CARESTREAM DIRECTVIEW Software V5.7E User Release Notes
CARESTREAM DIRECTVIEW Software V5.7E User Release Notes
CARESTREAM DIRECTVIEW Software V5.7E User Release Notes
PN AE6111
2017-03-27
Rev. B
The information contained herein is based on the experience and knowledge relating to the subject
matter gained by Carestream Health, Inc. prior to publication.
No patent license is granted by this information.
Carestream Health, Inc. reserves the right to change this information without notice, and makes no
warranty, express or implied, with respect to this information. Carestream Health shall not be liable for
any loss or damage, including consequential or special damages, resulting from any use of this
information, even if loss or damage is caused by Carestream Health’s negligence or other fault.
Technique Settings......................................................................................... 4
DNS IP Address ............................................................................................. 5
Extended Range............................................................................................. 5
System Configuration ..................................................................................... 5
Simultaneous Cassette Scanning .................................................................. 5
IHE Dose Reporting ....................................................................................... 5
Successful Harvest Message ......................................................................... 6
Bone Suppression .......................................................................................... 6
Snapshot Log Files......................................................................................... 6
DR Total Quality Tool (TQT) .......................................................................... 6
Multi-Format Printing ...................................................................................... 7
Different Image Destinations .......................................................................... 7
ROP Link ........................................................................................................ 8
Backup/Restore Function does not Save Detector Calibration Files .............. 8
Image Acquisition Screen does not Transition to Next Page when Study
Contains Multiple Pages ................................................................................. 8
Technique Settings
V5.7 supports CR techniques. When techniques are restored from an earlier version of software and
the backup contains site-specific custom Views, there will be no CR techniques. When any tabletop
View is selected by the user, the software will check for a CR technique first for that View (even if a
DR receptor will be used).
To avoid pop-up error messages that will be displayed even if the View is not done with CR, the Key
Operator must provide techniques for these Views. This can be done manually, or the Key Operator
can use the Technique Scaling feature in the Technique Settings menu.
189062
The software will allow the Key Operator to scale the techniques for individual patient sizes.
Workaround: The Key Operator must be sure to select all the patient sizes they wish to scale. The Key
Operator should also be sure that there are acceptable techniques for all the Views that might be
used.
Extended Range
In V5.6, Extended Range was set up with an HTML script accessible by Service. It allowed a setting
change that could increase the perceived bit depth of images. This information is not included in the
V5.6 backup file. You must manually configure Extended Range in the Destination Configuration
menu after upgrading to V5.7. Once configured in V5.7, it will be included in the V5.7 backup file.
System Configuration
Changes made in to some menu items in Key Operator/ System Configuration will require the system to
reboot before they will be applied. The Key Operator can make all changes and reboot after all changes
are made. The following menu items are affected:
Display Configuration Layout
Configuration
Prior Images Configuration
Customizable Field Descriptions
Delivery Preferences
HIS/RIS Configuration
Bone Suppression
185084
The Bone Suppression companion image can only be generated successfully when the image is oriented
“head up”. If a View is configured to automatically generate a Bone Suppression companion image, be
sure the Auto Rotation option is enabled. If there is a need to re-do the Bone Suppression companion
image, use the following steps:
1. Rotate the Standard image to the correct orientation.
2. Delete the auto-created Bone Suppression Companion image.
3. Request a new Bone Suppression Companion image manually.
186745
The Bone Suppression View is rotated 90 degrees when displayed on the Image Acquisition screen.
Workaround: Select Back or Main Menu and reselect the same patient.
18877
When performing a snapshot of the logs the system may freeze. If this occurs, restart the system.
179079, 179004
Several DR TQT test results are invalid for the CARESTREAM DRX 2530C Detector.
Workaround: None at this time.
Enterprise Wireless: The Detector does not Show up in the Image Acquisition
Screen.
202272
You may get a popup when disabling the detector that states that the detector could not be configured.
Workaround: Reboot the Console and the detector.