Professional Documents
Culture Documents
Troubleshooting: Fully Automated Urine Chemistry Analyzer
Troubleshooting: Fully Automated Urine Chemistry Analyzer
UC-3500
Troubleshooting
Read this manual when you encounter a problem, and to perform instrument maintenance.
The explanations in this manual assume that you have already read “General Information”.
Revision History
05/2016
Initial issue
Software version: V0.49
08/2016
Software version: Ver.1
03/2017
Software version: Ver.1
06/2017
Software version: Ver.1
10/2017
Software version: Ver.1
12/2017
Software version: Ver.1
05/2018
Software version: Ver.1
09/2019
Software version: Ver.1
3/56
UC-3500 Troubleshooting
Revision History
4/56
UC-3500 Troubleshooting
Table of Contents
Table of Contents
Revision History 3
Chapter 1 Troubleshooting 7
1.1 Types of errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2 Analysis error code list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.3 Trouble messages and remedial actions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
1.4 Warning message and remedial actions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
1.5 Causes of errors and remedial actions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
1.6 Remedial actions against U-RDU failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Chapter 2 Maintenance 23
2.1 List of maintenance tasks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
2.2 [MAINTENANCE] screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
2.3 Maintenance information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
2.4 Cleaning the dripping nozzle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
2.5 Cleaning the waste duct . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
2.6 Inspecting/cleaning the test strip holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
2.7 Disposing of analyzed test strips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
2.8 Disposing of waste solution in the waste bottle . . . . . . . . . . . . . . . . . . . . . . . . 32
2.9 Cleaning the washing solution bottle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
2.10 Inspecting around the test strip reaction line . . . . . . . . . . . . . . . . . . . . . . . . . . 33
2.11 Specific gravity calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
2.12 Replenishing test strips. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
2.13 Replenishing washing solution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
2.14 Disposing of waste solution in the waste bottle . . . . . . . . . . . . . . . . . . . . . . . . 46
2.15 Disposing of analyzed test strips in the dust box . . . . . . . . . . . . . . . . . . . . . . . 48
2.16 Replacing the printer paper . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
2.17 Replacing the fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
2.18 Managing for long term non-use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
2.19 Maintenance inspection checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Index 55
5/56
UC-3500 Troubleshooting
Table of Contents
6/56
UC-3500 Troubleshooting
Chapter 1 Troubleshooting
Chapter 1 Troubleshooting
● Analyzer
Analysis error code
Indicates that abnormalities occurred during analysis.
(➤P.8 “1.2 Analysis error code list”)
Trouble message
Indicates that abnormalities of functions occurred regardless of whether or not its during analysis. When an
analysis is being performed, the analysis stops immediately.
(➤P.12 “1.3 Trouble messages and remedial actions”)
Warning message
Indicates that analysis cannot start or continue due to the lack of test strip or washing solution, or that analysis
results could not be output to external media or the host computer.
(➤P.14 “1.4 Warning message and remedial actions”)
● Sampler section
Indicates an error number on the error display LED on the sampler section if an error occurs in the sampler
section.
7/56
UC-3500 Troubleshooting
Chapter 1 Troubleshooting
Analysis abnormalities
Error contents
Actions
Code Description Detailed information*
1100 Mechanical Trouble <a, 1> Remove racks from the sampler section
and restart the instrument.
<b, 1> Check if the reflection measurement
section moves back and forth during
analysis.
2200 Bad Sampling <a, 1, 2, 3> • Check the amount of sample (1 to 10
mL).
• Remove excessive foam on the
sample liquid surface.
• Check if the sample conductivity is
enough.
• Check if the dripping nozzle touches
the sample tubes during sample
aspiration.
<b, 1, 2> Check if the caps remain attached to the
sample tubes.
AA00 Nozzle Clogging <a, 1, 2> • Check if the viscosity of the sample is
too high.
• Centrifuge the sample.
BB00 Test Strip Edge <a, 1, 2, 3, 4> • Check if the number of test strip in the
Detection Error test strip receptacle does not exceed
<b, 1, 2, 3, 4> 100.
• Check if the desiccant was not
<c, 1, 2, 3, 4> inserted with the test strips.
• Loosen the test strip to prevent it from
popping up.
<d, 1> Check if the test strip holder is dirty.
EE00 Sampling Request <a> Check the connection with the carrier
Not Received system.
* It is not displayed on the touch panel, but displayed when outputting.
