Professional Documents
Culture Documents
1 333333endo All PDF
1 333333endo All PDF
CONTROL
RESOURCE AND CONSULTING
SERVICE
ENDOSCOPE SURVEY
1.0 Introduction 2
1.1 Emergence of a perceived issue/problem 2
6.0 Potential 6
Issues/Concerns
6.1 Expected or likely issues 6
6.2 Proposed management of concerns 7
6.3 Evidence 8
9.0 Recommendations 16
10.0 Implementation 18
i
Table of Contents
Page
12.0 References 20
Appendices
1. Alert Notice 22
2. Results of Endoscope Survey 23
3. HRICRCS recommendations for biological 39
monitoring of endoscopes and automated
endoscope reprocessors
4. Positive Test - action plan 40
5. Memorandum of Understanding proforma 42
6. Endoscope microbiological testing history 45
proforma
7. Endoscope processing history proforma 46
8. Questionnaire tool 47
ii
1.0 Introduction
A current clinical issue identified within our region is the provision of a ‘safe’
endoscopy service to customers. This involves cleaning, processing, storage, use
and monitoring activities of endoscopes and endoscope processing equipment. The
delivery of a safe endoscopy service is also reliant on nursing and medical
personnel being skilled to perform the procedures and activities associated with the
delivery of the procedure within a safe working environment. These activities are
required to meet current standards and guidelines with regard to use and
reprocessing of endoscopes in the clinical setting and provision of patient care.
HRICRCS are currently conducting site visits and audits of compliance of regional
hospitals’ Theatre/CSD units with AS 4187 and other relevant guidelines. This
included an overview of the endoscopy service, if provided. During these site visits
HRICRCS identified that processing and biological monitoring of owned and loan
endoscopes-especially flexible endoscopes was variable to the recommended
guidelines without explanation or justification for the variance. Monitoring activities
at some hospitals at best could be described ‘ad hoc’, or were non-existent or not
known. Some of the smaller hospitals within the region were delivering a service
with limited resources and equipment and compromised working environment.
2
3.0 Problem Statement
Endoscopes and the equipment used to process endoscopes within hospitals in the
region may not be undergoing appropriate processing and monitoring activities to
meet recommended guidelines. Are current recommended guidelines evidence
based and supported by scientific research, and can non-compliance affect patient
outcomes?
1. What, if any, is the significance of the microbiology results and what do they
mean? What are the implications for patient outcomes?
2. Are current recommended guidelines evidence based? if not, are they still
relevant?
a. Patient health
b. Staff health
c. Service provider liability
d. Endoscope provider liability
e. Endoscope life-span
f. Cost implications of all the above
by;
3
4.0 Key Stakeholders
AIM- To receive the best possible ‘care’. Most likely are not aware of, or witness
processing and monitoring activities and therefore have little comprehension of the
importance of these activities, and what is or should be occurring. Rely on
caregivers and hospital to ensure ‘safe environment is provided’ including provision
of cleaned reprocessed endoscopes.
AIM- Are able to provide the best possible care using the best possible equipment
within a safe working environment. This may be hindered by lack of ability to have
the best equipment or resources to hand due to financial constraints or inadequate
risk assessment strategy.
Directors of Nursing are most likely to be the Executive Sponsor for the Infection
Control Program within their agencies. Other Hospital Administration Officers are
involved if ethico-legal, and/or risk minimisation implications arise.
AIM- Are able to provide the best possible care within a costed framework, that is
‘’the best affordable care’. Resource decisions are based upon availability and
prioritisation of limited resources and a risk assessment of need to/no need to
comply with all or part of recommendations as determined by potential outcomes.
AIM- Are able to provide the best possible medical care. The provision of medical
care is supported by the hospitals providing current and up to date equipment (AS
approved), and experienced educated staff working. Able to have input into
equipment purchases and changes to procedure that affect the way in which they
provide their service to the patient.
4
4.5 HRICRCS-active participants-
Two consultants who work within a framework of assisting and supporting health
care facilities in the region to help prevent, minimise and control infection risks and
poor outcomes from non-compliance or breaches in infection control practices.
Responsible for conducting the review of endoscopic services, making
recommendations on the findings from this review and conducting the literature
search.
AIM- Organisations in the Hume Region provide a ‘safe environment’ for the delivery
of the endoscopy service to its customers. This includes all patients, staff and
providers of the service. In providing a safe environment the organisation engages
in risk assessment and resource management. Any adjustment within the
environment to comply with recommended standards and guidelines, will remain
compatible with the hospital philosophy and the philosophy of other key
stakeholders.
5
6.0 Potential issues or concerns
6.1 Active Key Stakeholders may have the following concerns-
6.1.2 Accountability of key stakeholders for patient outcomes arising from use
of scope is identified, especially if processing cannot be demonstrated to
have met current guidelines
6.1.4 Have practices identified which may lead to urgent review or cessation of
their endoscope service or de-commissioning of equipment
6.1.5 Change current activities undertaken within their own organisations with
regard to processing and monitoring of endoscopes
6
6.2 Proposed management of concerns
6.2.1 Fear that change will have a negative effect on current work dynamics.
This may hinder involvement in any initiatives resulting from the
investigation of endoscope processing. Key stakeholders may be
ambivalent towards being involved due to minimal knowledge of current
guidelines and the importance of meeting same.
6.2.2 Not want to be identified as the accountable body, this raises liability
concerns, which may have remained unresolved or even unrecognised to
date. Organisations may not want to open a potential ‘can of worms’ even
if it is in the best interest of the organisation and the patient.
