PM-9000Express: Patient Monitor
PM-9000Express: Patient Monitor
Technical specifications
PM-9000Express
Patient Monitor
Safety NeonatalMode: 40 - 135 mmHg accuracy within 3minutes.
IEC60601-1 approved, CE marking according to MDD93/42/EEC Range of diastolic pressure: Adult Mode: 10 - 210 mmHg Response time: Typical value: 2.9s, including the rising time and the
Pediatric Mode: 10 - 150 mmHg delay time
Dimension and Weight Neonatal Mode : 10 - 100 mmHg (adopting the FilterLine of standard length)
Dimension: 318(W) x 270 (H) x137 (D)mm Range of mean pressure: Adult Mode: 20 - 230mmHg Rising time: <190ms(rising from 10% to 90%)
12.5(W) x 10.6 (H) x 5.4 (D)inch Pediatric Mode: 20 - 165mmHg Delay time: 2.7s (typical value)
Weight: Standard Configuration : 4.7kg / 10.4lb. Neonatal Mode: 20 - 110mmHg Respiration rate 0-150 breaths/min
Maximum weight :7.5kg / 16.5lb. Accuracy of blood pressure measurement: Respiration rate accuracy: 0-70BrPM ±1BrPM
Operation Environment The mean error shall be less than ±5 mmHg 71-120BrPM ±2BrPM
Power Requirements: AC100-240V, 50/60 Hz,140VA The standard deviation shall be less than 8 mmHg 121-150BrPM ±3BrPM
Temperature: 0-40℃/ 32~104 oF Over-pressure protection: double safety protection Mode : Adult, neonate
Humidity: 15-95% non-condensing Resolution: 1mmHg
Patient Range Neonate, pediatric, and adult patients Alarm: Systolic, Diastolic, Mean Sidestream CO 2
PR from NIBP: Measurement 40~240 bpm Measurement range: 0-99mmHg
Performance Specifications Resolution 1 bpm Accuracy: ±2mmHg (0-40 mmHg)
Display: 12.1”color TFT(diagonal) Accuracy ±3bpm or ±3% whichever is greater ±5% of reading (41-76 mmHg)
Rolling and refreshing waveform display ±10% of reading (77-99 mmHg)
Resolution: 800x600 SpO 2 Sampling rate: 70 ml/min,100ml/min
Multi displays selectable, including: MINDRAY SpO 2 Sampling rate accuracy: 0.15 or 15ml/min
Standard display Measurement range: 0~100% Start-up time: 30 Sec. The module enters the warming up status after
Large-font Display Resolution: 1% the startup.
Trend coexist display Accuracy: ±2% (70-100%,Adult/Pediatric,non-motion); 1 minutes later, it enters the ready-to-measure status.
Bed-to-bed view display ±3% (70-100%,Neonate,non-motion); Mode: 10 minutes after start-up, the module reaches full
OxyCRG dynamic view display ±3% (70-100%,Adult/Pediatric/Neonate,motion); accuracy mode
Trace: 8 waveforms 0-69% unspecified Respiration rate: 0-120 rpm
Sweep Speed: 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s Alarm range: 1~100% Respiration rate accuracy: ±2 BrPM (0-70 BrPM )
Indicator: Alarm indicator light Pulse rate: Range: 20~254bpm ±5 BrPM (>70BrPM )
Power indicator light Resolution: 1bpm Response time: When measured with a neonatal watertrap
Battery indicator light Accuracy: ±3bpm(non-motion) and a 2.5 m-long
QRS beep and alarm sound neonatal sampling line:
±5bpm(motion) <3.5 s @ 100 ml/min
Interface: Parameter cable interface
Alarm range: 20~254bpm <4 s @ 70 ml/min
AC Power input socket When measured with an adult watertrap
Network interface NELLCOR SpO 2 and a 2.5 m-long adult
External VGA interface for connection of an sampling line:
Measurement range: 0~100%
Alternate Display <5.5 s @ 100 ml/min
Resolution: 1% <7 s @ 70 ml/min
Aux Output interface (Analog/ Defib-Synch/ Nurse Call) Accuracy: ±2% (70-100%, MAX-A,MAX-AL,MAX-N,MAX-P,MAX-I Delay time: When measured with a neonatal watertrap
