Professional Documents
Culture Documents
SPE/DOE 11638 A Method For Evaluation of Formation Damage Due To Fracturing Fluids
SPE/DOE 11638 A Method For Evaluation of Formation Damage Due To Fracturing Fluids
SPE/DOE 11638
This paper was presented at the 1983 SPE/DOE Symposium on Low Permeability held in Denver, Colorado, March 14-16, 1983. The material is subject
to correction by the author, Permission to copy is restricted to an abstract of not more than 300 words. Write SPE, 6200 North Central Expressway,
Drawer 64706, Dallas, Texas 75206.
333
2 A METHOD FOR EVALUATION OF FORMATION DAMAGE DUE TO FRACTURING FLUIDS SPE 11638
Various methods have been devised to measure ~Pc = ~ [6P c (initial) + 6P c (final)]
gas flow and core permeability. Two methods of
interest based on nonsteady state gas flow have Pi-CPo + ~ 6P t )
been described. Jones (17) measured pressure draw
down of pressure reservoir located upstream of a Also, the effective mean core pressure (P ) can be
core sample. More recently, Freeman and Bush (18) expressed as: c
described a method to measure low permeability
cores wi~h a nonsteady state gas flow through a
sample wlth a downstream pressure buildup test. '" ~ [pc (initial) + Pc (final) ]
Both methods reported rapid analysis times with
go?d.accuracy. The method described in this paper Pi + P + ~P )]
P. + (p
=~ [ 0 + lot
utlllzes a downstream pressure buildup (~P ) 2 2
in a known volume. A solenoid valve, installed on
~he outlet of a known volume, was closed period- =~ [Pi + (PO+~6Pt)J
lcally for a specific time interval and then
opened allowing equilibration to atmospheric Therefore, equation 4 can be rewritten in experi-
pressure. The pressure buildup was measured with mentally measured parameters.
a pressure transducer and the resulting pressure
(~Pt) displ~yed.o~ a strip c~art recorder. Figure
3 snows a slmpllfled schematlc of the essential
components and parameters.
If steady state flow th~ough a core is assumed, The above formulation assumes a constant temperature
the volume of an ideal gas (V ) passing through (T t ) and pressure equilibrium throughout the
the core in time (t) can be c~lculated by the bUildup volume. The temperature of the gas leaving
relationship: the core is approximately the core temperature (T )'
c
334
SPE 11638 VOLK, GALL, RAIBLE, CARROLL 3
However, even at elevated core temperatures, the by measuring gas permeabilities at 3 mean pressures
gas rapidly cools to Tt (ambient temperature) due and extrapolating to infinite mean pressure. Many
to the low heat capacity of the gas relative to of the permeability values used in the following
the plumbing. Pressure and temperature measurements discussion will be gas permeabilities at 1000 psig
at the exit of the cell verify the assumption re- (6.9 MPa). These values are very close to the
quired to use equation 5. Furthermore, flow rates Klinkenberg corrected permeabilities.
measured by conventional techniques agree with those
determined using equation 5. MOLECULAR WEIGHT DETERMINATIONS
Technique of Gas Flow Measurement High pressure liquid chromatography (HPLC)
Typically, regained-permeability experiments techniques (19) were used to determine the presence
involved slowly increasing flow. In order to as- and molecular weight distribution (MWD) of fractur-
certain the time required to establish steady state ing fluid polymers passing through the cores. A
flow, automated flow measurement was an important liquid chromatographic system, equipped with a
consideration in system design. Figure 4 is a HPLC pump, automatic injection refractive index
schematic of the auto-ranging flow monitoring detector and three Toyo Soda columns (two-G5000PW,
system. A timer was used to operate solenoid one-G3000PW) (20), was used to separate molecules
valve (SV'A'). This valve was periodically closed according to size. The method, commonly known as
for some time, t, allowing pressure build-up in gel permeation chromatography (GPC), elutes the
the downstream volume. The volume, previously largest molecules first with progressively smaller
referred to as V , includes the combined volume of molecule elution. Dextran standards with a known
valves, pressure t transducers, test cell downstream molecular weight (MW) were used to characterize
of the core outlet, and connecting tubing. This the col umn set. Dextran molecules, however, are
volume was measured as 90 cm 3 • Normal cycle time was more linear molecules than those of guar and guar
5 or 10 minutes with SV'A ' closure time of one derivatives (21) so that the absolute calibration
minute. However, these times could be selected in with respect to guar polymers is not precise.