6.2.4 Fear that review of current practices used by their organisation may
involve ceasing their endoscope service due to not meeting minimum
requirements
Assure key stakeholders that HRICRCS do not have the capacity to cease
their endoscopic services. HRICRCS will highly recommend changes and
upgrading to meet standards and guidelines, however the decision to
cease providing the endoscope service will ultimately remain with the
organisation and other key stakeholders providing the service.
7
6.2.5 Fear that review of current practices used by their organisation may
involve changing current practices to their endoscope service due to not
meeting minimum requirements
As above for both points 3 and 4, reassure that any changes made to
current practice is at the prerogative of key stakeholders.
6.3 Evidence
1. Information and feedback using the survey tool that demonstrates that
most hospitals within the region are not following the recommended
GENSA-1999 guidelines.
8
7.0 Literature Review
HRICRCS acknowledges that much has been written on the subject of endoscopes
and endoscopes processing, especially with reference to the use and concentration
of chemical disinfectants and differences of opinion in the timeframe required for
soaking of instruments to achieve high level disinfection. Lewis (1999, 34)
There has also been numerous articles written on the importance of cleaning
endoscopes to minimise bioburden and soil prior to the actual disinfection process.
Alfa, Degagne and Olson (1999, 2). In addition, studies on bronchoscopes have
been undertaken to assess the potential risk of transmission of mycobacteria strains
between cases. These studies have looked at potential reasons why contamination
of bronchoscopes with mycobacteria and other isolates has occurred, issues such
as rinsing water, ineffective alcohol and/or pressured air drying have been
investigated as likely causes of contamination. Muscarella (1996)
More recently with the advent of AER’s studies have looked at the efficiency of
manual cleaning and processing vs. manual cleaning and automated processing. A
study by Fraser, Zuckerman, Clouse, O’Rourke, Jones, Klasner and Murray (1993,
383-9) investigated the contamination levels post processing using microbiological
cultures taken from the scopes. The study showed that prior to an alcoholic rinse,
27% of endoscopes processed through AER’s, and 37% of endoscopes processed
manually grew gram negative and non-tuberculous mycobacteria, these levels
dropped to 10% and 27% respectively when alcoholic rinse was used at the end of
processing. Although the study stated that the results were not statistically
significant, of importance was that despite both manual and automated cleaning,
persistent contamination was apparent.
However, there is a paucity of evidence based research that examines the need to
microbiologically monitor endoscopes and endoscopic processing equipment. Most
current recommendations are empirical and are focussed on the test results
indicating endoscope processing failure. This failure does not indicate however if
there is a resultant infection risk to patients. Studies have yet to be conducted that
demonstrate a relationship between non-compliance with biological monitoring and
deleterious patient outcomes. This will be further expanded in the review of
literature
9
7.1 Conduction of Literature Review
Search words
Endoscope/ and: microbiological monitoring, cleaning, processing,
disinfection, sterilisation, automated reprocessors, environmental sampling,
infection control, quality assurance.
The search was not exhaustive, some articles found in the Cochrane Library
were not available in English text and only the abstract could be used as a
generalised reference.
10
7.3 Current Recommendations and Guidelines
7.3.1 Australian
7.3.2 Overseas
The ESGE and ESGNE (no year cited) protocols do not make reference to routine
microbiological testing. The BSG 1998 working party report and the ASGE-1996
Position Statement also make no reference to the need for microbiological testing.
In an APIC special communications article, Alvarado and Reichelderfer (1999, 149)
discuss quality control and purport that the CDC does not recommend routine
microbiologic sampling of inanimate hospital environment and patient care items.
However if an endoscopy related infection is observed focused testing would be
warranted.
The Asia pacific Congress of Digestive Endoscopy working party was convened in
Hong Kong in 2000 to form a mandate on disinfection of endoscopes and
accessories, the aim being to lay down ‘minimal standards’ for disinfection and
reprocessing of endoscopes and accessories. The working party report by Tandon
(2000, 71) highlights quality control measures including periodic bacteriologic
monitoring of the endoscope and accessories. HRICRCS again notes that no time
frames are identified for such monitoring.
11
examined the differences between guidelines and reprocessing recommendations
with respect to cleaning, quality assurance and evidence base for these
recommendations. In an overview of the study they stress the need for
‘experimental investigations’ of the cleaning and disinfection process, however
experimental investigations was not able to be determined to mean microbiological
testing, although HRICRCS believe it was inferred.
In an article written by Muscarella (2002, 285-9), a well known and respected doyen
of all things endoscope, he acknowledges that routine sampling is not generally
recommended by the CDC and other relevant bodies in the US, although sampling
of rinse water in some reprocessing settings is supported. Muscarella argues that
rinse water used in endoscopy processing should also be routinely sampled due to
reported outbreaks of gram negative bacteria nosocomial infection outbreaks in
endoscopy patients.
All guidelines and most of the articles cited emphasised the need for stringent
manual cleaning of endoscopes as the best possible way to minimise the
transmission of infection between endoscopy patients.
Some sources say that the risk of transmission of infection is minimal, Leung (2000,
74) reports that the incidence of transmission of infection was one in 1.8 million (28
reported cases) as estimated by ASGE. The infections were thought to have
resulted from breaches in recommended cleaning and disinfection procedures and
contaminated AER’s.
