Professional Documents
Culture Documents
Sem
Sem
• Assuming g scattering
gggives yyou contrast,, the
shorter the wavelength of the radiation, the
better your resolution
0.61λ
∂=
μ sin β
• Things to remember!
– e-
e have mass - can transfer momentum
momentum, behave
relativistically at high energies
– e- are waves - can constructively and destructively
interfere
Electron scattering
• 2p
primary
y classes of electron scattering:
g
– elastic scattering - incident electron will change
direction, but not kinetic energy, scales with Z2
– inelastic scattering - incident electron will change
direction, but also give up some kinetic energy to
the sample,
p , scales with Bethe expression
p
• will allow for elemental analysis
• Ultimately,
y scattering
g will lead to the image
g
contrast
Electron scattering probability
• Scattering
g cross section:
N 1
σ= ∝ 2
nincident nt argg ets E
• Mean free path:
MW
λ=
N A ρσ
• Total mean free path:
1 1 1
= + + ....
λtotall λa λb
Interaction Volumes
• From the mean free path, we can calculate
the “interaction
interaction volume”
volume of electrons and the
samples
– dependent upon incident electron energy, the
sample atomic mass, the sample density, and the
angle of incidence of the incoming electrons
Interaction Volumes - accelerating
voltage dependence (SEM samples)
2kV
10kV
30kV
Interaction Volumes - atomic mass
dependence
Al
Fe
W
Interaction Volumes - sample density
dependence
Au 100% dense
A 50% dense
Au
Au 25% dense
Interaction Volumes - incident
electron beam angle dependence
0 deg
g
(normal)
45 deg
60 deg
Elastic Scattering - SEM signals
• Most elastic scattering signals in the SEM are evident as
“Backscattered” electrons - comparable to reflection microscopy
• No energy
energ loss,
loss just
j st change in direction d
due
e to scattering and diffraction
- large dependence on surface topography and atomic number
Al
W
Elastic Scattering - TEM signals
Al
• Magnification
– M = Lcrt / Lspecimen
– Magnification depends on the size of
your monitor and the resolution you have
it set
– If monitor
it size
i andd resolution
l ti are fifixed,
d
can only change mag by changing
Lspecimen (decreasing the length of
beam scan on the sample)
• ONLY dependent upon the scan
coils, NOT the OL strength
More Magnification
• e- travel down electron column in helical path, should result in
image rotation when WD or focal length is changed
– Does not occur due to compensation by the scan coils
• Magnification calibration
– Completed using ASTM standards
– Also adjusts scan coils for squareness and roundness
• Digital Image Acquisition
– Pixel definition
definition, size
– Sampling: Pixel is not necessarily a fundamental unit of the specimen
being sampled - depends on spot size, interaction volume
Imaging - Depth of Field
• Depth of Field - biggest advantage of SEM
– Defined as the distance (D) between where the
crossover occurs and the beam broadening is
large enough that the broadening (r) is noticeable
((+ and -))
r 2r
tan α = D≈
D /2 α
• Types:
– C
Cold
ld cathode
th d (b
(best,
t needd
high vacuum)
– Thermal (bigger energy
spread))
– Schottky (thermal, but good
energy spread like cold
cathode, small current)
• Vacuum considerations:
– At 10-5 Pa, one monolayer
of gas is formed per second
– Run vacuum at 10-9 Pa to
avoid changes in Ew
– “flash” heat the tip to bake
off gas
– Use Schottk
Schottky FE
FEG Tip Materials
• Single
g crystal
y W sharpened
p to ~100nm radius
– W good due to high mechanical strength
• Needed due to high strains put on materials
under high fields
• How to increase e- current?
