Professional Documents
Culture Documents
Mills UPM 18010 Final
Mills UPM 18010 Final
A limitation of the above approach is that temperature No two Coriolis meters are identical and hence there are
and pressure variations are known to influence properties, small variations in natural frequency of the oscillating tubes.
other than fluid viscosity, that may also be critical to the This coupled with the wide range of process conditions any
overall measurement uncertainty. Meter 2s expected to work in means that potential errors
could be present. However, a simple zeroing process in
In Issue 9.2 of its Guidance Notes for Petroleum operation is claimed to eliminate these issues leaving the
Measurement under the Petroleum (Production) meter able to achieve its stated uncertainty in all operating
Regulations 2014, The UK Oil & Gas Authority (OGA) have conditions. Consider equation 1:
stipulated that temperature and pressure compensation
Copyright 2018, Letton Hall Group. This paper was developed for the UPM Forum, 21 – 22 February 2018, Houston, Texas, U.S.A., and is subject to correction by the author(s). The contents of the paper may not necessarily
reflect the views of the UPM Forum sponsors or administrator. Reproduction, distribution, or storage of any part of this paper for commercial purposes without the written consent of the Letton Hall Group is prohibited.
Non-commercial reproduction or distribution may be permitted, provided conspicuous acknowledgment of the UPM Forum and the author(s) is made. For more information, see www.upmforum.com.
Qm = FCF(Δtm +Δtlive_zero – Δtstored_zero) (1) pressure. To correct for pressure effects the user must input
the operating pressure into the flow computer or provide a
The mass flowrate, Qm, can be calculated using equation pressure measurement for an online correction.
1 where FCF is the flow calibration factor, Δt m is the measured
time difference caused by the mass flow of the fluid only, The performance of Coriolis flow meters is detailed in the ISO
Δtlive_zero is the measured time difference due to the live zero standard 10790 [2]. However, the performance of Coriolis
value and Δtstored_zero is the stored zero value from the meters meters in high viscosity fluids and the potential adverse effect
previous use. of flow profile / low Reynolds numbers are not addressed.
By zeroing a meter after installation, the stored zero value Joint Industry Project
then equals the live zero value therefore eliminating any zero In 2012, NEL proposed a Joint Industry Project (JIP)
effect from the meter. The mass flowrate can then be researching temperature, pressure and viscosity effects on
calculated using equation 2 and any measured time Coriolis flowmeters. The requirement for such a JIP was due
difference is only due to the mass flowrate only. to NEL’s vast knowledge and experiences with Coriolis
flowmeters from the main manufacturers.
Qm = FCF*Δtm (2)
Although it was previously thought that Coriolis meters
By zeroing a meter at process conditions, the user is were insensitive to temperature, pressure and viscosity, NEL
effectively calibrating out any effect of tube rigidity at those haVE a great deal of calibration data displaying otherwise.
process conditions. This means that any variations in meter Industry still appeared to be unaware of these significant
construction, thermal expansion or contraction of the meter effects.
body can be minimised.
As such, NEL formulated a large investigation
Typically, the zero value is given as a mass flowrate i.e. the programme as part of a JIP, to examine six Coriolis meters
mass flow that the meter would record at zero flow (two diameter sizes from the three main Coriolis
conditions. There is a limit to the value that would constitute manufacturers). The JIP included temperature calibrations
an acceptable zero. This differs by meter size and from 20 to 60 °C in oil, viscosity calibrations up to 300 cP in
manufacturer. oil and pressure tests in gas up to 60 bar.g.
Coriolis manufacturers have previously claimed to have One of the most important outcomes of the JIP, was that
negligible sensitivity to fluid viscosity. Some manufacturers there was a substantial requirement for calibrating Coriolis
now accept that Coriolis devices have a sensitivity to flow meters close to service conditions. Relying on the previous
profile / low Reynolds numbers with viscous fluids [1]. In highly methods of calibrating at ambient conditions in a laboratory
viscous fluids, it is possible to attain low Reynolds numbers and then deploying the Coriolis meters at elevated
with a moderate flow velocity relative to the fluid properties. conditions was deemed to be inappropriate for high
Thus the effects observed cannot solely be attributed to low accuracy, low uncertainty measurement. It was found that
fluid velocity. temperature, pressure and viscosity effects are significant
and can result in the meter deviating by far greater than the
In terms of pressure and temperature effects, Coriolis 0.25% fiscal measurement specification.
meters are not immune to physical changes due to variations
in operating conditions. It is known that the Young's modulus To meet this requirement, NEL proposed the design of a
of the flow tubes will alter with increasing / decreasing unique elevated pressure liquid flow facility in 2013. The
temperature and pressure. This change to the tube stiffness proposal received backing from Department for Business,
results in an increase / decrease in the ‘twisting’ or’ ‘phase Energy and Industrial Strategy (BEIS) and NEL’s parent
shift’ of the Coriolis device. company TUV SUD. The facility was designed, constructed,
commissioned and received accredidation within 18
Most Coriolis manufacturers have corrections months. The accrediting body is the United Kingdom
incorporated in the flow computer of the device for Accreditation Service (UKAS).
