Professional Documents
Culture Documents
Skanmobile User Manual
Skanmobile User Manual
User Responsibility
This product is designed to perform its intended function when operated in accordance with the instructions
provided in this manual and other accompanying labels and accessories and when assembled as per the
instructions provided. A defective unit should not be used. Parts that are broken, plainly worn, missing,
incomplete, distorted or contaminated should be replaced immediately.
There are no user repairable components / modules used as a part of this product. In case of a defective or
malfunctioning unit the repair should be carried out by a trained technician authorised by Skanray. The user
shall have the sole responsibility of any outcome originating from this product resulting from improper use, faulty
maintenance, damage or use of accessories / replacements parts not approved by Skanray.
Accessories
Use of accessories / replacement parts that are not recommended by Skanray could compromise the
performance of the unit if not compromise operator / patient safety.
WARNING
Skanray will not be able to honour any warranty claims due to the usage of accessories / replacement parts not
recommended by it. Additionally Skanray shall not be liable for any losses incurred due to the usage of any such
non-recommended accessories/parts. For Continued protection against risk of fire and electric shock, replace
the fuse with type 16A / 250V, Fast blow. Ensure the power inlet to the system is disconnected or mains switch
is switched off before replacing fuse.
For improving product quality and customer experience, we at Skanray are dedicated to upgrading
the technology and processes used in our product. The information provided in this document is
subject to change without prior notice. Please contact Skanray for latest updates on the product and
NOTE services.
Revision 6 Page I of IV
Revision 6 Page II of IV
Table Of Contents
1 Introduction ....................................................................................................................................................... 1
1.1 Your X-Ray Equipment .............................................................................................................................. 1
1.2 Indication for Use ...................................................................................................................................... 1
1.3 This Manual .............................................................................................................................................. 1
2 Safety and Precautions ..................................................................................................................................... 3
2.1 Patient and Operational Safety ................................................................................................................. 3
2.2 Conventions .............................................................................................................................................. 4
3 Know Your X-Ray Unit ...................................................................................................................................... 7
3.1 X-Ray System Components – A Checklist ................................................................................................ 7
3.2 Identification of Main Parts ........................................................................................................................ 7
3.3 System Labels, Controls and Indicators .................................................................................................. 10
3.4 Control and Indicators– Identifications (Integrated Console) .................................................................. 16
3.5 Control and Indicators– Identifications (External Console) ..................................................................... 17
3.6 Mechanical Dimensions: ......................................................................................................................... 19
4 Operating The Unit .......................................................................................................................................... 21
4.1 Before You Begin .................................................................................................................................... 21
4.2 Power ON And Initial Checks .................................................................................................................. 21
4.3 Modifying exposure parameters .............................................................................................................. 22
4.3.1 Parameter Modification Screen: kV ................................................................................................. 22
4.3.2 Parameter Modification Screen: mAs .............................................................................................. 22
4.3.3 Mode Selection Screen ................................................................................................................... 22
4.3.3.1 MEDIUM ................................................................................................................................. 23
4.3.3.2 THIN ....................................................................................................................................... 23
4.3.3.3 THICK ..................................................................................................................................... 23
4.3.4 Exposure History ............................................................................................................................. 24
4.3.5 Sleep Mode ..................................................................................................................................... 24
4.4 Positioning X-Ray Beam Using Collimator .............................................................................................. 24
4.5 Delivering an Exposure ........................................................................................................................... 25
4.6 User Default Parameter Configuration Settings ...................................................................................... 27
5 Maintenance ................................................................................................................................................... 29
5.1 Cleaning and Disinfecting ....................................................................................................................... 29
5.2 Caring for Your Equipment ...................................................................................................................... 29
5.3 Shipping and Long Term Storage ............................................................................................................ 29
5.4 Preventive Maintenance ......................................................................................................................... 30
5.5 Disposal of the Unit ................................................................................................................................. 30
6 Troubleshooting .............................................................................................................................................. 31
6.1 Errors & Warnings ................................................................................................................................... 31
7 Technical specifications ................................................................................................................................... 33
7.1 Tube-head Specifications ........................................................................................................................ 33
7.2 X-Ray Tube Insert Specifications ............................................................................................................ 33
7.3 Dimensions and Weight .......................................................................................................................... 35
7.4 Mains Power Requirements .................................................................................................................... 36
7.5 Environmental Conditions ....................................................................................................................... 36
Annex A: Exposure Time & mA List For Selected kV And mAs (230 V /110 V) .................................................. 37
Annex B: Basic X-Ray techniques used for Different Anatomy. ......................................................................... 41
Annex C: Declaration of Conformity ................................................................................................................... 67
Annex D: Guidance and Manufacturer’s Declaration ......................................................................................... 69
Annex E: Contact details .................................................................................................................................... 73
1 Introduction
This manual describes how to use the SKANMOBILE High frequency diagnostic X-Ray system. Read this
manual carefully before using the equipment.
This manual is applicable to the following variants of SKANMOBILE:
Page 1 of 73
SKANMOBILE‒ High Frequency
Chapter 2 Safety and Precautions
Diagnostic X-Ray system
Warranty of this equipment will be void in the event of any modification done to the equipment,
misuse of the equipment and opening or servicing by an unauthorized personnel.
WARNING
4) Patients should be provided with lead apron and thyroid collar while being exposed.
6) Operator should be at a distance of at least 2 meter away from the tube head while
carrying out the exposure.
7) Operator should not stand in front of collimator during exposure. Operator should stand
behind the Tube head assembly during exposure.
ELECTRICAL SAFETY 1) Always switch off the unit and remove the mains plug when cleaning and disinfecting the
unit.
2) Do not touch patient and accessible conductive parts simultaneously, to avoid Electric
shock.
3) The unit contains lethally high voltages. Do not attempt to open covers or repair the unit
on yourself or by non certified service personnel.
EXPLOSION SAFETY 1) This equipment must not be used in the presence of flammable or potentially explosive
disinfecting gases or vapours, which could ignite causing personal injury and/or damage to
the equipment. If such disinfectants are used, the vapour must be allowed to disperse before
using the equipment.
Page 3 of 73
SKANMOBILE‒ High Frequency
Chapter 2 Safety and Precautions
Diagnostic X-Ray system
ELECTRO MAGNETIC 1) Interference between the unit and other sensitive electronics can occur under extreme
INTERFERENCE
conditions. Do not use the X-Ray equipment in close conjunction with other sensitive devices
or devices which create high electromagnetic disturbance.
PHYSICAL INJURY 1) The Swivel arm lifts up suddenly if the gas spring actuation / push button is pressed
without the integration of Tube head assembly to the trolley.
2) Operator should be at a distance of at least 2 meter away from the tube head to avoid any
unintended movement.
INSTALLATION AND 1) Ensure that your X-Ray unit is assembled and installed inside the Hospital or clinic
SERVICE
building, by qualified service personnel. Consult the factory or your seller for installation of the
unit.
MOBILITY 1) Care must be taken for the movement and positioning of the system.
2) The system must be disconnected from the main power before moving.
3) Before moving system around, the system must be positioned to the parking position to
avoid unnecessary damage to the system.
4) While bringing the tube head into parking position, make sure tube head is positioned at
90 degrees to the swivel arm to avoid tube-head hitting the lower portion of trolley.
6) After the system is placed at the desired location, the wheels should be locked.
2.2 Conventions
WARNING - Warning statements describe conditions or actions that may result in personal injury or
loss of life.
CAUTION - Caution statements describe conditions or actions that may result in damage to the
equipment or software.
NOTE - Notes contain additional information on the use of the system.
Attention
This symbol invites the attention of the user towards a WARNING, CAUTION or NOTE.
Page 4 of 73
SKANMOBILE‒ High Frequency
Chapter 2 Safety and Precautions
Diagnostic X-Ray system
Take Note
This symbol points to an important detail / tip in the operation of the unit.
Protective Earth
Type of Insulation
High Voltage
Caution: X-Ray
X-Ray Source Assembly / Tube Head capable of generating X-Rays. This X-Ray unit may be
dangerous to patient & operators unless safe exposure factors and operating instructions are
observed.
Manufacturing date
Manufacturer's address
Temperature range
Requires special disposal methods. Consult local regulatory body for identifying proper disposal
method.
