Download as pdf or txt
Download as pdf or txt
You are on page 1of 44

c OMPUTED lOMOGRAPHY T(

ZAP-70 UPGRADE

DRS No. C872:K28


PICKER PART NO.T55E-1124

T H I S D O C U M E N T I S T H E P R O P E R T Y O F P I C K E R I N T E R N A T I O N A L , INC., A N D I S I N T E N D E D F O R E X C L U S I V E U S E B Y
C U R R E N T PICKER PERSONNEL. COPYING, DISCLOSURE TO OTHERS, OR O T H E R USE IS PROHIBITED W I T H O U T T H E
E X P R E S S W R I T T E N A U T H O R I Z A T I O N O F T H E P I C K E R I N T E R N A T I O N A L LAW D E P A R T M E N T . R E P O R T VIOLATIONS
O F T H I S R E Q U I R E M E N T T O P I C K E R I N T E R N A T I O N A L , I N C . : 595 M I N E R R O A D ; H I G H L A N D HTS., O H I O 44143.

REV. В
(C) 1996
PICKER
MORE THAN IMAGES. INSIGHT

© 1995/1996

PICKER INTERNATIONAL
HIGHLAND HEIGHTS, OHIO
Unpublished work-all rights reserved under
the copyright laws of the United States

This document prepared and distributed by


CT Technical Publications Department.
Address inquiries regarding this document to:

PICKER INTERNATIONAL
CT TECHNICAL PUBLICATIONS DEPARTMENT
595 Miner Road, Highland Heights, Ohio 44143
Phone: (216) 473-3000
DOCUMENT EFFECTIVITY

REVISION DATE COMMENTS

Prelim -1 02/13/95 ORIGINAL ISSUE

A 03/10/95 Released t o C C B

A-1 06/11/96 Modified to c o n f o r m to 4.4 Specifications.


7 upgrade paths in o n e d o c u m e n t , removed
sections contained in the QPROII upgrade
instruction:

G P U Board c h a n g e s
X S C Board c h a n g e s
LCDI PAL Upgrade
Microcontroller
S/W Installation procedure

В 07/26/96 Released t o C C B

a
LIST OF EFFECTIVE PAGES

PAGE DATE

ALL PAGES 02/13/95

ALL PAGES 03/10/95

ALL PAGES 06/11/96


NOTICE

The information contained in this manual conforms with the configuration of the equipment as of the date of manufacture.
Revisions to the equipment subsequent to the date of manufacture will be addressed in service updates distributed to the
Picker Technical Service Organization.

Picker International assumes no responsibility for the use, nor for any consequences which may result from the use or
interpretation of the information contained herein subsequent to the date of manufacture by anyone other than Picker
employees or authorized representatives of Picker.

TO THE USER OF THIS M A N U A L

THE USER OF THIS MANUAL IS DIRECTED TO READ AND CAREFULLY REVIEW THE INSTRUCTIONS, WARNINGS AND
CAUTIONS CONTAINED HEREIN PRIOR TO BEGINNING INSTALLATION OR SERVICE ACTIVITIES. WHILE YOU MAY
HAVE PREVIOUSLY INSTALLED OR SERVICED EQUIPMENT SIMILAR TO THAT DESCRIBED IN THIS MANUAL,
CHANGES IN DESIGN, MANUFACTURE OR PROCEDURE MAY HAVE OCCURRED WHICH SIGNIFICANTLY AFFECT THE
PRESENT INSTALLATION OR SERVICE.

THE INSTALLATION A N D SERVICE OF EQUIPMENT DESCRIBED HEREIN IS TO BE PREFORMED BY AUTHORIZED,


QUALIFIED PICKER INTERNATIONAL PERSONNEL. ASSEMBLERS AND OTHER PERSONNEL NOT EMPLOYED BY NOR
DIRECTLY AFFILIATED WITH PICKER INTERNATIONAL TECHNICAL SERVICES ARE DIRECTED TO CONTACT THE
LOCAL PICKER OFFICE BEFORE ATTEMPTING INSTALLATION OR SERVICE PROCEDURES.

This Document prepared and distributed by CT


Technical Publications Department. Address
inquiries regarding this document to:

PICKER INTERNATIONAL
CT TECHNICAL PUBLICATIONS DEPARTMENT
595 Miner Road
Highland Heights, Ohio 44143
Phone: (216) 473-3000
Diagnostic Imaging Systems

RADIATION WARNING

X-RAY AND GAMMA-RAYS ARE DANGEROUS TO BOTH O P E R A T O R AND OTHERS IN THE VICINITY UNLESS
ESTABLISHED SAFE EXPOSURE PROCEDURES ARE STRICTLY OBSERVED.

THE USEFUL AND SCATTERED BEAMS CAN PRODUCE SERIOUS OR FATAL BODILY INJURIES TO ANY PERSONS
IN THE SURROUNDING AREA IF USED BY AN UNSKILLED OPERATOR. ADEQUATE PRECAUTIONS MUST A L W A Y S
BE T A K E N TO AVOID EXPOSURE TO THE USEFUL BEAM, AS W E L L AS TO LEAKAGE RADIATION F R O M WITHIN THE
S O U R C E HOUSING OR TO SCATTERED RADIATION RESULTING FROM THE PASSAGE OF RADIATION T H R O U G H
MATTER.

T H O S E AUTHORIZED TO OPERATE, PARTICIPATE IN OR SUPERVISE THE OPERATION OF THE EQUIPMENT MUST


BE T H O R O U G H L Y FAMILIAR AND COMPLY COMPLETELY W I T H THE CURRENT ESTABLISHED SAFE EXPOSURE
F A C T O R S AND PROCEDURES DESCRIBED IN PUBLICATIONS, SUCH AS: SUBCHAPTER J OF TITLE 21 OF THE CODE
OF FEDERAL REGULATIONS, "DIAGNOSTIC X-RAY SYSTEMS A N D THEIR MAJOR COMPONENTS", A N D THE NATIONAL
COUNCIL ON RADIATION PROTECTION (NCRP) NO. 33, "MEDICAL X-RAY A N D G A M M A - R A Y PROTECTION FOR
ENERGIES UP TO 10 MEV-EQUIPMENT DESIGN AND USE", AS REVISED OR REPLACED IN THE FUTURE.

T H O S E RESPONSIBLE FOR PLANNING OF X-RAY AND GAMMA-RAY EQUIPMENT INSTALLATIONS M U S T BE


T H O R O U G H L Y FAMILIAR AND COMPLY COMPLETELY WITH N C R P NO. 49, "STRUCTURAL SHIELDING DESIGN AND
EVALUATION FOR MEDICAL OF X-RAYS AND GAMMA-RAYS OF ENERGIES UP TO 10 MEV", AS REVISED AND
REPLACED IN THE FUTURE.

FAILURE TO OBSERVE T H E S E W A R N I N G S MAY CAUSE SERIOUS OR FATAL BODILY INJURIES TO THE O P E R A T O R


OR T H O S E IN THE AREA.

Diagnostic Imaging Systems

MECHANICAL - ELECTRICAL WARNING

ALL OF THE MOVEABLE ASSEMBLIES A N D PARTS OF THIS EQUIPMENT SHOULD BE O P E R A T E D WITH C A R E AND
ROUTINELY INSPECTED IN A C C O R D A N C E WITH THE MANUFACTURER'S R E C O M M E N D A T I O N S CONTAINED IN THE
EQUIPMENT MANUALS.

