Bagmatic Novo: Service Manual For
Bagmatic Novo: Service Manual For
for
BAGMATIC NOVO
Version 1.0
September 2013
www.lmb.de
info@lmb.de
Service Manual – Bagmatic NOVO
PREFACE
Producer
Producer ....................................................................................................3
1 Overall description ...............................................................................5
2 Installation and handling ......................................................................6
3 Settings ................................................................................................7
3.1 USER MENU .............................................................. 7
3.2 SERVICE MENU.......................................................... 8
3.2.1 Balance calibration.................................................................... 9
3.2.2 Tube sensor calibration........................................................... 10
3.2.3 Clamp calibration .................................................................... 11
3.2.4 Clamp testing .......................................................................... 11
3.2.5 Communication test ................................................................ 12
3.2.6 Information about device ........................................................ 12
3.2.7 Battery info .............................................................................. 12
3.2.8 Balance test ............................................................................ 13
3.2.9 Motor test ................................................................................ 13
3.3 SUPERVISOR MENU .................................................. 13
3.3.1 Language menu ...................................................................... 14
3.3.2 Procedure configuration .......................................................... 15
3.3.3 Parameter configuration.......................................................... 17
3.3.4 DATA MANAGEMENT............................................................ 19
3.3.5 DATE & TIME ......................................................................... 19
3.3.6 PROFILES .............................................................................. 20
4 Troubleshooting .................................................................................21
4.1 BALANCE FAILED...................................................... 21
4.2 BATTERY ERROR: ..................................................... 21
4.3 SD CARD ............................................................... 22
4.4 CURRENT OF MOTOR FOR TRAY IS OVERLOADED ............... 23
4.5 BATTERY ERROR (BATTERY COMMUNICATION ERROR) ....... 23
4.6 WRONG WEIGHT ON TRAY .......................................... 24
4.7 OVER WEIGHT ERROR................................................ 24
4.8 PAUSE ................................................................... 25
4.9 STOP .................................................................... 25
4.10 LANGUAGE ERROR .................................................... 26
4.11 TUBE SENSOR.......................................................... 26
5 Wireless connectivity, infrastructure and SW operating requirements
27
6 Data Storage Information ...................................................................30
7 Battery Specification ..........................................................................32
4 LMB_DOC_S-MANUAL_BAGMATIC-NOVO – Revision status 00 / 09.13 – version 1.0 / September 2013
Service Manual – Bagmatic NOVO
1 Overall description
Use
Installation
- Unpack the device.
- Check the delivery content in accordance with delivery order and delivery
note.
- Connect the bar code reader device. Install barcode reader before switching
on.
- Connect supply with power outlet 230 V. (Recharge the battery first in case
the power supply indicates a red light)
- Put the tray on the swing mechanism. Pay attention that the pins on the
bottom of the tray snap in their counterparts in the swing mechanism.
- Press the button of the bar code reader for test. The barcode reader will
beep if properly connected.
- Plug the serial cable into the backside port RS 232 or connect LAN or USB
cable of the Bagmatic NOVO and then connect the opposite side of the
cable with a PC for the data transfer.
3 Settings
F1 F3
F1 F2
F3
F2 F4 F4
- Blood Taking
- Profile
- Last Result
- Turn Off
F1 F3
F2 F4
To enter the service menu press buttons F1 and F4 together and quick
after that F2 and F3.
+
Note: Interval between pressing the two buttons combinations should be less than
1 second!
Service Menu consists of the following functions:
CALIBRATION – balance calibration
TUBE SENSOR – tube sensor calibration
CLAMP – clamp testing
CLAMP CAL. – clamp calibration
COM TEST – communication check
DEVICE INFO – serial no, HW and SW info
BATTERY INFO – battery info
LOAD TEST – battery test
MOTOR – motor test
Place the device on stable and flat surface and remove everything from bag
tray.
Choose CALIBRATION in the menu and press NEXT (screen below).
F1 F3
F2 F4
When screen below appear, make sure that tray free and press NEXT.
F1 F3
F2 F4
Choose calibration weight (from 100g until 500g) and put it on bag tray. Wait for
5 seconds for the balance to stabilize and press NEXT.
