Download as pdf or txt
Download as pdf or txt
You are on page 1of 1

2.5.26. Nitrogen monoxide and nitrogen dioxide in gases EUROPEAN PHARMACOPOEIA 10.

01/2008:20526 prior to use. As the paramagnetic effect of oxygen is linear,


the instrument must have a suitable range with a readability of
0.1 per cent or better.
Calibration of the instrument. Make the setting in the
following manner :
2.5.26. NITROGEN MONOXIDE AND – set the zero by passing nitrogen R1 through the instrument
NITROGEN DIOXIDE IN GASES until a constant reading is obtained ;
– set the scale to 100 per cent by passing oxygen R through
Nitrogen monoxide and nitrogen dioxide in gases the instrument at the same flow rate as for nitrogen R1 until
are determined using a chemiluminescence analyser a constant reading is obtained.
(Figure 2.5.26.-1). Assay. Pass the gas to be examined through the instrument
The apparatus consists of the following : at a constant flow rate until a constant reading is obtained.
– a device for filtering, checking and controlling the flow of Record the concentration of oxygen in the gas to be examined.
the gas to be examined,
– a converter that reduces nitrogen dioxide to nitrogen 01/2017:20528
monoxide, to determine the combined content of nitrogen
monoxide and nitrogen dioxide. The efficiency of the
converter has to be verified prior to use,
– a controlled-flow-rate ozone generator ; the ozone is
produced by high-voltage electric discharges across 2.5.28. WATER IN GASES
two electrodes ; the ozone generator is supplied with
pure oxygen or with dehydrated ambient air and the Water in gases is determined using an electrolytic hygrometer,
concentration of ozone obtained must greatly exceed the described below.
maximum content of any detectable nitrogen oxides, The measuring cell consists of a thin film of diphosphorus
– a chamber in which nitrogen monoxide and ozone can pentoxide, between 2 coiled platinum wires which act as
react, electrodes. The water vapour in the gas to be examined is
absorbed by the diphosphorus pentoxide, which is transformed
– a system for detecting light radiation emitted at a
to phosphoric acid, an electrical conductor. A continuous
wavelength of 1.2 μm, consisting of a selective optical filter
voltage applied across the electrodes produces electrolysis of
and a photomultiplier tube.
the water and the regeneration of the diphosphorus pentoxide.
The resulting electric current, which is proportional to the
01/2009:20527 water content in the gas to be examined, is measured. This
system is self-calibrating since it obeys Faraday’s law.
Take a sample of the gas to be examined. Allow the gas to
stabilise at room temperature. Purge the cell continuously
until a stable reading is obtained. Measure the water content
2.5.27. OXYGEN IN GASES in the gas to be examined, making sure that the temperature
is constant throughout the device used to introduce the gas
Oxygen in gases is determined using a paramagnetic analyser. into the apparatus.
The principle of the method is based on the high paramagnetic The electrolytic hygrometer achieves accurate sample flows by
sensitivity of the oxygen molecule. Oxygen exerts a using a mass flow controller to deliver a constant volumetric
strong interaction on magnetic fields, which is measured flow rate to ensure that the water content is determined
electronically, amplified and converted to a reading of oxygen accurately. The calibration of the mass flow controller is
concentration. The measurement of oxygen concentration normally performed using nitrogen. When using gases other
is dependent upon the pressure and temperature and, if the than nitrogen for calibration, consult the manufacturer’s
analyser is not automatically compensated for variations in instructions for the appropriate conversion factors and ensure
temperature and pressure, it must be calibrated immediately that the correct cell is used for the type of gas to be examined.

Figure 2.5.26.-1. – Chemiluminescence analyser

176 See the information section on general monographs (cover pages)

You might also like