ONLY PROPERLY TRAINED AND QUALIFIED PERSONNEL SHOULD BE PERMITTED A C C E S S TO ANY INTERNAL
PARTS. LIVE ELECTRICAL TERMINALS ARE DEADLY; BE SURE LINE DISCONNECTS ARE O P E N E D AND O T H E R
A P P R O P R I A T E PRECAUTIONS ARE T A K E N BEFORE OPENING ACCESS DOORS, REMOVING ENCLOSURE PANELS,
OR ATTACHING ACCESSORIES.

DO NOT UNDER ANY CIRCUMSTANCES, REMOVE THE FLEXIBLE HIGH TENSION CABLES F R O M THE X-RAY TUBE
HOUSING OR HIGH TENSION GENERATOR AND/OR THE A C C E S S COVERS FROM THE G E N E R A T O R UNTIL THE MAIN
AND AUXILIARY P O W E R SUPPLIES HAVE BEEN DISCONNECTED.

FAILURE TO COMPLY WITH THE A B O V E MAY RESULT IN SERIOUS OR FATAL BODILY INJURIES TO THE O P E R A T O R
OR T H O S E IN THE AREA.
ELECTRICAL - GROUNDING INSTRUCTIONS

THE EQUIPMENT MUST BE G R O U N D E D TO AN EARTH G R O U N D BY A SEPARATE C O N D U C T O R . THE NEUTRAL SIDE


OF THE LINE IS NOT TO BE CONSIDERED THE EARTH GROUND. ON EQUIPMENT PROVIDED W I T H A LINE CORD,
THE EQUIPMENT MUST BE CONNECTED TO PROPERLY G R O U N D E D , THREE-PIN RECEPTACLE. DO NOT USE A
THREE-TO-TWO PIN ADAPTER.
INSTALLATION AND ENVIRONMENT

EXCEPT FOR INSTALLATIONS REQUIRING CERTIFICATION BY THE MANUFACTURER PER F E D E R A L S T A N D A R D S ,


SEE THAT A RADIATION PROTECTION SURVEY IS MADE BY A QUALIFIED EXPERT IN A C C O R D A N C E WITH N C R P 102,
SECTION 7, AS REVISED OR REPLACED IN THE FUTURE. PERFORM A SURVEY AFTER EVERY C H A N G E IN
EQUIPMENT, W O R K L O A D , OR OPERATING CONDITIONS WHICH MIGHT SIGNIFICANTLY INCREASE THE PROBABILITY
OF P E R S O N S RECEIVING MORE THAN THE MAXIMUM PERMISSIBLE DOSE EQUIVALENT.

d
PUBLICATION IMPROVEMENT RECOMMENDATION

Users of this publication are e n c o u r a g e d to report


errors, omissions a n d their r e c o m m e n d a t i o n s for

PICKER
improving the publication. This sheet is provided for
that purpose. To file a Publication Improvement
R e c o m m e n d a t i o n , fill in the information requested
below, fold, tape a n d mail this prepaid business reply
form.

T55-
This Publication Improvement DRS Publication Date
R e c o m m e n d a t i o n pertains to
m a n u a l ...

T h e following information w a s
Installation^
discovered during:(check one)
Other(specify)Q

Recommendation :
(Be specific by referencing your remarks to page numbers, etc.)

(Continue other side)


Your N a m e
(Please print)

Service Office
(City)(State)

Telephone.
(Area code) (Number)
е
i NO POSTAGE
I NECESSARY
IF MAILED
IN THE
UNITED STATES

BUSINESS REPLY MAIL


FIRST CLASS PERMIT NO. 19663 CLEVELAND, OH

P O S T A G E WILL BE PAID BY

PICKER INTERNATIONAL
CT TECHNICAL PUBLICATIONS DEPARTMENT

595 Miner Road Highland Heights, Ohio 44143

PICKER
C872:K28

TABLE OF CONTENTS
Introduction 1

Parts Required 2

Steps to follow in this upgrade 3

X A P Board Installation (according to s/n) 4

Expansion Board Installation ( P Q - C T only) 5

V I M Board Installation 6

128 MB M e s s a g e Passing M e m o r y 7

Hard Disk Drive Installation 8

Gantry Blower Modification 11

06/11/96
C872:K28

ZAP-70 UPGRADE INSTRUCTION

Introduction

This instruction describes the steps necessary to convert a P Q - 2 0 0 0 or P Q - C T system


currently running 4.2 or higher series software to Z A P - 7 0 or a P Q S s y s t e m currently
running 1.0 series or higher software to Z A P - 7 0 . Kit variations c o v e r e d by this instruction
include:

179740 ( P Q - C T , p r e - X A P upgrade, no Z A P to Z A P - 7 0 ) .
179740-A ( P Q - C T , p o s t - X A P upgrade, no Z A P to Z A P - 7 0 ) .
179740-B ( P Q - 2 0 0 0 , p r e - X A P upgrade, no Z A P to Z A P - 7 0 ) .
179740-C ( P Q - 2 0 0 0 , p o s t - X A P upgrade, n o Z A P t o Z A P - 7 0 ) .
179740-D (PQ-CT and PQ-2000, ZAP-16 to ZAP-70).
179740-E (PQS, no Z A P to ZAP-70).
179740-F (PQS, ZAP-16 to ZAP-70).

Prerequisites

1. Q P R O I I has been previously installed or is installed at the time of this upgrade.


2. A D S a n d A R D A S upgrades must be installed ( P Q - 2 0 0 0 , P Q - C T ) .
3. All air flow sensors should be functional.
4. T h e G P U Board must be a 5 1 2 K B b o a r d (P/N 1 7 4 3 6 3 or P/N 174439). For system
S/N 4 1 2 4 and system S/N 4 1 2 2 and below a n e w board m a y be required if Z A P - 1 6
is not installed. If a 2 5 6 K B G P U (P/N 171796) is installed or supplied with this kit,
contact Service Engineering ( P Q - 2 0 0 0 , P Q - C T ) .
5. T h e updated P M D C (P/N 171943-A or P/N 177821) must have been installed.

Tools required

Standard Service Engineers T o o l Kit

NOTE

See page 2 for Kit variation part quantities a n d page 3 for


steps to follow b a s e d on s y s t e m configuration.

06/11/96 Picker International, Highland Hts., O H , Confidential and Proprietary Information, Refer to Front Cover. 1
C872:K28

P a r t s r e q u i r e d (parts included m a y vary a c c o r d i n g t o prior u p g r a d e s , s/n, etc)

Kit v a r i a t i o n - * PQ-CT PQ-CT PQ-2000 PQ-2000 PQ-2000 PQS PQS


Part & no. i NO Z A P NO ZAP NO Z A P NO ZAP PQCT NO ZAP ZAP-16
PRE-XAP POST-XAP PRE-XAP POST-XAP ZAP-16
179740 179740A 179740B 179740C 179740D 179740E 179740F

PQ EB board 1 1 - - - - -
P/N 174227

2.0 GB Hard
Drive 1 1 1 1 -
P/N 174583

1.0 GB Hard 1
Drive - - - - -
P/N 174641

XAP 4 1 4 - - 1 -
P/N 177390

VIM 256 1 1 1 1 1 - -
P/N 177875-B

Microcontroller Ref. Ref. Ref. Ref. Ref. Ref. Ref.


