Professional Documents
Culture Documents
Flow Meter Lab Report
Flow Meter Lab Report
Checked by :
---------------------------
Date :
CONTENT
2.0 Abstract
2.1 Introduction
2.2 Objective
2.3 Theory
2.4 Experimental Procedure
2.5 Results and calculation
2.6 Discussion
2.7 Conclusion
2.8 Recommendations
2.9 References
2.10 Appendices
2.0 ABSTRACT
Flow meters are used in the industry to measure the volumetric flow rate of fluids.
Differential pressure type flow meters measure flow rate by introducing a constriction in
the flow. The pressure difference caused by the constriction is correlated to the flow rate
by using Bernoulli’s equation. If a constriction is placed in a pipe carrying a stream of
fluid, there will be an increase in velocity, and hence an increase in kinetic energy, at the
point of constriction. From energy balance as given by Bernoulli’s theorem, there must
be a corresponding reduction in pressure. Rate of discharge from the constriction can be
calculated by knowing this pressure reduction, the area available for flow at the
constriction, the density of the fluid and the coefficient of discharge C d. Coefficient of
discharge is the ratio of actual flow to the theoretical flow and makes allowances for
stream contraction and frictional effects. For smaller pipes orifice meter is a suitable
choice. In order to use any of these devices for measurement it is necessary to empirically
calibrate them. That is, pass a known volume through the meter and note the reading in
order to provide a standard for measuring other quantities. Controlling the flow in piping
systems is a significant issue in the chemical process industries. Obviously, in order to
control the flow in a pipe, the flow must be measured. This experiment will introduce
you to three devices that are used to measure flow. One, the rotmeter, is a simple
mechanical device that is designed to be read by an operator. It is rugged, relatively
inexpensive, and easily installed. The second, the orifice plate, can be set up to be read
locally or remotely using pressure transducers. Both are designed for flows that do not
contain significant amounts of solid material. The third, the magnetic flow meter, is a
more sophisticated device than either the rotmeter or the orifice plate. It requires that the
flowing material be electrically conductive, but can measure flows with suspended
material.
2.2 THEORY
Assuming a horizontal flow (neglecting minor elevation differences between measuring points)
the Bernoulli Equation can be modified to:
where
p = pressure
ρ = density
v = flow velocity
The equation can be adapted to vertical flow by adding elevation heights h₁ and h₂. Assuming
uniform velocity profiles in the upstream and downstream flow - the Continuity Equation can be
expressed as
q = v₁ A₁ = v₂ A₂ (2)
where
q = flow rate
A = flow area
Combining (1) and (2), assuming A₂ < A₁, gives the "ideal" equation:
For a given geometry (A), the flow rate can be determined by measuring the pressure difference
p₁ - p₂. The theoretical flow rate q will in practice be smaller (2 - 40%) due to geometrical
conditions. The ideal equation (3) can be modified with a discharge coefficient:
1 /2
q = c d A ₂[2( p ₁− p ₂)/ ρ(1−( A ₂/ A ₁)²)] (3b)
where
c d = discharge coefficient
The discharge coefficient cd is a function of the jet size - or orifice opening - the
area ratio = A vc / A₂
where
"Vena Contracta" is the minimum jet area that appears just downstream of the restriction. The
viscous effect is usually expressed in terms of the non-dimensional parameter Reynolds Number
- Re. Due to the Bernoulli and Continuity Equation the velocity of the fluid will be at its highest
and the pressure at the lowest in "Vena Contracta". After the metering device the velocity will
decrease to the same level as before the obstruction. The pressure recover to a pressure level
lower than the pressure before the obstruction and adds a head loss to the flow.
where
D₂ = orifice, venturi or nozzle inside diameter
d = D₂ / D₁ diameter ratio
π = 3.14
Equation (4) can be modified to mass flow for fluids by simply multiplying with the density:
When measuring the mass flow in gases, its necessary to considerate the pressure reduction and
change in density of the fluid. The formula above can be used with limitations for applications
with relatively small changes in pressure and density.
The orifice meter consists of a flat orifice plate with a circular hole drilled in it. There is a
pressure tap upstream from the orifice plate and another just downstream. There are in general
three methods of placing the taps. The coefficient of the meter depends upon the position of taps.
