Professional Documents
Culture Documents
PQ Upgrade PDF
PQ Upgrade PDF
ZAP-70 UPGRADE
T H I S D O C U M E N T I S T H E P R O P E R T Y O F P I C K E R I N T E R N A T I O N A L , INC., A N D I S I N T E N D E D F O R E X C L U S I V E U S E B Y
C U R R E N T PICKER PERSONNEL. COPYING, DISCLOSURE TO OTHERS, OR O T H E R USE IS PROHIBITED W I T H O U T T H E
E X P R E S S W R I T T E N A U T H O R I Z A T I O N O F T H E P I C K E R I N T E R N A T I O N A L LAW D E P A R T M E N T . R E P O R T VIOLATIONS
O F T H I S R E Q U I R E M E N T T O P I C K E R I N T E R N A T I O N A L , I N C . : 595 M I N E R R O A D ; H I G H L A N D HTS., O H I O 44143.
REV. В
(C) 1996
PICKER
MORE THAN IMAGES. INSIGHT
© 1995/1996
PICKER INTERNATIONAL
HIGHLAND HEIGHTS, OHIO
Unpublished work-all rights reserved under
the copyright laws of the United States
PICKER INTERNATIONAL
CT TECHNICAL PUBLICATIONS DEPARTMENT
595 Miner Road, Highland Heights, Ohio 44143
Phone: (216) 473-3000
DOCUMENT EFFECTIVITY
A 03/10/95 Released t o C C B
G P U Board c h a n g e s
X S C Board c h a n g e s
LCDI PAL Upgrade
Microcontroller
S/W Installation procedure
В 07/26/96 Released t o C C B
a
LIST OF EFFECTIVE PAGES
PAGE DATE
The information contained in this manual conforms with the configuration of the equipment as of the date of manufacture.
Revisions to the equipment subsequent to the date of manufacture will be addressed in service updates distributed to the
Picker Technical Service Organization.
Picker International assumes no responsibility for the use, nor for any consequences which may result from the use or
interpretation of the information contained herein subsequent to the date of manufacture by anyone other than Picker
employees or authorized representatives of Picker.
THE USER OF THIS MANUAL IS DIRECTED TO READ AND CAREFULLY REVIEW THE INSTRUCTIONS, WARNINGS AND
CAUTIONS CONTAINED HEREIN PRIOR TO BEGINNING INSTALLATION OR SERVICE ACTIVITIES. WHILE YOU MAY
HAVE PREVIOUSLY INSTALLED OR SERVICED EQUIPMENT SIMILAR TO THAT DESCRIBED IN THIS MANUAL,
CHANGES IN DESIGN, MANUFACTURE OR PROCEDURE MAY HAVE OCCURRED WHICH SIGNIFICANTLY AFFECT THE
PRESENT INSTALLATION OR SERVICE.
PICKER INTERNATIONAL
CT TECHNICAL PUBLICATIONS DEPARTMENT
595 Miner Road
Highland Heights, Ohio 44143
Phone: (216) 473-3000
Diagnostic Imaging Systems
RADIATION WARNING
X-RAY AND GAMMA-RAYS ARE DANGEROUS TO BOTH O P E R A T O R AND OTHERS IN THE VICINITY UNLESS
ESTABLISHED SAFE EXPOSURE PROCEDURES ARE STRICTLY OBSERVED.
THE USEFUL AND SCATTERED BEAMS CAN PRODUCE SERIOUS OR FATAL BODILY INJURIES TO ANY PERSONS
IN THE SURROUNDING AREA IF USED BY AN UNSKILLED OPERATOR. ADEQUATE PRECAUTIONS MUST A L W A Y S
BE T A K E N TO AVOID EXPOSURE TO THE USEFUL BEAM, AS W E L L AS TO LEAKAGE RADIATION F R O M WITHIN THE
S O U R C E HOUSING OR TO SCATTERED RADIATION RESULTING FROM THE PASSAGE OF RADIATION T H R O U G H
MATTER.
ALL OF THE MOVEABLE ASSEMBLIES A N D PARTS OF THIS EQUIPMENT SHOULD BE O P E R A T E D WITH C A R E AND
ROUTINELY INSPECTED IN A C C O R D A N C E WITH THE MANUFACTURER'S R E C O M M E N D A T I O N S CONTAINED IN THE
EQUIPMENT MANUALS.