0.1 minute multiples to provide flexibility in However, comparative MW curves and changes in the
properly scaling pressure build-up, ~P. Also, molecular size distribution of the broken polymers
two pressure transducers were used wit~ ranges of were possible using the column set.
0-100 torr and 0-1000 torr. Depending on sample
permeability and P., either the low or high range Figure 6 shows a composite of the MWD chro-
was used. 1 matograms for broad MW polydextran standards.
Each standard curve represents a MW range with the
A strip chart recorder displayed the pressure average MW indicated near the apex of each curve.
build-up from which flow rates could be calculated. The exclusion limit of the column set is approxi-
The flow system automatically increased downstream mately 3 to 5 million. Some of the T-2000 standard,
volume, V , as the pressure, ~P , increased to which has an extremely broad MW range, is excluded
some pres~t value in order not to exceed recorder from the column packing pores resulting in a
full scale (R f ). This was accomplished by monitor- bimodal curve. In the remaining discussion, refer-
ing the transdBcer voltage output. Normally, the ences to molecular weights of gels are for illustra-
voltage monitor was set at 90 and 95 percent of Rfs ' tive comparisons. Actual gel MW is not known.
If the transducer voltage increased to 90 percent
of R ,SV'B ' opened connecting the 160 cm 3 tank, Unbroken hydroxypropyl guar (HPG) elutes at
V , fa the downstream manifold. This increased the exclusion limit, indicating a MW greater than
tRe total volume to 250 cm 3 • Similarly, if bP t 5 million as shown in Figure 7. Broken gel,
increased to 95 percent of R ' SV1C 1 opened however, elutes within the separation range of the
connecting an additional 950 f ~m3 tank, V . When column set. Figure 7 also shows the MWD of HPG
either of these solenoid valves opened, ft remained fracturing fluid containing breaker and heated to
open until reset by the operator. An example of 343°K for various time durations. The composite
recorder response for a regained-permeability test chromatogram shows distributions of progressively
is shown in Figure 5. This figure shows a typical smaller molecular size as the duration of the oxi-
sequence of valve operation and the resultant ~Pt dation increases.
attenuation due to increased downstream volume.
The GPC system was useful for detection of
Nitrogen, with ideal gas properties, was used gels in fluid leak-off, as well as, estimating
for flow measurements. The system has a flow size of molecules which penetrated through core
measurement range of 4 orders of magnitude using samples. Broken gels were also characterized
all equipment parameters. Assuming 11 atmospheres prior to being used in regained-permeability
pressure drop (AP ) and mean pressure (Pc) of 6 experiments.
atmospheres acros~ a core, the system is ca~able .
of measuring .002 to 20 cm 3 /sec. For a tYPlcal Slze EXPERIMENTAL PROCEDURES
core used in these experiments, this corresponds
to a permeability range of 0.4 to 5000 J-ld. For Gel Preparation
unattended operation using the same criteria, the Hydroxypropyl guar (HPG) was prepared in 40
system would measure approximately 0.002 to 0.25 and 80 lb/1000 gal concentrations (4.8 gm/L and
cm 3 /sec or 0.4 to 60 J-ld for low flow operation. 9.6 gm/L). The polymers were hydrated in a
potassium dihydrogen phosphate solution brought to
Permeabilities measured with this system were pH 10 with 10 percent KOH. After hydration, the
effective or apparent gas permeabilities. Klin- solution pH was lowered to between 6 and 6.5.
