Before Contacting Us
Before Contacting Us
check the following when errors are found in the unit. Contact nearest branch or agency if
you cannot figure out the causes or if the errors are not listed in the following table.
The switch light is not lit in The battery for power failure Turn on the power breaker.
spite of pressing the I/O backup is low in charge and
switch the power breaker is turned
off.
Did not press the I/O switch Press the I/O switch for 0.5
for 0.5 consecutive sec or consecutive sec or longer.
longer
Nothing or only a part of a Backlight is turned off. Press the switch on the
display appears on the operation panel
message display monitor.
The connector to be Connect the connector
connected to the message properly.
display monitor is off.
The dialysis temperature The water supply Set the water supply
does not increase to the temperature is low temperature properly Refer
setting temperature to “21. Specification”.
The dialysis temperature The water supply Set the water supply
does not decrease to the temperature is high temperature at least the
setting temperature. setting temperature of the
unit °C.
The rinse movement does The DIA.MODE switch is Press the DIA.MODE
not start. ON. switch to proceed into the
rinse waiting process. Press
the RINSE switch once.
The system is in the rinse Press the RINSE switch
waiting process. once.
The blood pump does not The blood pump cover is Close the blood pump cover
rotate (when power is on or open. firmly.
in case of power failure).
The blood pump flow Turn the blood pump flow
display is “0 mL/min”. setting dial to the right to set
to the desired rate.
The blood pump does not The battery for power failure Charge the battery (energize
rotate (in case of power backup is low in charge. the unit for 48 hours or
failure). longer).
The heparin pump does not The heparin pump flow Set the heparin pump follow
rotate (when power is on or display is “0.00 mL/h”. as you desire.
in case of power failure).
The blood pump stopped. Move the blood pump.
The heparin pump does not The battery for power failure Charge the battery (energize
rotate (in case of power backup is low in charge. the unit for 48 hours or
failure). longer).
During the bicarbonate The stock solution Use the stock solution with a
dialysis, the actual dialysate concentration is high. proper concentration
concentration is higher than
the setting concentration The settings of the stock Set the dialysis type mode
solution to be used and the properly.
dialysis type mode are
different.
During the bicarbonate The stock solution Use the stock solution with a
dialysis, the actual dialysate concentration is low. proper concentration.
concentration is lower than
the setting concentration. The stock solution is not Make the air to flow better
aspirated sufficiently by loosening the cap for the
because the stock solution stock solution tank, etc.
tank is closed.
During the acetate dialysis, The stock solution Use the stock solution with a
the actual dialysate concentration is high. proper concentration
concentration is higher than
the setting concentration. The settings of the stock Set the dialysis type mode
solution to be used and the properly.
dialysis type mode are
different.
During the acetate dialysis, The stock solution Use the stock solution with a
the actual dialysate concentration is low. proper concentration.
concentration is lower than
the setting concentration. The settings of the stock Set the dialysis type mode
solution to be used and the properly.
dialysis type mode are
different.
Switch operation does not A switch whose operation Some switch operations do
work. was not accepted by the not work depending on a
system was pressed. screen display or process.
Read the instruction manual
again before operation.