ONLY PROPERLY TRAINED AND QUALIFIED PERSONNEL SHOULD BE PERMITTED A C C E S S TO ANY INTERNAL
PARTS. LIVE ELECTRICAL TERMINALS ARE DEADLY; BE SURE LINE DISCONNECTS ARE O P E N E D AND O T H E R
A P P R O P R I A T E PRECAUTIONS ARE T A K E N BEFORE OPENING ACCESS DOORS, REMOVING ENCLOSURE PANELS,
OR ATTACHING ACCESSORIES.
DO NOT UNDER ANY CIRCUMSTANCES, REMOVE THE FLEXIBLE HIGH TENSION CABLES F R O M THE X-RAY TUBE
HOUSING OR HIGH TENSION GENERATOR AND/OR THE A C C E S S COVERS FROM THE G E N E R A T O R UNTIL THE MAIN
AND AUXILIARY P O W E R SUPPLIES HAVE BEEN DISCONNECTED.
FAILURE TO COMPLY WITH THE A B O V E MAY RESULT IN SERIOUS OR FATAL BODILY INJURIES TO THE O P E R A T O R
OR T H O S E IN THE AREA.
ELECTRICAL - GROUNDING INSTRUCTIONS
d
PUBLICATION IMPROVEMENT RECOMMENDATION
PICKER
improving the publication. This sheet is provided for
that purpose. To file a Publication Improvement
R e c o m m e n d a t i o n , fill in the information requested
below, fold, tape a n d mail this prepaid business reply
form.
T55-
This Publication Improvement DRS Publication Date
R e c o m m e n d a t i o n pertains to
m a n u a l ...
T h e following information w a s
Installation^
discovered during:(check one)
Other(specify)Q
Recommendation :
(Be specific by referencing your remarks to page numbers, etc.)
Service Office
(City)(State)
Telephone.
(Area code) (Number)
е
i NO POSTAGE
I NECESSARY
IF MAILED
IN THE
UNITED STATES
P O S T A G E WILL BE PAID BY
PICKER INTERNATIONAL
CT TECHNICAL PUBLICATIONS DEPARTMENT
PICKER
C872:K28
TABLE OF CONTENTS
Introduction 1
Parts Required 2
V I M Board Installation 6
128 MB M e s s a g e Passing M e m o r y 7
06/11/96
C872:K28
Introduction
179740 ( P Q - C T , p r e - X A P upgrade, no Z A P to Z A P - 7 0 ) .
179740-A ( P Q - C T , p o s t - X A P upgrade, no Z A P to Z A P - 7 0 ) .
179740-B ( P Q - 2 0 0 0 , p r e - X A P upgrade, no Z A P to Z A P - 7 0 ) .
179740-C ( P Q - 2 0 0 0 , p o s t - X A P upgrade, n o Z A P t o Z A P - 7 0 ) .
179740-D (PQ-CT and PQ-2000, ZAP-16 to ZAP-70).
179740-E (PQS, no Z A P to ZAP-70).
179740-F (PQS, ZAP-16 to ZAP-70).
Prerequisites
Tools required
NOTE
06/11/96 Picker International, Highland Hts., O H , Confidential and Proprietary Information, Refer to Front Cover. 1
C872:K28
PQ EB board 1 1 - - - - -
P/N 174227
2.0 GB Hard
Drive 1 1 1 1 -
P/N 174583
1.0 GB Hard 1
Drive - - - - -
P/N 174641
XAP 4 1 4 - - 1 -
P/N 177390
VIM 256 1 1 1 1 1 - -
P/N 177875-B
128 MPM - - - - - 1 -
4.2 to 4.4 or
QPROII
Upgrade 1 1 1 1 1 1 1
Instruction
P/N T55E-1153
PQS 1.0 o r
4.3B1 t o 4.4 or
QPROII 1 1 1 1 1 1 1
Software
Upgrade
P/N T55E-1146
Comments: ZAP-70 must be installed on PQS as corequisite to QPROII upgrade (QPROII upgrade
kit variation 179760-B). All other ZAP-70 upgrades require QPROII as prerequisite.
2 Picker International, Highland Hts., OH, Confidential and Proprietary Information, Refer to Front Cover. 06/11 /96
C872:K28
ii. Boot s y s t e m :
Run s i t e _ c n f . Verify that the X A P a n d BP Expansion boards are
e n a b l e d . If not, enable t h e m a n d save the configuration.
