Professional Documents
Culture Documents
Hill-Rom Centra - In-Service Manual PDF
Hill-Rom Centra - In-Service Manual PDF
Hill-Rom Centra - In-Service Manual PDF
“GROUND LOSS” When the GROUND LOSS Light PATIENT CONTROLS If the Patient Controls are not
LIGHT FLASHES flashes, check all electrical plugs NOT FUNCTIONING functioning, ensure the power
to make sure that they are prop- cord is plugged in and make
erly grounded. sure that the controls at the
Nurse Control Panel are not
Flashing may also be caused by locked out. Also, check the
incorrect wiring polarity of the
wall outlet (ie: the “hot” and
neutral wires reversed).
SideCom connector, located
on the bed frame below the
mattress deck, to ensure that
Centra Bed
®
2 3
STANDARD FEATURES STANDARD FEATURES
SIDERAILS
NIGHT LIGHT RESTRAINT STRAP HOLDERS DRAINAGE BAG HOLDERS
2 3
STANDARD FEATURES STANDARD FEATURES
NURSE CALL ACTIVATION
There are three situations in which a nurse call could be activated: NURSE CONTROL PANEL
1. Patient call. 2. Patient leaves bed when Bed Exit 3. Communication Cable is
System is engaged. unattached.
NOTE Patient-controlled calls can be turned off at the nurse’s station while the other signals are turned off at the bed.
The specific message or call will differ by the type of Nurse Call system.
NURSE CALL
To place a nurse call simply depress the NURSE CALL Button.A
yellow light will illuminate to indicate that a call has been sent.
LIGHTING
To activate indirect room light, press the up arrow button in
the Siderail. To activate the exam or reading light, press the
down arrow button in the Siderail.
4 5
STANDARD FEATURES STANDARD FEATURES
NURSE CALL ACTIVATION
There are three situations in which a nurse call could be activated: NURSE CONTROL PANEL
1. Patient call. 2. Patient leaves bed when Bed Exit 3. Communication Cable is
System is engaged. unattached.
NOTE Patient-controlled calls can be turned off at the nurse’s station while the other signals are turned off at the bed.
The specific message or call will differ by the type of Nurse Call system.
NURSE CALL
To place a nurse call simply depress the NURSE CALL Button.A
yellow light will illuminate to indicate that a call has been sent.
LIGHTING
To activate indirect room light, press the up arrow button in
the Siderail. To activate the exam or reading light, press the
down arrow button in the Siderail.
4 5
OPTIONAL UPGRADEABLE FEATURES ACCESSORIES
DYNAMICAIRE For patients assessed to be at-risk of pressure ulceration, the INFUSION SUPPORT SYSTEM (ISS) SLIDING PERMANENT IV POLE
SLEEP SURFACE Surface should be placed in the Prevention Mode. This is The system consists of a The Sliding Permanent IV
The DynamicAire accomplished by activating the PREVENTION switch Mount Installation Kit (face Pole is mounted to the
Sleep Surface has located on the footboard control panel. In the Prevention or top), Offset Bar, and frame of the bed at the
two patient care Mode the amount of air within the dual air-bladder system Transfer Pole. Components head section. The sturdy
modes, Prevention of the Surface is controlled by the Surface itself to reduce may be purchased pole features a unique
and Comfort. The interface pressure across the patient’s body, especially the together or separately. mechanism which allows
choice of mode is best determined through an assessment bony prominences. The Surface adjusts in response to the pole to be moved
The Mount is the structural
of the patient as to their risk of pressure ulceration. changes in patient and bed positions. along a runner to either
interface between the bed
For patients assessed not to be at-risk, the Surface can be side of the bed. The pole
When the DynamicAire Sleep Surface is placed in the frame and the pole
placed in the Comfort Mode. This gives the patient and itself can be raised or
Prevention Mode, the green Indicator Light turns on, assembly. It consists of an
caregiver control of the firmness/softness of the Surface. This lowered with one hand.
reminding caregivers that the Surface is in the proper adaptor block, stop plate
is accomplished by pressing the COMFORT control in the mode for the at-risk patient. When in the Prevention Mode, and bolts, and rotational
DynamicAire control panel. The firmness/softness of the the patient and caregivers no longer have control of the lock screw. The block may
Surface is then controlled by the patient through the firmness/softness of the Surface. be mounted to the top
mattress controls located on each head-end Siderail. or face of either side of the
NOTE Mattress overlays are not required on the
Similarly, the firmness/softness can be controlled by headboard base.