Alternating current
Focal Spot
Page 5 of 73
SKANMOBILE‒ High Frequency
Chapter 3 Know Your X-Ray Unit
Diagnostic X-Ray system
Page 7 of 73
SKANMOBILE‒ High Frequency
Chapter 3 Know Your X-Ray Unit
Diagnostic X-Ray system
Illustration 2: TROLLEY
Illustration 3: TROLLEY
Page 8 of 73
SKANMOBILE‒ High Frequency
Chapter 3 Know Your X-Ray Unit
Diagnostic X-Ray system
Page 9 of 73
SKANMOBILE‒ High Frequency
Chapter 3 Know Your X-Ray Unit
Diagnostic X-Ray system
Page 10 of 73
SKANMOBILE‒ High Frequency
Chapter 3 Know Your X-Ray Unit
Diagnostic X-Ray system
LABEL 11: SYSTEM LABEL SKANMOBILE- 230V LABEL 12: SYSTEM LABEL SKANMOBILE-
(L9) 110V (L9)
Page 11 of 73
SKANMOBILE‒ High Frequency
Chapter 3 Know Your X-Ray Unit
Diagnostic X-Ray system
LABEL 24: X-RAY TUBE (L20) LABEL 25: TUBE HOUSING SERIAL NUMBER
(L21)
Page 12 of 73
SKANMOBILE‒ High Frequency
Chapter 3 Know Your X-Ray Unit
Diagnostic X-Ray system
Page 13 of 73
SKANMOBILE‒ High Frequency
Chapter 3 Know Your X-Ray Unit
Diagnostic X-Ray system
Page 14 of 73
SKANMOBILE‒ High Frequency
Chapter 3 Know Your X-Ray Unit
Diagnostic X-Ray system
Page 15 of 73
SKANMOBILE‒ High Frequency
Chapter 3 Know Your X-Ray Unit
Diagnostic X-Ray system
LED Indications
1) Yellow colour indicates system is getting prepared to take X-Ray by
PREP / READY LED preheating filament.
FAULT LED 1) When there is a fault in the system, FAULT LED will glow in Red colour.
KEY Functions
1) Increments the set kV in steps of 1 and jumps to steps of 5 for continuous
kV INCREMENT KEY pressing.
1) Increments the set mAs in steps of 0.1 and jumps to steps of 1 and then
mAs INCREMENT KEY to 10 for continuous pressing.
1) Decrements the set mAs in steps of 0.1 and jumps to steps of 1 and then
mAs DECREMENT KEY to 10 for continuous pressing.
RESET KEY 1) In fault state, this key resets the generator to standby mode. Some faults
cannot be reset by this key. They are reset only by switching off the power
Page 16 of 73
SKANMOBILE‒ High Frequency
Chapter 3 Know Your X-Ray Unit
Diagnostic X-Ray system
once and switching on again. These are called as “Power on reset faults”
and list is given in generator fault list Section 6.1.
1) Switches on and off the collimator lamp. Internal timer switches off the
LAMP KEY
light after 30seconds.
DISPLAY Indications
1) Displays default/set KV value in standby mode.
KV DISPLAY
2) Displays Error message Exx(where xx stands for code) in fault mode
LCD Display
LED Indications
1) Yellow colour indicates system is getting prepared to take X-Ray by
3) X-Ray on Indication
Page 17 of 73
SKANMOBILE‒ High Frequency
Chapter 3 Know Your X-Ray Unit
Diagnostic X-Ray system
FAULT LED 1) When there is a fault in the system, FAULT LED will glow in Red colour.
LAMP ON LED 1) When the lamp is switched ON, LAMP ON LED will glow in yellow colour
KEY Functions
1) Increments the set mAs in steps of 0.1 and jumps to steps of 1 and then to
mAs INCREMENT KEY 10 for continuous pressing.
1) Decrements the set mAs in steps of 0.1 and jumps to steps of 1 and then
mAs DECREMENT KEY to 10 for continuous pressing.
1) In fault state, this key resets the generator to standby mode. Some faults
cannot be reset by this key. They are reset only by switching off the power
RESET KEY
once and switching on again. These are called as “Power on reset faults” and
list is given in generator fault list Section 6.1.
1) Switches on and off the collimator lamp. Internal timer switches off the light
LAMP KEY
after 30seconds.
DISPLAY Indications
Page 18 of 73
SKANMOBILE‒ High Frequency
Chapter 3 Know Your X-Ray Unit
Diagnostic X-Ray system
+4º
Illustration 8: Overall Dimension (Min position at 23° ): 1330×1188×716 (±50mm)
−1º
Page 19 of 73
SKANMOBILE‒ High Frequency
Chapter 3 Know Your X-Ray Unit
Diagnostic X-Ray system
Illustration 9: Overall dimension (Max position at 139° ±4°): 2050×1110×716 (±50 mm)
All dimensions are in mm
Page 20 of 73
SKANMOBILE‒ High Frequency
Chapter 4 Operating The Unit
Diagnostic X-Ray system
Ensure that the operator has read and understood this manual regarding
operation of the system. Government regulators may require a licensed
operator to use this equipment. Check with your local seller regarding this.
CAUTION
You should be well acquainted with the radiation protection methods for both
the operator and patient before attempting to use this equipment.
1) Majority of repeat exposures and inferior X-Ray images are attributed to the
storage, handling, use and developing of X-Ray films rather than the
FILM DEVELOPMENT
equipment itself. Ensure that the image capture films are stored and used as
per instructions.
Let the patient know that he/she is going to be X-Rayed. Avoid X-Rays or take
necessary precautions when X-Raying pregnant patients.
On power up, integrated console will initially display the firmware version and then displays default
kV and mAs values. Default values will be 40 KV and 0.1 mAs or the previously configured value.
Page 21 of 73
SKANMOBILE‒ High Frequency
Chapter 4 Operating The Unit
Diagnostic X-Ray system
While this is displayed the console goes through a state of self test for making sure that all the
internal and external components of the console are working fine.
During this stage please do not press any keys on the keypad for they will be treated as a
keyboard error.
Immediately following a successful self test the console displays a screen similar to the one as
shown below. All the processes in this console starts from the home screen.
Default KV and mAs values will be displayed once the generator is ready to take X-Ray. Default
values will be 40 KV and 0.1 mAs or the previously configured value.
Page 22 of 73
SKANMOBILE‒ High Frequency
Chapter 4 Operating The Unit
Diagnostic X-Ray system
4.3.3.1 MEDIUM
Under this there is a list of 23 different Anatomy that the User can configure for each anatomy for a
particular image receptor.
S/N Anatomy S/N Anatomy S/N Anatomy S/N Anatomy S/N Anatomy S/N Anatomy
1 Chest AP 5 Hip 9 Ankle 13 Humerus 17 Hand 21 L-Spine Lat
2 Chest Lat 6 Femur 10 Foot 14 Elbow 18 T-Spine AP 22 C-Spine AP
3 Abdomen 7 Knee 11 Shoulder 15 Forearm 19 T-Spine Lat 23 C-Spine Lat
4 Pelvis 8 Tibia 12 Swimmers 16 Wrist 20 L-Spine AP
Use the kV INCREMENT / DECREMENT keys to navigate within the list. Press these buttons until
the desired setting is highlighted.
Press the SET key to use the highlighted mode. The default values are displayed. Pressing the
MODE key without saving returns to home screen without any changes.
Now set kV and mAs parameters and press SET key, automatically save the parameters into
particular anatomy.
4.3.3.2 THIN
Under this there is a list of 23 different Anatomy that the User can configure for each anatomy for a
particular image receptor.
S/N Anatomy S/N Anatomy S/N Anatomy S/N Anatomy S/N Anatomy S/N Anatomy
1 Chest AP 5 Hip 9 Ankle 13 Humerus 17 Hand 21 L-Spine Lat
2 Chest Lat 6 Femur 10 Foot 14 Elbow 18 T-Spine AP 22 C-Spine AP
3 Abdomen 7 Knee 11 Shoulder 15 Forearm 19 T-Spine Lat 23 C-Spine Lat
4 Pelvis 8 Tibia 12 Swimmers 16 Wrist 20 L-Spine AP
Use the kV INCREMENT/DECREMENT keys to navigate within the list. Press these buttons until the
desired desired setting is highlighted.
Press the SET key to use the highlighted mode. The default values are displayed. Pressing the
MODE key without saving returns to home screen without any changes.
