complete pH
and therefore to the solution’s acidity. The measuring point consists of the sensing
pH can (in theory) vary in water between element (pH electrode), an assembly, cable
0 – 14, with 0 being the acidic and 14 the and transmitter. This guide helps you
caustic end of the scale. with the selection of the right sensor and
assembly for your applications including
Application conditions for pH measurement the transmitter.
can be very different ranging, i.e. from
wastewater and chemical mixtures to
A B C
mounting assemblies, i.e.
in pipes, bypasses or small
tanks.
2
Contents
2. Checklist/Spec-Sheet. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
B
3.3 Application: High organic load. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.4 Application: Low conductivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
3.5 Application: Hygienic. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
3.6 Application: Heavy duty – abrasive. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
3.7 Application: High risk of buildup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
3.8 Application: Chemically demanding. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4. Selection of assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
4.1 Flow chart for assembly selection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
4.2 Immersion type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
4.3 Fixed installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
C
4.4 Retractable assembly (self-sealing piston) . . . . . . . . . . . . . . . . . . . . . . . . . . 44
4.5 Retractable assembly (with ball valves). . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
4.6 Required pH electrode length and total mounting length for
different assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Chemraz and Fluoraz are registered trademarks of Greene, Tweed & Co. Foundation Fieldbus is a registered trademark of
the Fieldbus FOUNDATION, Austin, TX USA. HART is a registered trademark of HART Communication Foundation, Austin,
TX USA. Kalrez, Teflon, and Viton are registered trademarks of E.I. Du Pont de Nemours & Co., Wilmington, USA. PEEK is a
registered trademark of Victrex.PROFIBUS is a registered trademark of the PROFIBUS User Organization. Tri-Clamp is a reg-
istered trademark of Ladish Company. Varivent is a registered trademark of GEA Tuchenhagen. EtherNet/IP is a trademark of
ODVA, Inc. Memosens, Orbisint, Cleanfit, Liquiline, Mycom, Top68 and W@M are trademarks of Endress+Hauser. All other
trademarks are the property of their respective owners.
3
A 1. Overview of pH electrodes and installation types
4
A
Revolutionary Memosens
technology
pH measurement has become to calibrate the sensor on its
easier and more reliable since own instead of the whole loop.
Endress+Hauser invented Decoupling of measurement and
Memosens. Inductive signal and calibration is possible. The “D” in
energy transmission without any the name of an Endress+Hauser
metallic contacts between sensor sensor indicates that it is available
head and cable connection ensures with Memosens (i.e. CPxxxD; D =
trouble-free operation even in digital).
humid environments. With the
galvanically decoupled system and See also section 5.1 on page 50 or
the storage of calibration data www.us.endress.com/memosens
in the sensor head it is possible
5
A 1. Overview of pH electrodes and installation types
1.2 pH electrodes
pH range 0 to 14 0 to 14 0 to 14
Temperature range 32 to 275°F / 0 to 135°C 32 to 275°F / 0 to 135°C 32 to 230°F / 0 to 110°C
Max. pressure Up to 232 psi / 16 bar Up to 188.5 psi / 13 bar Up to 188.5 psi / 13 bar
(with B-glass)
Min. conductivity 50 µS/cm 10 µS/cm 500 µS/cm
Organic content < 20 vol% < 20 vol% < 20 vol%
6
A
0 to 14 0 to 14 0 to 10 0 to 14
