Gas Chromatography-Mass Spectrometry: Glossary

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 7

Gas Chromatography–Mass Spectrometry

E Stauffer, Commissariat d’identification judiciaire, Police cantonale Fribourg, Fribourg, Switzerland


ã 2013 Elsevier Ltd. All rights reserved.

Glossary compounds, and to elucidate the structure and chemical


Chromatogram A graphical representation of properties of molecules. This is done by converting
chromatographic data. It is the record of detector response as components of a sample into rapidly moving gaseous ions
a function of time. and separating them on the basis of their mass/charge ratio.
Extracted ion chromatogram Graphical representation of Abbreviated MS.
the abundance of a single ion versus retention time (RT). Resolution In chromatography, the measure of the
Abbreviated EIC. separation of two components. It is a function of the peak
Extracted ion profile Graphical representation of the width and RT. In mass spectrometry, the measure of the
abundance of the sum of several ions versus RT. Abbreviated ability of the instrument to distinguish ions of different
EIP. masses.
Gas chromatography A separation technique in which the Retention time The length of time required for a compound
mobile phase is a gas. Gas chromatography is always carried to pass through a chromatographic column and be detected.
out in a column. Abbreviated GC. Total ion chromatogram Graphical representation of the
Gas chromatography–mass spectrometry A separation abundance of a full scan range versus RT. Abbreviated TIC.
technique (gas chromatography) coupled with a detection Two-dimensional comprehensive gas
technique (mass spectrometry). Abbreviated GC–MS. chromatography A procedure in which all of the separated
m/z ratio Mass-to-charge ratio: The mass of an ion divided sample components are subjected to an additional
by its charge. separation step. This is achieved by coupling (using a
Mass spectrometry An analytical technique that is used to thermal modulator) two different chromatographic
identify unknown compounds, to quantify known columns. Abbreviated GCGC.

Abbreviations MS Mass spectrometry or mass spectrometer


EIC Extracted ion chromatogram m/z Mass-to-charge ratio
EIP Extraction ion profile SIM Selected ion monitoring
FID Flame ionization detector TIC Total ion chromatogram
GC Gas chromatography or gas chromatograph TOF Time-of-flight
GCGC Two-dimensional comprehensive gas
chromatography

Introduction Mass spectrometry does not simply detect the presence or


absence of analytes; it provides structural information about
Chromatographic techniques have the power to separate the molecule, often leading to its identification. As such, it is
analytes from complex mixtures. However, separating these an extremely powerful instrument, but one which quickly
analytes would not be useful if it was not possible to either becomes impractical when analyzing complex mixtures of ana-
detect them or, better yet, to identify them. For this reason, lytes. As a matter of fact, its optimal use consists in analyzing
different detectors have been used throughout the years. With pure compounds. This is when the chances of identifying the
simple thin-layer chromatography, or paper chromatography, unknown compounds are greatest.
the naked eye, or an alternate light source, would suffice to Thus, the separation power of gas chromatography and the
detect some analytes. Eventually, a chemical reaction would be analytical power of mass spectrometry needed to be combined
triggered. On the more modern gas chromatograph, flame in one full analytical instrument in order to exploit the capa-
ionization (FID) or photoionization detectors are used, bilities of both techniques to their maximum. Chromatogra-
depending on the type of analytes to be detected. phy was invented in 1906 by Russian chemist Mikhail
However, the common ground of all these detectors is that Semenovich Tswett and mass spectrometry in 1913 by British
they only provide two types of information: the number of physicist Joseph John Thomson. Gas chromatography was
analytes and their concentration. Nowadays, particularly in the invented in 1952 by James and Martin, which allowed the
forensic field, this information does not suffice. One needs to first combined use of gas chromatography and mass spectrom-
know the identity of a compound or of a series of compounds etry as a hyphenated technique in 1956 by Roland Gohlke and
from a complex mixture. Fred McLafferty.

