PET Performance Measurements Using The NEMA NU 2-2001 Standard
PET Performance Measurements Using The NEMA NU 2-2001 Standard
1Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania; 2CTI PET Systems, Knoxville, Tennessee;
3PickerInternational, Cleveland, Ohio; 4Philips Medical Systems, Milpitas, California; 5Philips Medical Systems, Philadelphia,
Pennsylvania; 6Siemens Medical Systems, Hoffman Estates, Illinois; 7General Electric Medical Systems, Milwaukee, Wisconsin;
8Department of Radiology, University of Washington, Seattle, Washington; and 9Department of Physics, University of British
where Chot and Cbkgd are the average of the counts measured in the
hot sphere ROI and the average of the counts in all background
ROIs, respectively, and ahot /abkgd is the ratio of the activities in the
hot sphere and background. The cold sphere CRC (CRCcold) is
calculated as:
where Clung is the average of the counts in the lung insert ROI.
Transverse and coronal images are also reported because of the
difficulty in devising appropriate quantitative measures to express
all aspects of image quality.
scanners with axial sampling inadequate to perform the the NU 2-1994 standard. On the ADAC UGM C-PET
axial resolution measurement. The axial profile width mea- scanner with a lower energy threshold of 450 keV, the
surement involved acquiring many scans at fine axial inter- weighted average of the 3 source positions in a 19-cm-long
vals. In addition, the scan data were not reconstructed, so phantom is 25% ⫾ 3%, whereas the result from a single
effects of reconstruction processing on axial spatial resolu- measurement at 4.5 cm for the same phantom is 30% ⫾ 3%.
tion could not be assessed. Because the measurement was It is expected that other scanners would show a similarly
time-consuming and did not include processing effects, it higher scatter fraction for the single measurement. The
was decided to measure the axial resolution on all scanners, difference (20%) accounts entirely for the increase in scatter
regardless of sampling. fraction seen for 2D acquisition mode on the GE Advance
scanner in Table 3.
Intrinsic Scatter Fraction
An important consideration in interpreting the results of
A major motivation for using a longer phantom in the NU
the intrinsic scatter fraction measurement is that the scatter
2-2001 standard was to measure the effects of OFA, includ-
fraction, as measured with either the 19-cm-long or 70-cm-
ing increased scatter. Table 3 shows the results for the
long cylinder, is not representative of a realistic head or
scatter fraction measurement following the NU 2-1994 and
body size. It is useful for standard evaluations among scan-
NU 2-2001 protocols on scanners with different axial FOVs.
ners of different configurations. However, in the effort to
It can be seen that the measured scatter fraction for a
use 1 phantom, the diameter of both phantoms (20 cm) is a
scanner operating in 2D mode is only slightly increased
compromise between a brain and a body. This makes it
compared with that measured following the NU 2-1994
difficult to predict the absolute level of scatter that will be
procedure. However, scanners that operate in 3D mode
encountered in patient imaging situations. However, the
show markedly higher (35%– 40%) scatter fractions with
scatter fraction measured with the 20-cm-diameter phantom
the new standard.
on a particular PET system will indicate the relative level of
There are 2 differences between the NU 2-1994 and NU
scatter measured in a patient study.
2-2001 protocols for intrinsic scatter fraction, the longer
phantom and the single measurement at a radial position of Sensitivity
4.5 cm. The scatter fraction measured at a radial distance of The total sensitivity values for several scanners are
4.5 cm is somewhat higher than the area-weighted average shown in Table 4 for radial positions of 0 and 10 cm.
of scatter fraction measured for a line source at radial Results are also given using the uniform cylinder, as spec-
distances r ⫽ 0, 4.5, and 9.0 cm, as specified previously by ified in the NU 2-1994 standard. Because the source con-
TABLE 4
Comparison of Volume Sensitivity from NU 2-1994 and NU 2-2001 Standards
Data were acquired on GE Advance scanner. As point of reference, typical average activity concentration seen in clinical FDG study is
3– 4 kBq/mL, for 370-MBq injection in 70-kg patient after 1-h uptake period.
figurations and definitions of sensitivity differ significantly highly sensitive system may saturate at a relatively low
between the 2 standards, it is difficult to compare the results activity level, but the counting rate at this level may be
directly. However, the relative ordering of sensitivity values higher than that of a system with lower sensitivity at a
among scanners should be the same for the 2 measurements, higher activity level.
although there will not be a single scaling factor between A concern with using the line source to measure the
the results for the 2 standards. As an example, the sensitivity counting rate performance was whether it yields counting
values following the NU 2-1994 standard for the ADAC rates (and effective activity concentrations) that are clini-
MCD system, with a 37-cm axial FOV, are significantly cally relevant. Figure 2 shows a plot of the prompt rate as a
lower than would be expected from the NU 2-2001 values function of singles rate for phantom and patient data. On the
because the axial FOV is much longer than the older phan- top are shown the counting rates for the ADAC UGM
tom (length, 19 cm). Also, there is a radial dependence of C-PET scanner for the 70-cm-long line source in a 20-cm-
the NU 2-2001 sensitivity values for the coincidence diameter polyethylene cylinder (following the NU 2-2001
gamma camera that is almost absent in the 2 circular sys- standard) and whole-body patient studies. On the bottom are
tems. shown the counting rates for the University of Pennsylvania
It should be pointed out that the sensitivity is expressed as GSO brain scanner (G-PET) for the 20-cm-diameter ⫻
the rate of coincidence events per activity in the entire line 19-cm long water-filled cylinder (following the NU 2-1994
source. A scanner with a shorter axial FOV will, therefore, standard) and brain patient studies. There is good correla-
have a lower value than one with a longer FOV that en- tion between the measured counting rates for the 70-cm line
compasses more of the line source. An alternate scaling source and whole-body studies. Such a correlation has not
would have been to record the coincidence counting rate per been observed for the 19-cm-long uniform cylinder (9).