32 to 275°F / 0 to 135°C 32 to 176°F / 0 to 80°C 32 to 284°F / 0 to 140°C 35.6 to 167°F / 2 to 75°C
Up to 145 psi / 10 bar rel Up to 145 psi / 10 barrel Up to 87 psi / 6 barrel Up to 50 psi / 3.45 bar rel
with counter pressure
0.04 µS/cm 50 µS/cm 50 µS/cm 0.1 µS/cm
Higher level possible < 20 vol% < 20 vol% Not specified
depending on application
Glass Glass Enamel on stainless steel Stainless steel
Ceramic PTFE Ceramic PTFE
Liquid filled Gel filled, double chamber Liquid filled Gel
Ultrapure water, fat, dye, Wastewater, mining Food and pharma Pure water
food, process
7
A 1. Overview of pH electrodes and installation types
1.2 pH electrodes
Max. pressure Up to 145 psi / 10 barrel Up to 145 psi / 10 bar rel Up to 145 psi / 10 bar rel
pH range 0 to 14 0 to 14 0 to 14
Temperature range 5 to 275°F / -15 to 135°C 5 to 230°F / -15 to 110°C 5 to 266°F / -15 to 130°C
Min. conductivity 10 µS/cm 500 µS/cm 0.1 µS/cm
Organic content High level possible High level possible High level possible
depending on application depending on application depending on
application
Shaft material PEEK, chip sealing: EPDM PEEK , chip sealing: PEEK, chip sealing: EPDM
perfluorelastomer or perfluorelastomer
Diaphragm Ceramic Open junction Ceramic
Reference system Gel filled, double chamber Stabilized gel reference Liquid filled
Special options 3-A certificate, 3-A certificate
Certificate of Compliance
available
Application Non aqueous media, Non aqueous media, Non aqueous media,
galvanic, pharma, food, emulsions, suspensions, ultra-pure water, fat,
fermenter, process precipitation reactions dye, food, process
8
A
Up to 232 psi / 16 bar rel Up to 188.5 psi / 13 barrel Up to 188.5 psi / 13 bar rel
(with B glass)
pH: 0 to 14 pH: 0 to 14 pH: 0 to 14
ORP: -1500 to 1500 mV ORP: -1500 to 1500 mV ORP: -1500 to 1500 mV
rH: 0 to 42 rH: 0 to 42 rH: 0 to 42
32 to 275°F / 0 to 135°C 32 to 284°F / 0 to 140°C 32 to 230°F / 0 to 110°C
50 µS/cm 10 µS/cm 500 µS/cm
< 20 vol% < 20 vol% < 20 vol%
9
A 1. Overview of pH electrodes and installation types
10
A
11
A 1. Overview of pH electrodes and installation types
1.4 pH assemblies
(Type of electrode see table on page 48)
Max. operating 145 psi/10 barrel at 87 psi/6 barrel at 145 psi/10 barrel at
pressure/temp. 284°F/140°C, metal 68°F/20°C 32°F/0°C to 284°F/140°C
dependent on 116 psi/8 barrel at 0 psi/barrel at
material 140°F/60°C, PVDF 176°F/80°C
Materials in contact PVDF, 1.4404/316L PP 1.4435
with medium
Sealings EPDM/Viton ®, Chemraz®/ EPDM EPDM-FDA, FMP-FDA,
Fluoraz® silicone-FDA
Process connections Welding adapter for DN Thread G 1” DN 25 standard,
25 pipe NPT 1” DN 25 also for B.Braun port,
Flange DN 25 PN 16 Tri-Clamp® 1.5”; Tri-Clamp 2”
Flange ANSI 1” 150 lbs Dairy fitting DN 50
Flange JIS 10K 25 A DIN11851/2.5”
Thread FNPT ½” Varivent® DN 40-125/0.4”
APV DN 40-100/0.4”
Neumo BioControl D 50
Cleaning Spray cleaning Spray cleaning CPR31 3-A
connection G ½”
12
A
145 psi/10 bar rel at 58 psi/4 barrel at 145 psi/10 bar rel at 58 psi/4 barrel at
194°F/90°C, metal 68°F/20°C, 212°F/100°C, metal 68°F/20°C,
87 psi/6 barrel at 0 psi/barrel at 87 psi/6 barrel at 0 psi/barrel at
176°F/80°C, PVDF 176°F/80°C 68°F/20°C, PVDF 176°F/80°C
PVDF, 1.4571/316Ti, PP PVDF, 1.4404/316L PVC, 1.4404/316L
Monel
Viton EPDM EPDM/Viton/Chemraz/ EPDM
Fluoraz
M-NPT ½”, M-NPT ¾”, Flange DN 100 Flange DN 80 PN16 For sensor adapter, G1,
thread M 25 x 1.5 Adjustable flange DN Flange ANSI 3” 150 lbs NPT ¾”, G ¾”, PG13.5
100 Flange JIS 10K 80A
Suspension bracket
Pendulum frame
mounting
13
A 1. Overview of pH electrodes and installation types
1.4 pH assemblies
(Type of electrode see table on page 48)
Max. operating 232 psi/16 barrel at 232 psi/16 barrel at 58 psi/4 barrel at
pressure/temperature 284°F/140°C (metal and 284°F/140°C 248°F/120°C,
PEEK version) 174 psi/12 barrel static
Material in contact 1.4404/316L 1.4435/316L 1.4404/316L