Run a pilot plan (run plan in multimode).
iii. Install:
2 5 6 V I M Board ( P Q - 2 0 0 0 , PQ-CT) 6
S C S I HD A s s e m b l y - 2.05 GB ( P Q - 2 0 0 0 , P Q - C T ) 8
G a n t r y Blower Modification ( P Q - 2 0 0 0 , PQ-CT) 11
128 M P M ( P Q S only) 7
S C S I HD A s s e m b l y - 1.0 GB ( P Q S only) 8
2. F o r Z A P - 1 6 s y s t e m : ( P Q - 2 0 0 0 , PQ-CT)
i. Install:
2 5 6 V I M Board 6
3. F o r all s y s t e m s :
/dev/dsk60f
06/11/96 Picker International, Highland Hts., O H , Confidential and Proprietary Information, Refer to Front Cover. 3
C872:K28
CAUTION
5. If applicable, install the remaining X A P board in the slot w h e r e the airflow strip w a s
removed.
4 Picker International, Highland Hts., OH, Confidential and Proprietary Information, Refer to Front Cover. 06/11/96
C872:K28
CAUTION
2. R e m o v e the BP expansion board from its shipping box a n d install into the card cage
next to t h e existing BP Expansion board.
NOTE
06/11 /96 Picker International, Highland Hts., OH, Confidential and Proprietary Information, Refer to Front Cover. 5
C872:K28
VIM B O A R D INSTALLATION
Introduction
CAUTION
5 Picker International, Highland Hts., O H , Confidential and Proprietary Information, Refer to Front Cover. 06/11/96
C872:K28
128 M B M E S S A G E P A S S I N G M E M O R Y
Introduction
NOTE
CAUTION
4. Power up the s y s t e m . Run site_cnf. Verify that the M P M board is enabled; if not,
enable it a n d save the configuration.
06/11/96 Picker International, Highland Hts., O H , Confidential and Proprietary Information, Refer to Front Cover. 7
C872:K28
A 1.0 GB HP or Fujitsu hard drive (PQS) or 2.0 GB HP hard drive (PQ-2000, PQ-CT)
m a y be installed in the Operator's Console as a v o l u m e disk.
1.0GB, 2 . 0 G B H a r d D r i v e I n s t a l l a t i o n
3. Ensure that the brackets will fit easily into the console frame. T h e drive's rear
connector's should face t o w a r d the front of the console. S e e Figure 2. Do not
m o u n t the drive in the console at this time.
6. Insert the drive into the console above the card rack, aligning the bracket e n d s into
the slots t o w a r d the rear of the console. Install the front t w o retaining screws.
NOTE
8 Picker International, Highland Hts., O H , Confidential and Proprietary Information, Refer to Front Cover. 06/11/96
C872:K28
NOTE
JUMPER IN
STORAGE
POSITION-
SCSI ID TABLE
е
j PIN POSITION DRIVE
SCSI |D
j 10 11 12 CONFIGURATION
174636
SHORT = JUMPER IN ACTIVE POSITION
174696 OPEN = JUMPER IN STORAGE POSITION
F I G U R E 3 - B R A C K E T a n d J U M P E R C O N F I G U R A T I O N F O R 1 & 2 G B , 3.5" HP
06/11/96 Picker International, Highland Hts., OH, Confidential and Proprietary Information, Refer to Front Cover. 9
C872:K28
SCSI ID TABLE
Sf,S: in PIN POSITION DRIVE
CONFIGURATION
5-6 3-4 1-2
0 OPEN i OPEN OPEN IMAGE DRIVE #3
SEE NOTE
{ 4
5
! SHORT
SHORT;
OPEN
OPEN
OPEN
SHORT
SPIRAL DRIVE #1
IMAGE DRIVE #2
JUMPER DRIVE AT
THIS LOCATION
6 SHORT SHORT OPEN SYSTEM/IMAGE DRIVE #1
177923
F I G U R E 4 - B R A C K E T a n d J U M P E R C O N F I G U R A T I O N F O R 1 G B , 3.5" F U J I T S U
10 Picker International, Highland Hts., O H , Confidential and Proprietary Information, Refer to Front Cover. 06/11 /96
C872:K28
G A N T R Y B L O W E R M O D I F I C A T I O N - (For pre Z A P - 1 6 s y s t e m s )
Introduction
Procedure
2. Lift the gull wing at the rear of the gantry a n d r e m o v e the lower gantry cover.
3. Loosen the two captive screws and remove the gantry AC control cover.
4. C h a n g e blower operation to continuous high by removing the tow larger black blower
wires f r o m terminal #2 and connect t h e m to terminal #1 of the S S R 3 1 3 relay. T h i s
will leave the single, smaller black wire (card rack fans) alone on terminal # 2 . S e e
Figure 5.