activity inside the axial FOV. However, this scaling does Points lying below the phantom line correspond to patients
not reflect the fact that a scanner with a longer axial FOV whose attenuation is greater than that of a 20-cm-diameter
will require less time (have higher effective sensitivity) to cylinder. There is also good agreement between the mea-
scan a whole body than one with a smaller FOV, for the sured counting rates for the 20-cm cylinder and brain stud-
same count density. ies. These results indicate that the 70-cm line source in the
Counting Rate Performance 20-cm-diameter polyethylene cylinder is a reasonable ap-
Table 5 shows typical peak counting rates for a GE proximation to the clinical whole-body activity distribution
Advance scanner, operated in both 2D and 3D modes, for (albeit for thin patients), whereas the 20 ⫻ 19 cm long
the NU 2-1994 and NU 2-2001 standards. There is a dra- cylinder is a good representation of the clinical brain activ-
matic decrease in the peak true and NEC rates for 3D ity distribution.
acquisitions following the new standard compared with The measurement procedure specifies that the line source
those for the old standard. This is largely due to the increase be filled with sufficient activity to measure the peak true and
in randoms and dead time (decreased trues) as well as the peak NEC rates. The volume of the line source is approxi-
higher scatter fraction from the OFA associated with the mately 3 mL. For a 3D system operating without septa, the
longer phantom. There was inadequate activity in the line amount of activity required is 370 –550 MBq (⬃10 –15
source to achieve either the peak true or peak NEC rates in mCi). However, for 2D systems with septa, the amount of
2D mode on this scanner. activity required to reach the peak trues or peak NEC
It is important in comparing performance between scan- counting rates can be significantly higher (on the order of
ners to look both at the peak true and NEC counting rates 5.5–7.5 GBq [150 –200 mCi]); as has been seen, it can be
and at the activity levels where these peak rates occur. A difficult to achieve this activity level in the small volume of
activity remaining measured, with the line source activity Image Quality Measurement
being calculated from the difference of the 2 measured Figure 3 shows a representative image for a 6-min emis-
activities. The first has the advantage that the line source sion scan from the study with an 8:1 ratio of sphere-to-
activity is measured directly; however, manipulating the background activities taken on the ADAC UGM C-PET
line source in this manner can make it prone to leaks, and scanner. Also shown is the placement of ROIs for the
the response of the dose calibrator is sensitive to the geom- quantitative analysis.
etry of the coil inside the well. The second method is Because a single scan has limited statistics, due to the
straightforward, but care should be taken when filling the limit on acquisition time, the results can be expected to vary
line source to ensure that no activity is spilled. somewhat from scan to scan. Table 6 shows results for the
To compare the counting rate performance of various CRC, and Table 7 shows results for background variability
PET systems, it is necessary to specify the level of activity for 2 repeated scans taken on the ADAC UGM C-PET
to which the scanner is exposed. The total activity in the line scanner. Each emission scan was 6-min long and recon-
source is not a meaningful number if the FOV of the scanner structed with the 3D row action maximum likelihood algo-
is significantly less than the length of the source (70 cm), rithm (RAMLA) (12,13), using the clinical parameters. The
because not all of that activity is detected by the scanner. percentage difference in Tables 6 and 7 was calculated as
The activity inside the FOV could be an appropriate param- the absolute difference between the 2 scans, divided by the
eter but, for comparisons with other scanners, an activity average value. Both the background variability and the
concentration seemed more relevant. For that reason, the measured contrast of a given sphere vary between scans. For
effective activity concentration is used as the common de- this reason, the standard recommends that the measurement
nominator in this measurement. be repeated at least 3 times (with the durations of the
replicate scans increased to compensate for physical decay)
Accuracy of Corrections for Count Losses and the results averaged.
and Randoms The measured image quality will depend on the recon-
The results of measurements of the accuracy of correc- struction algorithm. Figure 4 shows the contrast versus
tions for count losses and randoms have been reported for 2 background variability for the 1.7-cm sphere for 2 recon-
different scanners (9,11). For both systems, a rather large struction algorithms (filtered backprojection and 3D
error (15%– 40%) was reported at or below activity levels RAMLA) over a range of reconstruction parameters. The
where the peak NEC rate is achieved. The use of a line
source for this measurement, rather than a distributed
source, may prove to be more challenging, from the point of TABLE 6
view of accurate corrections for random coincidences and Reproducibility of Image Quality Results: CRC
dead time. However, it should be noted that the activity Diameter %
levels where the NEC rate peaks may also be higher than (cm) Scan 1 Scan 2 Average Diff.
those seen clinically.
1.0 0.047 0.076 0.061 48
The large ROI centered on the phantom was chosen 1.3 0.16 0.19 0.17 18
rather than a smaller one, centered on the line source itself, 1.7 0.24 0.22 0.23 8.7
because random (and scattered) coincidences are globally 2.2 0.42 0.44 0.43 4.6
distributed throughout the FOV, not just in the region of 2.8 0.48 0.45 0.46 6.5
radioactivity. Without proper correction for randoms 3.7 0.48 0.47 0.48 2.1
throughout the image, image contrast will change as a
function of activity level. Using a larger ROI permits an Data were acquired on ADAC UGM C-PET scanner. Percentage
assessment of the global accuracy of the randoms subtrac- difference (% Diff.) was calculated as absolute difference between 2
scans, divided by average value.
tion.