Calibration Standards For Spectrophotometers PDF

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 26

Hellma. Where precision becomes an art.

Calibration standards
for spectrophotometers

©2008 by Hellma www.hellma-worldwide.com


Contents

Calibration of spectrophotometers 3

Traceability of the calibration and conformity to the regulations 4

Calibration standards for testing the wavelength accuracy 5


Holmium oxide liquid filter 6
Holmium oxide glass filter 7
Didymium glass filter 8

Calibration standards for testing the photometric accuracy 9


Potassium dichromate liquid filter 10
Potassium dichromate liquid filter (430 nm) 11
Neutral density glass filter 12
Didymium glass filter 13

Calibration standards for testing stray light 14


Potassium chloride liquid filter, sodium iodide liquid filter, sodium nitrite liquid filter 15

Calibration standard for testing the resolution 16


Toluene liquid filter 17

Calibration standards for microplate readers 18

Complete set liquid filter, solid filters 19

Product overview liquid filters, solid filters 22

FAQ Frequently asked questions 24

Handling 25

2
Calibration
of spectrophotometers

UV/Vis-spectrophotometry is one of the most commonly (GMP) or standard operating procedures in the pharma-
used techniques of instrumental analysis. For example, ceutical industry as required by the most important pharma-

Wavelength accuracy
spectrophotometry is used in clinical chemistry, phar- copeias (EP, DAB, USP), require a regular performance
maceutical industry, research and very often in quality testing of the UV/Vis spectrophotometers used. These checks
assurance. include the testing of the resolution, the wavelength
Regulations for quality control such as ISO 9000, Good accuracy as well as the testing of stray light and photo-
Laboratory Practice (GLP), Good Manufacturing Practice metric accuracy.

Liquid filter Solid filter

Photometric accuracy
With the Hellma liquid filters presented here for testing With the solid filters for testing spectrophotometers,
spectrophotometers it is possible to check the following in the following parameters can be checked:
the wavelength region of 198 - 650 nm:
Photometric accuracy
Wavelength accuracy in the visible spectral region
Photometric accuracy Wavelength accuracy
Stray light behaviour

Stray light
Resolution

The Hellma calibration standards and their


compliance with the most important regulations

Resolution
Material Test Region Filter type EP DAB USP ASTM Page

Holmium oxide (solution) Wavelength UV/Vis Liquid filter X X X X 6

Holmium oxide glass Wavelength UV/Vis Solid filter X X 7

Didymium glass Wavelength UV/Vis Solid filter X X 8


Complete sets

Didymium glass Photometric UV Solid filter 13


Accuracy

Potassium dichromate Photometric UV Liquid filter X X X X 10


(solution) Accuracy

Neutral density glass Photometric Vis Solid filter X X 12


Accuracy
Product overview

Toluene in Hexane (solution) Resolution UV Liquid filter X X 17

Potassium chloride (solution) Stray light UV Liquid filter X X X 15

Sodium iodide (solution) Stray light UV Liquid filter X 15

Sodium nitrite (solution) Stray light UV Liquid filter X 15

3
Traceability of the calibration
and conformity to the regulations

The creation of the Hellma Liquid filters are filled un- Hellma calibration standards
Calibration Standards is der controlled conditions are calibrated using a high-
carried out in accordance and permanently sealed. performance UV/Vis/NIR
with the regulations of the Solid filters are carefully spectrophotometer. This in-
NIST (National Institute of cleaned and built into the strument is used exclusively
Standards and Technology, holder in a stress-free for calibrating purposes and
USA), ASTM (American manner. The manufacturing is tested at regular intervals
Society for Testing and process is completed with for its accuracy.
Materials) and the Euro- the individual certification
pean Pharmacopeia. of the filters.

For testing our spectrophotomter, use is made, amongst


others, of primary calibration standards of the NIST to
which the measuring results can be referenced:
Photometric accuracy
SRM 930e and SRM 1930 neutral density glass filter
Wavelength accuracy
SRM 2034 Holmium oxide solution

The wavelength accuracy of the instrument is maintained


by the utilisation of the emission lines of deuterium and
mercury lamps. The linearity of the instrument is ensured
by the double aperture method.