However in contrast to these figures, Spach, Silverstein and Stamm (1993, 117-
128), in their review of 265 articles on transmission of infection by endoscopy,
identified that 281 and 96 cases of infection transmission via gastrointestinal scopes
and bronchoscopes had occurred respectively. (Spach et al 1993, 117) state
“Salmonella species and Pseudomonas aeruginosa were repeatedly identified as
the causative agents of infections transmitted by gastrointestinal endoscopy, and
Mycobacterium tuberculosis, atypical mycobacteria were the most common causes
of infections transmitted by bronchoscopy. One case of hepatitis B virus
transmission via endoscopy was documented”. Of the 84 cases of Salmonella
identified 1 patient died.
Gardner and Peel (1998, 247) cite that a study by Birnie et al, implicated
endoscopes with the transmission of Hepatitis B virus and studies by Wheeler et al,
Classen et al and Spach et al identified that endoscopes were responsible for
transmitting opportunistic pathogens from one patient to another patient. The
outcomes from these infections have ranged from colonisation to serious infection
and death. A report by Michele, Cronin, Graham, Dwyer Pope, Harrington,
Chaisson and Bishai (1997, 1093-5) demonstrated via DNA fingerprinting that
Mycobacterium tuberculosis was transmitted via a fiberoptic bronchoscope in a
health care facility in USA from one patient to another patient. A further report by
Bronowicki, Venard, Botte, Monhoven, Gastin, Chone, Hudziak, Rhin, Delanoe
12
LeFaou, Bigard and Gaucher (1997,237) reported two cases of Hepatitis C virus
were transmitted via a colonoscope used on a Hepatitis C positive patient.
A recent newspaper article “Unsterile Devices Prompt Warnings” in the Los Angeles
Times (Glionna 2003) was written in response to The California Department of
Health Services issuing warnings after two hospitals in the Sacramento area
advised 3,250 patients they would need to take blood tests as they were possibly
exposed to an infectious disease via an unsanitised chamber in an instrument called
a colonoscope. The article goes on to describe a history of incidents where
transmission or potential transmission of infection has occurred via an endoscope.
The article also highlights that manufactures have long been aware of problems with
cleaning and processing of endoscopes and have stated that “satisfactory cleaning
cannot be achieved” when filing for patent.
# There is conflicting advice from leading bodies and working parties as to the
need to microbiologically monitor endoscopes and endoscope reprocessors,
including rinsing water. There is no definite scientific evidence that
demonstrates microbiological monitoring prevents transmission of infection by
an endoscope.
13
8.0 Discussion of Survey Results
The results of the review demonstrate that biological monitoring of loan and hospital
owned endoscopes and the equipment in which they are processed varies greatly
between hospitals.
The GENSA –1999 guidelines indicated that routine biological testing should be
carried out on all endoscopes, however the results show that some endoscopes,
especially loan scopes are not biologically monitored within the recommended
timeframes, or are not being monitored at all. Of particular concern is the fact that
most endoscopes being processed via AER’s are not being tested 4 weekly
(monthly) as recommended, especially loaned scopes which had the least
monitoring activities undertaken.
The reasons why some organisations chose to not test any scopes or processing
equipment is not clear. Although the GENSA –1999 guideline is the most cited
reference by organisations for development of policy and procedure related to
cleaning and processing of scopes, many hospitals still failed to meet the
recommendations. Given that an alert notice was distributed 3 months prior to the
survey, hospitals cannot claim ignorance of these recommendations as an excuse
for non-compliance.
14
disinfection processes are not appropriate for rigid endoscopes. There are
no further recommendations for rigid endoscopes
Although rigid scopes are reported upon within the context of this
report, the aim of the review was to identify current biological
monitoring of flexible endoscopes, and endoscope processing in
accordance with GENSA –1999 and other relevant guidelines.
8.2 Conclusion
Endoscope processing and monitoring activities throughout the region are not
standardised and do not reflect current recommended guidelines.
Of concern is the ongoing use of open glutaraldehyde systems still in use despite
significant Occupational Health and safety concerns (not addressed in this report)
Rotating loan scopes are more than likely undergoing different processes at different
sites. There is concern that this practice is deleterious to the scope and may affect
performance and integrity of same over time (not addressed in this report)
15
9.0 Recommendations
Recommendation One
All Hospitals have a system for tracking endoscopes and endoscope equipment
and processing method to patients.
Rationale
Recommendation Two
Rationale
Recommendation Three
Develop endoscope history records that stay with the endoscope. One record
details the endoscope’s processing history, the other record details the
endoscope’s biological monitoring history.
Rationale
Records are kept with every endoscope. This details processing history
(AER/Steris and results of biological tests on that equipment) and microbiological
test undertaken on the endoscope itself. This will help determine if problems are
occurring with a particular piece of equipment, and will aid in tracking any
positive results back to patient via tracing systems.
16
Recommendation Four
Rationale
Recommendation Five
Rationale
Recommendation Six
Regular education and training sessions for existing staff and for orientation and
training for new staff reprocessing endoscopes
Rationale
Recommendation Seven
Rationale
Recommendation Eight
Rationale
17
global initiative and as a response to definitive requirement by regulatory bodies
for all endoscopes to be sterilised.
Recommendation Nine
Rationale
Rigid scopes are required to be sterile as they enter sterile body tissue and
cavities and are described as ‘critical items’ (page 41, NH&MRC-1996)
10.0 Implementation
Following ratification by the Advisory Committee and referral to the Hume Region
Nurse Executives for comment, the endoscope report with the above
recommendations will be distributed to hospitals within the Region.