P/N 177282

128 MPM - - - - - 1 -

4.2 to 4.4 or
QPROII
Upgrade 1 1 1 1 1 1 1
Instruction
P/N T55E-1153

PQS 1.0 o r
4.3B1 t o 4.4 or
QPROII 1 1 1 1 1 1 1
Software
Upgrade
P/N T55E-1146

Comments: ZAP-70 must be installed on PQS as corequisite to QPROII upgrade (QPROII upgrade
kit variation 179760-B). All other ZAP-70 upgrades require QPROII as prerequisite.

2 Picker International, Highland Hts., OH, Confidential and Proprietary Information, Refer to Front Cover. 06/11 /96
C872:K28

Steps to Follow in this Upgrade:

1. For non-ZAP systems:


Page
i. Install:
XAPs 4
E B s ( P Q - C T only) 5

ii. Boot s y s t e m :
Run s i t e _ c n f . Verify that the X A P a n d BP Expansion boards are
e n a b l e d . If not, enable t h e m a n d save the configuration.
Run a pilot plan (run plan in multimode).

iii. Install:
2 5 6 V I M Board ( P Q - 2 0 0 0 , PQ-CT) 6
S C S I HD A s s e m b l y - 2.05 GB ( P Q - 2 0 0 0 , P Q - C T ) 8
G a n t r y Blower Modification ( P Q - 2 0 0 0 , PQ-CT) 11
128 M P M ( P Q S only) 7
S C S I HD A s s e m b l y - 1.0 GB ( P Q S only) 8

2. F o r Z A P - 1 6 s y s t e m : ( P Q - 2 0 0 0 , PQ-CT)

i. Install:
2 5 6 V I M Board 6

3. F o r all s y s t e m s :

i. L o a d 4.4 Hardware a n d Software as follows:

a. For s y s t e m s starting with 4.2 software, refer to the 4.2 to 4.4 or


Q P R O I I Upgrade Instruction (P/N T 5 5 E - 1 1 5 3 ) for additional
required hardware a n d the software installation procedure.

b. For systems starting with P Q S 1.0 or 4.3B1 software, refer to the


P Q S 1.0 or 4.3B1 to 4.4 or Q P R O I I Software U p g r a d e (P/N
T 5 5 E - 1 1 4 6 ) for additional required hardware a n d the software
installation procedure.

c. For s y s t e m s already upgraded to 4.4 software, install the Rev. J


Microcontroller as per the appropriate Upgrade Kit listed in a a n d
b a b o v e . Run s w c o n f i g a n d reboot the s y s t e m . A n s w e r " n " to
all t a p e loads (site, diags, c a m e r a ) . A n s w e r "y" to Initialize Image
Partitions. T h e following prompt will be displayed: Are y o u sure
y o u want to destroy all data on image partitions? A n s w e r "y"-
A n o t h e r prompt is displayed:

/dev/dsk60f

Perform initialization of device? (yes/no/quit/all)

A n s w e r "n" to this prompt a n d a n s w e r "n" to all but /dev/dsk40f.

06/11/96 Picker International, Highland Hts., O H , Confidential and Proprietary Information, Refer to Front Cover. 3
C872:K28

ХАР B O A R D INSTALLATION - (АР REPLACEMENT on NON-ZAP S y s t e m s )

For PQ-CT s y s t e m s prior to s/n 4 2 6 5 , 4 X A P s are included.


For PQ-CT s y s t e m s after s/n 4 2 6 5 , 1 X A P is included.
For PQ-2000 s y s t e m s prior to s/n 4 2 6 5 , 4 X A P s are included.
For PQ-2000 s y s t e m s after s/n 4 2 6 5 , 0 X A P s are included (skip this section).
For P Q S s y s t e m s without ZAP-16, 1 X A P is included.
For P Q S s y s t e m s with ZAP-16, 0 X A P s are included (skip this section).

1. Ensure power is removed f r o m the s y s t e m .

CAUTION

AN ELECTROSTATIC HAZARD EXISTS FOR DEVICES ON


THE CIRCUIT BOARDS. USE A G R O U N D E D STATIC
WRIST BAND TO PREVENT ELECTROSTATIC DAMAGE
O F P C B C O M P O N E N T S . F A I L U R E T O C O M P L Y MAY
RESULT IN DAMAGE TO THE C O M P O N E N T S ON THE
CIRCUIT BOARDS.

2. R e m o v e the console front cover. R e m o v e o n e airflow strip to allow installation of the


X A P board(s).

3. R e m o v e one of the array processor boards f r o m the c o n s o l e a n d replace it with a


new X A P board (P/N 177390). Slots d e s i g n a t e d for array processor boards are 04,
0 5 , 17 and 18.

4. Repeat this procedure for each of the remaining AP b o a r d s in the s y s t e m .

5. If applicable, install the remaining X A P board in the slot w h e r e the airflow strip w a s
removed.

6. P a c k a g e the AP boards in the X A P shipping boxes, attach the shipping label a n d


mail the p a c k a g e back to Picker.

4 Picker International, Highland Hts., OH, Confidential and Proprietary Information, Refer to Front Cover. 06/11/96
C872:K28

EXPANSION BOARD INSTALLATION ( P Q - C T S y s t e m s only)

For P Q - C T s c a n n e r s without Z A P - 1 6 , an additional BP expansion board (P/N 174227) is


required to a c c o m m o d a t e spiral scanning.

1. Ensure p o w e r is removed from the s y s t e m .

CAUTION

AN ELECTROSTATIC HAZARD EXISTS FOR DEVICES ON


T H E C I R C U I T B O A R D S . USE A G R O U N D E D S T A T I C
WRIST BAND TO PREVENT ELECTROSTATIC DAMAGE
OF PCB COMPONENTS. FAILURE TO COMPLY MAY
RESULT IN DAMAGE TO THE COMPONENTS ON THE
CIRCUIT BOARDS.

2. R e m o v e the BP expansion board from its shipping box a n d install into the card cage
next to t h e existing BP Expansion board.

NOTE

T h e BP Expansion boards must be located next to o n e


another in the card cage. Both BP Expansion boards must be
next to the BackProjector board a n d all three boards must be
in slots 1 0 , 1 1 a n d 12 or 1 1 , 12 a n d 13 on the left side of the
card cage. S o m e shifting of other boards may be necessary.

3. Power up the s y s t e m . Run site_cnf. Verify X A P and BP Expansion boards are


enabled, if not, enable t h e m a n d save the configuration.

06/11 /96 Picker International, Highland Hts., OH, Confidential and Proprietary Information, Refer to Front Cover. 5
C872:K28

VIM B O A R D INSTALLATION

Introduction

For P Q - 2 0 0 0 a n d PQ-CT, the new 2 5 6 M B V I M board must be installed.

1. Ensure p o w e r is removed from the s y s t e m .

2. R e m o v e the cover f r o m the console a n d locate the M P M in the L O W E R S L O T (slot


09).