FIGURE 5 - SSR313
11
C872:K28
End of Instructions
1 2 P i c k e r International, Highland Hts., O H , Confidential and Proprietary Information, Refer to Front Cover. 06/11/96
c OMPUTED lOMOGRAPHY T(
ZAP-100 UPGRADE
T H I S D O C U M E N T I S T H E P R O P E R T Y O F P I C K E R I N T E R N A T I O N A L , INC., A N D I S I N T E N D E D F O R E X C L U S I V E U S E B Y
C U R R E N T P I C K E R P E R S O N N E L . C O P Y I N G , D I S C L O S U R E T O O T H E R S , O R O T H E R USE I S P R O H I B I T E D W I T H O U T T H E
E X P R E S S W R I T T E N A U T H O R I Z A T I O N O F T H E P I C K E R I N T E R N A T I O N A L LAW D E P A R T M E N T . REPORT VIOLATIONS
O F T H I S R E Q U I R E M E N T T O P I C K E R I N T E R N A T I O N A L , INC.: 595 M I N E R R O A D ; H I G H L A N D HTS., O H I O 44143.
© 1996 REV. В
Ш PICKER
MORE THAN IMAGES. INSIGHT
1995/1996
PICKER INTERNATIONAL
HIGHLAND HEIGHTS, OHIO
Unpublished work-all rights reserved under
the copyright laws of the United States
PICKER INTERNATIONAL
CT TECHNICAL PUBLICATIONS DEPARTMENT
595 Miner Road, Highland Heights, Ohio 44143
Phone: (216) 473-3000
C872:K20
TABLE OF CONTENTS
Introduction 1
Parts Required 2
V I M Board Installation 5
V I M Daughterboard Installation 7
G a n t r y Blower Modification 12
07/12/96
DOCUMENT EFFECTIVITY
A 03/10/95 Released t o C C B
G P U Board c h a n g e s
X S C Board c h a n g e s
LCDI PAL Upgrade
S/W Installation procedure
Microcontroller
В 07/26/96 Released t o C C B
LIST OF EFFECTIVE PAGES
PAGE DATE
The information contained in this manual conforms with the configuration of the equipment as of the date of manufacture.
Revisions to the equipment subsequent to the date of manufacture will be addressed in service updates distributed to the
Picker Technical Service Organization.
Picker International assumes no responsibility for the use, nor for any consequences which may result from the use or
interpretation of the information contained herein subsequent to the date of manufacture by anyone other than Picker
employees or authorized representatives of Picker.
THE USER OF THIS MANUAL IS DIRECTED TO READ AND CAREFULLY REVIEW THE INSTRUCTIONS, WARNINGS AND
CAUTIONS CONTAINED HEREIN PRIOR TO BEGINNING INSTALLATION OR SERVICE ACTIVITIES. WHILE YOU MAY
HAVE PREVIOUSLY INSTALLED OR SERVICED EQUIPMENT SIMILAR TO THAT DESCRIBED IN THIS MANUAL,
CHANGES IN DESIGN, MANUFACTURE OR PROCEDURE MAY HAVE OCCURRED WHICH SIGNIFICANTLY AFFECT THE
PRESENT INSTALLATION OR SERVICE.
PICKER INTERNATIONAL
CT TECHNICAL PUBLICATIONS DEPARTMENT
595 Miner Road
Highland Heights, Ohio 44143
Phone: (216) 473-3000
c
Diagnostic Imaging Systems
RADIATION WARNING
X-RAY AND GAMMA-RAYS ARE DANGEROUS TO BOTH OPERATOR AND OTHERS IN THE VICINITY UNLESS
ESTABLISHED SAFE EXPOSURE PROCEDURES ARE STRICTLY OBSERVED.
THE USEFUL AND SCATTERED BEAMS CAN PRODUCE SERIOUS OR FATAL BODILY INJURIES TO A N Y PERSONS
IN THE SURROUNDING AREA IF USED BY AN UNSKILLED OPERATOR. ADEQUATE PRECAUTIONS MUST A L W A Y S
BE T A K E N TO AVOID EXPOSURE TO THE USEFUL BEAM, AS WELL AS TO LEAKAGE RADIATION F R O M WITHIN THE
S O U R C E HOUSING OR TO SCATTERED RADIATION RESULTING FROM THE PASSAGE OF RADIATION T H R O U G H
MATTER.