4
Calibration standards for testing
the wavelength accuracy

Ideally, a standard for de- It should be noted that with In principle, the calibration At larger slit widths it is pos-
termining the wavelength all standards for testing standards for testing the sible that weaker peaks can-

Wavelength accuracy
accuracy makes use of the wavelength accuracy a wavelength accuracy can not be detected anymore.
small well-defined peaks of change of the slit width of also be measured for slit
several wavelengths in the the photometer can lead to widths that deviate from the
UV and visible spectra. a slight displacement of certificate.
the absorption maximum.
In comparison to holmium According to European Pharmacopeia checking
oxide-liquid filters, the hol- Because of the small band- of the wavelength accuracy is described as follows:
mium oxide glass filter has width of the peaks the height “Verify the wavelength scale using the absorption

Photometric accuracy
a somewhat weaker defined of the measured maximum maxima of holmium perchlorate solution R [...]. The
spectrum. Especially in the varies strongly with a change permitted tolerance is ±1 nm for the ultraviolet range
deep UV region, the holmi- of the slit width. Therefore and ±3 nm for the visible range”.
um peaks are superimposed standards for testing the
by the absorption edge of wavelength accuracy can European Pharmacopeia specifies four wavelengths
the host glasses. A similar generally not be used for for holmium perchlorate solution filters: 241.15 nm,
effect can be observed with testing the photometric 287.15 nm, 361.5 nm and 536.3 nm. The spectrum
the Didymium glass filter. accuracy. contains some additional peaks which can also be used
for checking the wavelength in principle.

Stray light
Resolution
Complete sets
Product overview

5
Wavelength accuracy

Holmium oxide liquid filter


667-UV5 holmium oxide dissolved in perchloric acid

The 667-UV5 filter holmium spectrum with many cha-


oxide dissolved in perchloric racteristic peaks. In their
acid is well suited for testing position and height the
the wavelength accuracy measured maxima depend
of photometers both in the on the spectral bandwidth
UV- and Vis-range. There and the selected slit width
holmium oxide shows a and vary accordingly.

536.3
1
451.29

241.15 640.68

416.30
287.15 361.5
278.12
485.25
0.5 345.41
Absorption

333.49

0
230 330 430 530 630
Wavelength in nm

Typical spectrum of holmium oxide dissolved in perchloric acid, measured with a slit width of 1 nm.

The filter with the holmi- at 241.15 nm, 287.15 nm, placed in the reference
um perchlorate solution is 361.5 nm and 536.3 nm is beam in order to better
placed in the measuring determined and compared suppress the background.
beam. The measurement with the values in the certifi- Ideally measurements
is carried out with a small cate. Deviations should not should be carried out with
slit width (e.g. 1 nm). be greater than ± 1 nm in the slit width shown in the
Optimally the average of the ultra-violet and ± 3 nm certificate. If this cannot
several measurements is in the visible range. If avai- be done, then at least the
calculated in order to avoid lable the reference filter smallest possible slit
errors in measurements 667-UV14 liquid filter with width should be used for
The exact position of peaks perchloric acid can be measuring.

Catalogue number 667-UV5

Description Holmium perchlorate filter, liquid filter certified

Use Testing the wavelength accuracy in the UV and visible range


between 240 nm and 650 nm for spectrophotometers with a spectral bandwidth (slit)
between 0.1 nm and 3 nm

Material Solution of holmium oxide in perchloric acid


Filled into Hellma precision SUPRASIL® quartz glass cuvettes

Theoretical 241.15 nm
Wavelength 287.15 nm
according to 361.5 nm
Pharm. Eur. 536.3 nm

6
Holmium oxide glass filter
666-F1 Holmium oxide doped glass

The light yellow filter mar-


ked F1 is a glass doped with

Wavelength accuracy
holmium oxide (Ho2O3). The
chemical element holmium
shows a series of small ab-
sorption peaks in UV and Vis
range.

Photometric accuracy
2

279.3
1 453.4
360.8
Absorption

536.4
637.5

Stray light
0
230 330 430 530 630
Wavelength in nm

Typical spectrum of a holmium oxide glass filter

Resolution
Complete sets

Catalogue number 666-F1

Description Holmium oxide glass filter, solid filter certified

Use Testing the wavelength accuracy in the UV and Vis spectrum (279 nm to 638 nm)
at spectral bandwiths up to 2 nm
Product overview

Material Holmium oxide doped glass

Filter thickness 2.1 mm

Wavelength at 279.3 nm
360.8 nm
453.4 nm
536.4 nm
637.5 nm

7
Wavelength accuracy

Didymium glass filter


666-F7 Neodymium oxide and Praseodymium oxide doped glass

The Didymium glass filter


marked with F7 made of
Schott BG 36 material can
be used for testing the wave-
length accuracy in the UV
and Vis spectrum.

2 512
875
Absorption

329 472

1 681

0
270 370 470 570 670 770 870
Wavelenght in nm

Typical spectrum of a Didymium glass filter

Didymium glass shows a In addition the use of the


spectrum with a number of filter as absorption standard
characteristic peaks. The is possible at wavelengths
measured maxima depend of 270 nm, 280 nm, 300 nm,
on their position and height 320 nm and 340 nm.
of the spectral width and the
selected slit width and vary
correspondingly.