Battery: Rechargeable Li-ion Battery : and MAX-FAST sensors); and a 2.5m-long
(for 2 pieces) Maximum 6.5 hours for charging; ±2.5% (70-100%, OxiCliq A,OxiCliq N,OxiCliq P and neonatal sampling line:
5 hours for continuous working OxiCliq I sensors); <3 s @ 100 ml/min
Rechargeable Lead-Acid Battery: ±3% (70-100%, D-YS, DS-100A, OXI-A/N and OXI-P/I <3.5 s @ 70 ml/min
(for 2 pieces) Maximum 12 hours for charging; When measured with an adult watertrap
sensors); and a 2.5m-long adult
2 hours for continuous working 3.5% (70-100%, MAX-R, D-YSE and D-YSPD sensors); sampling line:
Trend time: 1~96 hours 0-69% unspecified <5 s @ 100 ml/min
Alarm: User-adjustable High and Low limits 3-level audible and Alarm range: 0~100% <6.5 s @ 70 ml/min
visual alarm Pulse rate: Range: 20~300bpm
Networking: Connected to central monitoring system Resolution: 1bpm Mainstream CO 2
Recorder: Built-in, thermal array Accuracy: ±3bpm(20-250bpm); Method: Infrared Absorption
Waveform: 3 channels 251-300bpm unspecified Measuring Mode: Mainstream
Record mode: manual, on alarm, time-defined Unspecified: Alarm range: 20~300bpm Measurement range: EtCO 2 : 0-99mmHg
Paper width: 50mm InsCO 2 : 0-99mmHg
Print speed: 25mm/s, 50mm/s Temperature AwRR: 0-150BrPM
Resolution: EtCO 2 : 1mmHg
Measurement range: 0~50℃(32~122℉)
ECG InsCO 2 : 1mmHg
Resolution: 0.1℃
Lead type: 5-lead and 3-lead selectable, AwRR: 1BrPM
12-lead(including 3/5-lead)optional Accuracy: ±0.1℃ (excluding the sensor) Accuracy: CO 2 concentration: ±2mmHg (0-40mmHg)
Input: 10 leadwire cable: RA; LA; RL; LL; V1-V6 or R; L; N; F; C1-C6 Channel: Dual-channel. Provide T1; T2; ΔT ±5% of reading (41-76mmHg)
5 leadwire cable: RA; LA; RL; LL; V or R; L; N; F; C ±10% of reading (77-99mmHg)
3 leadwire cable: RA; LA; LL or R; L; F
AwRR: ±2BrPM
Lead selection: 12-lead: I; II; III; aVR; aVL; aVF; V1-V6 IBP
Alarm range: Same as Measurement range
5-lead: I; II; III; aVR; aVL; aVF; V Measurement range: -50~300mmHg
3-lead: I; II; III Channel: 2 channels
Multi-Gas/O 2
ECG waveform: 2 channels Pressure transducer: Sensitivity: 5μV/V/mmHg
Method: Infrared Absorption
Gain selection: x0.125;x0.25; x0.5; x1; x2; auto Impedance range: 300-3000Ω
Gas sorts: CO 2 , N 2 O, Des, Iso, Enf, Sev, Hal, O 2
Sweep speed: 12.5mm/s, 25mm/s, 50mm/s Pressure names: ART, PA, CVP, RAP, LAP, ICP,P1,P2
(Optional paramagnetic sensor)
Heart Rate range: Adult: 15-300bpm; neonate/ Pediatric: 15-350bpm Resolution: 1mmHg
Measurement range: CO 2 : 0~30%
Accuracy: ±1bpm or ±1%, whichever is greater Accuracy: ±2% or±1 mmHg, whichever is greater
N 2 O: 0~105%
Resolution: ±1bpm Alarm range: -10~300mmHg
O 2 : 0~105%
Filter: Diagnostic mode: 0.05-100Hz or 0.05-150Hz(optional PR from IBP: Range & Accuracy:
Enf, Iso, Hal: 0~30%
12-lead) 25 ~200 bpm ±1 BPM or ±1% whichever is greater Sev: 0~30%
Monitoring mode: 0.5-40 Hz 201 ~350 bpm ± 2% Des: 0~30%
Surgical mode: 1-20Hz Resolution: 1 bpm Data output: Fi and ET values
Protection: Withstand 4000VAC/50Hz voltage in isolation Respiration rate: ±1BrPM(2~60 BrPM)
Against electrosurgical interference and defibrillation Cardiac Output
61~100 BrPM (unspecified)
Scaling signal: 1mV±5% Other: Up to 4 waveforms displayed
Alarm range: 15-350 bpm Method: Thermodilution
Agent mixture detection
S-T segment detection: Measurement range: -2.0 mV~±2.0mV Measurement range: CO: 0.1-20 l./min.
