X-Mind Control Panel

17
ON / OFF MAIN SWITCH

CONTROL BUTTON

18 DISPLAY
ON/OFF
X-RAY KEY

7
RADIOGRAPHIC
DISTANCE INDICATOR

1 8
KEY TO DECREASE KEY TO INCREASE
EXPOSURE TIME EXPOSURE TIME

9
2 TUBEHEAD
TUBEHEAD TYPE SELECTION
INDICATOR

10
3 SELECTION OF TYPE
RADIOGRAPHIC OF PATIENT’S
VOLTAGE PHYSIQUE
INDICATOR

4
MEMORY

11
5 RADIOGRAPHIC
MAXILLARY CURRENT INDICATOR
LOWER TEETH

12
6 OCCLUSAL EXAM
MANDIBULARY
LOWER TEETH 13
POSTERIOR EXAM

X-RAY OUTPUT 14
SIGNAL BITE-WING EXAM

15
PAUSE INDICATOR DIGITAL
RADIOGRAPHIC
TECHNIQUE

MALFUNCTIONING 16
INDICATOR CONVENTIONAL
RADIOGRAPHIC
TECHNIQUE

17
KEY SWITCH

Trouble shooting 1 X-Mind ® DC


ERROR
CAUSE SOLUTION
MESSAGES
Check the “dip switch N°1” position.
Check the wiring:
RX1 Tube head is not connected or is out of
- C11; C12 connexion
order
- X RAY 1 connexion
- Tube head and pantograph connexions
Check the “dip switch N°3” position.
Check the wiring:
RX2 Tube head is not connected or is out of
- C21; C22 connexion
order
- X RAY 2 connexion
- Tube head and pantograph connexions
Check the software configuration
Corrupted EEPROM data  Press (5); (6) & (12)
Call Customer service
Check the software configuration
EEPROM data not saved properly  Press (5); (6) & (12)
Call Customer service
Check the Tube Head identification tag (XXX)
Check the Timer identification tag (YYY)
Line voltage value not included within the  Press (5); (6) & (13)
limit  Check the line voltage
 Press (5); (6) & (14)
 Check the system voltage
Check the Tube Head identification tag (XXX)
Check the Timer identification tag (YYY)
 Press (5); (6) & (13)
Line voltage value not included within the
 Check the line voltage
±15% nominal value
 Press (5); (6) & (14)
 Check the system voltage
 Take the measurement with a multimeter
The X-Ray key seems to be always pressed Make sure it is not jammed
Check the flex connexion between the timer PCB and
Anomaly on control panel
front panel
The exposure have been prematurely
Keep the trigger button pressed till the end of exposure
interrupted
Anomaly in the triac relay Call Customer service

Anomaly in electronic circuit Call Customer service

Anomaly in control circuit Call Customer service

Incorrect dip switch configuration Check the configuration


The control button does not correspond to
Chey Key 1 and Key 2 connexion
the selected tube head.
Check the wiring:
- C21; C22 connexion
The tube head do not work properly
- X RAY 2 connexion
Tube head and pantograph connexions
Set the “dip switch N°2” in ON position
The tube head is not in the correct mode
 ON is DC / OFF is AC
The tube head didn’t complete the exposure Repeat the exposure

X-Mind® DC Trouble shooting 2


Call Customer service
Problem in frequency regulation Call Customer service
Turn off the timer
Turn on keeping current key pressed
Select the tubehead to be calibrated with tube head
selection key (9)
Take the exposure X-Ray X-Ray until the acoustic
The tube head is not calibrated
signal stops

During this operation, X-Ray is active, so, it is


mandatory to adopt all the safety measures relevant to
radioprotection
Check the software configuration
EEPROM data not saved properly  Press (5); (6) & (12)
 Call Customer service

Check the software configuration


Corrupted EEPROM data  Press (5); (6) & (12)
Call Customer service
Overvoltage error Call Customer service

Anode voltage out of tolerance Call Customer service

Anode current out of tolerance Call Customer service


Check the wiring
Connexion problem
Call Customer service
Problem with reference voltage Call Customer service
All function disabled
Major error
Call Customer service

Tube head identification tag:

Fabbricato da de Götzen® S.r.l. OLGIATE OLONA (VA) - ITALY


model X-Mind® dc
70kVp 8mA Total filtration: 2mm Al at 70kV

Nominal Voltage: (XXX) V ±15% Class I Type B IP20


S/N: Manufactured Date:

Tube TOSHIBA DG-073B-DC S/N: 0.7


Complies with DHHS 21 CFR Subchapter J

Timer identification tag:


Fabbricato da de Götzen S.r.l. OLGIATE OLONA (VA) -ITALIA
®

model X-Mind®
Line Voltage: (YYY) V ∼ 50-60Hz Class I Type B IP20
Momentary current: A max Permanent current: A
S/N: Manufactured Date:
Complies with DHHS 21 CFR Subchapter J

Trouble shooting 3 X-Mind® DC


X-Mind® DC Trouble shooting 2