DynamicAire Sleep Surface and therefore should not be
caregivers through the FIRM and SOFT switch located in the
used on top of the Surface. To Install Pole Place the offset Pole into the ISS mount.
DynamicAire control panel.
Push the Transfer Pole into the Offset Bar. To secure system,
NOTE See page 8 for Surface cleaning instructions.
tighten the butterfly screw on the bed and the Offset Bar.
NOTE Insure that the Transfer Pole and Offset Bar are firmly
DYNAMICAIRE SLEEP SURFACE PATIENT CONTROLS secured to each other before removing them for transfers.
For bed models with lighting controls (383 and 385
SideCom) the firm/soft controls are housed in a pendant.
For bed models without lighting controls (384, 386, and 389 TRAPEZE SUPPORT BRACKET FRACTURE FRAME ADAPTORS
SideCom) the firm/soft controls are housed in the Siderail. The Trapeze Support The bed will accommodate
If the sleep surface is in Comfort mode, the patient can Bracket is mounted at the most fracture frame
regulate the firmness/softness of the mattress. head end of the bed frame equipment. For fracture
and will accommodate the frame equipment with 1⁄2''
installation of trapeze diameter support poles, the
equipment. The assembly is Hill-Rom adaptors can be
mounted with the provided used to allow the frame to
hex head screws and interface with the bed. The
BED EXIT SYSTEM SIDECOM lock-nuts. Holes are adaptors are placed in the
Two special sensor strips run The SideCom system is a modular communications provided on the bracket for IV sockets located at all
horizontally between the mattress package built into the Siderails within constant reach bolt-on or clamp-on type four corners of the bed.
deck and the mattress. When a of the patient. The basic package integrates bed controls trapeze units. The frame posts are inserted
patient’s weight is removed from and surface controls with the DynamicAire Sleep in each adaptor. Fracture
the mattress, the sensors send a Surface and ZoneAire Sleep Surface. The modular frame equipment with 3⁄4''
time-delayed electronic signal to the bed can be upgraded to nurse call, lighting, bed exit, and diameter support poles
nurse’s station if the bed is equipped entertainment. may be inserted into the IV
with the SideCom system. The signal sockets. An adaptor may be needed if additional frame
ENHANCEMATE
can vary to delay at 2, 4 or 6 post height is required.
The EnhanceMate allows
seconds. If desired, the system can be
patients who are unable to NOTE To determine whether adaptors are required, please
set up so that a signal will sound in
use Siderail embedded check with your Hill-Rom representative.
the patient’s room to indicate that
controls to operate bed
the patient left the bed.
functions and control the SIDERAIL PADS BED EXTENDER PATIENT PHONE
On/off, message time interval and room environment. The
audible bedside alarm controls are patient can operate the bed
located on the in one of two ways: voice
outside of the input or rehabilitation devices such as touchpads or sip and
SideCom unit behind puff. The patient can call the nurse, adjust the head and
a sliding panel in the knee sections of the bed, adjust the air surface, turn room
left head Siderail. This and reading lights on and off, operate the
system should be television, and send and receive phone calls.
used in conjunction
with your regular
safety procedures. Siderail pads provide a soft cover for The Bed Extender is mounted on The patient phone can be mounted
NOTE The Bed Exit the rails while maintaining access to the foot end of the bed to provide to either head-end Siderail for ease
system is intended as patient nurse controls. additional length for taller patients. of access by the patient. The phone
a reminder, not a restraint device. will also rest on the bedside cabinet
or overbed table for use by patient's
visitors.