with medium Alloy C22 Alloy C22 (no PVDF-PTFE) Alloy C4,
PEEK, PVDF, Titanium
PVDF conductive
Sealings EPDM/FKM/FFKM EPDM-FDA/FKM-FDA/ EPDM/Viton/Kalrez ®
FFKM-FDA
Operation Manual/ Manual/ Manual
pneumatic pneumatic
Process Clamp 2“, 2½“ Clamp 1½“, 2“, 2½“ Thread G1¼ ,
connections flange DN 40, DN 50, Aseptic DN 25, DN 50 G1½” external,
DN 80 Neumo Biocontrol D 65 thread NPT 1¼”,
Flange 2“, 3“ Neumo Bioconnect D 50, G1½” external,
Flange 10K50, 10K80 D 65 flange DN 32,
Thread NPT 1½“ Dairy fitting DN 50, DN 65 DN 40, DN 50
Thread G 1¼ Thread G 1¼ flange ANSI 1¼”,
Dairy fitting DN 50, DN 65 Varivent flange 1½”, 2” 150 lbs
Convertible to Yes Yes No
pneumatic
Sealing to process O-rings Gaskets Ball valve
Special options Immersion chamber Double chamber version Manual or with worm gear
version available with dynamic sealing handle
available,
certificates: EHEDG, USP
Class VI, ASME BPE, CoC,
FDA
Application Water, wastewater, process Food and pharma Water, wastewater,
processes process
14
A
145 psi/10 barrel at 212°F/100°C, 145 psi/10 barrel at 212°F/100°C, 87 psi/6 barrel at 176°F/80°C,
284°F/140°C max. 266°F/130°C max. 0 psi/0 barrel at 248°F/120°C
Various flow assemblies Flow chamber, optionally with Flow chamber, optionally with
PFA-lined, wiping seal wiping seal
3.1 certificates
Heavy duty and process Chemical industry, paper Paper industry, industrial water
application industry, sticky medium treatment
15
A 1. Overview of pH electrodes and installation types
1.5 Transmitters for pH measurement
Liquiline CM14
Liquiline CM14 is a basic
transmitter that offers all
you need to run a standard
measuring point. It fits into the
common cabinet cut-outs and is
easy to commission thanks to
digital Memosens technology.
The Memosens hot plug & play
concept allows to quickly install
and configure your digital
sensors.
16
A
17
A 1. Overview of pH electrodes and installation types
1.6 pH transmitters
Liquiline Liquiline
CM42 CM44/CM44R
18
A
pH glass, ORP, conductivity, oxygen pH glass, pH ISFET, conductivity, pH glass, pH ISFET, conductivity,
oxygen, chlorine, ORP ORP
No No Yes
2 line, LCD with dot matrix 2 line LCD Backlit LCD with dot matrix
IP65 / NEMA 4X Panel: IP54/Similar to NEMA 3S (front), Field: IP65 (Similar to NEMA 4)
IP30/Similar to NEMA 1 (housing)
Field: IP65/Similar to NEMA 4X
Compact device for panel and cabinet Cleaning with a timer, Logbook, data logger, DAT
mounting Chemoclean, PID controller... module, Chemoclean, controller,
NAMUR output contacts, different
user levels ...
19
A 2. Checklist
Customer contact data:
Name: Company:
Email: Telephone:
Medium pH range
Conductivity [μS/cm]
Suspended solids
Abrasives
Process pressure
Flow velocity
Installation in pipe
Bypass installation
Sample preparation
Transmitter 2-/4-wire
Ingress protection
Dosing to be controlled by
transmitter?
Automatic cleaning?
Multichannel device
EHEDG
3-A
SIL
3.1 certificate
20
A
Customer contact data:
Name: Company:
Email: Telephone:
21
3. Selection of pH electrodes according to applications
3.1 Flow chart for pH electrode selection
B
The selection of a pH electrode is primarily As some of the conditions might be difficult
based on chemical and physical behaviors to predict there is also the choice in the
of the process medium. Combinations with flow chart “unknown” From there you
process or industry specific requirements are directed to the individual chapters
such as “hygienic” will reduce the choice (3.2 – 3.8) with the indication of the
of pH electrodes suitable for certain recommended pH electrode including key
applications. However, the key criteria are advantages as well as application limits and
based on maximum expected lifetime and alternatives. Simplified to reduce
maintenance efforts such as calibration or complexity, you may find combinations that
refilling of KCl. require contact with specialists.