Now set kV and mAs parameters and press SET key, automatically save the parameters into
particular anatomy.
4.3.3.3 THICK
Under this there is a list of 23 different Anatomy that the User can configure for each anatomy for a
particular image receptor.
Page 23 of 73
SKANMOBILE‒ High Frequency
Chapter 4 Operating The Unit
Diagnostic X-Ray system
S/N Anatomy S/N Anatomy S/N Anatomy S/N Anatomy S/N Anatomy S/N Anatomy
1 Chest AP 5 Hip 9 Ankle 13 Humerus 17 Hand 21 L-Spine Lat
2 Chest Lat 6 Femur 10 Foot 14 Elbow 18 T-Spine AP 22 C-Spine AP
3 Abdomen 7 Knee 11 Shoulder 15 Forearm 19 T-Spine Lat 23 C-Spine Lat
4 Pelvis 8 Tibia 12 Swimmers 16 Wrist 20 L-Spine AP
Use the kV INCREMENT/DECREMENT keys to navigate within the list. Press these buttons until the
desired desired setting is highlighted.
Press the SET key to use the highlighted mode. The default values are displayed. Pressing the
MODE key without saving returns to home screen without any changes.
Now set kV and mAs parameters and press SET key, automatically save the parameters into
particular anatomy.
Adjust the light field using the 2 knobs(Dials) provided in collimator, as per the table below. The light
field adjusted will be the X-Ray field while taking X-Ray. Align the centre of the grid to centre of the
Image receptor. Grid is indicating centre of the X-Ray beam axis.
Page 24 of 73
SKANMOBILE‒ High Frequency
Chapter 4 Operating The Unit
Diagnostic X-Ray system
There is an automatic timer of 30 seconds for collimator light. This will switch off the light after 30
seconds. If the light field adjustment is still required, press the LAMP KEY again.
Press LAMP KEY or the preparation button in the Exposure Hand Switch to switch off the collimator
lamp, if it is on.
Press the actuator of the exposure handswitch to level I prep (preparation) as shown in Illustration
10. PREP/READY LED will glow in yellow colour and the buzzer will beep audibly in periodic fashion.
This indicates system is getting prepared to take X-Ray.
Page 25 of 73
SKANMOBILE‒ High Frequency
Chapter 4 Operating The Unit
Diagnostic X-Ray system
When the generator is ready to take X-Ray, PREP/READY LED will turn to green and buzzer will
beep audibly in periodic fashion at twice the rate of the beep during preparation mode. The external
console displays as shown below
Press the actuator of the exposure handswitch to level II exposure as shown in Illustration 10 to take
X-Ray and hold till the buzzer stops or X-RAY ON LED goes off. Buzzer and X-RAY ON LED will be
continuously ON throughout the exposure time. During exposure the external console displays as
shown below
Feed back kV and mAs values will be displayed after exposures for few seconds.
Generator enters cooling mode after taking each X-Ray. Generator cooling mode is indicated in
display as “CLG Prd”. Cooling period depends on the exposure time. The external console displays
as shown below.
Page 26 of 73
SKANMOBILE‒ High Frequency
Chapter 4 Operating The Unit
Diagnostic X-Ray system
There is a fan provided inside the generator for cooling, which needs specific time
to cool the generator electronics before taking next X- Ray. Power OFF/ON reset X- Ray exposures
Generator will display previously set kV and mAs values after the completion of cooling period and is
ready to take next exposure.
Press RESET KEY once to take the Generator to parameter display mode.
After one second delay, press RESET KEY once more to take the generator to standby mode
(Shows default kV and mAs) followed by mAs DECREMENT KEY immediately.
The generator enters into Default Parameter Configuration mode (It shows last configured
parameters if set, if not, then it shows 40 kV, 0.1 mAs).
After the values are set, press RESET KEY once followed by mAs DECREMENT KEY immediately.
The unit enters into standby mode with the previously configured parameters.
The configuration will be retained once the unit is switched OFF and ON and will be displayed as
default value henceforth.
If the operator wants to change the default values, repeat the above procedure.
Page 27 of 73
SKANMOBILE‒ High Frequency
Chapter 5 Maintenance
Diagnostic X-Ray system
5 Maintenance
5.1 Cleaning And Disinfecting
1) Use a soft cloth damped in a mild soap solution for cleaning the outside
surfaces of the unit.
Switch off the unit when leaving for the day or when not used for a long time.
Do not operate the swivel arm without pressing the gas spring actuating / push button.
Do not force the arm mechanisms or tube head into a position it is not designed for. There are
movement stoppers provided.
Do not hang external loads or weights on the tube head or extension arm. The arm and base units
are designed for its own weight and may not hold an additional weight.
Place the Tube head assembly to the parking position (23º angle of swivel arm) when not in use.
When not using for a long time, cover the unit with dust proof covers and ensure the unit is not
exposed to harsh environments
When re-starting after long (more than 3 months) storage, take the unit through a ‘seasoning’
procedure to ensure the tube head operates at its optimum. Following procedure explains the
seasoning process.
• Give 5 exposures
Page 29 of 73
SKANMOBILE‒ High Frequency
Chapter 5 Maintenance
Diagnostic X-Ray system
It is advised that the unit be subject to a maintenance schedule once every year after 1 st year of
usage
Do not open the unit. The unit can be refurbished only by a Skanray qualified technician.
Local governments would have rules and regulations on waste disposal of electronic goods. Please
follow the local guidelines.
Your distributor could also buy back the unit to be disposed off.
Contact the factory for shipping back units for disposal against relevant disposal charges.
Page 30 of 73
SKANMOBILE‒ High Frequency
Chapter 6 Troubleshooting
Diagnostic X-Ray system
6 Troubleshooting
6.1 Errors & Warnings
When generator is in fault mode, fault code and fault description will be displayed as “E” followed by 2 digit
fault code in kV DISPLAY and 3 digit fault description in mAs DISPLAY. Refer table below for complete list.
Page 31 of 73
SKANMOBILE‒ High Frequency
Chapter 6 Troubleshooting
Diagnostic X-Ray system
Listed below are the troubleshooting tips to help you recover from an error condition.
Error state with error code from Press RESET KEY. Generator will return to standby state.
1.
S / N 1 to S / N 20 If the problem persists, request service call
Switch off mains power. Wait for 2 minutes. Switch on mains
Error state with error code from
2. power.
S / N 21 to S / N 24
If the problem persists, request service call.
These faults are generated when defined operating
Error state with error code from specification/conditions are not met. These faults will be
3.
S / N 25 to S / N 30 automatically cleared if operating conditions comes back to
defined specifications.
Check if switch illuminates.
The unit does not power on when If switch is illuminating, then check the following.
4.
mains is switched on. Ensure that the mains cable connection to the power socket
in system is proper.