6 7
OPTIONAL UPGRADEABLE FEATURES ACCESSORIES
DYNAMICAIRE For patients assessed to be at-risk of pressure ulceration, the INFUSION SUPPORT SYSTEM (ISS) SLIDING PERMANENT IV POLE
SLEEP SURFACE Surface should be placed in the Prevention Mode. This is The system consists of a The Sliding Permanent IV
The DynamicAire accomplished by activating the PREVENTION switch Mount Installation Kit (face Pole is mounted to the
Sleep Surface has located on the footboard control panel. In the Prevention or top), Offset Bar, and frame of the bed at the
two patient care Mode the amount of air within the dual air-bladder system Transfer Pole. Components head section. The sturdy
modes, Prevention of the Surface is controlled by the Surface itself to reduce may be purchased pole features a unique
and Comfort. The interface pressure across the patient’s body, especially the together or separately. mechanism which allows
choice of mode is best determined through an assessment bony prominences. The Surface adjusts in response to the pole to be moved
The Mount is the structural
of the patient as to their risk of pressure ulceration. changes in patient and bed positions. along a runner to either
interface between the bed
For patients assessed not to be at-risk, the Surface can be side of the bed. The pole
When the DynamicAire Sleep Surface is placed in the frame and the pole
placed in the Comfort Mode. This gives the patient and itself can be raised or
Prevention Mode, the green Indicator Light turns on, assembly. It consists of an
caregiver control of the firmness/softness of the Surface. This lowered with one hand.
reminding caregivers that the Surface is in the proper adaptor block, stop plate
is accomplished by pressing the COMFORT control in the mode for the at-risk patient. When in the Prevention Mode, and bolts, and rotational
DynamicAire control panel. The firmness/softness of the the patient and caregivers no longer have control of the lock screw. The block may
Surface is then controlled by the patient through the firmness/softness of the Surface. be mounted to the top
mattress controls located on each head-end Siderail. or face of either side of the
NOTE Mattress overlays are not required on the
Similarly, the firmness/softness can be controlled by headboard base.
DynamicAire Sleep Surface and therefore should not be
caregivers through the FIRM and SOFT switch located in the
used on top of the Surface. To Install Pole Place the offset Pole into the ISS mount.
DynamicAire control panel.
Push the Transfer Pole into the Offset Bar. To secure system,
NOTE See page 8 for Surface cleaning instructions.
tighten the butterfly screw on the bed and the Offset Bar.
NOTE Insure that the Transfer Pole and Offset Bar are firmly
DYNAMICAIRE SLEEP SURFACE PATIENT CONTROLS secured to each other before removing them for transfers.
For bed models with lighting controls (383 and 385
SideCom) the firm/soft controls are housed in a pendant.
For bed models without lighting controls (384, 386, and 389 TRAPEZE SUPPORT BRACKET FRACTURE FRAME ADAPTORS
SideCom) the firm/soft controls are housed in the Siderail. The Trapeze Support The bed will accommodate
If the sleep surface is in Comfort mode, the patient can Bracket is mounted at the most fracture frame
regulate the firmness/softness of the mattress. head end of the bed frame equipment. For fracture
and will accommodate the frame equipment with 1⁄2''
installation of trapeze diameter support poles, the
equipment. The assembly is Hill-Rom adaptors can be
mounted with the provided used to allow the frame to
hex head screws and interface with the bed. The
BED EXIT SYSTEM SIDECOM lock-nuts. Holes are adaptors are placed in the
Two special sensor strips run The SideCom system is a modular communications provided on the bracket for IV sockets located at all
horizontally between the mattress package built into the Siderails within constant reach bolt-on or clamp-on type four corners of the bed.
deck and the mattress. When a of the patient. The basic package integrates bed controls trapeze units. The frame posts are inserted
patient’s weight is removed from and surface controls with the DynamicAire Sleep in each adaptor. Fracture
the mattress, the sensors send a Surface and ZoneAire Sleep Surface. The modular frame equipment with 3⁄4''
time-delayed electronic signal to the bed can be upgraded to nurse call, lighting, bed exit, and diameter support poles
nurse’s station if the bed is equipped entertainment. may be inserted into the IV
with the SideCom system. The signal sockets. An adaptor may be needed if additional frame
ENHANCEMATE
can vary to delay at 2, 4 or 6 post height is required.