8/56
UC-3500 Troubleshooting
Chapter 1 Troubleshooting
Test strip
Error contents
Actions
Code Description Detailed information*
4020 Foreign Object on <a, 1, 2, 3> Check if test strip is around the test strip
Internal Standards <b, 1, 2, 3, 4, 5> reaction line.
(➤P.33 “Chapter 2: 2.10 Inspecting
<c, 1, 2, 3>
around the test strip reaction line”)
<d, 1, 2, 3, 4, 5>
<e, 1, 2, 3, 4>
<f, 1, 2>
<g, 1, 2>
<h, 1, 2>
5010 Test Strip Position <a> • Check if the test strip holder is dirty.
Inappropriate <b, 1, 2> (➤P.29 “Chapter 2: 2.6 Inspecting/
cleaning the test strip holder”)
<c, 1>
• Check if test strip is around the test
5020 No Test Strip in Test <a, 1, 2, 3>
strip reaction line.
Strip Holder
(➤P.33 “Chapter 2: 2.10 Inspecting
5030 Insufficient Light <a, 1, 2, 3> around the test strip reaction line”)
Quantity (base is
insufficient)
5040 Cannot Find Test <a, 1, 2, 3>
Strip Edge
6010 Too Narrow Pitch <a, 1, 2, 3, 4> • Check if test strip other than specified
Between Pads <b, 1, 2, 3, 4> is used.
• Check if foreign objects are adhered
6020 Too Wide Test Strip <a, 1, 2>
to the test strip.
Edge Opening
• Check if the test strip holder is dirty.
6030 Too Wide Pitch <a, 1, 2>
Between Pads
6040 Reversed Test Strip <a, 1> Check the orientation of the test strip in
the test strip receptacle.
6050 Test Strip Type <a, 1, 2, 3, 4> Check if the type is correct.
Differs from
Specified (dropping
near the final pad)
7010 Cannot Find Pad <a, 1, 2, 3> Check if the LED on the reflective
photometry unit is lit.
7020 Test Strip Type <a, 1, 2, 3, 4> Check if the registered test strip type
Differs from <b, 1, 2, 3, 4> and the test strip type in the test strip
Specified receptacle match.
<c, 1, 2, 3, 4>
<d, 1, 2>
<e, 1, 2>
* It is not displayed on the touch panel, but displayed when outputting.
9/56
UC-3500 Troubleshooting
Chapter 1 Troubleshooting
Error contents
Actions
Code Description Detailed information*
7030 Test Strip Set <a, 1, 2, 3, 4, 5> Check the orientation of the test strip in
Backwards the test strip receptacle.
7050 Cannot Find Line <a, 1, 2> Check if test strip other than specified is
Mark used.
8010 Dropping Failure <a, 1, 2, 3, 4> Check if there is fluid leakage or cracks
(insufficient supply) <b, 1, 2, 3, 4> in tubes.
Flow cell
Error contents
Actions
Code Description Detailed information*
0301 Insufficient R Light None Remove racks from the sampler section
0302 Insufficient G Light None and restart the instrument.
10/56
UC-3500 Troubleshooting
Chapter 1 Troubleshooting
Other
Error contents
Actions
Code Description Detailed information*
00BB Barcode Read <a> Check the barcode label
Error (orientation, type, digits, dirt, blood
<b, 1>
stained, blurred or insufficient label
adhesion).
00CC Order Error <a> Check if data transmission from the host
computer takes too much time.
<b, 1> • Check if the communication cable is
connected properly.
• Check if there are abnormalities in the
host computer program.
<c> Check if the ID whose order is registered
by the host computer and the barcode
label match.
<d> Check if the test strip type (case,
number of spaces) registered by the
host computer is correct. (case, number
of spaces)
<e, 1> Match the test strip type registered by
the host computer and the test strips
used.
<f> Check if there are abnormalities in the
host computer program.
00DD Test Strip Expiration Check if test strip that has expired is
Error being used.
* It is not displayed on the touch panel, but displayed when outputting.
11/56
UC-3500 Troubleshooting
Chapter 1 Troubleshooting
Note:
Trouble message will be output to the built-in printer during abnormalities even if [REAL-TIME OUTPUT] is
set to [NO] on the [INT.PRINTER PARAMETER] screen.
Trouble contents
Check and remedial actions
Message Detailed information*
PROGRAM ERROR <a, 1> Check the surrounding environment (noise,
<b> unstable power, etc.).