Page 32 of 73
SKANMOBILE‒ High Frequency
Chapter 7 Technical specifications
Diagnostic X-Ray system
7 Technical specifications
7.1 Tube-head Specifications
Generator Type : High Frequency, DSP Controlled, Constant Potential (DC)
Control of High Voltage : Closed Loop
High Voltage Range : 40 kV – 100 kV Settable (Step size 1 kV)
Accuracy of High Voltage : < ± 5%
Generator Frequency : Upto 200kHz
High Voltage Ripple : Low frequency ripple < 4%
Page 33 of 73
SKANMOBILE‒ High Frequency
Chapter 7 Technical specifications
Diagnostic X-Ray system
Page 34 of 73
SKANMOBILE‒ High Frequency
Chapter 7 Technical specifications
Diagnostic X-Ray system
Illustration 14: Reference axis to which target angle and focal spot characteristics of the tube head
assembly
Page 35 of 73
SKANMOBILE‒ High Frequency
Chapter 7 Technical specifications
Diagnostic X-Ray system
Page 36 of 73
Chapter Annex A: Exposure Time & mA List For Selected kV And mAs SKANMOBILE‒ High Frequency
(230 V /110 V) Diagnostic X-Ray system
Annex A: Exposure Time & mA List For Selected kV And mAs (230 V /110 V)
Definitions:
kV Set kV
mAs Set mAs
Exp-Time(mSec) Calculated X-Ray ON time for set mAs, in milliseconds
mA Calculated mA for set mAs
3 40 0.3 10 30 32 50 0.1 10 10
4 40 0.4 10 40 33 50 0.2 10 20
5 40 0.5 10 50 34 50 0.3 10 30
6 40 0.6 10 60 35 50 0.4 10 40
7 40 0.7 10 70 36 50 0.5 10 50
8 40 0.8 10 80 37 50 0.6 10 60
9 40 0.9 10 90 38 50 0.7 10 70
10 40 1 10 100 39 50 0.8 10 80
12 40 3 30 100 41 50 1 12.50 80
13 40 4 40 100 42 50 2 25 80
14 40 5 50 100 43 50 3 37.50 80
15 40 6 60 100 44 50 4 50 80
16 40 7 70 100 45 50 5 62.50 80
17 40 8 80 100 46 50 6 75 80
18 40 9 90 100 47 50 7 87.50 80
22 40 40 800 50 51 50 20 400 50
23 40 50 1000 50 52 50 30 750 40
24 40 60 1200 50 53 50 40 1000 40
25 40 70 1400 50 54 50 50 1250 40
26 40 80 1600 50 55 50 60 1500 40
27 40 90 1800 50 56 50 70 1750 40
Page 37 of 73
Chapter Annex A: Exposure Time & mA List For Selected kV And mAs SKANMOBILE‒ High Frequency
(230 V /110 V) Diagnostic X-Ray system
Page 38 of 73
Chapter Annex A: Exposure Time & mA List For Selected kV And mAs SKANMOBILE‒ High Frequency
(230 V /110 V) Diagnostic X-Ray system
Page 39 of 73
Chapter Annex A: Exposure Time & mA List For Selected kV And mAs SKANMOBILE‒ High Frequency
(230 V /110 V) Diagnostic X-Ray system
Note:
The above indicated mA and exposure time are valid for nominal input voltage of 230 Vac / 110 Vac
If the input voltage goes above 230 Vac / 110 Vac, mA increases and exposure time decreases for
corresponding mAs.
If the input voltage goes below 230 Vac / 110 Vac, mA decreases and exposure time increases for
corresponding mAs.
Page 40 of 73
Users' Manual
SKANMOBILE‒ High Frequency Diagnostic X-Ray system
Page 41 of 73
SKANMOBILE‒ High Frequency Diagnostic X-Ray
Basic X-Ray techniques used for Different Anatomy.
system
Pictorial
S/N Image Radiation Protection kV mAs FFD cm Patient position
Representation
The patient lies supine with the shoulder raised up on a small pad
about 2 inches and the head supported, the trunk and legs are
angled away from the side to be imaged which is placed safely near
Direct lead rubber gonad protection
Infro – the edge of the trolley or couch top. The arm of the affected side is
using a "half apron". Avoid irradiating
3 superior 55 12 100 abducted 90 degrees and the elbow flexed to aid external rotation
the thyroid and breast tissue as much
Supine and supported (a drip stand can be a useful aid). A small cassette
as possible.
is placed using a small sandbag above the skin surface of the
shoulder well into the root of the neck which is fixed towards the
opposite side
The patient is seated with their back to the couch or horizontal
Direct lead rubber gonad protection bucky, the trunk is rotated 30 degrees to the affected side so that
Modified
using a "half apron". Avoid irradiating the blade of the scapula is parallel to the edge of the table, if
4 axial ( 65 8 150
the thyroid and breast tissue as much possible the tip of the elbow rests on the edge of the table top. The
Wallace)
as possible cassette is placed horizontally behind the humerus and in contact
with the arm.
The patient stands erect facing the cassette and is rotated into the
60 degree anterior oblique position, of the side under investigation,
Direct lead rubber gonad protection
the radiographers hand placed on the posterior skin surface should
Scapula using a "half apron". Avoid irradiating
5 65 9 100 be at right angles to the film, the median saggital plane at
Lateral the thyroid and breast tissue as much
approximately 60 degrees to the cassette. The arm of the affected
as possible
side should be brought posterior to the elbow flexed slightly and the
back of the hand placed on the buttock of the affected side.
The patient stands erect facing the film, leans forward to bring the
Direct lead rubber gonad protection
clavicle as close as possible and parallel to the film, coronal plane
using a "half apron". Avoid irradiating
6 Clavicle PA 65 6 100 approximately 20 degree anterior angle, median saggital plane 10
the thyroid and breast tissue as much
degrees forward into the anterior oblique position of the side under
as possible
investigation.
Page 42 of 73
SKANMOBILE‒ High Frequency Diagnostic X-Ray
Basic X-Ray techniques used for Different Anatomy.
system
Pictorial
S/N Image Radiation Protection kV mAs FFD cm Patient position
Representation
Direct lead rubber gonad protection
using a "half apron". Avoid irradiating The patient stands erect AP median saggital plane 90 degrees to
the thyroid and breast tissue as much the film coronal plane parallel to the film. Take care to prevent the
as possible. Rather than irradiating the patient leaning backwards and hunching up the shoulder. Two
Acromio –
whole width of the chest some centres exposure one each side / both sides are made, the first pair in the
7 Clavicular 65 8 100
prefer individually centred collimated normal relaxed position and the second with the patient holding a
Joints AP
images of each area. However this 1.5 kg wt in each hand to stress the joint and accentuate any
does not demonstrate the medial end subuxation or an unstable joint. Weights suspended from wrist band
of clavicles which may be involved in rather than being held reduces shoulder hunching.
an injury
The patient stands erect AP; the position of the film is adjusted to
Direct lead rubber gonad protection include the shoulder and elbow joints. The patients arm is abducted
Humerus using a "half apron". Avoid irradiating approximately 20 degrees and externally rotated to bring the distal
8 65 7 100
AP the thyroid and breast tissue as much epicondyles equidistant from the film, the patient may need to be
as possible rotated to the affected side. The humerus may need to be
positioned diagonally across the film to include the full length.
The patient stands erect PA, the position of the film is adjusted to
include the shoulder and elbow joints, with the humerus in contact
with the cassette, i.e. rotated into the anterior oblique position
Direct lead rubber gonad protection
approximately 20 degrees. The arm is abducted approximately 35
Humerus using a "half apron". Avoid irradiating
9 48 6 100 degrees away from the trunk and the palm of the hand placed on
Lateral the thyroid and breast tissue as much
the iliac crest to bring an imaginary line through distal humeral
as possible
epicondyles at 90 degrees to the film, elbow in true lateral position.
The humerus may need to be positioned diagonally across the film
to include the full length
Page 43 of 73
SKANMOBILE‒ High Frequency Diagnostic X-Ray
Basic X-Ray techniques used for Different Anatomy.
system
Pictorial
S/N Image Radiation Protection kV mAs FFD cm Patient position
Representation
The patient sits alongside the end of the table with the affected arm
fully extended and the hand supinated, the table top should be
Direct lead rubber gonad protection raised to the level of the lower border of the axilla so the whole arm
Forearm using a "half apron". Ensure lower is supported and parallel to the cassette. The patient should lean
10 50 6 100
AP limbs and torso are not below table top externally to bring the humeral epicondyles equidistant from the
in primary beam film, in this position the wrist will be externally rotated from AP. It is
not possible to view proximal and distal ends in true AP position, it
is best to ensure that the area of most interest is projected correctly
The patient sits alongside the end of the table with the affected
Direct lead rubber gonad protection
elbow flexed at 90 degrees and the hand rotated externally into the
using a "half apron".
Forearm true lateral position, the table top should be raised to the level of the
11 Ensure the lower limbs and torso is 50 6 100
Lateral lower border of the axilla so the whole arm is supported and parallel
not below the table top in the primary
to the cassette. The hand may need to be supported on a small pad
beam.
to bring to superimpose the humeral epicondyles.