FAILURE TO OBSERVE THESE W A R N I N G S MAY CAUSE SERIOUS OR FATAL BODILY INJURIES TO THE O P E R A T O R
OR T H O S E IN THE AREA.
ALL OF THE MOVEABLE ASSEMBLIES AND PARTS OF THIS EQUIPMENT SHOULD BE OPERATED W I T H CARE AND
ROUTINELY INSPECTED IN A C C O R D A N C E WITH THE MANUFACTURER'S RECOMMENDATIONS CONTAINED IN THE
EQUIPMENT MANUALS.
O N L Y PROPERLY TRAINED AND QUALIFIED PERSONNEL SHOULD BE PERMITTED ACCESS TO ANY INTERNAL
PARTS. LIVE ELECTRICAL TERMINALS ARE DEADLY; BE SURE LINE DISCONNECTS ARE O P E N E D AND O T H E R
APPROPRIATE PRECAUTIONS ARE TAKEN BEFORE OPENING A C C E S S DOORS, REMOVING ENCLOSURE PANELS,
OR ATTACHING ACCESSORIES.
DO NOT UNDER ANY CIRCUMSTANCES, REMOVE THE FLEXIBLE HIGH TENSION CABLES FROM THE X-RAY TUBE
HOUSING OR HIGH TENSION G E N E R A T O R AND/OR THE ACCESS C O V E R S FROM THE G E N E R A T O R UNTIL THE MAIN
A N D AUXILIARY POWER SUPPLIES HAVE BEEN DISCONNECTED.
FAILURE TO COMPLY WITH THE A B O V E MAY RESULT IN SERIOUS OR FATAL BODILY INJURIES TO THE O P E R A T O R
OR T H O S E IN THE AREA.
ELECTRICAL - GROUNDING INSTRUCTIONS
EXCEPT FOR INSTALLATIONS REQUIRING CERTIFICATION BY THE MANUFACTURER PER FEDERAL STANDARDS,
SEE T H A T A RADIATION PROTECTION SURVEY IS MADE BY A QUALIFIED EXPERT IN A C C O R D A N C E WITH NCRP 102,
SECTION 7, AS REVISED OR REPLACED IN THE FUTURE. PERFORM A SURVEY AFTER EVERY CHANGE IN
EQUIPMENT, W O R K L O A D , OR OPERATING CONDITIONS WHICH MIGHT SIGNIFICANTLY INCREASE THE PROBABILITY
OF PERSONS RECEIVING MORE THAN THE MAXIMUM PERMISSIBLE DOSE EQUIVALENT.
d
PUBLICATION IMPROVEMENT RECOMMENDATION
• PICKER
improving the publication. This sheet is provided for
that purpose. To file a Publication Improvement
R e c o m m e n d a t i o n , fill in the information requested
below, fold, tape a n d mail this prepaid business reply
form.
T55-
This Publication Improvement DRS Publication Date
R e c o m m e n d a t i o n pertains to
manual ...
T h e following information w a s
Installation^
discovered during:(check one)
Other(specify)Q
Recommendation:
(Be specific by referencing your remarks to page n u m b e r s , etc.)
Service Office
(City)(State)
Telephone
(Area code) (Number)
е
NO POSTAGE
NECESSARY
IF MAILED
IN THE
UNITED STATES
PICKER INTERNATIONAL
CT TECHNICAL PUBLICATIONS DEPARTMENT
€> PICKER
C872:K20
Introduction
Prerequisites
Tools required
07/12/96 Picker International, Highland Hts., OH, Confidential and Proprietary Information, Refer to Front Cover. 1
C872:K20
Parts required
Microcontroller
P/N 177282 Ref. Ref. Ref. Ref. Ref.
XAP
P/N 177390 4 - - - -
1K+ Display*
P/N 177632 1 1 1 1 1
VIM 256
P/N 177875-B - - - - 1
VIM 384
P/N 177875-C 1 1 - 1 -
VIM Daughterboard
P/N 310315 - - 1 - -
Boot system:
R u n s i t e _ c n f . Verify that the X A P a n d B P Expansion boards a r e e n a b l e d .
If not, enable t h e m a n d save the configuration.
R u n a pilot plan (run plan in m u l t i m o d e ) .