Catalogue number 666-F7

Description Didymium glass filter, solid filter certified

Use Testing the wavelength accuracy in the UV and Vis region (329 nm to 875 nm)
at a spectral bandwidth to 2 nm

Material Glass BG 36 of the SCHOTT Glaswerke

Filter thickness Approx. 1.2 mm

Wavelength at 329 nm
472 nm
512 nm
681 nm
875 nm

8
Photometric accuracy

Calibration standards for testing


the photometric accuracy

Various standards are Neutral density glass filters


available for determining are suitable for the visible

Wavelength accuracy
the photometric accuracy: spectrum; in the UV region,
the testing with Potassium-
dichromate is a proven
method. Typical standards
have broad peaks and broad
valleys.

Photometric accuracy
Stray light
Resolution
Complete sets
Product overview

9
Photometric accuracy

Potassium dichromate liquid filter


667-UV60 0.006 % Potassium dichromate dissolved in perchloric acid

667-UV60 Potassium dichro- 313 nm. The potassium di- With the measuring values
mate dissolved in perchloric chromate solution is kept in care must be taken that the
acid is suitable for testing an airtight precision cuvette measurements are carried
the photometric accuracy with a 10 mm light path. out with filter 667-UV14
(absorbance) of spectro- The absorption values of (0.001 N perchloric acid) in
photometers in the UV spec- the cuvette are measured the reference beam. The
trum. Here, potassium di- at the wavelengths 235 nm, measuring values of the
chromate shows a spectrum 257 nm, 313 nm and 350 nm reference filter are shown
with characteristic peak and recorded in the certi- separately in the certificate.
maxima at 257 nm and 350 ficate.
nm, minima at 235 nm and

The individually measured 257


0.9
absorption values are free
of systematic errors when
preparing the solution and 235 350
in the light path. As the
minimum and maximum in 0.5
the absorption spectra are
Absorption

relatively broad they can


be measured with a corre- 313
spondingly large slit width,
e.g. 2 nm. In this way, the
0
influence of noise on the
200 300 400 500
measuring-values can be
kept small. Wavelength in nm

Typical spectrum of a 0.006 % potassium dichromate solution

Catalogue number 667-UV60


Description Potassium dichromate filter, liquid filter certified
Use Testing the absorption behaviour in the UV spectrum (235 nm to 350 nm)
at a spectral bandwidth of 2 nm or less
Material 0.006 % Potassium dichromate dissolved in perchloric acid
Filled into Hellma precision SUPRASIL® quartz glass cuvettes
Wavelength 235 nm, 257 nm, 313 nm, 350 nm

10
Potassium dichromate liquid filter (430 nm)
667-UV600 0.06% Potassium dichromate dissolved in perchloric acid

The European Pharmacopeia mate in perchloric acid. filter 667-UV14 (0.001 N


requires, besides the testing The absorption value of this perchloric acid) in the refe-

Wavelength accuracy
of the photometric accuracy solution at 430 nm is deter- rence beam. The measuring
in the UV spectrum, also a mined and recorded in the value of the reference filter
testing at a wavelength in certificate. For this measu- is shown separately in the
the VIS spectrum (430 nm) ring value, care must also certificate.
with the aid of a 0.06 % so- be taken that the measure-
lution of Potassium dichro- ment is carried out with

Photometric accuracy
3

2
Absorption

430 nm

Stray light
1

0
380 420 460 500
Wavelength in nm

Typical spectrum of a 0.06 % potassium dichromate solution

Resolution
Complete sets

Catalogue number 667-UV600


Product overview

Description Potassium dichromate filter, liquid filter certified

Use Testing the absorption behaviour in the Vis spectrum (measuring wavelength 430 nm)
at a spectral bandwidth of 2 nm or less

Material 0.06 % Potassium dichromate dissolved in perchloric acid


filled into Hellma precision SUPRASIL® quartz glass cuvettes.

Wavelength 430 nm

11
Photometric accuracy

Neutral density glass filter


666-F2, 666-F3, 666-F4

For decades SCHOTT


neutral density glass filters
have been used in the Vis
spectrum for testing the
photometric accuracy and
linearity.

They provide a relatively The thickness of the filters


constant transmission over is determined such that
the visible wavelength spec- there is a stepping of the
trum and are calibrated at absorption values from
the following wavelengths: approx. 0.25 A, 0.5 A and 1 A.
440.0 nm, 465.0 nm,
546.1 nm, 590.0 nm, Also other thicknesses are
635.0 nm available on request.

If desired, other wave-


lengths are also possible.

1.5

1 666-F4
Absorption

0.5 666-F3
666-F2

0
400 450 500 550 600 650
Wavelength in nm

Typical spectrum of the neutral density glass filter, taken with a slit width of 1 nm

Catalogue number 666-F2


666-F3
666-F4

Description Neutral density glass filter, solid filter certified

Use Testing the absorption behaviour in the visible spectrum (440 nm to 635 nm)

Material Filter materials NG 11, NG 5, NG 4 from SCHOTT Glaswerke

12
Didymium glass filter
666-F7

The Didymium glass filter It can also be used for


marked F7 is usually a testing the photometric

Wavelength accuracy
standard for testing the accuracy.
wavelength accuracy in the
UV/Vis spectrum.