temperature and pressure variations. The robustness of these
corrections still requires further research and analysis. Elevate Pressure & Temperature Facility
The EPAT flow facility, located at NEL in East Kilbride
Furthermore, Coriolis meters have a resistance Scotland, consists of a high (6”) capacity and a low (3”)
thermometer (RTD) within the device that measures the capacity flow line. NEL can accommodate nominal pipe sizes
temperature of the flow tubes. There is no such sensor for from 0.5 to 10 inches and up to 10 m of horizontal straight
References 3.30
1.39
Compensation of Process Parameters on Coriolis Meters
1
with a View to Custody Transfer of Hydrocarbon 0 10 20 30 40 50 60 70 80 90 100 110
Products”. Paper 6.3, 9th South East Asia Hydrocarbon Fluid Temperature (°C)
Figures
1.25 1.25
1.00 1.00
0.75 0.75
0.50 0.50
% Err (Ref.Mass)
% Err (Ref.Mass)
0.25 0.25
0.00 0.00
-0.25 -0.25
-0.50 -0.50
-0.75 -0.75
-1.00 -1.00
-1.25 -1.25
-1.50 -1.50
0 10 20 30 40 50 60 70 80 90 100 10 100 1000 10000
Ref. Mass Flow [kg/s] Reynolds Number
600 cSt 1000 cSt 200 cSt 600 cSt 1000 cSt 1500 cSt
Figure 6 Coriolis 1 – Mass Error v Mass Flow wrt viscosity Figure 9 Coriolis 1 – Cor Mass Error v Re No wrt viscosity
1.00
1.50
0.75
1.25
0.50
1.00
0.75
0.25
% Err (Ref.Dens)
0.50
% Err (Ref.Mass)
0.00
0.25
0.00
-0.25
-0.25
-0.50
-0.50
-0.75
-0.75
-1.00
-1.00
-1.25 0 10 20 30 40 50 60 70 80 90 100
Ref. Mass Flow [kg/s]
-1.50
10 100 1000 10000
Reynolds Number
600 cSt (20degC) 1000 cSt (17degC)
1.50
0.25
1.25
0.00
1.00
0.75
% Err (Ref.Mass)
-0.25
0.50
% Err (Ref.Mass)
-0.50
0.25
0.00
-0.75
-0.25
-1.00
-0.50
-0.75
-1.25
-1.00
-1.50
-1.25
0 10 20 30 40 50 60 70 80
-1.50
Ref. Mass Flow, kg/s
0 10 20 30 40 50 60 70 80 90
Ref. Mass Flow, kg/s
Kerosene 10°C (3 cSt) Kerosene 20°C (2 cSt) Kerosene 30°C (2 cSt) Kerosene 40°C (2 cSt)
0.25
2.00
0.00
1.00
% Err (Ref.Mass)
-0.25
% Err (Ref.Mass)
-0.50 0.00
-0.75
-1.00
-1.00
-2.00
-1.25
-1.50 -3.00
0 10 20 30 40 50 60 70 80 0 10 20 30 40 50 60 70
Ref. Mass Flow, kg/s Ref. Temperature, °C
Primol 10°C (300 cSt) Primol 20°C (175 cSt) Primol 30°C (80 cSt) Primol 40°C (50 cSt) 20degC 1 40degC 1 60degC 1 20degC 2 40degC 2 60degC 2
Figure 12 Coriolis 2 – Mass Error v Re No wrt viscosity Figure 15 Coriolis 3 – Mass Error v Ref. Temp wrt Temp
1.00
0.50
0.75
0.25
0.50
0.25
0.00
0.00
% Err (Ref.Mass)
% Err (Ref.Mass)
-0.25
-0.25
-0.50
-0.50
-0.75
-0.75
-1.00
-1.25
-1.00
-1.50
-1.25
-1.75
-2.00
-1.50
0 50 100 150 200 250 300 350
100 1000 10000 100000 1000000
Ref. Mass Flow, T/hr
Reynolds Number
Kerosene 10°C (3 cSt) Kerosene 20°C (2 cSt) Kerosene 30°C (2 cSt) Kerosene 40°C (2 cSt) 2 bar.g 10 bar.g 20 bar.g 40 bar.g
Primol 10°C (300 cSt) Primol 20°C (175 cSt) Primol 30°C (80 cSt) Primol 40°C (50 cSt)
Figure 13 Coriolis 2 – Mass Error v Re No wrt viscosity Figure 16 Coriolis 4 – Mass Error v Mass Flow wrt Press
1.00
0.25
0.75
0.20
0.50
0.15
0.25
0.10 0.00
% Err (Ref.Mass)
% Err (Ref.Mass)
0.05 -0.25
-0.50
0.00
-0.75
-0.05
-1.00
-0.10
-1.25
-0.15 -1.50
-0.20 -1.75
-2.00
-0.25 0 5 10 15 20 25 30 35 40 45
0 5 10 15 20 25 30 35 40 45
Ref. Pressure, Bar.g
Ref. Mass Flow, kg/s
Figure 14 Coriolis 3 – Mass Error v Mass Flow wrt Temp Figure 17 Coriolis 4 – Mass Error v Pressure wrt Press
0.75
0.75
0.50
0.25 0.50
0.00
0.25
% Err (Ref.Dens)
% Err (Ref.Mass)
-0.25
-0.50 0.00
-0.75
-0.25
-1.00
-1.25 -0.50
-1.50
-0.75
-1.75
-2.00 -1.00
0 5 10 15 20 25 30 35 40 45 0 100 200 300 400 500 600 700 800
Ref. Pressure, Bar.g Ref. Mass Flow, kg/min
2 bar.g 10 bar.g 20 bar.g 40 bar.g NEL 4 bar.g NEL 10 bar.g NEL 40 bar.g NEL 60 bar.g
Figure 18 Coriolis 4 – Density Error v Pressure wrt Press Figure 21 Coriolis 5 – Mass Error v Mass Flow wrt Press
1.00
0.75
0.50
0.25
% Err (Ref.Mass)
0.00
-0.25
-0.50
-0.75
-1.00
0 100 200 300 400 500 600 700 800
Ref. Mass Flow, kg/min
0.00
-0.20
% Err (Ref.Mass)
-0.40
-0.60
-0.80
-1.00
0 10 20 30 40 50 60 70
Ref. Pressure, Bar.g