MAC value displayed
Alarm range: -2.0mV~±2.0 mV TB: 23-43℃
Arrhythmia analysis: YES TI: 0-27℃ Standard Configuration:
Alarm: YES, audible and visual alarm, alarm events recallable Resolution: CO: 0.1 l/min. ECG(Mindray 3/5 Lead);NIBP;RESP;Mindray-SpO 2 ;Dual-Temp;Lead Acid Battery
12 lead ECG analysis: YES TI: 0.1℃
Pace detection: YES TB: 0.1℃ Options:
ECG Configuration: Mortara ECG(M08A)-- 3/5/12 Lead; (Options Mortara S-T Accuracy: CO:± 5% or o.1l/min ECG(Mortara 3/5 lead with Extended ARR);
segment detection & Mortara Arrhythmia analysis) TB: ±0.1℃ ECG(Mortara 12 lead with Extended ARR);
TI: ±0.1℃ ECG(Mortara 3/5 lead with Extended ARR & 3/5 Lead ST );
Respiration ECG(Mortara 12lead with Extended ARR &12 lead ST );
Parameter output: Cardiac output
Method: Thoracic impedance Clinical proven SpO 2 ;Dual-IBP;CO;EtCO 2 ;Multi-gas/O 2 ;Thermal Recorder; Li-ion Battery,
Hemodynamics calculation
Operation modes: Auto/ Manual CF Card(power off storage)
Measurement range: Adult: 0~120 BrPM; Neonate/Pediatric: 0~150 BrPM
Apnea alarm: YES EtCO 2
Alarm: YES, audible and visual alarm, alarm events recallable Microstream CO 2
Measurement range: 0-99mmHg
NIBP Accuracy: 0-38 mmHg ±2 mmHg
Method: Automatic oscillometric 39-99 mmHg ±5% of reading + 0.08% for every 1 mmHg
Operation modes: Manual/Automatic/Continuous (above 38 mmHg)
Auto measure time: Adjustable Resolution: Waveform: 0.1mmHg
Measurement unit: mmHg/kPa selectable Value: 1 mmHg
Measurement types: Systolic, Diastolic, Mean Sampling rate: 50ml/min -7.5+15ml/min
Range of systolic pressure: Adult Mode: 40 - 270 mmHg Initialization time: 30 seconds (typical), reaches ±5% steady-state
Pediatric Mode: 40 - 200 mmHg
Microstream TM EtCO 2
Suitable for non-intubated and intubated
applications
For adult,pediatric,infant/neonate patients
Low sampling rate 50ml/min (± 7.5ml/min)
Rolling Stand
Large Font Display Mode
Fixed Memory Card Basic parameters (ECG, SpO 2, NIBP) in
Provides power off storage for 96-hour large font for distance view
graphic and trend, 800 NIBP records and 70
alarm records