2.5

Photometric accuracy
2
ca. 1.5 A
Absorption

ca. 1.3 A ca. 1.2 A


1.5

ca. 0.75 A
1

0.5
ca. 0.5 A
0

Stray light
270 280 290 300 310 320 330 340 350 360 370
Wavelength in nm
Wavelength
Typical spectrum of a Didymium glass filter between in and
270 nm nm 370 nm

The testing of the absorp- such that at 340 nm there


tion behaviour in the UV is an absorption of appro-
spectrum is possible for ximately 0.5 A. With reduc-
wavelengths of 270 nm, tions in the wavelength the

Resolution
280 nm, 300 nm, 320 nm absorption values increase
and 340 nm. As standard up to approx. 1.5 A at
the thickness of the filter 270 nm.
is set during production

Complete sets

Catalogue number 666-F7

Description Didymium glass filter, solid filter certified

Use Testing the photometric accuracy in the UV spectrum (270 to 340 nm)

Material Glass BG 36 from SCHOTT Glaswerke

Designation Didymium glass filter, certified


Product overview

Filter thickness Approx. 1.2 mm

Nominal value Depending on wavelength


of the absorption

Absorption value at 270 nm


280 nm
300 nm
320 nm
340 nm

13
Stray light

Calibration standards for testing


stray light

Stray light (false light) is Stray light is problematic as If it is desired to test the
light that is detected but it reduces the region of the instrument for stray light,
does not belong to the measurable absorption and then filters are required that
selected measuring wave- degrades the linear relation- have a very sharply limited
length. Causes of this defect ship of concentration and spectrum (so called cut-off
are light scattering, diffrac- absorption. filters). Below a fixed wave-
tion or a poor adjustment of length (cut-off wavelength)
the measuring instrument. Stray light can become a Hellma stray light-filters
problem at any wavelength do not transmit. The trans-
but the smaller the ener- mission values in this wave-
gy throughput, e.g. when length region are thus due
measuring in the UV range, to stray light.
the more the stray light in-
fluences the measurement
value.

In the European Pharmacopeia the testing For example, the absorption of a solution of potassium
for stray light is described as follows: chloride R (12 g/l) between 220 nm and 200 nm in a light
"The stray light can be measured at a defined wavelength path of 1 cm must rise steeply and at a wavelength of
with suitable filters or solutions: 198 nm must be greater than 2, measured against water
R as a compensation liquid."

In order to realistically estimate the amount of stray light,


a filter must be selected whose cut-off wavelength is as near
as possible above the required wavelength. The following
table shows the offered solutions for stray light measure-
ments for the cut-off wavelengths 200 nm, 259 nm and
385 nm or below.

Catalogue Content Cut-Off Certified


Number Wavelength Wavelengths

667.100-UV 667-UV1 Aqueous potassium chloride solution 200 nm 198 nm, 200 nm

667-UV12 Purified water (reference)

667.101-UV 667-UV10 Aqueous sodium iodide solution 259 nm 220 nm

667-UV12 Purified water (reference)

667.102-UV 667-UV11 Aqueous sodium nitrite solution 385 nm 340 nm, 370 nm

667-UV12 Purified water (reference)

667-UV12 667-UV12 Purified water filter 198 nm, 200 nm,


300 nm, 400 nm

14
Potassium chloride filter 667-UV1
Sodium iodide filter 667-UV10
Sodium nitrite filter 667-UV11

Wavelength accuracy
3

2,5
NaI
2 (667-UV10)

Absorption

Photometric accuracy
1,5 KCl NaNO2
(667-UV1) (667-UV11)
1
H2O
0,5 (667-UV12)

0
200 250 300 350 400 450 500

Wavelength
Wellenlängeininnm
nm

Stray light
These calibration standards Application Testing the maximum pecially when the measured
show a very high absorption The photometer is set to range of the photometer values are smaller than 0.02
below defined wavelengths a starting wavelength that The filter 667-UV12 is filled A, contact should be made
(cut-off-wavelengths) in the is 30 nm above the cut-off with purified water, whose with the customer service
UV spectrum. In this way wavelength of the filter to absorption is defined above of the instrument manufac-
the stray radiation can be be used. If, for example, 200 nm to the NIR practi- turer.
tested. The testing for stray potassium chloride is used cally only by the reflection
light is carried out with a then the start is at 230 losses at the two glass/air

Resolution
reference standard of puri- nm. Then a scan is carried transitions. With the certi-
fied water that is provided out down towards shorter fied values at 198 nm, 200
with every calibration filter wavelengths and the ab- nm, 300 nm and 400 nm it
set. The procedure is the sorption below the cut-off is possible to test the maxi-
same for all stray light wavelength is observed. mum range of the instru-
filters. The transmission that is ment for very low absorption
measured below the cut-off values. For clear deviations
wavelength is stray light. from the certified values, es-
Complete sets

