Pro-Test Series 2 Calibration
Pro-Test Series 2 Calibration
Pro-Test Series 2 Calibration
ISSUE 1
MAY 2005
-
2) Press and hold down the key down and turn the power switch ‘ON’ which is located
on the back of the display housing. Wait until the ‘TRANSDUCER SERIAL # ?????’ has
finished being displayed, then release the button and the following screen will be shown.
OK
+
UNITS
0.000 N.m
CALIBRATE
-
ZERO %
ZERO
Remove load and press the button.
4) Load the transducer to ‘FULL SCALE’ and allow to stabilise, then press the ‘CALIBRATE’
button. The reading will be updated to the Pro-Test series 2 full scale capacity.
5) Remove the load and take readings to the required standard or method.
7) Press ‘OK’ if no further adjustments are to be made. This will take you into the ‘TRACK’ mode
and calibration adjustment is now complete.
NOTE: - To exit without making any changes, press ‘OK’ when the calibrate screen is shown.