<c, 1, 2>
MOTOR CONTROLLER <a, 1, 2>
<b, 1, 2, 3>
FILE SYSTEM <a, 1, 2, 3>
<b, 1>
SROM ERROR <a, 1, 2>
<b, 1, 2, 3>
LAMP LEVEL <a, 1, 2> Check if the LED on the reflective photometry
unit is lit.
IMAGE CAPTURE CIRCUIT <a, 1> Remove racks from the sampler section and
<b, 1> restart the instrument.
<c, 1, 2>
<d>
<e>
<f, 1, 2>
CHAIN CONVEYER <a, 1> Check if there are any foreign objects around
the test strip reaction line or something
<b, 1>
hindering carriage.
<c, 1>
PICKUP S MOTOR <a, 1> Remove racks from the sampler section and
<b, 1> restart the instrument.
12/56
UC-3500 Troubleshooting
Chapter 1 Troubleshooting
Trouble contents
Check and remedial actions
Message Detailed information*
PICKUP PRESSURE <a, 1, 2> Remove racks from the sampler section and
PICKUP FACE SENSOR <a, 1, 2, 3, 4> restart the instrument.
MOTOR DRIVER <a, 1, 2, 3, 4> Remove racks from the sampler section and
<b, 1, 2, 3, 4> restart the instrument.
TOUCH PANEL <a, 1> Remove racks from the sampler section and
restart the instrument.
<b> • Check if there are foreign objects on the
touch panel.
• Restart without touching the touch panel.
CAPACITY SHORTAGE <a> Remove racks from the sampler section and
restart the instrument.
STRIP EDGE SENSOR <a, 1, 2> Remove racks from the sampler section and
restart the instrument.
<b, 1, 2> Check if there are foreign objects under the test
strip edge detection sensor.
* It is not displayed on the touch panel, but displayed when outputting.
13/56
UC-3500 Troubleshooting
Chapter 1 Troubleshooting
Message Actions
<<TEST STRIP EMPTY>> Replenish test strips.
PLEASE RELOAD AND PRESS [ENTER]. (➤P.37 “Chapter 2: 2.12 Replenishing test strips”)
<<WASHING SOLUTION EMPTY>> Prepare the washing solution for replenishment.
DON'T REFILL THE BUFFER TANK (➤P.44 “Chapter 2: 2.13 Replenishing washing solution”)
YET.
PLEASE PRESS [ENTER].
<<WASHING SOLUTION EMPTY>> Replenish the washing solution while keeping the message displayed.
PLEASE REFILL AND PRESS (➤P.44 “Chapter 2: 2.13 Replenishing washing solution”)
[ENTER].
<<WASTE SOLUTION FULL>> Wait (15 to 60 seconds) until disposal is enabled.
DON'T REMOVE THE WASTE (➤P.46 “Chapter 2: 2.14 Disposing of waste solution in the waste
SOLUTION YET. bottle”)
PLEASE PRESS [ENTER].
<<WASTE SOLUTION FULL>> Dispose of the waste bottle contents while keeping the message
PLEASE REMOVE THE WASTE displayed.
SOLUTION AND PRESS [ENTER]. (➤P.46 “Chapter 2: 2.14 Disposing of waste solution in the waste
bottle”)
<<NO RACK>> • Place the sample rack in the right sampler pool.
PLEASE SET THE SAMPLE RACK • Ensure that the grooves on the bottom right of the sample rack
AND PRESS [ENTER]. match the rails on the right side of the right sampler pool.
(➤Basic Operation “Chapter 3: 3.3 Preparing samples”)
<<NO TEST STRIP BOTTLE>> Place the test strip receptacle containing test strips used for analysis.
PLEASE SET THE TEST STRIP (➤P.37 “Chapter 2: 2.12 Replenishing test strips”)
BOTTLE AND PRESS [ENTER].
<<STRIPS NOT SPECIFIED>> Set the type of test strip for preferential use when using 2 or more
PLEASE SPECIFY THE TEST STRIP types of test strip.
TO BE USED AND PRESS [ENTER]. (➤P.41 “Chapter 2: 2.12.3 Setting test strip for priority use”)
<<CALIBRATION ERROR>> Arrange the sample tubes with specific gravity correction fluid in the
NO SAMPLE TUBE. sample rack from the left without spaces.
PLEASE PRESS [ENTER]. (➤P.34 “Chapter 2: 2.11 Specific gravity calibration”)
<<CALIBRATION ERROR>> Put 10 mL of the specific gravity correction fluid in the sample tube.
NO CALIBRATION BUFFER. (➤P.34 “Chapter 2: 2.11 Specific gravity calibration”)
PLEASE PRESS [ENTER].