Catalogue number 667-UV1


667-UV10
667-UV11
Product overview

Description Liquid filter with sharp cut-offs of transmission


at certain wavelength, certified

Use Testing for stray light in the UV spectrum (at wavelengths


from 198 nm to 385 nm, depending on the selected filter)

Material Aqueous solutions in Hellma precision SUPRASIL® quartz glass cuvettes

Reference filter 667-UV12 purified water filter

15
Resolution

Calibration standards for testing


the resolution

Spectral resolution is a The resolution for modern For high resolution photo-
measure of the ability of photometers depends di- meters, the spectral band-
an instrument to differen- rectly on the slit width. The width should not be more
tiate between two adjacent smaller the slit and the as- than 10 % of the natural
wavelengths. Two peaks sociated spectral bandwidth, bandwidth of the peak to be
usually are considered the higher the resolution. investigated.
resolved if the minimum
of the absorption between For wavelengths and ab- The regular testing of the
the two peaks is lower than sorption measurements, photometer for resolution
80% of the peak maxima. therefore, the resolution of ensures that, for instance,
the monochromator must be neighbouring peaks can
taken into account. However, be resolved and do not
when a small spectral band- merge with the peaks of the
width leads to a reduction neighbouring wavelength.
in the detectable measured In addition the absorption
light, then the signal-noise errors will be prevented.
relationship reduces.

16
Toluene liquid filter
667-UV6 Toluene in n-Hexane

In order to determine the Measurements of the reso- peak maximum and peak
resolution of a spectro- lution are carried out using minimum are determined.

Wavelength accuracy
photometer, the behaviour hexane reference filter Then the relationship bet-
of the absorption of a maxi- 667-UV9 which is provided ween the two is calculated.
mum at λmax (269 nm) with every filter set. The
to absorption of a minimum execution of the measu-
at λmin (266 nm) is shown. rement is the same for all
From this, one obtains details bandwidths. The wavelength
of the spectral bandwidth spectrum from 265 nm to
of the instrument used. 270 nm is scanned and the

Photometric accuracy
0.5

0.4
slit 0.25
slit 0.5
Absorption

Stray light
slit 1.0
0.3
slit 2.0
slit 3.0
slit 4.0

0.2

0.1
265 266 267 268 269 270

Resolution
Slit Ratio
width (Recommended Wavelength in nm
in nm values)
Typical spectra of the Toluene liquid filters with varying slit widths
0.25 2.3
For determining the resolu- kedly downwards (e.g. 15 %) conditions. Thus, especially
0.5 2.2
tion, the spectra of the given then contact should be ma- for small slit widths, a
1.0 2.0 slit widths are measured, de with the manufacturer of sufficiently large integration
Complete sets

2.0 1.4 the ratios are calculated the instrument. However, it time must be selected.
and the values are com- must be remembered that
3.0 1.1
pared with the certificate. the results are also depen-
4.0 1.0 If the ratios deviate mar- dent on the measurement

Catalogue number 667.200-UV


Product overview

Description Toluene liquid filter, certified

Use Testing the resolution of spectral bandwidths

Material Solution of toluene in hexane, filled into Hellma precision cuvettes of SUPRASIL® quartz glass

Liquid filter set 667-UV6 Toluene in n-hexane


667-UV9 Hexane (reference standard)

Wavelength 667-UV6 266 nm, 269 nm

17
Calibration standards for microplate readers

666.013
Calibration standards for microplate readers

With the calibration standard For determining the wave-


666.013 the photometric length accuracy of micro-
accuracy of microplate rea- plate readers, standard cali-
ders can be tested. It has the bration plates are available
dimensions of a microplate, in which the neutral density
whose 96 wells, have dia- glass strips are replaced by
meters of 6.6 mm. holmium oxide glass strips
or other glass (on request).
Neutral density glass strips
are located behind each 16
windows having varying ab-
sorption values (see table).
Each calibration standard is
provided with a certificate of
the calibration.

Catalogue number 666.013

Designation Calibration standard for microplate readers NG 11, NG 5, NG 4, NG 3 certified

Nominal value of 0.25


the absorption 0.5
1.0
1.5
2.5

Absorption 405 nm
values at 450 nm
490 nm
650 nm

18
Complete sets

Solid filters
666.000 Calibration filter set

The set 666.000 consists of


neutral density glass filters

Wavelength accuracy
for testing the absorption
values and a holmium oxide
glass filter for testing the
wavelength accuracy. The
glass filters are mounted in
precision frames of alumini-
um. They are designed such
that they fit into the holders

Photometric accuracy
for standard cuvettes with a
10 mm light path with which
the spectrophotometer is
equipped. The set is deli-
vered together with an empty
filter frame as reference in
a sturdy box.