Education sessions with relevant parties explaining the recommendations and the
documentation for recording will be available upon request.
18
11.0 Addressing the Problem Statement
1. What, if any, is the significance of the microbiology results and what do they
mean? What are the implications for patient outcomes?
2. Are current recommended guidelines evidence based? if not, are they still
relevant?
3. Will deleterious patient outcomes result from organisations that do not comply
with the recommended guidelines?
19
12.0 References
Alfa, M. Degagne, P. Olson, N. Worst –case soiling levels for patient used flexible
endoscopes before and after cleaning. American Journal of Infection Control Vol 27
(5) October 1999 pp1-12
Gardner, J. and Peel, M. Sterilisation, Disinfection and Infection Control. 3rd Ed.
Churchill Livingstone Publishers, Malaysia,1998
20
Gastroenterological Society of Australia (GESA). Gastroenterological Nurses
Society of Australia (GENSA). Infection Control in Endoscopy. Cowen, A. Editor.
Gastroenterological Society of Australia Publishers Sydney, 1999
Glionna, J. Unsterile Devices Prompt Warnings. The Los Angeles Times. February
13, 2003
National Health and Medical Research Council (NH&MRC) Infection Control in the
Health Care Setting. Australian Government Publishers, Canberra 1996
21
Appendix One
ALERT!
Microbiological testing of endoscopes (including bronchoscopes) and
automatic endoscope disinfectors
It has come to the attention of the Hume Region Infection Control Resource and
Consulting Services that some health care facilities in the region may not be aware
of the recommendations, relating to biological testing of scopes, contained in the
“Guidelines for Infection Control in Endoscopy” published by the Gastroenterological
Society of Australia –1999 (This document is commonly referred to as the GENSA-
guidelines).
4. If major changes are made in the Endoscopy Unit personnel responsible for
cleaning, or if there is a clinical suspicion of cross-infection related to
endoscopy, then further microbiological screening should be undertaken in
conjunction with a Clinical Microbiologist.
If you are processing endoscopes using automatic processors you may wish to
review your current procedures to determine if this recommendation is being
implemented in your facility.
Endoscope Questionnaire
The Hume Region Infection Control Resource and Consulting Service will be
forwarding a short questionnaire to all health care facilities in the region in the near
future to ascertain compliance with this recommendation with regard to facility
owned and loaned endoscopes.
The purpose of the questionnaire is to determine the current status within the region
in regard to processing and biological testing of scopes and will help identify and
determine, if any, regional strategies that may need to be implemented.
22
Appendix Two
In the event of current endoscopes being unavailable for use, only 5 (33%) of
organisations indicated that they would use interim loan endoscopes, yet 13 (87%)
had a strategy for interim loan as part of the contingency plan. Of the two hospitals
that did not state use of interim loan endoscopes, 1 (7%) indicated that they would
suspend the list, whilst the 1 (7%) remaining organisation did not have a
contingency plan.
The total number of gastroscopes in use in hospitals in the region is 38. Of these, 24
are owned by 10 hospitals, and 14 are loaned by 7 hospitals. Of the 7 hospitals
using the 14 loaned gastroscopes, there are 7 permanent and 7 on rotating loan.
Three of the 7 hospitals (43%) use permanent loan scopes and four (57%) use
rotating loan. This equates to four hospitals within the region using up to 7 loaned
gastroscopes. The number of actual rotating loan scopes is difficult to specify as the
same loan scope may be used by different hospitals, but each hospital has recorded
the scope.
There are currently 39 colonoscopes in use in hospitals throughout the region. Nine
hospitals own 21 of these scopes, whilst 18 are loaned by 8 hospitals. Of the 18
loan scopes, 9 are permanent loan and 9 are on rotating loan. Four of the 8
hospitals use rotating loan scopes. This equates to 4 hospitals within the region
using 9 rotating colonoscopes
The three hospitals providing a bronchoscopy service use 5 bronchoscopes, the two
hospitals providing the laryngoscopy service use 2 laryngoscopes and one hospital
providing the duodenoscopy service uses 3 duodenoscopes. All use and process
their own scopes.
Of the remaining loan scopes in use, one hospital uses a 1 rotating loan
laparoscope, and one hospital uses 1 rotating loan arthroscope.
23
The most cited reference by organisations for development of policy and procedure
related to cleaning and processing of scopes is GENSA-1999 with 11, followed by
automated endoscope processor manufacturers instructions 10 (two types of
processors). Next came AS 418-1998 with 8, endoscope manufacturers instructions
with 5, and ACORN which was cited 4 times. The NH&MRC-1996 and Hospital
Procedure Manual were cited only once each.
Of the 11 organisations using own scopes, 8 (73%) identified that they had a
documented procedure in place for biological monitoring of the scopes, 1 (9%)
organisation was in the progress of writing a procedure and 2 (18%) organisations
have no documented procedure.
Gastroscopes
One hospital is using a vented fume cabinet with glutaraldehyde for processing of
own scopes, and one hospital is using the same method for loan scopes. Two
hospitals are using vented fume cabinet with OPA for processing of loan scopes
Six hospitals are using automated endoscope reprocessors (AER’s), of these five
are using Soluscope with glutaraldehyde for owned and loan scopes, and one is
using a Medivator with OPA for own scopes. Seven hospitals are using a Steris
unit, six are processing own scopes and one is processing loan scopes.
