Infrared Thermometer: Description

You might also like

Download as docx, pdf, or txt
Download as docx, pdf, or txt
You are on page 1of 3

MS6520B

Description Display
Noncontact Themometer detects the infrared ray that Panel D:S
Infrared Thermometer an object emits, The instrument foculizes infrared
energy of the object onto a sensor through a lens,
ratio
1 2 3 4
changes the surface temperature into electric signal, a
microcomputer calculates and displays the Unit: mm
measurement temperature on
the LCD.
6 1500 300
900
• Single-spot Laser Sighting 7 13
• Backlit Display 2 5
5 S
• Current Temperature Plus MIN, MAX, 12 30
AVG Temperature Displays 6 150 90
• Preset Emissivity 0.95 8 D
11
Warning 3 D (distance):S (spot size) = 10:1
• Please read the following information 4
carefully before using the meter.Protection is
impaired if used in a manner not specified in Make sure that the target is larger than the meter's
this manual 10
9 visual spot size. The smaller the target, the closer you
• Do not clear the meter using solvents. should be to it. The relationship between distance and
• Keep the instrument clean, and do not get dust 9
spot size is 10:1
into detecting hole.
• Do not point laser directly at eye or indirectly 8 7
off reflective surfaces.
• Laser :class 2 <1mW/630-670nm 1 Laser indicator 8 Min value
•Laser radiation is classified according to
IEC 60825-1: 2014-05, Safety of laser products – 1 Screen 6 Sensor 2 Scan status 9 Max value
Part 1: Equipment classification and requirements. 3 Emissivity 10 Average value
2 Mode key 7 Laser
CAUTION 4 Low battery 11 Primary display
LASER RADIATION - DO NOT STARE INTO BEAM. 3 Backlight key 8 Trigger
AVOID EXPOSURE - LASER RADIATION IS EMTTED FROM THIS APERTURE . 5 °F unit 12 Surface scan
MAX OUTPUT<1mW, WAVELENGTH 630~670nm, CLASS 2 LASER PRODUCT 4 Laser key 9 Battery cover
6 °C unit 13 Hold status
WARNING 5 °C/°F key
Laser radiation -
laser 7 Secondary display
when open do not
stare into beam.

01 02 03 04
Operating Thermometer Note1
Make sure that the target is larger than the meter's
Note3 Specifications Warranty
• To measure temperature, pull and hold trigger. visual spot size When the battery voltage is lower, the battery symbol This produt will be free from defects in material
Release the trigger to hold a temperature appears, it indicates that we must change the battery. and workmanship for eighteen months from the
reading. The meter turns off when no activity is date of purchase.this warranty does not cover
detected for ten seconds. disposable batteries or damage for
• Press key to toggle between the MAX accident,neglect,misuse alteration
,MIN, AVG value. conntamination,or abnormal conditions of
• Press key to Active the laser. operation ro handling.
• Press key to light on or off the backlight.
• Press key to switch between °C or °F. OK

Note2
When ambient temperature changes quickly, must
wait 30 minutes to balance the temperature of the
55 instrument before use.

22 ºC 0 ºC
72 ºF 32 ºF
12

9 3
30Min
6

WARNING

When the battery voltage is lower, the battery symbol


appears, it indicates that we must change the battery.

05 06 07 08 09
HYS006762

Temperat
ure range -20°C~537°C or -4º F~932ºF

Accuracy
: -20°C~0°C(±3°C)
Assumes
ambient 0°C~500°C
Operatin
g ±(1.5%rdg+2°C)
temperat
ure
(23°C ±
2°C)
Optical Spectr
10:1 al 8-14 um
resolut
ion respo
nse
Repeatabi 1°C Response Approx.1s
lity time
Resoluti 0.1°C/0.2 Emissivity 0.95
on °F
Ambie
nt Relati
0~40°C 10%~90%
Operat ve
ing humi
Range dity
Conform Laser <1m
s to W 630-
Standards EN61326 670nm

You might also like