For purposes of identificati-


on, the set number is engra-
ved on each filter frame.

Stray light
The measured values of
the absorption and the wa-
velengths are shown for each
filter in the calibration certi-
ficate supplied. For purposes
of working in the laboratory,
the measurement values are
also shown on the inside of

Resolution
the lid of the box.

Complete sets

Catalogue number 666.000

Designation Solid filter set for testing


the photometric accuracy and the wavelength accuracy

Consisting of 666-F1 Holmium oxide glass filter


666-F2 Neutral density glass filter NG 11
Product overview

666-F3 Neutral density glass filter NG 5


666-F4 Neutral density glass filter NG 4
666-F0 Filter holder, empty as reference

Peakposition 666-F1 279.3 nm, 360.8 nm, 453.4 nm, 536.4 nm, 637.5 nm

Absorption value 666-F2 440 nm, 465 nm, 546.1 nm, 590 nm, 635 nm
certified at 666-F3 440 nm, 465 nm, 546.1 nm, 590 nm, 635 nm
666-F4 440 nm, 465 nm, 546.1 nm, 590 nm, 635 nm

All sets are also available individually! 19


Complete sets

Liquid filters
667.003-UV calibration filter set

The set 667.003-UV consists quartz cuvettes. The cuvet-


of a total of eight liquid fil- tes are permanently sealed.
ters and permits the com- The set is supplied in a
plete testing of the photo- sturdy box. For purposes
meters, as required of the of identification, a number
European Pharmacopeia. is engraved on each filter.
All liquid filters consist of The measured values are
calibration solutions that shown for each filter in the
are filled into Hellma calibration certificate.
SUPRASIL® precision

Catalogue number 667.003-UV

Designation Liquid filter set for testing photometers according to Ph. Eur.

Consisting of 667.100-UV Potassium chloride and reference filter


667.200-UV Toluene in hexane and reference filter
607.305-UV Potassium dichromate in perchloric acid and reference filter
667-UV5 Holmium oxide in perchloric acid

Wavelengths at 667.100-UV 198 nm, 200 nm (certified)


which measure- 667.200-UV 266 nm, 269 nm
ments are taken 607.305-UV 235 nm, 257 nm, 313 nm, 350 nm, 430 nm
667-UV5 241.15 nm, 287.15 nm, 361.5 nm, 536,3 nm

All filters are also available individually!

20
Liquid filters
Make up of the sets

Catalogue number Purpose Contents

Wavelength accuracy
667.003-UV 667.100-UV 667-UV1 Potassium chloride in purified water
Complete set Set for testing for stray light
for testing according to the European 667-UV12 Purified water reference filter
photometers Pharmacopeia
according to
the European 667.200-UV 667-UV6
Pharmacopeia Set for testing the resolution Toluene in n-hexane
according to the European
Pharmacopeia 667-UV9

Photometric accuracy
n-hexane reference filter

667.305-UV 667-UV60
Set for testing the photo- Potassium dichromate in perchloric acid
metric accuracy according to
the European Pharmacopeia 667-UV600
Potassium dichromate in perchloric acid (testing at 430 nm)

667-UV14
Perchloric acid reference filter

667-UV5 667-UV5

Stray light
Filter for testing the wavelength Holmium oxide in perchloric acid
accuracy according to the
European Pharamcopeia

667.101-UV 667-UV10 Sodium iodide in purified water


Set for testing
for stray light 667-UV12 Purified water reference filter

Resolution
667.102-UV 667-UV11 Sodium nitrite in purified water
Set for testing
for stray light 667-UV12 Purified water reference filter

Complete sets

667.307-UV
667.101-UV Set for 667-UV20 0.002 % Potassium dichromate in perchloric acid
Linearity filter
Testing for setlight
stray for
testing the absorption 667-UV40 0.004 % Potassium dichromate in perchloric acid
and linearity
667-UV60 0.006 % Potassium dichromate in perchloric acid

667-UV80 0.008 % Potassium dichromate in perchloric acid

667-UV100 0.01 % Potassium dichromate in perchloric acid


Product overview

667-UV14 Perchloric acid reference filter

667.400-UV 667-UV5 Holmium oxide in perchloric acid


Set for testing the
wavelength accuracy 667-UV14 Perchloric acid reference filter

All filters and filter sets are also available individually.