<<CALIBRATION ERROR>> Clean the dripping nozzles.
CALIB. BUFFER CANNOT BE (➤P.26 “Chapter 2: 2.4 Cleaning the dripping nozzle”)
SUCKED OUT.
PLEASE PRESS [ENTER].
<<CALIBRATION ERROR>> Please contact the Service Department of your local Sysmex service
LIQUID OR LIGHT IS INSUFFICIENT. representative.
PLEASE PRESS [ENTER].
14/56
UC-3500 Troubleshooting
Chapter 1 Troubleshooting
Message Actions
<<CALIBRATION ERROR>> Line up the specific gravity correction fluid in order of L/M/H gravities.
MEASURED VALUE INCONSISTENT (➤P.34 “Chapter 2: 2.11 Specific gravity calibration”)
WITH THE STANDARD.
PLEASE PRESS [ENTER].
<<CELL BLANK>> • Check if the washing solution tube is filled with washing solution.
[YES]:ABORT • Check if there is solution leakage or cracks in the washing solution
[NO]:IGNORE tube.
THE SAME ERROR HAPPENED Refer to the analysis error code list and take appropriate remedial
REPEATEDLY FOR 5 TIMES. actions.
PLEASE RESOLVE THE PROBLEM (➤P.8 “1.2 Analysis error code list”)
AND PRESS [ENTER].
<<MEASURING>>
PLEASE CHECK MEASUREMENT
RESULT.
PLEASE PRESS [ENTER].
<<WRITE IN FAILURE>> • Check if it is not write-protected.
PLEASE VERIFY THE EXTERNAL • Check if the capacity is full.
MEDIA WITH THE PERSONAL
COMPUTER.
PLEASE PRESS [ENTER].
<<ONLINE ERROR>> Please contact the Service Department of your local Sysmex service
AFTER RECEIVING [NAK], representative.
EXCEEDED THE NUMBER OF
RETRANSMISSIONS.
PLEASE PRESS [ENTER].
<<ONLINE ERROR>>
AFTER RECEIVING OTHER THAN
[ACK] OR [NAK], EXCEEDED THE
NUMBER OF RETRANSMISSIONS.
PLEASE PRESS [ENTER].
<<ONLINE ERROR>>
HOST IS NOT RESPONDING.
PLEASE PRESS [ENTER].
CLOSE COVER OF THE BOTTLE Close the bottle cover.
EXCHANGE PORT.
15/56
UC-3500 Troubleshooting
Chapter 1 Troubleshooting
Risk of infection
• Be sure to wear adequate personal protective equipment, such as protective gloves, a protective mask,
protective eyewear, and a lab coat when taking remedial actions for errors. Wash your hands with
antiseptic solution after completing the task.
There is a risk of infection by pathogens or other contaminants.
• Take appropriate care in handling samples. In the event that an infectious material gets in the eyes or an
open wound, rinse with large amounts of water and seek immediate medical attention.
Display
The display does not change after touching a touch panel item
Probable cause Actions
Case 1: Remove the object from the touch panel.
An object is on the touch panel.
Case 2: Clean the touch panel.
The touch panel is dirty.
16/56
UC-3500 Troubleshooting
Chapter 1 Troubleshooting
Analysis
The analysis result is wrong
Probable cause Actions
Case 1: Check if the test strip type is specified for the
The test strip used is inappropriate. instrument.
(➤General Information “Chapter 6: Supplies and
Accessories”)
Case 2: Check if the registered test strip type and the set test
The test strip type is not set properly. strip type in the test strip container match.
When 2 or more test strips are used, make sure to set
the priority test strip.
(➤P.41 “Chapter 2: 2.12.3 Setting test strip for priority
use”)
Case 3: Set the [NO.] or [#.] series for normal analysis.
The settings of measurement series are inappropriate.
17/56
UC-3500 Troubleshooting
Chapter 1 Troubleshooting
18/56
UC-3500 Troubleshooting
Chapter 1 Troubleshooting
At the time of U-RDU installation, we provide you with a set of parts to be used in the event of a U-RDU failure.
Prepare sufficient wiring cord for the number of instruments connected to the U-RDU.
If you are connecting 6 instruments, prepare 6 changeover wiring cords.
19/56
UC-3500 Troubleshooting
Chapter 1 Troubleshooting
Note:
To open/close the reagent tube, roll the wheel on the roller clamp.