F1 F3
F2 F4
F1 F3
F2 F4
F1 F3
F2 F4
If error appears, make sure that set calibration weight is the same like used
weight and then repeat procedure.
F1 F3 F1 F3
F2 F4 F2 F4
Screen above (left) will be shown, and with plus and minus buttons
tolerance (threshold) can be set. If tube is present current will be increased, if
tube is not present current will be decreased. D (difference) represents which
difference is needed between currents for tube detection. After pressing button F3
calibration will be done, and device will show on screen values. If difference is less
then set value, tube is not present in clamp and check box in right upper corner of
screen is not marked. Otherwise, if difference is bigger then set, it will be marked.
For better overview first remove the clamp cover. In menu choose CLAMP
CAL. option and press NEXT. First set the upper position of the clamp by pressing
the button F1 to move it and F2 to accept the position. Next step is to adjust the
lower position also using button F1 for movement and F3 for accepting the
position. If the procedure was successful it will be shown on screen message.
Select COM TEST in menu and press NEXT. By pressing the button F3
device will send the ACK message to server.
Select DEVICE INFO option in menu and press NEXT. On screen you can
see: serial number, software version, hardware version and donation counter.
Maximal number of donations is one million.
Select BATTERY INFO from menu and press NEXT. On screen you will see
following information:
1 – Relative charge status of the battery – ST
2 – Approximate time until empty battery – AT
3 – Battery capacity – C
4 – Battery voltage – V
5 – Battery serial number – SN
Select MOTOR option in menu and press NEXT. Motor will turn ON
automatically. Display will show PWM and time motor need for one full circle.
Buttons F1 and F2 are changing PWM value. Button F3 is used to exit to main
menu. If you exit the menu while the motor is not in initial position device will
turn on flashing green light and move the motor to initial position.
F1 F2
F3 F4
To enter the Supervisor Menu press buttons F1 and F4 together and quick
after that, F2 and F4.
Interval between pressing the 2 buttons combinations should be less than
1 second!
SUPERVISOR MENU:
LANGUAGE
PROC. CONFIG.
PARAM. CONFIG.
DATA MAN.
TIME&DATE
PROFILES
Select LANGUAGE option in menu and press NEXT. There are three options:
1 – ENGLISH
2 – USER
3 – COPY
Set clamp
This option enables us to control clamp status at the start of the donation. If this
option is checked clamp will be closed at the start of the donation.
Display in
This option enables us to choose display in grams or milliliters.
Flow alarm
Turn on/off alarm for too fast/slow flow.
Total weight
Bruto (checked) or neto (unchecked) measurement of weight.
To fast flow
Will be implemented in next software release
Stop alarm
Signalizes donation end (when checked).
Bidirectional
On/Off for bidirection.
Allow donation
If bidirection is turned on and there are no data for donor on the server, or donor
is not allowed to donate blood this option will allow the operator to continue with
donation process using the settings of current profile.
Hemoglobin
Turn On/Off barcode reading for haemoglobin.
Communication
Turn On/Off communication with server.
Mini flow
This option allows us to set up minimal allowed flow during donation process.
Buttons F3 and F4 are used to set up the value.
Maxi flow
This option allows us to set up maximal allowed flow during donation process.
Buttons F3 and F4 are used to set up the value.
Sleep time
Time in minutes. After this period of time, if device is in main menu and not used,
device will turn off automatically.
Interrupt time
Maximal pause time that will not cause error report. Time in seconds.
Buttons F3 and F4 are used to set up the value.
PC Port
Used to choose type of communication with server. Currently only RF
communication is available.
Start tilting
This option defines at which weight motor will start to work. Buttons F3 and F4
are used to set up the value.
Speed1, Speed2, Speed3, Speed4
This options defines at which weight motor will change speed. The larger the
weight, the slower the speed of rotation. Buttons F3 and F4 are used to set up the
value.
ID
During communication with server device needs to have valid ID. It should not be
allowed to two devices to have the same ID. Buttons F3 and F4 are used to set up
the value.
ID-0
Number of devices connected to one concentrator (device used for radio
communication).