Colonoscopes
Two hospitals are using vented fume cabinets for processing of loan colonoscopes,
one is using glutaraldehyde, and the other hospital is using OPA. Six hospitals are
using AER’s, of these, five hospitals are using Soluscope with glutaraldehyde for
processing of own and loaned scopes. The Medivator with OPA is used by one
hospital for processing own scopes. Please note that one hospital uses both
Soluscope and Steris for processing of loan colonoscopes
24
Bronchoscopes
The three hospitals providing bronchoscopy service, all process their own
bronchoscopes through a Steris unit.
Other scopes
The one hospital utilising the duodenoscopes, processes owned scopes in the Steris
unit. Cystoscopes are processed by four hospitals in the Steris unit, three hospitals
process their own scopes, whilst the other hospital processes a permanent loan
scope. Flexible laryngoscopes are processed in the Steris by the two hospitals who
own these scopes. The single laparoscope and arthroscope reported are rotating
loan scopes and are processed in the same hospital in an AER using OPA.
The timeframes for biological monitoring of units varied between organisations and
between processing methods utilised by each organisation. The 5 hospitals using
the Soluscope System for high disinfection all biologically monitored their units, of
these, 3 tested monthly, 1 bi-monthly and 1 quarterly. The only hospital using the
Medivator tested monthly. Hospitals using the Steris system for sterilisation of
endoscopes all biologically monitored their units, 1- fortnightly, 4- weekly and 2-
monthly. The organisations using the vented fume cabinets did not biologically
monitor the units.
Gastroscopes
Owned
Monitoring ranged from fortnightly to not tested at all, with one hospital giving no
answer. Of the three hospitals using the AER’s, two tested monthly, one gave no
answer. Of the six hospitals using the Steris System, one tested fortnightly, three
quarterly, one every 4 months and one hospital not tested the scope at all. The
remaining hospital using vented fume cabinet for processing, tested the scope
quarterly.
25
Loaned
Monitoring ranged from 3 monthly to not tested at all, with two organisations giving
no answer. Of the three hospitals using AER’s one tested quarterly, one did not
test, and one did not answer. The single hospital using the Steris system tested the
scope on arrival to the hospital. Of the three hospitals using vented fume cabinets,
one did not test, one didn’t know if they tested, and the other hospital gave no
answer.
Colonoscopes
Owned
Monitoring ranged from monthly to not tested at all, with one hospital giving no
answer. Of the three hospitals using the AER’s, one tested monthly, one hospital did
not test and the other one gave no answer. Of the six hospitals using the Steris
System, four tested quarterly, one every 4 months and one hospital did not test the
scope at all. No hospitals processed own scopes in vented fume cabinets
Loaned
Monitoring ranged from on arrival to not tested at all, with two organisations giving
no answer. Of the four hospitals using AER’s two did not test, and one did not
answer. Of the three hospitals using the Steris system, one tested the scope on
arrival to the hospital, two did not test scopes at all. Of the two hospitals using
vented fume cabinets, one did not test and the other hospital gave no answer.
Other endoscopes
Owned
The three hospitals using bronchoscopes all processed them in the Steris System
one tested 4 weekly, two tested monthly (please note that questionnaire format may
have affected this result with 4 weekly not given as a selection, the 4 weekly
response was added in other). The hospital using the duodenoscopes processed
the scope in the Steris unit and tested 4 weekly. Of the three hospitals using
cystoscopes, one tested monthly and two tested quarterly. The two hospitals using
laryngoscopes tested quarterly and processed the scopes in a Steris unit.
Loaned
The hospital using both loan laparoscope and arthroscope tested monthly and
processed the scopes in a Medivator unit. The only hospital using loan cystoscope
tested 4 monthly and the scope was processed in the Steris System.
26
Hume Region Infection Control
Resource and Consulting Service
Collated Endoscope Data
N Percentage
N=15
In the event of the need for interim loan of scopes (as part of your contingency plan)
are they/or would they be provided by?
27
Hume Region Infection Control
Resource and Consulting Service
Collated Endoscope Data
Gastroscopes 24 (10)
Colonoscopes 21 (9)
Bronchoscopes 5 (3)
Duodenosocopes 3 (1)
Cystoscopes 9 (3)
Laryngoscopes 4 (2)
Loaned Scopes P R P R
Bronchoscopes (0)
Sigmoidoscope** 1 (1)
P=Permanent loan R=Rotating loan
*Total number of reported scopes by each organisation, however same scope may have been
reported more than once, if rotating between hospitals.
**Sigmoidoscope was listed as loaned scope by one hospital but details of processing and monitoring
was not identified in returned questionnaire, therefore not further reference is made to this endoscope
What guidelines and standards do you use to formulate your hospital’s policy and
procedure documents regarding the cleaning and processing of owned and loan
endoscopes?
Name of Standard or Guideline Number of hospitals using this guideline
GENSA/GENCA 1998 11
NH&MRC 1
AS4187 8
ACORN 4
Scope Manufactures Instructions 5
Hospital Procedure manual 1
Gallay Scientific 1
Device Technology 9
28
Hume Region Infection Control
Resource and Consulting Service
Collated Endoscope Data
Does your hospital have a documented procedure in place that identifies how often
biological monitoring of your own scopes is to occur?