22
B
µ
266 F
232 psi
23
3. Selection of pH electrodes according to applications
3.2 Application: Standard
B First Choice
Orbisint Memosens Orbipac
CPS11/CPS11D CPS16D CPF81/CPF81D pH electrode
application
24
Application: Standard
Conditions Process Typ. Liquids
• Conductivity > 50 µS/cm • Neutralization • Wastewater
• Organic content < 20 vol % • Water treatment
B
• NON-hygienic or abrasive
Ceraliquid
CPS41/CPS41D pH electrode
application
Standard
• Fast response time due to ceramic
diaphragm and liquid filling
• More soiling-resistant due to continuous
flushing of diaphragm
25
3. Selection of pH electrodes according to applications
3.3 Application: High organic load
B First Choice
Tophit Tophit
CPS441/CPS441D pH electrode CPS471/CPS471D pH electrode
High organic
load
Advantages • No aging effect of ISFET chip caused by • No aging effect of ISFET chip caused by
organics organics
• Stable and fast measurement due to • Up to 95% organic content possible
liquid reference
Technical data
• Process temperature 5°F to 275°F/-15°C to 135°C 5°F to 275°F/-15°C to 135°C
• Process pressure Up to 145 psi/10 barrel, KCl vessel CPY7 Up to 145 psi/10 barrel
with counter pressure necessary
• pH range 0 to 14 0 to 14
• Sensor lengths 120, 225, 360 and 425 mm 120, 225, 360 and 425 mm
• Transmission Memosens and TOP68 Memosens and TOP68
Application limits • Presence of hot ‣ Glass type • Remark: Generally
caustics, i.e. CPS41/CPS41D automatic
during “CIP” cleaning with
• Soiling media ‣ CPS41/CPS41D Topclean (see
and/or page 51)
automatic recommended
cleaning with
Topclean
(see page 51)
‣ = alternative product
26
Application: High organic load
Conditions Process/Industry Typ. Liquids
• Organic content > 20 vol % • Dye and pigment • Impregnating resin
• NO fibers, particles or risk of production
B
poisoning, i.e. by S2-,CN- or
NH3
• NON-hygienic or abrasive
High organic
load
27
3. Selection of pH electrodes according to applications
3.4 Application: Low conductivity
B First Choice
Ceraliquid Purisys
CPS41/CPS41D ph electrode CPF201 pH electrode
conductivity
Low
28
Application: Low conductivity
Conditions Process/Industry Typ. Liquids
• Conductivity < 50 µS/cm • Power [Pharma] • Boiler water
• NO fibers, particles or risk of • Pure/ultra pure
B
poisoning, i.e. by S2-,CN- or water “WFI”
NH3
• NON-abrasive
Orbisint
CPS11/CPS11D pH electrode
conductivity
(with salt ring)
Low
• No electrolyte refilling necessary
2 to 12 (A-glass)
120, 225, 360 and 425 mm
Memosens and TOP68
• Limited life time ‣ CPS41/CPS41D
of approx. 6
months until
electrolyte
reservoir is used
up
29
3. Selection of pH electrodes according to applications
3.5 Application: Hygienic
B First Choice
Technical data
• Process temperature 32°F to 275°F/0°C to 135°C 5°F to 275°F/-15°C to 135°C
30
Application: Hygienic
Conditions Process/Industry Typ. Liquids
• Organic content < 20 vol % • Food • Fermentation
• NON-abrasive • Pharma • WFI (water for
B
injection)
application
• Less risk of breakage
Hygienic
• Direct mounting in process with
hygienic process connection
• Fast response
• Highly viscous media
5°F to 176°F/-15°C to 80°C (A-glass), 5°F to 275°F/-15°C to 135°C 32°F to 284°F/0°C to 140°C
32°F to 275°F/0°C to 135°C (B-glass)
Up to 145 psi/10 barrel, KCl vessel Up to 145 psi/10 barrel, KCl vessel
CPY7 with counter pressure CPY7 with counter pressure Up to 87 psi/6 barrel
necessary necessary
2 to 12 (A-glass), 0 to 14 (B-glass) 0 to 14 0 to 10
120, 225, 360 and 425 mm 120, 225, 360 and 425 mm
Memosens and TOP68 Memosens and TOP68 Memosens
• Risk of glass ‣ Non-glass • Reduced ‣ Retract sensor • Manual ‣ Gel-filled
breakage ISFET sensor resistance during cleaning refilling of CPS71/
CPS441/ against hot cycle or use electrolyte CPS71D,
CPS441D caustic (CIP) CPS41/CPS41D, vessel CPS76D, or
• Chance of ‣ Gel-filled CPS71/CPS71D, necessary CPS471/
continuous CPS71/ CPS76D • Significantly CPS471D