The patient sits alongside the end of the table with the affected arm
Direct lead rubber gonad protection fully extended and the hand supinated, the table top should be
using a "half apron". Ensure lower raised to the level of the lower border of the axilla so the whole arm
12 Elbow AP 48 6 100
limbs and torso are not below table top is supported and parallel to the cassette. The patient should lean
in primary beam externally to bring the humeral epicondyles equidistant from the
film, in this position the wrist will be externally rotated from AP
The patient sits alongside the end of the table with the affected
Direct lead rubber gonad protection elbow flexed at 90 degrees and the hand rotated externally into the
Elbow using a "half apron". Ensure lower true lateral position, the table top should be raised to the level of the
13 50 7 100
Lateral limbs and torso are not below table top lower border of the axilla so the whole arm is supported and parallel
in primary beam to the cassette. The hand may need to be supported on a small pad
to bring to superimpose the humeral epicondyles
Page 44 of 73
SKANMOBILE‒ High Frequency Diagnostic X-Ray
Basic X-Ray techniques used for Different Anatomy.
system
Pictorial
S/N Image Radiation Protection kV mAs FFD cm Patient position
Representation
The patient sits alongside the end of the table with the affected
Direct lead rubber gonad protection elbow flexed at 90 degrees and the hand rotated externally into the
Elbow
using a "half apron". Ensure lower true lateral position, the table top should be raised to the level of the
14 head of 55 6 100
limbs and torso are not below table top lower border of the axilla so the whole arm is supported and parallel
radius
in primary beam to the cassette. The hand may need to be supported on a small pad
to bring to superimpose the humeral epicondyles
Direct lead rubber gonad protection The patient sits alongside the long edge of the table arm extended
using a "half apron". Ensure lower palm downwards, or at the end of the table with the elbow flexed at
16 Wrist PA 46 3 100
limbs and torso are not below table top 90 degrees, hand and wrist flat on the cassette with the fingers
in primary beam flexed to maximize contact of the wrist with the cassette
Direct lead rubber gonad protection The patient sits alongside the long edge of the table arm extended
Wrist using a "half apron". Ensure lower palm downwards, or at the end of the table with he elbow flexed at
17 48 4 100
Lateral limbs and torso are not below table top 90 degrees, hand and wrist flat rotated external to bring the radio
in primary beam and ulna styloid processes in vertical alignment.
The patient sits alongside the long edge of the table arm extended
Direct lead rubber gonad protection palm downwards, or at the end of the table with he elbow flexed at
Wrist using a "half apron". Ensure lower 90 degrees, hand and wrist flat on the cassette with the fingers
18 45 5 100
Oblique limbs and torso are not below table top flexed to maximize contact of the wrist with the cassette. The wrist
in primary beam and forearm are then externally rotated 45 degrees and supported
on a foam pad.
Page 45 of 73
SKANMOBILE‒ High Frequency Diagnostic X-Ray
Basic X-Ray techniques used for Different Anatomy.
system
Pictorial
S/N Image Radiation Protection kV mAs FFD cm Patient position
Representation
Wrist The patient sits alongside the long edge of the table arm extended
Direct lead rubber gonad protection
Angled for palm downwards, or at the end of the table with he elbow flexed at
using a "half apron". Ensure lower
19 Scaphoid + 50 5 100 90 degrees, hand and wrist flat on the cassette with the fingers
limbs and torso are not below table top
Ulnar flexed to maximize contact of the wrist with the cassette. The hand
in primary beam
deviation and fingers are then maximally deviated laterally (ulna deviation).
The patient sits alongside the end of the table elbow flexed palm
Direct lead rubber gonad protection
Hand downwards, or at the end of the table with he elbow flexed at 90
using a "half apron". Ensure lower
20 Dorsi- 45 5 100 degrees, hand and wrist flat on the cassette the elbow may need a
limbs and torso are not below table top
Palmar small pad to promote contact of the hand with the cassette. The
in primary beam
fingers can be spread slightly
The patient sits alongside the end of the table elbow flexed palm
Direct lead rubber gonad protection
downwards, or at the end of the table with he elbow flexed at 90
Hand DP using a "half apron". Ensure lower
21 45 5 100 degrees, hand and wrist flat on the cassette the hand is then
Oblique limbs and torso are not below table top
rotated laterally 35 to 45 degrees laterally and supported so that the
in primary beam
fingers are parallel to the film
The patient sits alongside the end of the table elbow flexed and the
Direct lead rubber gonad protection hand in true lateral position the fingers are then flexed and the
Finger using a "half apron". Ensure lower finger in question extended parallel to the film, a radio lucent pointer
22 50 5 100
Lateral limbs and torso are not below table top may aid maintenance of the extension. For fingers other than the
in primary beam middle the hand is rotated to the lateral position bringing the finger
in question nearest the film or in contact with it for the 2nd and 5th
Page 46 of 73
SKANMOBILE‒ High Frequency Diagnostic X-Ray
Basic X-Ray techniques used for Different Anatomy.
system
Pictorial
S/N Image Radiation Protection kV mAs FFD cm Patient position
Representation
The PA position is easiest for the patient however the AP projection
produces less magnification and better detail as the thumb is in
contact with the cassette. PA Position, The patient sits alongside
the end of the table elbow flexed palm downwards, or at the end of
Direct lead rubber gonad protection
the table with he elbow flexed at 90 degrees, the hand is rotated
Thumb using a "half apron". Ensure lower
23 45 5 100 into the true later position projecting the thumb clear of
PA/AP limbs and torso are not below table top
superimposition with the fifth metacarpal. AP Position, The patient
in primary beam
stand with their back to the table and the arm is extended and
internally rotted to bring the palmar aspect of the thumb facing
upwards, adjust the table height to a comfortable position, adjust
the rotation of the arm to bring the thumb into true AP position
The patient sits alongside the end of the table elbow flexed palm
downwards, or at the end of the table with he elbow flexed at 90
Direct lead rubber gonad protection
degrees, hand and wrist flat on the cassette the elbow may need a
Thumb using a "half apron". Ensure lower
24 50 5 100 small pad to promote contact of the hand with the cassette. The
Lateral limbs and torso are not below table top
fingers can be spread slightly. The thumb is abducted and the hand
in primary beam
internally rotated and supported in position such that the thumb is
lateral and in contact with the cassette.
The patient lies supine on the table with the femur under
examination aligned to the midline of the table. The femur is aligned
25 Femur AP Direct lead rubber gonad protection 50 8 100 in slight internal rotation to bring the distal femoral condyles
equidistant from the film, ensure the knee joint is projected onto the
film
26 Femur Direct lead rubber gonad protection 52 10 100 (Non Trauma)The patient lies on the affected side and the upper
Lateral limb is flexed maximally and drawn up over and in front of the
affected limb and supported suitably. the knee of the affected side is
flexed 20 degrees to aid stability, and the long axis of the femur is
aligned to the long axis of the table.(Trauma cases require a
Page 47 of 73
SKANMOBILE‒ High Frequency Diagnostic X-Ray
Basic X-Ray techniques used for Different Anatomy.
system
Pictorial
S/N Image Radiation Protection kV mAs FFD cm Patient position
Representation
horizontal ray lateral)With the patient supine on the table and the
affected limb aligned to the long axis of the table. A small support is
placed under the knee to raise the knee off the table, the unaffected
leg is flexed 90 degrees at the hip and knee and supported
The patient lies supine on the table legs extended and the affected
limb aligned to the long axis of the table. The leg is rotated inwards
27 Knee AP Direct lead rubber gonad protection 48 6 100 until the femoral condyles are equidistant from the film or the patella
is midway between the femoral condyles. The tibia should be
parallel to the film
(Non Trauma)The patient lies on the affected side and the upper
limb is flexed maximally and drawn up over and in front of the
affected limb and supported suitably. The knee of the affected side
is flexed 45 degrees. The femoral condyles should be in vertical
Knee
28 Direct lead rubber gonad protection 50 7 100 alignment and the tibia parallel to the table. (Trauma cases require
Lateral
a horizontal ray lateral)With the patient supine on the table and the
affected limb aligned to the long axis of the table. A small support is
placed under the knee to raise the knee and support it in
approximately 25 degrees of flexion
Patient Prone, The patient lies prone on the table with the long axis
Knee of the leg aligned to the long axis of the table. The knee is flexed 45
29 Intercondyl Direct lead rubber gonad protection 55 7 100 degrees and the ankle supported. Patient supine, The patient sits
ar notch on the table with the long axis of the leg aligned to the long axis of
the table, the knee is flexed 45 degrees
Page 48 of 73
SKANMOBILE‒ High Frequency Diagnostic X-Ray
Basic X-Ray techniques used for Different Anatomy.
system
Pictorial
S/N Image Radiation Protection kV mAs FFD cm Patient position
Representation
The patient lies prone on the table with the leg aligned to the long
axis of the table, and supported so that the femoral condyles are
Knee PA Direct lead rubber gonad protection
30 55 7 100 equidistant from the film and the patella centred midway between
Patella using a "half apron".
the condyles. Note this position may note be possible on an injured
patient
The patient sits on the table with the long axis of the leg in line with
the long axis of the table, the affected limb is flexed to bring the
Knee Axial internal angle to 45 degrees, the tibia and fibula are in line with the
Direct lead rubber gonad protection
31 Patella 55 7 100 femur There are several variations to this projection in position of
using a "half apron".
