Download as pdf or txt
Download as pdf or txt
You are on page 1of 8

Operating & Maintenance

Moisture Analyzer
AMETEK
3050-OLV
Specification
• SN : 8150303 (AT-27076 A)
8150304 (AT-27076 B)
• Manufacture/Type : AMETEK/3050-OLV
• Range : 0.1 – 2500 ppmV
• Accuracy : ± 10% from 1 – 2500 ppmV
• Power : 220 VAC
• Signal Output : 4-20mA, Dry Contact
• Connection/Protocol : RS-232, RS-485 Modbus RTU/ASCII
• Configuration : AMETEK 3050 Configurator Software
Analyzer Components
• Moisture sensor: Quartz Crystal Microbalance (QCM)
Quartz crystal that has been coated with proprietary hygroscopic coating. The
coating absorb moisture from a sample gas stream changing the mass of the
coating. Changes in mass are detected as changes in natural resonance frequency
of the oscillator
• Dryer: to dry reference gas to less than 0.025 ppmv
Replace periodically ~ 1 year

• Moisture generator: a device capable of generating a known precise


moisture level.
Typically 50 ppmV.
Used for analyzer’s proper operation verification & calibration
Operation
• Allowable inlet pressure: 1.3 – 3.3 Bar (20 – 50 psi) gauge
• Exhaust pressure: 0 – 1 Bar (0 – 15 psi) gauge
• Sample gas temperature: 0 – 100°C (32° – 212°F)
• Sample gas flow: approximately 150 sccm
• Ambient analyzer temperature limits: 5° – 50°C (41° – 122°F)
• Voltage/Power: 230 ±10% VAC, 50/60 Hz, 150 W max.

› Start Up
• Purged sample line with ultra high purity nitrogen, set pressure @20 psig
• Turn on power source
• Set sample pressure 25 psig (20 – 50 psig)
• Let analyzer warm up and stabilized (estimate time : 2 hours)
• Working sensor temperature approximately 60°F
Installation Scheme
Flow Diagram
Configurator Software
Use the General tab to view the
current configuration and define
parameters for the analyzer and for PC
communications
THANK YOU