Flange location - Tap location 1 inch upstream and 1 inch downstream from face of
orifice
"Vena Contracta" location - Tap location 1 pipe diameter (actual inside) upstream and 0.3
to 0.8 pipe diameter downstream from face of orifice
Pipe location - Tap location 2.5 times nominal pipe diameter upstream and 8 times
nominal pipe diameter downstream from face of orifice
The discharge coefficient - cd - varies considerably with changes in area ratio and the Reynolds
number. A discharge coefficient cd = 0.60 may be taken as standard, but the value varies
noticeably at low values of the Reynolds number.
The pressure recovery is limited for an orifice plate and the permanent pressure loss depends
primarily on the area ratio. For an area ratio of 0.5, the head loss is about 70 - 75% of the orifice
differential.
The orifice meter is recommended for clean and dirty liquids and some slurry services.
The rangeability is 4 to 1
The pressure loss is medium
Typical accuracy is 2 to 4% of full scale
The required upstream diameter is 10 to 30
The viscosity effect is high
The relative cost is low
The fluid slows down in a cone with smaller angle (5 - 70 ) where most of the kinetic energy is
converted back to pressure energy. Because of the cone and the gradual reduction in the area
there is no "Vena Contracta". The flow area is at a minimum at the throat. High pressure and
energy recovery makes the venturi meter suitable where only small pressure heads are available.
A discharge coefficient c d = 0.975 can be indicated as standard, but the value varies noticeably at
low values of the Reynolds number.
The pressure recovery is much better for the venturi meter than for the orifice plate.
The venturi tube is suitable for clean, dirty and viscous liquid and some slurry services.
The rangeability is 4 to 1
Pressure loss is low
Typical accuracy is 1% of full range
Required upstream pipe length 5 to 20 diameters
Viscosity effect is high
Relative cost is medium
From the continuity and Bernoulli equations, show that the ideal (inviscid, incompressible) flow
rate through a Venturi tube or an orifice is:
d2 2 P
Qideal
4 d 4
1
D
where D and d are the diameters at the inlet and the throat, respectively, is the fluid density,
and P is the pressure drop from inlet to throat. The discharge coefficient is a correction factor
between the
Qactual
Cd
actual flow rate and the ideal flow rate, and is simply defined as: Qideal
Hence if Cd of a Venturi tube or an orifice is determined, the actual flow rate can
be predicted by measuring the pressure drop across the device.
2.3 OBJECTIVE
To verify the accuracy of a commercial flow meter for measuring flow rates, and to calibrate a
Venturi tube and an orifice as flow meters by determining their discharge coefficients
2.4 PROCEDURE
1. Turn on the pump and adjust the control valve on the hydraulic bench, the exit valve of the
Venturi-orifice assembly unit and the bleed valve on the manifold to bleed off all air bubbles.
2. Adjust the valves to obtain a flow rate of 22 liters/min as indicated by the commercial flow
meter.
3. Read the manometers that indicate the static pressure change across the Venturi and the
orifice.
4. Close the drain valve at the bottom of the tank of the hydraulic bench. Collect a known
volume of water in the tank and record the time required to do so.
5. Decrease the flow rate by 1 liter/min, and repeat steps 3 and 4, until zero flow rate.
APPARATUS
The Venturi Meter - in the venturi meter the fluid is accelerated through a converging cone of
angle 15-200 and the pressure difference between the upstream side of the cone and the throat is
measured and provides a signal for the rate of flow.