CAUTION

AN ELECTROSTATIC HAZARD EXISTS FOR DEVICES ON


THE CIRCUIT BOARDS. USE A G R O U N D E D STATIC
WRIST BAND TO PREVENT ELECTROSTATIC DAMAGE
OF PCB COMPONENTS. FAILURE TO C O M P L Y MAY
RESULT IN DAMAGE TO THE COMPONENTS ON THE
CIRCUIT BOARDS.

3. If s y s t e m already has Z A P - 1 6 (2 M P M s . ) : R e m o v e the M P M that is located in the


L O W E R S L O T (slot 0 9 ) . If n o n - Z A P s y s t e m : Install V I M in slot 09 a n d the M P M in
slot 19.

4. R e m o v e the V I M board (P/N 177875-B) f r o m its p a c k a g e . Ensure that pins 1 a n d


2 of J 3 4 are j u m p e r e d . This enables the reset button on the b o a r d . S e e Figure 1.
Install the V I M w h e r e the M P M w a s r e m o v e d . S e n d the old M P M board back to
Picker in the V I M shipping p a c k a g e using the provided shipping label.

FIGURE 1 - 256MB VIM B O A R D JUMPER VERIFICATION

5 Picker International, Highland Hts., O H , Confidential and Proprietary Information, Refer to Front Cover. 06/11/96
C872:K28

128 M B M E S S A G E P A S S I N G M E M O R Y

Introduction

For P Q S systems with no Z A P , a 128 M P M (P/N 177997) is required to a c c o m o d a t e


spiral scanning.

1. Ensure power is removed from the s y s t e m .

2. R e m o v e the cover from the console a n d locate the L O W E R M P M S L O T (slot 09).

NOTE

Slot 09 starts at the right side a n d counts d o w n to slot 0 0 .


T h e next slot is 19 a n d c o u n t s d o w n to slot 10 at the left side.

CAUTION

AN ELECTROSTATIC HAZARD EXISTS FOR DEVICES ON


THE CIRCUIT BOARDS. USE A G R O U N D E D STATIC
WRIST BAND TO PREVENT ELECTROSTATIC DAMAGE
OF PCB COMPONENTS. FAILURE TO COMPLY MAY
RESULT IN DAMAGE TO THE COMPONENTS ON THE
CIRCUIT BOARDS.

3. R e m o v e the 128 MB M e s s a g e Passing M e m o r y (P/N 177997) from its p a c k a g e a n d


install it in slot 09, pressing firmly until it is completely seated. Tighten the t w o
captive mounting screws at the top a n d bottom of the P C B . S e e figure 2.

4. Power up the s y s t e m . Run site_cnf. Verify that the M P M board is enabled; if not,
enable it a n d save the configuration.

06/11/96 Picker International, Highland Hts., O H , Confidential and Proprietary Information, Refer to Front Cover. 7
C872:K28

HARD DISK DRIVE I N S T A L L A T I O N :

A 1.0 GB HP or Fujitsu hard drive (PQS) or 2.0 GB HP hard drive (PQ-2000, PQ-CT)
m a y be installed in the Operator's Console as a v o l u m e disk.

1.0GB, 2 . 0 G B H a r d D r i v e I n s t a l l a t i o n

1. Ensure p o w e r is r e m o v e d f r o m the s y s t e m . R e m o v e the front cover from the


Operator's Console. R e m o v e the drive from the shipping package.

2. Attach the brackets to the hard drive. S e e Figures 3 or 4.

3. Ensure that the brackets will fit easily into the console frame. T h e drive's rear
connector's should face t o w a r d the front of the console. S e e Figure 2. Do not
m o u n t the drive in the console at this time.

4. Ensure the S C S I T e r m i n a t o r Resistor (jumper for 1 GB and 2 GB HP) is r e m o v e d


f r o m the drive. S e e T a b l e 1 a n d Figure 3 for disk drive configuration.

5. Locate an o p e n (unused) p o w e r connector a n d associated ground connector from


the wiring harness. Cabling from the factory is well tied back, and removal of s o m e
tie w r a p s m a y be necessary. Plug the power connector into the back of the hard
drive a n d attach the g r o u n d cable.

6. Insert the drive into the console above the card rack, aligning the bracket e n d s into
the slots t o w a r d the rear of the console. Install the front t w o retaining screws.

NOTE

T h e s a m e j u m p e r configuration is used for both the 1 and 2 GB


H P S C S I Hard Drives.

T h e hard drive is to be used as a v o l u m e drive a n d therefore must


be installed at S C S I address 4.

8 Picker International, Highland Hts., O H , Confidential and Proprietary Information, Refer to Front Cover. 06/11/96
C872:K28

NOTE

REMOVE THE TERMINATOR JUMPERS.


IF JUMPERS ARE NOT REMOVED, THE
CPB SCSI FUSE WILL BLOW.

REFER TO TABLE FOR


SCSI ID CONFIGURATION
JUMPER IN
n
ACTIVE POSITION

JUMPER IN
STORAGE
POSITION-

SCSI ID TABLE

е
j PIN POSITION DRIVE
SCSI |D

j 10 11 12 CONFIGURATION

0 ! OPEN OPEN OPEN IMAGE DRIVE #3


SEE NOTE
4 SHORT! OPEN OPEN SPIRAL DRIVE #1 JUMPER DRIVE AT
5 SHORT OPEN SHORT IMAGE DRIVE #2 THIS LOCATION
SHORT SHORT OPEN SYSTEM/IMAGE DRIVE #1
CD

174636
SHORT = JUMPER IN ACTIVE POSITION
174696 OPEN = JUMPER IN STORAGE POSITION

NOTE: THE FOURTH DRIVE ON A Q-SERIES SYSTEM CAN BE


EITHER A SPIRAL OR IMAGE DRIVE

F I G U R E 3 - B R A C K E T a n d J U M P E R C O N F I G U R A T I O N F O R 1 & 2 G B , 3.5" HP

06/11/96 Picker International, Highland Hts., OH, Confidential and Proprietary Information, Refer to Front Cover. 9
C872:K28

SCSI ID TABLE
Sf,S: in PIN POSITION DRIVE
CONFIGURATION
5-6 3-4 1-2
0 OPEN i OPEN OPEN IMAGE DRIVE #3
SEE NOTE

{ 4

5
! SHORT

SHORT;
OPEN

OPEN
OPEN

SHORT
SPIRAL DRIVE #1

IMAGE DRIVE #2
JUMPER DRIVE AT
THIS LOCATION
6 SHORT SHORT OPEN SYSTEM/IMAGE DRIVE #1

SHORT = J U M P E R IN ACTIVE POSITION


OPEN = J U M P E R IN S T O R A G E POSITION

NOTE: THE FOURTH DRIVE ON A Q-SERIES SYSTEM C A N BE


EITHER A SPIRAL OR IMAGE DRIVE

177923

F I G U R E 4 - B R A C K E T a n d J U M P E R C O N F I G U R A T I O N F O R 1 G B , 3.5" F U J I T S U

10 Picker International, Highland Hts., O H , Confidential and Proprietary Information, Refer to Front Cover. 06/11 /96
C872:K28

G A N T R Y B L O W E R M O D I F I C A T I O N - (For pre Z A P - 1 6 s y s t e m s )

Introduction

T h e gantry blowers will be modified to operate continuously on P Q - 2 0 0 0 a n d P Q - C T .