outflow of KCl CPS71D, • Soiling media ‣ CPS41/CPS41D, higher
traces CPS76D CPS71/CPS71D, investment
• Manual or CPS471/ CPS76D and/ costs than
refilling of CPS471D or standard
electrolyte automatic electrode
vessel cleaning
necessary [see page 51]
31
3. Selection of pH electrodes according to applications
3.6 Application: Heavy duty – abrasive
B First Choice
Orbipac Tophit
CPF81/CPF81D pH electrode CPS491/CPS491D pH electrode
Advantages • Flat membrane prevents glass abrasion • Sensor design allows mounting of sensing
• Double chamber reference offers element in “flow shadow”
protection against poisoning
Technical data
• Process temperature 32°F to 176°F/0°C to 80°C 5°F to 230°F/ -15°C to 110°C
• Process pressure Up to 145 psi/10 barrel (176°F/80°C) Up to 145 psi/10 barrel
• pH range 0 to 14 0 to 14
• Sensor lengths 120, 225, 360 and 425 mm
• Transmission Memosens, TOP68 and fixed cable Memosens and TOP68
application
Heavy duty
32
Application: Heavy duty – abrasive
Conditions Process/Industry Typ. Liquids
• Conductivity > 50 µS/cm • Mining • Slurries
• Organic content < 20 vol %
• NON-hygienic
application
Heavy duty
33
3. Selection of pH electrodes according to applications
3.7 Application: High risk of buildup
B First Choice
Orbipore Memosens Orbisint Memosens
CPS91/CPS91D CPS96D CPS11/CPS11D CPS16D
pH electrode pH/ORP electrode pH electrode pH/ORP electrode
34
Application: High risk of buildup
Conditions Process/Industry Typ. Liquids
• Conductivity > 50 µS/cm • Pulp and paper • Paper bleaching
• Organic content < 20 vol % • Power and energy • Emulsions
B
• NON-hygienic or abrasive • Flue gas
desulfurization
of buildup
High risk
35
3. Selection of pH electrodes according to applications
3.8 Application: Chemically demanding
B First Choice
(BT version)
36
Application: Chemically demanding
Conditions Process/Industry Typ. Liquids
• Risk of poisoning, i.e. by • All • HCN production
S2-,CN- or NH3 • Chemical process
B
• Conductivity > 50 µS/cm solutions
• Organic content < 20 vol %
• NON-hygienic or abrasive
Ceragel Memosens
CPS71/CPS71D CPS76D
pH electrode pH/ORP electrode
0 to 14
120, 225, 360 and 425 mm
Memosens and TOP68
• For fastest ‣ CPS41/CPS41D
response time and
best poison
resistance
demanding
Chemically
37
4. Selection of assembly
4.1 Flow chart for assembly selection
General considerations
Correct assembly selection requires For automatic cleaning and calibration
consideration of the installation and using Topclean and Topcal, you must
application conditions as well as pH select pneumatically driven retractable
electrode selection. assemblies.
C
assemblies are available in different technology to operate with 2 pH electrodes
versions suitable for either gel or liquid- (one in the application and the 2nd in the
filled electrodes. Retrofitting from one calibration/recovering cycle) retractable
version to another is either impossible or assemblies are recommended unless sensor
requires considerable effort. Make sure that is installed in bypass. This ensures the
you select an assembly that matches the exchange of pH electrodes under process
length of the pH electrode. See table 4.6 on conditions.
page 48 for details.
For hygienic applications (i.e. FDA
For chemically demanding and safety conformity or EHEDG certificates) you will
relevant applications, ball valve sealed find the selection in the technical data).
assemblies should be used due to complete
mechanical isolation during the exchange
process.
38
START
no
C
Installation in open yes Immersion type
channel or in (pressurized) See Page 40
vessel from top?
no
No removal of pH yes
electrode during Fixed installation
operation required? See page 42
no
Chemically yes
Retractable assembly
demanding self-sealing piston
or hygienic? See page 44
no
Abrasives
no yes
Retractable assembly
Contact media, fibers particles
with ball valves
Endress+Hauser present or high
See page 46
availability?
