Professional Documents
Culture Documents
Temperature Measurement Lab Report PDF
Temperature Measurement Lab Report PDF
Prepared by:
Raveendra Sampath Kumara Sarukkali Kankanamge
800226-A298
rsksk@kth.se
International College for Business and Technology
Sri Lanka
26-05-2011
1
Q 01) Finding the ambient temperature
Vreading = 0.7896 mV
Vreading = V1 – V2
Temperature at J2 = 0 °C
Thus, V2 = 0 mV
Hence, V1 = 0.7896mV
2
o
Voltage (mV) vs Temperature ( C)
0.795
0.79
0.785
y = 0.041x - 0.03
0.78
0.775
Voltage
0.77
0.765
0.76
0.755
0.75
0.745
18.8 19 19.2 19.4 19.6 19.8 20 20.2
Temperature
V2 = 0.7896 mV
Vreading = 2.568
Vreading = V1 – V2
2.568 = V1 – 0.7896
V1 = 3.358
Hence, temperature of J1 = 80 °C
3
Q 03) If the polarity of the thermocouple placed in ice water had been reversed,
what would be the
Voltage :
Temperature :
Voltage :
Temperature :
Q 04)
4
a) If the junction is at ambient air
Vreading = V1 – V2
Vreading =0
In this case junction J1 is in the ice and junction J2 is in the ambient condition, which is
19.99 °C.
Thus, V1 = 0 V2 = 0.7896
Vreading = V1 – V2
Vreading = 0 - 0.7896
Vreading = - 0.7896
5
c) If the junction is at hot water
In this case junction J1 is in hot water and it is 80 °C, and junction J2 is in ambient
condition and it is 19.99 °C.
Vreading = V1 – V2
Vreading = 2.5684
Q 05) Connect the single thermocouple with the extremely long junction to a channel of the
instrument, and configure the logger reading voltage on the display. By inserting the junction
only partly into the hot or cold water, or by heating the wire in between your fingers, try to
figure out which part of the junction has the temperature displayed by the instrument. Is it the
tip, the end towards the insulation or perhaps an average of the whole length?
It is whole length
Q 06) Insert the thermocouple with a short junction to the bottom of the thin copper tube
(which could be a thermometer well of a test rig). Insert the copper tube two, five, ten and 20
cm into the ice-water and read the temperature for each insertion length, waiting till the
temperature has stabilized completely before each reading. Redo the test using the brass tube.
Finally, redo the test with only the (coiled) thermocouple wire without a tube. Compare the
results. Which of the two tubes is the better thermometer well? Why? What insertion length
is necessary to make a correct reading in the three cases?
In this case junction J1 is in ice water (0 °C), and junction J2 is in ambient condition
(19.99 °C)
Thus, V1 = 0 V2 = 0.7896
Vreading = V1 – V2
Vreading = 0 – 0.7896
Vreading = – 0.7896
6
Thermocouple 2cm 5cm 10cm
Copper tube -0.7516mV -0.77022mV -0.78921mV
Brass tube -0.7586mV -0.78185mV -0.78921mV
Bare wire -0.78573mV -0.78921mV -0.78921mV
0.03
Error (mV)
0.02
0.01
0
0 0.5 1 1.5 2 2.5 3 3.5
-0.01
Inserted Length (cm)
Error decreases with the insertion length increases. Minimum error happens when bare
wire is used. But Brass is better for thermometer well.
7
Q 08) Connect a Pt100 sensor to the logger. Make a drawing below, and explain how the
measurement is done. Read the manual of the logger to see how Pt100 measurements are
conducted. Illustrate the use both 2 wire as well as 4 wire measurements. What is the
difference between these measurements?
The above figure illustrates two wires configuration. It is only used when high accuracy
is not required as the resistance of the connecting wires is always included with that of
the sensor leading to errors in the signal. The values of the lead resistance can only be
determined in a separate measurement without the resistance thermometer sensor and
therefore a continuous correction during the temperature measurement is not possible.
The four wire resistance thermometer configuration even further increases the accuracy
and reliability of the resistance being measured. In the diagram above a standard two
terminal RTD is used with another pair of wires to form an additional loop that cancels
out the lead resistance.
8
Q 09) Configure the logger to read the temperature of the Pt100 sensor while having the
sensor in the air. What is the temperature for the 2 wire setup?
PT100 has a resistance of 100 ohms at 0 °C and 138.4 ohms at 100 °C.
(138.4 − 100)
resistance = Temperature + 100
100
(138.4 − 100)
107.8 = Temperature + 100
100
Temperature = 20.31 °C
Q 10) Redo point 9 above but insert resistance of the each of the wire, simulating (very) long
connection cables. What is the new reading? Conclusion? What would be an alternative
connection which may be better?
(138.4 − 100)
resistance = Temperature + 100
100
9
Temperature = 25.52 °C
Conclusion: When long connection cables are used, two wires configuration should not
be used, since resistance of the cables cause an error to the final reading.
Alternative connection
Four wires configuration can be used, if long cables are used.
Q 11) Read the temperature of ice-water with the illustrated setup using thermocouple and a
Pt100-element.
What is reading of the thermocouple?
………………………mV
………………………°C
-What is reading of the Pt100-element?
………………………Ω
………………………°C
10
Thermocouple reading = Vreading
Vreading = V3 + V1 – V2
Vreading = -0.7896 mV
V3 =0
11
Calculated values of V2
o
Voltage (mV) vs Temperature ( C)
1.04
1.035
1.03
1.025
y = 0.041x - 0.033
1.02
1.015
Voltage
1.01
1.005
0.995
0.99
0.985
24.8 25 25.2 25.4 25.6 25.8 26 26.2
Temperature
V2 = 0.041×25.52 + 0.033
V2 = 1.07932
Vreading = V3 + V1 – V2
-0.7896 = 0 + V1 – 1.07932
V1 = 0.28972
12
Voltage (mV) vs Temperature (oC)
0.315
0.31
0.305
0.3
y = 0.039x
0.295
Voltage
0.29
0.285
0.28
0.275
0.27
6.8 7 7.2 7.4 7.6 7.8 8 8.2
Temperature
Voltage = 0.039×temperature
0.28972 = 0.039×temperature
13