Open Close
(1) ANALYZER 1
(1) (2) ANALYZER 2
(2) (3) ANALYZER 3
(3) (4) ANALYZER 4
(4) (5) ANALYZER 5
(7)
(5) (8) (6) ANALYZER 6
(6)
(7) REAG #1
(8) REAG #2
(9) (9) REAG #3
Yellow (10)
tube
(10) REAG #4
20/56
UC-3500 Troubleshooting
Chapter 1 Troubleshooting
1 Prepare the washing solution bottle for 3 Connect the washing solution tube
UC-3500 and dispensing set. coming from the dispensing set, to the
Refill the washing solution bottle for UC-3500 with connector where you just removed the
saline, and attach the dispensing set. tube above.
Caution!
• Add only the saline specified in General
Information “Chapter 7 Reagents and Test
Strips” to the washing solution bottle. When
filling, take care that foreign objects do not get
on the dispensing set and other parts.
There is a risk of instrument failure and
inaccurate analysis results.
(➤General Information “Chapter 7: 7.5 Washing
solution (saline)”)
• Rinse the inside of the bottle once a month.
Tube
21/56
UC-3500 Troubleshooting
Chapter 1 Troubleshooting
1 Remove the green connector from the 4 After the instrument starts up, touch
level sensor connection port (LS). [PRIME WASH] twice on the
[MAINTENANCE] screen to fill the
instrument with washing solution.
The instrument can be used once the instrument is
filled with washing solution.
Information
When multiple UC-3500 units are connected to the
U-RDU, similarly perform the procedures in
“1.6.3 Switching the connection of the washing
solution tube” and “1.6.4 Switching the
connection of the connector” for the other UC-
3500 units.
Information
Keep the removed connector so as not to lose it.
The connector is necessary when using the U-RDU
again.
22/56
UC-3500 Troubleshooting
Chapter 2 Maintenance
Chapter 2 Maintenance
This chapter explains an overview of instrument maintenance and those procedures, replacing and
replenishing supplies, and checking the operation of this instrument.
Risk of infection
Be sure to wear adequate personal protective equipment, such as protective gloves, a protective mask,
protective eyewear, and a lab coat when working. Wash your hands with antiseptic solution after
completing the task.
There is a risk of infection by pathogens or other contaminants.
23/56
UC-3500 Troubleshooting
Chapter 2 Maintenance
[MAINTENANCE] screen
[S.G. CALIBRATION] Set values to replace analysis values when reporting specific gravity calibration or
specific gravity analysis results using specific gravity fluid.
(➤P.34 “2.11 Specific gravity calibration”)
[INFORMATION] Outputs maintenance information to the built-in printer or external media.
(➤P.25 “2.3 Maintenance information”)
[PRIME WASH] Washing solution is continuously sent to the duct until filled.
Used for cleaning the dripping nozzle and washing tank, and removing solution from
washing solution duct.
(➤P.26 “2.4 Cleaning the dripping nozzle”)
[MOVE STRIP Touch [EXECUTE] to send test strip reaction line at the interval set in [TIME
HOLDER] INTERVAL].
Automatically stops after 2 rotations.
(➤P.29 “2.6 Inspecting/cleaning the test strip holder”)
[FLOW WASH] Touch [AUTO WASH] to aspirate CELLCLEAN U from the sample tube, and auto rinse
the waste duct.
(➤P.29 “2.6 Inspecting/cleaning the test strip holder”)
24/56
UC-3500 Troubleshooting
Chapter 2 Maintenance
Note:
A Sysmex service representative may ask for [SAVE COMMUNICATION LOG] or [MAINTENANCE
INFORMATION OUTPUT] when there are analysis errors or trouble.
[INFORMATION] screen
[PRINT BASIC INFO.] Outputs adjusted operation log and instrument values (operation count for
each unit) to the built-in printer.
[PRINT TROUBLE INFO.] Outputs trouble logs to the built-in printer.
[PRINT ERROR INFO.] Outputs analysis error logs to the built-in printer.
[PRINT SETTINGS] Outputs values of each screen to the built-in printer.
[MAINTENANCE Outputs communication logs with the host computer and sample carrier
INFORMATION OUTPUT] system saved with [SAVE COMMUNICATION LOG] to external media.
[SAVE COMMUNICATION LOG] Saves communication logs with the host computer and sampler section on
the instrument. Outputs the saved content to external media with
[MAINTENANCE INFORMATION OUTPUT].
[LOGON LOG OUTPUT] Outputs logon information to external media.