596 Encyclopedia of Forensic Sciences, Second Edition http://dx.doi.org/10.1016/B978-0-12-382165-2.00249-X


Methods | Gas Chromatography–Mass Spectrometry 597

The typical setup of a gas chromatograph–mass spectrome- of information. For each point on the curve, it is possible to
ter is shown in Figure 1. The sample is injected in the gas extract a mass spectrum, providing the structural information
chromatograph, separated through the column, then goes of the analyte(s) eluting at that time. Figure 2 illustrates this.
through the transfer line and is analyzed in the mass spectrom- This information consists of a series of mass-to-charge ratio
eter. A computer controls the overall instrument. (m/z) peaks with their relative abundance. The most important
Gas chromatography–mass spectrometry (GC–MS) does of these peaks is the molecular ion peak, which corresponds
not simply add the capability of one to another; it literally to the unfragmented compound, and the base peak, the most
multiplies both capabilities and leads to a full analytical in- intense peak of the spectrum, normalized to 100%. The other
strument. While the specificities of gas chromatography alone peaks represent the fragments of the molecule, fragments with
(with a simple detector such as FID) and mass spectrometry distribution and relative abundances that are characteristic of
alone are not great, the combination of both results in an the structure of the molecule.
enormous increase in specificity. If two compounds were not As such, in addition to RT data, one benefits from structural
to be separated by the chromatographic process, there is a good data to identify an analyte. If each analyte in the complex
chance that they would exhibit different structural data and, mixture is completely separated from the others, it is then pos-
thus, be distinguished through the mass spectrometer. Simi- sible to identify with certainty the nature of each analyte. This is
larly, two compounds exhibiting a very close mass spectrum where the power of GC–MS comes into play: the capability of
may have a good chance of being separated by the chromato- identifying almost every component from a complex mixture.
graphic process, thus allowing for their distinction. However,
one will always find compounds that have very close chro-
matographic and structural characteristics, for which GC–MS Different Ways of Looking at Data
will not be able to make a distinction. This is the case with
some structural isomers. One great advantage of GC–MS lies in the different ways it is
possible to acquire and look at data. Typically, one acquires
data using a full scan range, which means that the mass spec-
Multiplied Powers trometer is configured to acquire all the m/z within a given
range (e.g., 33–400 m/z). The resulting data is called a total ion
When considering a regular chromatogram, obtained with a chromatogram (TIC). It can be seen as a chromatogram iden-
simple detector such as FID, the sole information available is tical to one that would be obtained with another simple detec-
the curve constituting the chromatogram. One can extract the tor, except that each data entry making the chromatographic
retention time (RT) and the peak area, thus the concentration pattern contains a full mass spectrum.
of the analyte if a calibration curve is available. However, it is The great advantage of the TIC is that all the analytes
very possible that another analyte elutes exactly at the same present in a given sample are detected and recorded. Thus,
time as the analyte of interest. As such, it is not possible to be the chromatogram truly and fully represents the overall con-
sure that the peak in question is the sought-after analyte. tent and the analyst has a good panoramic view of what is in
When considering a chromatogram obtained from a the sample. A few disadvantages also exist. The first one is that
GC–MS, the resulting data actually hides another whole layer carrying out a full scan takes time and the resolution of

Pressure regulator

Injection port

Inlet

Flow Gas filters Carrier


controller
gas
supply
Column

Mass Oven
Computer spectrometer Gas chromatograph
controller and
data acquisition
Transfer line
Figure 1 Typical setup of a gas chromatograph–mass spectrometer. Reproduced from Stauffer E, Dolan JA, and Newman R (2008) Fire Debris
Analysis, p. 246. Burlington, MA: Elsevier Academic Press. Copyright ã 2008, Elsevier.
598 Methods | Gas Chromatography–Mass Spectrometry

105
100
1,2,4-trimethylbenzene
Abundance

Relative abundance
80

60
120
40

20 77 91
51 65
0
0 50 100 150
3 000 000 Mass-to-charge ratio (m/z)