N= 15
Denominator 11 (organisations using own scopes)
Yes 8 73%
No 2 18%
In progress 1 9%
*Note that one respondent who answered no to use of own scopes, answered yes to this
question, but endoscopes are permanent loan and respondent represented in no
answer/not applicable
**Reponses not included in % as respondents use only loaned endoscopes
Does your hospital have a documented procedure in place that identifies how often
biological monitoring of loaned endoscopes is to occur?
N= 15
Denominator 11 (organisations using loan scopes)
Yes 2* 18%
No 9 82%
*Two of the organisations who responded yes to this question who do not routinely use
loaned scopes indicated that they would;
1. test loan scopes on arrival to organisation and as per own scope routine for the
period of the loan
2. test scopes on arrival to the organisation
Respondents represented in no answer/not applicable
29
Hume Region Infection Control
Resource and Consulting Service
Collated Endoscope Data
Does the hospital undertake responsibility for biological monitoring of loan scopes?
N=15
Denominator 11 (organisations using loan scopes)
Yes 3* 27%
No 8 73%
*Two of the organisations who responded yes to this question who do not routinely use
loaned scopes indicated that they would;
3. test loan scopes on arrival to organisation and as per own scope routine for the
period of the loan
4. test scopes on arrival to the organisation
Respondents represented in no answer/not applicable
Does the provider of endoscopic service (brings own scopes) take responsibility for
biological monitoring of endoscopes?
N=15
Denominator 8 (organisations using rotating loan scopes)
Yes 3 37%
No 5 63%
N=8
Denominator 8 (organisations answering yes to above)
Yes 1* 13%
No 7** 87%
*The respondent answering yes to this question, answered no to the above. The intent of
the question was to identify formalised agreements affirming service provider responsibility,
not formalising an agreement for service provider not to have responsibility
**One respondent answered no to this question when response should have been no
answer or not applicable
30
Hume Region Infection Control
Resource and Consulting Service
Collated Endoscope Data
If agreement is not yet formalised, are you in the process of formalising the
agreement?
N=7
Denominator 8 (organisations answering no to above)
Yes 2 29%
No 5 71%
*Same respondent who indicated that formalised agreement was for service provider not to
have responsibility for biological monitoring-Respondent not included in the data
31
Hume Region Infection Control
Resource and Consulting Service
Collated Endoscope Data
Processing Method Fortnightly 4 weekly Monthly 3 Monthly 4 Monthly 6 Monthly On Arrival Not tested Don’t Know No Answer
Soluscope using 1 1
Glutaraldehyde
Medivator using 1
OPA
Steris using 1 3 1 1
Paracetic acid
VFC using 1
Glutaraldehyde
VFC using
OPA
Processing Method Fortnightly 4 weekly Monthly 3 Monthly 4 Monthly 6 Monthly On Arrival Not tested Don’t Know No Answer
Soluscope using 1 1
Glutaraldehyde
Medivator using 1
OPA
Steris using 4 1 1
Paracetic acid
VFC using
Glutaraldehyde
VFC using
OPA
32
Hume Region Infection Control
Resource and Consulting Service
Collated Endoscope Data
Processing Method Fortnightly 4 weekly Monthly 3 Monthly 4 Monthly 6 Monthly On Arrival Not tested Don’t Know No Answer
Soluscope using
Glutaraldehyde
Medivator using
OPA
Steris using 1 2
Paracetic acid
VFC using
Glutaraldehyde
VFC using
OPA
Processing Method Fortnightly 4 weekly Monthly 3 Monthly 4 Monthly 6 Monthly On Arrival Not tested Don’t Know No Answer
Soluscope using
Glutaraldehyde
Medivator using
OPA
Steris using 1
Paracetic acid
VFC using
Glutaraldehyde
VFC using
OPA
33
Hume Region Infection Control
Resource and Consulting Service
Collated Endoscope Data
Processing Method Fortnightly 4 weekly Monthly 3 Monthly 4 Monthly 6 Monthly On Arrival Not tested Don’t Know No Answer
Soluscope using
Glutaraldehyde
Medivator using
OPA
Steris using 1 2
Paracetic acid
VFC using
Glutaraldehyde
VFC using
OPA
Processing Method Fortnightly 4 weekly Monthly 3 Monthly 4 Monthly 6 Monthly On Arrival Not tested Don’t Know No Answer
Soluscope using
Glutaraldehyde
Medivator using
OPA
Steris using 2
Paracetic acid
VFC using
Glutaraldehyde
VFC using
OPA
34
Hume Region Infection Control
Resource and Consulting Service
Collated Endoscope Data
LOANED SCOPES
Processing Method Fortnightly 4 weekly Monthly 3 Monthly 4 Monthly 6 Monthly On Arrival Not tested Don’t Know No Answer
Soluscope using 1 1 1
Glutaraldehyde
Medivator using
OPA
Steris using 1
Paracetic acid
VFC using 1*
Glutaraldehyde
VFC using 1 1
OPA
*Not tested by the organisation-owner responsibility
Processing Method Fortnightly 4 weekly Monthly 3 Monthly 4 Monthly 6 Monthly On Arrival Not tested Don’t Know No Answer
Soluscope using 2,1a 1
Glutaraldehyde
Medivator using
OPA
Steris using 1, 1,1b
Paracetic acid
VFC using 1*
Glutaraldehyde
VFC using 1
OPA
*Not tested by the organisation-owner responsibility
a,b denotes same organisation, but may use different processes for colonoscopes
35
Hume Region Infection Control
Resource and Consulting Service
Collated Endoscope Data
LOANED SCOPES
Processing Method Fortnightly 4 weekly Monthly 3 Monthly 4 Monthly 6 Monthly On Arrival Not tested Don’t Know No Answer
Soluscope using
Glutaraldehyde
Medivator using 1
OPA
Steris using
Paracetic acid
VFC using
Glutaraldehyde
VFC using
OPA
Processing Method Fortnightly 4 weekly Monthly 3 Monthly 4 Monthly 6 Monthly On Arrival Not tested Don’t Know No Answer
Soluscope using
Glutaraldehyde
Medivator using
OPA
Steris using 1
Paracetic acid
VFC using
Glutaraldehyde
VFC using
OPA
36
Hume Region Infection Control
Resource and Consulting Service
Collated Endoscope Data
LOANED SCOPES
Processing Method Fortnightly 4 weekly Monthly 3 Monthly 4 Monthly 6 Monthly On Arrival Not tested Don’t Know No Answer
Soluscope using
Glutaraldehyde
Medivator using 1
OPA
Steris using
Paracetic acid
VFC using
Glutaraldehyde
VFC using
OPA
Note one organisation indicated that they would test all loaned scopes on arrival if in use as interim replacements
37
Hume Region Infection Control
Resource and Consulting Service
Collated Endoscope Data
How often is the equipment used to process hospital owned and loaned scopes biologically monitored?