"Skyline" the patient for minimizing radiation dose to the gonads by avoiding
directing the central ray in line with the rest of the body and comfort
and ease of performing the examination.
The patient lies supine on the table with the long axis of the tibia
and fibula in line with the long axis of the table. If the leg is
positioned in true anatomical normal position with neither the knee
Tibia & Direct lead rubber gonad protection
32 48 7 100 or ankle joint will be projected in the "normal " AP position, therefore
Fibula AP using a "half apron".
it is probably best to position the joint nearest the suspected
anomaly in correct AP position, i.e. when the knee is in true AP
position the ankle will be internally rotated more than normal
The patient lies supine on the table with the long axis of the tibia
and fibula in line with the long axis of the table and then rotates
externally to the affected side. If the leg is positioned in true
Tibia & anatomical normal position with neither the knee or ankle joint will
Direct lead rubber gonad protection
33 Fibula 50 7 100 be projected in the "normal " lateral position, therefore it is probably
using a "half apron".
Lateral best to position the joint nearest the suspected anomaly in correct
lateral position, i.e. when the knee is in true lateral position the
ankle will be externally rotated more than normal
Page 49 of 73
SKANMOBILE‒ High Frequency Diagnostic X-Ray
Basic X-Ray techniques used for Different Anatomy.
system
Pictorial
S/N Image Radiation Protection kV mAs FFD cm Patient position
Representation
The patient lies supine on the table legs extended, long axis of the
limb in line with the long axis of the table, the foot is dorsi flexed to
bring the plantar aspect 90 to the film. The patient then rotates
externally onto the affected side and flexes the knee slightly to aid
stability; a small pad under the knee will aid correct positioning. The
Ankle Direct lead rubber gonad protection
34 50 8 100 position is correct when the tibial and fibula malleoli are not in
Lateral using a "half apron".
vertical alignment but with the tibial malleolus approximately one
centimetre anterior to the fibula malleolus approximately. Note that
the foot is extremely mobile and rotating the foot externally may not
rotate the tibia and fibula, when rotating the ankle the whole leg has
to be rotated.
The patient lies supine on the table legs extended, long axis of the
limb in line with the long axis of the table, the foot is dorsi flexed to
bring the plantar aspect 90 to the film. The patient then rotates
externally onto the affected side and flexes the knee slightly to aid
Ankle AP & stability; a small pad under the knee will aid correct positioning. The
Direct lead rubber gonad protection
35 Oblique 48 6 100 position is correct when the tibial and fibula malleoli are not in
using a "half apron".
(mortice) vertical alignment but with the tibial malleolus approximately one
centimetre anterior to the fibula malleolus approximately. Note that
the foot is extremely mobile and rotating the foot externally may not
rotate the tibia and fibula, when rotating the ankle the whole leg has
to be rotated.
The patient sits on the table legs extended, long axis of the limb in
Ankle
line with the long axis of the table, the foot is dorsi flexed to bring
Axial for Direct lead rubber gonad protection
36 50 7 100 the plantar aspect 90 to the film, a bandage round the foot held by
Calcaneu using a "half apron".
the patient may help to maintain this dorsi-flexion of the foot. the
m
long axis of the foot is positioned vertically
Page 50 of 73
SKANMOBILE‒ High Frequency Diagnostic X-Ray
Basic X-Ray techniques used for Different Anatomy.
system
Pictorial
S/N Image Radiation Protection kV mAs FFD cm Patient position
Representation
The patients lies or sits on the table legs extended, the affected
Direct lead rubber gonad protection
37 Foot DP 45 6 100 limb is flexed at the knee to bring the plantar aspect of the foot in
using a "half apron".
flat contact with the cassette
The patients lies or sits on the table legs extended, the affected
Foot DP Direct lead rubber gonad protection limb is flexed at the knee to bring the plantar aspect of the foot in
38 48 7 100
Oblique using a "half apron". flat contact with the cassette, the foot is then rotated medially until
the dorsal skin surface is parallel to the cassette
The patient lies supine on the table legs extended, long axis of the
limb in line with the long axis of the table, the foot is dorsi flexed to
Foot Direct lead rubber gonad protection
39 48 7 100 bring the plantar aspect 90 to the film. The patient then rotates
Lateral using a "half apron".
externally onto the affected side and flexes the knee slightly to aid
stability; a small pad under the knee will aid correct positioning.
Page 51 of 73
SKANMOBILE‒ High Frequency Diagnostic X-Ray
Basic X-Ray techniques used for Different Anatomy.
system
Axial Skeleton
Pictorial
S/N Image Radiation Protection kV mAs FFD cm Patient position
Representation
The 28 day rule should be applied for female
patients. In males direct lead rubber gonad
The patient lies supine on the table with the mid-
protection can be applied at all times In females
saggital line in the center of the table. The legs are
direct lead rubber gonad protection is generally not
internally rotated approximately ten degrees by
1 Pelvis AP advised on the first examination, however if the 70 70 100
separating the heels 5 cm and bring the toes
patient is subsequently examined gonad protection
together in order to bring the femoral necks parallel
may be used if the area of interest is not obscured.
to the film to reduce foreshortening
On small patients the omission of a grid will reduced
the exposure
The 28 day rule should be applied for female
patients. In males direct lead rubber gonad
The patient lies supine on the table with the mid-
protection can be applied at all times In females
saggital line in the center of the table. The legs are
Hip joint direct lead rubber gonad protection is generally not
internally rotated approximately ten degrees by
2 AP advised on the first examination, however if the 60 50 100
separating the heels 5 cm and bring the toes
(implant) patient is subsequently examined gonad protection
together in order to bring the femoral necks parallel
may be used if the area of interest is not obscured.
to the film to reduce foreshortening
On small patients the omission of a grid will reduced
the exposure
The patient lies supine on the table, A.S.I.S.s
The 28 day rule should be applied for female
equidistant from the table top with the midsaggital
patients. In males direct lead rubber gonad
line in the center of the table, the affected limb is
protection can be applied at all times In females
turned inwards 10 degrees and supported using a
direct lead rubber gonad protection is generally not
sandbag or similar. The unaffected limb is flexed 90
Pelvis Hip advised on the first examination, however if the
3 70 60 100 degrees at the hip and knee and supported. A
lateral patient is subsequently examined gonad protection
24×30 cm grid is placed upright with the medial
may be used if the area of interest is not obscured.
side well into the waist and parallel to the neck of
On small patients the omission of a grid will reduced
the femur. A small pad under the patients buttocks
the exposure
raising the of the table may ease positioning
Page 52 of 73
SKANMOBILE‒ High Frequency Diagnostic X-Ray
Basic X-Ray techniques used for Different Anatomy.
system
Pictorial
S/N Image Radiation Protection kV mAs FFD cm Patient position
Representation
The 28 day rule should be applied for female
patients. In males direct lead rubber gonad The patient lies supine on the table, A.S.I.S.s
protection can be applied at all times In females equidistant from the table top with the midsaggital
Pelvis Hips direct lead rubber gonad protection is generally not line in the center of the table. The hips and knees
4 "Frogs" advised on the first examination, however if the 70 60 100 are flexed and the soles of the feet placed together
Lateral patient is subsequently examined gonad protection in the midline, the knees are forced down towards
may be used if the area of interest is not obscured. the table top as far as possible. Ensure leg
On small patients the omission of a grid will reduced positioning is symmetrical
the exposure
The 28 day rule should be applied for female The patient lies supine on the table with the
patients. In males direct lead rubber gonad midsaggital line in the centre of the table. The
Pelvis Hip protection can be applied at all times In females affected side is raised 45 degrees and supported
Judets direct lead rubber gonad protection is generally not on pads with the legs extended or minimally flexed
5 view of advised on the first examination, however if the 80 50 100 to aid support. A second projection may be made
Acetabulu patient is subsequently examined gonad protection with the patient rotated 45 degrees onto the
m(1) may be used if the area of interest is not obscured. affected side and the central ray directed to a point
On small patients the omission of a grid will reduced midway along an imaginary line from the ASIS to
the exposure the symphysis on the affected side
The 28 day rule should be applied for female
patients. In males direct lead rubber gonad
Pelvis protection can be applied at all times In females
Leonard direct lead rubber gonad protection is generally not
6 George advised on the first examination, however if the 80 50 100
lateral patient is subsequently examined gonad protection
may be used if the area of interest is not obscured.