39
4. Selection of assembly
4.2 Immersion type
First Choice
Flexdip Dipfit
CYA112 CPA111
C
Immersion
type
Advantages • Modular system for a wide variety of • 3 electrode slots for redundant
sensor mounting, i.e. pH, turbidity measurement
• Flexible immersion depth by chemically
resistant pipes
• Spray cleaning head as option
Technical data
• Process temperature 32°F to 140°F/0 to 60°C, PVC; 32°F to 14°F to 176°F/-10 to 80°C
194°F/0 to 90°C, 316L
• Process pressure 0 to 58 psi/4 bar rel 0 to 58 psi/4 bar rel
• Wetted materials PVC; 316L, EPDM, Viton PP, EPDM
Process connection Different holder systems, float ball, Flange DN 100, adjustable flange DN
chain from Nylon hanging, pendulum 100, hanging bracket, pendulum frame
frame mounting mounting
• Immersion depth 23.6 in. to 11.8 ft./600 to 3600 mm 19.7 in. to 9.8 ft./500 to 3000 mm
Application limits • Spray cleaning ‣ CPA111 • Long immersion ‣ CYA112 with
available on [for CPF81/ depth or high float installation
request CPF81D] lateral load like from side of the
• Long immersion ‣ CPA140 agitation vessel if possible
depth or high
lateral load like
‣ = alternative product agitation on
request
40
Installation in open channels, basins
and in closed vessels from top
Immersion assemblies are usually used for installation in open
channels and basins fixed by chains or on a rail. Version with
flanges can as well be used for installation of the sensor from
the top of a vessel. Typical applications are, i.e. municipal and
industrial wastewater.
Dipfit
CPA140
Immersion
type
• 3 electrode slots for redundant
measurement
• Robust process sealing due to the bayonet
mounting method
41
4. Selection of assembly
4.3 Fixed installation
First Choice
Flowfit Flowfit
CPA250 CPA240
• Low cost PP, flow through style • Certificate according to NACE available
• Easy calibration using the detachable • Different design (axis) of inlet and outlet
Fixed
• Process connection Thread G1, NPT 1”, flange DN 25; Thread or flanges DN 25; 90° or 180°
90° or 180° between in- and outflow between in- and outflow
• Immersion depth
42
Installation in pipes/bypass with flow
through or insertion type assembly
Suitable for processes which do not need frequent replacement
or calibration of pH electrode. For applications with pressure in pipe
medium, flow has to be interrupted in the pipe or bypass to get
access to the sensor by an external valve.
Unifit Ecofit
CPA442 CPA640
installation
Fixed
14°F to 284°F/-10 to 140°C 14°F to 284°F/-10 to 140°C
0 to 87 psi/6 bar rel 0 to 87 psi/6 bar rel
1.4435/316L, EPDM-FDA, FPM-FDA, PVDF, 1.4571/316L
silicone-FDA
Thread G 1¼”, NPT, Tri-Clamp, Varivent, Thread G 1¾” or M25x1.5, NPT ½”, ¾”
Bioconnect, AVP, DN 25 and DN 50 dairy
fitting
0.98”, 1.96”, 3.34”/25 mm, 50 mm or 85
mm
• Change of sensor ‣ Retractable • Only thread ‣ CPA442
under pressure in CPA875 process
hygienic design connections
43
4. Selection of assembly
4.4 Retractable assembly (self-sealing piston)
Advantages • Flexible retractable assembly for a broad • Non-metal material offers good chemical
range of applications compatibility against most acids and
• Intelligent functions for highest process caustics
and personnel safety • Fully retractable in PP + PVDF i.e. for
water treatment
Technical data
Retractable assem.
self-sealing piston
44
Installation in pipe and vessels with
self-sealing piston
Manual and pneumatic
Cleanfit Cleanfit
CPA875 CPA472D
• Hygienic design 3-A, FDA-listed materials • Various materials available for best
and EHEDG certificate for a broad range of chemical compatibility
hygienic applications • Heavy duty version available with flow
• Double chamber with patented sealing chamber and sight glass as option
principle for 100% sterile processes • High immersion depth up to 11 inches /
280 mm
Retractable assem.