43
100
Decane
57

Relative abundance
80

60

40 29
71
85
2 000 000 20
99 142
0
0 50 100 150
Mass-to-charge ratio (m/z)

128
100 Naphthalene

Relative abundance
80

1 000 000 60

40

20
51
0
0 50 100 150
Mass-to-charge ratio (m/z)

0
2 3 4 5 6 7 8 9 10 11 12 13 14 15 Time
Carbon # 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25

Figure 2 Total ion chromatogram of a 75% evaporated gasoline sample. Each point on the chromatogram contains a full mass spectrum,
as shown with the peaks of 1,2,4-trimethylbenzene, decane and naphthalene.

the chromatogram may not be optimal, thus not allowing for a matter of fact, it is important to have a good view of the
full separation of some analytes. Second, if one is only inter- content given off by the sample’s matrix, since it is taken
ested in the presence of a given or a set of given analytes, these into consideration in the interpretation of the results. How-
may not stand out within the overall pattern of the mixture. ever, it is recommended in some instances, after performing
Third, the sensitivity of the full scan range is not as good as a full scan range, to carry out a second analysis, under a SIM
the one that would result from the scan of just a few ions. The mode, to improve the sensitivity of the GC–MS and to
more m/z are scanned, the worse the sensitivity. Thus, some provide further data.
trace analytes may simply not be detected or distinguished When using a full scan, it is still possible to benefit from the
from the background signal. mass spectral data. Starting from the TIC, one can selectively
As an alternative, it is possible to use a selected ion moni- display some individual m/z, leading to an extracted ion chro-
toring (SIM) rather than a full scan. In this situation, the user matogram (EIC), as shown in Figure 3, or a sum of several m/z,
decides which ions must be monitored. The MS will then scan leading to an extracted ion profile (EIP), as shown in Figure 4.
only for these m/z and ignore all of the others. This allows for a When comparing Figures 3 and 4 to Figure 2, one quickly
quasi-complete suppression of the matrix interferences: only realizes the advantage of extracted ions in interpreting chro-
the analytes sought after are detected. It also significantly im- matograms and identifying classes of compounds.
proves the sensitivity since each scan covers only a couple of This provides the advantage of first having a complete view
dozens m/z rather than a full range of 300 or more m/z, for of the sample through the full scan and second, looking at
example. One major disadvantage in this configuration is specific compound-related patterns through the EIC or EIP.
that the analyst completely loses the overall view of the sample. This allows the data to be filtered through the partial removal
Because only a partial image of what is in the sample is of matrix interferences or other analytes and for the pertinent
obtained, it is not possible to take into account the overall data to be revealed. The great advantage of EIC and EIP is that it
composition of the sample during the interpretation of the does not filter at the data acquisition, thus the full data always
results. If the full scan corresponds to a wide-angle view, SIM remains available.
would correspond to a super telephoto view.
Some analyses are better carried out using SIM and some
others using a full scan. Analyzing a sample of unknowns Requirements
would naturally require a full scan, since one does not know
in advance which analytes are sought after. Conversely, per- In order to combine a gas chromatograph to a mass spectrom-
forming routine analysis for some specific known metabo- eter, a few conditions must be met.
lites in trace amounts, for which their presence and quantity First of all, the sample must be suitable for injection into
are not influenced by the content of the matrix, is best the mass spectrometer. This is naturally achieved because the
performed using a SIM configuration. In fire debris analysis, mobile gas phase exiting the GC with the analytes is perfectly
for example, even though the analytes sought after are quite suitable for entry into the MS, where it will first be ionized and
known, it is not recommended to carry out only SIM. As a then filtered, to be finally detected.
Methods | Gas Chromatography–Mass Spectrometry 599

Abundance

50 000
Alkane profile
(ions 57 + 71 + 85 + 99)

25 000

Cycloalkane profile
20 000
(ions 55 + 69 + 83)