On small patients the omission of a grid will reduced
the exposure
Page 53 of 73
SKANMOBILE‒ High Frequency Diagnostic X-Ray
Basic X-Ray techniques used for Different Anatomy.
system
Pictorial
S/N Image Radiation Protection kV mAs FFD cm Patient position
Representation
The patient lies supine on the table with the long
The 28 day rule should be applied for female
axis of the body in line with the centre of the table,
patients. In males direct lead rubber gonad
Sacrum legs and knees extended. The anterior superior iliac
7 protection can be applied at all times On small 70 60 100
AP spines should be equidistant from the table
patients the omission of a grid will reduced the
ensuring there is no rotation of the mid saggital
exposure.
plane
The 28 day rule should be applied for female The patient lies on the table with the mid saggital
Sacrum & patients. In90 males direct lead rubber gonad plane parallel to the table, support under the waist
8 Coccyx protection can be applied at all times On small 85 80 100 may be required. The long axis of the patient
Lateral patients the omission of a grid will reduce the should be in line with the long axis of the table, the
exposure. hips and knees flexed to aid comfort and stability
Page 54 of 73
SKANMOBILE‒ High Frequency Diagnostic X-Ray
Basic X-Ray techniques used for Different Anatomy.
system
Pictorial
S/N Image Radiation Protection kV mAs FFD cm Patient position
Representation
The patient lies on the table with the mid saggital
Lumbar plane parallel to the table, support under the waist
The 28 day rule should be applied for female
Sacral and a pad between the knees may be required. The
11 patients. In males and females direct lead rubber 75 45 100
articulation long axis of the patient should be in line with the
gonad protection can be applied at all times
Lateral long axis of the table, the hips and knees flexed to
aid comfort and stability.
Page 55 of 73
SKANMOBILE‒ High Frequency Diagnostic X-Ray
Basic X-Ray techniques used for Different Anatomy.
system
Pictorial
S/N Image Radiation Protection kV mAs FFD cm Patient position
Representation
The patient lies supine on the table midline aligned
to the midline of the table, the ASISs and the
Acromio clavicular joints are equidistant from the
The 28 day rule should be applied for female
Thoracic table ensuring the midsaggital plane is at 90
15 patients. In males and females direct lead rubber 66 30 100
Spine AP degrees to the table. The neck is extended to avoid
gonad protection can be applied at all times
superimposition of the mandible on the upper
thoracic spine, flexion of the hips and knees may
help reduce the thoracic curve
The patient lies on one side on the table, mid-
saggital plane parallel to the table top and mid
coronal plane midline aligned to the midline of the
Thoracic The 28 day rule should be applied for female
16 75 50 100 table, the hips and knees flexed 45 degrees to
Spine patients. In males and females direct lead rubber
support the patient and minimise movement, raising
Lateral gonad protection can be applied at all times
the arms well above the head maximises
visualisation of the upper vertebrae
The patient stands erect with the median saggital
Thoracic plane parallel to the bucky face, the arm nearest
The 28 day rule should be applied for female
Spine the bucky is raised the elbow flexed and the
17 patients. In males and females direct lead rubber 75 50 100
Upper forearm rested on the top of the head. The other
gonad protection can be applied at all times
Lateral arm is positioned hanging loosely at the patients
side, a weight may help lower the shoulder
Page 56 of 73
SKANMOBILE‒ High Frequency Diagnostic X-Ray
Basic X-Ray techniques used for Different Anatomy.
system
Pictorial
S/N Image Radiation Protection kV mAs FFD cm Patient position
Representation
The patient stands erect in the AP position median
saggital plane at 90 degrees to the film with the
In males and females direct lead rubber gonad patient’s midline in line with the center of the bucky,
Cervical
protection can be applied at all times, avoid the feet slightly apart to aid stability, coronal plane
18 spine 1-2 65 12 100
breast tissue particularly in young females; collimate parallel to the film. The mouth is opened wide, chin
AP
to prevent irradiation of the eyes. is raised until an imaginary line from the center of
the mouth to a point level with the mastoid tips is
horizontal
The patient stands erect in the AP position median
Cervical saggital plane at 90 degrees to the film with the
In males and females direct lead rubber gonad
spine 3-7 patient’s midline in line with the center of the bucky,
protection can be applied at all times, avoid the
19 AP 65 12 100 feet slightly apart to aid stability, coronal plane
breast tissue particularly in young females; collimate
(Swimmer' parallel to the film. The chin is raised until an
to prevent irradiation of the eyes.
s View) imaginary line from the upper lip to the external
occipital protuberance is horizontal
Page 57 of 73
SKANMOBILE‒ High Frequency Diagnostic X-Ray
Basic X-Ray techniques used for Different Anatomy.
system
Pictorial
S/N Image Radiation Protection kV mAs FFD cm Patient position
Representation
In males and females direct lead rubber gonad The patient stands or sits facing the bucky then
Cervical
protection can be applied at all times, avoid the turns 45 degrees to bring the median saggital plane
21 Anterior 68 14 100
breast tissue particularly in young females; collimate at 45 degrees to the bucky, the head is then rotated
Oblique
to prevent irradiation of the eyes. parallel to the film, with chin is raised slightly.
Page 58 of 73
SKANMOBILE‒ High Frequency Diagnostic X-Ray
Basic X-Ray techniques used for Different Anatomy.
system
SKULL: Supine Trolley Technique
Pictorial
S/N Image Radiation Protection kV mAs FFD cm Patient position
Representation
Skull The patient lies supine on the table midsaggital plane aligned
Lateral Collimate to limit irradiation of the central to the table, the head is supported on a small pad, side of
3 75 32 75
(Horizontal thyroid gland interest nearest the film. the median saggital plane must be
Ray) parallel to the film which is placed along side the head
Page 59 of 73
SKANMOBILE‒ High Frequency Diagnostic X-Ray
Basic X-Ray techniques used for Different Anatomy.
system
Pictorial
S/N Image Radiation Protection kV mAs FFD cm Patient position
Representation
place vertical above the vertex as in the picture below
Skull The patient lies supine on the trolley the midsaggital plane
Facial- aligned to the long axis of the trolley, interpupilary line parallel to
Collimate to limit irradiation of the
5 bones 65 45 100 the floor. The cassette is supported alongside the affected side of
thyroid gland
Lateral / the face parallel to the midsaggital plane. the chin is raised to
Mastoid bring the orbital meatal line vertical
Page 60 of 73
SKANMOBILE‒ High Frequency Diagnostic X-Ray
Basic X-Ray techniques used for Different Anatomy.
system
Erect Bucky Technique
Pictorial
S/N Image Radiation Protection kV mAs FFD cm Patient position
Representation
The patient sits erect facing the bucky, midsaggital plane in the
Skull
Avoid irradiation of the thyroid region, midline of the film,interpupilary line parallel to the floor. The chin
2 Reverse 75 40 100
direct lead rubber protection my be used is lowered 30 degrees from the PA position to bring the orbital
Townes
meatal line at 60 degrees to the film
Take care with this technique, the patient may suffer vaso vagal
reactions or vertigo, it is not suitable for trauma patients. The
patient sits erect with their back to the erect bucky, a small pillow
Direct lead rubber thyroid protection
3 Skull SMV 75 45 100 is placed behind the shoulders and the patient extends the neck
advisable
until the orbital meatal baseline is parallel to the film, the
interpupilary line parallel to the floor and the median saggital
plane at 90 degrees to the film
Page 61 of 73
SKANMOBILE‒ High Frequency Diagnostic X-Ray
Basic X-Ray techniques used for Different Anatomy.
system
Pictorial
S/N Image Radiation Protection kV mAs FFD cm Patient position
Representation
Take care with this technique, the patient may suffer vaso vagal
reactions or vertigo, it is not suitable for trauma patients. The
PNS patient sits erect with their back to the erect bucky, a small pillow
Direct lead rubber thyroid protection
4 water's 60 55 100 is placed behind the shoulders and the patient extends the neck
advisable
view until the orbital meatal baseline is parallel to the film, the
interpupilary line parallel to the floor and the median saggital
plane at 90 degrees to the film
Page 62 of 73
SKANMOBILE‒ High Frequency Diagnostic X-Ray
Basic X-Ray techniques used for Different Anatomy.
system
Pictorial
S/N Image Radiation Protection kV mAs FFD cm Patient position
Representation
The patient sits erect facing the erect film, the head is then
turned so that the side of the face is in contact with the bucky,
Nose Direct lead rubber thyroid protection
9 55 2 100 median saggital plane parallel to the film and interpupillary line
Lateral advisable.
parallel to the floor. This position requires the patient to sit upright
as close as possible to the bucky.