Procedure

1. Ensure power is r e m o v e d from the system.

2. Lift the gull wing at the rear of the gantry a n d r e m o v e the lower gantry cover.

3. Loosen the two captive screws and remove the gantry AC control cover.
4. C h a n g e blower operation to continuous high by removing the tow larger black blower
wires f r o m terminal #2 and connect t h e m to terminal #1 of the S S R 3 1 3 relay. T h i s
will leave the single, smaller black wire (card rack fans) alone on terminal # 2 . S e e
Figure 5.

5. R e a s s e m b l e the system a n d verify that the blowers operate continuously at high


s p e e d , by performing a scan.

FIGURE 5 - SSR313

11
C872:K28

End of Instructions

1 2 P i c k e r International, Highland Hts., O H , Confidential and Proprietary Information, Refer to Front Cover. 06/11/96
c OMPUTED lOMOGRAPHY T(

ZAP-100 UPGRADE

DRS No. C872:K20 i l l PICKER PART No.T55E-1097

T H I S D O C U M E N T I S T H E P R O P E R T Y O F P I C K E R I N T E R N A T I O N A L , INC., A N D I S I N T E N D E D F O R E X C L U S I V E U S E B Y
C U R R E N T P I C K E R P E R S O N N E L . C O P Y I N G , D I S C L O S U R E T O O T H E R S , O R O T H E R USE I S P R O H I B I T E D W I T H O U T T H E
E X P R E S S W R I T T E N A U T H O R I Z A T I O N O F T H E P I C K E R I N T E R N A T I O N A L LAW D E P A R T M E N T . REPORT VIOLATIONS
O F T H I S R E Q U I R E M E N T T O P I C K E R I N T E R N A T I O N A L , INC.: 595 M I N E R R O A D ; H I G H L A N D HTS., O H I O 44143.

© 1996 REV. В
Ш PICKER
MORE THAN IMAGES. INSIGHT

1995/1996

PICKER INTERNATIONAL
HIGHLAND HEIGHTS, OHIO
Unpublished work-all rights reserved under
the copyright laws of the United States

This document prepared and distributed by


CT Technical Publications Department.
Address inquiries regarding this document to:

PICKER INTERNATIONAL
CT TECHNICAL PUBLICATIONS DEPARTMENT
595 Miner Road, Highland Heights, Ohio 44143
Phone: (216) 473-3000
C872:K20

TABLE OF CONTENTS
Introduction 1

Parts Required 2

S t e p s to follow in this upgrade 2

V I M Board Installation 5

V I M Daughterboard Installation 7

1K+ Display Board Upgrade 8

X A P Board Installation (according to s/n) 9

Hard Disk Drive Installation 10

G a n t r y Blower Modification 12

07/12/96
DOCUMENT EFFECTIVITY

REVISION DATE COMMENTS

Prelim -1 02/13/95 O R I G I N A L ISSUE

A 03/10/95 Released t o C C B

A-1 07/12/96 Modified to conform to 4.4 Specifications.


7 upgrade paths in o n e document, removed
sections contained in the QPROII upgrade
instruction:

G P U Board c h a n g e s
X S C Board c h a n g e s
LCDI PAL Upgrade
S/W Installation procedure
Microcontroller

В 07/26/96 Released t o C C B
LIST OF EFFECTIVE PAGES

PAGE DATE

ALL PAGES 02/13/95

ALL PAGES 03/10/95

ALL PAGES 07/12/96


NOTICE

The information contained in this manual conforms with the configuration of the equipment as of the date of manufacture.
Revisions to the equipment subsequent to the date of manufacture will be addressed in service updates distributed to the
Picker Technical Service Organization.

Picker International assumes no responsibility for the use, nor for any consequences which may result from the use or
interpretation of the information contained herein subsequent to the date of manufacture by anyone other than Picker
employees or authorized representatives of Picker.

TO THE USER OF THIS M A N U A L

THE USER OF THIS MANUAL IS DIRECTED TO READ AND CAREFULLY REVIEW THE INSTRUCTIONS, WARNINGS AND
CAUTIONS CONTAINED HEREIN PRIOR TO BEGINNING INSTALLATION OR SERVICE ACTIVITIES. WHILE YOU MAY
HAVE PREVIOUSLY INSTALLED OR SERVICED EQUIPMENT SIMILAR TO THAT DESCRIBED IN THIS MANUAL,
CHANGES IN DESIGN, MANUFACTURE OR PROCEDURE MAY HAVE OCCURRED WHICH SIGNIFICANTLY AFFECT THE
PRESENT INSTALLATION OR SERVICE.

THE INSTALLATION AND SERVICE OF EQUIPMENT DESCRIBED HEREIN IS TO BE PREFORMED BY AUTHORIZED,


QUALIFIED PICKER INTERNATIONAL PERSONNEL. ASSEMBLERS AND OTHER PERSONNEL NOT EMPLOYED BY NOR
DIRECTLY AFFILIATED WITH PICKER INTERNATIONAL TECHNICAL SERVICES ARE DIRECTED TO CONTACT THE
LOCAL PICKER OFFICE BEFORE ATTEMPTING INSTALLATION OR SERVICE PROCEDURES.

This Document prepared and distributed by CT


Technical Publications Department. Address
inquiries regarding this document to:

PICKER INTERNATIONAL
CT TECHNICAL PUBLICATIONS DEPARTMENT
595 Miner Road
Highland Heights, Ohio 44143
Phone: (216) 473-3000

c
Diagnostic Imaging Systems

RADIATION WARNING

X-RAY AND GAMMA-RAYS ARE DANGEROUS TO BOTH OPERATOR AND OTHERS IN THE VICINITY UNLESS
ESTABLISHED SAFE EXPOSURE PROCEDURES ARE STRICTLY OBSERVED.

THE USEFUL AND SCATTERED BEAMS CAN PRODUCE SERIOUS OR FATAL BODILY INJURIES TO A N Y PERSONS
IN THE SURROUNDING AREA IF USED BY AN UNSKILLED OPERATOR. ADEQUATE PRECAUTIONS MUST A L W A Y S
BE T A K E N TO AVOID EXPOSURE TO THE USEFUL BEAM, AS WELL AS TO LEAKAGE RADIATION F R O M WITHIN THE
S O U R C E HOUSING OR TO SCATTERED RADIATION RESULTING FROM THE PASSAGE OF RADIATION T H R O U G H
MATTER.

T H O S E AUTHORIZED TO OPERATE, PARTICIPATE IN OR SUPERVISE THE OPERATION OF THE EQUIPMENT MUST


BE T H O R O U G H L Y FAMILIAR AND COMPLY COMPLETELY WITH THE CURRENT ESTABLISHED SAFE EXPOSURE
F A C T O R S AND PROCEDURES DESCRIBED IN PUBLICATIONS, SUCH AS: SUBCHAPTER J OF TITLE 21 OF THE C O D E
OF FEDERAL REGULATIONS, "DIAGNOSTIC X-RAY SYSTEMS AND THEIR MAJOR COMPONENTS", AND THE NATIONAL
COUNCIL ON RADIATION PROTECTION (NCRP) NO. 33, "MEDICAL X-RAY AND GAMMA-RAY PROTECTION FOR
ENERGIES UP TO 10 MEV-EQUIPMENT DESIGN AND USE", AS REVISED OR REPLACED IN THE FUTURE.