ken berg corrected permeabilities were determined Typical viscosities measured with a Fann 35
335
4 A METHOD FOR EVALUATION OF FORMATION DAMAGE DUE TO FRACTURING FLUIDS SPE 11638
336
SPE 11638 VOlK, GAll, RAIBLE, CARROLL 5
The permeability reduction was most likely due to Figure 12 illustrates typical fluid loss
the insoluble material produced on breaking which curves observed for wall-building gels on cores.
formed a filter cake. This residue would not be The apparent negative spurt loss is due to the
evident in a GPC chromatogram. Removing 0.14 cm presence of the core which provided additional
from the core surface restored the permeability to resistance to fluid flow. As expected, the slope
36.6 lJd. of the two fluid loss curves approach the origin
at zero slope, typical of Darcy flow. The per-
Broken HPG gel was found to pass through both meability difference between the two core was
a core and a core with filter cake buildup in primarily responsible for the different slopes.
te s t s run a t 343 0 K• Us i ng C1GE- 21 ( 6499. 7) , Fluid loss coefficients were calculated to be 1.3
broken HPG gel also penetrated the core with X 10- 3 ft/m;n~ for C1GE-21 (6499.7) and 0.3 X 10- 3
approximately the same MWD as the original solution ft/mi for MWX-l (7884C). In general, no dynamic
(Figure 8). If a gel coating of unbroken HPG filtration effects were observed. This is due to
was first formed on the surface of the core, the low shear rate of the fluid as it passes over
broken gel passed through the core with a lower the core and the rather large flow slot (0.5 cm)
MWD than that in the solution (Figure 9). There- relative to the filter cake thickness (0.1 cm or
fore, the filter cake has effectively a much less),
smaller pore structure than does the core. The
core essentially passes molecules of molecular CONCLUSIONS
weight of 10 5 -10 6 without significant attenuation.
The filter cake, however, begins retaining molecules (1) An apparatus is presented for continuously
having molecular weights greater than 10 4 • monitoring changing permeability of a
core over a wide range for extended periods
Time Required to Restore Permeability of time. This is accomplished automat;
The time required to regain stable permea- cally while maintaining the core at in-situ
bility should depend on the type of fluid used to conditions. The principle behind this
saturate the core, temperature, original per- technique can be extended to lower or
meability of the core, and the pressure of the higher flow rates if desired.
clean-up gas. Table 2 lists the time needed to
obtain 80 percent of the final regained-permea- (2) Depth of invasion of unbroken HPG is very
bility, t 8' using N2 at 1000 psig (6.9 MPa). The small, most likely less than 0.02 cm into
value of ~ percent was an arbitrary choice to rock with permeabilities less than 120 ~d
facilitate comparisons. (gas) and 600 psig per cm pressure gradient
ac ros s the core.
As expected, the cores cleaned up consi-
derably faster at 343°K, partially due to the (3) Regardless of the rock permeability (0.6
lower viscosity of water (brine or filtrate) at to 115 ~d), the permanent damage remains
elevated temperatures. Exposure to HPG increased approximately constant at a given tempera-
the regained-permeability time. ture (17-20 percent at 343°K).
A correlation between the ratio of clean-up (4) Degraded HPG passes through tight sand-
time after exposure to gel, teo (gel), and clean- stone without significantly altering
up time after exposure to brine, tso (brine), and even the high molecular weight fractions
the initial permeability (k.) can be seen in (MW 0 f 10 5 - 10 6 ) •
Figure 10 for tests perform~d with a confining
stress of 3000 psig (20.7 MPa). Until more work (5} Only degraded HPG molecules with molecular
can be accomplished, the exact nature of all the weights of 10 4 or less pass through un-
factors influencing the relative rate of clean-up hindered if a gel filter cake is allowed
is not known. Some possible factors, however, to form on the core surface.