Processing Method Fortnightly Weekly Monthly 2 Monthly 3 Monthly 4 Monthly 6 Monthly Not tested No
Applicable
Soluscope 3 1 1
Medivator 1
Steris 1 4 2
VFC 3
Table 13 Utilisation of specific processing method for each type of scope per organisation
Processing Gastroscopes Colonoscopes Bronchoscopes Duodenoscopes Cystoscopes Laryngoscopes Laparoscopes Arthroscopes Total
Method
Own Loan Own Loan Own Loan Own Loan Own Loan Own Loan Own Loan Own Loan Own Loan
Soluscope 2 2 2 3 0 4 5
Medivator 1 0 1 0 0 1 1 2 2
Steris 6 2 6 3 3 1 4\3 1 2 22 7
VFC 1 3 0 2 0 3 5
TOTAL 10 7 9 8 3 1 3 1 2 1 1
38
HRICRCS recommendations for
Biological monitoring of endoscopes and endoscopes
reprocessors (AER’s)
*Recommendations derived form GENSA guidelines 1999. Research and review of the validity of
biological monitoring is occurring at present, and challenges to the validity of the GENSA
recommendations may become apparent in the future. However, until this research is established
and there is evidence to prove or disprove the validity of such testing, GENSA recommendations
should be followed. In Australia this body is recognised as having the leading experts in this field. It
is worth noting GENSA guidelines also only make reference to disinfection procedures and biological
monitoring, however HRICRCS recommends that flexible endoscopes processed via sterilisation are
monitored in the same way as disinfected endoscopes until further research proves that testing is not
required.
*Disinfection reprocessors as per GENSA guidelines, other units as per manufacturers instructions or
organisation discretion as units are not automated.
39
Positive Test Result
Action Plan
The person nominated to receive the test results (Refer Endoscopy Service- Memorandum of
Understanding) should contact the Director of Medical Services and or the Infection Control
Practitioner, and person in charge of Endoscopy Unit where the endoscope has been used to
formulate a working party and action plan.
Positive test results should be identified as a ‘potential critical incident’ and although it is not
mandatory to report results as described above, the Department of Human Services suggests
that the Regional Office in your area be contacted in the event the issue progresses to a critical
incident, with the possibility of media exposure and community knowledge occurring. As such
the organisation should implement the critical incident management plan and identify the
contact /liaison person for the organisation.
In the event of a positive test result identify the organism and level of growth. Consult
microbiologist to determine possible origin of the isolate (water borne, environmental, skin flora
etc) to eliminate contamination at time of sampling as source of positive result. Discuss the
significance of the result with microbiologist and potential outcomes for patients exposed to;
the endoscope with a positive test result, or endoscopes processed in an automated machine
with positive test result. Re-test to determine if isolate
is still insitu.
Actions
Contact other organisations (if any) that have used the same endoscope following the positive
test result.
Contact the owner of the endoscope (if not another organisation). Discuss clinical significance
and actions your organisation is taking in response to positive test result.
Dependent on the clinical significance of the results the organisation must decide upon actions
that reflect the risk management strategy and critical incident action plan. This may or may not
include contact tracing of patients exposed to or potentially exposed to an endoscope with a
positive result, or endoscopes processed through automated reprocessor with positive result.
40
If required, withdraw the endoscope/endoscope processor from use until tests prove the unit to
be biologically safe for use as determined by your working party. Review the need to import
temporary equipment, or postpone endoscopy lists until issue is resolved.
Write up report of the incident, actions taken and any outcomes. Keep Regional Office of
DHS, other organisations and parties involved in this issue informed of progress and outcomes.
41
(Insert Name of Organisation) Endoscopy Service
MEMORANDUM OF UNDERSTANDING
This memorandum of understanding outlines the responsibilities of the nominated parties
for the cleaning, processing and biological monitoring of the rotating loan endoscopes
used within this organization.
Responsibilities
Endoscope Owner
The endoscope owner will ensure that the endoscope is cleaned, processed,
dried, handled, stored and biologically monitored in accordance with
recommended standards and guidelines. The owner will ensure that
automated units, if used to process the endoscope, are biologically,
chemically and physically monitored in accordance with manufacturers
instructions and with recommended standards and guidelines.