[STRIP CHANGE LOG Outputs reagent replacement log to external media.
OUTPUT]
[VERSION INFORMATION] Displays instrument version information.
25/56
UC-3500 Troubleshooting
Chapter 2 Maintenance
Warning!
Do not insert your hand in the dripping arm section while the instrument is in operation. Doing so may result
in injury.
Information
When nozzle edges contact sample tubes due to bending or position gaps, contact your local Sysmex
service representative.
26/56
UC-3500 Troubleshooting
Chapter 2 Maintenance
Warning!
Be sure to wear adequate personal protective
equipment, such as protective gloves, a protective
mask, protective eyewear, and a lab coat when
opening the seal of CELLCLEAN U. Avoid contact
with skin, eyes, and clothing, because
CELLCLEAN U is a strong alkaline detergent. In
the event that CELLCLEAN U comes into contact Analysis mode indicator LED
with your body or clothes, rinse immediately with
copious amounts of running water. CELLCLEAN U
may cause skin injury and damage to clothing.
If CELLCLEAN U gets in the eye, rinse with large Mode switch button
amounts of water, and seek immediate medical
attention.
Note:
Set only the necessary number of sample tubes,
because all the sample tubes inserted into the
sample rack are aspirated.
Caution!
• Load the rack horizontally with respect to the
instrument. If loaded at an angle, correct
operation cannot be assured.
• Check that no dirt or foreign objects are in the
left and right sampler pools and the analysis
line. Also check that no dirt or foreign objects
are stuck to the bottom of the rack and that the
rack is not damaged or deformed.
27/56
UC-3500 Troubleshooting
Chapter 2 Maintenance
28/56
UC-3500 Troubleshooting
Chapter 2 Maintenance
Warning!
Do not touch moving test strip holder. Fingers may become caught.
Caution!
• Never touch parts inside the instrument besides the test strip holder.
• Analysis quality may deteriorate if cleaned improperly.
• Use neutral detergent when using detergent.
Using any other detergents may deteriorate the surface or cause the instrument to malfunction.
• Use paper towels or other cloth without dust when cleaning.
3 Touch [EXECUTE].
The test strip holder moves at the designated time
interval.
Fastener
29/56
UC-3500 Troubleshooting
Chapter 2 Maintenance
Risk of infection
Be sure to wear adequate personal protective equipment, such as protective gloves, a protective mask,
protective eyewear, and a lab coat when disposing of analyzed test strips. Wash your hands with antiseptic
solution after completing the task.
There is a risk of infection by pathogens or other contaminants.
Information
Turn OFF the instrument power before replacing the test strip dust box.
Note:
A message is displayed on the touch panel when the number of analyzed test strips reaches the number
set in [DUST COUNT] during analysis. Check the test strip waste section. The amount of analyzed test
strips can be checked from the test strip waste section view port.
(➤P.48 “2.15 Disposing of analyzed test strips in the dust box”)
View port
1 Press the upper part of the test strip 2 Remove the test strip dust box with
waste section. analyzed test strips.
The lock is released to open the test strip waste
section.
30/56
UC-3500 Troubleshooting
Chapter 2 Maintenance
3 Insert a new test strip dust box. 4 Close the test strip waste section.
31/56
UC-3500 Troubleshooting
Chapter 2 Maintenance
Risk of infection
• Be sure to wear adequate personal protective equipment, such as protective gloves, a protective mask,
protective eyewear, and a lab coat when disposing of waste solution. Wash your hands with antiseptic
solution after completing the task.
There is a risk of infection by pathogens or other contaminants.
• Waste solution and used protective gloves must be separated from general waste and treated according
to applicable local and national regulations.
32/56
UC-3500 Troubleshooting
Chapter 2 Maintenance
1 Remove the washing solution bottle cap 4 Replenish the washing solution.
and pull out the electrode.
Information
Use the saline specified in “General Information” for
the washing solution.
(➤General Information “Chapter 7: 7.5 Washing
solution (saline)”)
Electrode
5 Insert the electrode into the washing
solution bottle and close the bottle cap.
6
7
8
2 Discard the washing solution in the
bottle. 9
10
3 Wash the inside of the washing solution 11
bottle with purified water. 12
13
2.10 Inspecting around the test strip reaction line
Analysis errors may be caused by analyzed test strips around the test strip reaction line.
Inspect around the test strip reaction line and remove any objects.
1 Turn OFF the power. 3 Remove any test strips or objects that
have fallen around the test strip reaction
line.