21
Product overview

Liquid filters
Overview

Catalogue Use consisting Content Wavelength


Number of
667.003-UV Complete set for testing 667.100-UV Potassium chloride and 198 nm; 200 nm
photometers according reference filter (certified)
to Ph. Eur.
667.200-UV Toluene in hexane and 266 nm; 269 nm
reference filter

667.305-UV Potassium dichromate in 235 nm; 257 nm; 313 nm;


HClO4 and reference filter 350 nm; 430 nm
667-UV5 Holmium oxide in 241.15 nm; 287.15 nm;
perchloric acid 361.5 nm; 536.3 nm

667.100-UV Set for testing 667-UV1 Potassium chloride 200 nm (cut-off)


to stray light according
to Ph. Eur. 667-UV12 Purified water 198 nm; 200 nm
(reference filter) (certified)
667.200-UV Set for testing the resolution 667-UV6 Toluene in hexane 266 nm; 269 nm
according to Ph. Eur.
667-UV9 Hexane (reference filter)
667.305-UV Set for testing 667-UV60 Potassium dichromate in 235 nm; 257 nm;
the photometric accuracy HClO4 313 nm; 350 nm
according to Ph. Eur.
667-UV600 Potassium dichromate 430 nm
in HClO4 (430 nm)
667-UV14 Perchloric acid (reference filter)

667.307-UV Filter set for testing 667-UV20 0.002 % Potassium 235 nm; 257 nm;
the absorption and linearity 667-UV40 0. 004 % dichromate in 313 nm; 350 nm
perchloric acid
667-UV60 0.006 %
667-UV80 0.008 %
667-UV100 0.01 %
667-UV14 Perchloric acid reference filter
667-UV5 Testing the wavelength 667-UV5 Holmium oxide in perchloric 241.15 nm; 287.15 nm;
accuracy according to Ph. Eur. acid 361.5 nm; 536.3 nm

667.101-UV Set for testing 667-UV10 Sodium iodide 259 nm (cut-off)


for stray light 220 nm (certified)
667-UV12 Purified water (reference filter)
667.102-UV Set for testing 667-UV11 Sodium nitrite 385 nm (cut-off)
for stray light 340 nm; 370 nm
667-UV12 Purified water (reference filter) (certified)

667.400-UV Set for testing the 667-UV5 Holmium oxide in perchloric acid 241.15 nm; 287.15 nm;
wavelength accuracy 361.5 nm; 536.3 nm
667-UV14 Perchloric acid (reference filter)

All filters and filter sets are also available individually.


22
Solid filters
Overview

Catalogue Use consisting Contents Wavelength


number of
666.000 Complete set for testing the 666-F1 Holmium oxide glass filter 271.3 nm; 360.8 nm;
photometric accuracy and the 453.4 nm; 536.4 nm;
wavelength accuracy 637.5 nm

666-F2 Neutral density glass filter 440 nm; 465 nm;


NG 11 546.1 nm; 590 nm;
635 nm

666-F3 Neutral density glass filter 440 nm; 465 nm;


NG 5 546.1 nm; 590 nm;
635 nm

666-F4 Neutral density glass filter 440 nm; 465 nm;


NG 4 546,1 nm; 590 nm;
635 nm

666-F0 Filter holder,


empty as reference

Individually available
666-F1 Testing the wavelength accuracy Holmium oxide glass filter 271.3 nm; 360.8 nm;
453.4 nm; 536.4 nm;
637. 5 nm

666-F2 Testing the photometric accuracy Neutral density glass filter 440 nm; 465 nm;
(Nominal value of the absorption 0.25 A) NG 11 546.1 nm; 590 nm;
635 nm

666-F3 Testing the photometric accuracy Neutral density glass filter 440 nm; 465 nm;
(Nominal value of the absorption 0.5 A) NG 5 546.1 nm; 590 nm;
635 nm

666-F4 Testing the photometric accuracy Neutral density glass filter 440 nm; 465 nm;
(Nominal value of the absorption 1 A) NG 4 546.1 nm; 590 nm;
635 nm

666-F7 Testing the photometric accuracy Didymium glass filter BG 36 270 nm; 280 nm;
300 nm; 320 nm;
340 nm

666-F7 Testing the wavelength accuracy Didymium glass filter BG 36 329 nm; 472 nm;
512 nm; 681 nm;
875 nm

666-F0 Filter holder, empty as reference

Within the technical possibilities and on request we can also deliver


23
other wavelength and absorption ranges as special products.
FAQ
Frequently asked question

What does ”traceability“ mean?


Measurement results have Accuracy of those standards USA (National Institute of to be calibrated against
to be internationally com- is assured by an unbroken Standards and Technology). a standard whose metro-
parable. This demand is chain of international com- ”Traceability“ describes logical quality was also
assured by the relation of parison measurements. the procedure of comparing certified against a standard
the measurements results Our calibration standards the display of a measuring of higher order, creating
to standard references, and calibration standard device with the original a calibration hierarchy
normally maintained by a measurement equipment reference source for the from the national working
national standard body. is traceable to international measurement in one or standards through to a
accepted standards of NIST, more steps. Each step has laboratory.

Traceability is characterised by some basic essentials:

An unbroken chain of comparisons which can Laboratories or others who accomplish one or
be traced back to stated national or international more steps in the measuring chain have to prove
references, approved by the parties involved. their technical competence.