2.5 RESULTS AND CALCULATIONS
Flow Rate
Flow
Rotmeter Volume Time Calculate using
Manometer Reading (mm) Rate, Q
(L/min) (L) (s) Bernoulli's
(L/min)
Equation (L/min)
A B C D E F G H I J Venturi Orifice
287 280 276 280 283 284 284 264 273 272 5 3 35.00 5.146 3.26 8.41
313 307 272 293 299 304 303 219 253 250 10 3 15.96 11.278 6.17 18.86
351 337 267 312 324 334 330 155 224 218 15 3 9.42 19.108 9.59 30.76
401 379 259 336 356 373 370 66 182 174 20 3 8.40 21.429 10.35 34.09
B) Determination of The Loss Coefficient When Fluid Flows Through a 90 degree
Elbow
Differential Piezometer
Volume Time Flow Rate, Q V
Head ∆h (mm) Loss Coefficient
(L) (s) (L/min) (m/s)
Elbow (hi-hj)
3 35.00 5.146 1 0.1615 0.75
3 15.96 11.278 3 0.3540 0.47
3 9.42 19.108 6 0.5997 0.33
3 8.40 21.429 8 0.6726 0.35
Tapping Venturi Meter Diameter Orifice Meter
A = 26 mm G = 26 mm
B = 21.6 mm H = 16 mm
C = 16 mm
D = 20 mm
E = 22mm
F = 26 mm
Cd A2
Qv = x√2 g ∆ h
A 2
√ 1−( 2 )
A1
π D 2 π (0.026)
2
A1= = =5.31 x 10−4 m 2
4 4
2
π D2 π ( 0.016 )
A2= = =2.01 x 10−4 m2
4 4
Q actual 5.146 L/min
C d= = =1.03
Q ideal 5 L/min
For orifice:
√ 1−
( 2.01 x 10−4
5.31 x 10−4
m3 1000 l
)
Qo =8.41 x 10−3 (
s 1 m3 )( 160mins )
L
¿ 8.41
min
For venturi:
√ 1−
( 2.01 x 10
5.31 x 10−4
m3 1000 l
)
Qv =5.43 x 10−5 (
s 1 m3 )( 160mins )
L
¿ 3.26
min
Volumetric flow,
Q = 5 L/min
3
L 1min
= (5 ¿( 1 m ) ( )
min 1000 L 60 s
= 8.3333 x 10−5 m3 / s
D2
Area of 90˚ elbow, A=π ( )
4
¿ π ¿)
= 0.0005310 m2
Q
Average velocity, v =
A
8.3333 x 10−5
=
0.0005310
= 0.1567 m/s
2g
Loss of coefficient, k = h x
v2
2 x 9.81
= 0.001 x
0.15672
= 0.799
D2
Area of 90˚ elbow, A=π ( )
4
¿ π ¿)
= 0.0005310 m 2
Q
Average velocity, v =
A
8.577 x 10−5
=
0.0005310
= 0.1615 m/s
2g
Loss of coefficient, k = h x
v2
2 x 9.81
= 0.001 x
0.16152
= 0.752
2.6 DISCUSSION
From the experiment, we got the value of the flow rate is 5.146 L/min,
3.26 L/min for venture and 8.41 L/min for orifice with rotmeter value of 5 L/min. For
rotmeter reading of 10 L/min, the flow rate is 11.278 L/min, 6.17 L/min for venturi and
18.86 L/min for orifice. For rotmeter reading of 15 L/min, the flow rate is 19.108 L/min,
9.59 L/min for venturi and 30.76 L/min for orifice. For rotmeter reading of 20 L/min, the
flow rate is 21.429 L/min, 10.35 L/min for venturi and 34.09 L/min for orifice. All of
these vaue might be difference from the actual value. This happens because of some error
that we had done during conducting this experiment.
First of all, the error might cause from the parallax error that happen
during taking the reading of the manometer. This will affect the whole calculation of the
experiment. Next, the error might come from the air bubbles that contain on the stream
of fluid flow through the apparatus. It might be that we did not notice about the presence
of the air bubble. This will affect the flow of the fluid and the reading of the manometer.
Other than that the error might be happen when we were taken the time for
the 3 L of fluid flow. It might happen that we start or stop the time at slightly did not
same as we started to flow or stop the flow of the fluid.
2.7 CONCLUSION
In this experiment, there are three basic type of flowmeter which are rotmeter,
orifice, and venture. The readings of volume flow rate in all flowmeter should have the
same value to the ideal flow rate. The values of volume flow rate that we obtained for
rotmeter was 5, 10, 15, and 20 L/min, venture (3.26, 6.17, 9.59 and 10.35 L/min) and
orifice (8.41, 18.86, 30.76 and 34.09 L/min). Based on the values of the volume flow
rate, rotmeter has the most nearest value to ideal volume flow rate. For the determination
of the loss coefficient when fluid flow through 90 degree elbow was determined and the
value is 0.75, 0.47, 0.33 and 0.35 when the volume reach 3 L.
2.8 RECOMMENDATIONS
After complete doing the experiment, do calculation and obtain a result, there are
some recommendations in order to obtain more accurate result in future. Firstly, avoid
parallax error when reading the volume of water at each flow by placing the eyes directly
to the meniscus point. This is the important steps because the value of reading will give
the main result. To prevent from error, the student must look carefully and more than one
student read that value to ensure the value is correct. Besides that, make sure that there is
no air bubbles inside the equipment during the experiment. This will happen by flow the
water inside the equipment in longer time and connect the water supply correctly and
ensure there is no leakage inside the system. Another step in order to obtain more
accurate result is make sure measure the time correctly by using a stopwatch. The normal
digital clock because this will give less accurate result. Finally, make sure the calculation
was doing right and not make any error when doing the calculations.
2.9 REFERENCES
2.10 APPENDICES