2 Open the right side cover.
Test strip reaction line
33/56
UC-3500 Troubleshooting
Chapter 2 Maintenance
Caution!
• Specific gravity correction fluid (SG Calibrator) should be used at room temperature.
• Do not use specific gravity correction fluid other than the specified one.
Note:
• The correction values for the specific gravity meter (refractive index specific gravity analysis) are saved
in the instrument even if power is turned OFF, making it unnecessary to calibrate every day.
• It is recommended to perform periodic calibration (once a month) to maintain analysis efficiency.
[MAINTENANCE] screen
34/56
UC-3500 Troubleshooting
Chapter 2 Maintenance
7 Touch [EXECUTE].
Note: The analysis starts.
35/56
UC-3500 Troubleshooting
Chapter 2 Maintenance
Note:
The refractive index is a large value in order from L
(low), M (middle) and H (high).
10 Touch [YES].
[*** ADOPTED ***] is output to the built-in printer.
Note:
The range of specific gravity values is 1.000 to
1.050. When the obtained result is out of the
analysis range, 0.999 or below is reported as
1.000, and 1.051 or above as 1.050.
36/56
UC-3500 Troubleshooting
Chapter 2 Maintenance
1 Check the type of insufficient test strip 5 Match the placement positions in the
and touch [ENTER]. bottle rack and touch [REGISTER] on the
The [BOTTLE EXCHANGE] screen appears and the top of the test strip container display.
bottle rack is unlocked. The [REAGENT EXCHANGE] dialog box appears.
37/56
UC-3500 Troubleshooting
Chapter 2 Maintenance
Information
Keep the removed desiccant in a dry and well-closed
container such as a used test strip container
temporarily.
38/56
UC-3500 Troubleshooting
Chapter 2 Maintenance
Caution!
Clean the test strip receptacle once every 3 days.
Analysis errors may be generated or correct analysis results may not be obtained due to the effects of dust
from the test strip (filter paper dust).
1 Check the type of insufficient test strip 5 Match the placement positions in the
and touch [ENTER]. bottle rack and touch [REGISTER] on
The [BOTTLE EXCHANGE] screen appears and the the top of the test strip receptacle
bottle rack is unlocked. display.
The [REAGENT EXCHANGE] dialog box appears.
2 Open the bottle cover.
7 Touch [OK].
Information about the test strip appears on the bottle
display.
39/56
UC-3500 Troubleshooting
Chapter 2 Maintenance
9 Open the lid of the test strip receptacle 11 Put the test strips into the test strip
desiccant receptacle, and insert the receptacle.
desiccant.
Information
Information Insert with the padded side facing down.
Use the desiccant in the test strip container to add.
Desiccant
40/56
UC-3500 Troubleshooting
Chapter 2 Maintenance
1 Press the BTL button while the [MEASURE] or [MENU] screens are displayed.
The [BOTTLE EXCHANGE] screen appears.
Note:
When all of the test strip types registered for the test strip receptacles are the same, they are not checked
and the priority cannot be changed.
41/56
UC-3500 Troubleshooting
Chapter 2 Maintenance
Note:
Priority is effective for different types of test strip. When there is the same type of test strip, all of them are
checked and the priority is set to the registered order.
42/56
UC-3500 Troubleshooting
Chapter 2 Maintenance
[MEASURE] screen
3 Touch the test strip for priority use and place a check.
4 Touch [SAVE].
Sets the test strip for priority use.
43/56
UC-3500 Troubleshooting
Chapter 2 Maintenance
Note:
When the analysis is completed, this warning message is not displayed. Proceed to step 2.
Electrode
44/56
UC-3500 Troubleshooting
Chapter 2 Maintenance
Information
Use the saline specified in “General Information” for
the washing solution.
(➤General Information “Chapter 7: 7.5 Washing
solution (saline)”)
5 Touch [ENTER].
When test strip removal restarts, analysis is
continued.
45/56
UC-3500 Troubleshooting
Chapter 2 Maintenance
Note:
When the analysis is completed, this warning message is not displayed. Proceed to step 2.
46/56
UC-3500 Troubleshooting
Chapter 2 Maintenance
Caution!
Place the waste bottle below the bottom of the
instrument.
5 Touch [ENTER].
When test strip removal restarts, analysis is
continued.
47/56
UC-3500 Troubleshooting
Chapter 2 Maintenance
1 Press the upper part of the test strip 3 Insert a new test strip dust box.
waste section.
The lock is released to open the test strip waste
section.