Install:
VIM Board 4
S C S I HD A s s e m b l y - 2.05 GB 8
1 K + Display B o a r d 6
G a n t r y B l o w e r Modification 10
Picker International, Highland Hts., OH, Confidential and Proprietary Information, Refer to Front Cover. 07/12/96
C872:K20
1. Install:
V I M Board 4
S C S I HD A s s e m b l y - 2.05 GB 8
1K+ Display Board 6
1. Install:
V I M Daughterboard 4
1K+ Display Board 6
For PQ 2000S s y s t e m s :
1. Install:
V I M Board 4
1K+ Display Board 6
F o r all s y s t e m s :
/dev/dsk60f
Perform initialization of device? (yes/no/quit/all)
07/12/96 Picker International, Highland Hts., OH, Confidential and Proprietary Information, Refer to Front Cover. 3
C872-.K20
viii. Run the protocol editor from the ELTP and reset all
customized feet-first protocols.
4 Picker International, Highland Hts., OH, Confidential and Proprietary Information, Refer to Front Cover. 07/12/96
C872:K20
VIM B O A R D INSTALLATION
Introduction
NOTE
2. Z A P - 1 6 systems only:
CAUTION
A N E L E C T R O S T A T I C H A Z A R D EXISTS F O R D E V I C E S O N
THE CIRCUIT BOARDS. USE A GROUNDED STATIC
WRIST BAND TO PREVENT ELECTROSTATIC DAMAGE
OF PCB C O M P O N E N T S . FAILURE TO COMPLY MAY
RESULT IN DAMAGE TO THE COMPONENTS ON THE
CIRCUIT BOARDS.
3. Z A P - 1 6 systems only:
07/12/96 Picker International, Highland Hts., O H , Confidential and Proprietary Information, Refer to Front Cover. 5
C872:K20
6 Picker International, Highland Hts., OH, Confidential and Proprietary Information, Refer to Front Cover. 07/12/96
C872:K20
VIM D A U G H T E R B O A R D INSTALLATION
CAUTION
07/12/96 Picker International, Highland Hts., O H , Confidential and Proprietary Information, Refer to Front Cover. 7
C872.K20
Introduction
CAUTION
2. Locate the old display board a n d remove the cables from the board. R e m o v e the
board f r o m the console. R e m o v e both boards if R D S is installed.
3. Install the n e w 1 K + display board (both if RDS) into the slot w h e r e the old display
board w a s . Attach the cables.
4. Return the old display board to Picker International using the enclosed shipping
labels.
8 Picker International, Highland Hts., O H , Confidential and Proprietary Information, Refer to Front Cover. 07/1 2/96
C872:K20
CAUTION
2. R e m o v e the console front cover. R e m o v e one airflow strip to allow installation of the
X A P board(s).
3. R e m o v e one of the array processor boards from the console a n d replace it with a
new X A P board (P/N 177390).
5. If applicable, install the remaining X A P board in the slot w h e r e the airflow strip w a s
removed.
6. Package the AP boards in the X A P shipping boxes, attach the shipping label a n d
mail the p a c k a g e back to Picker.
07/12/96 Picker International, Highland Hts., OH, Confidential and Proprietary Information, Refer to Front Cover. 9
C872:K20
3. Ensure that the brackets will fit easily into the console f r a m e . T h e drive's rear
connector's should face toward the front of the console. S e e Figure 3. Do not
mount the drive in the console at this time.
6. Insert the drive into the console above the card rack, aligning the bracket e n d s into
the slots t o w a r d the rear of the console. Install the front two retaining screws.
10 Picker International, Highland Hts., O H , Confidential and Proprietary Information, Refer to Front Cover. 07/12/96
C872:K20
NOTE
SCSI ID TABLE
PIN P O S I T I O N DRIVE
S C S I ID
10 11 12 CONFIGURATION
SHORT: OPEN
OPEN
SHORT
SPIRAL DRIVE #1
IMAGE DRIVE #2
JUMPER DRIVE AT
THIS LOCATION
6 SHORT! SHORT OPEN SYSTEM/IMAGE DRIVE #1
174636
SHORT = JUMPER IN ACTIVE POSITION
174696 OPEN = JUMPER IN STORAGE POSITION
F I G U R E 3 - B R A C K E T a n d J U M P E R C O N F I G U R A T I O N F O R 2 G B , 3.5" HP
07/12/96 Picker International, Highland Hts., O H , Confidential and Proprietary Information, Refer to Front Cover. 11
C872:K20
Introduction
Procedure
2. Lift the gull wing at the rear of the gantry a n d remove the lower gantry cover.
FIGURE 4 - SSR313
End of Instructions
12 Picker International, Highland Hts., OH, Confidential and Proprietary Information, Refer to Front Cover. 07/12/96