– W crystal
t l direction
di ti <310>310 tto give
i llowestt work k
function direction the best look at the first anode
– Coat W crystal with ZrO (Ew = 2.8eV)
Electron Lenses
• How direct electrons into a point? – Electromagnets
B = magnetic
ti flflux d
density,
it f(I)
B = H in vacuum
F = -e (B x v) using right hand rule
Iron core wrapped N times with a
conductor, pass I (current) through
the conductors
B is
i hhomogeneous b
between
t th
the
gaps
Produces “Fringing
Fringing Fields”
Fields near
the optic axis - steer the e-
Lenses - General
• Electron source
• Condenser and
Objective Lenses
• Faraday
F d C Cup
• Steering lens
• Ray traces
• Detectors
• Electronics
• Imaging
Control of Lenses
• Effects of working
di t
distance (WD)
– Longer WD, larger beam
diameter
Control of Lenses
• Effects of Condenser
lens strength
– Stronger CL, smaller spot
size
Lens Aberrations
Aperture Diffraction
Astigmatism
Spherical aberration
Chromatic aberration
Astigmatism
Imaging Defects
• Tilt distortion
– Mag is lower
perpendicular to tilt
axis (more specimen
is mapped) than
parallel to the tilt axis
– Tilt correction
– Dynamic
D i ffocus
correction
• Moire effects
• Scan distortion
– Misaligned scan coils
give rise to distortions
at the edges of the
scan
Imaging Detectors - SE
• Everhart - Thornley
Detector
– Converts e- into
light, then back into
electrical signals
– Positively biased
Faraday cage
attracts low E SEs
– Also collect BSEs
present within line
of sight
More SE Imaging
Al
Fe
W
Backscattered electrons - atomic
mass dependence
for pure
p
compounds
2kV
10kV
No significant
accelerating voltage 30kV
dependence on BSE
BSE – tilt angle dependence
0 deg
(normal)
Higher tilts,
tilts more BSE
60 deg
BSE detectors
Solid-state detector
-p-n junction like a solar
cell
-detect electronic signal to
determine contrast
Solid State BSE detectors
• Flat, thin detectors placed directly
above sample, allows for high BSE
collection
ll ti efficiency
ffi i
• Signal is sent through an amplifier
to change the brightness and
contrast
t t off the
th image
i
SE1 is the HR
signal - need to
minimize the
other signals for
HR imaging
More HRSEM
• Signal to pixel ratios
– At low mag (<1,000x), pixels are generally smaller than the
interaction volume
– At medium mag (10,000x), overlap of pixels and interaction
volumes begins to occur
• For 1nm, 10kV probe into Si, interaction volume is 10,000x
larger than the probe size
– At high mag (100,000x), FOV is smaller than the interaction
volume, BSE and SE2 signals get averaged out in the noise
– Bright edges help with contrast
• HRSEM at low kV
– Interaction volume smaller
– Most of the emitted e e- have high spatial res
– Brightness falls off too, must balance
Setting up HRSEM
1. Reduce contamination!!
2. Thin, fine grained metal coating (increase
SE1)
3. Small spot size
4. Low kV ((balance with spot
p size to maintain
contrast)
5. Short WD
6. Snorkel lens
7. TTL detector
8. STIGMATE!!!!
HRSEM
Low-k / Ru surface
Stereomicroscopy
• SEM has large depth of field compared to LM
• 3-D info is lost when displaying a 2-D image
– Can be recovered by stereomicroscopy
• Stereomicroscopy:
py
– Collect (2) 2-D images at different tilts (6° separation) -
sample must be at eucentric height!
– Displaying stereopairs:
• Blue / red 3-d glasses (Image J, instructions in text)
• Stereoviewer used with p printed imagesg
– Can also be done quantitatively using geometry to determine
3-D information about the samples
20 μm
10 μm
200 μm
Voltage and Magnetic Contrast
• If high local fields are present, they
mayy be strong g enough
g to either
significantly repel e- or attract them
back into the surface
• Voltage Contrast
– Useful for failure analysis of
electronic devices - bias the device
in the SEM, look for bright and dark
areas to see where the device is
failing
• Which areas will appear bright
vs. dark?
– Also visible during “charging”
Magnetic Contrast - Type 1
• Interactions of SE1
with leakage B
F = -ev x B
Magnetic Contrast - Type 2
• Interactions of e- with B
inside the specimen
Crystallographic Contrast and
Channeling
• e- can be Bragg diffracted by
individual crystals (grains)
– Directionality induced by
diffraction gives rise to light / dark
areas
– Contrast determination?
Threshold equation?
– Best to use large solid angle BSE
detector
• e- channeling can also give rise to
diffraction patterns
– Strongly affected by surface
topography
– Better to determine
crystallography with EBSD
(discussed later in the semester)