10 000

0
2 3 4 5 6 7 8 9 10 11 12 13 14 15 Time
Carbon # 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25
Aromatic profile
(ions 91 + 105 + 119)
1 000 000

500 000

Indanes profile
300 000 (ions 117 + 118 + 131 + 132)

200 000

100 000

Polynuclear aromatic profile


500 000
(ions 128 + 142 + 156)

250 000

0
5 6 7 8 9 10 Time
Carbon # 9 10 11 12 13 14 15

Figure 3 Example of extracted ion profiles from a 75% evaporated gasoline sample. One can clearly see the patterns exhibited by the different
classes of compounds. Reproduced from Stauffer E, Dolan JA, and Newman R (2008) Fire Debris Analysis, p. 316. Burlington, MA: Elsevier
Academic Press. Copyright ã 2008, Elsevier.

Second, it is important to realize that in order to obtain a coeluting compounds, the resolution of the chromatogram
mass spectrum, a certain amount of time is required. This time must be high enough to allow for the separation of the peaks.
depends on the type of detector and the scan range (typically Thus, if two compounds elute within 4 s and the MS makes one
something like 30–400 m/z). The larger the scan range, the full scan every 4 s, the compounds will not be separated and
longer the scan takes. In order to obtain full separation of the specificity of the instrument will strongly suffer.
600 Methods | Gas Chromatography–Mass Spectrometry

Abundance C3-Alkylbenzenes
(ion 105)

500 000

C4-Alkylbenzenes
300 000 (ion 119)

100 000
0

200 000 C1-Indanes


(ion 117)

100 000

40 000 C2-Indanes
(ion 131)
20 000

200 000
C1-Naphthalenes
(ion 142)
100 000

15 000 C2-Naphthalenes
(ion 156)
10 000

5000

0
6 7 8 9 10 Time

Carbon # 9 10 11 12 13 14 15

Figure 4 Example of extracted ion chromatograms from a 75% evaporated gasoline sample. One can clearly see the patterns exhibited
by the different ions representative of organic compounds. Reproduced from Stauffer E, Dolan JA, and Newman R (2008) Fire Debris Analysis, p. 317.
Burlington, MA: Elsevier Academic Press. Copyright ã 2008, Elsevier.

In practice, all modern MS exhibit a scan rate high enough rate can only be achieved with a different MS technology: time-
(2–10 Hz) to allow for a good chromatographic resolution. of-flight mass spectrometry (TOF-MS).
However, the chromatographic resolution obtained with a
regular GC–MS is definitely lower than that obtained from a
GC-FID for example. The reason is that the FID has a scan rate Instrumentation
of approximately 200 Hz. Actual MS scan rates are not consid-
ered an issue for regular GC–MS; however, they become a There are very few variations regarding the gas chromatograph.
concern with new methods such as two-dimensional compre- All modern chromatographs use a capillary column with a
hensive gas chromatography–mass spectrometry (GCGC–MS). coated stationary phase, which offers a very high resolution.
In this case, all the components coming out of the first gas The injection techniques, split or splitless, are also available on
chromatograph are subjected to a second separation (based on all gas instruments.
different characteristics) in another chromatograph. The scan rate The main variation found in a GC–MS lies in the MS,
must be of at least 50 Hz, thus allowing for enough MS scans which offers different major alternatives. There are three
during the time required for the second separation. This scan types of mass spectrometers available on the market today:
Methods | Gas Chromatography–Mass Spectrometry 601