T H O S E RESPONSIBLE FOR PLANNING OF X-RAY AND GAMMA-RAY EQUIPMENT INSTALLATIONS MUST BE


T H O R O U G H L Y FAMILIAR AND COMPLY COMPLETELY WITH NCRP NO. 49, "STRUCTURAL SHIELDING DESIGN AND
EVALUATION FOR MEDICAL OF X-RAYS AND GAMMA-RAYS OF ENERGIES UP TO 10 MEV", AS REVISED AND
REPLACED IN THE FUTURE.

FAILURE TO OBSERVE THESE W A R N I N G S MAY CAUSE SERIOUS OR FATAL BODILY INJURIES TO THE O P E R A T O R
OR T H O S E IN THE AREA.

Diagnostic Imaging Systems

MECHANICAL - ELECTRICAL WARNING

ALL OF THE MOVEABLE ASSEMBLIES AND PARTS OF THIS EQUIPMENT SHOULD BE OPERATED W I T H CARE AND
ROUTINELY INSPECTED IN A C C O R D A N C E WITH THE MANUFACTURER'S RECOMMENDATIONS CONTAINED IN THE
EQUIPMENT MANUALS.

O N L Y PROPERLY TRAINED AND QUALIFIED PERSONNEL SHOULD BE PERMITTED ACCESS TO ANY INTERNAL
PARTS. LIVE ELECTRICAL TERMINALS ARE DEADLY; BE SURE LINE DISCONNECTS ARE O P E N E D AND O T H E R
APPROPRIATE PRECAUTIONS ARE TAKEN BEFORE OPENING A C C E S S DOORS, REMOVING ENCLOSURE PANELS,
OR ATTACHING ACCESSORIES.

DO NOT UNDER ANY CIRCUMSTANCES, REMOVE THE FLEXIBLE HIGH TENSION CABLES FROM THE X-RAY TUBE
HOUSING OR HIGH TENSION G E N E R A T O R AND/OR THE ACCESS C O V E R S FROM THE G E N E R A T O R UNTIL THE MAIN
A N D AUXILIARY POWER SUPPLIES HAVE BEEN DISCONNECTED.

FAILURE TO COMPLY WITH THE A B O V E MAY RESULT IN SERIOUS OR FATAL BODILY INJURIES TO THE O P E R A T O R
OR T H O S E IN THE AREA.
ELECTRICAL - GROUNDING INSTRUCTIONS

THE EQUIPMENT MUST BE G R O U N D E D TO AN EARTH GROUND BY A SEPARATE C O N D U C T O R . THE NEUTRAL SIDE


OF THE LINE IS NOT TO BE CONSIDERED THE EARTH GROUND. ON EQUIPMENT PROVIDED WITH A LINE C O R D ,
THE EQUIPMENT MUST BE CONNECTED TO PROPERLY GROUNDED, THREE-PIN RECEPTACLE. DO NOT USE A
THREE-TO-TWO PIN ADAPTER.
INSTALLATION AND ENVIRONMENT

EXCEPT FOR INSTALLATIONS REQUIRING CERTIFICATION BY THE MANUFACTURER PER FEDERAL STANDARDS,
SEE T H A T A RADIATION PROTECTION SURVEY IS MADE BY A QUALIFIED EXPERT IN A C C O R D A N C E WITH NCRP 102,
SECTION 7, AS REVISED OR REPLACED IN THE FUTURE. PERFORM A SURVEY AFTER EVERY CHANGE IN
EQUIPMENT, W O R K L O A D , OR OPERATING CONDITIONS WHICH MIGHT SIGNIFICANTLY INCREASE THE PROBABILITY
OF PERSONS RECEIVING MORE THAN THE MAXIMUM PERMISSIBLE DOSE EQUIVALENT.

d
PUBLICATION IMPROVEMENT RECOMMENDATION

Users of this publication are e n c o u r a g e d to report


errors, omissions a n d their recommendations for

• PICKER
improving the publication. This sheet is provided for
that purpose. To file a Publication Improvement
R e c o m m e n d a t i o n , fill in the information requested
below, fold, tape a n d mail this prepaid business reply
form.

T55-
This Publication Improvement DRS Publication Date
R e c o m m e n d a t i o n pertains to
manual ...

T h e following information w a s
Installation^
discovered during:(check one)
Other(specify)Q

Recommendation:
(Be specific by referencing your remarks to page n u m b e r s , etc.)

(Continue other side)


Your N a m e
(Please print)

Service Office
(City)(State)

Telephone
(Area code) (Number)
е
NO POSTAGE
NECESSARY
IF MAILED
IN THE
UNITED STATES

BUSINESS REPLY MAIL


FIRST CLASS PERMIT NO. 19663 CLEVELAND, OH

POSTAGE WILL BE PAID BY

PICKER INTERNATIONAL
CT TECHNICAL PUBLICATIONS DEPARTMENT

595 Miner Road Highland Heights, Ohio 4 4 1 4 3

€> PICKER
C872:K20

ZAP-100 UPGRADE INSTRUCTIONS

Introduction

T h e s e instructions describe the steps n e c e s s a r y to upgrade a PQ-2000 or PQ 5 0 0 0


system to Z A P 100 functionality. T h e instructions cover non-ZAP PQ-2000 s y s t e m s (both
pre a n d post XAP), ZAP-16 a n d ZAP-70 PQ-2000 or P Q 5 0 0 0 systems, a n d ZAP-70
PQ 2000S systems. Five kit variations:

179685: PQ-2000, non-ZAP, pre X A P to ZAP-100.


179685-A: PQ-2000, non-ZAP, post X A P to ZAP-100.
179685-B: PQ-2000/5000, ZAP-70 to ZAP-100.
179685-C: PQ-2000, ZAP-16 to ZAP-100.
179685-D: PQ 2 0 0 0 S , ZAP-70 to ZAP-100.

Prerequisites

1. QPROII has been installed prior to installing this kit.


2. Rhino or Mark III Х-Ray tube is installed.
3. A D S and A R D A S u p g r a d e s must be installed (PQ-2000).
4. All air flow s e n s o r s should be functional.
5. The G P U Board must be a 5 1 2 K B b o a r d (P/N 1 7 4 3 6 3 or P/N 174439). For system
S/N 4 1 2 4 a n d s y s t e m S/N 4 1 2 2 a n d below a new board may be required if ZAP-16
is not installed. If a 2 5 6 K B G P U (P/N 171796) is installed or supplied with the kit,
contact Service Engineering (PQ-2000, PQ-CT).
6. The updated P M D C (P/N 171943-A or P/N 177821) must have been installed.

Tools required

Standard Service Engineer Tool Kit

07/12/96 Picker International, Highland Hts., OH, Confidential and Proprietary Information, Refer to Front Cover. 1
C872:K20

Parts required

Kit v a r i a t i o n s PQ-2000 PQ-2000 PQ-2000/5000 PQ-2000 PQ-2000S


Part & no. 4 NO ZAP NO ZAP ZAP-70 TO ZAP-16 TO ZAP-70 TO
PRE-XAP POST-XAP ZAP-100 ZAP-100 ZAP-100
179685 179685-A 179685-B 179685-С 179685-D

2.0 GB Hard Drive


P/N 174583 1 1 - - -

Microcontroller
P/N 177282 Ref. Ref. Ref. Ref. Ref.