The EnhanceMate allows
seconds. If desired, the system can be
patients who are unable to NOTE To determine whether adaptors are required, please
set up so that a signal will sound in
use Siderail embedded check with your Hill-Rom representative.
the patient’s room to indicate that
controls to operate bed
the patient left the bed.
functions and control the SIDERAIL PADS BED EXTENDER PATIENT PHONE
On/off, message time interval and room environment. The
audible bedside alarm controls are patient can operate the bed
located on the in one of two ways: voice
outside of the input or rehabilitation devices such as touchpads or sip and
SideCom unit behind puff. The patient can call the nurse, adjust the head and
a sliding panel in the knee sections of the bed, adjust the air surface, turn room
left head Siderail. This and reading lights on and off, operate the
system should be television, and send and receive phone calls.
used in conjunction
with your regular
safety procedures. Siderail pads provide a soft cover for The Bed Extender is mounted on The patient phone can be mounted
NOTE The Bed Exit the rails while maintaining access to the foot end of the bed to provide to either head-end Siderail for ease
system is intended as patient nurse controls. additional length for taller patients. of access by the patient. The phone
a reminder, not a restraint device. will also rest on the bedside cabinet
or overbed table for use by patient's
visitors.
6 7
DYNAMICAIRE SLEEP SURFACE CLEANING INSTRUCTIONS SAFETY TIPS
REMOVAL CLEANING MEDICAL FLUID SPILLS BED POSITIONS ELECTRICAL SAFETY
To remove the sleep surface, raise the foot end of the bed Fluid spills should be wiped up as soon as possible. Some Always leave the bed in the low position when the patient Policies and procedures must be established to train and
frame manually and disconnect all four quick disconnect fluids used in the hospital environment, such as betadine, will is unattended. This could reduce the possibility of patient educate your staff on the risks associated with electric
hoses and detach the strap. leave a permanent stain, but most will not. falls and the severity of resultant injury. equipment. At any time, it is not prudent or necessary for
Temporary stains can be removed by wiping vigorously with SIDERAILS/RESTRAINTS/PATIENT MONITORING staff personnel to have their entire body within the confines
To reconnect, replace mattress onto the frame with the hoses
a rag or sponge that is wet with a standard household The Siderails should always be in a full upright position and of the bed. Whenever a bed is being cleaned or serviced it
at the foot end of the bed. Insert the hoses and the strap
cleaning solution. Whatever cleaning solution is used, it must latched when a patient is unattended. When raising the should be unplugged from its power source. If service
through the bed frame and re-attach the strap
be diluted according to the manufacturer’s instructions. Siderails, an audible “click” should indicate that the Siderails personnel need to get under the bed, the Hi-Lo portion
and hoses.
are completely raised and locked in place. must be blocked up as an added precaution (refer to
NOTE Never remove the mattress from the bed frame CLEANING BLOOD AND EXCRETA Centra Bed Service Manual).
without disconnecting the hoses and straps at the foot of If possible, excesss blood and excreta should be wiped up Hill-Rom recognizes that certain healthcare situations may
when wet, since the cleaning process is more difficult when indicate the need for specialized Siderail configurations. In LOCKOUT CONTROLS
the bed.
these substances are dry and caked onto the ticking. response to this need, we offer, upon request, several Whenever a patient should be restricted from operating
NOTE Before the bed is cleaned, and especially before the patient controls, activate the appropriate Siderail
Siderail accessories.
touching any electrical components, it should be Cleaning of blood and excreta from the ticking can be Lockouts in the Nurse Control Panel at the foot end of the
disconnected from the electrical outlet. Since the bed is accomplished by wiping vigorously with rag or sponge that Siderails are intended to be a reminder, not a patient bed. This is especially true for patients in traction since it can
electric, the excess use of cleaning fluids should be is wet with a standard household cleaning solution. restraining device. Hill-Rom recommends the appropriate prevent a patient or visitor from accidentally
avoided. medical personnel determine the level of restraint activating a control, which would cause the bed to move
The rag or sponge must be rinsed periodically until, after
necessary to ensure a patient will remain safely in bed. and possibly result in an injury. The Siderail Lockouts are for
NOTE Care must be taken to secure the foot section to the wiping, the rag or sponge shows no evidence of the
Consult the restraint manufacturer’s instructions for use to the convenience of the staff and the safety of the patient
head section during any of the above procedures to substance being cleaned. Rinsing may be required five to
verify the correct application of each restraining device. and should be used when necessary.