Measurement uncertainty has to be calculated Calibrations should be repeated after reasonable
for each intermediate reference standard using periods of times. The length of this period depends
approved methods. It has to be stated in a way that on variables such as the required certainty, the
allows calculating the over-all uncertainty for frequency of use, the way of using it or the constancy
the measuring chain. of the device. .

All calibration steps have to be accomplished
following generally accepted and described methods,
results have to be documented.

The peak positions found filters do depend on their How often have the in the first two years of
with the calibration stand- thickness, and so they can calibration filters to be use. Afterwards an interval
ard is beyond tolerance – be available on customer recalibrated? of recalibration should be
what is the problem? request. chosen which seems to be
To check the correctness suitable based on the values
Does the setting of the slit Where do the greyish of the values given on the obtained this way.
match with the specifica- shadows on the holmium certificate, the calibra-
tions on the certificate? oxide filter come from? tion standards should be
Other slit settings can Do they affect the recalibrated at regular
cause errors in the peak measurement? intervals. The periodicity of
maxima. these intervals should be
The holmium oxide glass is determined by the user and
Do you supply neutral slightly hygroscopic, so the depends on the laboratory
density glass filters with covering is a kind of water environment and the condi-
absorbance values diffe- film. Measurements are tions of use, just like the
rent from those which are not affected. The filter may total lifetime of the filters.
specified in the Hellma be wiped with alcohol and To define a statistical basis
catalogue? (e.g. Abs = 0.8 a soft tissue to remove the for establishing the re-
instead of 1.0) film. As a basic principle certification interval it is
the filters should be stored recommended to have all
The absorbance values of in a dry place. calibration standards re-
the neutral density glass calibrated every 12 months

24
Handling

The calibration standards section 7.6 “Control of and corrosion on the opti- All filters can also be
are to be considered mea- Inspection and Measuring cal surfaces can easily in- ordered separately, e.g.

Wavelength accuracy
suring equipment in accor- Equipment“). The calibra- troduce substantial errors. as replacement parts, by
dance with international tion standards must be their catalogue number.
standards and should be treated with special care Extensive handling in- The filters are sequentially
recalibrated at regular if they are to retain their structions are provided numbered and come com-
intervals (see EN ISO 9001, validity. Scratches, dirt, with each set of filters. plete with a certificate.

Photometric accuracy
Usually customer choose a recalibration of the solid The periodicity of these intervals should be determined by
filters every 12 months for the first two years of use and the user and depends on the laboratory environment and
after that every 24 months, recalibration of the liquid the conditions of use. Filters or sets of filters which are
filters not later than every 12 months. sent to us to be recalibrated will be cleaned and calibrated.
They will then be returned to you with a new certificate.

Influence of Other factors influencing Preparing for height in the lower third.
temperature the measurement a measurement When using single-beam
photometers and especially

Stray light
The uncertainty of the certi- Dirt and dust as well as Because of the variety of diode array photometers
fied measurement results damage (scratches, ble- possible designs of spectro- connected to a separate
is only valid for measuring mishes) to the polished photometers the application cell holder via a fibre-optic
temperatures as stated on surfaces can considerably of a calibration standard cable, you should also con-
the certificate. It is recom- affect the measured values. for checking an instrument sider that any stray radia-
mended to store the filters Store the filters in the case can only be given in general tion or vibrations (moving
after use at room tempera- and avoid any contamination terms. the fibre-optic cables) may
ture in their storage case of the polished windows. affect the measurements.
in a dry, dust-free environ- There should always be a The spectrophotometer

Resolution
ment. recalibration associated is switched on about half Current versions of the
with the cleaning of the an hour before starting the safety data sheets for any
For reasons of safety and standards. measurements in order used dangerous goods are
especially with the cells to achieve a constant available to download on
containing liquids, take Safety instructions instrument temperature the internet:
care not to expose them to (please follow the manu- www.hellma-worldwide.com
temperatures below 0 °C The filters containing liquids facturers’ instructions).
or above 50 °C when storing carry a label on one side The measurement should
Complete sets

and during transportation. with the chemical formula be carried out in a room
of the substances contained with low humidity and the
inside. If a filter breaks, temperature which is given
please follow the corres- on the filters certificate.
ponding safety instructions
(e.g. safety data sheet). The filters fit into cell hol-
ders for standard cells
with 10 mm light path. They
Product overview

should always be placed


in the cell holder in the
same orientation, i.e. with
the Hellma logo pointing
towards the light source.
The centre of the light beam
should pass through the
filter always at the same

25
Hellma GmbH & Co. KG
P.O. Box 1163
79371 Müllheim / Germany
phone + 49 7631 182-0
fax + 49 7631 13546
info@hellma-worldwide.com
www.hellma-worldwide.com

145/08-GB-2 Printed in Germany Subject to technical changes


SUPRASIL® is a registered trademark of Heraeus Quarzglas GmbH & Co. KG.

You might also like