exploited by using EIC and EIP to interpret the complex chro-


matograms. Because of the type of organic molecules, it is
possible to extract different profiles using several specific ions
and, thus, to facilitate the interpretation of the chromatograms.
Fire debris analysis is mostly concerned with qualitative, maybe
semi-quantitative, analysis.
In drug analysis, the forensic scientist is presented with
unknown samples of powder or liquid and must determine
the nature and quantity of its components. Not all drugs can be
analyzed through GC–MS, given that some are sensitive to
the heat of the injector, for example. However, the analysis
can be carried out on most drugs. With many drugs, such as
cocaine and heroin seized from the street, GC–MS provides a
very thorough analysis. Determining that an active substance
is present and at what quantity is an important element for
Figure 5 An Agilent 6890N-5975 gas chromatograph–mass
the prosecution. However, identifying the adulterants and the
spectrometer. An Agilent 7683B autosampler is present above the GC
contaminations/impurities also present brings a whole new
on the left. Reproduced from Stauffer E, Dolan JA, and Newman R (2008)
Fire Debris Analysis, p. 246. Burlington, MA: Elsevier Academic Press. edge to the investigation. As a matter of fact, adulterants and
Copyright ã 2008, Elsevier. contaminations/impurities are used to perform drug profiling,
a method capable of establishing links between individual
seizures and identifying the drug network. GC–MS is a pri-
quadrupole, ion trap, and TOF. Each presents its advantages mordial tool in this regard, since it allows for a complete
and drawbacks and some are more suitable for some types of separation and identification of unknowns in a substance.
analysis than others. Figure 5 shows a typical GC–MS setup Because MS is not always optimal in quantitative analysis, it
in a forensic laboratory. is often used in conjunction with other detectors, such as FID,
With the computer controlling system and the autosam- to obtain both qualitative and quantitative accurate results.
pler, a modern GC–MS is a highly capable instrument that In explosive analysis, the goal is to analyze seized sub-
can guarantee a very high throughput. This provides tremen- stances or residues of possible explosives. Again, the scientist
dous relief in terms of the manpower requirement for a not necessarily being aware of the type of substance involved,
forensic laboratory. The convenience of its use is also an GC–MS’s versatility brings a great advantage to the identifica-
important factor contributing to its reputation as a gold tion of explosives and residues. It is considered as a sufficient
standard. Basically, it is possible to fully automate the sepa- technique to allow for the identification of many explosives
ration of 150 complex mixtures, including the identification without any other testing.
of their analytes. In other words, one can let the instrument Forensic toxicology is a wide field concerned with the
run by itself for several days without worrying about it and detection of drugs, metabolites, or toxins in the human body.
just concentrate on the interpretation of the results. This ranges from simple medication to alcohol consumption
and other drug-related substances. The samples are body fluids
such as blood and urine and other body samples, such as hair,
Forensic Applications tissues, or fingernails. The separation and identification power
of GC–MS presents an interesting advantage, even though not
The applications of GC–MS to the forensic field are extremely all substances can be run through GC–MS. A forensic toxicol-
diverse. Basically, all analytes that can go through a gas chro- ogy lab includes many other analytical instruments such as
matograph can be analyzed by GC–MS. So, everything that is liquid chromatography, capillary electrophoresis, and immu-
volatile under 300–500  C, does not decompose at these tem- noassay. As such, GC–MS is one important tool among others,
peratures, and exhibits a molecular weight below 600 amu (up such as liquid chromatography–mass spectrometry (LC–MS),
to 1000 with new MS) can be analyzed by GC–MS. This leaves in a forensic toxicology laboratory.
room for many different applications, which is the reason why In environmental crime investigation, one is often con-
GC–MS is one of the most versatile analytical instruments that fronted with hydrocarbon spills or contaminations. Because
one can find in a forensic laboratory. hydrocarbons are extremely complex mixtures of hundreds of
The most common applications in forensic science are the components, the separation power exhibited by GC and the
analysis of fire debris, drugs, explosives, and toxins. Other characterization power of MS are fully necessary. As a matter of
applications, less often encountered, include the analysis of fact, environmental laboratories often use specific parameters,
hydrocarbons in environmental crime investigations, vegetable in terms of separation, that are quite different from parameters
oil residues, and, when using pyrolysis-GC–MS, paints, fibers, used in traditional forensic settings, because of the complexity
and plastics. of the substances they are dealing with. The ultimate goal of an
In fire debris analysis, ignitable liquid residues extracted environmental investigation is not to simply demonstrate that
from fire debris samples are analyzed in order to detect for the pollution exists, as one rarely needs a GC–MS for that;
the presence of an ignitable liquid, possibly used as an accel- however, it is to identify the source of the pollution. In this
erant at a fire scene. These samples consist of complex mixtures scope, different techniques of oil spill fingerprinting, for exam-
of hydrocarbons, often unresolved. The power of MS is fully ple, have been developed over the years.
602 Methods | Gas Chromatography–Mass Spectrometry