ensure that the foot section does not drop back onto the ten times for puddled, dried blood.
frame while cleaning. Whenever “high profile” patients (typically, the frail, elderly INSTANT CPR RELEASE
A final wipe should be done using clean disinfecting
and medicated or confused) are involved, Hill-Rom The emergency head release,“Instant CPR Release,” is to
solution, and a clean rag or sponge. Allow 30 to 60 minutes
CLEANING recommends the following minimum actions: be used by health care professionals only.
for the ticking to dry completely.
To clean under the frame, manually lift the foot end frame 1. Develop guidelines for all high profile patients that indicate: BED EXIT SYSTEM
until it’s secure. Whatever cleaning or disinfecting solution is used, it must be
diluted according to the manufacturer’s instructions. • Which patients may need to be restrained and the The Bed Exit System can assist the nurse in monitoring
To clean directly beneath the sleep surface, lift the mattress appropriate restraint to utilize. certain patient activities that could lead to patient falls.
from the head end of the bed. DAMAGE REPAIR • The proper method to monitor a patient, whether
Repair of tears or holes in the upper mattress cover section restrained or not, including time interval, visual check of PARTS AND ACCESSORIES
GENERAL CLEANING is not recommended due to leak potential of the repair. restraint, etc. Use only Hill-Rom parts and accessories. Do not modify the
General cleaning of the sleep surface mattress cover should 2. Develop training programs for all caregivers concerning bed without authorization from Hill-Rom.
be done by wiping down with a sponge or rag that has Tears or punctures in the air chambers can usually be
repaired. A repair kit is available from Hill-Rom’s Service the proper use and application of restraints. MATTRESSES
been dampened in a cleaning solution. 3. Maintain the bed at its lowest position whenever a The use of mattresses other than Hill-Rom’s may substantially
Organization. An air chamber can withstand up to 15
The cleaning solution can be any standard household typical needle punctures without significantly affecting caregiver is not in the room. reduce the effectiveness of the safety features and systems
cleaner. For disinfecting purposes, any phenolic or performance. 4. Clarify the need for restraint devices to families or incorporated into Hill-Rom beds.
quaternary type cleaners may be used. Whatever cleaning guardians.
PREVENTIVE MAINTENANCE
solution is used, it must be used and diluted according to BRAKES Annual preventive maintenance must be performed to
the manufacturer’s instructions. Brakes should always be set when the bed is occupied and insure all bed features are functioning as originally designed.
Clorox® brand bleach can be used for disinfecting purposes, in position, especially during patient transfer. Patients often Particular attention must be addressed on safety features,
but should not be used in concentrations greater than 1 use the bed for support when getting out of bed and could including but not limited to:
part bleach to 10 parts water. be injured if the bed unexpectedly moves. After setting the • Siderail latching mechanisms.
brake, push and pull the bed sideways to ensure stability. • Caster braking systems.
Care should be taken to wipe up excess cleaning solutions.
FLUIDS • Electrical cords and components.
Damp areas can take 15 to 30 minutes to dry.
When spills occur on the Siderails or the foot end areas of • All controls return to OFF or neutral position when released.
LINENS • Controls or cabling entanglement of bed mechanisms
the bed, immediately:
Although the bottom sheet can be flat, fitted sheets are in Siderails.
• Clean the fluid from the bed.
recommended. Proper sizing of fitted sheets is important to • Proper operation of the Siderail Lockout functions.
• Check the bed controls (i.e. Head, Knee, Hi-Lo) to see if
insure good retention. The best retention will be achieved if
they are operable.
the fitted sheet extends 2 inches or more under the corners
• Have the maintenance department check the internal
of the surface.
electronics.