include greater polymer deposition on the surfaces
of the more permeable cores since filtrate volume (6) For core of the permeabilities used in
is considerably greater for these cores. this study, with HPG, there appears to be
In. addition, the cores with higher permeability a correlation between the relative rate
may have greater depth of invasion of the frac- of clean-up and the formation permeability,
turing fluid polymer causing them to clean-up more
slowly. The final gas regained-permeability may (7) Humidification of the gas used in regained-
not be a strong function of filter cake thickness permeability tests does not seem to have
or depth of invasion, but it may have some in- much effect on the rate of ultimate
fluence on the relative clean-up time. permeability attained.
Fluid Loss
In general, brine flow through a core or a
core with a gel build-up was linear with time.
Degraded HPG (fluid viscosity of 3 cp or less at
479 sec-i) also passed through a core and filter
cake with a flow rate that was linear with time as
shown in Figure 11. Although degraded HPG solu-
tion possessed some solids, apparently the rate of
deposition on the existing filter cake did not
significantly alter the cake permeability or
thickness within the experimental time limits.
337
6 A METHOD FOR EVALUATION OF FORMATION DAMAGE DUE TO FRACTURING FLUIDS SPE 11638
Time required for core permeability to 10. Simon, D.E., and Derby, J.A., IIFormation
increase to 80 percent of its final Damage is Evaluated by New Instruments New
value after saturation Methods,1I The Oil and Gas Journal, Sept. 20,
1976, pp 209-214.
Vi seas ity, cp
11. Pence, S.A., "Evaluating Formation Damage in
Volume of an ideal gas at pressure P and Low Permeability Sandstone", Paper SPE 5638
T cm 3 c presented at the SPE-AIME 50th Annual
c' Fall Meeting, Dallas, Tx, Sept. 28-0ct. 1
Vt Measurement volume downstream of core, cm 3 1975. '
Held by Four Unconventional Gas Sources," 14. White, J.L., and Means, J. 0., "Polysac-
Oil and Gas Journal, Vol. 76, (1978), charide Derivatives Provide High Viscosity
pp 47-54. and Low Friction at Low Temperatures ," J.
Pet. Tec h. , ., 1975. pp 1067 - 1073 •
3. Fast, C.R., Holman, G.B., and Calvin, R.J.,
liThe Application of Massive Hydraulic 15. McDaniel, R.R., Houx, M.R., and Barringer,
Fracturing to the Tight Muddy IJI Forma- O.K., "A New Genertion of Solid Free Frac-
tion, Wattenberg Field, Colorado," J. turing Fluids," Paper SPE 5641 presented at
Pet. Tech., (Jan. 1977), pp 10-16. the 50th Annual Fall SPE Meeting, Dallas,
TX., Sept. 28-0ct. 1,1975.
338
SPE 11638 VOLK, GALL, RAIBLE, CARROLL 7
16. Free, D.L., Fredrick, A.F., and Thompson, 19. Yau, W.W., Kirkland, J.Jo, and Bly, D.O.,
J.E., "Fracturing with a High Viscosity Modern Size Exclusion Liquid Chromatography,
Crosslinked-Gel Continuous Fracturing John Wiley and Sons, Inc., 1979.
Techni que, Paper 5640 presented at the
II
50th Annual Fall SPE Meeting, Dallas, 20. Hashimoto, T., Sasaki, H., Aiura, M., and
TX., Sept 28-0ct. 1, 1975. Kato, T., "High Speed Aqueous Gel Permeation
Chromatography," J. Polym. Sci., Polymer
17. Jones, S.C., "A Rapid Accurate Unsteady State Phy sic sEd i t i on, Vo 1. 16 , (1978), pp
Kl i nkenberg Permeabi 1i ty ," Soc. Pet. Eng. 1789-1800.
J., (Oct. 1972), pp 383-397.