self-sealing piston
14 to 284°F / -10 to 140°C -4 to +284°F / -20 to 140°C,
short time up to 320°F / 160°C
Manual: 0 to 116 psi / 8 bar rel, 0 to 145 psi / 10 bar rel
Pneu.: 0 to 232 psi / 16 bar rel SS 316L, alloy C4, titanium, PVDF,
1.4435/316L, Alloy C22 PVDF conductive, PEEK
45
4. Selection of assembly
4.5 Retractable assembly (with ball valves)
Cleanfit Cleanfit
CPA450 CPA473
Advantages • Variable immersion depth up to • Open sensor protection guard and tape
27.5”/700 mm wiper prevents fiber buildup around the
• Open sensor protection guard prevents sensor, i.e. media in pulp & paper, mining
fiber buildup around the sensor
• Increased safety due to high pressure
protection
Technical data
• Process temperature 32°F to 266°F/0 to 130°C 32°F to 266°F/0 to 130°C
• Process pressure 0 to 58 psi/4 barrel retraction; 0 to 232 0 to 145 psi/10 bar rel
psi/16 barrel static
• Wetted materials SS 316L and Alloy C4, Titanium, EPDM, SS 316L , FPM, Kalrez
FPM, Kalrez
• Process connection G 1¼”, G 1½” NPT ½”, flanges DN 32, G 1¼”, dairy DN 50, flanges DN
Retractable assem.
46
Installation in pipe and vessels with
ball valve
• Ball valve offers safety process sealing
• Manual and/or pneumatic
Cleanfit
CPA474
Up to 8.15”/207 mm
Manual/pneumatic
• PP/PVDF/PEEK ‣ CPA473
chemically not
compatible
• Not compatible ‣ Use CPA640 as
with CPF81/ adapter
CPF81D
• Higher immersion ‣ On request
length
47
4. Selection of assembly
4.6 Required pH electrode length and total immersion depth
for different assemblies
Glass sensors
C
Maximum
immersion CPS11/ CPS71/ CPS91/
CPS41/
depth2) CPS11D / CPS71D / CPS91D /
CPS41D 1)
CPS16D CPS76D CPS96D
CPA111 see note 3) 120 mm 120 mm 120 mm 120 mm
CYA112 see note 3) 120 mm 120 mm 120 mm 120 mm
CPA140 see note 3) 120 mm 120 mm 120 mm 120 mm
CPA240 not applicable 120 mm 120 mm 120 mm 120 mm
CPA250 not applicable 120 mm 120 mm 120 mm 120 mm
CPA442 73 mm 120 mm 120 mm 120 mm 120 mm
CPA640 85 mm 120 mm 120 mm 120 mm 120 mm
CPA450 see note 3) 120 mm n/a 120 mm 120 mm
short 146 mm 225 mm 360 mm 225 mm 225 mm
CPA472D
long 280 mm 360 mm n/a 360 mm 360 mm
short 100 mm 225 mm 425 mm 225 mm 225 mm
CPA473
long 230 mm 360 mm n/a 360 mm 360 mm
short 76 mm 225 mm 425 mm 225 mm 225 mm
CPA474
long 207 mm 360 mm n/a 360 mm 360 mm
basic 120 mm n/a 120 mm 120 mm
36 mm
short 225 mm 225 mm 225 mm 225 mm
basic
78 mm 225 mm n/a 225 mm 225 mm
CPA871 long
immer-
225 mm n/a 225 mm 225 mm
sion 135 mm
360 mm 360 mm 360 mm 360 mm
chamber
single
chamber 36 mm 225 mm 225 mm 225 mm 225 mm
short
single
225 mm n/a 225 mm 225 mm
CPA875 chamber 78 mm
360 mm 360 mm 360 mm 360 mm
long
225 mm n/a 225 mm 225 mm
double
78 mm 360 mm 360 mm 360 mm 360 mm
chamber
360 mm n/a 360 mm 360 mm
48
Max. length
see Note 2)
ISFET
CPS441/
CPS441D 1)
CPS471/
CPS471D
CPS491/
CPS491D Notes:
1) Liquid-filled [KCl]
C
120 mm 120 mm 120 mm 2) The indicated length is the maximum length
which must be considered i.e. for installation
120 mm 120 mm 120 mm in pipes to ensure mechanical space;
120 mm 120 mm 120 mm depending on process connection it may be
shorter for the individual assembly
120 mm 120 mm 120 mm (see drawing)
3) Depending on length of immersion assembly
120 mm 120 mm 120 mm
120 mm 120 mm 120 mm
120 mm 120 mm 120 mm
n/a 120 mm 120 mm
360 mm 225 mm 225 mm
n/a 360 mm 360 mm
425 mm 225 mm 225 mm
n/a 360 mm 360 mm
425 mm 225 mm 225 mm
n/a 360 mm 360 mm
n/a 120 mm 120 mm
225 mm 225 mm 225 mm
49
5. Life cycle management of pH loops
5.1 Calibration concept for the laboratory using Memosens
and Memobase Plus
In the case of Memosens technology, the one and the measurement is available
conversion from analog to digital signals again immediately. The most important
takes place inside the sensor. That is why maintenance task is then carried out in
the sensor is the only component which the comfortable environment of the lab
must be inspected and calibrated regularly. pertaining to the plant, where all the
The cable and the transmitter have much essentials are directly available. There
less impact on the measurement process the sensors are cleaned, conditioned and
than in the case of an analog system, which calibrated under optimum conditions.