Magnets are used on the underside of the surface to assist • Fluids remaining on the electronic controls may cause
in the retention of additional sheets and blankets. Better corrosion, which may cause the electronic components
retention will be achieved if the number of layers folded to fail. These component failures may cause the bed to
under the magnets is minimized. move or operate on its own at a time that may be
injurious to the patient or staff.
8 9
DYNAMICAIRE SLEEP SURFACE CLEANING INSTRUCTIONS SAFETY TIPS
REMOVAL CLEANING MEDICAL FLUID SPILLS BED POSITIONS ELECTRICAL SAFETY
To remove the sleep surface, raise the foot end of the bed Fluid spills should be wiped up as soon as possible. Some Always leave the bed in the low position when the patient Policies and procedures must be established to train and
frame manually and disconnect all four quick disconnect fluids used in the hospital environment, such as betadine, will is unattended. This could reduce the possibility of patient educate your staff on the risks associated with electric
hoses and detach the strap. leave a permanent stain, but most will not. falls and the severity of resultant injury. equipment. At any time, it is not prudent or necessary for
Temporary stains can be removed by wiping vigorously with SIDERAILS/RESTRAINTS/PATIENT MONITORING staff personnel to have their entire body within the confines
To reconnect, replace mattress onto the frame with the hoses
a rag or sponge that is wet with a standard household The Siderails should always be in a full upright position and of the bed. Whenever a bed is being cleaned or serviced it
at the foot end of the bed. Insert the hoses and the strap
cleaning solution. Whatever cleaning solution is used, it must latched when a patient is unattended. When raising the should be unplugged from its power source. If service
through the bed frame and re-attach the strap
be diluted according to the manufacturer’s instructions. Siderails, an audible “click” should indicate that the Siderails personnel need to get under the bed, the Hi-Lo portion
and hoses.
are completely raised and locked in place. must be blocked up as an added precaution (refer to
NOTE Never remove the mattress from the bed frame CLEANING BLOOD AND EXCRETA Centra Bed Service Manual).
without disconnecting the hoses and straps at the foot of If possible, excesss blood and excreta should be wiped up Hill-Rom recognizes that certain healthcare situations may
when wet, since the cleaning process is more difficult when indicate the need for specialized Siderail configurations. In LOCKOUT CONTROLS
the bed.
these substances are dry and caked onto the ticking. response to this need, we offer, upon request, several Whenever a patient should be restricted from operating
NOTE Before the bed is cleaned, and especially before the patient controls, activate the appropriate Siderail
Siderail accessories.
touching any electrical components, it should be Cleaning of blood and excreta from the ticking can be Lockouts in the Nurse Control Panel at the foot end of the
disconnected from the electrical outlet. Since the bed is accomplished by wiping vigorously with rag or sponge that Siderails are intended to be a reminder, not a patient bed. This is especially true for patients in traction since it can
electric, the excess use of cleaning fluids should be is wet with a standard household cleaning solution. restraining device. Hill-Rom recommends the appropriate prevent a patient or visitor from accidentally
avoided. medical personnel determine the level of restraint activating a control, which would cause the bed to move
The rag or sponge must be rinsed periodically until, after
necessary to ensure a patient will remain safely in bed. and possibly result in an injury. The Siderail Lockouts are for
NOTE Care must be taken to secure the foot section to the wiping, the rag or sponge shows no evidence of the
Consult the restraint manufacturer’s instructions for use to the convenience of the staff and the safety of the patient
head section during any of the above procedures to substance being cleaned. Rinsing may be required five to
verify the correct application of each restraining device. and should be used when necessary.
ensure that the foot section does not drop back onto the ten times for puddled, dried blood.
frame while cleaning. Whenever “high profile” patients (typically, the frail, elderly INSTANT CPR RELEASE
A final wipe should be done using clean disinfecting
and medicated or confused) are involved, Hill-Rom The emergency head release,“Instant CPR Release,” is to
solution, and a clean rag or sponge. Allow 30 to 60 minutes
CLEANING recommends the following minimum actions: be used by health care professionals only.
for the ticking to dry completely.