XAP
P/N 177390 4 - - - -

1K+ Display*
P/N 177632 1 1 1 1 1

VIM 256
P/N 177875-B - - - - 1

VIM 384
P/N 177875-C 1 1 - 1 -

VIM Daughterboard
P/N 310315 - - 1 - -

4.2 to 4.4 or QPROII


Upgrade Instruction
P/N T55E-1153 1 1 1 1 1

PQS 1.0 or 4.3B1 t o


4.4 or QPROII
Software Upgrade 1 1 1 1 1
P/N T55E-1146

Comments: * Quantity 2 for RDS systems.

Steps to Follow in this Upgrade:

For non-ZAP systems:


Page
1. Install:
XAPs 7

Boot system:
R u n s i t e _ c n f . Verify that the X A P a n d B P Expansion boards a r e e n a b l e d .
If not, enable t h e m a n d save the configuration.
R u n a pilot plan (run plan in m u l t i m o d e ) .

Install:
VIM Board 4
S C S I HD A s s e m b l y - 2.05 GB 8
1 K + Display B o a r d 6
G a n t r y B l o w e r Modification 10

Picker International, Highland Hts., OH, Confidential and Proprietary Information, Refer to Front Cover. 07/12/96
C872:K20

For ZAP-16 systems:

1. Install:
V I M Board 4
S C S I HD A s s e m b l y - 2.05 GB 8
1K+ Display Board 6

For PQ2000/5000 ZAP-70 s y s t e m s :

1. Install:
V I M Daughterboard 4
1K+ Display Board 6

For PQ 2000S s y s t e m s :

1. Install:
V I M Board 4
1K+ Display Board 6

F o r all s y s t e m s :

1. Load 4.4 Hardware and Software as follows:

a. For s y s t e m s starting with 4.2 software, refer to the 4.2 to 4.4 or


Q P R O I I Upgrade Instruction (P/N T 5 5 E - 1 1 5 3 ) for additional
required hardware and the software installation procedure.

b. For s y s t e m s starting with 4.3B1 software, refer to the P Q S 1.0 or


4.3B1 to 4.4 or Q P R O I I Software U p g r a d e (P/N T 5 5 E - 1 1 4 6 ) for
additional required hardware a n d the software installation
procedure.

c. For s y s t e m s already upgraded to 4.4 software:

i. Install the Rev. J Microcontroller as per the appropriate


Upgrade Kit listed in a a n d b a b o v e .

ii. Run s w _ c o n f i g and reboot the s y s t e m . A n s w e r " n " to all


device loads (site, diags, c a m e r a ) .

iii. A n s w e r "y" to Initialize Image Partitions. T h e following p r o m p t


will be displayed: Are y o u sure y o u w a n t to destroy all d a t a on
image partitions? A n s w e r "y". A n o t h e r prompt is displayed:

/dev/dsk60f
Perform initialization of device? (yes/no/quit/all)

A n s w e r "n" to this prompt a n d a n s w e r "n" to all but


/dev/dsk40f.

07/12/96 Picker International, Highland Hts., OH, Confidential and Proprietary Information, Refer to Front Cover. 3
C872-.K20

iv. Boot the system, c h o o s e option 5 - Regulus User M o d e .

v. T y p e s h o w f e e t at the IQ> prompt. Feet-first protocols that


existed before loading software with the previous Z A P option
will be displayed. Note all c u s t o m i z e d feet-first protocols.

vi. T y p e i q s t a r t at the IQ> prompt.

vii. W h e n the E L T P body is displayed, type i n i _ p r o t at the IQ>


prompt. This is n e c e s s a r y to install Z A P - 1 0 0 protocols. A n y
existing customized protocols will be lost by taking this action.

viii. Run the protocol editor from the ELTP and reset all
customized feet-first protocols.

d. C h e c k system operation as per steps 3 3 , 34, 35 a n d 36 in the 4.2


to 4.4 or Q P R O I I Upgrade Instruction (P/N T 5 5 E - 1 1 5 3 ) .

4 Picker International, Highland Hts., OH, Confidential and Proprietary Information, Refer to Front Cover. 07/12/96
C872:K20

VIM B O A R D INSTALLATION

Introduction

In order for Z A P - 1 0 0 to function properly, the new 3 8 4 M B VIM board (PQ-2000) or a


2 5 6 M B V I M board (PQ 2 0 0 0 S ) must be installed.

NOTE

For systems upgrading from a 2 5 6 M B V I M (ZAP-70 for


example) see page 7 for the 1 2 8 M B daughterboard
installation to the existing V I M board.

1. Ensure power is removed from the s y s t e m .

2. Z A P - 1 6 systems only:

R e m o v e the cover f r o m the console a n d locate the M P M that is in the L O W E R S L O T


(slot 0 9 ) .

CAUTION

A N E L E C T R O S T A T I C H A Z A R D EXISTS F O R D E V I C E S O N
THE CIRCUIT BOARDS. USE A GROUNDED STATIC
WRIST BAND TO PREVENT ELECTROSTATIC DAMAGE
OF PCB C O M P O N E N T S . FAILURE TO COMPLY MAY
RESULT IN DAMAGE TO THE COMPONENTS ON THE
CIRCUIT BOARDS.

3. Z A P - 1 6 systems only:

R e m o v e the M P M that is located in the L O W E R S L O T (slot 0 9 ) .

4. R e m o v e the V I M board (P/N 177875-B for PQ 2 0 0 0 S and P/N 1 7 7 8 7 5 - C for P Q -


2000) f r o m its package. Ensure that pins 1 and 2 of J 3 4 are j u m p e r e d . T h i s enables
the reset button on the board. S e e Figure 1 on the next page. Install the V I M w h e r e
the M P M w a s r e m o v e d . S e n d the old M P M board back to Picker in the V I M shipping
p a c k a g e using the provided shipping label.

07/12/96 Picker International, Highland Hts., O H , Confidential and Proprietary Information, Refer to Front Cover. 5
C872:K20

FIGURE 1 - VIM B O A R D J U M P E R VERIFICATION

6 Picker International, Highland Hts., OH, Confidential and Proprietary Information, Refer to Front Cover. 07/12/96
C872:K20

VIM D A U G H T E R B O A R D INSTALLATION

To upgrade an existing 2 5 6 M B V I M board to 3 8 4 M B , the installation of a 1 2 8 M B


daughterboard is required.

1. Ensure power is removed f r o m the s y s t e m . R e m o v e the cover from the console a n d


locate the VIM (P/N 177875-B).

CAUTION

AN ELECTROSTATIC HAZARD EXISTS FOR DEVICES ON


THE CIRCUIT BOARDS. USE A G R O U N D E D STATIC
WRIST BAND TO PREVENT ELECTROSTATIC DAMAGE
OF PCB C O M P O N E N T S . FAILURE TO COMPLY MAY
RESULT IN DAMAGE TO THE COMPONENTS ON THE
CIRCUIT BOARDS.

2. R e m o v e the V I M board a n d place on an anti-static surface, face up. R e m o v e the


daughterboard (P/N 177876) f r o m the p a c k a g e . T h e daughterboard will be installed
in the middle left location. S e e Figure 2.