is sensitive to moisture and electromagnetic
interference, among other things. The lab calibration concept is supported
by the Memobase Plus sensor and data
Memosens sensors not only store process management software. Memobase Plus
data which can be used for predictive stores all sensor and calibration data in a
maintenance, but most importantly store database which can be used as a server.
current calibration data such as the slope This offers visualization, reporting and
and zero point of the pH electrodes. This data export functionality. The software
has enabled the development of the lab supports not only pH glass electrodes and
calibration concept using the Memobase pH ISFET electrodes, but also sensors for
Plus software. ORP, conductivity, dissolved oxygen and
free chlorine. Memobase Plus is available
Calibration on the applications site takes in 12 languages and is linked to the W@M®
time. During the calibration you have no portal from Endress+Hauser. As a result,
monitoring of the process pH value. With professional life cycle management of all
Memosens you can simply replace the sensors used in the process is possible.
sensor in the process with a pre-calibrated
50
5.2 Fully automatic measuring, calibration, sterilization and
cleaning
51
5. Life cycle management of pH loops
5.3 Lifetime of pH electrodes
You might be wondering why there are so while dilution of the reference solution
many different pH electrodes and options changes the potential of the reference
available. The answer lies in the measuring system. The latter effect is the reason why a
principle. On the one hand the sensing pH electrode has to be calibrated regularly.
element, glass or ISFET (ion selective field Problems associated with water egress of
effect transistor) is directly in contact connectors or ground leaks are no longer an
with the medium. Any deposits, abrasive issue due to Memosens technology.
particles, mechanical stress and aggressive
chemicals will have an impact on the What is the lifetime of the sensor? It
measurement accuracy and/or life time of depends, sensor lifetime is based on sensor
the sensor. Additionally, the diaphragm choice, cleaning intervals and of course the
of the pH electrode brings the reference application. Because of this pH electrodes
system in direct contact with the medium. have a life cycle and are therefore
Ions which react with the silver reference considered consumable items.
wire such as sulfides and cyanides can
destroy the reference system. Plugging of
the diaphragm interrupts the measurement
52
5.4 Accreditation for permanent buffer laboratory
53
5.5 Steam/water analysis systems
Steam production consumes a high amount Endress+Hauser offers the full scope of
of energy within industrial processes. equipment for the analysis of pure water
The usage of high quality water in boiler for such boiler applications. As pressure
applications of power plants and utility and temperature are in most cases too
departments prevents corrosion processes high to measure directly in the process a
and buildup. This ensures keeping the boiler sample conditioner is needed in front of
efficiency high and therefore contributes to the analytical panels. This is as well in the
energy saving. Endress+Hauser offering.
54
Notes
55
Supplementary documentation Links
• Parameter overview • Application Selection Software
FA00007C/24/en www.us.endress.com/applicator
• Overview of all components
www.us.endress.com/pH
• Memosens technology
www.us.endress.com/memosens
• Topclean
www.us.endress.com/cpc30
• Topcal
www.us.endress.com/cpc310
Endress+Hauser, Inc. Endress+Hauser Canada Endress+Hauser México S.A. de C.V. For international locations
2350 Endress Place 1075 Sutton Drive Fernando Montes de Oca 21 Edificio A Piso 3 visit: www.endress.com/worldwide
Greenwood, IN 46143 Burlington, ON L7L 5Z8 Fracc. Industrial San Nicolás
Tel: 317-535-7138 Tel: 905-681-9292 54030 Tlalnepantla de Baz
888-ENDRESS 800-668-3199 Estado de México
Fax: 317-535-8498 Fax: 905-681-9444 México
info@us.endress.com info@ca.endress.com Tel: +52 55 5321 2080
www.us.endress.com www.ca.endress.com Fax: +52 55 5321 2099
eh.mexico@mx.endress.com
www.mx.endress.com