To clean under the frame, manually lift the foot end frame 1. Develop guidelines for all high profile patients that indicate: BED EXIT SYSTEM
until it’s secure. Whatever cleaning or disinfecting solution is used, it must be
diluted according to the manufacturer’s instructions. • Which patients may need to be restrained and the The Bed Exit System can assist the nurse in monitoring
To clean directly beneath the sleep surface, lift the mattress appropriate restraint to utilize. certain patient activities that could lead to patient falls.
from the head end of the bed. DAMAGE REPAIR • The proper method to monitor a patient, whether
Repair of tears or holes in the upper mattress cover section restrained or not, including time interval, visual check of PARTS AND ACCESSORIES
GENERAL CLEANING is not recommended due to leak potential of the repair. restraint, etc. Use only Hill-Rom parts and accessories. Do not modify the
General cleaning of the sleep surface mattress cover should 2. Develop training programs for all caregivers concerning bed without authorization from Hill-Rom.
be done by wiping down with a sponge or rag that has Tears or punctures in the air chambers can usually be
repaired. A repair kit is available from Hill-Rom’s Service the proper use and application of restraints. MATTRESSES
been dampened in a cleaning solution. 3. Maintain the bed at its lowest position whenever a The use of mattresses other than Hill-Rom’s may substantially
Organization. An air chamber can withstand up to 15
The cleaning solution can be any standard household typical needle punctures without significantly affecting caregiver is not in the room. reduce the effectiveness of the safety features and systems
cleaner. For disinfecting purposes, any phenolic or performance. 4. Clarify the need for restraint devices to families or incorporated into Hill-Rom beds.
quaternary type cleaners may be used. Whatever cleaning guardians.
PREVENTIVE MAINTENANCE
solution is used, it must be used and diluted according to BRAKES Annual preventive maintenance must be performed to
the manufacturer’s instructions. Brakes should always be set when the bed is occupied and insure all bed features are functioning as originally designed.
Clorox® brand bleach can be used for disinfecting purposes, in position, especially during patient transfer. Patients often Particular attention must be addressed on safety features,
but should not be used in concentrations greater than 1 use the bed for support when getting out of bed and could including but not limited to:
part bleach to 10 parts water. be injured if the bed unexpectedly moves. After setting the • Siderail latching mechanisms.
brake, push and pull the bed sideways to ensure stability. • Caster braking systems.
Care should be taken to wipe up excess cleaning solutions.
FLUIDS • Electrical cords and components.
Damp areas can take 15 to 30 minutes to dry.
When spills occur on the Siderails or the foot end areas of • All controls return to OFF or neutral position when released.
LINENS • Controls or cabling entanglement of bed mechanisms
the bed, immediately:
Although the bottom sheet can be flat, fitted sheets are in Siderails.
• Clean the fluid from the bed.
recommended. Proper sizing of fitted sheets is important to • Proper operation of the Siderail Lockout functions.
• Check the bed controls (i.e. Head, Knee, Hi-Lo) to see if
insure good retention. The best retention will be achieved if
they are operable.
the fitted sheet extends 2 inches or more under the corners
• Have the maintenance department check the internal
of the surface.
electronics.
Magnets are used on the underside of the surface to assist • Fluids remaining on the electronic controls may cause
in the retention of additional sheets and blankets. Better corrosion, which may cause the electronic components
retention will be achieved if the number of layers folded to fail. These component failures may cause the bed to
under the magnets is minimized. move or operate on its own at a time that may be
injurious to the patient or staff.
8 9
TROUBLESHOOTING
PROBLEM SOLUTION PROBLEM SOLUTION
“GROUND LOSS” When the GROUND LOSS Light PATIENT CONTROLS If the Patient Controls are not
LIGHT FLASHES flashes, check all electrical plugs NOT FUNCTIONING functioning, ensure the power
to make sure that they are prop- cord is plugged in and make
erly grounded. sure that the controls at the
Nurse Control Panel are not
Flashing may also be caused by locked out. Also, check the
incorrect wiring polarity of the
wall outlet (ie: the “hot” and
neutral wires reversed).
SideCom connector, located
on the bed frame below the
mattress deck, to ensure that
Centra Bed
®