5 Touch [ENTER].
Test strip dust box The counter for the number of accumulated test strip
in the dust box is reset to “0”.
(➤Basic Operation “Chapter 5: 5.3.3 Setting warning
for analysis error”)
48/56
UC-3500 Troubleshooting
Chapter 2 Maintenance
Information
Use only specified printer paper.
Using non-designated paper may lower printing quality or printer malfunctions.
New printer
paper
2 Remove the printer paper shaft. 4 Close the printer paper cover.
Pull the edge of the printer paper a few centimeters
Printer paper shaft from the cutter section of the printer until it clicks, and
firmly close the printer paper cover.
Thermal head
Warning!
Do not touch inside the printer or the cutter section.
It may cause printer malfunctions or injuries. 5 Touch [NO PAPER].
Immediately after printing, the thermal head and
[PAPER FEED] is displayed when printer paper is fed.
surrounding area is hot and may cause burns.
Output restarts when there is data waiting to be
printed.
49/56
UC-3500 Troubleshooting
Chapter 2 Maintenance
Warning!
• Make sure to unplug the power cable before replacing a fuse.
Otherwise, there is a risk of electrical shock.
• Use a fuse of the specified type and rating.
Otherwise, an electrical malfunction could result causing smoke or a fire.
Information
If the fuse blows again immediately after it is replaced, there may be a fault in the instrument. Contact your
local Sysmex service representative.
1 Confirm the start/stop button lamp of 3 Unplug the power cable from the power
the analyzer is not lit. outlet.
2 Turn OFF the power switch of the 4 Unplug the power cable plug.
analyzer.
50/56
UC-3500 Troubleshooting
Chapter 2 Maintenance
Power switch
Hook
Fuse holder
2 Remove the old fuse from the fuse
holder.
Fuse
Fuse holder
51/56
UC-3500 Troubleshooting
Chapter 2 Maintenance
Caution!
Store in a location where the instrument will be protected from high temperature, humidity, fire, dust and
direct sunlight.
1 Remove the washing solution bottle cap 6 Touch [PRIME WASH] on the
and pull out the electrode. [MAINTENANCE] screen.
The washing solution is removed from the duct.
Electrode
8 Clean the section that touches samples.
Sterilize or clean the section that touches samples
well to prevent the growth of bacteria.
(➤P.27 “2.5 Cleaning the waste duct”)
52/56
UC-3500 Troubleshooting
Example:
UC-3500 Troubleshooting
Cleaning the waste duct
Inspecting/cleaning the test strip
holder
Disposing of analyzed test strips
Disposing of waste solution
Signed
* We recommend that our customers prepare a checklist that suits their operating environment.
Maintenance
53/56
Chapter 2 Maintenance
54/56
UC-3500 Troubleshooting
Index
Index
A T
Analysis error code list.........................................8 Test strip holder cleaning and inspection .......... 29
Analysis abnormalities.....................................8 Test strip replenishment .................................... 37
Flow cell.........................................................10 Trouble messages and remedial actions........... 12
Other..............................................................11 Types of errors .................................................... 7
Test strip ..........................................................9
Analysis settings U
Specific gravity calibration .............................34 Using test strip container................................... 37
Analysis trouble error ...........................................8 Using test strip receptacle................................. 39
Analyzed test strips disposal..............................30
W
C Warning messages and remedial actions ......... 14
Causes of errors and remedial actions ..............16 Washing solution bottle cleaning....................... 33
Analysis .........................................................17 Washing solution replenishment ....................... 44
Display...........................................................16 Waste bottle waste solution disposal .......... 32, 46
Output............................................................17 Waste duct cleaning .......................................... 27
Print ...............................................................17
D
Disposing the test strips dust box ......................48
Dripping nozzle cleaning....................................26
E
Errors related to analysis ...................................17
Errors related to display .....................................16
Errors related to flow cell ...................................10
Errors related to output ......................................17
Errors related to print .........................................17
Errors related to test strip.....................................9
I
Inspection around the test strip reaction line .....33
M
Maintenance ......................................................23
Task list..........................................................23
Maintenance information....................................25
Maintenance inspection checklist ......................53
[MAINTENANCE] screen ...................................24
Managing for long term non-use ........................52
P
Printer paper replacement .................................49
R
Remedial actions against U-RDU failure ...........19
Replacing the fuses ...........................................50
S
Setting test strip for priority use .........................41
Specific gravity calibration .................................34
55/56
UC-3500 Troubleshooting
Index
56/56
UC-3500 Troubleshooting