Base TS-TN: Gaseous Sulfur-Nitrogen Compounds Analyzer
Base TS-TN: Gaseous Sulfur-Nitrogen Compounds Analyzer
GASEOUS SULFUR-NITROGEN
COMPOUNDS ANALYZER
Rev. 2.0
BAGGI BASE® offer solutions for measurement and analysis in gases, solids, liquids, steam and multiphase:
Temperature
Humidity
Velocity – Flow
Pressure
Level – Interface
Components Analysis
Sampling and Filtration
Data Acquisition
Data Communication
Vision systems
B.A.G.G.I . SrL – V.le Campania,29 20133 Milano Italy – Tel. +39 02715547 – Fax +39 027490571 – baggi@baggi.com
1 Introduction
The BAGGI® BASE instrument BASE TS-TN offers accurate measurements of sulphur compounds
(typically H2S, SO2 , CS2 and COS) concentration in a mixture of gases.
In this application, the analyzer is also able to calculate the Air Demand output signal, for driving the
combustion air to the furnace.
It is suitable for outdoors installation (ATEX zone 2).
Maintenance is reduced to a minimum due to:
Very long life UV lamp
Direct mount on the process pipe
Ruggedized components
Controlled sulfur condensation
Cell clean-up with instrument air
No moving parts
Self-monitoring
Calibration facility
Remote control
B.A.G.G.I . SrL – V.le Campania,29 20133 Milano Italy – Tel. +39 02715547 – Fax +39 027490571 – baggi@baggi.com
2 Architecture
Embedded computer
The implementation of the BASE TS-TN analyzer follows the general philosophy of the BASE
Instruments Series.
The raw input data from the sensors (UV spectrophotometer) are processed by algorithms provided by
BAGGI, running in an embedded computer that is the heart of the system.
The computer, together with the sensors and the power supply, is within an enclosure provided with a
protective purge system and a Vortex cooler (connected to the plant instrument air system).
The following figure shows the computer’s monitor with the functional keys:
Spectrophotomer
B.A.G.G.I . SrL – V.le Campania,29 20133 Milano Italy – Tel. +39 02715547 – Fax +39 027490571 – baggi@baggi.com
Pyrolisis Furnace
The measurement of the total sulphur content in the stream can be achieved by means of a pyrolysis
furnace placed before the analytical point.
Using air as a carrier we get an oxidation of the sulphur compounds in the stream:
Otherwise, using H2 as a carrier, we get a reduction where the sulphur compounds in the sample are
converted to H2S:
We have the capability to measure the H2S and SO2 concentrations and therefore work out the total
sulphur content in the stream.
Otherwise it is possible to choose the measurement of H2S, SO2, COS and CS2 concentrations
separately in the stream.
B.A.G.G.I . SrL – V.le Campania,29 20133 Milano Italy – Tel. +39 02715547 – Fax +39 027490571 – baggi@baggi.com
3 Principle of Operation
As anticipated, the measurement is based upon the Beer-Lambert law.
This one is an empirical relationship that relates the absorption of light (or UV waves) to the properties of
the material through which the light (UV) is travelling.
The measurement is centered on the wave length where the investigated material has maximum energy
absorption.
The sample, introduced in the pyrolysis tube, is oxidized with an excess of O2. As a result, the Sulphur
compounds are converted into SO2.
The same happens when the H2 carrier gas acts as a reducing (instead of oxidizing) agent.
The reaction products pass into the optical cell. Here two options are possible, according to the
instrument model: the SO2 concentration is calculated either by the absorbance spectrum in the UV band
(Beer-Lambert law) or by molecular fluorescence caused when excited by Ultraviolet light.
The analyzer establishes the intensity of the signal transmitted by the lamp of the spectrophotometer and
measures the intensity of the signal received by a photodiode array.
The signal is transmitted at wavelenghts where the absorbance of the measured substance is maximal.
Then the application software calculates the concentration according to the measured values and the
above formulas. A multi-compounds analysis is possible, because each compound has its unique
absorbance spectrum.
4 Installation
This drawing illustrates a typical cabinet front end configuration.
B.A.G.G.I . SrL – V.le Campania,29 20133 Milano Italy – Tel. +39 02715547 – Fax +39 027490571 – baggi@baggi.com
5 Technical Specifications
Instrument Specification
Power:
- Standard: 90-264 VAC, 47-63 Hz; 6A max
Environment:
- 0° to 40°C (32° to 104°F)
- 0° to 55°C (32° to 131°F), with vortex cooler
Dimensions (without sample system):
- Skid-mount 500mm H x 400mm W x 250mm D
(19,68” H x 15,74” W x 9,84 D)
Mounting:
- wall (standard)
Approximate Weight (without sample system):
- 15 Kg
Detector:
- PDA Spectrophotometer
Modem:
- As option: dual channel system with IR cell to be - Field-configurable; 300 to 33.6k baud rate (optional)
integrated with UV main channel. Ethernet Card :
Analogue Inputs: - Two 10/100 mbps with RJ-45 port
- Four inputs filtered with transient protection, user WIFI Card :
- scalable and assignable (optional) - One Integrated WIFI card 11Mbit/s
Analogue Outputs: Cooling system :
- Three isolated analogue outputs, 4 – 20 mA (standard); - Passive Heatsink (Fanless)
- Three additional isolated analogue outputs (optional) Operative System :
Digital Inputs: - Microsoft Windows XP Embedded
- Six digital inputs; user assignable (optional)
Digital Outputs: Software
- Four isolated relay output signal: Fault, operation, Compound detection :
Calibration, lamp replacement - Real-Time concentrations measuring
- Four additional relay output (optional) Ranges :
Enclosure Protection: - Total Sulphur: 0 – 500 ppmv
- IP66 - Others on request
Compliances: Response time :
- EN61326, EN 61010-1, EN60079-2 - 90% of final value in 10 sec. (typical)
- ATEX, IECEx ( optional) Accuracy :
II 2 G Ex px II T6 - ±1% F.S.
II 3 G Ex pz II T6
Repeatability :
- ±0.5% F.S.
Integrated Computer Specification Sensitivity :
Display: - ±0.15% F.S.
- 7” touchscreen 262K colour display Calibrations Events :
- Resolution:800*480 - Automatic For 0
Storage: - Manual for Span
- 1GB CFII Gating Options:
Serial Communication Ports: - Fixed-Time, Slope and Automatic gating of peaks
One serial ports with RS-232 Password protection :
(optional) One serial ports with RS-232, RS-422 and RS - 2 Levels: User, Maintenance
485 standard is available with: Modbus, Profibus and
FieldbusFoundation Protocol
Parallel Printer Port:
- One parallel port available for printed For all trademarks refer to the owners.
reports(optional) The information contained herein are subject to change
without notice by B.A.G.G.I.Srl
KPP170223AR0MG_BASE TS-TN page 6
B.A.G.G.I . SrL – V.le Campania,29 20133 Milano Italy – Tel. +39 02715547 – Fax +39 027490571 – baggi@baggi.com