The analysis of vegetable oils (or rather their derivates) by Further Reading
GC–MS has been well known for many years in the food industry.
In fire investigation, a method has been developed to recover Gohlke RS and McLafferty FW (1993) Early gas chromatography/mass spectrometry.
Journal of the American Society for Mass Spectrometry 4: 367–371.
vegetable oil residues from a fire that could have been triggered by
Levine B (ed.) (2010) Principles of Forensic Toxicology, 3rd edn. Washington, DC:
the autoignition of oil residues. This method uses GC–MS American Association for Clinical Chemistry.
to analyze the extract and to identify the type of oil involved. Silverstein RM, Webster FX, and Kiemle D (2005) Spectrometric Identification of
GC–MS is clearly the best instrument to carry out this type of Organic Compounds, 7th edn. Hoboken, NJ: John Wiley and Sons.
analysis, because of its combined separation and analytical Smith FP (ed.) (2005) Handbook of Drug Analysis. Burlington, MA: Elsevier Academic
Press.
power. Sparkman OD, Penton Z, and Kitson FG (2011) Gas Chromatography and Mass
Finally, many other samples can be analyzed by GC–MS. Spectrometry: A Practical Guide, 2nd edn. Burlington, MA: Elsevier Academic Press.
This is the case with plastics, polymers, paints, and fibers Stauffer E, Dolan JA, and Newman R (2008) Fire Debris Analysis. Burlington, MA:
whose composition can be analyzed after being pyrolyzed. Elsevier Academic Press.
Terrettaz-Zufferey AL, Ratle F, Ribaux O, Esseiva P, and Kanevski M (2007) Pattern
This provides important information regarding a common
detection in forensic case data using graph theory: Application to heroin cutting
source between a questioned and a comparison samples. In agents. Forensic Science International 167: 242–246.
this regard, pyrolysis-GC–MS was developed, allowing for Wang Z and Stout S (2006) Oil Spill Environmental Forensics: Fingerprinting and
the direct injection of pyrolysis products into the GC. This Source Identification. Burlington, MA: Elsevier Academic Press.
technique, less often encountered, has also been used for Wong RC and Tse HY (eds.) (2005) Drugs of Abuse: Body Fluid Testing. Totowa, NJ:
Humana Press.
many years in different crime laboratories around the world. Yinon J and Zitrin S (1996) Modern Methods and Applications in Analysis of
Explosives. Hoboken, NJ: John Wiley and Sons.
Zoro JA and Hadley K (1976) Organic mass spectrometry in forensic science. Journal of
See also: Chemistry/Trace/Drugs of Abuse: Analysis of the Forensic Science Society 16(2): 103–114.
Controlled Substances; Chemistry/Trace/Explosives: Explosives:
Analysis; Chemistry/Trace/Fire Investigation: Analysis of Fire
Debris; Interpretation of Fire Debris Analysis; Methods: Gas Relevant Websites
Chromatography; Liquid Chromatography–Mass Spectrometry; Mass
Spectrometry; Toxicology/Drugs of Abuse: Drugs in Hair; Validation http://www.agilent.com – Agilent
http://www.perkinelmer.com – PerkinElmer Inc.
of Twelve Chemical Spot Tests for the Detection of Drugs of Abuse. http://www.restek.com – Restek Corporation

You might also like