Endoscope User
The organization using the endoscopes ensure that the endoscope is cleaned,
processed, dried and handled in accordance with recommended standards
and guidelines. The automated units, if used to process the endoscopes, are
biologically, chemically and physically monitored in accordance with
manufacturers instructions and with recommended standards and guidelines.
42
Biological Monitoring
……………………………………………………………………………
43
MEMORANDUM OF UNDERSTANDING
Agreement
I, …………………………………………………………………………………
(Owner of endoscope e.g. surgeon, or representative of organization if endoscope
owned by that organization e.g. NUM theatre)
On behalf of,
…………………………………………………………………………………
(myself, organization)
agrees to
…………………………………………………………………………………
(myself, organization)
being the nominated party responsible for undertaking the biological monitoring
of the endoscope as outlined on page 2.
Signed………………………………………………………………………………..
………………………………………………………………………………………
Date………./……../………..
Signed………………………………………………………………………………..
………………………………………………………………………………………
Date………./……../………..
44
ENDOSCOPE RECORD
MICROBIOLOGICAL TESTING HISTORY
To be kept with endoscope
EXAMPLE ONLY
ENDOSCOPE ID NUMBER 657854321-Gastroscope-pentax
AND TYPE
ROUTINE MICROBIOLOGICAL MONITORING MONTHLY-if endoscope is processed at any site in an Automated Endoscope Reprocessor
4 MONTHLY-if processed by sterilisation or manual disinfection process
Name of party responsible Date this sample Sample taken by Testing Agent Pathology Endoscope Signed Additional
for testing of endoscope taken Used Department biological comments
(as identified on MoU) processing sample test result
and date
sample
taken
45
ENDOSCOPE RECORD
PROCESSING HISTORY
To be kept with endoscope
EXAMPLE ONLY
ENDOSCOPE ID NUMBER 657854321- Gastroscope-pentax
AND TYPE
Name of Hospital or Date Endoscope Type of Chemical Agent Processing Unit’s last Signed Additional comments
Endoscopy Unit used Processing Unit Used known biological test
and ID Number result and date test
taken
46
Hume Region Infection Control
Resource and Consulting Service
Endoscope Questionnaire
Please print clearly your responses
Name of Organisation
………………………………………………………………………………
……………………………………………………………………………….
……………………………………………………………………………….
……………………………………………………………………………….
Endoscopes
(please circle correct answer)
Permanent loan
(scopes remain on the hospital premises) Yes No
Rotating loan
(scopes arrive and leave with endoscopy service Yes No
provider at each visit)
Interim loan
(short term - if own scopes are unavailable for period Yes No
of time i.e. for repair)
In the event of the need for interim loan of scopes (as part of your
contingency plan) are they/or would they be provided by?
(please tick)
47
Hume Region Infection Control
Resource and Consulting Service
Endoscope Questionnaire
Gastroscopes ………………
Colonoscopes ………………
Bronchoscopes ………………
Loaned Scopes
Gastroscopes ………………
Colonoscopes ………………
Bronchoscopes ………………
What guidelines and standards do you use to formulate your hospital’s policy and
procedure documents regarding the cleaning and processing of owned and loan
endoscopes?
48
Hume Region Infection Control
Resource and Consulting Service
Endoscope Questionnaire
Processing-Hospital Owned Scopes
How do you currently process own endoscopes?
(use the information listed below to complete the Table 1)
Processing Unit
Medivator 1
Soluscope 2
Steris 3
Vented fume cabinet 4
Other unit (specify) 5 …………………………………………………..,
Disinfectant/sterilant
Glutaraldehyde A
OPA B
Paracetic acid C
Other (specify) D ……………………………………………………
(*example only)
Scope type Processing Unit Disinfectant/sterilant Biological
Monitoring
Gastroscopes 2 A 1
*This indicates that the gastroscopes are processed using soluscope with glutaraldehyde
and the scopes are tested monthly
Please list appropriate letter and number beside each listed scope type (refer example given above)
Table 1
Scope type Processing Unit Disinfectant/sterilant Biological
Monitoring
Gastroscopes
Colonoscopes
Bronchoscopes
Other (specify)
Does your hospital have a documented procedure in place that identifies how often
biological monitoring of your own scopes is to occur?
Yes No
49
Hume Region Infection Control
Resource and Consulting Service
Endoscope Questionnaire
Processing-Loan Scopes
Table 2
Scope type Processing Unit Disinfectant/sterilant Biological
Monitoring
Gastroscopes
Colonoscopes
Bronchoscopes
Other (specify)
Does your hospital have a documented procedure in place that identifies how often
biological monitoring of loaned endoscopes is to occur?
Yes No
Does the hospital undertake responsibility for biological monitoring of loan scopes?
Yes No
Does the provider of endoscopic service (brings own scopes) take responsibility for
biological monitoring of endoscopes?
Yes No
Yes No
If agreement is not yet formalised, are you in the process of formalising the
agreement?
Yes No
50
Hume Region Infection Control
Resource and Consulting Service
Endoscope Questionnaire
How often is the equipment used to process both own and loaned
scopes biologically monitored?
Medivator
Soluscope
Steris
Other (specify) ………………………………….
………………………………………………………………………………………………
………………………………………………………………………………………………
………………………………………………………………………………………………
………………………………………………………………………………………………
………………………………………………………………………………………………
………………………………………………………………………………………………
………………………………………………………………………………………………
51