SYS. WE.
Maximal starting weight on bag tray before starting donation.
SAVE
Saves the settings in memory.
Select option DATA MAN. and press next to enter barcode settings menu with
following options:
1.DATA PER DAY
2.DATA AT START
3.DATA DURING
4.DATA AT END
If user needs to read same barcode it is needed to enter desired menu and check
desired barcodes for reading. For example, if user needs to read haemoglobin
barcode at the end on donation then in submenu 4 he needs to check
haemoglobin barcode for reading.
This menu is used for setting up the date and time in the device. Buttons F1 and
F2 are used for selecting DATE or TIME and then F3 to confirm.
In submenu, button F1 is used to select the field and F3 and F4 to change value.
Button F2 is used to confirm entered value.
3.3.6 PROFILES
Submenus are:
1. Active
2. EDIT
3. VIEW
4. SAVE
Active
Select the profile you want to edit or use.
EDIT
Submenus:
1. Volume – blood volume
2. N.DUR. – nominal donation time
3. M.DUR. – maximal donation time
4. BAGNo – number of bags (1-7)
5. TubeNo – number of tubes (1-7)
6. SAVE – save selected settings in memory
VIEW
View the selected profile
Save
Save the settings in device memory.
4 Troubleshooting
1.Battery error!
2.BattAdrNack error!
3.BattDatNack error!
4.ReadErrInfo error!
4.3 SD Card
In this case please turn off device and call service or technical
support.
Repeat this action again and if device again shows same massage
turn off device and contact technical support.
Picture 7 shows over weight error. This means that if flow is much
greater than defined and it lasts longer than 3 seconds, device
will show this error (picture). User should check tray and bugs
system, and then presses button F3 if he accepts this condition.
Then device will check tray, and if weight in allowed range
donation will be continued. Otherwise, if user presses button F4,
donation will be stopped. If weight o tray in allowed range, but
device shows this message please contact technical support.
4.8 Pause
4.9 Stop
Operating Wireless
BNC ERYSYS
Systems Link (v2.02) (v 3.6.1.0)
(v4.0)
Windows 8 (64-
bit only)
Windows 7
Windows Vista
Windows XP
Minimum Hardware
RAM 2 GB or more
Other
Anti-Virus TCP ports (80, 15000) must be open in firewall
Additional
Adobe Reader 10 or newer
software needed
Regular work:
7 Battery Specification
Description:
Rechargeable Smart Lithium Ion Battery Pack
General Precautions
Handling
Avoid shorting the battery
Storage
Store in a cool, dry and well-ventilated area.
Disposal
Regulations vary for different countries. Dispose of in
accordance with local regulations.
Nominal Voltage
The battery nominal operating voltage is 14.4V.
Rated Capacity:
The initial capacity is ≥ 2755mAh (based on a CV charge of
16.8V ±50mV with a current limit of 1500mA and a 580mA
discharge to 10.00V @ 25C, within 1 hour of charge).
Initial Impedance
The internal impedance of a fully charged battery shall be <
240mΩ when measured across the positive and negative battery
terminals at 1kHz at 20°C.
Discharge
Discharge Temperature Limits: As shown below, ≤ 80%RH
Charge
Charge Temperature Limits: As Shown below, ≤ 80%RH
The battery shall be capable of continuous charge at 16.8V, as
shown in the graph below.
Storage
Storage Temperature Limits: -20°C to 60°C, ≤ 80%RH
The battery packs should be stored in an environment with low
humidity, free from corrosive gas at a recommended temperature
range <21°C. Extended exposure to temperatures above 45°C
could degrade battery performance and life.
Thermal Fuse
A Thermal Fuse is fitted in series with the charge/ discharge path
to protect the battery from over temperature. This device goes
open circuit if the cell case temperature reaches the fuse's
temperature rating of 93°C (+0°C, -5°C). The fuse is non-re-
settable rendering the battery pack non-functional.
Mechanical Specifications
Weight
Approximately 0.24 Kg.
Mating Connector
The recommended interconnection mating male connector is AMP
P/N 5787422-1 or 5787446-1.