21. Barth, Ho G., and Smith, D.A., "High Per-
18. Freeman, D.L., and Bush, D.C., "Low Per- formance Size Exclusion Chromatography
meability Laboratory Measurements by of Guar Gum," J. Chromatogro, 206, (1981),
Nonsteady State and Conventional Methods," pp 410-415.
Paper SPE 10075 presented at the SPE 56th
Annual Fall Technical Conference and
Exhibition, San Antonio, TX., Oct. 5-7, 1981.
Confining stress 3000 psig (20.7 MPa) and Fluid Pressure 2000 psig (13.8 MPa)
CIGE-21
6499.1 28.8 26.2 8 295 80 (9.6)
CIGE-21
6499.5 35.5 29.5 19 343 40 (4.8)
MWX-1
7955 0.58 0.48 18 343 40 (4.8)
Confining stress 1700 ps i 9 (11.7 MPa) and Fluid Pressure 700 psig (4.8 MPa)
MWX-l
7884A 5.3 4.4 17 343 40 (4.8)
MWX-l
7884C 4.5 3.6 20 343 40 (4.8)
CIGE-21
6499.7 115 91.6 20 343 SO (9.6)
*Perrneability measured at 1000 psig (6.9 r~Pa) nitrogen pressure
339
rHydrostatic stress
I
~"""","""'.,..,J..__/Rubber sleeve
E,ces~
fluid
Bock pressure
regulator
Displacement
Pressure tank
-----0
test cell
[1n.CUJ
Li quid fraction
collector or Fractur ing fluid
B eCO'de,
Transducer
readout
~::----
J
I
H ig h pressure
____-1--11, 1 <0,d,,1
Re
Water
Nitrogen
humidifier
14_
16
I
Il
>
>
+
fD
><f.
12
No oxidant
~IO Solvent
w
~ en peak
w z
o 24 Hours
~
8 a..
i= en
w
0::
ttl 0::
6 > o
+ f-
o
l ><f. W
r-
:i 1
w
o
~ Elution volume- 53 ml
I I I I 4
0 20 40 60 80 100 - Molecular size
RECORDER SCALE Fig. 7-Composile chromatogram showing molecular weight distribution change with time as H PG gel reacts
with oxidant breaker.
Fig. 5-Example of recorder response for regained·permeability test.
w
en Sol vent
z Solvent w peak
o peak en
Cl.
en z
o
W a..
0:: en
w
0:: 0::
o
r- 0::
o o Broken HPG
w f-
r- o diluted by
w W 1/2 Broken H PG
o r- t h roug h core
w
o
Exclusion
limit
a:::
Solvent
I.iJ
(1) pea k
z
o IJ)
E
a..
(1) ~ 20
w Orig ina I broken
HPG\
01
a:::
r:
a::: I
o ~
I- W
u ~
W
I- 0
W
I.iJ I-
o Fluid passed «
...J
through gel ~ ::J
bu i Idup and :::E
::> 10
i:.··qre <.>
,~;~~ <.>
«
25 ml Elution volume ----IIi"' 53ml
- - Molecular size
Fig. 9-Molecular weight distribution of broken HPG gel passed through core and gel filter cake compared with
MW of the original broken HPG gel.
o 10 20 30 40
TIME, minutes
tsO(GEL) / tSO( BR I NE)
Fig. 11-Broken gel filtrate passage through a core with previously
deposited filter cake.
o
6.-------------------------------------------
0
G')
1:>
(J)
II
fTt 5
::0 N
3: 0
fTt
1:>
CD
IJ)
r E
-i ~ 4
-< (,>J
0
01
~
J:
<.9
W
1::
Q. 3: 3
.f:>
0 o
W
I-
«
...J
::J
~ 2
~~--------------------------~----------~ u
u
«
Fig. 10-Relalive cleanup lime vs. gas regained-permeability at 1.000
pSlg.
o~------~--------~-------L------~------~
2 4 6 8 10
TIME, minutes 1/2
Fig. 12-Typical fluid·loss curves for unbroken HPG gel.