4. Align the 4 holes on the d a u g h t e r b o a r d with the 4 tabs on the V I M b o a r d . Push AT


E A C H T A B with moderate pressure until the d a u g h t e r b o a r d seats. DO N O T push
in the middle of the board. Press again a b o v e the connector.

FIGURE 2 - D A U G H T E R B O A R D I N S T A L L A T I O N (256MB TO 384MB U P G R A D E )

5. Attach the -C label over the -B of the V I M a n d re-install.

07/12/96 Picker International, Highland Hts., O H , Confidential and Proprietary Information, Refer to Front Cover. 7
C872.K20

1K+ DISPLAY BOARD UPGRADE

Introduction

T h e 1K+ Display Board is e q u i p p e d with additional m e m o r y to load more 3D and M P R


images.

1. Ensure all p o w e r is r e m o v e d f r o m the s y s t e m .

CAUTION

AN ELECTROSTATIC HAZARD EXISTS FOR DEVICES


U S E D O N C I R C U I T B O A R D S INSIDE T H E C O N S O L E . U S E
GROUNDED STATIC WRIST BAND TO PREVENT
ELECTROSTATIC DAMAGE OF PCB COMPONENTS.
FAILURE TO COMPLY MAY RESULT IN DAMAGE TO
ELECTRONIC COMPONENTS ON THE CIRCUIT BOARDS.

2. Locate the old display board a n d remove the cables from the board. R e m o v e the
board f r o m the console. R e m o v e both boards if R D S is installed.

3. Install the n e w 1 K + display board (both if RDS) into the slot w h e r e the old display
board w a s . Attach the cables.

4. Return the old display board to Picker International using the enclosed shipping
labels.

8 Picker International, Highland Hts., O H , Confidential and Proprietary Information, Refer to Front Cover. 07/1 2/96
C872:K20

XAP B O A R D INSTALLATION - (AP REPLACEMENT on NON-ZAP Systems)

Skip this section if system is currently at ZAP-16 or ZAP-70

For P Q - 2 0 0 0 s y s t e m s prior to s/n 4 2 6 5 , 4 X A P s are included.

1. Ensure p o w e r is r e m o v e d from the s y s t e m .

CAUTION

AN ELECTROSTATIC HAZARD EXISTS FOR DEVICES ON


T H E C I R C U I T B O A R D S . USE A G R O U N D E D S T A T I C
WRIST BAND TO PREVENT ELECTROSTATIC DAMAGE
OF PCB COMPONENTS. FAILURE TO COMPLY MAY
RESULT IN DAMAGE TO THE COMPONENTS ON THE
CIRCUIT BOARDS.

2. R e m o v e the console front cover. R e m o v e one airflow strip to allow installation of the
X A P board(s).

3. R e m o v e one of the array processor boards from the console a n d replace it with a
new X A P board (P/N 177390).

4. Repeat this procedure for each of the remaining AP boards in the s y s t e m .

5. If applicable, install the remaining X A P board in the slot w h e r e the airflow strip w a s
removed.

6. Package the AP boards in the X A P shipping boxes, attach the shipping label a n d
mail the p a c k a g e back to Picker.

07/12/96 Picker International, Highland Hts., OH, Confidential and Proprietary Information, Refer to Front Cover. 9
C872:K20

H A R D D I S K D R I V E I N S T A L L A T I O N - 2.05 G B (3.5" HP)

A 2.05 GB hard drive is included if the s y s t e m is not currently running Z A P - 1 6 or Z A P - 7 0 .


Do not m a k e this drive the s y s t e m drive. This drive is intended to be used as the V o l u m e
drive (Spiral Disk) and should be j u m p e r e d at SCSI ID 4.

Skip this section if the system is currently running ZAP-16 or ZAP-70.

2.0GB Hard Drive Installation

1. Ensure p o w e r is r e m o v e d f r o m the system. R e m o v e the front c o v e r from the


Operator's Console. R e m o v e the drive from the shipping package.

2. Attach the brackets to the hard drive. See Figure 3.

3. Ensure that the brackets will fit easily into the console f r a m e . T h e drive's rear
connector's should face toward the front of the console. S e e Figure 3. Do not
mount the drive in the console at this time.

4. Ensure the S C S I T e r m i n a t o r Resistor (jumper for 2 GB HP) is r e m o v e d from the


drive. See Table 1 a n d Figure 3 for disk drive configuration.

5. Locate an o p e n (unused) p o w e r connector a n d associated g r o u n d connector f r o m


the wiring harness. Cabling from the factory is well tied back, a n d removal of s o m e
tie w r a p s m a y be necessary. Plug the power connector into the back of the hard
drive and attach the ground cable.

6. Insert the drive into the console above the card rack, aligning the bracket e n d s into
the slots t o w a r d the rear of the console. Install the front two retaining screws.

10 Picker International, Highland Hts., O H , Confidential and Proprietary Information, Refer to Front Cover. 07/12/96
C872:K20

NOTE

REMOVE T H E TERMINATOR JUMPERS.


IF JUMPERS ARE NOT REMOVED, THE
CPB SCSI FUSE WILL BLOW.

SCSI ID TABLE
PIN P O S I T I O N DRIVE
S C S I ID
10 11 12 CONFIGURATION

0 OPEN OPEN OPEN IMAGE DRIVE #3


SEE NOTE
{ 4
5
SHORT OPEN

SHORT: OPEN
OPEN

SHORT
SPIRAL DRIVE #1

IMAGE DRIVE #2
JUMPER DRIVE AT
THIS LOCATION
6 SHORT! SHORT OPEN SYSTEM/IMAGE DRIVE #1
174636
SHORT = JUMPER IN ACTIVE POSITION
174696 OPEN = JUMPER IN STORAGE POSITION

NOTE: THE FOURTH DRIVE ON A Q-SERIES SYSTEM CAN BE


EITHER A SPIRAL OR IMAGE DRIVE

F I G U R E 3 - B R A C K E T a n d J U M P E R C O N F I G U R A T I O N F O R 2 G B , 3.5" HP

07/12/96 Picker International, Highland Hts., O H , Confidential and Proprietary Information, Refer to Front Cover. 11
C872:K20

G A N T R Y B L O W E R M O D I F I C A T I O N - (For pre Z A P - 1 6 systems)

Introduction

T h e gantry blowers will be modified to operate continuously ( P Q - 2 0 0 0 only).

Procedure

1. Ensure power is r e m o v e d from the s y s t e m .

2. Lift the gull wing at the rear of the gantry a n d remove the lower gantry cover.

3. Loosen the t w o captive s c r e w s a n d remove the gantry AC control cover.


4. C h a n g e blower operation to c o n t i n u o u s high by removing the t o w larger black blower
wires from terminal #2 a n d c o n n e c t t h e m to terminal #1 of the S S R 3 1 3 relay. This
will leave the single, smaller black wire (card rack fans) alone on terminal # 2 . See
Figure 4.

5. R e a s s e m b l e the s y s t e m a n d verify that the blowers operate continuously at high


s p e e d , by performing a s c a n .

FIGURE 4 - SSR313

End of Instructions

12 Picker International, Highland Hts., OH, Confidential and Proprietary Information, Refer to Front Cover. 07/12/96