Page 63 of 73
SKANMOBILE‒ High Frequency Diagnostic X-Ray
Basic X-Ray techniques used for Different Anatomy.
system
Pictorial
S/N Image Radiation Protection kV mAs FFD cm Patient position
Representation
Page 64 of 73
SKANMOBILE‒ High Frequency Diagnostic X-Ray
Basic X-Ray techniques used for Different Anatomy.
system
Chest
Pictorial
S/N Image Radiation Protection kV mAs FFD cm Patient position
Representation
The patient stands (feet slightly apart for stability) erect left
(normally unless right lateral is indicated) side to the film / bucky,
Direct lead rubber protection using a half median saggital plane parallel to the film. The patient may need
Chest apron around the waist. Correct to flex forward from the waist to ensure the limits of the
2 70 10 100
Lateral centering should avoid irradiation of the examination fit on the film. The patients arms are folded across
thyroid gland the top of the head and the elbows gripped with the hand of the
opposite side. The chin is raised up. The top of the film should be
positioned about 3 cm above the tops of the shoulders
The patient stands erect facing the bucky, from here the patient is
rotated 45degrees with the affected side away from the film, the
Chest arm of the affected side is raised and placed across the head
Anterior Direct lead rubber waist level protection, which may need to be turned towards the affected side for
3 oblique, posterior at low kV, anteriorly at higher 70 10 100 comfort and ease of positioning. Ensure the opposite site anterior
Ribs kV chest wall is in contact with the bucky to aid immobilization. The
(Axillary) erect anterior oblique position reduces the dose to breast tissue
compared with the posterior oblique, however the ribs are further
from the film which may compromise image quality
Page 65 of 73
SKANMOBILE‒ High Frequency Diagnostic X-Ray
Basic X-Ray techniques used for Different Anatomy.
system
Pictorial
S/N Image Radiation Protection kV mAs FFD cm Patient position
Representation
Sternum
The patient lies prone on the table with the left side raised (RAO)
PA Rt Direct lead rubber waist level protection,
4 70 30 100 approximately 15 degrees, arm and knee of the raised side to aid
Anterior avoid irradiation of thyroid
stability
Oblique
Page 66 of 73
Users' Manual
SKANMOBILE‒ High Frequency Diagnostic X-Ray system
EN/IEC 60601-1 ed3.0:2005+CORR.1 Medical electrical equipment – Part 1: General requirements for
(2006)+CORR.2 (2007) basic safety and essential performance
Page 67 of 73
SKANMOBILE‒ High Frequency
Declaration of Conformity
Diagnostic X-Ray system
C3: Marking:
The products described herein are conform to the following regulatory markings
a) UL safety marking.
b) CE marking.
C4: Declaration
Skanray Technologies Private limited declares that the products described herein meet all the applicable
Essential Requirements of the EC Medical Device Directive 93/42/EEC in Annex I, For Class IIb products
described herein, the product is manufactured, inspected, tested, and released in accordance with the
approved quality assurance system established in accordance with ISO 9001:2008, ISO 13485:2003
Annex II of the EC Medical Device Directive under the supervision of the UL INTERNATIONAL (UK) LTD,
a Notified Body carrying the Notified Body No. 0843.
Page 68 of 73
Users' Manual
SKANMOBILE‒ High Frequency Diagnostic X-Ray system
SKANMOBILE-XXX is tested as per applicable IEC standards, to be used under electromagnetic environment
specified below. The customer or the user of SKANMOBILE-XXX should assure that it is used in such an
environment. Where XXX Variants = 230 Vac, 110 Vac
Electromagnetic
Immunity test EN 60601 test level Compliance level
environment – guidance
Electrical fast ± 2 kV for power supply ± 2 kV for power supply Mains power quality should be
transient/burst lines lines that of a typical commercial or
EN 61000-4-4: 2004 ± 1 kV for Signal lines ± 1 kV for Signal lines hospital environment.
interruptions and (> 95 % dip in UT) (> 95 % dip in UT) Mains power quality should be
voltage variations for 0,5 cycle for 0,5 cycle that of a typical commercial or
hospital environment. If the
on power supply 40 % UT 40 % UT
user of the SKANMOBILE-XXX
input lines (60 % dip in UT) (60 % dip in UT)
requires continued operation
EN 61000-4-11: for 5 cycles for 5 cycles
Page 69 of 73
SKANMOBILE‒ High Frequency
Guidance and Manufacturer’s Declaration
Diagnostic X-Ray system
Electromagnetic
Immunity test EN 60601 test level Compliance level
environment – guidance
70 % UT 70 % UT
during power mains
(30 % dip in UT) (30 % dip in UT) interruptions, it is
for 25 cycles for 25 cycles recommended that the
2004
< 5 % UT < 5 % UT SKANMOBILE-XXX be
Power frequency
Power frequency magnetic
(50/60 Hz) fields should be at levels
magnetic field characteristic of a typical
Table 2: Guidance and Manufacturer’s Declaration – Electromagnetic Immunity – For all EQUIPMENT and SYSTEMS
Page 70 of 73
SKANMOBILE‒ High Frequency
Guidance and Manufacturer’s Declaration
Diagnostic X-Ray system
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which
SKANMOBILE-XXX is used exceeds the applicable RF compliance level above, the SKANMOBILE-XXX
should be observed to verify normal operation. If abnormal performance is observed, additional measures may
be necessary, such as reorienting or relocating the SKANMOBILE-XXX.
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Table 3: Guidance and manufacturer’s declaration – electromagnetic immunity – for all EQUIPMENT and SYSTEMS that are
not LIFE-SUPPORTING
Page 71 of 73
Users' Manual
SKANMOBILE‒ High Frequency Diagnostic X-Ray system
e-mail : office@skanray.com
Web : www.skanray.com
24 Hour Hotline : +91 9901144411 (Customer Complaint and Incident Reporting Only)
e-mail : inquiries@skanrayhealthcare.com
Web : www.skanrayhealthcare.com
1800-233-5858
Page 73 of 73
Title
Customer Feedback Form
Department: Sales Format # 70-0001 Rev 3 Page 1of 2
Notes: 1) * Details are essential. 2) Attach additional sheet if required. 3) Read all instructions before filling up.
4) Completed format can be sent to Skanray Technologies Pvt Ltd Plot No. 15-17, Hebbal Industrial Area
Mysore 570 016, India or visit website www.skanray.com “Contact Us” web page for posting your feedback
online. 5) For any more clarification please contact Skanray Technologies.
a) Feedback by
Customer Name*
Street: City:
Customer Address* State: Pin/Zip Code: Country:
b) Product details
Product / (Full System) Type/Model*
Product Serial Number*
Product Application
Installed Site Address
c) *Feedback
Interaction with Sales team/Distributors* Fully Satisfied Satisfied Unsatisfied
Product Condition when received* Fully Satisfied Satisfied Unsatisfied
Technical details for product Installation (if applicable) Fully Satisfied Satisfied Unsatisfied
Technical details for product Operation* Fully Satisfied Satisfied Unsatisfied
Product meets Intended use & performance* Fully Satisfied Satisfied Unsatisfied
Technical details for product maintenance* Fully Satisfied Satisfied Unsatisfied
Response from service team* Fully Satisfied Satisfied Unsatisfied
Quality of service* Fully Satisfied Satisfied Unsatisfied
Notes if any:
d) Receipt of feedback
*Receiver Name: *Signature: *Date:
(c) Feedback:
Check respective check boxes for each feedback points. Feed back can be sent after actual usage of 1 to 3
months.
Notes: This field is for recording any relevant information provided by the customer.